Professional Documents
Culture Documents
(BM-055L) Service Manual (ENG - V1.0)
(BM-055L) Service Manual (ENG - V1.0)
(BM-055L) Service Manual (ENG - V1.0)
BM-055L
Table of Contents
Table of Contents
Ⅰ. Installation of product
1. Installation Environment ------------------ 5 3) Explanation about Error
2. Installation of product ---------------------- 6 (1) Door Error ------------------------------ 23
- Enter into expert mode ----------------------- 9 (2) Cancel Error --------------------------- 23
3. Check of installation (3) Injection Heater Error ---------------- 24
1) Check output signals of heater ---------- 10 (4) Chamber Heater Error -------------- 24
2) Confirm operational status of each part (5) Charge H2O2 -------------------------- 25
--------------------------------------------------- 10 (6) H2O2 Supply Error ------------------ 25
3) Check of oil level ---------------------------- 11 (7) Wet Over Error ------------------------ 26
4) Charge of sterilant -------------------------- 14 4) Verification of sterilizing power
5) Self Test --------------------------------------- 15 (1) Chemical method -------------------- 27
4. Education for installation (2) Biological method ------------------- 28
1) Preparation for use ------------------------- 17 5) Cautions ------------------------------------- 29
2) How to use ------------------------------------ 18
Ⅲ. Manufacturer Contact
- Address and Contact ------------------- 110
Installation Environment
1. Install in indoor location.
2. Install on flat floor.
3. There should be no fixed objects within 1 meter from the product.
4. Power
1) Use ground connected power.
2) Use power within ±10% of rated power.
5. Temperature & humidity environment
1) 5℃~30℃: 80% ↓
2) 31℃~40℃: 50% ↓
Installation of product
1. Remove side & back cover
2. Check whether screw is loose or not through tightening the clamp of each part by hand.
Vacuum valve
Vacuum gauge
Temperature sensor
DC power switch
Start-up screen
2. Caution
1) Only the people who can install the product & do technical service are allowed to enter into export mode and
users are not allowed to enter into expert mode except special cases.
turning on injection valve is prohibited in the vacuum state of the inside of chamber (H202 flows
into inside of chamber excessively).
Check of installation
1. Check output signals of heater
1) Confirm if a, b, c, d of heater output signals on the picture occurs.
① a, d (Injection heater)
② b, c (Chamber heater)
d
a
c
2. Confirm operational status of each part
b
1) Enter into expert mode
2) Confirm ON operational status of each part and then, OFF
① Vacuum pump ON->Confirm operating sound of vacuum pump->vacuum pump OFF
② Vacuum valve ON->Confirm operating sound of vacuum valve->vacuum valve OFF
③ Vent valve ON->Confirm operating sound of vent valve->vent valve OFF
④ Door open->Injection valve ON->Confirm operating sound of injection valve->
Injection valve OFF->Door close ① ②
⑤ Plasma ON->Visual confirmation of plasma occurrence->Plasma OFF ④ ③
⑥ FAN ON->Confirm FAN ventilation->FAN OFF ⑤ ⑥
Operating injection valve is prohibited in the vacuum state of the inside of chamber.
2. Way off
- Touch to the sterilizer text(a) of main screen
- Touch to the Supp Manager text(b) of Expert Mode
b - Touch to the Pump Oil, Vacc Fiter, CabonFiter Text
- Touch to the Exchange
a - Twice touch to the Back
6. Self Test
1) Touch self-diagnosis
2) Confirm message of self-diagnosis completion
3) Confirm output of printing paper.
Printouts
How to use
1. Remove the moisture of object for sterilizing after washing completely. 1*
2. Pack the object for sterilizing. 2*
3. Open the door and put the packed object for sterilizing evenly in the inside of chamber 3*
4. Close the door and touch standard sterilization or surface sterilization on LCD Indication window by hand
Sterilizing process
Start-up screen Completion screen Start-up screen
proceeding
1*. Product life of pump in equipment is to be shorten if moisture is not removed completely.
2*. Use exclusive pouch & sterilized cotton for plasma sterilizer, set the temperature of pouch sealing
appliance at 125℃
3*. Pay attention that the amount of object for sterilizing does not exceed about 70% space of the inside of
chamber.
※ What is difference between standard sterilization & Advanced & surface sterilization sterilization?
▷ Standard sterilization : It’s a process for sterilizing up to the inside of object for sterilizing
(Need about 36 minutes)
▷ Warm Up sterilization : It’s a process for sterilizing Much of cold metal of object for sterilizing
(Need about 54 minutes)
▷ Advanced sterilization : Advanced than the standard sterilization process Sterile improved
process (Need about 50 minutes)
(Complex surgical instruments for sterilization)
▷ Surface sterilization: It’s a process for sterilizing the surface of object for sterilizing
(Need about 24 minutes)
(It is used for simple appliances and urgent sterilization owing to its short time of sterilization).
Gouge part Scope Line part Dressing Forceps part Lever part
Plier part Light Cable part Suction Tube part Currette part
DRILL part Burr part(Line, Tip) Clamp part Hook / Loops part
Handle part
If you do not push cancel button and power off, you may have burn skin due to residual sterilant.
After cancellation, you should complete through the finishing process of sterilization without fail.
Pay attention that do not touch during sterilization process except above reasons.
1. Door Error
▷ Cause: If implement the sterilization in the state of door open
2. Cancel Error
▷ Cause: If push ‘Cancel’ button by user’s mistake.
▷ Action: If it’s within 2 hours from main power supply of product, wait until warming
up and contact the supplier if it is occurred during the use in the state of main power
supplied.
▷ Action: If it’s within 2 hours from main power supply of product, wait until warming
up and contact the supplier if it is occurred during the use in the state of main
power supplied.
5. Change H2O2
▷ Cause: If any shortage of sterilant in the inside of product
▷ Action:
- Empty the inside of chamber and implement self-
diagnosis. If it is completed within normal time (5
minutes), confirm the moisture of inserted object
for sterilizing, dry, remove the materials which can
absorb moisture and then, sterilize again.
- If wet over Error in self-diagnosis occurred,
contact the supplier.
Self Test
⑴ Chemical method
Chemical Indicator (CI) :
- It’s equipment for confirming the success or failure of sterilization after sterilization and it’s confirmed by
chemical method.
Judge whether the success or failure of sterilization by confirming if the color of CI changed or not.
The color is changed as hydrogen peroxide applied on CI with certain amount and judge the changed CI
means the success of sterilization
Before sterilization
After sterilization
(Success of sterilization)
Cautions
① Never include the things which can absorb sterilant such as sponge, paper, wood, fabric etc. in the objects
for sterilizing.
- It absorbs sterilant and the sterilant may insert into the inside of chamber excessively
- The absorbed sterilant remained even after completion of sterilizing process and may cause smell &
reaction on hand.
② The object for sterilizing should be washed & dried completely without fail before sterilization and then,
implement the sterilization.
- If moisture remained, wet over error may occur.
- If moisture injected into vacuum pump excessively, the lives of pump oil & filter is to be shorten.
③ Fill the object for sterilizing in the inside of chamber with the amount of below 70% space and then,
implement the sterilization.
- If Fill the object for sterilizing too much for sterilization, sterilizing power may decline or sterilant may insert
excessively.
- It’s better to have good space for inserting hand in the chamber after the object for sterilizing inserted into
chamber.
Cautions
④ Use sterilized pouch & non-woven fabric only for low-temperature plasma sterilizer.
- Pouch
▷ Different sealing temperature of sealing appliance (EO-GAS: 170℃, low-temperature plasma: 125℃)
▷ If use EO-GAS & steam sterilizer pouch, particle size is different and effect the penetration of sterilant.
As a result, the sterilizing power may decline or is not working
- Sterilized fabric
▷ For low-temperature plasma, it is commonly accepted material of polypropylene.
▷ Never use the sterilized fabric which coated on one side.
⑤ Use BI & CI indicator only for low-temperature plasma sterilizer without fail.
- If it’s not only for low-temperature plasma, accurate sterilizing power cannot be guaranteed.
⑥ BI should be cultured for 48 hours without fail after sterilization
- Confirm the result of cultivation
▷ Success of sterilization: Maintain the existing purple color
▷ Failure of sterilization: Confirm the virus cultured to yellow color
⑩ If sterilant contacts with skin, wash it with running water by using soap immediately.
-If sterilant contacts with hand, it reacts with white color after few seconds and may form blister on it by the
extent of it.
-If wash it with running water & soap before reacting with white color, you can prevent the reaction.
-Even reacted with white color, it can be itself again as time pass by (Excessive blister occurrence may leave
scar on skin).
-First aid
A. If enter into eye If it enters into eye, wash it with a plenty of water for 15 minutes immediately.
Get doctor’s examination & treatment at once.
B. If contact with skin Take off contaminated clothes & shoes and remove it.
Wash it with a plenty of soapy water and remove chemical substance.
Clean contaminated cloths & shoes before re-use of it.
Get doctor’s examination & treatment at once
⑪ When sterilize a lot of the object for sterilizing which has cold metal component, wet over error may occur.
- If you intend to sterilize the object for sterilizing which has a lot of metal component or big volume object for
sterilizing which has metal component, it is recommended that you put it in chamber for about 5~10 minutes
and after warning up, implement the sterilization.
-If sterilize the object of cold metal component in big Q’ty, the sterilant may be condensed in metal component
and inserted excessively. After completion of sterilizing process, the sterilant may remain.
-When sterilize cold object for sterilizing in big Q’ty, not alone metal component, sterilize after warming up for
about 5~10 minutes and it helps to strengthen sterilizing power, can decrease the cause for Error occurrence.
⑫ Keep the sterilant refrigerated endways
-Special cap for gas emission inserted in the upper part of sterilant bottle and if keep it upside down, no gas
emission and the sterilant may flow out.
a
Choose Relevant part
c
Touch Sterilizer once Touch Consumables Touch Replace confirm once.
management once.
Insert silicone ring Insert silicone ring complete inserting of silicone ring Remove foreign substance
-After confirming the temperature of injection chamber declined, separate the valve by using spanner.
Separate nut Separate nut Separate mail connector Separate mail connector
Be careful about the burn skin by the residual of hydrogen peroxide in the tube
(Work with wearing protection gloves).
-Assemble the valve by using spanner with a care about the direction of valve for replacement.
-Fix 2 bolts (a) at the bottom side of valve by flat screwdriver.
-Peel off the sheath of power supply line with about 5mm length, insert sleeve, compress by compressor and
connect the line with the valve
(confirm if compression is done without problem through pulling out the line by hand)
Tighten nut Fix bolt Tighten nut Connect power supply line
a
Separate power supply line Separate bolt Separate heater Fix bolt after inserting heater Connect power supply line
a c
b
Must! Work after DC power OFF (If separate or insert the connector in the state of power ON, faulty
vacuum gauge may occur).
c
a
c d g
a
b
e f
Remove oil cap Charge oil Upper limit check Insert oil cap
b b
e
a b
f
g
c d
a b
Remove bolt (Down) Remove bolt (up) Remove clamp (Down) Remove clamp (up)
Insert center ring Fix clamp (Down) Insert center ring Fix clamp (up)
Remove the clamp b Remove the filter Insert the replacement filter
a
Fixed filter fixed Vent valve
Door Error
Slide59
Corrective measure for error – Door error
1. Door Error
Necessary item: Door switch, Cable, Multi meter, Flat screwdriver, Hot-melt gun.
1) If implement the sterilization in the state of door opening
-Action: Close the door and sterilize again
2) Poor quality of door switch
-Confirm: Set up the multi tester to make buzz sound -> Let the measuring bar of multi tester contact with
switch terminal block (a, b) -> Repeat door switch (c) ON & OFF -> Judge door switch quality by multi tester
-Action: Separate the left cover->Remove hot melt (d) at the inside of door switch product ->Separate door
switch to the front side of product -> Separate the connected terminal block (a, b) ->Insert again the
terminal block (a, b) into door switch for replacement ->Insert the switch at the front side of product ->
Fix by hot melt
d e
c a
f
b
Cancel
Slide62
Corrective measure for error- Cancel Error
2. Cancel Error
1) If user push ‘Cancel’
-Action: After settlement of user’s cancel purpose, implement the sterilization again.
2) If the part of ‘Cancel’ touched by careless handling
-Action: Be careful about touching on the part of ‘Cancel’ and implement the sterilization again.
Be careful that no touch ‘cancels’ on the way of sterilizing process except above reasons.
Injection
Heater Error
a b
Push for 3 seconds Push once Push 4 times Push 5 times Push 1 times Confirm the set-up value
b
a
221IH
222IH
a b
-Action: Main power OFF-> Separate all the lines (a) connected at SSR board -> separate 4 bolts (b, c, d, e)
for SSR board fixing -> Separate SSR board -> Locate SSR board for replacement at the separated place ->
Insert 4 bolts (b, c, d, e) for replacement -> Connect all separated lines (a) at SSR terminal block as per the
drawing.
Before main power ON, reconfirm all the lines connected at terminal block
(Refer to P.101 of the drawing)
a d
c e
b
c
-Action: Main power OFF-> Separate the covers of left & right side-> separate all the lines (d) connected at
terminal block of temperature controller in the middle by Phillips screwdriver ->Separate 2 bolts (e, f) for
temperature controller fixing -> Separate the controller toward front side of temperature controller
(Left side of product) -> Insert temperature controller for replacement -> Fix 2 bolts (e, f) for fixing ->
Connect all separated lines (d) as per the drawing.
-Set-up the temperature controller after completing the replacement of temperature controller
(refer to next page)
e f
Before main power ON, reconfirm all the lines connected at terminal block
(Refer to P.101 of the drawing)
Push 3 seconds
Push 5 times Push 2 times Push once Push 3times Push once Push once
SET
SET
SET
relevant mode.
- Release to the start-up state by pushing for long
after mode enter
- Store the set-up value
SET
2) ◁ : Change input position
3) ▽ : Page forward (down) & Input down
4) △ Page backward (up) & Input up
Actual temperature
Temperature sensor
Chamber
Heater Error
Slide75
Corrective measure for error – Chamber Heater Error
4. Chamber Heater Error
1) Ill-posed problem of temperature controller #1 (Fiducial set-up value)
- Confirm: If temperature does not increase even after more than 2 hour’s pass from the time of main power
supply to the product, confirm it the set-up value of temperature controller (a) is 55
- Action: If temperature set-up is not 55, push the cursor (b) on the below end of temperature controller front
side, set-up as 55 and confirm temperature increase for about 2 hours.
a b
Push for 3 Push once Push 4 Push 5 Push once Confirm the set-up value.
seconds times times
a b
- Action: Cut off main power supply and let the covers of side & back opening
Separate the power line of heater cable from terminal block by using flat screwdriver
(No.227, No.234) Pull out the line from duct and cut the cable tie of which power line fixed.
a b
-Action: Main power OFF-> Separate all the lines (a) connected at SSR board -> Separate 4 bolts (b, c, d, e)
for SSR board fixing -> Separate SSR board -> Locate SSR board for replacement at the separated
place -> Insert 4 bolts (b, c, d, e) for replacement -> Connect all separated lines (a) at SSR terminal
block as per the drawing.
Before main power ON, reconfirm all the lines
a d
connected at terminal block
b (Refer to P.101of the drawing)
c e
-Action: Main power OFF-> Separate the covers of left & right side-> Separate all the lines (d) connected at
terminal block of temperature controller in the middle by Phillips screwdriver -> Separate 2 bolts (e, f)
for temperature controller fixing -> Separate the controller toward front side of
temperature controller (Left side of product) -> Insert temperature controller for
replacement -> Fix 2 bolts (e, f) for fixing -> Connect all separated lines (d) as per the
drawing.
Before main power ON, reconfirm all the lines connected at terminal block
f
(Refer to P.101 of the drawing)
e
-Set-up the temperature controller after completing the replacement of temperature
d controller (refer to next page)
Push 3 seconds
Push 5 times Push 2 times Push once Push 3 times Push once Push once
SET
SET
mode enter
- Store the set-up value
2) ◁ : Change input position
SET
3) ▽ : Page forward (down) & Input down
4) △ Page backward (up) & Input up
Actual temperature
▷Action: Main Power OFF-> Separate the covers of left & right side -> Separate the connected lines at No.
18, 19, 20 of temperature controller terminal block in the middle from right side by Phillips
screwdriver -> Remove the cable tie which fix temperature sensor -> Wait until injection
chamber temperature cooling down-> Separate temperature sensor (a) by using spanner ->
Insert temperature sensor for replacement -> Fix temperature sensor by using spanner ->
Wiring temperature sensor line to temperature controller with fixing by cable tie -> Connect the
red line of sensor at No.18 on the back side of temperature controller -> Connect the blue line
at No.19 & 20 respectively (The blue lines can be reversed each other) -> Main power ON ->
Confirm actual temperature value on temperature controller
Temperature Sensor line
Change H2O2
Slide83
Measure for error - Charge H2O2
5. Change H2O2
Necessary item: Sterilant, multi meter, Sensor cable, Universal spanner, Flat screwdriver
1) If any shortage of sterilant in the inside of product
-Action: Supplement sterilant and implement sterilization.
2) Disconnection of level sensor line
- Confirm: Separate the cover of left side-> Confirm if LED (a) lighting of main board level sensor bloc is on or
not (If sterilant charged and sensor line connected, LED is lighting on) -> If LED is lighting off, confirm whether
it is disconnected or not by applying multi tester at level senor line at the upper part of
sterilant storage tank and No.1 (b) & No.2 (c) of main board level sensor bloc
b
a
d
c
- Action: Remove hot melt & nut for fixing sensor terminal bloc at the upper part (d) of
sterilant shortage tank -> Separate the sensor line which is connected at No.1 & No.2 of main board level
sensor bloc -> Regardless of polarity, connect the sensor line for replacement at No.1 & No.2 the sensor line
for replacement at No.1 & No.2 of main board level sensor bloc -> Tighter terminal block of level sensor at
the upper part of sterilant storage tank -> Fix by hot melt.
H2O2 Supply
Error
Slide86
Measure for error- H2O2 Supply Error
6. H2O2 Supply Error
Necessary item: Injection valve, Universal spanner, Nipper, Phillips screwdriver, Cable tie, Sleeve & Compressor
1) If any shortage of sterilant during the implementation of sterilizing process
-Action: Sterilize again after supplement of sterilant
2) Defect of injection valve
-Confirm: Enter into expert mode ->ON/OFF of injection valve for 2~3 times (Confirm operating sound)->If no
operating sound heard or the same error occurs at the time of re-sterilization after the supplement of sterilant
& operating sound heard, judge as defect of injection valve
-Action: Main power OFF -> Separate the cover of left side -> Wait until the
temperature of injection valve cooling down (about 1 hour) -> Cut
out the power line of injection valve with about 20cm distance
from valve -> Separate valve by using spanner.
Cut the power line
Separate nut Separate nut Separate mail connector Separate mail connector
-Assemble the valve by using spanner with caring about direction of valve for replacement
-Fix 2 bolts at the bottom of valve by phillips screwdriver
-Peel off about 5mm of power supply line, insert the sleeve, compress by compressor and connect the valve
line (Confirm if compression is done without problem through pulling out the line by hand)
Tighten nut Fix bolt Tighten nut Connect power supply line
a
Slide89
Measure for error- Wet Over Error
7. Wet Over Error
1) If the inserted object for sterilizing contains a plenty of moisture
-Action: Remove the moisture of object for sterilizing and sterilize again
2) If the material which can absorb moisture inserted (Paper, Fabric, Wood, Sponge, Powder etc.)
-Action: Remove the materials which can absorb moisture and sterilize again
※ Confirm 1, 2 in common: Empty the inside of chamber, implement self-diagnosis and if it is completed
within normal time (5 minutes), judge it is inserting problem
◈ Check procedure if Wet over Error occurs after the implementation of self-diagnosis
Necessary item: Stopwatch
-Close the door after confirming empty state of chamber inside without foreign substance -> Prepare
stopwatch -> Enter into expert mode -> Vacuum pump ON -> Measure the time by stopwatch at the same
time of vacuum valve ON at the upper part of left side electronic display and measure the time to reach
0.5 Torr-> Vacuum valve OFF -> Vacuum pump OFF-> Vent valve ON-> Confirm the door open ->
Vent valve OFF
1-1) If rising over than 0.2 Torr for 10 seconds, check the below
① Open the door and if door gasket floated or foreign substance contained, remove it.
② Open the door and confirm if moisture or residual of sterilant exist in the inside of chamber
-If moisture or residual of sterilant exist, let the vacuum to 0.5 Torr for 4~5 times and remove.
Vacuum vale
Vacuum gauge
Temperature sensor
1-2) If rising below 0.2 Torr for 10 seconds, check the below
① Enter into expert mode-> Vacuum pump ON -> Vacuum valve ON-> Confirm 0.5 Torr at pressing meter ->
Vacuum pump OFF-> Confirm pressure rising numerical value (Vacuum valve should be opening)
② If rising over than 0.2 Torr for 10 seconds, check the assembling state of plasma chamber
-Tighten NW25 flange (a) at both ends of plasma chamber as much as possible
③ Confirm clamp tightening state of each part
Plasma chamber clamp
Vacuum
pump clamp
b a c
Sure ! DC power OFF and will work
(Power connector from the state when you remove or insert a vacuum gauge defects)
c d g
a
b
e f
-Securely place the vacuum pump in product in line with vibration dampener bolts-> Fix pump by tightening
4 vibration dampener fixing bolts (a, b, c, d) -> Assembler power supply line at terminal block (e) ->
Fix plasma chamber by clamp (f) with inserting center ring into the bottom of vacuum valve ->
Fix the bellows at inhale part with inserting center ring.
e
a b
f
g
c d
b a c
Sure ! DC power OFF and will work
(Power connector from the state when you remove or insert a vacuum gauge defects)
- Method
③ If USB memory inserted, Log Data back-up automatically.
The matter that require attention at this moment is you do not back up without separate request of
Head office (manufacturer) since the data does not remain in the product and store in USB memory only in
the state of back-up.
④ Remove USB memory
- Start-up screen->Insert USB memory->LOG DATA back-up
Sterilize
PACKING OPEN CHECK