Professional Documents
Culture Documents
Unit 4
Unit 4
4.0 Introduction
4.1 Objectives
4.2.5 Instrumentation
4.3.3 Instrumentation
4.4.2 Instrumentation
4.0 INTRODUCTION
You would recall that spectroscopy is the study of interaction of electromagnetic
(EM) radiation with matter. The nature of the spectrum depends on the type of
electromagnetic radiation and the nature of matter e.g., atoms or molecules with
which it interacts. In this unit we are going to take up molecular spectroscopy
i.e., the study of interaction of EM radiation with molecular systems. We would
begin the unit with UV-VIS spectroscopy i.e., the study of interaction of UV-
VIS radiation with different species. Herein we would discuss the origin of the
spectrum in terms of the interaction of radiation with different types of absorbing
species. This will be followed by a discussion on the characteristics of the
spectrum and the instrumentation used for obtaining the spectrum. Then we
would take up the principle of UV-VIS spectrophotometry as an analytical tool.
In the third part of the unit, we would take up vibration spectroscopy i.e.,
the study of interaction of EM radiation in the IR region with the molecules.
Herein, we would take up two types of interaction i.e., the absorption and
scattering of radiation. These give rise to IR and Raman spectra respectively.
We would explain the origin and characteristics of the IR and Raman spectra
and the instrumentation required for them. Having discussed these
molecular spectroscopic methods, we would outline their environmental
applications.
4.1 OBJECTIVES
After studying this unit you will be able to:
For now, it is important to note that the molecules are much more complex
and their electronic, vibrational and rotational energies are quantised. A
given electronic energy level has severalvibrational energy levels in it and
each of the vibrational energy level in turn has many rotational energy
76
levels in it. A schematic energy level diagram for a molecule showing Molecular Spectroscopy
different quantised energy levels is given in Fig. 1.1.
Fig. 1.1: Schematic energy level diagram showing quantised electronic, vibrational,
and rotational energy levels.
When a photon of a given wavelength interacts with the molecule it may cause
a transition amongst the electronic energy levels if its energy matches with the
difference in the energies of these levels.In order to obtain a UV-VIS spectrum
the sample being studied is irradiated with the electromagnetic radiation varied
over a range of wavelength. A monochromatic radiation i.e., a radiation of a
single wavelength is employed at a time. This process is called scanning. The
amount of the radiation absorbed at each wavelength is measured and plotted
against the wavelength to obtain the spectrum. Thus, a typical UV spectrum is
a plot of wavelength or frequency versus the intensity of absorption.
The Omax refers to the wavelength of the most absorbed radiation and is a
measure of the difference in the electronic energy levels involved in the
transition. The intensity on the other hand is indicative of the probability of
the transition i.e., whether the transition is allowed or not. It is also is a measure
of the concentration of the absorbing species. The relationship between the
intensity of absorption and the concentration is explained later.
The third characteristic of the spectral band is its width and is generally
expressed as full width at half maximum (FWHM) i.e., the width of the spectral
band at half its height (i.e., the intensity). Thewidth of the spectral band depends
on type of the instrument used and the intrinsic nature of the species being
analysed. The instrumental component can be minimised but not eliminated
whereas the natural line width due to the nature of the system cannot be
removed.
The first two characteristics of UV-VIS spectrum i.e., the position and the
intensity depend on the structure and concentration of the absorbing species
in solution. Therefore, these spectra are extensively used in the characterisation
and in the quantitative estimations of the analyte.
Po = Pr + Pa + Pt … (1.1)
78
The effect of reflection can be compensated by passing equal intensities of Molecular Spectroscopy
beams through the solution and through the solvent contained in the same or
similar container and comparing the transmitted radiations. We can then write
the above equation as follows:
Po = Pa + Pt. … (1.2)
Lambert’s Law
Lambert (1760) and Bouguer independently studied the decrease in the intensity
of radiation when it passes through a substance and made the following
observations:
ii) The intensity of the transmitted light decreases exponentially when the
thickness of the substance, through which the light is passing, increases
linearly.
Let Po represent the radiant power of incident light and P represent the radiant
power of transmitted light after passing through a cuvette of thickness b. Now,
if consider a small slab of cuvette having thickness dx, then the change in
power (dPx), is proportional to the power of incident light (Px) multiplied by
thethickness (dx) of the slab through which the light is passed. That is,
where k is the proportionality constant, and the negative sign indicates that
radiant power decreases with absorption. Eq. 1.3 can be rearranged to Eq. 1.4.
dPx
k dx … (1.4)
Px
Integrating Eq. 1.4 within the limits of Po to P for intensity and 0 to b for the
thickness we can write,
p b
dPx
³p Px k ³ dx
0
… (1.5)
0
Solving, we get
P
ln kb … (1.6)
Po
79
Spectroscopic Methods Eq.1.6 is the mathematical expression for Bouguer-Lambert law or Lambert’s
law. Changing the logarithms in this equation to base 10 and rearranging we
get,
P k
log o b k 'b … (1.7)
P 2.303
Note that the ratio P/Po has been inverted to remove the negative sign.
Beer’s Law
P
or ln k "c
Po … (1.10)
Po k"
or log 10 c
P 2.303
Eq. 1.10 is the mathematical expression for Beer’s law. The Lambert’s and
Beer’s laws are combined and are expressed as per Eq. 1.11.
P
log o abc … (1.11)
P
In this expression, ‘a’ is a constant (combining two constants k¢, k² and the
numerical factor) and is called absorptivity (earlier called extinction
coefficient) whose value depends on unit of concentration used and is a function
of wavelength of the monochromatic light used. The concentration is generally
expressed in terms of grams per dm3 and b in cm. Therefore, it has the units of
cm-1g-1 dm3.
80
However, if the concentration is expressed as mol dm3 and b in cm then Molecular Spectroscopy
absorptivityis called molar absorptivity (earlier called molar extinction
coefficient) and is expressed as e. Its units are cm1 mol-1dm3. The modified
expression for the Beer-Lambert’s law then becomes,
P
log 0 Hbc … (1.12) Absorbance is defined as the
P logarithm of the ratio of the
The term log Po/P is called absorbance and is represented as ‘A’. intensities of the incident and
transmitted radiations.
P
A log 0 … (1.13)
P
The expressions for Beer-Lambert’s law then becomes,
P
log 0 A abc or Hbc ...(1.14)
P
The absorbance, A is related to another important term called transmittance
which is defined as the fraction of the incident radiation transmitted by the
absorbing medium. Mathematically,
P
Transmittance, T P … (1.15)
0
Example 1
The molar absorptivity of a substance is 1.0 × 104 cm1 mol1 dm3. Calculate the
transmittance through a cuvette of path length 2.0 cm containing 1.5 × 106mol
dm3 solution of the substance.
Solution
Given:
e = 1.0 × 104 cm1 mol1 dm3, c = 1.5 × 106 mol dm3 and b = 2.0 cm
From the Lambert-Beer’s law expression given above, you can note that
absorbanceis a direct measure of the concentration of the analyte if the thickness
of the absorbing medium is kept constant. This proportionality is exploited in
the determination of analyte concentration. However, the following factors
81
Spectroscopic Methods affect the validity of the Beer’s law and need to be controlled to take up
quantitative determinations.
z Presence of Electrolytes
z Temperature
SAQ 1
SAQ 2
In the absence of interfering substances, the measurements are made at The wavelength chosen for
the wavelength of the maxima ( O max) of the largest peak in the spectrum quantitative work is called the
analytical wavelength.
of the analyte. At this wavelength the absorbance is most sensitive to the
concentration. However, in some cases where the reagent, metal and
products all absorb light, it may so happen that at O max there is not much
of difference in the absorbance value for pure reagent and the metal
complex. In such cases we need to identify such a wavelength at which
there is large difference in the absorbance values of pure reagent and the
complex.
SAQ 3
4.2.5 Instrumentation
Today a wide range of instruments are available for making molecular
absorption measurements in the UV-visible range. These vary from simple
84
and inexpensive machines for routine work to highly sophisticated devices Molecular Spectroscopy
that can be used for specialised work and of course the routine jobs also.
However, the basic components of these instruments remain the same. The
five essential components of UV-VIS instruments are as follows
A. Radiation Sources
A radiation source must provide a stable high energy output over a broad range Glass prisms and lenses can
of wavelengths. There is no inexpensive source available that may provide be used in the visible region.
However, since glass absorbs
stable output over the entire UV-visible range (190 nm to 780 nm). For ultraviolet light, a quartz or
measurements in the UV region, electric discharge sources like hydrogen or a fused silica is a better choice
deuterium lamp are used. In these, the excitation of the gaseous molecules is for the material of the prism
because it can be used in both
brought about by the passage of electrons through the gas at low pressures. A the regions.
The gratings used for the
hydrogen lamp is commonly used in the spectrophotometers and gives light in
ultraviolet and visible region
the wavelength region of 160-375 nm. generally contain about 1200-
1400 grooves/mm.
Fig. 1.6 Radiation sources: a) deuterium lamp for UV range b) tungsten lamp for
visible range
85
Spectroscopic Methods On the other hand, a tungsten filament lamp is used as the radiation source for
visible range. This consists of a thin, coiled tungsten wire that is sealed in an
evacuated glass bulb. This gives radiations in the range of 350-2200 nm.
Prism Monochromators
You know that a prism disperses sunlight into seven different colours because
the radiations of different colours are refracted to different extent due to the
difference in the refractive index of glass for different wavelengths. Shorter
wavelengths are refracted more than longer wavelengths.
In a prism monochromator, shown in Fig.1.7, a fine beam of the light from the
source is obtained by passing through an entrance slit. This is then collimated
on the prism with the help of a lens. The refracted beams are then focused on
an exit slit. The prism is then rotated in a predetermined way to provide the
desired wavelength from the exit slit.
Grating Monochromators
C. Sample holder
The UV-VIS absorption spectra are usually determined either in vapour phase
or in solution. Tomeasure the UV spectrum of the analyte it is taken in a cell
called a cuvette which is transparent to the wavelength of light passing through
it. A variety of quartz cuvettes are available for the spectral determination in
the vapour phase. These are of varying path lengths and are equipped with
inlet and outlets. For measurements on solutions in the visible region the
cuvettes made of glass can also be used. However, since glass absorbs the
ultraviolet radiations, these cannot be used in the UV region. Therefore, most
of the spectrophotometers employ quartz cuvettes, Fig 1.9 as these can be
used for both visible and UV region. Usually square cuvettes having internal
path length 1.0 cm are used, though cuvettes of much smaller path lengths say
of 0.1 mm or 0.05 mm are also available.
Fig. 1.9: Quartz cuvettes for measurements in solution and in vapour phase
D. Detectors
The detectors are used to convert a light signal to an electrical signal which The eye of a common person
can be suitably measured and transformed into an output. There are three types is quite sensitive to notice
differences in radiant power
of detectors which are used in modern spectrophotometers. These are described transmitted through two
in the following paragraphs. coloured solutions. We can
say that eye is a natural
1. Phototube photosensitive detector in the
visible range. Therefore, early
A phototube consists of a photoemissive cathode and an anode in an instruments used eye or
photographic plate as the
evacuated tube with a quartz window as shown in Fig.1.10 (a). These two
detector.
electrodes are subjected to high voltage (about 100 V) difference. When a
photon enters the tube and strikes the cathode, an electron is ejected and is
attracted to the anode resulting in a flow of current which can be amplified
and measured. The response of the photoemissive material is wavelength
dependent and different phototubes are available for different regions of
the spectrum.
87
Spectroscopic Methods 2. Photomultiplier (PM) Tube
(a) (b)
Fig. 1.10: Detectors of UV-VIS radiation; a) Phototube and b) Photomultiplier tube
When a radiation falls on the cathode an electron is emitted from it. This is
accelerated towards the next photoemissive electrode by the potential difference
between the two. Here, it releases many more secondary electrons. These, in
turn are accelerated to the next electrode where each secondary electron releases
more electrons. The process continuous upto about 10 stages of amplification.
The final output of the photomultiplier tube gives a much larger signal than
the photocell.
The electrical signal from the detector is suitably amplified or processed before
it is sent to the recorder to give an output. The subtraction of the solvent
spectrum from that of the solution is done electronically. The output plot
between the wavelength and the intensity of absorption is the resultant of the
subtraction process and is characteristic of the absorbing species.
Fig. 1.14: The Jablonski diagram showing the phenomena of fluorescence and
phosphorescence
89
Spectroscopic Methods The electron spins in the excited state achieved by absorption of radiation
may either be parallel or antiparallel. Accordingly, this may be a triplet (parallel)
or a singlet (antiparallel) state. These are designated as T1or S1, respectively as
the case may be. The spin states are schematically depicted in Fig.1.15.
The energies of the two excited spin states are also different due to the difference
in the interactions between electrons: the energy of the triplet state usually
being lower than that of the singlet state. In the Jablonski diagram two excited
singlet states (S1 and S2) and a triplet state (T1) are shown.
The absorption of a photon of suitable energy causes the molecule to get excited
from the ground state to one of the excited states. This process is called as
excitation or activation and is governed by Franck-Condon principle.
According to this principle, the electronic transition takes place so fast (~1015s)
that the molecule does not get an opportunity to execute a vibration, i.e., when
the electrons are excited the internuclear distance does not change. The basis
for the principle is that the nuclei are very massive as compared to the electrons
and therefore move very slowly. The implication of Frank -Condon principle
is that the transition from the ground state to excited state can be represented
by vertical arrows in the Jablonski diagram.
The electronic transition takes place from the state of lowest vibrational energy
of an electronic ground state to any of the vibrational levels of the excited
electronic states as shown by the vertical upward arrows in the diagram. The
excited state achieved would depend on the energy of the photon absorbed. In
the diagram the two excited states as S1 and S2 have been shown. These are
obtained by the absorption of the radiation in the region of wavelength O1 and
O2 respectively.
z Nonradiative deactivation
z Radiative deactivation
Vibrational Relaxation
In the higher vibrational levels of an excited state, the molecule rapidly loses
(in < 10-12 s) its excess vibrational energy by collision with other molecules
and falls to the lowest vibrational level of the excited state. This nonradiative
mode of relaxation is called vibrational relaxation,and the energy lost in this
process is dissipated as heat to the surroundings.
Internal Conversion
If the intensity of exciting
Once the molecules that are excited to an electronic state (say S2)higherlightthanis kept constant as its
the S1 state reach the vibrational ground state in the electronic level, these can
wavelength is changed, the
plot
pass to a higher vibrational level of a lower excited state (S1) which has the of emission against
exciting wavelength is
same energy. This process is referred to as internal conversion. The moleculeknown as the corrected
can continue to lose energy in this state in a nonradiative way (vibrational excitation spectrum.
relaxation) until it reaches the lowest vibrational level in this excited state.
Intersystem Crossing
z Radiative Deactivation
When the molecule in the excited state (S1) relaxes down to the lowest
vibrational level it may emit a photon and come down to the electronic ground
state (S0). This process is called fluorescence and takes about 109 s. Another
radiative relaxation process can arise from the excited molecule that had crossed
over to the triplet excited state by intersystem crossing and has relaxed to the
vibrational ground state in the triplet excited state. In such a case the molecule
emits a photon and comes down to a vibrational mode of the electronic ground
state, S0. This phenomenon is called phosphorescence. As the transition from
a triplet state to a singlet state is theoretically forbidden, it does not take place In fluorescence, the spin
readily. Thus, while the fluorescence emission can take place within 109"106 multiplicities of the ground
and emissive excited states
seconds, the transition from an excited triplet state to the ground state in case are the same whereas for
of phosphorescence requires at least 104 seconds and may take as long as 102 phosphorescence, the excited
seconds. and ground states have
different spin multiplicities.
Since the fluorescence emission takes place only after the excited molecule
has relaxed to the vibrational ground state of the S1 state i.e., after having lost
part of its energy, the energy of the emitted radiation is lower than that of the
excitation radiation. This means that the wavelength of fluorescence emission
would be greater than the excitation wavelength. Further, since the energy of
the triplet excited state of the molecule is lower than that of the associated
91
Spectroscopic Methods singlet state; the transitions to the ground state are associated with the emission
of light of still lower energy. As a consequence the phosphorescence occurs at
longer wavelengths than fluorescence. Fig.1.16 shows the excitation,
fluorescence and phosphorescence spectra of phenanthrene
Fig. 1.16:The excitation (E), fluorescence (F) and phosphorescence (P) spectra of
phenanthrene
Since a given analyte can fluoresce only after it has absorbed radiation, an
excitation spectrum consists of the wavelengths of light that the analyte is
able to absorb. In other words generally the excitation spectrum of a molecule
is the same as its UV-VIS absorption spectrum. Generally, the maximum in
the fluorescence spectrum of a compound occurs at longer wavelength than
the maximum in the absorption spectrum. The wavelength difference between
the absorption and fluorescence maxima is called the Stokes shift.
SAQ 4
iii) a molecule lowers its vibrational energy by losing its excess energy
as a photon.
93
Spectroscopic Methods
SAQ 5
4.3.3 Instrumentation
All fluorescence instruments use essentially the same components as are used
in absorption spectrophotometers. However, the geometric arrangement of the
components is somewhat different. This is due to the reason that any transmitted
radiation is not measured along with the fluorescence. The absorption and
transmission of radiant energy occur only along the direction of the incident
light whereas the fluorescence radiation emanates in all directions. The
detection of transmitted radiation is avoided by placing the detector at a right
angle to the transmitted beam, as shown in Fig. 1.18.
A. Sources
The source in fluorescence spectrometer must be more intense than that required
for UV-VIS absorption spectroscopy because the fluorescence intensity, IF is
directly proportional to the source power. An increase in P0 will produce a
94
larger signal for a given concentration and thereby improve sensitivity. The Molecular Spectroscopy
tungsten filament and deuterium lamps used in absorption spectrophotometers
are generally not suitable for fluorescence instruments as they lack the desired
intensity.
B. Wavelength Selectors
The low-cost instruments designed for routine determinations are simple filter
fluorimeters. Such instruments are used when it is sufficient to measure fluorescence
intensity at a single excitation and emission wavelength. These employ fixed filters
to isolate both the excitation and emission wavelengths. In order to isolate one
particular wavelength from a source what we need is just a pair of cut-off filters.
Absorption filters are comprised of a suitably absorbing substance or substances
dispersed in gelatin, glass or plastic. The filter fluorimeters are used primarily in
environmental field screening, hospital or clinical settings and other applications
in which low cost and small size are crucial.
C. Sample holder
z The optical surfaces of the cell should not be touched with hand as it
invariably leaves an invisible film that may change the light transmission
and reflection characteristics of the cell, especially in the ultraviolet region.
z The cuvettes should be handled only at the top portions of the side plates
that do not face the optical axis.
95
Spectroscopic Methods z The samples should be transferred to the cuvettes with the help of a dropper
or a pipette.
z The cuvettes should be rinsed with the analyte solution before filling and
the overfilling should be avoided.
D. Detectors
E. Output Devices
The output from the detector is suitably amplified and displayed on a read out
device like a meter or digital display. The sensitivity of the amplifier can be
changed so as to be able to analyse samples of varying concentrations. In
some instruments the display can be adjusted to directly give the output in
terms of the concentration. Nowadays, the instruments have microprocessor
controlled electronics that provides outputs compatible with the printers and
computers whereby minimising the possibility of operator error in transferring
data.
SAQ 7
b) Photomultiplier tubes with low noise and high sensitivity are preferred
over simple photo tubes so as to detect …………... signals.
The Raman spectroscopy also involves vibrational energy levels but it differs
96
from IR spectroscopy on many counts. These include the nature of radiation Molecular Spectroscopy
used, the mechanism of interaction between the radiation and matter, the
necessary condition for the interaction, required instrumentation and the
resulting spectra etc. Further, even the information available from it is different;
in fact it is complimentary to the one available from IR spectroscopy. The UV-
VIS spectroscopy discussed above involved absorption of radiation whereas
the fluorescence spectroscopy is based on emission of radiation. The Raman
spectroscopy on the other hand involves scattering of radiation.
Fig. 1.19: Raman spectra for CCl4 using 488 nm laser source
You may note that the intense peak in the centre of the spectrum has the same
wavelength (or wavenumber) as that of the incident radiation. This is the
Rayleigh peak and the other signals on either side of the Rayleigh peak are the
Raman lines. The lines to the left of Rayleigh peak and having lower value of
wavenumber are the Stokes lines while the ones to the right and having higher
value of wavenumber are the anti-Stokes lines. Stokes lines are at lower energy
while the anti-Stokes lines are at energy greater than the Rayleigh peak. You
may note two more features in the spectrum firstly, the Stokes and anti-Stokes
lines are equidistant from the Rayleigh line and secondly, the Stokes lines are
much more intense as compared to the anti-Stokes lines.
'Q ( Q s Q 0 ) cm 1 …(1.19)
Where, Q s and Q 0 are the wavenumbers of the source (or incident) radiation
and the observed scattered lines respectively. It is obvious that the Raman
shifts of the Stokes lines would be positive while for anti-Stokes lines, these
would be negative.
When a molecule is placed in a static electric field, the positive and negatively
98
charged particles constituting the molecule get attracted to the opposite poles Molecular Spectroscopy
of the applied field. This leads to a charge separation and consequently to the
development of an induced dipole moment. In simple words, we say that the
molecule has become polarised. This tendency of the molecule to get polarised
is called polarisability i.e., the ability to get polarised. The magnitude of the
induced dipole moment is proportional to the strength of the applied field and
is given by the following equation.
P DE … (1.20)
The proportionality constant D is called polarisability- a measure of the ease The polarisability (D) of the
with which the molecule (or a bond in it) can get polarised. The polarisability molecule depends on the
bond length; shorter bonds
can be anisotropic, i.e., the electrons of a bond may be polarised to different being difficult to polarise
extents when the field is applied in different directions. For example, the than longer bonds.
hydrogen molecule gets polarised to different extents when the field is applied
in the direction of the bond or in the direction perpendicular to it, as shown in
the margin. What would happen if we place a molecule in an oscillating electric
field like that of an electromagnetic radiation? Let us see.
where E is the electrical field strength at any time t and E0 is the amplitude of
the wave. When such a radiation is made to interact with a molecule it would
induce a dipole moment in it. Further, as the electrical field of the radiation is
oscillating the resulting induced dipole moment would also oscillate at the The necessary condition to
same frequency as that of the radiation. That is, observe vibration Raman
spectrum is that the
P DE DE0 cos( 2SQ ext ) … (1.22) polarisability of the bond
must change in the course of
This oscillating dipole would generate a radiation of the frequency, Q ex i.e vibration
Rayleigh scattering.
As you are aware, the molecules are never stationary and always undergo
vibration (and may be rotation) motions. These motions may cause a change
in the polarisability of the molecule. When such a system vibrating with a
frequency of Q vib interacts with the oscillatory electric field of the
electromagnetic radiation then the scattered radiation may have three different
components. One of these has frequency equal to the incident frequency ()
while the frequencies of the other two are equal to and respectively. The first
component with a frequency of is the reason for the Rayleigh scattering. The
other two components are the sources of the anti-Stokes and Stokes lines
respectively.
The three components in the scattered radiation are observed only when the
vibration causes a change in the polarisability. If the polarisability does not
change in the course of vibration only Rayleigh scattering is observed.
Therefore, we can conclude that in order to observe Stokes and anti-Stokes
lines the polarisability of the bond must change in the course of vibration.
99
Spectroscopic Methods Let us raise a question. Since both the IR and Raman spectra involve transitions
amongst the same vibrational levels, would the two spectra be the same? As a
first response, the answer to the question about the similarity of IR and Raman
spectra raised above, could be in a firm affirmative or one may respond, ‘may
be’. However in actual practice the two spectra are generally not identical and
in case of a number of molecules, the spectra in fact do not have anything in
common. The Raman and IR spectra for a simple molecule CCl4 is given in
Fig. 1.20. You may note that only some of the signals are common in the two
spectra and also the common signals are of different relative intensities.
Fig. 1.20: IR and Raman spectra of CCl4; note the similarities and the differences
You would recall that only those modes of vibration are IR active which have
an associated change in dipole moment. Similarly only those vibrational modes
are Raman active which are associated with a change in polarisability.
100
4.4.2 Instrumentation for Raman Spectroscopy Molecular Spectroscopy
Like the IR instruments, the Raman spectrometer, has the following basic
components.
As Raman experiments
always involve the
measurement of small energy
shifts of the order of 100 to
3000 cm1 from the excitation
energy, a monochromatic
source is highly desirable and
lasers meet this requirement
the most.
Fig. 1.21: Schematic representation of the experimental set up of a dispersion Raman Unlike most chemical
spectrometer analysis techniques, Raman
spectrometry does not require
You may find the arrangements to be similar to that of the corresponding IR any special preparation of the
sample. Since Raman
instruments with the only difference being that the scattered radiation is spectrometry involves only
collected at right angles to the radiation and it is passed through a filter before illuminating a sample with a
sending to a transducer for detection. The Raman spectrometers differ from laser and collecting the
scattered photons, no contact
the IR instruments in terms of the sources, the sample handling devices and with the sample is needed at
the transducers used for detecting the scattered radiation. The first ever Raman all. This makes Raman
instrument was constructed in 1928 which used the Sun light focused by a spectroscopy a non-
destructive technique.
telescope to achieve a high enough intensity in the scattered signals. The
spectrometers that followed used mercury arc lamps as a light source. With
the advent of lasers with highly desirable properties of high intensity, single
wavelength and coherence, the modern spectrometers almost exclusively use
laser sources.
For analysing gaseous samples, the cuvette consists of a cylindrical glass tube
with mirrors on both the ends; one of the mirrors has a small window to let the
incident radiation to pass through. When the cuvette containing the sample is
placed in the radiation path the incident laser beam enters the sample through
the window and undergoes multiple reflections in the sample. The scattered
light at the right angles to the sample is then suitably collected and measured.
The original Raman instrument that lead to the discovery of the Raman effect
used eyes as the detector. In initial versions of the spectrometers, photographic
plates were used to detect the scattered radiation. This was followed by more
sensitive photomultiplier tubes (PMTs) that allowed electronic data collection
and manipulation. However, today most of the Raman spectrometers are FT
instruments which use cooled germanium photoconductors as transducers.
SAQ 7
What is Raman shift? The Raman shift positions of Stokes and anti-Stokes
lines are equal but opposite. Comment.
...................................................................................................................
...................................................................................................................
...................................................................................................................
...................................................................................................................
...................................................................................................................
SAQ 8
II. The Raman spectra are generally taken for aqueous solutions only.
SAQ 9
I. When the collisions between the photons and molecules are ……….,
the photons would be…………..with the same frequency.
102
II. If in the event of an inelastic collision with photons the molecules get Molecular Spectroscopy
103
Spectroscopic Methods
4.6 LET US SUM UP
In this unit we have discussed about molecular spectroscopy i.e., the study of
interaction of EM radiation with molecular systems. We began the unit with a
discussion on UV-VIS spectroscopy wherein we discussed the origin of the
UV-VIS spectrum. Then we discussed the characteristics of the spectrum i.e.,
the position of the signal, its intensity and width. This was followed by the
description of the instrumentation used and the explanation of the principle of
UV-VIS spectrophotometry as an analytical tool.
104
10. Describe briefly different mechanisms of nonradiative relaxation of an Molecular Spectroscopy
excited electronic state.
11. The Raman spectrum of an amino acid cystine was obtained by using a
488 nm laser. The observed Raman shift of the most intense Stokes line is
105 cm1. Compute the positions of the most intense Stokes and the
corresponding anti-Stokes lines.
12. State mutual exclusion principle for deciding IR and Raman activities.
ANSWERS
Self Assessment Questions
1. P0 = 85.4 b = 2.00cm
P = 20.3 c = 1 × 10-4M e =?
Since A= e b c; º = A/b.c
0.624
Substantially the value H
2.0 cm u1 u104 M
1 1 3
=> 3120 cm mol dm
1 1 3
2. º =14000 cm mol dm
c= ?b = 1cm A= 0.85
A = ecbc = A/ e.b
c = 0.85/14000 × b
= 6.07 ×10-5M
5. a) ii)
b) i)
6. The fluorescence emission takes place only after the excited molecule has
relaxed to the vibrational ground state of the S1 state i.e., after having lost
part of its energy. As the energy of the emitted radiation is lower than that
of the excitation radiation it is observed at a longer wavelength than that
of the excitation radiation
7. a) source power
b) weak optical
105
Spectroscopic Methods 8. Raman shift refers to shift in the observed scattered frequency from the
frequency of the excitation radiation. Raman shift, 'Q is defined as
Δν ( νs ν0 ) cm 1
Both the Stokes lines and anti-Stokes lines arise from inelastic collisions
with the incident radiation. A Stokes line arises when the radiation excites
a given vibration mode whereas the corresponding anti-Stokes line arises
due to relaxation of the same vibration. The Raman shifts of the two are
equal because they involve same vibration and are opposite because one
involves excitation while the other concerns relaxation.
9. a) True
b) False
b) lesser
c) polarisability
Terminal Questions
1. Some of the factors responsible for the deviation from Beer and Lambert’s
law are as follows:
z Presence of Electrolytes
z Temperature
2. The atomic weight of iron is 56, therefore the molar concentration of the
solution containing 1.00 microgram of iron (II) present in 10 mL of solution
will be
1u 10 6 g 1000mL
1.786 u10 6 mol per dm .
3
u
56 10mL
0.02
H A / b.c
1.786 mol dm 3 u 1 cm
= 1.12 × 104
The spin multiplicities of the singlet and triplet states are 1 and 3,
respectively.
107
Spectroscopic Methods z In internal conversion the molecule is transferred from an excited
singlet state to another singlet state of the same energy.
Since the Raman shift of the most intense Stokes line is at 105 cm-1 therefore
its wavenumber would be
12. The mutual exclusion principle states that, “for a molecule having a center
of symmetry the Raman active vibrations are IR inactive and vice versa”.
108