Professional Documents
Culture Documents
SM 617210009 Wa
SM 617210009 Wa
Brand: GE Medical
Model: Carestation 620 A1
SN: SM617210009WA
Technical Observation:
“Please do Checkout” message appears on screen
Work Done:
Conducted detailed assessment
Conducted Checkout
Ventilator Leak test Failed
o Cannot pressurize circuit
“Is patient Y open? Leak in bellows?” alarm message appears on screen
Replaced new set of breathing circuit
o Same results yielded
Replaced new bellows
o Same results yielded
Replaced new O2 sensor
o Same results yielded
Remarks:
Unit can still be used using Bypass mode
“Please do Checkout” message remains on screen