Shimadzu Sonavision Versa I.G.
Shimadzu Sonavision Versa I.G.
Shimadzu Sonavision Versa I.G.
SonialVision Versa
combined with CH-200
Revision
Rev D2010 :
SN
SonialVision Versa Installation Guide Page 2
Introduction
This steering guide is used to help FSE installing the Sonialvision Versa system. It can be
used as a guide to facilitate the installation process.
It also has the current useful information and tips that are sometimes not available in the
manuals.
System Software
Dar-8000i
Ixworks
IAPDB
DDM
FPGA0
DAR FPGA1
FPGA2
Panel
GIM
GIM Remote
Side Station
UD-CONT (cabinet )
UD CONSOLE
B-40 DATA
NEX-SH
CLC-MAIN
ZS-100
Remote
Local
CH-200/M
BR-120/M
Others CXDI-50G
Page 4
Generator UD150B-40
TUBE1 G292
TUBE2 P324DK-85
Starter SA-60
Tube Heat cooler HE-125 Option
Table ZS-100IR
Table Collimator R-50
CTM CH-200
CTM Collimator R-30H
Unit Unit name S/N
System Options
Side Station
TomoSynthesis (Safire system only)
RSM-DSA
Slot Radiography (Safire system only)
Dual Energy (Safire system only)
Dicom Print
Dicom Send
MWL
MPPS
Page 7
Photo timing kit (SPT-C42) option 2 ( in case of system with Direct lens)
PHOTO CONT 2002 pcb
PHOTO I/F 2002 pcb
Cables, spacers, screws
Page 8
1. Fluoroscopy unit kit: Refer to chapter 8 of B-40 installation manual for detail.
Set sw4-6= OFF on UD cont-2002 PCB (when combined with two mA power 2002 PCB).
BH Filter
The Gold filter is not necessary to install. It has been
noticed in the field that the gold filter reduced the
image contrast. Always measure HVL to confirm that it
is 2.9mm at 80kv ( FDA )
Connect CN15 from SGIM PCB to TV7-TV2 on Fluoro 2002 PCB. In B-40 option page,
set DC fluoro to TV and Pulse fluoro to TV
Page 10
DAR-8000i Connection
Monitors Connection
Refer to DAR-8000i installation manual
CH-200 connection
Refer to CH-200 installation
Page 11
Arcnet Connection
Attention: CH-200 arcnet connection is not used!!!!! Make sure to configure it with B-30 gen-
erator not the B-40 generator.
SonialVision Versa Installation Guide Page 14
II Cable Connection:
The II comes with extra cables that are not used. Connect them as follows:
Page 16
SA-60 Connection
The SA-60 comes prewired from the factory for the 1.5MHU x-ray tube. Therefore before
power up the SA-60, the proper wiring for each x-ray tube must be performed first. Otherwise,
x-ray tube will be damaged!!!!!!!
SA-60 Prgramming
SN 2007-08-07
TUBE 1 TUBE 2
SW2-5 = ON SW2-6=ON
SET HIGH SPEED HOLD FOR 3MIN
SW3-5,8=ON ( SEE NOTE)
DISPLAY MODE
STARTOR
SW2-2,3,4,=ON
TYPE R OR
Q
TUBE 1 OR
R STATOR Q STATOR 2
SW3-3=OFF SW3-3= ON
TUBE 1 TUBE 2
WRITE DATA TO MEMORY SW2-5=ON SW2-6=ON
SW2-1= ON
WAIT FOR 2 SEC.
SW2-1=OFF
PUSH RESET
After programming, always check the content if the programming has been performed cor-
rectly. Set the dipswitch sw2 and sw3 to display mode using the flow chart on previous page
and the table below as a guide.
Note: SA-60 shipping second quarter of 2009, will have a new CPU pcb (NEXSH) that allows
programming by PC instead of dip switches progamming !!
Page 21
Unused cable
• The B-40 accepts 3 phases voltage ranging from 380 to 480V without using auto XAT-2H
transformer. Always match T1 and T2 transformers according to the line voltage.
Ex: if the line voltage is 380V, tap T1 and T2 to -200 and +180 to get a total of 380V.
• We recommend disconnecting X-ray high voltage portion when power up for the first
time during installation. T his allows you to check basic function without generating
high voltage to the inverter circuit of the generator and doesn’t heat filament to the X-Ray
tube. To do so:
-Disconnect J2 (MGL) on terminal 2002 pcb
-Disconnect JS1 cable on UD cont 2002 pcb
-Disconnect J2 on mA power 2002 pcb
-Set sw4-4= on on UD cont 2002 pcb
After Power ON, always check to confirm that L0, L100 of the B-40 should be 100V (+/- 5V) AC.
The B-40 is always boot up in Japanese language during installation. You need to initialize it to
English language and replace 3 Japanese labels to English labels on the Console. Lots of
people can’t locate English labels: they are attached in a small bag of the B-40 console
SonialVision Versa Installation Guide Page 22
The System Preset CD (which includes B-40 system configuration setting (initial data), APRs for
DR fluoro , Table Bucky APRs, Wall Bucky APRs and General Radiography APRs. Please load
the system preset data from CD to the B-40 in order to save a few days of setup. Refer to B-40
installation manual, chapter 9 section 9.22. Use command name: BKUP. Select 3 to upload ->
Select File -> transfer->Xmodem -> send. From Select File, the system has a time out of about 15
sec to do this process; otherwise it will time out and won’t upload. If it happen, you need to reset
the B-40, and redo the upload again
B-40 Setting
RATIO KV MA
MAXKV 150 534
MINKV 80 1000
MINRATING 80 KW
FREQUENCY 60
Input Tube chart on the next page if G292 is cooling unit option
TECH
TECH NO TECH CODE NAME TUBE FLUO PH CODE PH DETECT PH FIELD TECHSYSTEM
0 01h GR 1 OFF 0 0 0 1
1 17h XDR 1 0 1 0 0 14
2 01h GR 2 OFF 0 0 0 1
3 02h BU1 1 OFF 5 1 3 4
4 05h BU2 2 OFF 5 2 3 3
5 04h PL 1 OFF 0 0 0 04
SonialVision Versa Installation Guide Page 24
Edit mode
Main Page: Tube Info
Anode High
Focus L&S
kV MAX 150
kV MIN 40
HU Factor (%) 90
Cine Factor
(%) 90
Page 2: Tube :HU
B-40 Setting
FLUO
FLUO NO NAME FOCUS HLC PULSE FLUORO PULSE RATE PULSE WIDTH
0 DOSN 0 0 0
1 DOSL 0 0 0
2 PULSE N 0 0 1 30 10
3 PULSE B 0 0 1 15 16
4
5
6
COMMUNICATION
UD-COMM I/F
PORT 0 PORT 1
Not used Not used
VIA ARCNET
DAR-6000 MECHA 1 MECHA 2 CH-200 #1 CH-200 #2 RAD DR
used used Not used Not used Not used Not used
FPD GRID#1 FPD GRID#2 FDA CONSOLE
Not used Not used used
SonialVision Versa Installation Guide Page 26
B-40 Setting
Refer to
FPD Canon 50G DR Tray for BR-120 wall bucky Connection and Table Top
SonialVision Versa Installation Guide Page 33
Note: B-40 software v2.20 has new feature: FVR auto calibration. Select AUTO-INIT to start calibration.
Page 34
TUBE1: G292
FOCUS mA 40kV 60kV 80kV 100kV 125kV 150kV
S
100
125
160
200
250
320
FOCUS mA 40kV 60kV 80kV 100kV 125kV 150kV
100
L
200
250
320
400
500
630
SonialVision Versa Installation Guide Page 35
TUBE2: P324DK-85
FOCUS mA 40kV 60kV 80kV 100kV 125kV 150kV
S
100
125
160
200
250
320
FOCUS mA 40kV 60kV 80kV 100kV 125kV 150kV
100
L
200
250
320
400
500
630
Page 36
DOSN (Fluoro 0)
50kv 60kv 80kv 90kv 100kv 120kv
FmA 0.4 0.6 7.3 6.8 6.0 5.2
FVF
Calibrate FVF at each KV point (IBS = OFF) so that the FmA readout matches the Set FmA in the FmA curve table.
DOSL (Fluoro1)
50kv 60kv 80kv 90kv 100kv 120kv
FmA 0.2 0.3 3.65 3.4 3.0 2.6
FVF
Calibrate FVF at each KV point (IBS = OFF) so that the FmA readout matches the Set FmA in the FmA curve table
PULSE N (Fluoro 2) - Refer to the back of the document how to calibrate using auto calibration
50kv 60kv 80kv 90kv 100kv 120kv
Pulse mA P_P 6 12 30 24 24 24
Average Display 1.8 3.6 9 7.2 7.2 7.2
Value
FVF
Bariatric Only
PULSE B (Fluoro 3) -Refer to the back of the document how to calibrate using auto calibration
50kv 60kv 80kv 90kv 100kv 120kv
Amplitude of 6 12 30 24 24 24
FmA
Average Display 1.35 2.7 6.75 5.4 5.4 5.4
Value
FVF
Note: Always confirm that the skin dose never exceeds 10R/min.
Always measure to confirm that the Maximum skin dose never exceeds 10R/min.
The method to measure maximum skin dose is to install the probe at 30cm from table top and measure Dose at
Max KV.
SonialVision Versa Installation Guide Page 38
Syntax:
Ex:
Let’s change 4:BONE N to PULSE B ( B stands for Bariatric)
At the UD prompt:
UD>fluoname
Then type
UD>fluoname 4 PULSE B
Let’s explain how to do Method#2: Go to FVR->Select small filament->assign 2mA, 3mA, 6mA, 8mA, 10mA and 12mA
SonialVision Versa Installation Guide Page 40
• At 2mA, calibrate the FVR at 40KV and 60KV. You don’t need to calibrate the whole line ( just 40kv and 60kv
for 2mA station) interpolate between 40KV and 60KV. I use the above example: (3.98+3.93)/2= 3.96. Write
3.955 on a paper
• At 3mA, Calibrate the FVR at 60KV point.. Write your calibrated result on a paper. In this example is it 3.96
• At 6mA, calibrate the FVR at 100KV and 125KV. Then interpolate the data between the two points.
And so on for 8mA, 10mA and 12mA.
Pulse Fluoro is a Pulse waveform (ON, OFF, ON etc..), the Generator can display only Average value of the pulse
waveform.
How the Average value is is calculated from a pulse wave form for 30 F/S.
Go to service mode->setup->Fluo
On line 03, select edit and add the pulse fluoro mode as shown below. Then rename it from BONE N to
PULSE B.
SonialVision Versa Installation Guide Page 43
Always measure the maximum skin dose so that it doesn’t go higher than 10R/min
SID=1100mm
Probe- =30cm from tabletop
Go to B-40 service mode -> signal and set the signal as follows.
The configuration is done in DAR-8000i service mode. How to log in service mode:
Boot up the Dar-8000i and push the shift key and hold it until log in screen appears
Username: shimadzu
Password: Dar8k
Double click Service application. Perform system configuration according to your system
combination such as camera type, table type etc.. (Refer to DAR-8000i installation M517-
E380 starting section 5.1)
SonialVision Versa Installation Guide Page 46
DAR-8000i
IP address:
Subnet Mask:
Default gateway:
Dicom Print
Name:
IP address:
AE Title:
Port Number:
Film Orientation:
Film Size:
Dicom MWL
Name:
IP address:
AE Title:
Port Number:
Select unprocessed/ processed Dicom auto storage: select processed, update and Exit
Select Edit “Modalities Values for work list query. Change number to 3 and add CR. Update
and Exit
SonialVision Versa Installation Guide Page 49
UD>flashtbl w Select: 6
Setting Flash Value for pulse Fluoro
Flash Volume value [1] 20
Flash Volume value [2] 20
Flash Volume value [3] 20
Flash Volume value [4] 20
Flash Volume value [5] 20
Flash Volume value [6] 20
Flash Volume value [7] 20
Flash Volume value [8] 20
Flash Volume value [9] 20
Flash Volume value [10] 20
Flash Volume value [11] 20
Flash Volume value [12] 20
Flash Volume value [13] 20
Flash Volume value [14] 20
Flash Volume value [15] 20
Flash Volume value [16] 20
Flash Volume value [17] 20
Flash Volume value [18] 20
Flash Volume value [19] 0
Flash Volume value [20] 0
SonialVision Versa Installation Guide Page 50
UD> flashopt 12
Page 51
TV imaging adjustment
Tips:
ο Always remove Aluminum shipping plate from II during installation.
When installing Mu Metal shield in front of the II, make sure that the Mu Metal shield doesn’t get in con-
tact with the II glass. when table is tilting, the MU metal shield may get in contact with the II which causes
the image to be fuzzy
Acrylic = 19 cm
• IBS Response
ο DOSN ( Fluo0) : up=50 down= 50
ο DOSL ( Fluo1) : up=50 down= 50
ο PULS (Fluo2) : up=25 down= 25
Page 53
Symptoms:
IBS fluoroscopy balance point is not stabilized after DR radiography
IBS fluoroscopy balance point is not stabilized after changing I.I. field size.
DR radiography image brightness is not stabilized.
DR radiography image brightness after Pulse Fluoroscopy is different from the brightness after
DC fluoroscopy.
Procedure:
1)Power OFF the system.
2)Remove the cover of DAR-8000i Camera head. Confirm the potentiometer R19 on top board in the
Camera head as shown.
3)Adjust R19 to get approximately 80 ohm (from 80 ohm to 100 ohm) between pin No.1 and pin No.3
of R19. ( pin No.1 and pin No.2 are shorted ) (R19 value is max. 200 ohm 30 turns)
Note) If this value is too much, Iris may hunting, Fluoroscopy IBS KV value may hunting.
SonialVision Versa Installation Guide Page 54
Note:
When you execute DR Radiography Iris adjustment by [Manual Setting], you should make Radiography after
Fluoroscopy, and check window level.
When you execute Fluoroscopy Iris Adjustment by [Manual Setting], you should make Fluoroscopy after Radiog-
raphy, and check Fluoroscopy kV.
When you execute Pulse Fluoroscopy Iris Adjustment by using [System Manual] function, system does not set
value for all pulse rates at once. You should set same value as “Iris Control Level” of 30f/s at other pulse
rate by [Manual Setting]. When you use [Manual Setting] mode, after finishing one pulse rate setting you
should exit from the mode, and re-enter to the mode for next pulse rate setting.
Page 55
Digital Tomo uses II with Dar-8000i to acquire an image where as conventional Tomo uses Cassette
or Canon DR to get an image.
Digital Tomo uses the same Technique as Spot/SerialTechnique (XDR) so it is not necessary to cre-
ate an extra technique. All we need is to add extra APR for digital Tomo under XDR technique.
Calibrate the iris until you see white dots about 100 dots
The height adjustment is done by connecting the PC to ZS-100IR cabinet . Refer to Z-100IR installation
manual
Page 58
Memory shot is performed in the B-40 service mode. Set the acrylic according to the table below
and perform fluoro at least 3~5 sec until IBS is stabilized and then register the FKV. Repeat it on
each FOV and Each fluoro mode (fluo0, fluo1, fluo2).
At the highest thickness, set the kv to maximum (120kv).
Memory-
FOV 7cm 14cm 21cm 28cm
DOSN Max 120kv
Large 120kv
Middle 120kv
Small 120kv
DOSL Max 120kv
Large 120kv
Middle 120kv
Small 120kv
PULSE N Max 120kv
Large 120kv
Middle 120kv
Small 120kv
PULSE B Max 120kv
Large 120kv
Middle 120kv
Small 120kv
Fluoro KV, mA fluctuation countermeasure
IBS margin function to countermeasure IBS KV, mA fluctuation
Required Software Revision:
V1.86 and later
Installation Procedure
On the PC, Run Tera term program and set the baud rate: 9600, 8, odd, 2, 1, none
Connect the PC to UD-cont 2002 pcb using null modem cable ( 9 pins)
Enable “ibsmargin” function as described below:
UD>ibsmargin
1 2 3 4 5 6
FkV 50 70 80 90 100 115
Margin 54 54 50 50 50 50
1 2 3 4 5 6
FkV 50 70 80 90 100 115
Margin 54 54 50 50 50 50
UD>
UD>ibsmargin on p
1 2 3 4 5 6
FkV 50 70 80 90 100 115
Margin 54 54 50 50 50 50
1 2 3 4 5 6
FkV 50 70 80 90 100 115
Margin 54 54 50 50 50 50
Page 60
On ZS-100IR system the FmA curve is higher than YSF-300 system. To prevent the FOC error, perform the
following procedure:
Connect PC to UD-cont 2002 pcb with Teraterm and
UD> Null Modem cable (9600, 8, Odd, 2, None).
UD>flashdelay w At the UD prompt, type flashdelay w
IFV flash delay table description.
------------------------------------------------------
Table 1: Radiography Large focus.
Table 2: Radiography Small focus.
Table 3: N/A
Table 4: Fluoroscopy- Continuous.
Table 5: Fluoroscopy- Pulse Large focus.
Table 6: Fluoroscopy- Pulse Small focus.
Type 4 to select con-
Table 7: Fluoroscopy- Pulse(Large FmA) Large focus. tinuous fluoro
Table 8: Fluoroscopy- Pulse(Large FmA) Small focus.
------------------------------------------------------
UD>
UD>flashopt 2
UD>
Confirm that DC FLUO
UD> DELAY [ON}. Disconnect
the PC and Reset the gen-
erator.
*Now Flash Opt Settings.
EXAMPLE
MEMORY SHOT F-T R-T
Fluoro X-ray Condition Bar AVE SER Low KV AVE IVR AVE (S)
Barium AVE
(FkV) (S)(0) SER (S)(0) (+1) (S)(0)
Acrylic
Mid- mA/
Max Large Small kV mA/msec kV mA/msec kV mA/msec kV
Fluoro0=DOS dle msec
N 7cm 57 58 59 60 68 200/100 65 200/100 75 200/200 68 200/200
14cm 65 67 68 70 80 200/100 75 200/100 82 200/200 80 200/200
21cm 77 82 86 88 93 200/100 85 200/100 88 200/200 87 200/200
28cm 120 120 120 120 105 160/100 95 160/100 110 160/200 105 160/200
Fluoro X-ray Condition Barium XL (L) Low kV AVE (L) Barium AVE
IVR XL (L) (+1)
(FkV) (0) (0) (S)(0)
Acrylic
Mid-
Max Large Small kV mA/msec kV mA/msec kV mA/msec kV mA
dle
Fluoro1=DOSL
7cm 60 61 63 64 68 200/200 65 200/200 75 200/200 68 200/200
14cm 72 72 74 78 80 200/200 75 200/200 82 320/200 80 200/200
21cm 86 89 100 110 87 320/200 85 320/200 90 320/200 87 200/200
28cm 120 120 120 120 105 400/200 95 400/200 110 400/200 105 160/200
Fluoro X-ray Condition Bar XL SER (L) Low KV AVE Myelo AVE(S) Barium AVE
(FkV) (0) SER (L)(0) (+2) (S)(0)
Acrylic
Mid-
Max Large Small kV mA/msec kV mA/msec kV mA/msec kV mA
Fluoro2=PULS dle
EN 7cm 50 50 53 53 68 200/100 65 200/100 75 200/200 68 200/200
14cm 60 60 64 66 80 200/100 75 200/100 82 200/200 80 200/200
21cm 73 76 79 84 93 320/100 85 320/100 88 200/200 87 200/200
28cm 120 120 120 120 110 400/100 95 500/100 110 160/200 105 160/200
Fluoro X-ray Condition Low kV AVE (S) Myelo XL (L) Barium AVE
15 FPS
(FkV) (0) (+2) (S)(0)
Acrylic
Mid-
Max Large Small kV mA/msec kV mA/msec kV mA/msec kV mA
Fluoro3=PULS dle
EB 7cm 52 53 55 56 65 200/200 75 400/32 75 200/200 68 200/200
14cm 62 65 67 70 75 200/200 85 400/32 82 320/200 80 200/200
21cm 77 82 88 99 85 200/200 95 500/32 95 320/200 87 200/200
28cm 120 120 120 120 95 160/200 115 500/32 110 400/200 105 160/200
Page 62
FVR mode:
The auto cal starts at lowest mA station with
80KV , 40ms
Procedure
1. Select GR technique
2. Go to FVR mode
3. Select tube # ( 1 or 2)
4. Select small or large filament
5. Select Auto (init) : first time for the new tube
6. Select start
7. Select Yes
FVR auto calibration
FVR auto calibration
Tips:
During calibration, do not cancel auto cal in the middle of the procedure, it won’t save the mA
station that it finished calibrating.
Auto ( init) is used when the tube has never been calibrated
Auto (Regular) is used during preventive maintenance ( tube already calibrated in the past)
FVF Pulse Fluoro mA auto calibration
Software Revision V2.20
FVF mode:
This auto cal mode is independent of the Dar-8000i.
It can be turned OFF during this cal mode.
Auto cal is valid for Pulse fluoro mode only.
( DC fluoro is still in manual)
Procedure
1. Select XDR technique
2. Go to FVF mode
3. Select Pulse Fluoro mode (fluo2)
4. Input the Pulse Fluoro mA curve as recommended in the fluorospeed 300 or Sonial vision series install guide or image
quality manual
5. Select Auto (Simple)
6. Select the first FVF position at 50KV
7. Perform Pulse Fluoroscopy, the FVF starts adjusting automatically until it finds the target mA . After the target mA is
reached, X-Ray is stopped
8. Select the second FVF position at 60KV and repeat step 7. and repeat until 120KV position.