11/02/2010 SonialVision Versa Installation Guide

System Setup guide

SonialVision Versa
combined with CH-200

Revision
Rev D2010 :

SN
SonialVision Versa Installation Guide Page 2

Introduction

Sonialvision Versa is a combination of UD150B/V-40 X-ray generator, the ZS-100IR table,


the DAR-8000i Digital system, CH-200 (automatic) or CH-200M (manual) with wall stand BR
-120 or BR-120M (manual). Sometimes, Canon CXDI-50G is also combined to replace cas-
settes of the table bucky and wall bucky.

This steering guide is used to help FSE installing the Sonialvision Versa system. It can be
used as a guide to facilitate the installation process.

It also has the current useful information and tips that are sometimes not available in the
manuals.

Update: March 2010


1) B-40 V2.20 has FVR auto calibration and FVF auto calibration with pulse fluoro mA. FVF for DC fluoro
remains the same.
2) SA-60 High Speed Hold is removed. Use Low Speed Hold.
Page 3

System Software

This sheet is used to record system software revision information

S/W Name Rev Update Update Update Update

Dar-8000i
Ixworks
IAPDB
DDM
FPGA0
DAR FPGA1
FPGA2
Panel
GIM
GIM Remote
Side Station
UD-CONT (cabinet )
UD CONSOLE
B-40 DATA
NEX-SH
CLC-MAIN
ZS-100
Remote
Local
CH-200/M
BR-120/M
Others CXDI-50G
Page 4

System Serial numbers

This sheet is used to record system serial number of each component

Unit name S/N


DR system Dar-8000i

Generator UD150B-40
TUBE1 G292
TUBE2 P324DK-85
Starter SA-60
Tube Heat cooler HE-125 Option
Table ZS-100IR
Table Collimator R-50

CTM CH-200
CTM Collimator R-30H
Unit Unit name S/N

Monitor SMD 18101SCS


SMD 18101SCS
SMD 18101SCS Option
SMD 18101SCS Option
Page 5

Typical Room Layout


SonialVision Versa Installation Guide Page 6

System Options

Side Station
TomoSynthesis (Safire system only)
RSM-DSA
Slot Radiography (Safire system only)
Dual Energy (Safire system only)
Dicom Print
Dicom Send
MWL
MPPS
Page 7

B-40 Option Kits (Pre-installed from factory 2010)


The UD150B-40 is an 80kw X-Ray generator. It is shipped as barebones unit from the
factory. The option units for R/F system consist of:
Fluoro unit kit:
Fluoro 2002 pcb
mA power 2002 (#2)
MGF contactor, and Capacitor
Cables, spacers, screws
UD-DAR 8000 connection kit:
SGIM power supply
SGIM pcb
DAR I/F pcb
Cables, spacers, screws
Communication I/F kit:
Com I/F 2002 pcb
422 I/F 2002 pcb
Cables, spacers, screws
Photo timing kit (SPT-C42M) option 1 ( in case of system with Tandem lens)
PHOTO CONT 2002 pcb
PHOTO I/F 2002 pcb
PH power 2002 pcb
Cables, spacers, screws

Photo timing kit (SPT-C42) option 2 ( in case of system with Direct lens)
PHOTO CONT 2002 pcb
PHOTO I/F 2002 pcb
Cables, spacers, screws
Page 8

B-40 kit installation (Pre-installed from factory 2010)

1. Fluoroscopy unit kit: Refer to chapter 8 of B-40 installation manual for detail.
Set sw4-6= OFF on UD cont-2002 PCB (when combined with two mA power 2002 PCB).

2. UD-DAR 8000 connection kit:


Refer to UD-DAR 8000 connection kit installation manual M517-E397 for detail. The installa-
tion manual requests you to flash the GIM code. It was already flashed from the factory. The
generator software should be version 1.74 or later.

3. Communication I/F Kit:


c. Refer to chapter 13 of B-40 installation manual for detail.
d. For jumper setting on Com I/F PCB refer to Arcnet communication setting section.
Confirm that SW2-2=ON, the rest OFF

4. Photo Timing Kit:


f. Refer to chapter 6 of B-40 installation manual for detail
g. Set PH field Option= ALL ( select all field combination)
Set to AVERAGE mode in Photo Gain

5. FDA dosimeter kit:


i. Refer to Ext console installation manual M501-E375 for detail
Page 9

BH Filter
The Gold filter is not necessary to install. It has been
noticed in the field that the gold filter reduced the
image contrast. Always measure HVL to confirm that it
is 2.9mm at 80kv ( FDA )

II Photo cable Connection

In case of Direct type optic is used:


Connect Photo cell cable from optical distributer to PH /IF pcb

In case of Tandem type optic is used:


Connect PMT cable from optical distributer to PH Power 2002 pcb

TV signal Connection for IBS

Connect CN15 from SGIM PCB to TV7-TV2 on Fluoro 2002 PCB. In B-40 option page,
set DC fluoro to TV and Pulse fluoro to TV
Page 10

DAR-8000i Connection

Refer to DAR-8000i installation manual

Scan Converter Connection


Refer to DAR-8000i installation manual (when
connecting VCR or DVD recorder)

Monitors Connection
Refer to DAR-8000i installation manual

FDA Dosimeter connection


Refer to Ext Console installation manual M514-E068

CH-200 connection
Refer to CH-200 installation
Page 11

Camera head orientation

Refer to DAR-8000i image quality adjustment manual


SonialVision Versa Installation Guide Page 12

System Power Connection


Page 13

Arcnet Connection
Attention: CH-200 arcnet connection is not used!!!!! Make sure to configure it with B-30 gen-
erator not the B-40 generator.
SonialVision Versa Installation Guide Page 14

ZS-100IR and B-40 Connection


Page 15

II Cable Connection:
The II comes with extra cables that are not used. Connect them as follows:
Page 16

SA-60 Connection

The SA-60 comes prewired from the factory for the 1.5MHU x-ray tube. Therefore before
power up the SA-60, the proper wiring for each x-ray tube must be performed first. Otherwise,
x-ray tube will be damaged!!!!!!!

Step1: Primary input voltage


The primary input voltage should be between 208~240V. There is tap at the back of the SA-60
to match the input voltage.

Step2: Secondary output voltage for Each X-Ray tube


The secondary output voltage should be set according to each x-ray tube. The picture below
shows secondary wiring from the factory.
Page 17

SA-60 NEXSC pcb


SonialVision Versa Installation Guide Page 18

SA-60 Connection according to each x-ray tube

SA-60 Prgramming
SN 2007-08-07

START INITIALIZE CPU


SW2-8=ON
SW3-1=ON
POWER ON AND WAIT FOR 20 SEC. TURN SW3-1=OFF
AND PUSH RESET
YES
NEW
INSTALLATION

NO TURN ON OPTION MODE


SW2-2=ON
SETUP MODE (SW2-8= ON)

ENABLE LOW SPEED ROTATION HOLD


TUBE 1 OR NO SW3-3=ON
TUBE 2

YES PUSH RESET

TUBE 1 TUBE 2
SW2-5 = ON SW2-6=ON
SET HIGH SPEED HOLD FOR 3MIN
SW3-5,8=ON ( SEE NOTE)

TURN ON GENERAL MODE


SW2-4=ON
WRITE DATA TO MEMORY
SW2-1= ON
WAIT FOR 2 SEC.
PUSH RESET SW2-1=OFF

ANODE PUSH RESET


CODE

TURN OFF OPTION MODE


SW3= ALL OFF
ANODE CODE=3 ANODE CODE=9 SW2-8=ON, REST OFF
SW3=4,5 ON SW3=4,7 ON

DISPLAY MODE
STARTOR
SW2-2,3,4,=ON
TYPE R OR
Q

TUBE 1 OR
R STATOR Q STATOR 2
SW3-3=OFF SW3-3= ON

TUBE 1 TUBE 2
WRITE DATA TO MEMORY SW2-5=ON SW2-6=ON
SW2-1= ON
WAIT FOR 2 SEC.
SW2-1=OFF
PUSH RESET

CONFIRM YOUR PROGRAMING


TURN OFF GENERAL MODE
SW2-4=OFF, SW3= ALL OFF

SET SW-2=ALL OFF


SW3=ALL OFF
PUSH RESET
SonialVision Versa Installation Guide Page 20

SA-60 programming confirmation

After programming, always check the content if the programming has been performed cor-
rectly. Set the dipswitch sw2 and sw3 to display mode using the flow chart on previous page
and the table below as a guide.

Note: SA-60 shipping second quarter of 2009, will have a new CPU pcb (NEXSH) that allows
programming by PC instead of dip switches progamming !!
Page 21

Unused cable

A flat dark grey cable J82 is not used!!!!!!!

Before Power ON:

• The B-40 accepts 3 phases voltage ranging from 380 to 480V without using auto XAT-2H
transformer. Always match T1 and T2 transformers according to the line voltage.
Ex: if the line voltage is 380V, tap T1 and T2 to -200 and +180 to get a total of 380V.
• We recommend disconnecting X-ray high voltage portion when power up for the first
time during installation. T his allows you to check basic function without generating
high voltage to the inverter circuit of the generator and doesn’t heat filament to the X-Ray
tube. To do so:
-Disconnect J2 (MGL) on terminal 2002 pcb
-Disconnect JS1 cable on UD cont 2002 pcb
-Disconnect J2 on mA power 2002 pcb
-Set sw4-4= on on UD cont 2002 pcb

After Power ON:

After Power ON, always check to confirm that L0, L100 of the B-40 should be 100V (+/- 5V) AC.
The B-40 is always boot up in Japanese language during installation. You need to initialize it to
English language and replace 3 Japanese labels to English labels on the Console. Lots of
people can’t locate English labels: they are attached in a small bag of the B-40 console
SonialVision Versa Installation Guide Page 22

Load B-40 System Preset CD

The System Preset CD (which includes B-40 system configuration setting (initial data), APRs for
DR fluoro , Table Bucky APRs, Wall Bucky APRs and General Radiography APRs. Please load
the system preset data from CD to the B-40 in order to save a few days of setup. Refer to B-40
installation manual, chapter 9 section 9.22. Use command name: BKUP. Select 3 to upload ->
Select File -> transfer->Xmodem -> send. From Select File, the system has a time out of about 15
sec to do this process; otherwise it will time out and won’t upload. If it happen, you need to reset
the B-40, and redo the upload again

Program SA-60 High Speed Starter


The High Speed Starter must be programmed to set tube code, Rotor type, High Speed hold time
etc... Refer to the Appendix on the instruction how to program it. Note: the instruction in the SA-61
installation manual is at times difficult to understand. Attention: Before power up the High
Speed Starter, confirm that the High Speed voltage (HS), Normal Speed (NS), Brake (BS).
The X-Ray tube Rotor may get damaged if it is not set correctly!
Page 23

B-40 Setting

RATIO KV MA
MAXKV 150 534
MINKV 80 1000
MINRATING 80 KW

FREQUENCY 60

TUBE Tube Code


TUBE 1 40 G292(no cooling unit)
TUBE 2 30 P-324DK

Input Tube chart on the next page if G292 is cooling unit option

TECH
TECH NO TECH CODE NAME TUBE FLUO PH CODE PH DETECT PH FIELD TECHSYSTEM
0 01h GR 1 OFF 0 0 0 1
1 17h XDR 1 0 1 0 0 14
2 01h GR 2 OFF 0 0 0 1
3 02h BU1 1 OFF 5 1 3 4
4 05h BU2 2 OFF 5 2 3 3
5 04h PL 1 OFF 0 0 0 04
SonialVision Versa Installation Guide Page 24

G292 with Heat exchanger tube chart


Tube E1 Extra Tube Data Mode-1 Edit

Edit mode
Main Page: Tube Info
Anode High
Focus L&S
kV MAX 150
kV MIN 40
HU Factor (%) 90
Cine Factor
(%) 90
Page 2: Tube :HU

HU MAX 600 kHU


HU CURVE
kHU 600 420 240 150 0
1 2 3 4 5
sec 0 2 9 25 99

Page 3: Tube :Small Focus


Emission
kV 40 60 80 0 0
mA 255 350 420 0 0
Ratings
kV 125 125 125 125 125
0.1/
0.01/3 0.03/10 20 0.3/50 1 [sec]
125 125 125 125
mA 365 330 320 280 225
0.1/
0.01/3 0.03/10 20 0.3/50 1 [sec]
170 110 70 50
IF max 540
mA max 320

Page 4: Tube :Large Focus


Emission
kV 40 60 80 0 0
106
mA 630 880 0 0 0
Ratings
kV 125 125 125 125 125
0.1/
0.01/3 0.03/10 20 0.3/50 1 [sec]
125 125 125 125
mA 960 900 760 630 480
0.1/
0.01/3 0.03/10 20 0.3/50 1 [sec]
320 170 120 80
IF max 540
mA max 630
SonialVision Versa Installation Guide Page 25

B-40 Setting

FLUO
FLUO NO NAME FOCUS HLC PULSE FLUORO PULSE RATE PULSE WIDTH
0 DOSN 0 0 0
1 DOSL 0 0 0
2 PULSE N 0 0 1 30 10
3 PULSE B 0 0 1 15 16
4
5
6

COMMUNICATION
UD-COMM I/F
PORT 0 PORT 1
Not used Not used
VIA ARCNET
DAR-6000 MECHA 1 MECHA 2 CH-200 #1 CH-200 #2 RAD DR
used used Not used Not used Not used Not used
FPD GRID#1 FPD GRID#2 FDA CONSOLE
Not used Not used used
SonialVision Versa Installation Guide Page 26

B-40 Setting

OPTION B-40 ver 2.1 or later PAGE 1


NDD Not used YSF Not used
MEASURED Not used IRIS WIDE Not used
FCR NDD Not used ZS-5Y Not used
TOMO I/F Not used HLC SOUND Continious
PAGE 2
PHOTO BACKUP 0 PH PHIELD All
BA COMP Not used RCCD Not used
FSP II FIELD Max DEFAULT IBS On
SPOT ANODE High SEC UNIT Not used
PAGE 3
IDR 700 Not used CVS Not used
FLUO TIMER FDA X DENSITY RAD Not used
ION CHAMBER Not used X DENSITY FLUO Not used
SET FMA DISP Not used
PAGE 4
IBS DC FLUORO TV FOCUS SLIDE Not used
IBS Pulse FLUORO TV FLASH CALIB Not used
DEFAULT AEC On FPD SYSTEM Not used
DEFAULT RMA FEEDBACK On 0.5 SEC EXPOSURE Not used
PAGE 5
FLUORO TIME UNIT MinSec BG COLOR IN STEP Not used
CH-200 AUTO POSITION Not used AFTER X-RAY ILLUMI Not used
SERIAL LOCK Unlock READY UP SOUND ON On
TUBE COLOR IN PROTKEY Default AUTO READY UP Not used
PAGE 6
EXTENDED ICONS Not used EXTENDED SOUND Not used
EXTENDED TABS Not used EXTENDED MEMORY- Not used
ADJUST ILLUMINATION Not used ILLUMINATION STANDBY Tube Default
IBS SHIFT used (0) SET FCR STEP NO Not used
PAGE 7
IFV LIMIT CONTROL Not used DM_M4 FOR TUBE Not used
SEC SHUTTLE Default COMM LOCK BUTTON Not used
COMPRESSION CONE Not used NEW PULSE FLUORO Not used
FCR_DM_M4 Not used NOT RESET STEP SELECT Not used
Page 27

CH-200 Registration Setting

Go to CH-200 service mode by:


Setting sw2-4= on of Perip SH pcb

CH-200 system setting:

CH-200 layout = standard


Collimator type = R-30
CH-200 ceiling height = Normal
Generator type = B30 ( important)
Pin control option = OFF

Table system setting:

Table type= off

Stand system setting:

Table type= BR-120


Layout= long
Direction= left (or right depend on site)
Tracking option= on (or off)
Stand alignment position= center
Tilt option= off

Calibrate the potentiometers positions of the CH-200 according to CH-200


installation manual chapter 11
Page 28

CH-200 (BR-120 wall bucky with cassette) interface with B-40


• Do not connect J25 from CH-200 cabinet to B-40 Terminal pcb CN9
• B-40 Terminal pcb CN9 is used by ZS-100IR
• Set SW2=1,3,4,5,6 to ON and 2,7,8 to OFF on Terminal 2002 pcb
• Set SW1=2, 5 to ON and the rest to OFF on Terminal 2002 pcb
• Modify J25 cable connected in CH-200 cabinet: Cut B1 wire from J25. Connect B1 to cp9 on CHC-200
pcb.
• Add a jumper on J20 pin B11 to GND of CHC-200 pcb and set dip switch SW2=1 to ON of PERIP-SH
( Tube 2 selection)
• On the other end of J25 cable, cut the connector and use only 3 wires: B3, B6, A9 pins (B3: Green-1black
color ), (B6:orange-2black), (A9: grey-2 red)
Page 29

In Case Canon 50G is installed with the system

Go to CH-200 service mode by:


Setting sw2-4= on of Perip SH pcb

CH-200 system setting:

CH-200 layout = standard


Collimator type = R-30
CH-200 ceiling height = Normal
Generator type = B30 ( important)
Pin control option = OFF

Table system setting:

Table type= off

Stand system setting:

Table type= BR-120


Layout= long
Direction= left (or right depend on site)
Tracking option= on (or off)
Stand alignment position= center
Tilt option= off

Calibrate the potentiometers positions of the CH-200 according to CH-200


installation manual chapter 11, starting section 11.5.
For FPD caon DR tray registration, refer to manual M511-E344
Page 30

FPD Canon 50G DR Tray for BR-120 wall bucky Connection


Page 31

Canon 50G Interface with ZS-100IR for Table bucky Connection

Refer to

CANON FPD TRAY KIT


INSTALLATION MANUAL ZS-100I / ZS-100IR
CANON FPD TRAY KIT INSTALLATION MANUAL
Page 32

FPD Canon 50G DR Tray for BR-120 wall bucky Connection and Table Top
SonialVision Versa Installation Guide Page 33

FVR Calibration (v2.20 has FVR autocalibration. Refer to end of Doc):

Radiographic mA Calibration (FVR)


The Radiographic mA calibration is done in FVR mode. Select Measure Option: use ( turn back to NO use in User
mode operation) so that the Console will becomes a readout display during calibration. Set exposure time to 50
mS. You can connect the scope to TKV and TmA on UD cont 2002 pcb to observe KV and mA waveform if you need
to see the KV and RmA waveforms:
TKV, 1v= 20 KV
TmA, 1v=200mA for Radiography, 1v=1mA for DC fluoro and 1v=10mA for Pulse fluoro.
By default, the B-40 assign 6 mA stations for Large Filament and 6 mA stations for small Filament as shown below.
You can add more mA stations if requested by the customer. The feedback must be OFF during calibration.

Note: B-40 software v2.20 has new feature: FVR auto calibration. Select AUTO-INIT to start calibration.
Page 34

FVR Calibration: Tube 1

TUBE1: G292
FOCUS mA 40kV 60kV 80kV 100kV 125kV 150kV

S
100
125
160
200
250
320
FOCUS mA 40kV 60kV 80kV 100kV 125kV 150kV

100
L
200
250
320
400
500
630
SonialVision Versa Installation Guide Page 35

FVR Calibration: Tube 2

TUBE2: P324DK-85
FOCUS mA 40kV 60kV 80kV 100kV 125kV 150kV

S
100
125
160
200
250
320
FOCUS mA 40kV 60kV 80kV 100kV 125kV 150kV

100
L
200
250
320
400
500
630
Page 36

FVF Calibration: Tube 1

The DC fluoro/ Pulse fluoro mA calibration is done in FVF


mode. The B-40 has 3 fluoro modes:
Fluo 0= DOSN
Fluo 1= DOSL
Fluo 2= PULSE N
Fluo 3= PULSE B
.
Each Fluoro mode has its own FmA curve, which depends
on the X-Ray tube type. The screen below shows an
example of a FVF screen.
Page 37

FVF Calibration: Tube 1 (V2.20 has Pulse Fluoro mA autocalibration)

DOSN (Fluoro 0)
50kv 60kv 80kv 90kv 100kv 120kv
FmA 0.4 0.6 7.3 6.8 6.0 5.2
FVF
Calibrate FVF at each KV point (IBS = OFF) so that the FmA readout matches the Set FmA in the FmA curve table.

DOSL (Fluoro1)
50kv 60kv 80kv 90kv 100kv 120kv
FmA 0.2 0.3 3.65 3.4 3.0 2.6
FVF
Calibrate FVF at each KV point (IBS = OFF) so that the FmA readout matches the Set FmA in the FmA curve table

PULSE N (Fluoro 2) - Refer to the back of the document how to calibrate using auto calibration
50kv 60kv 80kv 90kv 100kv 120kv
Pulse mA P_P 6 12 30 24 24 24
Average Display 1.8 3.6 9 7.2 7.2 7.2
Value
FVF

Bariatric Only

PULSE B (Fluoro 3) -Refer to the back of the document how to calibrate using auto calibration
50kv 60kv 80kv 90kv 100kv 120kv
Amplitude of 6 12 30 24 24 24
FmA
Average Display 1.35 2.7 6.75 5.4 5.4 5.4
Value
FVF

Note: Always confirm that the skin dose never exceeds 10R/min.
Always measure to confirm that the Maximum skin dose never exceeds 10R/min.
The method to measure maximum skin dose is to install the probe at 30cm from table top and measure Dose at
Max KV.
SonialVision Versa Installation Guide Page 38

How to Rename Pulse Mode

How to change Pulse Fluoro mode name ex: Bone N to Pulse B

Syntax:

UD> fluoname [x] [Newname]

Where x= fluoro mode number such as 1,2,3 etc..

Ex:
Let’s change 4:BONE N to PULSE B ( B stands for Bariatric)
At the UD prompt:

UD>fluoname

*Display Fluoroscopy name


1: DOSN
2: DOSL
3: PULSE N
4: BONE N
..
..
..
..
8: user1

Then type

UD>fluoname 4 PULSE B

Reboot the generator to update.


Page 39

How to Calibrate Pulse Fluoro mA (V2.20 has Pulse Fluoro mA autocal)

How to calibrate Pulse fluoro:


At 50 KV, we adjust FmA to be equal to 2 mA
At 60 KV, we adjust FmA to be equal to 3 mA
At 80 KV, we adjust FmA to be equal to 12 mA
At 90 KV, we adjust FmA to be equal to 10 mA
At 100KV, we adjust FmA to be equal to 8 mA
At 115KV, we adjust FmA to be equal to 6 mA

There is three ways to adjust pulse fluoro mA


Method #1 is to use the oscilloscope on TmA waveform 1v= 10mA.
Method #2 is to use FVR to adjust Pulse Fluoro. Method 3 is AUTO CAL (Recommended).

Let’s explain how to do Method#2: Go to FVR->Select small filament->assign 2mA, 3mA, 6mA, 8mA, 10mA and 12mA
SonialVision Versa Installation Guide Page 40

How to Calibrate Pulse Fluoro mA continued

• At 2mA, calibrate the FVR at 40KV and 60KV. You don’t need to calibrate the whole line ( just 40kv and 60kv
for 2mA station) interpolate between 40KV and 60KV. I use the above example: (3.98+3.93)/2= 3.96. Write
3.955 on a paper
• At 3mA, Calibrate the FVR at 60KV point.. Write your calibrated result on a paper. In this example is it 3.96
• At 6mA, calibrate the FVR at 100KV and 125KV. Then interpolate the data between the two points.
And so on for 8mA, 10mA and 12mA.

Pulse Fluoro is a Pulse waveform (ON, OFF, ON etc..), the Generator can display only Average value of the pulse
waveform.

How the Average value is is calculated from a pulse wave form for 30 F/S.

Ex: at 12mA, pulse width= 6.3 mS (0.0063Sec)


(12*0.0063)*30 + Wave tail. Wave tail is negligible.
Page 41

How to Calibrate Pulse Fluoro mA continued


SonialVision Versa Installation Guide Page 42

How to Calibrate Pulse Fluoro mA continued

Bariatric option ( Bariatric only)


If Bariatric application is needed, you need to add one more fluoro mode called “Pulse B”. B stands for
Bariatric.

Go to service mode->setup->Fluo
On line 03, select edit and add the pulse fluoro mode as shown below. Then rename it from BONE N to
PULSE B.
SonialVision Versa Installation Guide Page 43

Maximum Skin Dose Measurement

Always measure the maximum skin dose so that it doesn’t go higher than 10R/min

SID=1100mm
Probe- =30cm from tabletop

IBS OFF IBS ON


PULSE N30 50KV 60KV 80KV 90KV 100KV 120KV 120KV
R/min

PULSE N15 50KV 60KV 80KV 90KV 100KV 120KV 120KV


R/min

PULSE L15 50KV 60KV 80KV 90KV 100KV 120KV 120KV


R/min

PULSE L7.5 50KV 60KV 80KV 90KV 100KV 120KV 120KV


R/min
SonialVision Versa Installation Guide Page 44

Signal setting FOR ZS-100 system only on B-40

Go to B-40 service mode -> signal and set the signal as follows.

UD-CONT I/F OUT-5 item 14 = 741 ( original= 526)


UD-CONT I/F OUT-5 item 12 = 542 ( original= 600)
FLUO I/F IN-2 item 12= 703 ( original= 37)
Set SW1-5=ON, ( all the rest OFF) on Terminal 2002 (B-40)
Connect XR1MA, XR1MB on Terminal 2002 (B-40)
Set SW2=1, 2,3,4,6 ON (all the rest OFF) on Terminal 2002 (B-40).
Note: If using Canon tray: set SW1=1, 2, 5 ON (all the rest OFF) on Termi-
nal 2002 (B-40). SW2=1,2,5:OFF SW2=3,4,6:ON

ZS-100IR configuration and position calibration


Refer to ZS-100IR installation manual M506-E327
Page 45

DAR-8000i System configuration

The configuration is done in DAR-8000i service mode. How to log in service mode:
Boot up the Dar-8000i and push the shift key and hold it until log in screen appears
Username: shimadzu
Password: Dar8k

Double click Service application. Perform system configuration according to your system
combination such as camera type, table type etc.. (Refer to DAR-8000i installation M517-
E380 starting section 5.1)
SonialVision Versa Installation Guide Page 46

DAR-8000i Network Info Collection


Refer to DAR-8000i installation M517-E380 starting section 5

The following information need to be obtained from the hospital IT department:

DAR-8000i
IP address:
Subnet Mask:
Default gateway:

Dicom Print
Name:
IP address:
AE Title:
Port Number:
Film Orientation:
Film Size:

Dicom Store (send)


Name:
IP address:
AE Title:
Port Number:

Dicom MWL
Name:
IP address:
AE Title:
Port Number:

DAR-8000i Dicom Print and Dicom SEND setting


Refer to DAR-8000i installation M517-E380 for step by step procedure

DAR-8000i MWL setting


Refer to DAR-8000i installation M517-E380 for step by step procedure
Page 47

How to add CR modality on DAR-8000i


By Default, the DAR-8000i has 2 modalities RF and XA. Some hospitals still use CR.
To enable CR modality, Run the utility located at the Desktop.
SonialVision Versa Installation Guide Page 48

Select unprocessed/ processed Dicom auto storage: select processed, update and Exit

Select Edit “Modalities Values for work list query. Change number to 3 and add CR. Update
and Exit
SonialVision Versa Installation Guide Page 49

HVL ( Haft Value Layer)

Always measure HVL to confirm that it is at least 2.9mm at 80kv


( FDA )

Setting Flash Value (Preset from factory)

UD>flashtbl w Select: 6
Setting Flash Value for pulse Fluoro
Flash Volume value [1] 20
Flash Volume value [2] 20
Flash Volume value [3] 20
Flash Volume value [4] 20
Flash Volume value [5] 20
Flash Volume value [6] 20
Flash Volume value [7] 20
Flash Volume value [8] 20
Flash Volume value [9] 20
Flash Volume value [10] 20
Flash Volume value [11] 20
Flash Volume value [12] 20
Flash Volume value [13] 20
Flash Volume value [14] 20
Flash Volume value [15] 20
Flash Volume value [16] 20
Flash Volume value [17] 20
Flash Volume value [18] 20
Flash Volume value [19] 0
Flash Volume value [20] 0
SonialVision Versa Installation Guide Page 50

Setting Delay for filament heating (Preset from factory)

Setting Delay for Filament Heating


UD> flashdelay w
Input 6
Input 20

Enabling Delay and Flash of Pulse Fluoro (preset from factory)

UD> flashopt 12
Page 51

Monitor Adjustment (SIEMENS LCD)

MONITOR ADJUSTMENT (LCD BY SIEMENS) SMD 18101SCS


Brighness 45
Contrast 60
Backlight 100
RGB Relation 1
SERVICE LEVEL 2 (Press up key once and down key twice within 1 sec while in
the main menu)
LUT 2
SonialVision Versa Installation Guide Page 52

TV imaging adjustment

Tips:
ο Always remove Aluminum shipping plate from II during installation.
When installing Mu Metal shield in front of the II, make sure that the Mu Metal shield doesn’t get in con-
tact with the II glass. when table is tilting, the MU metal shield may get in contact with the II which causes
the image to be fuzzy

Imaging set up condition:


• Set PKV/PDT
ο 50kv = 1000
ο 90kv = 1000
ο 100kv= 1000
110kv= 1000

• Short time correction release


Turn VR14 on PH-POWER 2002 fully CCW (ZS-100IR: Tandem type)

• Offset signal release


ZS-100IR: Tandem type= in standby, CP 18 on PH power 2002 should be 0V +/- 10mV. VR13 (OFFSET1)

• Minimum SID ( ZS-100IR FFD=1100mm)

• Grid in (it is not easy to remove the Grid from ZS-100IR)

Acrylic = 19 cm

• IBS Response
ο DOSN ( Fluo0) : up=50 down= 50
ο DOSL ( Fluo1) : up=50 down= 50
ο PULS (Fluo2) : up=25 down= 25
Page 53

Before Iris calibration, improve camera iris motor torque

DAR-8000i TV iris motor torque improvement to countermeasure unstable iris aperture

Symptoms:
IBS fluoroscopy balance point is not stabilized after DR radiography
IBS fluoroscopy balance point is not stabilized after changing I.I. field size.
DR radiography image brightness is not stabilized.
DR radiography image brightness after Pulse Fluoroscopy is different from the brightness after
DC fluoroscopy.
Procedure:
1)Power OFF the system.
2)Remove the cover of DAR-8000i Camera head. Confirm the potentiometer R19 on top board in the
Camera head as shown.
3)Adjust R19 to get approximately 80 ohm (from 80 ohm to 100 ohm) between pin No.1 and pin No.3
of R19. ( pin No.1 and pin No.2 are shorted ) (R19 value is max. 200 ohm 30 turns)

Note) If this value is too much, Iris may hunting, Fluoroscopy IBS KV value may hunting.
SonialVision Versa Installation Guide Page 54

Camera Iris motor torque improvement (continued)

4)Reinstall the cover of Camera head, and power ON.


5)Carry out Iris adjustment and ABS (IBS) adjustment of Image Quality Adjustment.
< Iris adjustment: All mode of Fluoroscopy and DR radiography >
6)Confirm that IBS balance KV and DR Radiography brightness stabilizes.
7)Confirm that IBS Fluoroscopy KV is not hunting when performing fluoroscopy for at least 5 sec.

Note:
When you execute DR Radiography Iris adjustment by [Manual Setting], you should make Radiography after
Fluoroscopy, and check window level.
When you execute Fluoroscopy Iris Adjustment by [Manual Setting], you should make Fluoroscopy after Radiog-
raphy, and check Fluoroscopy kV.
When you execute Pulse Fluoroscopy Iris Adjustment by using [System Manual] function, system does not set
value for all pulse rates at once. You should set same value as “Iris Control Level” of 30f/s at other pulse
rate by [Manual Setting]. When you use [Manual Setting] mode, after finishing one pulse rate setting you
should exit from the mode, and re-enter to the mode for next pulse rate setting.
Page 55

Fluoro IBS Adjustment

FLUORO IBS stabilization

II 16” 12” 9” 7.5”


DOSN 16” 75kv 77kv 79kv 81kv
PULS 16” 75kv 77kv 79kv 81kv
Note: The image Quality M517-E398 is set to 82kv, 83kv, 86kv, 90kv for DOSN and 85kv, 86kv, 88kv
and 90kv for PULS. These setting reduce patient thickness coverage. It means that Fkv will reach Max KV
(120KV) with large patient size and don’t have coverage for extra large patient size.

Refer to image Quality M517-E398 adjustment procedure.

Spot/Serial Dose Adjustment

SPOT DOSE (AEC= ON)


PHV-DR
KV, mA 16” 12” 9” 7.5”
85kv, 320mA 19ms 25ms 31ms 34ms
PHV-DR
Calibrate PHV-DR until exposure time is
the same as the above value
Spot/Serial Iris Adjustment

SPOT Radiography Iris (AEC= ON)


Set Window=48 and L=2250 by using the mouse

KV, mA 16” 12” 9” 7.5”


85kv, 320mA 2250 2250 2250 2250
IRIS (in DAR-8000I)
Calibrate the iris until you see white dots about 100 dots
SonialVision Versa Installation Guide Page 56

How to Distinguish between Digital Tomo and Conventional Tomo

Digital Tomo uses II with Dar-8000i to acquire an image where as conventional Tomo uses Cassette
or Canon DR to get an image.

Digital Tomo uses the same Technique as Spot/SerialTechnique (XDR) so it is not necessary to cre-
ate an extra technique. All we need is to add extra APR for digital Tomo under XDR technique.

Conventional Tomo Technique is (PL).


SonialVision Versa Installation Guide Page 57

Digital Tomo adjustment

Digital Tomo Iris (AEC= OFF)


Set Window=48 and L=2500 by using the mouse
Acrylic= 10 cm
Layer height= 5 cm
PHVDR set “1” at all FOV

KV, mA 16” 12” 9” 7.5”


60kv, 1mAs 2500 2500 2500 2500
IRIS (in DAR-8000I)

Calibrate the iris until you see white dots about 100 dots

Digital Tomo height adjustment

The height adjustment is done by connecting the PC to ZS-100IR cabinet . Refer to Z-100IR installation
manual
Page 58

Memoryshot Adjustment (F-T curve)

Memory shot is performed in the B-40 service mode. Set the acrylic according to the table below
and perform fluoro at least 3~5 sec until IBS is stabilized and then register the FKV. Repeat it on
each FOV and Each fluoro mode (fluo0, fluo1, fluo2).
At the highest thickness, set the kv to maximum (120kv).

Memory-
FOV 7cm 14cm 21cm 28cm
DOSN Max 120kv
Large 120kv
Middle 120kv
Small 120kv
DOSL Max 120kv
Large 120kv
Middle 120kv
Small 120kv
PULSE N Max 120kv
Large 120kv
Middle 120kv
Small 120kv
PULSE B Max 120kv
Large 120kv
Middle 120kv
Small 120kv
Fluoro KV, mA fluctuation countermeasure
IBS margin function to countermeasure IBS KV, mA fluctuation
Required Software Revision:
V1.86 and later

Installation Procedure

On the PC, Run Tera term program and set the baud rate: 9600, 8, odd, 2, 1, none
Connect the PC to UD-cont 2002 pcb using null modem cable ( 9 pins)
Enable “ibsmargin” function as described below:

1. To display current ibsmargin setting (by default, ibsmargin is disable)

UD>ibsmargin

*Display FkV of IBS margin parameter (DC Fluo.)[Disable].

1 2 3 4 5 6
FkV 50 70 80 90 100 115
Margin 54 54 50 50 50 50

*Display FkV of IBS margin parameter(Pulse Fluo.)[Disable].

1 2 3 4 5 6
FkV 50 70 80 90 100 115
Margin 54 54 50 50 50 50

2. To enable ibsmargin function on DC fluoro and Pulse fluoro

Enter “ibsmargin on d or p” command in order to enable ibsmargin function.


UD>ibsmargin on d

*IBS margin parameter(DC Fluo.) is Enable.

UD>
UD>ibsmargin on p

*IBS margin parameter(Pulse Fluo.) is Enable. Check


Status; “Enable”
3. After enable ibsmargin, confirm that ibsmargin function is enable And setting values of
Red characters
UD>ibsmargin

*Display FkV of IBS margin parameter(DC Fluo.)[Enable].

1 2 3 4 5 6
FkV 50 70 80 90 100 115
Margin 54 54 50 50 50 50

*Display FkV of IBS margin parameter(Pulse Fluo.)[Enable].

1 2 3 4 5 6
FkV 50 70 80 90 100 115
Margin 54 54 50 50 50 50
Page 60

FOC error countermeasure


Preheat adjustment to prevent intermittent FOC error (Fluoro over Current)

On ZS-100IR system the FmA curve is higher than YSF-300 system. To prevent the FOC error, perform the
following procedure:
Connect PC to UD-cont 2002 pcb with Teraterm and
UD> Null Modem cable (9600, 8, Odd, 2, None).
UD>flashdelay w At the UD prompt, type flashdelay w
IFV flash delay table description.
------------------------------------------------------
Table 1: Radiography Large focus.
Table 2: Radiography Small focus.
Table 3: N/A
Table 4: Fluoroscopy- Continuous.
Table 5: Fluoroscopy- Pulse Large focus.
Table 6: Fluoroscopy- Pulse Small focus.
Type 4 to select con-
Table 7: Fluoroscopy- Pulse(Large FmA) Large focus. tinuous fluoro
Table 8: Fluoroscopy- Pulse(Large FmA) Small focus.
------------------------------------------------------

[IFV Flash Delay Table Write]


IFV Flash Delay Table number = [1-8]: 4

[ Table 4: Fluoroscopy- Continuous.]

*Flash delay value: 10 Type 20


[ Table 4: Fluoroscopy- Continuous.]
*Flash delay value: 10 20
[ Table 4: Fluoroscopy- Continuous.]
*Flash delay value: 20
>> Results:

Table 4: Fluoroscopy- Continuous.


Type flashopt 2
*IFV Flash Delay 4 : 20

UD>
UD>flashopt 2
UD>
Confirm that DC FLUO
UD> DELAY [ON}. Disconnect
the PC and Reset the gen-
erator.
*Now Flash Opt Settings.

0x0001: DC FLUO FLASH [OFF]


0x0002: DC FLUO DELAY [ON]
0x0004: PULSE FLUO FLASH [OFF]
0x0008: PULSE FLUO DELAY [OFF]
UD>
Page 61

Example of Memory shot Typical system setup

EXAMPLE
MEMORY SHOT F-T R-T
Fluoro X-ray Condition Bar AVE SER Low KV AVE IVR AVE (S)
Barium AVE
(FkV) (S)(0) SER (S)(0) (+1) (S)(0)
Acrylic
Mid- mA/
Max Large Small kV mA/msec kV mA/msec kV mA/msec kV
Fluoro0=DOS dle msec
N 7cm 57 58 59 60 68 200/100 65 200/100 75 200/200 68 200/200
14cm 65 67 68 70 80 200/100 75 200/100 82 200/200 80 200/200
21cm 77 82 86 88 93 200/100 85 200/100 88 200/200 87 200/200
28cm 120 120 120 120 105 160/100 95 160/100 110 160/200 105 160/200
Fluoro X-ray Condition Barium XL (L) Low kV AVE (L) Barium AVE
IVR XL (L) (+1)
(FkV) (0) (0) (S)(0)
Acrylic
Mid-
Max Large Small kV mA/msec kV mA/msec kV mA/msec kV mA
dle
Fluoro1=DOSL
7cm 60 61 63 64 68 200/200 65 200/200 75 200/200 68 200/200
14cm 72 72 74 78 80 200/200 75 200/200 82 320/200 80 200/200
21cm 86 89 100 110 87 320/200 85 320/200 90 320/200 87 200/200
28cm 120 120 120 120 105 400/200 95 400/200 110 400/200 105 160/200
Fluoro X-ray Condition Bar XL SER (L) Low KV AVE Myelo AVE(S) Barium AVE
(FkV) (0) SER (L)(0) (+2) (S)(0)
Acrylic
Mid-
Max Large Small kV mA/msec kV mA/msec kV mA/msec kV mA
Fluoro2=PULS dle
EN 7cm 50 50 53 53 68 200/100 65 200/100 75 200/200 68 200/200
14cm 60 60 64 66 80 200/100 75 200/100 82 200/200 80 200/200
21cm 73 76 79 84 93 320/100 85 320/100 88 200/200 87 200/200
28cm 120 120 120 120 110 400/100 95 500/100 110 160/200 105 160/200
Fluoro X-ray Condition Low kV AVE (S) Myelo XL (L) Barium AVE
15 FPS
(FkV) (0) (+2) (S)(0)
Acrylic
Mid-
Max Large Small kV mA/msec kV mA/msec kV mA/msec kV mA
Fluoro3=PULS dle
EB 7cm 52 53 55 56 65 200/200 75 400/32 75 200/200 68 200/200
14cm 62 65 67 70 75 200/200 85 400/32 82 320/200 80 200/200
21cm 77 82 88 99 85 200/200 95 500/32 95 320/200 87 200/200
28cm 120 120 120 120 95 160/200 115 500/32 110 400/200 105 160/200
Page 62

Memory shot setup

MEMORY SHOT F-T R-T


Fluoro X-ray Condition Bar AVE SER Low KV AVE IVR AVE (S) Barium AVE (S)
(FkV) (S)(0) SER (S)(0) (+1) (0)
Acrylic
Mid-
Fluoro0= Max Large Small kV mA/msec kV mA/msec kV mA/msec kV mA/msec
dle
7cm
DOSN
14cm
21cm
28cm 120 120 120 120
Fluoro X-ray Condition Barium XL (L) Low kV AVE (L) Barium AVE (S)
IVR XL (L) (+1)
(FkV) (0) (0) (0)
Acrylic
Mid-
Fluoro1= Max Large Small kV mA/msec kV mA/msec kV mA/msec kV mA
dle
7cm
DOSL
14cm
21cm
28cm 120 120 120 120
Fluoro X-ray Condition Bar XL SER (L) Low KV AVE Myelo AVE(S) Barium AVE (S)
(FkV) (0) SER (L)(0) (+2) (0)
Acrylic
Mid-
Fluoro2= Max Large Small kV mA/msec kV mA/msec kV mA/msec kV mA
dle
7cm
PULSEN
14cm
21cm
28cm 120 120 120 120
Fluoro X-ray Condition Low kV AVE (S) Myelo XL (L) Barium AVE (S)
15 FPS
(FkV) (0) (+2) (0)
Acrylic
Mid-
Fluoro3= Max Large Small kV mA/msec kV mA/msec kV mA/msec kV mA
dle
7cm
PULSEB
14cm
21cm
28cm 120 120 120 120
FVR auto calibration
Software Revision V2.20

FVR mode:
The auto cal starts at lowest mA station with
80KV , 40ms

Procedure
1. Select GR technique
2. Go to FVR mode
3. Select tube # ( 1 or 2)
4. Select small or large filament
5. Select Auto (init) : first time for the new tube
6. Select start
7. Select Yes
FVR auto calibration
FVR auto calibration

It takes between 17~20min for each filament ( 6 mA stations)


About 40 min per tube.

Tips:
During calibration, do not cancel auto cal in the middle of the procedure, it won’t save the mA
station that it finished calibrating.

Auto ( init) is used when the tube has never been calibrated
Auto (Regular) is used during preventive maintenance ( tube already calibrated in the past)
FVF Pulse Fluoro mA auto calibration
Software Revision V2.20

FVF mode:
This auto cal mode is independent of the Dar-8000i.
It can be turned OFF during this cal mode.
Auto cal is valid for Pulse fluoro mode only.
( DC fluoro is still in manual)

Procedure
1. Select XDR technique
2. Go to FVF mode
3. Select Pulse Fluoro mode (fluo2)
4. Input the Pulse Fluoro mA curve as recommended in the fluorospeed 300 or Sonial vision series install guide or image
quality manual
5. Select Auto (Simple)
6. Select the first FVF position at 50KV
7. Perform Pulse Fluoroscopy, the FVF starts adjusting automatically until it finds the target mA . After the target mA is
reached, X-Ray is stopped
8. Select the second FVF position at 60KV and repeat step 7. and repeat until 120KV position.