Professional Documents
Culture Documents
Total Organic Carbon Analysis For Cleaning Validation in Pharmaceutical Manufacturing
Total Organic Carbon Analysis For Cleaning Validation in Pharmaceutical Manufacturing
Total Organic Carbon Analysis For Cleaning Validation in Pharmaceutical Manufacturing
2 December 1999
World Leader
in
Total Organic Carbon
Analysis
Model A-2000 Wide Range Total Organic Carbon Analyzer
INTRODUCTION METHODOLOGY
In the pharmaceutical industry, Good Manufacturing TOC analysis involves the oxidation of carbon and
Practices (GMP) require that the cleaning of drug the detection of the resulting carbon dioxide. A
manufacturing equipment be validated.1 Many number of different oxidation techniques exist,
different validation techniques can demonstrate that including photocatalytic oxidation, chemical
the manufacturing equipment is cleaned and oxidation, and high-temperature combustion. In this
essentially free from residual active drug substances study, an Anatel A-2000 Wide-Range TOC Analyzer,
and all cleaning agents. equipped with an autosampler, was used. The Anatel
A-2000 Wide-Range Analyzer measures TOC in
Common analytical techniques in the validation accordance with ASTM methods D 4779-88 and
process include HPLC, spectrophotometry (UV/Vis) D 4839-88. It measures TOC directly by adding
and TOC. HPLC and UV/Vis are classified as phosphoric acid to the water sample to reduce the pH
specific methods that identify and measure to approximately 2 to 3. At this low pH any inorganic
appropriate active and substances. TOC is classified carbon that is present is liberated as CO2 into a
as a non-specific method and is ideal for detecting nitrogen carrier gas and is directly measured by a
all carbon-containing compounds, including active non-dispersive infrared (NDIR) detector. Any
species, excipients, and cleaning agent(s). remaining carbon in the sample is assumed to be
TOC. A sodium persulfate oxidant is then added to
The disadvantage of specific methods, particularly the sample, and in the presence of UV radiation, the
HLPC, is that a new procedure must be developed for remaining carbon is oxidized to CO2. The amount of
every active drug substance that is manufactured. CO2 generated is then measured by the NDIR to
This development process can be very time determine the amount of TOC originally present in
consuming and tedious, plus important sampling the water.
issues also must be considered. In addition, HPLC
analyses must be performed in a relatively short time
period after sampling to avoid any chemical Key Words:
deterioration of the active substance. Finally, the Cleaning Validation
sensitivity of HPLC methods can be limited by the Total Organic Carbon
presence of degradation products.
Method Detection Limit
TOC analysis can be adapted to any drug compound Precision and Accuracy
or cleaning agent that contains carbon. The method Linearity
is sensitive to the ppb range and is less time Swab Recovery
consuming than HPLC or UV/Vis. USP TOC
methods are standard for Water for Injection and
Purified Water2, and simple modifications of these
methods can be used for cleaning validation.
Anatel Corporation
2200 Central Avenue, Boulder, CO 80301; Phone (303) 442-5533 Fax (303) 447-8365; E-mail: info@anatel.com; http://www.anatel.com
©Copyright 1999, Anatel Corporation. All Rights Reserved. Printed in USA.
Information, descriptions and specifications are subject to change without notice.
Application Note 12.99.2 December 1999
40 y = 0.0355x + 1.1983
2
LINEARITY R = 0.9787
30
20
between the measured compound concentration and 15
the TOC response of the analyzer. We evaluated four 10
y = 415.76x + 16.997
R2). Three representative examples of active 10000 2
R = 0.9982
substances were also tested for linearity: an excipient 7500
(sucrose), an antibiotic (vancomycin), and endotoxin. 5000
Results are shown in Figures 5-7. All three
2500
compounds demonstrated excellent linearity with
0
correlation coefficients (R2) ranging from 0.9996 to
0 5 10 15 20 25
0.9998.
Triton-X 100 Concentration (ppm)
7000
6000
5000
4000 12000
Measured TOC (ppb)
Anatel Corporation
2200 Central Avenue, Boulder, CO 80301; Phone (303) 442-5533 Fax (303) 447-8365; E-mail: info@anatel.com; http://www.anatel.com
©Copyright 1999, Anatel Corporation. All Rights Reserved. Printed in USA.
Information, descriptions and specifications are subject to change without notice.
Application Note 12.99.2 December 1999
8000
y = 0.9287x + 30.8
LOQ 151 ppb
6000 R2 = 0.9998
5000 Table 1: Calculated TOC averages from 10 blank
4000
3000
vials
2000
1000
0
PRECISION AND ACCURACY
0 2000 4000 6000 8000
To demonstrate the precision and accuracy for this
Endotoxin Concentration (ppb)
TOC method, a representative solution of CIP-100 as
1000 ppb, or 1 ppm as carbon, was analyzed
Figure 7: Linearity of endotoxin sequentially ten times. This carbon concentration was
chosen to evaluate these method parameters because,
in general, TOC residual limits are typically around
METHOD DETECTION LIMIT AND 1 ppm. Results are listed in Table 2. At this TOC
LIMIT OF QUANTITATION level, the precision was ± 1% and the accuracy was
± 5%.
We determined the method detection limit (MDL) by
measuring the TOC response of the method blank. Vial Number Measured TOC (ppb)
A method blank consists of the sampling vial, swab, 1 1041
and recovery solution. In this study the recovery
1 1025
solution was low TOC (< 25 ppb) water. 10 pre-
1 1039
cleaned vials were filled with of the low TOC water.
One swab was placed in each vial (Texwipe Alpha 1 1057
Swab TX761; tips cut off). Solutions were vortexed 1 1054
and allowed to stand for one hour prior to analysis. 2 1034
Four replicates from each vial were analyzed. The 2 1042
four replicates from each of the ten blank vials were 2 1048
averaged. These 10 values were averaged again and a 2 1054
standard deviation was calculated. The standard 2 1055
deviation was multiplied by the Student t number for Average 1045
n-1 degrees of freedom (3.25 for n=10), at 99% Standard Deviation 10.5
confidence levels to determine the method detection % CV (precision) 1.0%
limit. The MDL was calculated to be 50 ppb. The % recovery based on 105%
LOQ was calculated by multiplying the MDL by 3. A 1 ppm C (accuracy)
value of 150 ppb was obtained ( see Table 1).
Table 2: Calculated accuracy and precision from ten
replicates of a 1 ppm CIP-100 solution as carbon
Anatel Corporation
2200 Central Avenue, Boulder, CO 80301; Phone (303) 442-5533 Fax (303) 447-8365; E-mail: info@anatel.com; http://www.anatel.com
©Copyright 1999, Anatel Corporation. All Rights Reserved. Printed in USA.
Information, descriptions and specifications are subject to change without notice.
Application Note 12.99.2 December 1999
SWAB RECOVERY
Stainless steel plates were used in the swab recovery test to simulate manufacturing equipment. One side of each
plate was spiked with a solution of active substance or cleaning agent. The plates were allowed to dry completely
overnight at room temperature. A Texwipe alpha swab TX761 was moistened with low TOC (< 25 ppb) water and
the spiked plate surface was swabbed both vertically and horizontally. The swab end was cut off, placed into a vial
to which we added 40-mL of low TOC water. The vial was capped tight, vortexed, and allowed to stand for one hour
prior to analysis. The same volume of each solution that was spiked onto the plates was separately spiked directly
into 40-mL of low TOC water and analyzed. The percent recoveries of the different substances are listed in Table 3.
Reported values are the average of three individual swab samples for each substance. The swab recoveries varied
between 79.3% to 95.9%
CONCLUSIONS
This study demonstrates that the Anatel A-2000 Wide-Range TOC Analyzer is suitable for measuring organic
residues on stainless steel surfaces, and that it is a reliable tool for cleaning validation as demonstrated by surface
residue recoveries of 79%-96%. The Anatel TOC analyzer offers low limits of detection, excellent linearity,
precision, and accuracy. All of these TOC results, with the exception of Alconox and Triton-X 100, were generated
using the same TOC method, making this TOC technology a low cost and less time consuming alternative for
cleaning validation.
REFERENCES
1. Food and Drug Administration’s Current Good Manufacturing Practice Regulations; 21 CFR 211.220
Anatel Corporation
2200 Central Avenue, Boulder, CO 80301; Phone (303) 442-5533 Fax (303) 447-8365; E-mail: info@anatel.com; http://www.anatel.com
©Copyright 1999, Anatel Corporation. All Rights Reserved. Printed in USA.
Information, descriptions and specifications are subject to change without notice.