(Can Help With Electrode Selection) Impedance - and - Noise - of - Passive - and - Active - Dry - EEG - Electrodes - A - Review
(Can Help With Electrode Selection) Impedance - and - Noise - of - Passive - and - Active - Dry - EEG - Electrodes - A - Review
wet electrodes, and consequently they should have a higher to as 1/ f noise or pink noise, is dominant at lower frequen-
Z in,amp for clinical applications. cies. Although flicker noise is observed in most materials,
Due to rapid advances in the field, the impedance of the precise source is still unknown. Fortunately, the chopping
non-invasive dry contact electrodes is approximately limited technique helps to eliminate this noise in amplifiers with the
to that of the upper-most layer of the skin, which is approxi- help of modulation. The flicker-noise voltage power in a single
mately 1 M. This impedance is lower for invasive dry contact MOSFETs (metal oxide semiconductor field effect transistors)
electrodes, as they penetrate the stratum corneum to make a is modeled as in equation (3) [50]:
low-impedance contact. In many cases, invasive dry contact K
electrodes have a similar, or even lower, impedance than do Vn,2 f = (3)
Cox W L f
wet electrodes. Nevertheless, due to advances in dry contact
electrodes, they are compatible with chopper amplifiers, with where W and L are the width and length of the channel
boosted input impedance, and insulated electrodes still exhibit of the MOSFET, respectively; Cox the capacitor between
high impedance in the EEG bandwidth. This is why circuit gate and substrate; K the process-dependent constant; and f
designers should utilize other techniques to use non-contact the frequency. Although this is the flicker noise of a single
and insulated electrodes, as these electrodes have an approx- MOSFET, an increase in noise density can be observed at
imate impedance above 1 G at 1 Hz, whereas the nominal lower frequencies for a single small metal electrode or one
input impedance of EEG chopper amplifiers is only a few of another material, proving the existence of flicker noise in
hundred mega-ohms. other materials [29]. Finally, it is worth noting that bipolar
When Z 1 is very high, high-impedance nodes behave like and junction gate field-effect transistor (JFET) transistors have
an antenna and absorb environmental noise, especially at a lower flicker noise [51], [52]. As MOSFETs are much
50/60 Hz. Designers thus employ inherent buffers to reduce better for integration and the cost of production is less,
the length of wiring nodes with high impedance. The buffer circuit designers still use MOSFET, instead of bipolar or JFET
is used to take a signal with high impedance and transmit it transistors.
at the same value, but with a lower impedance, at the cost The thermal or white noise voltage power in a MOSFET
of power consumption. In other words, they consume power gate can be calculated according to equation (4). In this
solely to convert impedance [40]–[43]. In some cases, instead equation, c is approximately 2/3 for a strong inversion bias,
of using a buffer, circuit designers utilize amplifiers. As a k is the Boltzmann constant, gm the transconductance of
result, they have not only an input signal with lower impedance the MOSFET, and T refers to room temperature in Kelvin.
at the output of the amplifier, but it is also of much higher In addition, the thermal noise of a single resistor can be
amplitude. Consequently, they achieve a higher SNR, if the calculated according to equation (5), where R is the value of
dominant noise doesn’t come from the electrode itself. The the resistor. Suppose that the impedance of a dry electrode
first low-power active electrode application-specific integrated is limited to the epidermis layer, which is approximately
circuit (ASIC) system for EEG monitoring was designed 1 M. In this case, just the thermal noise of this tissue is
in 2011 [44]. Although some active electrodes were already 125 nV. If the desired EEG bandwidth is up to 200 Hz,
available, they used off-the-shelf components. Furthermore, the total input-referred noise will be approximately 1.7 μVrms
a low output impedance amplifier helps to suppress cable if this impedance is constant throughout this bandwidth. This
artifacts and eliminate the need of using a shield for wires [45]. would be just the thermal noise in this bandwidth, without
Overall, the performance of active electrodes is highly any other noises of the electrode, if the electrode impedance
superior to that of passive dry electrodes, but at the expense on the skin is dominated by the resistance of the epidermis.
of increased power consumption. ADS1299 is one of the However, if the contact between the electrode and skin is
conventional amplifiers used to implement an EEG ampli- good, the capacitive behavior of the stratum corneum becomes
fier and can also be used to measure the impedance of an dominant and the impedance value decreases as the frequency
electrode [46], [47]. increases.
4kT c
III. N OISE OF VARIOUS M ATERIALS AND D EVICES v n2 M,t h = (4)
gm
Many different noise sources have been reported in the v n2 R,t h = 4kT R (5)
literature for wet and dry electrodes. Unfortunately, dry elec-
trodes are more susceptible to noise due to their weak contact, A simplified electrical model for an active electrode with
which leads to higher impedance. Such high impedance is the noise sources is depicted in Fig. 6. In this figure, Z c is
consequence of the increment of Z se , while Z s and Z e remain the active shield-to-electrode capacitance, and thus it can
constant. Among the numerous sources, the most important are be modeled as a capacitor. Furthermore, v n and i n are
motion artifacts, thermal artifacts, flicker, line noise, and the the input-referred noise of the amplifier and the total cur-
half-cell effect at the skin-to- electrode interface [29], [48]. rent modulated noise in the node v i , respectively. The total
Generally, the accepted noise for electrodes is less than input-referred noise has two different sources: the thermal
2 μVrms in the 1-100 Hz range when it is immersed in saline noise of the component and the noise of amplifier, as shown
solution [49]. in equation 6. The effect of the noise of the amplifier can
Thermal and flicker noise are two well-known noise sources be calculated as in equation 7, whereas the input-referred
in electrodes and/or amplifiers. Flicker noise, which is referred thermal noise is calculated according to equation 8 at the input
HABIBZADEH TONEKABONY SHAD et al.: IMPEDANCE AND NOISE OF PASSIVE AND ACTIVE DRY EEG ELECTRODES: A REVIEW 14569
TABLE I
C OMPARISON B ETWEEN Z1 , Zse AND Ze OF VARIOUS T YPES OF E LECTRODES
We survey several dry electrodes in the following para- on these sites, it is not a good choice for patients with dense
graphs. They were chosen based on their impedance values hair, due to the absence of pins to penetrate hair, and this
(Z s , Z se and Z e ), fundamental material, coating material, article contains no information about the dimension of the
size, and shape. Their mechanical properties, for example electrode.
stretchability and flexibility, are also investigated. Further- Since moisture or water can reduce the impedance of the
more, several electrodes that can be fabricated by 3D printers epidermis, a self-wetting paper electrode was proposed in [64].
are discussed because their mode of production makes them The water content in the corneum acts as an electrolyte,
widely accessible. as it contains ions, and helps to increase the effective contact
1) Non-Invasive Dry Electrodes: The reverse-curve-arch- area between the electrode and the skin. They used a porous
shaped dry electrode is one of the shapes proposed in [62]. fiber, which can absorb the moisture of the dermal layer. The
It is made of sterling silver and a 3D printer was used in porous fiber behaves like a reservoir when there is a barrier
its manufacture. There are five arches of 10 mm2 , with a on top. This novel approach helped them to achieve a SNR
thickness of 2.5 mm. The Z se of this electrode for hairy similar to that of wet electrodes. EEG measurements while
sites and bare skin are just below 300 k and approximately it was attached at the FP1 position showed the mean mea-
100 k, respectively. sured impedance to be 240 ± 51 k, whereas the measured
In [63], a flexible polymer-based dry electrode was pro- impedance for the wet electrode was 110 ± 52 k. The size
duced for high user comfort. Moreover, 44% carbon was used of the dry EEG electrode was not reported in the article.
to manufacture this active dry EEG electrode. The length, This electrode is not suitable for hairy sites, like other dry
thickness, and density of the pins to penetrate hair are critical. electrodes without bristles or pins. The Z se of this electrode,
Thus various structures were tested to find the best model. after dropping a small amount of saline to wet the surface, was
The final circular active EEG electrode is 5 mm in length approximately 60 k, and there were only negligible changes
and 13 mm in diameter. Despite all of these considerations, in this impedance during a 48-h experiment.
the final electrode has an impedance that is 10 times higher In [65], a dry polymer foam-based electrode for EEG
than that of conventional wet electrodes. The reported Z 1 measurements was introduced. It was made of an electrically
of the electrode is 1 M.cm2 at 10 Hz, while its Z e is conductive polymer foam covered with a conductive fabric.
approximately 10 .cm2 . The electrode has the same Z se impedance on the forehead and
In [57], a 3D-printed dry electrode was proposed with superior impedance at hairy sites. Moreover, skin preparation
different length pins, to accommodate people with different was not needed, whereas the reported wet electrode impedance
hair densities. They did not report the Z se in their paper, but in this article required skin preparation. The reported Z se was
their Z e was comparable to that of wet electrodes. All of the approximately 10 k on the forehead and 20 k at 1 Hz.
samples showed an impedance of less than 4 k at 10 Hz. Furthermore, the proposed electrode is relatively inexpensive
Their experiments show that their electrodes could be used and was able to adapt to an irregular scalp surface.
for BCIs, which do not require a high SNR. Finally, many studies have tested coated pin-shape dry
Another type of dry electrode is the bristle dry electrode, electrodes. However, a manufactured pin-shaped dry electrode
proposed in [59]. They are comprised of a silver-coated poly- may not possess good electrical properties. Consequently,
mer that is flexible, low-cost, and low-impedance, resulting the authors of these studies coated them with conductive
in a passive electrode with 10 mm long bristles. Due to their materials, such as silver, tin, stainless steel, or gold-plated
long bristles, they can easily penetrate dense hairy sites and silver [66]. Although this technique may allow the production
establish good contact with the skin. This helped them to of inexpensive dry electrodes, they may lose their properties
achieve a Z se of approximately 80 k, while the dimension after long-term use. Thus, their performance must be checked
was 12 mm × 12 mm. regularly over the long term. Several examples of non-invasive
A dry electrode based on a conductive stretchable Ag dry electrodes are shown in Fig. 7.
nanowire/polydimethylsiloxane (NWs/PDMS) composite mat- 2) Invasive Dry Electrodes: As previously discussed, the
erial has also been produced by 3D printing [60]. This dry main bottleneck of high-impedance dry electrodes is related
electrode was utilized as an active electrode and achieves to the epidermis layers, especially the stratum corneum. Thus
very good results in experiments on steady-state visual evoked some research has focused on penetrating this layer with
potentials (SSVEP). The measured Z se was approximately micro- or nano-needles, which results in a large reduction in
10 k at 10 Hz for F3 and Fp3 sites according to the measured impedance. Thus, they show higher signal quality
international 10 - 20 system. In spite of its good performance than conventional dry electrodes, which absorb the signal from
HABIBZADEH TONEKABONY SHAD et al.: IMPEDANCE AND NOISE OF PASSIVE AND ACTIVE DRY EEG ELECTRODES: A REVIEW 14571
Fig. 7. Examples of non-invasive dry electrodes: A) 3D-printed electrode coated with silver paint [57], B) bristle dry electrode [59], C) Ag NWs/PDMS
flexible dry electrode [60], D) g.SAHARASYS [61], E) reverse-curve-arch-shaped dry electrode [62], and F) micro-needle arrays dry electrode [63].
the top of the stratum corneum. Consequently, these electrodes approximately 1-1.5 mm. The diameter of the circular sub-
perform much better, and are sometimes called non-invasive, strate was 10 mm. The impedance of each pillar was approx-
as the user does not experience any pain. Nevertheless, they are imately 7.5 k at 40 Hz, while there were many pillars in
somewhat invasive. Indeed, to avoid pain, the micro-needles each electrode. The final Z e was less than 8 . Its measured
should not reach the dermis. Thus, it is best that their pene- impedance (Z 1 ) on the skin was less than 100 k, which is
tration does not exceed 200 μm [67]. However, for reasons of comparable to that of wet electrodes.
safety, the penetration length should ideally be decreased to Several articles have reported that CNTs may be toxic in
70 μm, as the thickness of the epidermis layer reported in the some cases. Furthermore, they cannot be grown beyond a cer-
medical literature is just 70 μm [35]. tain height due to their mechanical properties. Consequently,
Micro-needle arrays (MNAs) show low impedance and good they cannot sufficiently penetrate dense hair to make a good
contact with the skin. They can be categorized into two major contact with the scalp. Given the high potential risk of CNTs,
sub-categories. The first is stiff micro-needles, with a stiff but their good properties, there is an extensive literature in
substrate, such as silicon [63], titanium/titanium nitride [68], this field that has focused on various parameters that affect
or brass [69]. They do not generally make a good contact with the toxicity of CNTs, including size, length, agglomeration,
the skin, and thus need high pressure to achieve acceptable and impurities [74]–[78].
contact between the skin and the electrode. Thus, users will In [34], a 100-200 μm tall micro-tip was used, which was
likely not feel comfortable during the experiment. The second coated with IrO to improve its contact with skin. The width
sub-category consists of dry electrodes with flexible polymer of the micro-tip was 150-200 μm, which easily penetrates the
micro-needles and substrate, such as polydimethylsiloxane stratum corneum. The dimensions of the square electrode were
(PDMS) [70] or SU-8 [71] to increase user comfort. However, 8 mm × 8 mm. The Z e of this electrode was approximately
although patients experience more comfort during the mea- 100 .cm2 . Moreover, it achieved the same performance as
surements, the skin-to-electrode impedance is higher, as their Ag/AgCl. However, below a frequency of 3 Hz, its electrode
nonrigid micro-needles cannot easily penetrate the stratum performance was more capacitive than Ag/AgCl.
corneum. Therefore, a dry electrode with stiff needles and a In 2009, an innovative dry electrode was proposed that
flexible substrate constitutes a good option. had micro-spike electrodes, consisting of a micro-pillar with
After penetration into the skin, the material that is used a micro-tip on top. The micro-pillar was used to penetrate
on the outer layer of the electrode plays the most important the hair, while the micro-tips were designed to penetrate the
role in the final impedance. Various materials have been stratum corneum [79]. The final version of this dry electrode
investigated to identify the best quality of the recorded sig- had a Z se of 7-25 k. In [67], the performance of an electrode
nal, including platinum, PEDOT/PSS [72], sintered Ag/AgCl, with the same structure for hairy sites was tested and good
disposable Ag/AgCl, silver, gold-plated silver, stainless steel, results were achieved. In addition, they recorded EEG signals
tin [57], [66], and iridium oxide [34]. for open and closed eyes without any skin preparation or
Multi-wall carbon nanotubes (MWCNTs), which have con- conductive gel. In 2018 [69], an ERP (event-related potential)
siderably higher conductivity, were used in [49]. Indeed, was successfully recorded, which included P300, for unusual
MWCNT of 10-15 μm in length have been used to penetrate tasks from hairy sites.
the stratum corneum. Although the impedance of this electrode A motion interference-insensitive flexible dry electrode was
is not reported in the article, it is expected to be quite low. produced in [55]. It obtained a performance similar to that
A vertical patterned carbon nanotube was proposed in [73]. of wet electrodes, achieving a normalized impedance on skin
MWCNTs were grown vertically on a 2-mm thick stain- half of that of wet electrodes. Its normalized Z 1 was 61.2 ±
less steel foil substrate. The height of each pillar was 31.3 .cm2 for a diameter of 1.2 cm.
14572 IEEE SENSORS JOURNAL, VOL. 20, NO. 24, DECEMBER 15, 2020
TABLE II
I NPUT-R EFERRED N OISE AND I NPUT I MPEDANCE OF EEG A MPLIFIERS
Despite the use of such devices, weak contacts between VI. P ERFORMANCE C OMPARISON
electrodes and skin are still a problem. A low-impedance, Fundamentally, there is a trade-off between noise, user
skin-grabbing dry electrode was proposed in [90], [91] to comfort, movement artifacts and sensor complexity. Generally,
solve this problem. In this technique, numerous micro-teeth dry electrode signals show a slight amount signal degradation
are used to penetrate the epidermis to achieve a strong con- in comparison with wet electrodes and due to the weak contact,
nection between electrode and skin. Through this approach, it’s more susceptible to artefacts [94].
the electrode has comparatively lower sensitivity to motion According to measurements, the noise of dry electrodes
artifacts. follows the equation 8 and they can be equal to wet elec-
As decreasing impedance leads to reducing input-referred trodes [29]. Due to huge progress in dry electrodes technology,
noise, all techniques which reduce impedance will help they comply with the needs of clinical EEG applications
to reduce noise in the electrode. The acceptable electrode and higher user comfort is achieved for dry electrodes in
impedance for EEG recording should be kept to less than comparison with wet electrodes [95]–[97]. This shows that
5 k [39]. If we assume that the electrode impedance is 5 k, most problems can be solved with careful circuit design [23].
then the Z 1 will be dominated by skin impedance when Finally, for prolonged use, the signal quality of wet elec-
it is wet, and it is approximately 100 k. Furthermore, trodes tend to degrade whereas the signal quality of dry
the thermal noise will be 40 nV, according to equation (5). electrodes will not degrade, and it even gets better for a period.
The total noise of such an electrode in the bandwidth of As the skin under the electrode starts to sweat, the contact
0.5−100 Hz is 400 nV. According to the measurements, flicker between electrode and skin will be improved [98]. This is
noise is dominant at lower frequencies, which leads to higher good for long term recording whereas it’s a disadvantage of
noise in the desired bandwidth. Therefore, in this bandwidth, dry electrodes in time constrained laboratory applications.
the input-referred noise without calculation of the noise of the
amplifier will be greater than 400 nV. VII. G UIDELINE TO R EDUCE N OISE AND I MPEDANCE OF
E LECTRODES AND C OMPATIBLE A MPLIFIERS
B. Amplifier Noise-Reduction Techniques We have introduced various dry-electrode structures.
Although we have focused more on electrodes than ampli- Although pin-shaped and MNA dry electrodes are the most
fiers, the noise and input impedance of amplifiers play a common structures for hairy sites, several flexible and flat
significant role in the design of dry electrodes, while their electrodes have also been investigated due to their good
input impedance can limit the performance of the electrode impedance at other sites. Flat dry electrodes are not a good
and their high noise may lead to loss of brain signal. Therefore, choice for use on hairy sites because of their weak contact
we discuss the required noise properties and those of state-of- with skin and their inability to penetrate hairy sites. Since the
the-art amplifiers in the following paragraphs. reference electrodes are generally used on the ear, flat dry
The International Electrotechnical Commission (IEC) stan- electrodes can be utilized as a good reference.
dared IEC80601 defines a series of technical standards for There are many dry electrodes that are manufactured with
the safety and effectiveness of medical electrical equipment. 3D printers, offering the optimal solution for home-made
According to the International Federation of Clinical Neuro- electrodes in the near future. Thus, they could be a very inex-
physiology (IFCN), the total input-referred noise from 0.5 to pensive solution to manufacture the infrastructure. They can
100 Hz should be less than 0.5 μVrms or 1.5 μVpp [39]. be coated with other materials to increase their conductivity
In addition, according to IEC80601-2-26, the input-referred or manufactured with 3D printers with conductive material at
noise per channel of an amplifier should be less than 6 μVpp the cost of an increase in the price of the product.
or 2.1 μVrms in the bandwidth from 0.5 to 50 Hz [92]. Therefore, pin-shaped dry electrodes are the best choice for
The lowest noise can be achieved using the chopping a non-invasive model. By contrast, MNA electrodes exhibit
technique with the same power and area consumption. Accord- comparatively lower impedance, but they require special con-
ing to Table II, most research has achieved noise lower sideration, as they come into contact with skin and are intended
than 2 μVrms from 0.5 to 100 Hz. For example, in [93], to penetrate the first layer. A flexible substrate is preferable
the authors achieved 0.44 μVrms from 0.5 to 250 Hz, while to maximize user comfort. In addition, the MNAs or pins
their input impedance was 1 G at 60 Hz. Their total power should be sufficiently rigid to penetrate the stratum corneum
consumption was 5.4 μW. Although most state-of-the-art and/or hair. By combining these structures, a low-impedance
amplifiers have slightly higher noise in the bandwidth of inter- dry electrode with high user comfort can be achieved, making
est, according to IFCN standards, it is achievable with higher it suitable for long-term monitoring.
power consumption solely in the first stage amplifier if it limits Until now, invasive dry electrodes have not been very
the performance of the amplifier. Therefore, the bottleneck in popular because their impedance is generally limited by skin
designing an EEG amplifier suitable for dry electrodes is the impedance. This highlights the pressing need for high input
input impedance of the amplifier. impedance amplifiers. We propose three EEG amplifiers that
Finally, important parameters of the amplifier are not limited could be compatible with invasive, non-invasive, and insulated
to noise and input impedance. The common mode rejection dry electrodes that comply with two international standards in
ratio, power supply rejection ratio, tolerable offset, area con- Table III. Their features are based on electrode impedance on
sumption, and power consumption also affect the performance the skin of (Z 1 ), their sensitivity to power line noise, and their
of an EEG system. meeting international standards.
14574 IEEE SENSORS JOURNAL, VOL. 20, NO. 24, DECEMBER 15, 2020
TABLE III
S TANDARDS AND G UIDELINE TO D ESIGN A C OMPATIBLE A MPLIFIER FOR D RY E LECTRODES
Although the amplitude of input signals is less than 1 mVpp, For an amplifier with 40 dB differential gain, it means that it
artifacts may saturate the output and limit the performance attenuates 600 mV common-mode signal to 6 mV which is
of the amplifier. Therefore, a higher input range is proposed. acceptable value for most EEG applications.
Input referred noise is based on the required specification of
Add
the application. It can be a little higher or lower, depending on CM RR = = Add (d B) − Adc (d B) (9)
the sensitivity of the final application to noise. Since insulated Adc
electrodes are capacitively-coupled to amplifiers, there is no According to equation 8, the input-referred noise of insu-
concern about the offset. Acceptable attenuation according lated electrodes due to a very large g1 is not dependent on C1 .
to equation 1 defines the input impedance of the proposed Therefore, increasing C1 simply decreases the impedance of
amplifier. Power consumption is proposed according to that the electrode (Z 1 ) and thus power-line noise will be reduced.
of state-of-the-art amplifiers. In addition, a 100-G input impedance amplifier using the
A 0.05 Hz cut-off frequency leads to approximately a 20-s chopping technique has not yet been reported. The proposed
recovery time when there is an overload or artifacts [29]. chopper amplifier in [100] has an input impedance of 1.6 G
Dry contact and non-contact electrodes have a weaker contact at 20 Hz, which is among the highest reported with this
with skin than wet electrodes. Consequently, they are more structure, which is still much smaller than insulated electrodes.
susceptible to artifacts. As a result, a 0.5 Hz cut-off frequency Therefore, an insulated dry electrode with smaller Z 1 is
is proposed. This frequency can be tuned according to the another way to fulfill the requirements. Thus, it would be
purpose of the application. If the signal below this frequency compatible with chopping amplifiers, which leads to lower
is important, this frequency should be reduced and then an noise and substantially greater user comfort.
additional circuit will be required to improve the recovery
time.
The CMRR of an amplifier is equal to equation 9 where VIII. D ISCUSSION
Add and Adc are differential and common-mode gain of the The performance of dry electrodes is not limited solely
amplifier [99]. The High impedance node behaves like an by their impedance and noise. Indeed, offset, drift, phase,
antenna and it absorbs environmental noise [21]. Although this cost of production, bio-compatibility, and user comfort must
problem can be mitigated by utilizing a low-output-impedance also receive serious consideration. In addition, one of the
active electrode, if there is a mismatch between electrodes, disadvantages of dry electrodes is their sensitivity to motion
this common-mode interference and motion artifact will be artifacts, especially if they have weak contact with skin.
converted to a differential signal. Eventually, the amplitude Although techniques which penetrate the stratum corneum
of this signal may be much larger than that of the EEG may help to reduce this effect, it remains a major challenge
signal. This problem is more critical when EEG recording and in the progress of dry electrodes. EEG head caps have helped
brain stimulation are performed simultaneously. In this case, to decrease this effect, but they are not compatible with all
the common-mode input signal may be up to 650 mVpp [100]. dry electrodes and further development is necessary to ensure
In [101] The readers can find more information about CMRR good contact at all sites. For example, electrodes do not make
and different sources for common mode signals reported in good contact with the scalp at the location of the visual cortex,
biomedical signals. As a rule of thumb, an amplifier with a even with a head cap. In spite of this potential shortcoming,
high enough CMRR is essential to prevent output saturation. compatibility of the designed electrode with head caps can
We propose a CMRR of 80 dB or more for BCI applications. still be a good feature to reduce motion artifacts.
HABIBZADEH TONEKABONY SHAD et al.: IMPEDANCE AND NOISE OF PASSIVE AND ACTIVE DRY EEG ELECTRODES: A REVIEW 14575
[35] H. Hori, G. Moretti, A. Rebora, and F. Crovato, “The thickness of [58] Z. Zhao et al., “Signal quality and electrode-skin impedance evaluation
human scalp: Normal and bald,” J. Investigative Dermatol., vol. 58, in the context of wearable electroencephalographic systems,” in Proc.
no. 6, pp. 396–399, Jun. 1972. 40th Annu. Int. Conf. IEEE Eng. Med. Biol. Soc. (EMBC), Jul. 2018,
[36] B. A. Taheri, R. T. Knight, and R. L. Smith, “A dry elec- pp. 4965–4968.
trode for EEG recording,” Electroencephalogr. Clin. Neurophys- [59] C. Grozea, C. D. Voinescu, and S. Fazli, “Bristle-sensors—Low-
iol., vol. 90, no. 5, pp. 376–383, May 1994. [Online]. Available: cost flexible passive dry EEG electrodes for neurofeedback and BCI
http://www.sciencedirect.com/science/article/pii/0013469494900531 applications,” J. Neural Eng., vol. 8, no. 2, Mar. 2011, Art. no. 025008,
[37] Q. Fan, K. A. A. Makinwa, and J. H. Huijsing, Capacitively- doi: 10.1088/1741-2560/8/2/025008.
Coupled Chopper Amplifiers. Cham, Switzerland: Springer, 2017, doi: [60] Z. Wang et al., “A multichannel EEG acquisition system with novel
10.1007/978-3-319-47391-8. ag NWs/PDMS flexible dry electrodes,” in Proc. 40th Annu. Int. Conf.
[38] R. Wu, J. H. Huijsing, and K. A. A. Makinwa, Precision Instrumen- IEEE Eng. Med. Biol. Soc. (EMBC), Jul. 2018, pp. 1299–1302.
tation Amplifiers and Read-Out Integrated Circuits. New York, NY, [61] g.tec Medical Engineering. Accessed: Jul. 23, 2019. [Online]. Avail-
USA: Springer, 2013, doi: 10.1007/978-1-4614-3731-4. able: https://www.gtec.at
[39] M. R. Nuwer et al., “IFCN standards for digital record- [62] J. S. Lee, K. S. Park, J. H. Kim, and C. M. Han, “Reverse-curve-
ing of clinical EEG,” Electroencephalogr. Clin. Neurophysiol., arch-shaped dry EEG electrode for increased skin–electrode contact
vol. 106, no. 3, pp. 259–261, Mar. 1998. [Online]. Available: area on hairy scalps,” Electron. Lett., vol. 51, no. 21, pp. 1643–1645,
http://www.sciencedirect.com/science/article/pii/S0013469497001065 Oct. 2015, doi: 10.1049/el.2015.1873.
[40] F. Z. Padmadinata, J. J. Veerhoek, G. J. A. van Dijk, and [63] Y.-H. Chen et al., “Soft, comfortable polymer dry electrodes for
J. H. Huijsing, “Microelectronic skin electrode,” Sens. Actuators B, high quality ECG and EEG recording,” Sensors, vol. 14, no. 12,
Chem., vol. 1, nos. 1–6, pp. 491–494, Jan. 1990. [Online]. Available: pp. 23758–23780, Dec. 2014, doi: 10.3390/s141223758.
http://www.sciencedirect.com/science/article/pii/092540059080257Z [64] X. Guo et al., “A self-wetting paper electrode for ubiquitous bio-
[41] S. Nishimura, Y. Tomita, and T. Horiuchi, “Clinical application of an potential monitoring,” IEEE Sensors J., vol. 17, no. 9, pp. 2654–2661,
active electrode using an operational amplifier,” IEEE Trans. Biomed. May 2017.
Eng., vol. 39, no. 10, pp. 1096–1099, Oct. 1992. [65] C.-T. Lin, L.-D. Liao, Y.-H. Liu, I.-J. Wang, B.-S. Lin, and J.-Y. Chang,
[42] T. Degen, S. Torrent, and H. Jackel, “Low-noise two-wired buffer elec- “Novel dry polymer foam electrodes for long-term EEG measurement,”
trodes for bioelectric amplifiers,” IEEE Trans. Biomed. Eng., vol. 54, IEEE Trans. Biomed. Eng., vol. 58, no. 5, pp. 1200–1207, May 2011.
no. 7, pp. 1328–1332, Jul. 2007. [66] P. Tallgren, S. Vanhatalo, K. Kaila, and J. Voipio, “Evalua-
[43] M. Fernandez and R. Pallas-Areny, “A simple active electrode for tion of commercially available electrodes and gels for recording
power line interference reduction in high resolution biopotential mea- of slow EEG potentials,” Clin. Neurophysiol., vol. 116, no. 4,
surements,” in Proc. 18th Annu. Int. Conf. IEEE Eng. Med. Biol. Soc., pp. 799–806, Apr. 2005. [Online]. Available: http://www.sciencedirect.
vol. 1, Oct. 1996, pp. 97–98. com/science/article/pii/S1388245704003906
[44] J. Xu, R. F. Yazicioglu, B. Grundlehner, P. Harpe, K. A. A. Makinwa, [67] M. Arai, Y. Kudo, and N. Miki, “Electroencephalogram measurement
and C. V. Hoof, “A 160 μw 8-channel active electrode system for from the hairy part of the scalp using polymer-based dry microneedle
EEG monitoring,” IEEE Trans. Biomed. Circuits Syst., vol. 5, no. 6, electrodes,” in Proc. 37th Annu. Int. Conf. IEEE Eng. Med. Biol. Soc.
pp. 555–567, Dec. 2011. (EMBC), Aug. 2015, pp. 3165–3168.
[45] A. C. MettingVanRijn, A. P. Kuiper, T. E. Dankers, and [68] P. Fiedler et al., “Multichannel EEG with novel Ti/TiN dry
C. A. Grimbergen, “Low-cost active electrode improves the resolution electrodes,” Sens. Actuators A, Phys., vol. 221, pp. 139–147,
in biopotential recordings,” in Proc. 18th Annu. Int. Conf. IEEE Eng. Jan. 2015. [Online]. Available: http://www.sciencedirect.com/science/
Med. Biol. Soc., vol. 1, Oct. 1996, pp. 101–102. article/pii/S0924424714004440
[69] Y. Yoshida, Y. Kudo, E. Hoshino, Y. Minagawa, and N. Miki,
[46] A. von Luhmann, H. Wabnitz, T. Sander, and K.-R. Müller, “M3BA:
“Preparation-free measurement of event-related potential in oddball
A mobile, modular, multimodal biosignal acquisition architecture for
miniaturized EEG-NIRS-Based hybrid BCI and monitoring,” IEEE tasks from hairy parts using candle-like dry microneedle electrodes,”
Trans. Biomed. Eng., vol. 64, no. 6, pp. 1199–1210, Jun. 2017. in Proc. 40th Annu. Int. Conf. IEEE Eng. Med. Biol. Soc. (EMBC),
Jul. 2018, pp. 4685–4688.
[47] Y. Zou, V. Nathan, and R. Jafari, “Automatic identification of artifact-
[70] C.-Y. Chen, C.-L. Chang, T.-F. Chien, and C.-H. Luo, “Flexible
related independent components for artifact removal in EEG record-
PDMS electrode for one-point wearable wireless bio-potential
ings,” IEEE J. Biomed. Health Informat., vol. 20, no. 1, pp. 73–81,
acquisition,” Sens. Actuators A, Phys., vol. 203, pp. 20–28,
Jan. 2016.
Dec. 2013. [Online]. Available: http://www.sciencedirect.com/
[48] E. Huigen, “Noise in biopotential recording using surface electrodes,” science/article/pii/S0924424713003944
Univ. Amsterdam Sect. Med. Phys., Amsterdam, The Netherlands,
[71] Y. Nishinaka, R. Jun, G. Prihandana, and N. Miki, “Fabrication of
Tech. Rep., 2000.
polymeric dry microneedle electrodes coated with nanoporous pary-
[49] G. Ruffini et al., “ENOBIO dry electrophysiology electrode; first lene,” in Proc. 17th Int. Conf. Solid-State Sensors, Actuat. Microsyst.
human trial plus wireless electrode system,” in Proc. 29th Annu. Int. (TRANSDUCERS & EUROSENSORS XXVII), Transducers & Eurosen-
Conf. IEEE Eng. Med. Biol. Soc., Aug. 2007, pp. 6689–6693. sors XXVII, 2013, pp. 1326–1327.
[50] B. Razavi, Design of Analog CMOS Integrated Circuits. New York, [72] X. Crispin et al., “The origin of the high conductivity of poly(3,4-
NY, USA: McGraw-Hill, 2005. ethylenedioxythiophene)–poly(styrenesulfonate) (PEDOT–PSS) plastic
[51] A. J. Deen and O. Marinov, “Effect of forward and reverse substrate electrodes,” Chem. Mater., vol. 18, no. 18, pp. 4354–4360, Sep. 2006.
biasing on low-frequency noise in silicon PMOSFETs,” IEEE Trans. [73] M. J. Abu-Saude and B. I. Morshed, “Patterned vertical carbon nan-
Electron Devices, vol. 49, no. 3, pp. 409–413, Mar. 2002. otube dry electrodes for impedimetric sensing and stimulation,” IEEE
[52] F. A. Levinzon, “Measurement of low-frequency noise of modern low- Sensors J., vol. 15, no. 10, pp. 5851–5858, Oct. 2015.
noise junction field effect transistors,” IEEE Trans. Instrum. Meas., [74] C.-W. Lam, J. T. James, R. McCluskey, S. Arepalli, and R. L. Hunter,
vol. 54, no. 6, pp. 2427–2432, Dec. 2005. “A review of carbon nanotube toxicity and assessment of potential
[53] A. Searle and L. Kirkup, “A direct comparison of wet, dry and insu- occupational and environmental health risks,” Crit. Rev. Toxicol.,
lating bioelectric recording electrodes,” Physiol. Meas., vol. 21, no. 2, vol. 36, no. 3, pp. 189–217, Jan. 2006.
p. 271, May 2000. [75] C. P. Firme and P. R. Bandaru, “Toxicity issues in the application of
[54] C. M. Lopez et al., “An implantable 455-active-electrode 52-channel carbon nanotubes to biological systems,” Nanomed. Nanotechnol., Biol.
CMOS neural probe,” IEEE J. Solid-State Circuits, vol. 49, no. 1, Med., vol. 6, no. 2, pp. 245–256, Apr. 2010.
pp. 248–261, Jan. 2014. [76] W. Yang, P. Thordarson, J. J. Gooding, S. P. Ringer, and F. Braet,
[55] H. Zhang et al., “A motion interference-insensitive flexible dry elec- “Carbon nanotubes for biological and biomedical applications,” Nan-
trode,” IEEE Trans. Biomed. Eng., vol. 63, no. 6, pp. 1136–1144, otechnol., vol. 18, no. 41, 2007, Art. no. 412001.
Jun. 2016. [77] R. Alshehri, A. M. Ilyas, A. Hasan, A. Arnaout, F. Ahmed, and
[56] S. L. Kappel and P. Kidmose, “Study of impedance spectra for dry A. Memic, “Carbon nanotubes in biomedical applications: Factors,
and wet EarEEG electrodes,” in Proc. 37th Annu. Int. Conf. IEEE Eng. mechanisms, and remedies of toxicity: Miniperspective,” J. Med.
Med. Biol. Soc. (EMBC), Aug. 2015, pp. 3161–3164. Chem., vol. 59, no. 18, pp. 8149–8167, Sep. 2016.
[57] S. Krachunov and A. Casson, “3D printed dry EEG electrodes,” [78] N. Saito et al., “Safe clinical use of carbon nanotubes as innovative
Sensors, vol. 16, no. 10, p. 1635, Oct. 2016, doi: 10.3390/s16101635. biomaterials,” Chem. Rev., vol. 114, no. 11, pp. 6040–6079, Jun. 2014.
HABIBZADEH TONEKABONY SHAD et al.: IMPEDANCE AND NOISE OF PASSIVE AND ACTIVE DRY EEG ELECTRODES: A REVIEW 14577
[79] W. C. Ng et al., “Micro-spike EEG electrode and the vacuum- [100] H. Chandrakumar and D. Marković, “An 80-mVpp linear-input range,
casting technology for mass production,” J. Mater. Process. Tech- 1.6-G input impedance, low-power chopper amplifier for closed-loop
nol., vol. 209, no. 9, pp. 4434–4438, May 2009. [Online]. Available: neural recording that is tolerant to 650-mVpp common-mode interfer-
http://www.sciencedirect.com/science/article/pii/S092401360800770X ence,” IEEE J. Solid-State Circuits, vol. 52, no. 11, pp. 2811–2828,
[80] H. Chandrakumar and D. Marković, “5.5 A 2 μW 40 mVpp linear- Nov. 2017.
input-range chopper- stabilized bio-signal amplifier with boosted input [101] N. V. Thakor and J. G. Webster, “Ground-free ECG recording with
impedance of 300 MO and electrode-offset filtering,” in IEEE Int. two electrodes,” IEEE Trans. Biomed. Eng., vol. BME-27, no. 12,
Solid-State Circuits Conf. (ISSCC) Dig. Tech. Papers, Jan. 2016, pp. 699–704, Dec. 1980.
pp. 96–97.
[81] Q. Fan, F. Sebastiano,
√ J. H. Huijsing, and K. A. A. Makinwa,
“A 1.8 μw 60 nv/ Hz capacitively-coupled chopper instrumentation Erwin Habibzadeh Tonekabony Shad recei-
amplifier in 65 nm CMOS for wireless sensor nodes,” IEEE J. Solid- ved the B.Sc. degree in electrical engineering
State Circuits, vol. 46, no. 7, pp. 1534–1543, Jul. 2011. from Guilan University, Iran, in 2015, and the
[82] Y. M. Chi, C. Maier, and G. Cauwenberghs, “Ultra-high input M.Sc. degree in analog integrated circuit design
impedance, low noise integrated amplifier for noncontact biopotential from the Iran University of Science and Technol-
sensing,” IEEE J. Emerg. Sel. Topics Circuits Syst., vol. 1, no. 4, ogy in 2018. He has been working on biomedical
pp. 526–535, Dec. 2011. amplifiers since receiving the M.Sc. degree. He is
[83] C. Harland, T. Clark, and R. Prance, “Electric potential probes-new currently pursuing the Ph.D. degree with the
directions in the remote sensing of the human body,” Meas. Sci. Norwegian University of Science and Technology
Technol., vol. 13, no. 2, p. 163, 2001. (NTNU). He is working on chopper amplifiers
[84] M. Saad, M. El-Nozahi, and H. Ragai, “A chopper capacitive feedback for EEG signal acquisition in a commercially
instrumentation amplifier with input impedance boosting technique,” available 0.18 μm CMOS technology. He is also a part of the David
in Proc. IEEE 59th Int. Midwest Symp. Circuits Syst. (MWSCAS), versus Goliath: FlexEEG project, which works on EEG applications with
Oct. 2016, pp. 1–4. dry electrodes.
[85] J. Xu et al., “A wearable 8-channel active-electrode EEG/ETI acqui-
sition system for body area networks,” IEEE J. Solid-State Circuits,
vol. 49, no. 9, pp. 2005–2016, Sep. 2014. Marta Molinas (Member, IEEE) received the
[86] J. Xu, B. Busze, C. Van Hoof, K. A. A. Makinwa, and R. F. Yazicioglu, B.Sc. degree in electromechanical engineering
“A 15-channel digital active electrode system for multi-parameter from the National University of Asuncion, Asun-
biopotential measurement,” IEEE J. Solid-State Circuits, vol. 50, no. 9, cion, Paraguay, in 1992, the M.Sc. degree in
pp. 2090–2100, Sep. 2015. information engineering from Ryukyu University,
[87] X. Zhou, Q. Li, S. Kilsgaard, F. Moradi, S. L. Kappel, and P. Kidmose, Nishihara, Japan, in 1997, and the Ph.D. degree
“A wearable ear-EEG recording system based on dry-contact active in electrical engineering from the Tokyo Institute
electrodes,” in Proc. IEEE Symp. VLSI Circuits (VLSI-Circuits), of Technology, Tokyo, Japan, in 2000. She was a
Jun. 2016, pp. 1–2. Guest Researcher with the University of Padova,
[88] V. Nathan and R. Jafari, “Reducing the noise level of EEG signal Italy, in 1998. From 2004 to 2007, she was
acquisition through reconfiguration of dry contact electrodes,” in Proc. a Postdoctoral Researcher with the Norwegian
IEEE Biomed. Circuits Syst. Conf. (BioCAS), Oct. 2014, pp. 572–575. University of Science and Technology (NTNU), Trondheim, Norway,
[89] V. Nathan and R. Jafari, “Design principles and dynamic front end where she was a Professor with the Department of Electric Power Engi-
reconfiguration for low noise EEG acquisition with finger based neering, from 2008 to 2014. She is currently a Professor with the Depart-
dry electrodes,” IEEE Trans. Biomed. Circuits Syst., vol. 9, no. 5, ment of Engineering Cybernetics, NTNU. Her research interests include
pp. 631–640, Oct. 2015. stability of complex power electronics systems, harmonics, oscillatory
[90] M. Sun, W. Jia, W. Liang, and R. J. Sclabassi, “A low- phenomena, and nonstationary signals from humans and machines. She
impedance, skin-grabbing, and gel-free EEG electrode,” in Proc. serves as an Editor of the IEEE JOURNAL OF EMERGING AND SELECTED
Annu. Int. Conf. IEEE Eng. Med. Biol. Soc. (EMBC), Aug. 2012, TOPICS IN POWER ELECTRONICS and the IEEE TRANSACTIONS ON ENERGY
pp. 1992–1995. CONVERSION. She is an Associate Editor of the IEEE TRANSACTIONS
[91] B. Luan, W. Jia, P. D. Thirumala, J. Balzer, D. Gao, and M. Sun, OF POWER ELECTRONICS and the IEEE TRANSACTIONS ON INDUSTRIAL
“A feasibility study on a single-unit wireless EEG sensor,” in Proc. ELECTRONICS.
12th Int. Conf. Signal Process. (ICSP), Oct. 2014, pp. 2282–2285, doi:
10.1109/icosp.2014.7015401.
[92] Medical Electrical Equipment—Part 2-26: Particular Requirements for Trond Ytterdal (Senior Member, IEEE) received
the Basic Safety and Essential Performance of Electroencephalographs, the M.Sc. and Ph.D. degrees in electrical engi-
Standard IEC 60601-2-26, May 2019. neering from the Norwegian Institute of Tech-
[93] U. Ha, J. Lee, M. Kim, T. Roh, S. Choi, and H.-J. Yoo, “An nology in 1990 and 1995, respectively. He was
EEG-NIRS multimodal SoC for accurate anesthesia depth monitor- employed as a Research Associate at the
ing,” IEEE J. Solid-State Circuits, vol. 53, no. 6, pp. 1830–1843, Department of Electrical Engineering, University
Jun. 2018. of Virginia, from 1995 to 1996, and as a Research
[94] Y.-H. Chen et al., “Soft, comfortable polymer dry electrodes for Scientist at the Electrical, Computer, and Sys-
high quality ECG and EEG recording,” Sensors, vol. 14, no. 12, tems Engineering Department, Rensselaer Poly-
pp. 23758–23780, 2014. technic Institute, Troy, New York, from 1996 to
[95] H. Hinrichs, M. Scholz, A. K. Baum, J. W. Y. Kam, R. T. Knight, 1997. From 1997 to 2001, he worked as a Senior
and H.-J. Heinze, “Comparison between a wireless dry electrode EEG ASIC Designer at Nordic Semiconductor, Trondheim, Norway. Since
system with a conventional wired wet electrode EEG system for clinical 2001, he has been on the Faculty of the Norwegian University of Science
applications,” Sci. Rep., vol. 10, no. 1, pp. 1–14, Dec. 2020. and Technology (NTNU), where he is a Professor with the Department
[96] J. W. Y. Kam et al., “Systematic comparison between a wireless of Electronics and Telecommunications. He present research interests
EEG system with dry electrodes and a wired EEG system with wet include design of analog integrated circuits, behavioral modeling and
electrodes,” NeuroImage, vol. 184, pp. 119–129, Jan. 2019. simulation of mixed-signal systems, modeling of nanoscale transistors,
[97] Y.-H. Yu, S.-W. Lu, L.-D. Liao, and C.-T. Lin, “Design, fabrication, and novel device structures for application in circuit simulators. He has
and experimental validation of novel flexible silicon-based dry sensors authored or coauthored more than 200 scientific articles in international
for electroencephalography signal measurements,” IEEE J. Transl. Eng. journals and conference proceedings. He is a coauthor of the books
Health Med., vol. 2, pp. 1–7, 2014. Semiconductor Device Modeling for VLSI (Prentice Hall, 1993), Introduc-
[98] T. Yamamoto and Y. Yamamoto, “Analysis for the change of skin tion to Device Modeling and Circuit Simulation (Wiley, 1998), and Device
impedance,” Med. Biol. Eng. Comput., vol. 15, no. 3, pp. 219–227, Modeling for Analog and RF CMOS Circuit Design (Wiley, 2003), and has
May 1977. been a contributor to several other books published internationally. He is
[99] W. M. C. Sansen, Analog Design Essentials. New York, NY, USA: also a co-developer of the circuit simulator AIM-Spice. He is a member
Springer, 2006. of the Norwegian Academy of Technological Sciences.