ANSR Salmonella Lot No.

: 273608
Product No. 9870 Exp. Date: Nov 02, 2019

A representative sample of this lot was tested according to Neogen Standard Assay Procedures. Testing was performed in
compliance with the product information sheet. This product met or exceeded established performance specifications as listed
below.

Performance
SENSITIVITY SPECIFICATION RESULT DECISION
Salmonella typhimurium 104 cfu/mL All Replicates Positive Pass

Salmonella enteritidis 104 cfu/mL All Replicates Positive Pass

Salmonella newport 104 cfu/mL All Replicates Positive Pass

Salmonella typhimurium 103 cfu/mL At Least Half of Replicates Positive Pass

Salmonella enteritidis 103 cfu/mL At Least Half of Replicates Positive Pass

Salmonella newport 103 cfu/mL At Least Half of Replicates Positive Pass

SPECIFICITY SPECIFICATION RESULT DECISION


Escherichia coli >105 cfu/mL All Replicates Negative Pass

Citrobacter freundii >105 cfu/mL All Replicates Negative Pass

Hafnia alvei >105 cfu/mL All Replicates Negative Pass

The Lysis and Lysis Buffer supplied with this lot are universal components and have lot numbers and expiry dates different than
what is defined on the Certificate of Analysis. The data above was generated using all components within the product lot.

Variances in data may occur. External factors such as stress from shipping, storage conditions, handling or contamination of
components may influence results. Neogen Corporation certifies that this lot has met all internal quality control specifications for this
product.

Caution: All kit assembly is a defined matrix and individual components should not be mixed with other kits.

Date: May 7, 2019

CofA-9870