Homocysteine

CHEMISTRY PARAMETERS FOR HITACHI 7180


Standby 37.0 Operator ID: 1 Thu 10/10/02 08:36
Workplace Reagent Calibration QC Utility
Calculated Test Special Wash Report Format Communication Trace Start
System Maintenance Maintenance Log Application Cond.
Analyze Calibration Range Standard Sample
Stop
Select Test: HCY Diluent: Water 0
Stop
Test Name: HCY Assay/Time/Point: 2 Point End 10 27 31 0 0
D
No: 1 Wavelength(2nd/Primary): 700 340
App. Code: 1
Sample Volume Reagent
Class 1 Class 2 R1: 190 0 00001 99 Sample
Normal: 14.0 0.0 0 3.0 0.0 0 R2: 28 0 00001 99 Track
Decrease: 6.0 0.0 0 3.0 0.0 0 R3: 20 0 00001 99
Print
Increase: 20.0 0.0 0 3.0 0.0 0 R4: 0 0 00001 99

Alarm
Class1 Class2 Dec/Inc
ia
Abs. Limit: 5000 5000 Decrease Twin Test:
Help
Prozone Limit: 0 0 Lower

Delete Read Parameters… Add Application With Parameters… OK


Sleep
Exit

Standby 37.0 Operator ID: 1 Thu 10/10/02 08:36


Sy
Workplace Reagent Calibration QC Utility
Calculated Test Special Wash Report Format Communication Trace Start
System Maintenance Maintenance Log Application Cond.
Analyze Calibration Range Standard Sample
Stop
Select Test: HCY
Stop
Calibration Type: Linear
Point: 2 Span Point: 2
Weight: 0
Auto Calibration
Time Out Clange Over SD Limit: 999.9 Sample
Blank 24 hr Lot No. : Duplicate Limit: 99 % 32000 Abs. Track
Span 0 hr Bottle: Sensitivity Range: -99999 - 99999
Print
s
2 Point 100 hr S1 Abs. Range: -32000 32000
Full 0 hr
Alarm

Help

Delete OK
Sleep
Exit
The Calibration Application screen is displayed.
Homocysteine
CHEMISTRY PARAMETERS FOR HITACHI 7180
Standby 37.0 Operator ID: 1 Thu 10/10/02 08:36
Workplace Reagent Calibration QC Utility
Calculated Test Special Wash Report Format Communication Trace Start
System Maintenance Maintenance Log Application Cond.
Analyze Calibration Range Standard Sample
Stop
Select Test: HCY Report Name: HCY
Stop
Test Name: HCY Control Interval: 0 Unit: umol/L Instrument Factor(Y=aX+b)
D
No: 1 Data Mode: On Board a= 1.0 b= 0.0
App. Code: 1
Select Sample Type: Ser/Pl Use Qualitative Table
Expected Ranges (1) 0.0 ? Sample
Age Male Female (2) 0.0 - Track
1 Years 6.0 - 17.0 5.0 - 14.0 (3) 0.0 +-
Print
50 Years -99999 - 99999 -99999 - 99999 (4) 0.0 +
-99999 - 99999 -99999 - 99999 (5) 0.0 ++
Alarm
Default Age: 0Years – 1Years (6) +++
ia
Default Sex: Male Repeat Range: 0 - 50.0
Help
Class 1 Technical Range: 0 - 50.0 Class 2 Technical Range: -99999 - 99999

Delete OK
Sleep
Exit
The Range Application screen is displayed.

Standby 37.0 Operator ID: 1 Thu 10/10/02 08:36


Sy
Workplace Reagent Calibration QC Utility
Calculated Test Special Wash Report Format Communication Trace Start
System Maintenance Maintenance Log Application Cond.
Analyze Calibration Range Standard Sample
Stop
Select Test: HCY
Stop

<Standard> (1) (2) (3) (4) (5) (6)


Calibration Code: Wa # 0 0 0 0
Concentration: 0.00 *
Position: * * Sample
Track
Sample Volume: 15.0 15.0 3.0 3.0 3.0 3.0
Print
s
Diluted Sample Volume: 0.0 0.0 0.0 0.0 0.0 0.0
Diluent Volume: 0 0 0 0 0 0
Alarm

Help

Delete OK
Sleep
Exit
The Standard Application screen is displayed.