Professional Documents
Culture Documents
Ref CAX
Ref CAX
PROBE
Instead of calibrating probes for refinery liquors and massecuites in brix units by
taking samples and using the regression method, a simpler procedure is to
calibrate the probe against capacitance using a calibrator. In this method, the effect
of changing conductivity due to variations in ash is minimised.
The method still involves use of the regression method but a calibrator type CAX is
used to simulate different massecuite or liquor conditions. 20 calibration points are
used as follows: (refer to the manual for calibration procedure)
1 98.6 32 2 0 6 0
2 98.6 32 3 0 6 0
3 98.6 32 6 0 6 0
4 87.9 47 2 0 8 0
5 87.9 47 3 0 8 0
6 87.9 47 6 0 8 0
7 72.1 69 2 0 C 0
8 72.1 69 3 0 C 0
9 72.1 69 6 0 C 0
10 50.0 100 2 0 0 1
11 50.0 100 3 0 0 1
12 50.0 100 6 0 0 1
13 27.1 132 2 0 6 1
14 27.1 132 3 0 6 1
15 27.1 132 6 0 6 1
16 27.1 132 9 0 6 1
17 0.7 169 2 0 C 1
18 0.7 169 3 0 C 1
19 0.7 169 6 0 C 1
20 0.7 169 9 0 C 1
Go to "new calibration" and set calibrator as above for samples 1 to 20. Once this
is entered, go to "add analyses" and key in the percent values for samples 1 to 20.
Once this is entered, go to "regression". The output will then be calibrated to give a
0-100% readout for a capacitance of 170 to 30 pf. This will cover a range of
approximately 50 to 100 brix. K1 and K2 can then be set in the process to adjust
zero and span for the 4-20 ma signal to cover the required range. K1 is the
indicated value required for 4 ma (zero) and K2 the value required for 20
mA output.