M E E T I N G A B S T R AC T S
Conclusions With either IBTR or MCBC the size of the second tumour is O6
important, patients with cancers >2 cm in diameter being at a significantly Comparison of film screen, computed radiography and direct digital
greater risk of death. Lead-time bias from the mode of detection may be a mammography in the Southern Derby screening programme
factor in these results. The frequency of surveillance mammography should M Bagnall, AE Turnbull, S Puri
be considered for maximum benefit. Breast Unit, Royal Derby Hospital, Derby, UK
Breast Cancer Research 2010, 12(Suppl 3):O6 (doi: 10.1186/bcr2653)
O4 Introduction Our screening programme has been fully digital since April
Reduced breast biopsy rates with a combined high temporal and high 2009 and utilises a combination of computed radiography (CR) (Fuji Profect)
spatial resolution MR imaging protocol at 3 Tesla and three direct digital mammography (DDM) mammographic systems (Fuji
K Pinker1, W Bogner2, S Gruber2, G Grabner2, S Trattnig2, TH Helbich1 Amulet, GE DS and GE Essential). This study compares the performance of
1
Department of Radiology, Division of Molecular and Gender Imaging, Medical our CR and DDM systems against film screen (FS) mammography in a real-life
University Vienna, Austria; 2Department of Radiology, MR Centre of Excellence, screening environment.
Medical University Vienna, Austria Methods Southern Derbyshire screening episodes between April 2007 and
Breast Cancer Research 2010, 12(Suppl 3):O4 (doi: 10.1186/bcr2651) April 2010 were interrogated on NBSS using a Crystal report. Recall and
cancer detection rates were collected according to mammography type.
Purpose To develop a 3.0 Tesla breast imaging protocol that combines high Data were analysed for significance using the chi-squared test.
temporal and spatial resolution 3D MR sequences for quantitative time Results A total of 66,989 screening episodes were recorded, 22,039 FS (recall
course and morphological analysis of breast lesions. rate 3.23%, cancer detection 8.08/1,000), 30,739 CR (recall rate 2.73%, cancer
Materials and methods One hundred and sixty-five breast lesions classified detection 7.87/1,000) and 14,157 DDM (recall rate 3.87%, cancer detection
by mammography or ultrasound as BIRADS 4 and 5 were included in this 9.89/1,000). Recall rates compared with FS were statistically significantly
prospective IRB-approved study. The MRI protocol consisted of a coronal T2- different for CR (P = 0.0007) and DDM (P = 0.0016). Cancer detection rates
weighted TIRM and a coronal combined high temporal and spatial resolution compared with FS mammography were not significantly different for either
T1-weighted sequence before and after application of a standard-dose CR (P = 0.833) or DDM (P = 0.081), although there was a significant difference
Gd-DOTA (VIBE with a high temporal resolution of SI 1.7 mm isotropic; TA found between CR and DDM (P = 0.035).
3.45 min for 17 measurements; FLASH with high spatial resolution of SI 1 mm Conclusions DDM in our programme has detected an increased number of
isotropic; TA 2 min). Lesion size and morphology were assessed according to cancers but at an increased recall rate. CR has detected a similar number of
the BIRADS classification. ROIs for suspicious areas were manually drawn and cancers as FS mammography at a lower recall rate and the study supports
evaluated for contrast-enhancement behavior by plotting intensity courses the use of CR in our programme. As an interim measure in changing to a fully
against time. Sensitivity and specificity with a 95% confidence interval and DDM service it has allowed us to spread costs whilst adopting a single digital
the negative predictive value (NPV) and positive predictive value (PPV) workflow, which is efficient and safe for our clients.
were calculated. Diagnostic accuracy was assessed. The histopathological
diagnoses were used as the standard of reference.
Results All malignant breast lesions were identified correctly with a sensitivity
of 100%, a specificity of 84% and a diagnostic accuracy of 95.7%. PPV was
0.94 and a NPV 1. All seven false positive lesions were lesions with atypia.
Conclusions The proposed combined 3 Tesla MR imaging protocol,
comprising both high temporal and spatial resolution, enabled an accurate
detection and assessment of breast lesions with high sensitivity and POSTER PRESENTATIONS
specificity reducing false positive breast biopsies
P1
Breast density as a predictor of breast cancer risk
G Lip1, N Zakharova2, SW Duffy2, MGC Gillan1, FJ Gilbert1
1
O5 University of Aberdeen, UK; 2Queen Mary University of London, UK
MR visible only lesions: what are the predictors for malignant outcome? Breast Cancer Research 2010, 12(Suppl 3):P1 (doi: 10.1186/bcr2654)
M Bhattacharyya, F Ng, W Teh
Northwick Park Hospital, Harrow, UK Introduction As a part of a retrospective study of computer-aided detection
Breast Cancer Research 2010, 12(Suppl 3):O5 (doi: 10.1186/bcr2652) in breast cancer screening, we recorded the percentage of the breast density
on a visual analogue scale on 4,866 routine screening mammograms taken in
Introduction To correlate pathological outcomes of MRI vacuum biopsies on 1996 as part of the North East of Scotland Breast Screening Service. In these
MRI visible only breast lesions with lesion morphology, time-enhancement women, 284 breast cancers were diagnosed from the time of the original
curves and clinical indications to determine the use of these as predictors mammograms up until February 2010.
for malignancy. Methods The main goal was to evaluate breast density as a breast cancer risk
Methods A retrospective analysis of 277 patients referred for MRI-guided factor by detection mode, histology and time since the original mammogram.
vacuum biopsies of impalpable breast lesions visible only on MRI was The association of density with risk of breast cancer was assessed using
performed. All patients had a minimum follow-up period of 11 months. MRI logistic regression giving odds ratios per 10% increase in density, and by
biopsies were undertaken on a 1.5 T magnet using a minimum of 12 passes comparison of continuous mean densities between particular groups of
of vacuum-assisted biopsies. The pathological findings were correlated cancers and those who did not develop breast cancer.
against BI-RADS appearances and time-enhancement characteristic of the Results After adjusting for age, breast density was significantly associated
lesions and against the clinical indications for MRI examination. with cancer in the first 6 years after the original mammogram (OR per 10%
Results A total of 286 vacuum biopsies were undertaken. Eighty-one were density = 1.08, 95% CI = 1.01 to 1.15, P = 0.03), but not with cancer more than
malignant (28.3%), of which 72.8% are masses and 27.2% are nonmasses. Only 6 years after the mammogram (OR = 1.01, 95% CI = 0.93 to 1.10, P = 0.8).
two malignant lesions had a type 1 curve (2.5%), compared with malignant There was a statistically significant association between histological type
lesions with type 2 (54.3%) and type 3 curves (43.2%). Both malignant lesions of tumour and breast density (P = 0.02). The mean densities for no cancer,
with type 1 curve had a suspicious morphology. Nonmalignant lesions with invasive ductal carcinoma, DCIS and invasive lobular carcinoma were 32%
type 3 enhancement included lymph nodes, fibroadenomatoid hyperplasia, (SD 23%), 33% (21%), 35% (26%) and 44% (24%), respectively. The effect of
papillary lesions, fibrocystic change and lobular neoplasia. density on risk varied significantly by detection mode (P = 0.02), with highest
Conclusions Lesion morphology and time-enhancement curves are useful densities being observed in interval cancers arising more than 1 year after
predictors of malignancy and can be used to develop an algorithm to help the mammogram (55% compared with 32% in those with no cancer).
direct appropriate biopsy of MRI-detected lesions. We recommend that in Conclusions Breast density as measured by visual analogue assessment is
the absence of suspicious morphology, only lesions with type 2 and type 3 strongly associated with late interval cancers and with lobular carcinoma. Its
curves should be subjected to MRI-guided biopsy. predictive value for risk, however, declines with time since its measurement.
Breast Cancer Research 2010, Volume 12 Suppl 3 S3
http://breast-cancer-research.com/supplements/12/S3
P10 P12
Preliminary assessment of ShearWave™ elastography features in Texture analysis applied to full-field digital mammography: ability to
predicting breast lesion malignancy discriminate between invasive ductal and invasive lobular breast cancer –
DO Cosgrove1, C Doré2, C Tourasse2, R Ohlinger3, H Madjar4, R Hooley5, preliminary results
C Balu-Maestro6, E Mendelson7, W Svensson1,8, A Tardivon9, A Cossi10, L Barke10,11, M Mayerhoefer1, N Perry2, S Milner2, K Mokbel2, S Duffy3, K Pinker1,2
L Larsen10,12 1
Department of Radiology, Division of Molecular and Gender Imaging, Medical
1
Imperial College, Hammersmith Hospital, London, UK; 2Jean Mermoz Private University Vienna, Austria; 2Princess Grace Hospital, The London Breast Institute,
Hospital, Lyon, France; 3University Hospital, Frauenklinik, Greifswald, Germany; London, UK; 3Wolfson Institute, Queen Mary College, University of London, UK
4
German Diagnostic Clinic, Wiesbaden, Germany; 5Yale Medical Center, New Breast Cancer Research 2010, 12(Suppl 3):P12 (doi: 10.1186/bcr2665)
Haven, CT, USA; 6Centre Antoine Lacassagne, Nice, France; 7Northwestern
Memorial Hospital, Chicago, IL, USA; 8Charing Cross Hospital, London, UK; 9Centre Purpose To determine texture features of IDC and invasive lobular carcinoma
Antoine Lacassagne, Nice, France; 10Institut Curie, Paris, France; 11Sally Jobe Breast (ILC) of the breast on full-field digital mammography (FFDM). To evaluate the
Center, Denver, CO, USA; 12University of Southern California, Los Angeles, CA, USA ability of texture analysis to differentiate between those tumor types.
Breast Cancer Research 2010, 12(Suppl 3):P10 (doi: 10.1186/bcr2663) Materials and methods Fourteen IDC and nine ILC imaged with FFDM were
included in this study. For each lesion the ROI was manually defined covering
Introduction ShearWave™ elastography (SWE) provides a quantitative the lesion and 1 cm normal-appearing breast tissue around the lesion.
measurement of tissue stiffness with high spatial resolution and may Texture features derived from the grey-level histogram, the co-occurrence
improve characterization of breast masses. The goal of this study was to matrix, the run-length matrix, the absolute gradient, the autoregressive
evaluate the reproducibility of SWE and the impact of adding SWE features model, and the wavelet transform were calculated for the ROIs. Fisher
to the BI-RADS classification of breast masses from the first 1,000 cases in a coefficients were calculated to determine which texture features were best-
prospective multicenter trial. suited for distinguishing between IDC and ILC. Based on the combination
Methods SWE studies were performed on a prototype of the Aixplorer of those five texture features with the highest Fisher coefficients, lesion
system (Supersonic Imagine, Aix-en-Provence, France). A subset of 192 breast classification was performed, using linear discriminant analysis (LDA) and
lesions (42.71% malignant) was analyzed. Reproducibility of SWE images and principal component analysis (PCA) classifiers, as well as a k-means clustering
measurements was assessed; logistic regression analysis was performed to algorithm. Classification accuracy was used as the primary outcome measure.
predict the pathology findings. SWE features were added to the ultrasound Results Of the five texture features with the highest Fisher coefficients, the
BI-RADS to generate models that were challenged by comparing the areas top four were derived from the wavelet transform. Using LDA and PCA,
under the ROC curves (Az), and the sensitivity and specificity scores. classification accuracies of 82.6% (19 of 23 lesions) and 78.3% (18 of 23
Results In the preliminary analysis, intra-operator reproducibility of SWE size lesions) were achieved, respectively. k-means clustering also yielded a similar
(R ≥0.93) and mean elasticity (R = 0.88) measurements were in near-perfect classification accuracy of 82.6% (19 of 23 lesions).
agreement. Using the best three-variable model (BIRADS + elasticity shape Conclusions Texture features, best suited for discrimination between ILC and
+ maximum elasticity), the Az increased from 0.77 to 0.93 and specificity IDC, are derived from the wavelet transform. Texture analysis of breast cancer
increased from 61.8% to 87.3%, although sensitivity decreased from 92.7% cases imaged with FFDM allows a good degree of accuracy of discrimination
to 87.8%. Adding more variables did not effect further improvements. between IDC and ILC.
Conclusions In this ongoing study, SWE provided reproducible information
(elasticity values and SWE mapping) that improved the characterization
of breast lesions. These features are directly linked to the characteristics of
SWE: local quantification and millimeter resolution. Further evaluation of the P13
study is in progress. Sensitivity of integral computer-aided detection with full-field digital
mammography for detection of breast cancer according to different
histopathological tumor types and appearances
K Pinker1,2, N Perry2, S Milner2, K Mokbel2, S Duffy3
1
Department of Radiology, Division of Molecular and Gender Imaging, Medical
P11 University Vienna, Austria; 2Princess Grace Hospital, The London Breast Institute,
To recognise the imaging of complications from Macrolane™ injection London, UK; 3Wolfson Institute, Queen Mary College, University of London, UK
SE McWilliams1, L Wilding2 Breast Cancer Research 2010, 12(Suppl 3):P13 (doi: 10.1186/bcr2666)
1
Guy’s and St Thomas’ NHS Trust, London, UK; 2West Middlesex University Hospital,
London, UK Purpose To retrospectively evaluate the sensitivity of computer-aided
Breast Cancer Research 2010, 12(Suppl 3):P11 (doi: 10.1186/bcr2664) detection (CAD) for full-field digital mammography (FFDM) in 360 breast
cancers with regard to mammographic appearance and histopathological
Introduction Macrolane™ volume restoration factor (VRF) (Q-Med) is in vogue tumor type.
for use in breast augmentation. There is sparse literature on the imaging Materials and methods Three hundred and sixty breast cancers imaged with
characteristics of complications resulting from Macrolane™ injection. FFDM were evaluated retrospectively using CAD. A CAD mark was scored true
Methods This study describes the multimodality imaging of 12 patients positive (TP) if it correctly indicated a malignant lesion. All other CAD marks
from three centres who had breast complications as a result of Macrolane™ were considered false. Cancer cases were classified as microcalcifications (64),
injection. The patients had all undergone Macrolane™ injection at varying masses (196), or both (100). Histopathological findings were classified as IDC,
intervals prior to presentation. Macrolane™ consists of non-animal ILC, DCIS or other. Sensitivity values for CAD according to mammographic
stabilised hyaluronic acid (NASHA™) and is approved for nonsurgical breast appearance and, histopathological findings were analyzed using chi-squared
augmentation. Treatment involves injection of up to 100 ml fluid, superficial tests.
to the pectoral muscle and deep to the glandular breast disc, and is semi- Results A TP mark was observed on 319 out of the 360 cancers (89%).
permanent, usually only lasting up to 1 year before a top-up injection is Calcifications were significantly (P <0.001) more likely to be marked than
required. masses, 163/164 (99%) compared with 244/296 (82%). The probability
Results Presenting symptoms included breast lumps, breast pain and axillary of a mass TP mark was significantly related to histological type (P <0.001),
pain. The complications seen on imaging included abscess formation, in- with sensitivity being greatest for IDC at 76% (167/221) compared with
determinate masses, subpectoral collections and complex cystic collections. 67% (38/57) for ILC, 31% (16/51) for DCIS and 74% (23/31) for other types.
The fluid and indeterminate masses were not all resorbed in the timeframe The probability of a calcification TP mark was also significantly related to
expected of Macrolane™ and remained in situ in some cases for at least histological type, with sensitivity being greatest for DCIS at 75% (38/51),
2 years. compared with 43% (96/221) for IDC, 33% (19/57) for ILC and 32% (10/31)
Conclusions Macrolane™ injection is associated with long-term complica- for other types.
tions and creates diagnostic challenges on breast imaging that are only Conclusions CAD prompted the significant majority of radiological
beginning to emerge in the breast clinics and that may warrant triple abnormalities related to cancers and was most accurate for detection of
assessment. calcification and DCIS, less so for mass lesions and ILC.
Breast Cancer Research 2010, Volume 12 Suppl 3 S6
http://breast-cancer-research.com/supplements/12/S3
Figure 1 (abstract P15). Left: tumour visualisation with conventional analysis DCE-MRI. Middle: CAD curve shape map (red, wash-out; green, plateau; blue,
persistent). Right: kinetic curve.
Breast Cancer Research 2010, Volume 12 Suppl 3 S7
http://breast-cancer-research.com/supplements/12/S3
staging were recruited into this study following informed consent. The MRI P20
was undertaken on a 1.5T GE CVi/NVi (Milwaukee, WI, USA) and a 3T Philips Negative predictive value for atypia and malignancy of 14-gauge core
Achieva (Best, the Netherlands). T2W, dynamic T1W (voxel size 0.85 x 1.19 x biopsy of breast papillomas
2 mm – 1.5 T MRI, and 0.6 x 0.6 x 2 mm – 3T MRI) and high-resolution fat- GT Mataka, MJ Pearson, AJ Maxwell
suppressed T1W postcontrast sequences (single-dose contrast) were carried Royal Bolton Hospital, Bolton, UK
out. The confidence level in morphology and contrast kinetics (three-point Breast Cancer Research 2010, 12(Suppl 3):P20 (doi: 10.1186/bcr2673)
scale) and conspicuity for each lesion (five-point scale, –2 to +2) was assessed
by a single observer (SKAR). Introduction Papillary lesions of the breast may be heterogeneous and
Results Seventeen patients were included in the study. Eleven patients had associated with atypia or frank malignancy. Current practice is to perform
one or more lesions, giving 22 lesions. The confidence level in assessing wider sampling of lesions showing benign papilloma on core biopsy because
morphology was high in 16/22 and 19/22 and in assessing contrast kinetics of the risk of underestimation of disease. The literature, however, is unclear
was high in 12/22 and 16/22 in 1.5T and 3T examinations, respectively. The about the negative predictive value (NPV) of 14-gauge needle core biopsy
mean and standard deviation of the conspicuity score are 1.09 ± 0.88 for 3T. for atypia or malignancy.
Conclusions The confidence in characterising and conspicuity of the breast Methods A retrospective review of image-guided biopsies performed over
lesions is improved and no lesions identified at 1.5T were missed at 3T MRI. an 11-year period from January 1999 to December 2009 was undertaken.
3T MRI can be used safely in clinical practice. We identified cases with a 14-gauge core biopsy diagnosis of benign
papilloma. Patients with atypia or malignancy on core biopsy were excluded.
The imaging features and number of core samples were documented. All
P19 patients subsequently underwent lesion excision.
Effect of region of interest size in quantitative diffusion-weighted Results Seventy-eight patients had a diagnosis of a benign papilloma on
magnetic resonance imaging of the breast core biopsy. Subsequent excision was with vacuum-assisted biopsy in 48 and
N AlRashidi, T Gagliardi, T Ahearn, T Redpath, F Gilbert surgery in 30. Twenty-eight patients with microcalcification had a stereo-
University of Aberdeen, UK tactic-guided biopsy. Fifty patients with a mass had ultrasound-guided
Breast Cancer Research 2010, 12(Suppl 3):P19 (doi: 10.1186/bcr2672) biopsy. Atypical ductal hyperplasia was found in three out of 28 (11%) who
had microcalcification (mean number of 10 cores) and one out of 50 (2%)
Introduction In breast MRI, morphological and dynamic enhancement with a mass (mean number of three cores). Seventy-four (95%) patients had
features determine whether a lesion is benign or malignant but specificity a benign papilloma only.
is low. Diffusion-weighted magnetic resonance imaging (DW-MRI) measures Conclusions The NPV for atypia and malignancy of 14-gauge core biopsy of
microscopic motion of water and gives quantitative measurement known papillomas is 95% in this series. Underestimation of disease is more common in
as the apparent diffusion coefficient (ADC). This study was conducted to lesions presenting with microcalcification. The current practice of wider sampling
determine whether the whole of the lesion should be included within the of all papillary lesions diagnosed on 14-gauge core biopsy should continue.
region of interest (ROI) or whether a small ROI would differentiate benign
from malignant disease.
Methods Fifteen female patients with 15 suspicious lesions were imaged on P21
a 3T MRI machine (Philips HealthCare, Best, the Netherlands). DWI-MRI was Educational abstract
performed with b-values of 0, 150, 800 s/mm2 using single-shot SE-EPI (TR/ Educational abstract not submitted for online publication.
TE = 9,543 ms/50 ms). The ROI of the lesion and of fibroglandular tissue was Breast Cancer Research 2010, 12(Suppl 3):P21 (doi: 10.1186/bcr2674)
used to calculate ADC values. Histology or follow-up data were available for
all lesions.
Results The mean ADC value of malignant lesions (13) from two small ROIs P22
was 0.954 ± 0.145 mm2/second and for benign (2) was 1.69 ± 0.17 mm2/ Educational abstract
second (Figure 1a). The ADC values for the whole lesion were 1.027 ± Educational abstract not submitted for online publication.
0.23 mm2/second and 1.78 ± 0.293 mm2/second, respectively (Figure 1b). Breast Cancer Research 2010, 12(Suppl 3):P22 (doi: 10.1186/bcr2675)
Conclusions There is a significant difference between ADC values from large
and small ROIs (P <0.05), with small ROIs giving greater differentiation. DWI is
a promising technique to improve specificity of breast MRI.
Figure 1 (abstract P19). Box plot of ADC values for benign, malignant and normal tissue using (a) two small ROIs and (b) a large ROI.
Breast Cancer Research 2010, Volume 12 Suppl 3 S8
http://breast-cancer-research.com/supplements/12/S3
P23 P25
Axillary lymph node fine needle aspiration in breast cancer staging: Educational abstract
diagnostic impact of a second 20G spinal needle Educational abstract not submitted for online publication.
A Leaver1, M Cockburn1, J Gholkar1, CE Holmes1, N Sibal1, B Kaye1, L McLean1, V Breast Cancer Research 2010, 12(Suppl 3):P25 (doi: 10.1186/bcr2678)
Wadehra2
1
Royal Victoria Infirmary Breast Screening Unit, Newcastle-upon-Tyne, UK; 2Royal
Victoria Infirmary Cytology Department, Newcastle-upon-Tyne, UK P26
Breast Cancer Research 2010, 12(Suppl 3):P23 (doi: 10.1186/bcr2676) Educational abstract
Educational abstract not submitted for online publication.
Introduction All breast cancer patients in our centre undergo ipsilateral Breast Cancer Research 2010, 12(Suppl 3):P26 (doi: 10.1186/bcr2679)
axillary ultrasound, followed by fine needle aspiration (FNA) where
appropriate in line with NICE guidelines. We use two needle passes into a
suspicious node, without suction. Improved preoperative detection of nodal P27
metastasis allows patients improved triage to appropriate axillary surgery. Effect of the introduction of preoperative MRI scans for lobular cancer in
We present our analysis of the impact of the second needle in our axillary an individual breast unit
FNA procedure. H Cain, J Sinclair, A Martin, T Fasih, D Browell, K Clark, L Lunt, A Redman
Methods All breast cancer patients undergoing axillary FNA from April 2010 Queen Elizabeth Hospital, Gateshead, UK
to July 2010 were included, where possible. The first FNA was labelled ‘1’ and Breast Cancer Research 2010, 12(Suppl 3):P27 (doi: 10.1186/bcr2680)
the second ‘2’. Individual and overall FNA results were compared with final
surgical pathology, where available. Introduction NICE guidelines suggest that patients with lobular breast
Results The study included 27 female patients. There was a difference in cancer should be offered an MRI scan to measure the tumour size and to
the cytology grading (described LN0 to LN5) allocated between the first exclude multifocality and contralateral tumours. The aim of this study was to
and second needle in five cases (19%). The second needle increased sample ascertain the effect of this guideline.
adequacy on three (11%) occasions. Of the 17 patients that had axillary Methods This guideline was introduced in January 2009. Patients with
surgery, three had no lymph node metastases and the preoperative FNA lobular cancer were selected from before and after this date. Cases were
cytology was LN2. There were no false positive cytology results. Preoperative reviewed and compared for type of surgery, positive resection margins and
cytology was LN5 overall in 11/14 (79%) patients with nodal metastases, with alteration in patient management. The number of additional targeted USS
LN5 obtained only in the second pass in 4/14 (29%) cases (first-pass results: and biopsies was also recorded.
LN2/LN2/LN0/LN0). No complications were reported. Results Sixty-nine patents were included in the study, of which 22 had
Conclusions A second needle pass into suspicious axillary lymph nodes preoperative MRI scans. There was no significant difference in mastectomy
in breast cancer patients has been validated by this study, increasing our rates (MRI = 45.5% vs. no MRI 57.5%, P = 0.44) and no significant difference
preoperative rate of detection of lymph node metastases. in positive margins following WLE (MRI 41% vs. no MRI 35%, P = 0.23). Of
the 22 MRIs 11 additional findings were reported, six in the contralateral
breast, leading to nine targeted USS and eight further core biopsies. Three
P24 of these core biopsies confirmed malignancy. Two MRI scans demonstrated
Fine needle aspiration versus touch imprint cytology in ultrasound- multifocality and one diagnosed contralateral DCIS. Four patients’
guided core of breast masses management was altered due to the MRI result, one of these was due to an
A Leaver1, S Athey1, A O’Brien1, AJ Potterton1, A Redman1, L Lunt1, JD Hemming2 increase in size. There was one case of multifocality that was invisible on all
1
Queen Elizabeth Hospital Breast Screening Unit, Gateshead, UK; 2Queen Elizabeth imaging.
Hospital Cytology Department, Gateshead, UK Conclusions This study has shown that the guideline for preoperative
Breast Cancer Research 2010, 12(Suppl 3):P24 (doi: 10.1186/bcr2677) MRI scanning in lobular cancer will alter the surgical management in
approximately 20% of patients. An additional 20% of patients will undergo
Introduction The aim of this study was to compare the results of fine additional imaging and biopsies that do not alter management.
needle aspiration (FNAC, current practice) with imprint cytology (CIC) from
ultrasound-guided cores taken in the one-stop clinic setting. CIC allows
same-day results with one biopsy procedure. Literature suggests CIC provides P28
a higher sensitivity/C5 rate, although there are few direct comparisons of CIC Educational abstract
with FNAC. Educational abstract not submitted for online publication.
Methods From October 2009 to April 2010, wherever possible, CIC was Breast Cancer Research 2010, 12(Suppl 3):P28 (doi: 10.1186/bcr2681)
performed in patients undergoing both ultrasound-guided FNAC and
core biopsy. CIC slides were independently reported blind to FNA results.
Results were compared with core biopsy histology and therapeutic excision P29
histology, when available. An audit to assess the impact of introducing a protocol of bilateral whole
Results The cases included 90 female patients with 93 masses (54 malignant, breast and axillary ultrasound for assessment of screen-detected cancers
39 benign masses based on final histology). In the malignant group the IT Perera1, LS Wilkinson2
1
C5 rate was 61% (33/54) for FNAC, 83% (45/54) for CIC and 85% (46/54) for Guy’s and St Thomas’ Hospitals NHS Trust, London, UK; 2St George’s Hospital,
the combination of FNA and CIC. No cancers in this sample had an initially London, UK
benign core result with malignant FNA or CIC result. In the benign group the Breast Cancer Research 2010, 12(Suppl 3):P29 (doi: 10.1186/bcr2682)
C2 rate was 41% (16/39) for FNA, 69% (27/39) for CIC, and 64% (25/39) for the
combination of FNA and CIC. There were no false positive C5 results. Imprints Introduction: Our protocol for assessment of women with screen-detected
were easy to perform and there was no damage to the core biopsy material. malignancy was changed to include bilateral whole breast and axillary
Cytologists encountered no problems interpreting CIC. ultrasound (BBUS) following a prospective study that confirmed the benefit
Conclusions The use of CIC in ultrasound-guided core biopsies in place of of this in 2002 [1]. This audit assesses the impact of introducing this change.
conventional FNAC has been validated by this study. No significant obstacle Methods Biopsy results for all women with screen-detected cancer diagnosed
to adopting core imprint cytology has been identified. between April 2003 and March 2009 were reviewed to identify cases where
multiple biopsies had been performed. The reason for additional biopsy and
subsequent management were recorded. The data were compared with
control data (2001) obtained prior to the introduction of the protocol, and
with national outcome data for screen-detected cancer.
Results A total of 199,307 women were screened during the audit period,
and 1,700 women were diagnosed with breast cancer. Table 1 demonstrates
the findings for women diagnosed with cancer who had additional biopsies,
Breast Cancer Research 2010, Volume 12 Suppl 3 S9
http://breast-cancer-research.com/supplements/12/S3
P30
Does a second image-guided biopsy contribute to the management of P33
B3 lesions? Current clinical use of the [18F]FDG PET/CT in breast cancer patients: an
JK Juttla1, W Teh2 audit of local referral patterns
1
North London Breast Screening Service, Edgware, UK; 2Northwick Park Hospital, NJ Laurence, J Searle, J Bristol, ID Lyburn
Harrow, UK Cheltenham General Hospital, Cheltenham, UK
Breast Cancer Research 2010, 12(Suppl 3):P30 (doi: 10.1186/bcr2683) Breast Cancer Research 2010, 12(Suppl 3):P33 (doi: 10.1186/bcr2686)
Introduction B3 (benign with uncertain malignant potential) lesions such Introduction [18F]FDG PET/CT scanning has been shown to be an established,
as atypias and papillary lesions have final surgical outcome of malignancy accurate method of detecting many primary and metastatic tumours.
in about 20% of vacuum-assisted and 45% of image-guided core biopsies. However, widespread use of PET/CT scanning in the UK remains generally
We examine the benefits of undertaking a second image-guided biopsy in limited, as it is expensive and restricted to a few tertiary centres. Cheltenham
these cases. General Hospital receives referrals for patients from across the southwest of
Methods A retrospective study of all B3 lesions diagnosed between January England and Wales for [18F]FDG PET/CT. We looked at the referral patterns of
2008 and December 2009 at the North London Breast Screening Service. patients with breast cancer to see whether they reflected current indications
Results Out of 235 biopsies in 167 women over a 2-year period, there were for [18F]FDG PET/CT and to see whether the findings impacted on clinical
191 B3 outcomes of which 59 had a final malignant outcome (30.8%). decision-making.
Following MDT discussion, 60/191 B3 lesions were subjected to a second Methods Ninety-five [18F]FDG PET/CT scans from 2006 to 2010 were
biopsy (25.5%) and one lesion a third biopsy. These yielded a malignant requested for breast cancer patients. The indications and results of these
diagnosis in 11 (18.3%) cases. Lesions successfully upgraded to malignancy were recorded and analysed.
included three microcalcification (MCC), three masses, three masses with Results The following indications were found – staging scans in patients
MCC and one cyst with solid elements. Cancers diagnosed on second with locoregional disease: 22%; equivocal radiology: 17%; specific concern of
biopsies were one invasive ductal, one invasive papillary carcinoma and nine recurrent disease: 17%; diagnostic/therapeutic conflict with second primary:
ductal carcinoma in situ (DCIS). All second biopsies were performed using a 15%; staging of locally advanced disease: 11%; response to endocrine or
vacuum device with the initial biopsy method being 14G core biopsy in eight chemotherapy: 9%; stability/speed of progression of metastatic disease:
cases and vacuum-assisted in three cases. Second biopsies were therefore 8%. On reviewing clinical records/follow-up, the results of [18F]FDG PET/
equally useful in upgrading MCC and masses and especially useful for DCIS. CT in many circumstances gave information about metastatic disease in
Conclusions Second-time image-guided biopsies can successfully yield locoregional recurrence, resolved issues surrounding equivocal radiology
a definitive malignant diagnosis in B3 lesions (18.3%) and thus enable and ensured that patients with second operable cancers were not subjected
therapeutic rather than diagnostic surgery to be undertaken. Second to major surgery only to succumb to their breast metastasis.
biopsies are equally useful in the management of MCC, masses and masses Conclusions [18F]FDG PET/CT can give valuable information in patients with
with MCC. breast cancer, particularly in scenarios with equivocal findings on other
imaging and regarding disease progression. The technique often contributed
to management decisions in complex cases.
P31
Educational abstract
Educational abstract not submitted for online publication.
Breast Cancer Research 2010, 12(Suppl 3):P31 (doi: 10.1186/bcr2684)
Breast Cancer Research 2010, Volume 12 Suppl 3 S10
http://breast-cancer-research.com/supplements/12/S3
P34 P37
Incident cancer cases: what can we learn from the previous screening Role of imaging in gynaecomastia: results of a Royal College of
round? Radiologists Breast Group Annual Scientific Meeting 2009 survey
JM Scarth, G Markham, S Kulaveerasingam A Leaver, A Yeung, A Redman
North London Breast Screening Unit, London, UK Queen Elizabeth Hospital Breast Screening Unit, Gateshead, UK
Breast Cancer Research 2010, 12(Suppl 3):P34 (doi: 10.1186/bcr2687) Breast Cancer Research 2010, 12(Suppl 3):P37 (doi: 10.1186/bcr2690)
Methods A retrospective analysis of all incident cancers with their previous Introduction Gynaecomastia is a benign enlargement of male breast tissue
screening round mammograms diagnosed between April 2006 and March that has no proven excess risk of breast cancer. Clinical evaluation is required
2008 at the North London Breast Screening unit was performed. to exclude breast cancer, but the role of imaging in the male breast is unclear.
Results Two hundred and forty cancers were reviewed. In 187 cases (77.9%) Our aim was to determine practice in other units, review literature and
the previous round mammograms were normal (group A), in 53 cases formulate informed and realistic departmental guidelines.
(22.1%) an abnormality was detected (group B). Of these, five (9.4%) were Methods A questionnaire was created, and copies placed on delegates’ seats.
classified as normal/benign, 40 (75.5%) were uncertain and eight (15.1%) Descriptive statistics applied.
were suspicious. There was no significant difference in the size of the lesions Results Of approximately 160 delegates, 90 questionnaires were returned
between the two groups; there was, however, a significant increase in size from at least 58 different units. Delegates reported (estimated) greater than
in the lesions in group B on the subsequent mammograms (P <0.0001). Of 864.75 years collective experience, each performing (on average) 4.3 male
the lesions in group B, 25 (47%) of the cases had microcalcifications only on breast assessments per month. Sixty-five per cent (58/89) of delegates that
the previous mammograms; this is higher than previously published data responded reported routine imaging in clinically typical gynaecomastia,
of 27% and 17% [1,2] with all of these cases being subsequently diagnosed rising to 89% (79/89) in clinically typical unilateral gynaecomastia. However,
as either DCIS or IDC. Group A had less non-invasive (53, 29%) and grade 3 78% (68/87) of responding delegates agreed with the statement ‘Imaging is
tumours (26, 20%) compared with group B (non-invasive = 18, 34%; grade 3 = not necessary in cases of clinically typical gynaecomastia’. Three delegates
9, 29%) and more grade 1 tumours (51, 39% vs. 8, 26%) with a similar amount reported experiencing a case of ultrasonically typical gynaecomastia that
of grade 2 tumours between the groups (41% and 45%). subsequently proved to be breast cancer. Imaging protocol and biopsy
Conclusions In 22% of incident cancer cases an abnormality is present on the practice varied greatly between units. Delegates volunteered concerns that
previous screening round mammogram, and the most frequently overlooked the NHS breast service is inequitable between males and females, and that
lesions are microcalcifications. guidelines were needed to rationalise imaging.
References Conclusions Based upon questionnaire findings and literature review,
1. Maxwell AJ, et al.: Breast 2001, 10:392-398. guidelines have been now been drawn up in our unit. There is great variation
2. O’Flynn EAM, et al.: Breast Cancer Res 2009, 11(Suppl 2):P8. evident in UK gynaecomastia imaging practice. National discussion and
agreement on evidence-based guidelines could help rationalize use of
precious NHS resources and reduce imaging of this benign condition.
P35
Vacuum-assisted core biopsy of B3 lesions showing atypia on needle core
biopsy: a worthwhile exercise? P38
M Al-Attar, S Tennant, E Denton, H Khan, L Grosvenor, D Lister Malignant microcalcification: prediction of excision margins by
University Hospitals of Leicester, UK separating calcified and noncalcified core biopsies
Breast Cancer Research 2010, 12(Suppl 3):P35 (doi: 10.1186/bcr2688) H Cain, A Mooney, J Westgarth, T Fasih, D Browell, K Clark, D Hemming, L Lunt
Queen Elizabeth Hospital, Gateshead, UK
Introduction There has been increasing interest in the use of vacuum- Breast Cancer Research 2010, 12(Suppl 3):P38 (doi: 10.1186/bcr2691)
assisted core biopsy (VACB) over the past two decades. There remains some
uncertainty about its role in the presence of cytological or architectural Introduction Suspicious microcalcification is investigated by vacuum-
atypia on needle core biopsy (NCB). We have recently been offering VACB assisted core biopsy. Many patients, in whom malignancy is diagnosed, go
for selected B3 cases with atypia. MDT discussion, where technical suitability on to have breast-conserving surgery where positive resection margins
and potential value of VACB is debated, is used to select appropriate cases. require re-operation. The aim of this study was to determine whether there is
Methods A retrospective case review of all B3 lesions diagnosed at NCB an increased risk of positive margins when malignancy is identified in cores
between 1 March 2008 and 1 March 2010. without microcalcification.
Results A total of 166 B3 lesions were diagnosed: 123 underwent surgical Methods In this institute, core biopsy samples are separated into those
biopsy, of which 30 were malignant (eight invasive, 22 non invasive), 10 had containing calcification and those not. Sixty-eight consecutive patients
LCIS and 83 were benign. Forty-three B3 lesions underwent VACB: 18/43 undergoing breast-conserving surgery were selected. From the histology
lesions had shown atypia at NCB and following VACB, seven were upgraded report, the presence of malignancy in each group of cores was recorded.
to DCIS, three were downgraded to B2, six remained as B3 with atypia The adequacy of margins following the local excision and the need for re-
(5/6 had surgery – three malignant, two benign). VACB failed in two cases. operation was then analysed.
Twenty-five out of 43 cases had shown no atypia at NCB. One out of 25 was Results Sixty-eight patients were included in the study. The mean age at
upgraded to B4 on VACB, but no malignancy was seen at surgery. Twenty- diagnosis was 56.7 (range 40 to 77). All 68 patients had malignancy in the
four out of 25 cases were downgraded to B2. cores containing calcification. Twenty-three (34%) patients had malignancy
Conclusions Our results show a potential benefit of this technique in in these cores alone and nine (39%) of these patients required further
selected cases. We were able to upgrade 39% of B3 lesions with atypia, surgery. Forty-five patients had malignancy in both of the sets of cores, 24
thereby allowing immediate definitive surgery. We were able to avoid (53%) required further surgery for inadequate margins. Although there is a
surgical excision in 3/18 (17%) of B3 lesions with atypia. difference between the re-operation rate, this is not statistically significant
(P = 0.3124, Fisher’s exact test).
Conclusions Malignancy is commonly diagnosed in cores that do not
P36 contain microcalcification. Although there appears to be a higher rate of
Educational abstract inadequate margins in this group (53% vs. 39%), in this study the difference
Educational abstract not submitted for online publication. was not statistically significant. Following the findings of this initial study, we
Breast Cancer Research 2010, 12(Suppl 3):P36 (doi: 10.1186/bcr2689) are now increasing the sample size of the group.
Breast Cancer Research 2010, Volume 12 Suppl 3 S11
http://breast-cancer-research.com/supplements/12/S3
P39 P41
Initial experience with MicroPure™, a new ultrasound image processing Is digital better?
function to improve calcification visualisation AA Duncan1, ME Wheaton1, N Rogers2, O Kearins2
1
P Sankaye, G Porter, S Doyle, S Steel Warwickshire, Solihull and Coventry Breast Screening Unit, Coventry, UK; 2West
Primrose Breast Unit, Derriford Hospital, Plymouth, UK Midlands QA Reference Centre, Birmingham, UK
Breast Cancer Research 2010, 12(Suppl 3):P39 (doi: 10.1186/bcr2692) Breast Cancer Research 2010, 12(Suppl 3):P41 (doi: 10.1186/bcr2694)
Introduction Mammographic calcification may be the only sign of breast Introduction During the changeover from analogue to digital screening in
malignancy often requiring diagnostic biopsy. If visible, this calcification the UK, reassurance is needed to confirm that the outcomes with digital are
may be biopsied using ultrasound guidance or using X-ray guidance if equal to or better than analogue screening.
not. Ultrasound biopsy is quicker, cheaper and more comfortable for the Methods Warwickshire, Solihull & Coventry Breast Screening Service
patient. Any technique that improves ultrasound calcification visualisation commenced the phased conversion to digital screening in 2005, with a single
is desirable. MicroPure™ (Toshiba Medical Systems Corporation, Otawara, digital machine on one of three mobiles. This retrospective study compares
Japan) is a new ultrasound processing technology designed to achieve this. the screening outcomes of 138,173 women aged between 49 and 70 years
Methods We prospectively audited our experience of calcification screened on analogue or digital imaging between April 2005 and March
detection with B-mode ultrasound and MicroPure™ between April and July 2009 inclusive. Approximately one-third of these were screened using digital,
2010. Twenty-five women presenting with a dominant mammographic the remainder on analogue.
abnormality of calcification were studied. Targeted imaging with both Results The results show no difference in the rates of screen-detected cancer
techniques was performed with a Toshiba Aplio™ XG ultrasound machine in prevalent or incident groups between analogue and digital, and rates were
using a 12 MHz linear probe. Technical support to optimise imaging was stable with time. Similar proportions of invasive and non-invasive cancers
provided during our audit by Toshiba. We recorded ultrasound visibility with were detected in both groups and tumour size was not significantly different.
these two techniques and histological diagnosis. No significant increase in cancer detection in younger compared with older
Results Of 24 screening and one symptomatic woman examined, 11 (44%) women was seen in the digital group, and digital did not diagnose a higher
were diagnosed with breast malignancy (six invasive cancers, four DCIS, one proportion of lobular cancers. No difference in interval cancer rates between
LCIS), and 14 (56%) were benign. Overall four (16%) (three malignant and the two methods of screening was seen. On radiological subclassification of
one benign) calcifications were visualised by ultrasound. All were detectable interval cancers into normal/benign, uncertain and suspicious, significantly
using both B-mode and MicroPure™. Subjectively all four were felt to be fewer interval cancers were classified as uncertain in the digital than the
more conspicuous using B-mode than MicroPure™. analogue screening group. See Table 1.
Conclusions Our initial experience has demonstrated MicroPure™ to be
no better at detecting benign or malignant mammographic calcification
than B-mode ultrasound. MicroPure™ would only be useful if it detects Table 1 (abstract P41)
calcifications that are not visualised with B-mode ultrasound therefore Analogue Digital
reducing X-ray-guided biopsies.
n % (95% CI) n % (95% CI)
Normal/benign 97 73.5% (66.0 to 81.0%) 72 85.7% (7.02 to 93.2%)
P40
A retrospective review of triple-negative breast cancer cases: are there Uncertain 20 15.2% (9.0 to 21.3%) 3 3.6% (–0.4 to 7.5%)
common features in the clinical presentation, imaging and risk factors? Suspicious 15 11.4% (5.9 to 16.8%) 9 10.7% (4.1 to 17.3%)
N Trevett1, L Edwards2,3, K Gower-Thomas3,4, V Shah2,3
1
Welsh School of Medicine, Cardiff, UK; 2Royal Gwent Hospital, Newport, UK; 3Breast
Test Wales, Cardiff, UK; 4Royal Glamorgan Hospital, Llantrisant, UK
Breast Cancer Research 2010, 12(Suppl 3):P40 (doi: 10.1186/bcr2693) Conclusions These results are reassuring that digital diagnoses similar
cancers to analogue screening, and suggest that digital may allow more
Introduction Triple-negative breast cancer (TNBC) cases comprise definitive interval cancer classification.
approximately 15% of newly diagnosed breast cancers and are associated
with poor prognosis and limited treatment options. In this retrospective
study from South Wales, 81 patients with breast cancer found to be ER, PR P42
and HER2 negative were reviewed to determine whether there are common Variability in film reader estimates of breast density in the PERFORMS
imaging and pathological findings. scheme
Methods Patients identified from pathological databases at two hospitals IT Darker, AG Gale, Y Chen
included symptomatic and screening cases. Clinical records were reviewed to Loughborough University, Loughborough, UK
determine age at diagnosis, family history and clinical findings; pathological Breast Cancer Research 2010, 12(Suppl 3):P42 (doi: 10.1186/bcr2695)
reports to identify size, grade, type and nodal status; and imaging studies to
determine breast density, lesion type, classification and size. Introduction All UK screening personnel are invited to take part annually
Results Eighty per cent of patients were aged over 50 years (range 50 to in the PERFORMS self-assessment scheme where they make several
89 years). Nineteen per cent reported a family history. Eighty-nine per cent judgements about series of challenging recent screening cases. As part of
were grade 3 tumours, 89% were ductal type of which 30% had associated this process they assess the density of each case. Density is a factor known
DCIS. The majority were large tumours (78% over 20 mm). Fifty-two per to be associated with a greater risk of developing breast cancer and thus
cent were node positive (20% had more than four nodes positive). Thirty- accurate density judgements may well presage the facility to proffer
nine per cent had associated vascular invasion. Calcification was a dominant improved follow-up for individual women. The present study examines the
mammographic feature (37%), 28% had well-defined masses. degree of variability amongst film reader estimates of breast density on a
Conclusions Previous studies have found well-defined masses to be a large number of cases.
dominant imaging feature but this study has found malignant calcification to Methods Data were examined from the most recent 2 years of the PERFORMS
be more common. The tumours were mainly large and frequently associated scheme where breast density estimates were made for each case examined
with vascular invasion, possibly contributing to the poor prognosis despite using a three-point rating scale of fatty, mixed density, and dense. These data
being node negative in nearly one-half of the cases. There appears to be an comprised information from 444 individuals (mainly consultant radiologists,
association with family history and to be common in the over 50s, contrary advanced practitioners and breast physicians) who had all examined the
to current thinking. A national prospectively collected database TNBC could same 240 difficult cases.
aid understanding of this group. Results The inter-rater reliability, corrected for chance agreements, was
assessed using kappa. Overall, the degree of agreement across cases
on breast density category was significantly greater than no agreement
(P <0.0001). However, only a moderate degree of inter-rater reliability was
Breast Cancer Research 2010, Volume 12 Suppl 3 S12
http://breast-cancer-research.com/supplements/12/S3
exhibited, κ = 0.470. There were significant differences between the levels of P46
agreement amongst the ratings of the radiologists, advanced practitioners Breast histoscanning: the development of a novel technique to improve
and others (all P <0.05). tissue characterization during breast ultrasound
Conclusions The low agreement rates between participants for density LS Wilkinson, C Coleman, CM Pagliari, P Skippage, V Thomas, R Given-Wilson
ratings were surprising. That there were differences between the occupa- St George’s Healthcare NHS Trust, London, UK
tional groupings may reflect breast screening experience. Breast Cancer Research 2010, 12(Suppl 3):P46 (doi: 10.1186/bcr2699)
P48 P51
Benefits of CT-angiography localisation in the surgical planning of deep Educational abstract
inferior epigastric perforator flap breast reconstruction Educational abstract not submitted for online publication.
NC Chhaya1, P Sarbeng1, S Stuart1, F Angullia2, A Mosahebi2, A Malhotra1 Breast Cancer Research 2010, 12(Suppl 3):P51 (doi: 10.1186/bcr2704)
1
Department of Radiology, Royal Free Hospital, London, UK; 2Department of Plastic
Surgery, Royal Free Hospital, London, UK
Breast Cancer Research 2010, 12(Suppl 3):P48 (doi: 10.1186/bcr2701) P52
Benign solitary breast masses in the prevalent screening round: do they
Introduction The aim of breast reconstruction in the postmastectomy cancer contribute to a high recall rate?
patient is to restore breast contour and dimensions whilst minimising the N Hoskins1, L Edwards2, K Gower-Thomas2, V Shah2
1
cosmetic impact. The deep inferior epigastric perforator (DIEP)-flap is a Welsh School of Medicine, Cardiff, UK; 2Breast Test Wales, Cardiff, UK
complex but state-of-the-art procedure that provides a durable and natural Breast Cancer Research 2010, 12(Suppl 3):P52 (doi: 10.1186/bcr2705)
result. It is rapidly becoming the preferred choice at many institutions,
including our regional plastic surgery unit. In achieving superior cosmesis Background The Breast Test Wales prevalent round recall rate between 2003
without sacrificing abdominal wall musculature, a successful DIEP-flap and 2006 was 9.07% (above the NHSBSP target of 7%) and remains high.
requires painstaking and time-consuming microdissection of the inferior This study was based on the hypothesis that recall of benign solitary masses
epigastric artery perforators. This relies on a high level of surgical expertise might be a major contributor to this as no prior imaging is available.
and results in prolonged anaesthetic times. As greater volumes of breast Methods Prospectively collected data from Breast Test Wales (South East
reconstructive surgery are performed, there will be increasing requirements Wales) identified all prevalent screens in a 3-year cycle recalled for a benign
for such preoperative imaging. We describe the required optimisation of CTA mass lesion confirmed by core biopsy. All women attended a subsequent
protocols to obtain the pertinent information and demonstrate how best to screen and remained free of cancer. Mammograms were retrospectively
convey this complex information to our surgical colleagues. reviewed and the lesions were re-evaluated by applying criteria typical of
Methods Since 2009 we have provided CT-angiography in the preoperative a benign mass.
planning for DIEP-flap breast reconstruction. We explored the implications of Results A total of 2,322 women following a prevalent screen were recalled;
CT-angiography to optimise localisation of arterial perforators and identified 2,069 were returned to routine recall without biopsy (cysts are included in
the benefits of this imaging-guided approach. this group), 186 were diagnosed with cancer and 105 had a benign biopsy
Results A total of 60 female patients have benefitted from CTA-guided where mammography had been considered benign or probably benign.
perforator localisation, providing valuable procedural-planning information The benign to malignant biopsy ratio was 1:1.8. A total 46.6% (n = 49)
to our surgical colleagues. We have shown benefits in terms of markedly lesions on retrospective review of mammography showed typical benign
shorter operative duration with consequently reduced hospital stays and characteristics. Dense breast composition and overlying glandular tissue
morbidity. Two patients had unsuspected metastatic disease identified, were noted to correlate with higher rates of retrospectively indicated recall.
precluding reconstructive surgery. Conclusions Stricter adherence to applying classification of benign solitary
Conclusions An imaging-guided approach optimises preoperative planning. lesions could reduce the recall rate and decrease the psychological distress
Accurate identification of arterial perforators enables targeted intraoperative for these women without adversely compromising the cancer detection rate.
localisation. This results in decreased operative time and patient morbidity, The impact on the overall recall rate would be small but would significantly
providing benefits for the cost of healthcare provision. improve the benign to malignant biopsy ratio. The issue of breast density
and overlying tissue may be resolved with the advent of digital applications
P49 such as tomosynthesis.
Interval breast cancers: positive influencing behaviour
A Kumar, G Penn, D Meacock, H Yeang
Wirral University Teaching Hospital, Liverpool, UK P53
Breast Cancer Research 2010, 12(Suppl 3):P49 (doi: 10.1186/bcr2702) Breast MRI screening for high-risk family history: the Sheffield experience
AV Genever, E Surgey, AL Ford, O Hatsiopoulou, CE Ingram, IJ Jolley
Introduction Studies estimate that 10 to 32% of breast cancers are over- Breast Unit, Sheffield Teaching Hospitals NHS Trust, Sheffield, UK
looked in mammograms. One of the important causes of interval breast Breast Cancer Research 2010, 12(Suppl 3):P53 (doi: 10.1186/bcr2706)
cancer is fatigue and loss of concentration. The likelihood of fatigue increases
with the duration of a reporting session and errors are more likely to occur Introduction Around 5% of breast cancers can be attributed to gene
towards the end of the session. The purpose of this talk is to address the lack mutations. NICE guidelines have recently advocated the use of MRI screening
of awareness of the issues that drive reporting performance. in high-risk young women. We have retrospectively audited our unit’s
Methods A retrospective study was carried out on interval breast cancers experience in this field.
from a 2-year period. This identified 90 histopathologically proven interval Methods All eligible women were identified from the family history database.
breast cancers. Each interval cancer mammogram was reviewed by two Notes, imaging and pathology were reviewed.
blinded consultant radiologists and placed into one of three categories on Results One hundred and thirty-three breast MRI scans were performed
retrospective radiological review: Category 1: normal; Category 2: uncertain; on 91 women with a high-risk family history between 2007 and 2010.
Category 3: suspicious. Each case was then analysed further to correlate the Sixteen women were recalled for assessment (one woman was recalled
interval categories with their position on the mammography roller viewer. twice). The total recall rate was 12.7%. Of the recalled patients, four had
Results Of the 90 interval breast cancers, 59 (66%) were Category 1, 14 (16%) normal ultrasound (US) and follow-up imaging has remained unchanged.
were Category 2 and 17 (18%) were Category 3. Statistical analyses with Thirteen patients had corresponding US-detected abnormalities. Twelve
one-way ANOVA test revealed the presence of clinical significance between were biopsied, the other was a normal intramammary lymph node. Eight of
Category 3 cases and their position in the roller viewer (P <0.021). the biopsies were benign (benign core biopsy rate 6%). Four biopsies were
Conclusions Fatigue is identified as an important factor responsible for malignant (age range of women 35 to 45), giving a cancer detection rate
missing breast cancer. Aside from making changes in double-reporting of 3%. Three of these were solitary lesions (8 mm, 11 mm and 16 mm). One
techniques, human performance factors such as nutrition, rest, behaviour patient had multifocal malignancy, the largest single lesion being 16 mm.
prior to reporting and environment have to be considered. A national All were node-negative ductal carcinoma. Two patients were oestrogen
programme has to be implemented to facilitate a plan to raise awareness of receptor-positive, all were herceptin receptor-negative. Only the extensive
these factors in the NHS culture. malignant change could be seen on conventional mammography.
Conclusions We suggest that MRI screening is beneficial in these patients,
P50 and although our recall rate lies a little above what is to be recommended
Educational abstract by the NHSBSP (7 to 10%) we feel this can be attributed to the steep learning
Educational abstract not submitted for online publication. curve that introducing a new screening technique to a service invariably
Breast Cancer Research 2010, 12(Suppl 3):P50 (doi: 10.1186/bcr2703) brings.
Breast Cancer Research 2010, Volume 12 Suppl 3 S14
http://breast-cancer-research.com/supplements/12/S3
P56
Educational abstract P59
Educational abstract not submitted for online publication. Educational abstract
Breast Cancer Research 2010, 12(Suppl 3):P56 (doi: 10.1186/bcr2709) Educational abstract not submitted for online publication.
Breast Cancer Research 2010, 12(Suppl 3):P59 (doi: 10.1186/bcr2712)
P57
What is involved in a comprehensive breast MRI service? Implications for P60
service provision Breast cancer screening in the over 70s
ND Forester, N Sharma, BJG Dall D Eddy1, L Edwards2, K Gower-Thomas2
1
Leeds Teaching Hospitals Trust, Leeds, UK Welsh School of Medicine, Cardiff, UK; 2Breast Test Wales, Cardiff, UK
Breast Cancer Research 2010, 12(Suppl 3):P57 (doi: 10.1186/bcr2710) Breast Cancer Research 2010, 12(Suppl 3):P60 (doi: 10.1186/bcr2713)
Introduction MRI as an adjunct to triple assessment is well established, Introduction Routine invitation for breast screening currently ceases in Wales
but often identifies additional lesions within the breast necessitating over the age of 70 despite the incidence of breast cancer remaining high in
further characterisation. This study addresses the additional investigations this group. Attendance for screening aged over 70 is low compared with the
generated by MRI. invited population, often attributed to lack of awareness of eligibility and
Methods A retrospective review of the MRI database between 2006 and how to access screening. If screening is to be extended, the motivation to
2009 identifying patients requiring investigations post MRI. attend and outcomes need to be understood.
Results Over 4 years, 1,119 MRIs were performed on 717 patients, with 102 Methods This audit of prospectively collected data from Breast Test Wales
recalled for second-look ultrasound. Three patients were lost to follow-up. A Screening (South East Wales) identified all women over 70 years who
total 124 incidental lesions were identified on MRI. Ultrasound identified 68 attended. In those where cancers were detected, age, pathological data,
lesions, with definitive diagnosis following core biopsy/FNAC (62), surgery previous screening history, family history, clinically palpability and breast
(two) core biopsy + repeat MRI (one) and MRI biopsy (three). Twenty-two symptoms were recorded.
lesions not identified by USS were assessed with X-ray-guided biopsy Results A total of 5,736 women attended aged 71 to 92 years in a 3-year
(two), MRI biopsy (13), interval MRI (five) or surgery (two). Nineteen MRI screening round (1 April 2006 to 31 March 2009) with numbers attending
lesions following normal ultrasound had routine follow-up. Fifteen lesions decreasing with higher age. In total, 295 (5.1%) were recalled to assessment
(12 patients) did not have follow-up USS, as recommended following and 81 (1.4%) were diagnosed with cancer. A total of 61.5% of the cancers
MDT discussion. Sixty lesions (44 patients) were malignant; MRI identified were grade 2, 77% <20 mm and 84% node-negative. The majority had an
a primary in four patients presenting with lymphadenopathy and in 39 excellent or good Nottingham Prognostic Index, mirroring national data for
patients identified additional foci that changed management. Malignant the younger invited population. Of those diagnosed with cancer, 61.6% had
Breast Cancer Research 2010, Volume 12 Suppl 3 S15
http://breast-cancer-research.com/supplements/12/S3
four or more previous screens; 28.9% of women reported a family history; be assessed via sentinel lymph node procedures (SLNP) [1]. This can be
55.4% were clinically palpable and of these 49% had themselves noticed a time consuming, however, and it would be useful to identify a subset of
change in the breast. quantitative nodal features on ultrasound in order to predict metastatic
Conclusions This study indicates that women continue to attend regularly involvement and avoid the SLNP. Peer review states that cortical thickness is
after invitation ceases. Clinical concerns and family history awareness may be one of the best predictive characteristics [1,2]. We wanted to interrogate our
motivators but as cancers detected have good prognostic indices, screening own data with a specific focus on this feature.
this group may be beneficial. Methods A retrospective study of 454 patients audited between August
2007 and October 2009. One hundred and thirty-seven had proven breast
cancer and underwent axillary node sampling.
P61 Results In 75 cases the cortical thickness was recorded. Comparing with
Are there features on imaging or core biopsy that can predict tumour- sentinel node biopsy or postoperative pathology, the results are as shown
positive margins after breast-conserving surgery? in Figure 1.
N Hill1, L Edwards2, IJ Monypenny2, K Gower-Thomas2 Conclusions The study confirms that the nodal cortical thickness correlates
1
Welsh School of Medicine, Cardiff, UK; 2Breast Test Wales, Cardiff, UK well with the presence of disease. It is easy to measure and appears to be
Breast Cancer Research 2010, 12(Suppl 3):P61 (doi: 10.1186/bcr2714) a reliable indicator. Further, the minimum cortical thickness for positivity
(27 mm) will help us to grade our degree of suspicion in future.
Introduction Twenty-two per cent of women required further surgery References
following breast-conserving surgery (BCS) in the UK in 2008 because of 1. Kuenen-Boumeester V, Menke-Pluymers M, de Kanter A, et al.: Ultrasound
involved margins. This study was carried out to see whether it was possible to guided fine needle aspiration cytology of axillary lymph nodes in breast
predict the likelihood of this from preoperative information and thus reduce cancer patients. A pre-operative staging procedure. Eur J Cancer 2003,
the re-excision rate. 39:170-174.
Methods Women having BCS following a preoperative diagnosis of invasive 2. Damera A, Evans AJ, Cornford EJ, et al.: Diagnosis of axillary nodal
cancer were identified from prospectively collected data at Breast Test Wales metastases by ultrasound guided core biopsy in primary operable breast
for South East Wales, over a 3-year screening round (2006 to 2009). In cases cancer. Br J Cancer 2003, 89:1310-1313.
where DCIS was found at or within 2 mm of a margin, preoperative imaging
was reviewed to look for peripheral calcification.
Results One hundred and twelve out of 844 women with an invasive tumour
had involved margins after BCS. Fifty-nine women had DCIS at the margin, of P64
which 60% had calcification within 10 mm of the periphery of the tumour on Breast cancer in 35 to 39 year olds and imaging: is changing to
mammographic review and 30% had DCIS as well as invasive cancer on core ultrasound without mammography going to be safe? What are the
biopsy. The postsurgical pathological size was greater than the mammographic workload implications?
size in 79% and greater than the ultrasound size in over 90% of cases. This was G Rubin, C Zammit
not significantly changed on further review of the films. On mammographic Brighton and Sussex University Hospitals NHS Trust, Brighton, UK
review, 12% had dense breasts, 88% being fatty or fatty/glandular. Breast Cancer Research 2010, 12(Suppl 3):P64 (doi: 10.1186/bcr2717)
Conclusions Imaging is underestimating the true size despite background
breast tissue being fatty. The presence of calcification on the mammogram Introduction New guidelines suggest that ultrasound should replace
in such a high number of cases where DCIS is at the margin may be a useful mammography as the primary imaging test for 35 to 39 year olds in symptomatic
predictor, but requires further correlation with the presence of histological outpatient clinics. We currently use clinical examination, clinically guided fine
calcification on the pathological specimen, which is ongoing. needle aspiration and mammograms for women aged 35+ as initial triple
assessment. We wanted to understand the implications of introducing the
guidelines, based on our current practice.
P62 Methods We reviewed our workload and cancer detection in symptomatic 35
Educational abstract to 39 year olds attending breast outpatients in our district general hospital over
Educational abstract not submitted for online publication. the past 18 months.
Breast Cancer Research 2010, 12(Suppl 3):P62 (doi: 10.1186/bcr2715) Results Mammograms were taken on all symptomatic patients unless pregnant
or breastfeeding. Seven hundred and fourteen patients had mammograms
and 442 patients had ultrasound. This was performed on most patients found
P63 to have clinically palpable lumps, and on anyone with a clinical examination
Ultrasound of the axilla: analysing nodal cortical thickness graded as uncertain suspicious or malignant, or with abnormal mammograms,
PM Riley, S Datta, H Alwan-Walker, S Limdi or with abnormal results on clinically guided needle cytology. Fourteen patients
Salford Royal Foundation Trust, Manchester, UK were found to have breast cancer. All 14 had ultrasounds graded as uncertain,
Breast Cancer Research 2010, 12(Suppl 3):P63 (doi: 10.1186/bcr2716) suspicious or malignant, an indication for diagnostic core biopsy. Thirteen cancer
patients had mammograms. Two were graded as benign and 11 as uncertain or
Introduction Prognosis in breast cancer is dependent upon axillary worse. Five cancer patients had clinically normal or benign breast examinations,
lymph node status. For breast-conserving surgery, lymph node status can including one with a benign mammogram report.