Professional Documents
Culture Documents
CL-900i Series Installation Guidance - V1.0 - EN
CL-900i Series Installation Guidance - V1.0 - EN
CL-900i Series
Chemiluminescence Immunoassay
Analyzer
Installation Guidance
Table of Contents
Table of Contents ................................................................................................................................... 2
Chapter 1 Check before Installation .................................................................................................. 3
1.1 Installation Environment ...................................................................................................... 3
1.2 Space and Accessibility Requirements for Unpacking ................................................... 3
1.2.1 Space and Accessibility Requirements for Installation ....................................... 3
1.2.2 Power Supply and Noise ......................................................................................... 4
1.2.3 Drainage Check (if draining water through a sewer)........................................... 4
1.2.4 Recommended PC Configuration .......................................................................... 4
1.3 Configuration Check ............................................................................................................ 5
Chapter 2 Instrument installation ...................................................................................................... 6
2.1 Tools ....................................................................................................................................... 6
2.2 Installation Procedure .......................................................................................................... 6
2.2.1 Unpacking .................................................................................................................. 6
2.2.2 Fluidic Connection .................................................................................................. 16
2.2.3 Connect the computer............................................................................................ 19
2.2.4 Software Installation ............................................................................................... 19
Chapter 3 Power on and alignment ................................................................................................. 20
3.1 Preparation for Powering On ............................................................................................ 20
3.2 Initial Startup ....................................................................................................................... 20
3.3 Main Unit Position Confirmation and Alignment ............................................................ 21
3.4 Load and Prime Wash Buffer............................................................................................ 24
3.5 Probe Vertical Pos. to Mixing Position Alignment .......................................................... 25
3.6 Substrate Prime .................................................................................................................. 26
3.7 Setting up............................................................................................................................. 31
3.7.1 Load and check the consumables ....................................................................... 31
3.7.2 Importing and Configuring Chemistry Parameters ............................................ 33
Chapter 4 System Performance Test .............................................................................................. 36
4.1 DCF Diagnosis .................................................................................................................... 36
4.2 Substrate Background detection ...................................................................................... 36
4.3 System Repeatability ......................................................................................................... 36
4.4 Dispersion Wash Effects ................................................................................................... 37
4.5 Repeatability Test ............................................................................................................... 37
Chapter 5 LIS and Remote Help ....................................................................................................... 39
5.1 LIS Connection ................................................................................................................... 39
Appendix A Prepare Installation Reagent Pack ........................................................................... 40
Appendix B Comparison of altitude and atmospheric pressure ............................................. 41
Power line
≥700mm
Communication
Waste tube line
Waste tank
2782m
2474m
operation
m
LIS host
UPS
PC for
≥500mm
m
CL-900i ≥350mm
WARNING
Make sure the power socket is grounded correctly. Incorrect grounding may lead to electric shock
or equipment damage.
Check if the power socket outputs voltage meeting the specified requirements and has a proper
fuse installed.
⚫ Rated input power of analyzer: 500VA. The instrument should be connected to a power
socket with load no less than 2.5A.
⚫ The ground voltage should be <5V.
⚫ If the user is going to use a UPS to power the instrument, make sure that the UPS can
provide power supply greater than or equal to 1500VA (analyzer + computer + printer).
⚫ The installation site should be kept away from big noise and power supply interference.
⚫ Keep the system away from brush-type motors and electrical contact devices that are
frequently switched on and off.
⚫ Do not use such devices as mobile phones and radio transmitters near the system.
BIOHAZARD
Dispose of liquid waste according to the local regulations.
Item Description
Make sure that the C drive is over 30G and D drive over 100G,
and the disk file system is of NTFS format.
Deselect the two options at the bottom of the disk properties
window: “Compress drive to save disk space” and “Allow
Indexing Service to index this disk for fast file searching”.
The operating system installed on the computer must be an
Operating system activated Microsoft Windows 10 HOME (64 bits & version above
1607).
Except for the operating system, other application software
must not be installed or reserved on the computer. If an anti-
Application software virus application has been installed, then remove the
automatic scheduled scanning and add the operating software
and BSLOG to the trust list.
Screen saver and system Turn off the screen saver and BS Special Power Policy power
standby scheme, and then disable the hibernation option.
17” touchscreen monitor or above, 16:9 or 4:3 with resolution
Screen display properties
of 1280×1024.
Automatic synchronization Disable the Automatically synchronize with an Internet time
with Internet time server server option.
Automatic updates Turn off the automatic updates.
If you are going to use the auto startup function, perform
Auto startup setup necessary settings for BIOS and network adapters while
referring to their operation manuals.
2.1 Tools
Hexagon wrenches (M3~M6), 13# box spanner, tape measure, 200mm spanner, cross
screwdriver, flathead screwdriver and snap-off knife
Figure 2 Tools
2.2.1 Unpacking
1. Check against packing list: Check the instrument configuration against the packing list
before unpacking according to section 1.3 System Configuration.
Analyzer SN
and packing list
2. Unloading: Unload the instrument from the truck. It is recommended to use a forklift.
Note: There is a tilt warning label on the main unit package, which can change in color for
tilting. When the center of the label turns red, it means the package has been seriously tilted
and instrument damage may be caused. Contact the logistics company immediately to
prevent unexpected results.
3. Transfer: Use a forklift to transport the instrument to a location suitable for unpacking.
Note: Keep the proper distance between the instrument and the wall to facilitate the removal
of the package.
4. Unpacking analyzer: Cut the packing ropes with scissors, remove the upper cover of the
main box, →remove the protective plates on both sides, →and remove the boarding. Deliver
the removed packaging materials to the customer for safekeeping.
NOTE
After unpacking, remind the user to keep the packaging materials for 3 months for use in
returning or replacing the instrument.
CAUTION
After remove the accessory foam, remind the user to keep the accessory foam for 3 months
for use in returning or replacing the instrument.
tightened;
(2) Lift the instrument and move it to the test bench by 4 persons;
CAUTION
Because the analyzer is heavy, it is recommended to carry it 4 people together and walk
with caution.
Be careful of bruises! Be careful not to hit other fragile items! Be careful not to scratch the
instrument!
(3) After the instrument is placed on the test bench, adjust the anchor of the instrument
to make the machine evenly bear the force;
Figure 20 Remove the fixing plate for the probe drive assembly
g. Remove the fixing plate for waste drainage
Figure 23 Sealing Tapes around the Aspirate Plate of the Reagent Chamber
i. There are 7 fixing plates, make sure all fixing plates are removed
(2) Connect the other end of the waste tubes to the waste tank cover and install the waste
sensor.
CL-900i Chemiluminescence
Immunoassay Analyzer
Waste
drain
5. Ensure that all electric connectors are away from and higher than hydropneumatic system
parts.
6. Ensure that all electric connectors are higher than and away from fluidic parts and prevent
fluidic connectors from facing electric components.
BIOHAZARD
Dispose of liquid waste according to the local regulations.
3. Take two wash buffer tanks. Remove the cover of the wash buffer tank and place the wash
buffer tank assembly into the tank and tighten the cover.
Figure 29 Placing the Wash Buffer Tank Cover Assembly into the Tank and Tighten the Cover
4. Place the wash buffer tank and complete the connection of the wash buffer tank.
The layout of the waste tank and the wash buffer tank can be adjusted slightly according to the
actual situation of the customer.
2. Enter the username “ServiceUser” and password “#BS8A#SEU” in the login window
Confirm and align the positions in the red frame above. For detailed steps, please refer to
the Alignment Guide.
Regarding the Reagent Disk Barcode Scanning, only need to place 3-5 reagent bottles in
reagent disk to confirm.
Regarding the Sample Disk Barcode Scanning, only need to place 5-10 sample tubes with
barcode in sample rack to confirm.
One reagent bottle with 4 chambers is needed on probe-reagent disk position alignment.
The list of reagent bottle can be used is as below:
FT3 T3 Anti-Tg Anti-TPO
rT3 E2 Anti-Hbe Anti-Hbc
HIV Anti-TP Anti-HCV VB12
Confirm and align the positions in the red frame above. For detailed steps, please refer to
the Alignment Guide.
Confirm and align the positions in the red frame above. For detailed steps, please refer to
the Alignment Guide.
Confirm and align the positions in the red frame above. For detailed steps, please refer to
the Alignment Guide.
2) Select Utility > Maintenance > Alignment > Fluidic Alignment; select 12. Priming Wash
Buffer Tubes, and prime the wash buffer according to the prompt information;
Attention: Wash buffer prime should be executed BEFORE sampling probe position alignment, or
the position alignment will be INACCURATE!
Finish the VLP of probe to mixing position 1/2 alignment refer to Alignment Guide.
Attention: please STRICTLY FILLOW the procedure to complete substrate tube washing
and priming, or else the substrate background test might be FAILED!
Connect the
substrate drain
tube to the
connector under
the substrate
Remove the substrate tube holder
holder (note to protect the
substrate tube)
4) In the 2nd step, clean the substrate with purified water. Fill two bottles of clean substrate with
ultra-pure water, load the two bottles to the positions of substrate L and substrate R, and
then tap Clean on the software to confirm that you have followed the prompts and tap OK to
access the cleaning screen. Execution Times is set to 16 by default. The system
automatically starts to clean substrate L and substrate R with ultra-pure water; during the
cleaning process, observe the screen prompts become "No bubbles detected in the substrate
tubes", observe that the substrate tubes are inserted into the positions correctly, and there
is no leakage;
5) Tap Continue to go to the 3rd step, this step is to empty the ultra-pure water in the substrate
tubes, remove the clean substrate bottles containing ultra-pure water of substrate L and
substrate R, tap Empty, prompting that the substrate position should be vacant. After
confirming, enter the Empty screen. set Execution Times to the default value. Empty the
ultra-pure water in channel L and channel R. Observe that the screen should prompt
"Bubbles detected in the substrate tubes"; exit the screen after completion
6) Tap Continue to go to the 4th step, and use acidic lotion (System Wash Solution) to clean
the substrate tubes;
Note: System Wash Solution should be diluted manually by full-filling the bottle of the
solution with ultra-pure water.
Load the diluted System Wash Solution bottles on substrate position L and R, loose the
bottle caps, then tap Clean on software, prompting to confirm that the two substrate bottles
have been loaded with the acidic lotion. Enter the Clean screen, and set Execution Times
to the default value. The system automatically starts cleaning. Clean the substrate channels
with acidic lotion, the screen should prompt "No bubbles detected in the substrate tubes".
Exit the screen after completion;
7) Tap Continue to go to the 5th step, this step is to empty the acidic lotion in the substrate
tubes; tap Empty, and take out the System Detection Solution bottles of substrate L and
substrate R position as prompted. Tap OK to access the Empty screen. Set Execution
Times to the default value. Empty the acidic lotion in channel L and channel R. Observe that
the screen prompts "Bubbles detected in the substrate tubes"; exit the screen after
completion;
8) Tap Continue to go to the 6th step, this step is to clean the substrate tubes with ultra-pure
water. Tap Clean, and replace the ultra-pure water in the two clean substrate bottles as
prompted. Place ultra-pure water at the positions of substrate L and substrate R. Enter the
Clean screen, and set Execution Times to the default value. The system automatically
starts cleaning. Clean the substrate channels L and R with acidic lotion. The screen prompts
"No bubbles detected in the substrate tubes". Exit the screen after completion;
Attention: It is MUST to discard the former ultra-pure water in the clean substrate
bottles, and rinse the bottles with ultra-pure water, then fill the bottles with new ultra-
pure water. Or else the substrate tube might not be cleaned completely.
9) Tap Continue to go to the 7th step, this step is to empty the ultra-pure water in the substrate
tubes. Tap Empty, and take out the substrate bottles containing ultra-pure water of substrate
L and substrate R. Tap OK and enter the Empty screen. set Execution Times to the default
value. Tap Start to empty the ultra-pure water in channels L and R;
10) After the substrate cleaning is completed, the screen prompts to restore the tubing
installation. Since we need to prime the substrate later, you can temporarily leave it. Tap
Continue to complete the process;
same way. After loading, loosen the substrate cap, and prepare for the next step;
Note:
1. Before loading the substrate, tear the aluminum foil seal at the bottom of the substrate bottle;
2. If the reader cannot scan the bar code, perform the “Add and Enter Key” operation on the bar
code manual to scan three bar codes and see if it can be restored.
13) Select Utility →Alignment→ Fluidic Alignment → 4. Priming the substrate tubes (Note:
Do NOT enter 2. Cleaning the substrate tubes by mistake), tap Continue, and treat the
substrate tubes as prompted. Because the last step "Cleaning the substrate tubes" does not
restore the tubes, tap OK in this step;
14) In the second step, according to the substrate actually loaded by the client, you can select
"Substrate L", "Substrate R" or "Substrate L & Substrate R", prompting to confirm the loading
of the substrate, and tap OK. The system automatically enters the Priming screen and starts
priming. Prime substrate L and substrate R for 50 times (The process fix the priming time as
25, execute this process for twice to realize 50 times priming), respectively, and the tube
statuses should be "Normal".
15) The system exits the screen automatically when the operation is done. (If the priming is
interrupted midway, in order to avoid abnormality of the subsequent tests, you need to restart
Note:
a) Carefully install the screws. Do not drop screws into the dispersion carousel. Remove the
protection cover for the IO port of the dispersion carousel at the last step.
b) Protect the substrate tube from bending. Protect the hose at the substrate joint and the joint
outlet from contamination.
c) Do NOT forget to re-plug the substrate drain tube plug and fix the tube back to the original
position, and tighten the connector.
d) Make sure the screws of substrate installing base are tightened.
Insertion depth of
substrate drain tube
must be over 2 mm
Confirm it is
tightened
Remove
Fix with at last
cable tie
Fasten the
screws. Do not
drop the screws
in the dispersion
carousel.
3.7 Setting up
Figure 49 Pushing it forward after confirming that the cuvettes are placed well
3. Load the CD80 detergent
1) Select Reagent > Consumable Management;
2) Select CD80 Detergent, tap Load, input the corresponding information and margin of
CD80, and load the CD80 detergent;
Figure 53 Import
Select Load Default to display all Mindray reagent chemistries in the left column. Select Add All
to add all chemistries to the right column.
Select Import and then select Exit. The default chemistry parameters are imported.
3. Record the Relative Extreme Difference, which should be less than 1.5%.
The Sample Aspirate Volume (ul) is 15, and the Sample Dispense Volume (ul) (ul) is 10;
The Reagent Aspirate Volume (ul) is 0, and the Reagent Dispense Volume (ul) is 0;
4. Tap Start, and wait for the results.
5. The instrument automatically starts detection, and displays the results after completion. The
Item Index
Sample RLU 400000-2000000
Sample CV CV ≤ 1.5%
CV≤5%
Fill the test results into the Basic Performance Test Record file attached, it will show if the
repeatability test is pass or failed:
Basic
Performance Test Record.xlsx