Download as pdf or txt
Download as pdf or txt
You are on page 1of 42

IVD Global Technical Support Dept.

CL-900i Series
Chemiluminescence Immunoassay
Analyzer

Installation Guidance

IVD Global Technical Support Department

Document No.: Page 1 of 42


IVD Global Technical Support Dept.

Table of Contents
Table of Contents ................................................................................................................................... 2
Chapter 1 Check before Installation .................................................................................................. 3
1.1 Installation Environment ...................................................................................................... 3
1.2 Space and Accessibility Requirements for Unpacking ................................................... 3
1.2.1 Space and Accessibility Requirements for Installation ....................................... 3
1.2.2 Power Supply and Noise ......................................................................................... 4
1.2.3 Drainage Check (if draining water through a sewer)........................................... 4
1.2.4 Recommended PC Configuration .......................................................................... 4
1.3 Configuration Check ............................................................................................................ 5
Chapter 2 Instrument installation ...................................................................................................... 6
2.1 Tools ....................................................................................................................................... 6
2.2 Installation Procedure .......................................................................................................... 6
2.2.1 Unpacking .................................................................................................................. 6
2.2.2 Fluidic Connection .................................................................................................. 16
2.2.3 Connect the computer............................................................................................ 19
2.2.4 Software Installation ............................................................................................... 19
Chapter 3 Power on and alignment ................................................................................................. 20
3.1 Preparation for Powering On ............................................................................................ 20
3.2 Initial Startup ....................................................................................................................... 20
3.3 Main Unit Position Confirmation and Alignment ............................................................ 21
3.4 Load and Prime Wash Buffer............................................................................................ 24
3.5 Probe Vertical Pos. to Mixing Position Alignment .......................................................... 25
3.6 Substrate Prime .................................................................................................................. 26
3.7 Setting up............................................................................................................................. 31
3.7.1 Load and check the consumables ....................................................................... 31
3.7.2 Importing and Configuring Chemistry Parameters ............................................ 33
Chapter 4 System Performance Test .............................................................................................. 36
4.1 DCF Diagnosis .................................................................................................................... 36
4.2 Substrate Background detection ...................................................................................... 36
4.3 System Repeatability ......................................................................................................... 36
4.4 Dispersion Wash Effects ................................................................................................... 37
4.5 Repeatability Test ............................................................................................................... 37
Chapter 5 LIS and Remote Help ....................................................................................................... 39
5.1 LIS Connection ................................................................................................................... 39
Appendix A Prepare Installation Reagent Pack ........................................................................... 40
Appendix B Comparison of altitude and atmospheric pressure ............................................. 41

No.: Version: V1.0 Page 2 of 42


IVD Global Technical Support Dept.

Chapter 1 Check before Installation

1.1 Installation Environment


⚫ Altitude: -400-3,000m.
⚫ The system is for indoor use only.
⚫ The bearing platform (or ground) should be level (with gradient less than 1/200).
⚫ The bearing platform (or ground) should be able to support at least 150Kg weight.
⚫ The installation site should be well ventilated.
⚫ The installation site should be free of dust.
⚫ The installation side should not be in direct sun.
⚫ The installation site should be kept away from a heat or draft source.
⚫ The installation site should be free of corrosive gas and flammable gas.
⚫ The bearing platform (or ground) should be free of vibration.
⚫ Operating temperature: 15°C-30°C with fluctuation <2°C/H. Provide air conditioning
equipment if the room temperature does not meet the requirements.
⚫ Relative humidity: 35%-85% RH, without condensation.

1.2 Space and Accessibility Requirements for Unpacking


⚫ Dimensions of whole unit after packing: 1130 mm (length) × 910 mm (depth) × 760 mm
(height)
⚫ Unpacking space requirement: After putting down the instrument with the packing box, the
distance from the wall is ≥ 0.5 m for unpacking operation.

1.2.1 Space and Accessibility Requirements for Installation


⚫ Dimensions of whole unit before packing: 860mm (length) × 740 mm (depth) × 560mm
(height). 5417m
5279m
m
m Circuit breaker

Power line
≥700mm
Communication
Waste tube line

Waste tank
2782m
2474m

operation
m

LIS host

UPS
PC for

≥500mm
m

CL-900i ≥350mm

Figure 1 Space and accessibility requirements for installation of the CL-900i

No.: Version: V1.0 Page 3 of 42


IVD Global Technical Support Dept.

1.2.2 Power Supply and Noise


⚫ Power supply: 100V-240V~ 50Hz, 100V/240V~ 60Hz. Voltage fluctuation: +/-10%. Line
frequency: +/-1Hz. Three-wire power cord with good grounding performance.

WARNING
Make sure the power socket is grounded correctly. Incorrect grounding may lead to electric shock
or equipment damage.
Check if the power socket outputs voltage meeting the specified requirements and has a proper
fuse installed.

⚫ Rated input power of analyzer: 500VA. The instrument should be connected to a power
socket with load no less than 2.5A.
⚫ The ground voltage should be <5V.
⚫ If the user is going to use a UPS to power the instrument, make sure that the UPS can
provide power supply greater than or equal to 1500VA (analyzer + computer + printer).
⚫ The installation site should be kept away from big noise and power supply interference.
⚫ Keep the system away from brush-type motors and electrical contact devices that are
frequently switched on and off.
⚫ Do not use such devices as mobile phones and radio transmitters near the system.

1.2.3 Drainage Check (if draining water through a sewer)


⚫ The chemiluminescence immunoassay analyzer can directly discharge the waste liquid to
the sewer. The length of the waste tube is less than 5 m and greater than 1 m.
⚫ The waste outlet should be no less than 8mm wide.
⚫ The waste outlet should be lower than sewer for at least 0.5m.

BIOHAZARD
Dispose of liquid waste according to the local regulations.

1.2.4 Recommended PC Configuration


Item Description
CPU At least Intel P4, 2.8GHz
Random access memory (RAM) At least 4GB
The computer is connected to the analyzer through a network
adapter. If you are going to connect the computer with the LIS
Network adapter
or Internet, you should prepare another network adapter (Intel
gigabit network adapter)
At least 500GB, with SATA interface.
Hard disk defragment Install the operating system in the C drive and the operating
software of the instrument in the D drive.

No.: Version: V1.0 Page 4 of 42


IVD Global Technical Support Dept.

Item Description
Make sure that the C drive is over 30G and D drive over 100G,
and the disk file system is of NTFS format.
Deselect the two options at the bottom of the disk properties
window: “Compress drive to save disk space” and “Allow
Indexing Service to index this disk for fast file searching”.
The operating system installed on the computer must be an
Operating system activated Microsoft Windows 10 HOME (64 bits & version above
1607).
Except for the operating system, other application software
must not be installed or reserved on the computer. If an anti-
Application software virus application has been installed, then remove the
automatic scheduled scanning and add the operating software
and BSLOG to the trust list.
Screen saver and system Turn off the screen saver and BS Special Power Policy power
standby scheme, and then disable the hibernation option.
17” touchscreen monitor or above, 16:9 or 4:3 with resolution
Screen display properties
of 1280×1024.
Automatic synchronization Disable the Automatically synchronize with an Internet time
with Internet time server server option.
Automatic updates Turn off the automatic updates.
If you are going to use the auto startup function, perform
Auto startup setup necessary settings for BIOS and network adapters while
referring to their operation manuals.

1.3 Configuration Check


Table 1 Configuration list

• CL-900i/CL-920i Chemiluminescence Immunoassay Analyzer ------------------------------ 1


• Accessory kit ----------------------------------------------------------------------------------------------- 1
• Computer (Optional) ------------------------------------------------------------------------------------- 1
• Display monitor (Optional)------------------------------------------------------------------------------ 1
• Waste tank ------------------------------------------------------------------------------------------------- 2
• Optional modules
Printer ------------------------------------------------------------------------------------------------------ 1

No.: Version: V1.0 Page 5 of 42


IVD Global Technical Support Dept.

Chapter 2 Instrument installation

2.1 Tools
Hexagon wrenches (M3~M6), 13# box spanner, tape measure, 200mm spanner, cross
screwdriver, flathead screwdriver and snap-off knife

Figure 2 Tools

2.2 Installation Procedure

2.2.1 Unpacking
1. Check against packing list: Check the instrument configuration against the packing list
before unpacking according to section 1.3 System Configuration.

Analyzer SN
and packing list

Figure 3 Serial No. and packing list

No.: Version: V1.0 Page 6 of 42


IVD Global Technical Support Dept.

2. Unloading: Unload the instrument from the truck. It is recommended to use a forklift.
Note: There is a tilt warning label on the main unit package, which can change in color for
tilting. When the center of the label turns red, it means the package has been seriously tilted
and instrument damage may be caused. Contact the logistics company immediately to
prevent unexpected results.

If it turns red here, a


serious tilt occurs during
transportation!

Figure 4 Tilt warning label in normal conditions

3. Transfer: Use a forklift to transport the instrument to a location suitable for unpacking.
Note: Keep the proper distance between the instrument and the wall to facilitate the removal
of the package.

Figure 5 Transferring Instrument to a Location Suitable for Unpacking

No.: Version: V1.0 Page 7 of 42


IVD Global Technical Support Dept.

4. Unpacking analyzer: Cut the packing ropes with scissors, remove the upper cover of the
main box, →remove the protective plates on both sides, →and remove the boarding. Deliver
the removed packaging materials to the customer for safekeeping.

NOTE
After unpacking, remind the user to keep the packaging materials for 3 months for use in
returning or replacing the instrument.

Figure 6 Outer Package Removed

5. Remove the instrument protective film and accessory foam.

CAUTION
After remove the accessory foam, remind the user to keep the accessory foam for 3 months
for use in returning or replacing the instrument.

Figure 7 Accessory foam removed

6. Carry the analyzer to the desk:


(1) Confirm that the carrying handles at the four ends of the instrument have been

No.: Version: V1.0 Page 8 of 42


IVD Global Technical Support Dept.

tightened;
(2) Lift the instrument and move it to the test bench by 4 persons;

CAUTION
Because the analyzer is heavy, it is recommended to carry it 4 people together and walk
with caution.
Be careful of bruises! Be careful not to hit other fragile items! Be careful not to scratch the
instrument!

Figure 8 Carry the analyzer to the desk

(3) After the instrument is placed on the test bench, adjust the anchor of the instrument
to make the machine evenly bear the force;

Figure 9 Adjusting the anchor of the Instrument

7. Removing Clear protective accessories :


(1) Unscrew the carrying handles of the instrument and hand them to the customer for
safekeeping for subsequent transfer;
(2) Remove all the fixing tapes on the surface of the instrument;
(3) Open the front door, and remove the protective foam of the front door;

No.: Version: V1.0 Page 9 of 42


IVD Global Technical Support Dept.

Remove the protective


foam of the front door

Figure 10 Remove the protective foam of the front door


(4) Remove the screws of the transparent cover to open it;

2.Pull this place to remove


the transparent cover.

1.Open the front left door. Remove the


hexagon screw which is on the bottom of
the beam. Fix the screw after
maintenance.

Figure 11 Remove the screws of the transparent cover

No.: Version: V1.0 Page 10 of 42


IVD Global Technical Support Dept.

Figure 12 Remove the transparent cover


(5) Remove all fixing plates as follow :
a. Remove the fixing plate for fixing gripper Y axis. First, remove the screw from
the position below;

Figure 13 Remove the screw from the position below


Remove the fixing plate.

Push foreward the gripper X


axis movement part and remove
another screw. Then remove
the fixing plate.

Figure 14 Remove the fixing plate

No.: Version: V1.0 Page 11 of 42


IVD Global Technical Support Dept.

b. Remove the fixing plate for the drawer assembly

Figure 15 Remove the fixing plate for the drawer assembly


c. Remove the fixing plate for fixing gripper X axis

Remove the screw witch is on


the rightside of the griping

Push foreward the gripper


X axis movement part and
remove another screw.
Then remove the fixing
plate.

Figure 16 Remove the fixing plate for fixing gripper X axis


d. Remove the fixing plate for fixing gripper Z axis

No.: Version: V1.0 Page 12 of 42


IVD Global Technical Support Dept.

1.Loose this screw, no need to remove it.


Re-tighten it after remove the fixing plate.
2.Remove another screw which
is on the leftside of the gripper

Figure 17 Remove the fixing plate for fixing gripper Z axis


e. Unplug the rubber covers on both sides, remove the hex socket screws on both
sides, and press the top two sides of the front vertical plate. When removing the
pop-up, take care not to pull the wiring on both sides.

Figure 18 The rubber covers

No.: Version: V1.0 Page 13 of 42


IVD Global Technical Support Dept.

Figure 19 Removing the Front Vertical Plate


f. Remove the fixing plate of the Probe drive assembly

Figure 20 Remove the fixing plate for the probe drive assembly
g. Remove the fixing plate for waste drainage

No.: Version: V1.0 Page 14 of 42


IVD Global Technical Support Dept.

Figure 21 Remove the fixing plate for waste drainage


h. Remove the sealing tapes around the incubation block, dispersion, mixing
position, wash well, and aspirate plate of the reagent chamber in turn, and keep
the small hole opaque disk of the incubation block for subsequent use.

Figure 22 Sealing Tapes around the Incubation Block, Dispersion, Mixing


Position, and Wash Well

Figure 23 Sealing Tapes around the Aspirate Plate of the Reagent Chamber
i. There are 7 fixing plates, make sure all fixing plates are removed

No.: Version: V1.0 Page 15 of 42


IVD Global Technical Support Dept.

Figure 24 All the fixing plates

2.2.2 Fluidic Connection


2.2.2.1 Connecting Waste Tubes
The instrument has two waste drainage outlets: waste outlet 1 and waste outlet 2. According to
the actual situation of the customer, you can choose to use the waste tank, or directly discharge
the waste.
1. Use the waste tank
(1) Take out the waste tank cover assembly and the waste sensor assembly, and connect
one end of the two waste tubes to “waste outlet 1” and “waste outlet 2” of the instrument, and
connect the waste sensor at the same time.

Figure 25 Waste Tube and Waste Sensor Connection

No.: Version: V1.0 Page 16 of 42


IVD Global Technical Support Dept.

(2) Connect the other end of the waste tubes to the waste tank cover and install the waste
sensor.

Figure 26 Waste tank connection


(3) Place the waste tank and complete the connection of the waste tank.
NOTE: Do not connect the tubes of waste outlet 1 and waste outlet 2, which will cause waste
to be flowed backward!
2. Directly discharge the waste
If the waste can be discharged through sewer, follow the steps below:

CL-900i Chemiluminescence
Immunoassay Analyzer

Waste
drain

Figure 27 Waste discharge


NOTE: Do not connect the tubes of waste outlet 1 and waste outlet 2, which will cause waste
to be flowed backward!

Requirements for connecting waste tubes:


1. Insert the waste tube into the sewer, and make sure that the tube is about 1m~5m long.
2. Make sure that the waste outlet is no less than 50cm high.
3. Cover the waste outlet to eliminate noise and prevent bubbles overflow.
4. When installing the tubes, prevent them from being bent, twisted, or pressed, and keep them
away from any sharp-edged objects or others that may scratch the tubes.

No.: Version: V1.0 Page 17 of 42


IVD Global Technical Support Dept.

5. Ensure that all electric connectors are away from and higher than hydropneumatic system
parts.
6. Ensure that all electric connectors are higher than and away from fluidic parts and prevent
fluidic connectors from facing electric components.

BIOHAZARD
Dispose of liquid waste according to the local regulations.

2.2.2.2 Connect Wash buffer tank


1. Take out the wash buffer tank assembly.
2. Connect the tubes of the wash buffer tank assembly to the wash buffer inlets of the
instrument;
Note: The tube with mark 1 is connected to wash buffer tank 1; the tube with mark 2 is
connected to wash buffer tank 2. Do not mistake them!

Figure 28 Wash Buffer Tube Connection

3. Take two wash buffer tanks. Remove the cover of the wash buffer tank and place the wash
buffer tank assembly into the tank and tighten the cover.

Figure 29 Placing the Wash Buffer Tank Cover Assembly into the Tank and Tighten the Cover
4. Place the wash buffer tank and complete the connection of the wash buffer tank.

No.: Version: V1.0 Page 18 of 42


IVD Global Technical Support Dept.

2.2.2.3 Final Connection Effects

Figure 30 Final connection effects

The layout of the waste tank and the wash buffer tank can be adjusted slightly according to the
actual situation of the customer.

2.2.3 Connect the computer.


Connect the PC with analyzer
The network cable is used to connect the analyzing unit. The IP of the network adapter on the
PC should be set as 192.168.23.3.

2.2.4 Software Installation


Complete the software installation according to the SetupGuide in software package.

No.: Version: V1.0 Page 19 of 42


IVD Global Technical Support Dept.

Chapter 3 Power on and alignment

3.1 Preparation for Powering On


➢ Load solid waste containers.

Figure 31 Load solid waste container

➢ Power on the instrument

3.2 Initial Startup


1. Start up the computer. The CL-900i Operating Software is run automatically. See the figure
below.

Figure 32 Start up CL-900i

2. Enter the username “ServiceUser” and password “#BS8A#SEU” in the login window

No.: Version: V1.0 Page 20 of 42


IVD Global Technical Support Dept.

Figure 33 Login window of CL-900i

3.3 Main Unit Position Confirmation and Alignment


1. Preparation for position confirmation and alignment.
Execute Home. If it enters the "Idle" state smoothly and there is no fault alarm, you can
refer to the Debugging Guide to confirm and debug the key positions. Otherwise, the fault
should be checked.

Figure 34 Home system

2. Procedure of position confirmation and alignment


(1) Position confirmation and alignment of sample probe
Select Utility > Maintenance > Alignment > Sampling System Alignment

No.: Version: V1.0 Page 21 of 42


IVD Global Technical Support Dept.

Figure 35 Position confirmation and alignment of sample probe

Confirm and align the positions in the red frame above. For detailed steps, please refer to
the Alignment Guide.
Regarding the Reagent Disk Barcode Scanning, only need to place 3-5 reagent bottles in
reagent disk to confirm.
Regarding the Sample Disk Barcode Scanning, only need to place 5-10 sample tubes with
barcode in sample rack to confirm.
One reagent bottle with 4 chambers is needed on probe-reagent disk position alignment.
The list of reagent bottle can be used is as below:
FT3 T3 Anti-Tg Anti-TPO
rT3 E2 Anti-Hbe Anti-Hbc
HIV Anti-TP Anti-HCV VB12

(2) Position confirmation and alignment of gripper


Select Utility > Maintenance > Alignment > Transport System Alignment.

No.: Version: V1.0 Page 22 of 42


IVD Global Technical Support Dept.

Figure 36 Position confirmation and alignment of gripper

Confirm and align the positions in the red frame above. For detailed steps, please refer to
the Alignment Guide.

(3) Position alignment of dispersion probe


Select Utility > Maintenance > Alignment > Dispersion System Alignment.

No.: Version: V1.0 Page 23 of 42


IVD Global Technical Support Dept.

Figure 37 Position confirmation and alignment of dispersion probe

Confirm and align the positions in the red frame above. For detailed steps, please refer to
the Alignment Guide.

(4) Position confirmation and alignment of shielding cover


Select Utility > Maintenance > Alignment > Photometer System Alignment.

Figure 38 Position confirmation and alignment of shielding cover

Confirm and align the positions in the red frame above. For detailed steps, please refer to
the Alignment Guide.

3.4 Load and Prime Wash Buffer


1) In the Consumables Management screen, select Wash Buffer 1 and Wash Buffer 2,
respectively, and tap Load to load the wash buffer;

No.: Version: V1.0 Page 24 of 42


IVD Global Technical Support Dept.

Figure 39 Load the wash buffer

2) Select Utility > Maintenance > Alignment > Fluidic Alignment; select 12. Priming Wash
Buffer Tubes, and prime the wash buffer according to the prompt information;
Attention: Wash buffer prime should be executed BEFORE sampling probe position alignment, or
the position alignment will be INACCURATE!

3.5 Probe Vertical Pos. to Mixing Position Alignment


Select Utility > Maintenance > Alignment > Sampling System Alignment

Figure 40 Position confirmation and alignment of sample probe

No.: Version: V1.0 Page 25 of 42


IVD Global Technical Support Dept.

Finish the VLP of probe to mixing position 1/2 alignment refer to Alignment Guide.

3.6 Substrate Prime

Attention: please STRICTLY FILLOW the procedure to complete substrate tube washing
and priming, or else the substrate background test might be FAILED!

Clean the substrate tube


1) Prepare 2 bottles of acidic lotion and 2 clean substrate bottles;
2) The positions of two substrate spikes are vacant;
3) Select Utility > Maintenance > Alignment > Fluidics Alignment > 2. Clean Substrate
Tubes, and operate as prompted.

Figure 41 Prompts before cleaning substrate tubes

Here is the substrate drain


tube, take out it and pull out
the white plug before use.

No.: Version: V1.0 Page 26 of 42


IVD Global Technical Support Dept.

Protect dispersion IO port

Connect the
substrate drain
tube to the
connector under
the substrate
Remove the substrate tube holder
holder (note to protect the
substrate tube)

Figure 42 Preparation before cleaning substrate tubes

4) In the 2nd step, clean the substrate with purified water. Fill two bottles of clean substrate with
ultra-pure water, load the two bottles to the positions of substrate L and substrate R, and
then tap Clean on the software to confirm that you have followed the prompts and tap OK to
access the cleaning screen. Execution Times is set to 16 by default. The system
automatically starts to clean substrate L and substrate R with ultra-pure water; during the
cleaning process, observe the screen prompts become "No bubbles detected in the substrate
tubes", observe that the substrate tubes are inserted into the positions correctly, and there
is no leakage;

Figure 43 Clean Substrate Tubes Screen

5) Tap Continue to go to the 3rd step, this step is to empty the ultra-pure water in the substrate
tubes, remove the clean substrate bottles containing ultra-pure water of substrate L and
substrate R, tap Empty, prompting that the substrate position should be vacant. After
confirming, enter the Empty screen. set Execution Times to the default value. Empty the

No.: Version: V1.0 Page 27 of 42


IVD Global Technical Support Dept.

ultra-pure water in channel L and channel R. Observe that the screen should prompt
"Bubbles detected in the substrate tubes"; exit the screen after completion
6) Tap Continue to go to the 4th step, and use acidic lotion (System Wash Solution) to clean
the substrate tubes;
Note: System Wash Solution should be diluted manually by full-filling the bottle of the
solution with ultra-pure water.
Load the diluted System Wash Solution bottles on substrate position L and R, loose the
bottle caps, then tap Clean on software, prompting to confirm that the two substrate bottles
have been loaded with the acidic lotion. Enter the Clean screen, and set Execution Times
to the default value. The system automatically starts cleaning. Clean the substrate channels
with acidic lotion, the screen should prompt "No bubbles detected in the substrate tubes".
Exit the screen after completion;
7) Tap Continue to go to the 5th step, this step is to empty the acidic lotion in the substrate
tubes; tap Empty, and take out the System Detection Solution bottles of substrate L and
substrate R position as prompted. Tap OK to access the Empty screen. Set Execution
Times to the default value. Empty the acidic lotion in channel L and channel R. Observe that
the screen prompts "Bubbles detected in the substrate tubes"; exit the screen after
completion;
8) Tap Continue to go to the 6th step, this step is to clean the substrate tubes with ultra-pure
water. Tap Clean, and replace the ultra-pure water in the two clean substrate bottles as
prompted. Place ultra-pure water at the positions of substrate L and substrate R. Enter the
Clean screen, and set Execution Times to the default value. The system automatically
starts cleaning. Clean the substrate channels L and R with acidic lotion. The screen prompts
"No bubbles detected in the substrate tubes". Exit the screen after completion;
Attention: It is MUST to discard the former ultra-pure water in the clean substrate
bottles, and rinse the bottles with ultra-pure water, then fill the bottles with new ultra-
pure water. Or else the substrate tube might not be cleaned completely.
9) Tap Continue to go to the 7th step, this step is to empty the ultra-pure water in the substrate
tubes. Tap Empty, and take out the substrate bottles containing ultra-pure water of substrate
L and substrate R. Tap OK and enter the Empty screen. set Execution Times to the default
value. Tap Start to empty the ultra-pure water in channels L and R;
10) After the substrate cleaning is completed, the screen prompts to restore the tubing
installation. Since we need to prime the substrate later, you can temporarily leave it. Tap
Continue to complete the process;

Prime the substrate.


11) Prepare two bottles of well-balanced substrate with balance time as long as 6 hours and use
them before the expiration date;
12) Exit the Alignment screen, Select Reagents →> Consumables Management > Substrate
L, and tap Load. The Load screen is displayed. Use the handheld bar code reader to scan
the bar code of the substrate bottle, identify the substrate bottle (tear the aluminum foil seal
at the bottom of the substrate bottle) and load the bottle to the position of substrate L; if two
bottles of substrate should be loaded, load the bottle to the position of substrate R in the

No.: Version: V1.0 Page 28 of 42


IVD Global Technical Support Dept.

same way. After loading, loosen the substrate cap, and prepare for the next step;

Figure 44 Loading the Substrate

Note:
1. Before loading the substrate, tear the aluminum foil seal at the bottom of the substrate bottle;
2. If the reader cannot scan the bar code, perform the “Add and Enter Key” operation on the bar
code manual to scan three bar codes and see if it can be restored.

Figure 45 Bar code manual

13) Select Utility →Alignment→ Fluidic Alignment → 4. Priming the substrate tubes (Note:
Do NOT enter 2. Cleaning the substrate tubes by mistake), tap Continue, and treat the
substrate tubes as prompted. Because the last step "Cleaning the substrate tubes" does not
restore the tubes, tap OK in this step;
14) In the second step, according to the substrate actually loaded by the client, you can select
"Substrate L", "Substrate R" or "Substrate L & Substrate R", prompting to confirm the loading
of the substrate, and tap OK. The system automatically enters the Priming screen and starts
priming. Prime substrate L and substrate R for 50 times (The process fix the priming time as
25, execute this process for twice to realize 50 times priming), respectively, and the tube
statuses should be "Normal".
15) The system exits the screen automatically when the operation is done. (If the priming is
interrupted midway, in order to avoid abnormality of the subsequent tests, you need to restart

No.: Version: V1.0 Page 29 of 42


IVD Global Technical Support Dept.

the process to perform the complete priming process.);


16) Follow the instructions in the third step to restore the substrate tubes;

Figure 46 Prompts for restoring substrate tubes

Note:
a) Carefully install the screws. Do not drop screws into the dispersion carousel. Remove the
protection cover for the IO port of the dispersion carousel at the last step.
b) Protect the substrate tube from bending. Protect the hose at the substrate joint and the joint
outlet from contamination.
c) Do NOT forget to re-plug the substrate drain tube plug and fix the tube back to the original
position, and tighten the connector.
d) Make sure the screws of substrate installing base are tightened.

Insertion depth of
substrate drain tube
must be over 2 mm

Confirm it is
tightened

Remove
Fix with at last
cable tie

Fasten the
screws. Do not
drop the screws
in the dispersion
carousel.

Figure 47 Diagram of restoring substrate tubes

No.: Version: V1.0 Page 30 of 42


IVD Product Global Technical Support Department

3.7 Setting up

3.7.1 Load and check the consumables


1. Loading the waste container:
Before startup, the waste container has been loaded. The sensor checks that there is a waste
container and automatically refreshes the volume to start counting.
2. Loading the cuvette tray:
Select Left Tray, tap Load, and then pull out the left drawer assembly manually, load the cuvette
tray, and push it forward after confirming that it is placed well. Load the cuvettes in the right tray
with the same method.

Figure 48 Loading the cuvette tray

Figure 49 Pushing it forward after confirming that the cuvettes are placed well
3. Load the CD80 detergent
1) Select Reagent > Consumable Management;

Document No.: Page 31 of 42


IVD Global Technical Support Dept.

Figure 50 Consumable management screen

2) Select CD80 Detergent, tap Load, input the corresponding information and margin of
CD80, and load the CD80 detergent;

Figure 51 Load the CD80 Detergent

4. Load the waste tank


Before connecting the hydropneumatic system tubes during installation, you have loaded the
water tank.
5. Load the substrate
Before priming the substrate, you have loaded the substrate.
6. Besides the Consumable Management screen, you can query the detailed information of
consumables on the Reagent Overview screen

No.: Version: V1.0 Page 32 of 42


IVD Global Technical Support Dept.

Figure 52 Reagent overview

3.7.2 Importing and Configuring Chemistry Parameters


Select Utility -> Chemistries -> Import.

Figure 53 Import

Select Load Default to display all Mindray reagent chemistries in the left column. Select Add All
to add all chemistries to the right column.

No.: Version: V1.0 Page 33 of 42


IVD Global Technical Support Dept.

Figure 54 Import default chemistry parameters

Select Import and then select Exit. The default chemistry parameters are imported.

Figure 55 Select Import and Exit

Check the imported chemistries on the Chemistries screen.

No.: Version: V1.0 Page 34 of 42


IVD Global Technical Support Dept.

Figure 56 Check the imported chemistries

No.: Version: V1.0 Page 35 of 42


IVD Global Technical Support Dept.

Chapter 4 System Performance Test

4.1 DCF Diagnosis


1. Enter Utility-Maintenance-Diagnostics- Photometer Diagnosis -DCF diagnosis.
2. Parameter settings: Set the Test Cycle as 20 and select Start to start the test.

3. Record the Relative Extreme Difference, which should be less than 1.5%.

Figure 57 DCF stability test

4.2 Substrate Background detection


1. Select Utility > Maintenance > Basic Performance > Substrate;
2. Select Substrate Background;
3. Tick the Dispersion checkbox, tap Start, and wait for the results;
4. The instrument automatically starts detection, and displays the results after completion.
Item Index
Substrate background RLU 3000-6000
Substrate background SD SD<100

4.3 System Repeatability


Prepare the sample: Add 1 ml System Detection Solution A (0.002 ug/ml) to the microtube and
load it to 1# sample position;
1. Select Utility > Maintenance > Basic performance > Photon Counting;
2. Select System Repeatability;
3. Confirmation:

No.: Version: V1.0 Page 36 of 42


IVD Global Technical Support Dept.

The Sample Aspirate Volume (ul) is 15, and the Sample Dispense Volume (ul) (ul) is 10;
The Reagent Aspirate Volume (ul) is 0, and the Reagent Dispense Volume (ul) is 0;
4. Tap Start, and wait for the results.
5. The instrument automatically starts detection, and displays the results after completion. The
Item Index
Sample RLU 400000-2000000
Sample CV CV ≤ 1.5%

4.4 Dispersion Wash Effects


Prepare reagents: Prepare an empty reagent bottle and add the following solutions to 2#, 3#,
and 4# cells of the reagent bottle:
Add 4 ml of Wash Buffer to 2# cell of the reagent bottle;
Add 4 ml of System Detection Solution B (20 ug/ml) to 3# cell of the reagent bottle;
Add 4 ml of System Detection Solution A (0.002 ug/ml) to 4# cell of the reagent bottle;
Place the filled reagent bottle in 2# position of the reagent carousel and close the reagent
carousel;
Prepare the sample: Add 5 ml of the Wash Buffer to the tube and place it in 2# sample position.
1. Select Utility > Maintenance > Basic performance > Dispersion;
2. Select Dispersion Wash Effects;
3. Tap Start, and wait for the results;
4. The instrument automatically starts detection, and displays the results after completion.
Item Index
Mean<1.00ppm
Dispersion wash effects
SD ≤ 0.15ppm
Note: If the test does not pass, analyze the cause and add each sample to the specified volume,
and then test again.

4.5 Repeatability Test


Pick at least 1 item that will be test by user in the list below, finish the calibration at first.
Anti-Tg Anti-TPO LH HCG
HBsAg HIV Anti-TP Anti-HCV
Take a quality control as sample to perform 15 times repeatability test, the repeatability test
should satisfy the following indices:

CV≤5%
Fill the test results into the Basic Performance Test Record file attached, it will show if the
repeatability test is pass or failed:

No.: Version: V1.0 Page 37 of 42


IVD Global Technical Support Dept.

Basic
Performance Test Record.xlsx

No.: Version: V1.0 Page 38 of 42


IVD Product Global Technical Support Department

Chapter 5 LIS and Remote Help

5.1 LIS Connection


The service engineer should assist the hospital in connecting the LIS software.

Document No.: Page 39 of 42


IVD Global Technical Support Dept.

Appendix A Prepare Installation Reagent Pack


Please prepare these consumables according to the following steps.
1. The consumables and reagent are prepared by the agent.
2. Service personnel prepare the AP enzyme and acid wash solution according to the Part
Number.
3. Two clean substrate bottles accompanied with the instrument can be used directly.
4. Use the cuvettes and solid waste container in the accessory kit.
The solution required for installation should be prepared first as follows:
Person in
No. Name Quantity Code Function Solutions
Charge
1 Wash buffer 1 tank / Consumable, dispersion
Consumable, Luminous
2 Substrate 1 bottle / Notify the
reagent
agent to Distributor,
Consumable,
3 CD80 (40ml) 1 bottle / purchase Service
wash solution
before personnel
4 Reagent 1 /
installation
5 Calibrator 1 / Clinic precision test
6 Control 1 /
Acid wash solution Service
7 1 bottle 105-009143-00 Wash substrate tubes Request it
(System Wash Solution) personnel
according
AP enzyme(5ml) Service
8 1 box 105-009141-00 to PN
(System Detection Solution) Basic Performance Test personnel
9 Empty Reagent Bottle 1 piece 115-023170-00 N/A
10 Clean substrate bottle 2 pieces 105-005389-00 Wash substrate tubes Shipped N/A
11 Cuvette 2 trays / Consumable, reaction with N/A
Consumable, holding instrument
12 Solid Waste Container 1 piece / N/A
discarded cuvettes
Wash substrate tubes the user N/A
13 Deionized water some /
prepared

No.: Version: V1.0 Page 40 of 42


IVD Global Technical Support Dept.

Appendix B Comparison of altitude and


atmospheric pressure

Altitude Atmospheric Pressure(kPa) Atmospheric Pressure(psi)


(m)
-400 105.9 15.4
-300 104.8 15.2
-200 103.6 15.0
-100 102.5 14.9
0 101.3 14.7
100 100.2 14.5
200 99.1 14.4
300 98.0 14.2
400 96.9 14.0
500 95.8 13.9
600 94.7 13.7
700 93.6 13.6
800 92.6 13.4
900 91.5 13.3
1000 90.5 13.1
1100 89.4 13.0
1200 88.4 12.8
1300 87.4 12.7
1400 86.4 12.5
1500 85.4 12.4
1600 84.4 12.2
1700 83.4 12.1
1800 82.4 12.0
1900 81.5 11.8
2000 80.5 11.7
2100 79.6 11.5
2200 78.6 11.4
2300 77.7 11.3
2400 76.8 11.1
2500 75.9 11.0

No.: Version: V1.0 Page 41 of 42


IVD Global Technical Support Dept.

Altitude Atmospheric Pressure(kPa) Atmospheric Pressure(psi)


(m)
2600 75.0 10.9
2700 74.1 10.7
2800 73.2 10.6
2900 72.3 10.5
3000 71.4 10.4
3100 70.6 10.2
3200 69.7 10.1
3300 68.9 10.0
3400 68.0 9.9
3500 67.2 9.7
3600 66.4 9.6
3700 65.6 9.5
3800 64.8 9.4
3900 64.0 9.3
4000 63.2 9.2
4100 62.4 9.0
4200 61.6 8.9
4300 60.8 8.8
4400 60.1 8.7
4500 59.3 8.6
4600 58.6 8.5
4700 57.8 8.4
4800 57.1 8.3
4900 56.4 8.2
5000 55.6 8.1

No.: Version: V1.0 Page 42 of 42

You might also like