PKL PPC142 Operation Manual
PKL PPC142 Operation Manual
Microplate Reader
PKL PPC142
Operation Manual
Page 0 of 41
Microplate Reader Operation Manual
Content
1. Introduction ..................................................................................... 2
1.1 Summary ................................................................................. 2
1.2 Sign, labels and pictures ......................................................... 3
1.3 Warranty and service .............................................................. 4
2. Safety and precaution ..................................................................... 4
2.1 Safety classification................................................................. 4
2.2 Safety ...................................................................................... 4
2.3 Notes....................................................................................... 4
3. Installation ....................................................................................... 5
4. Working principle and structure drawing ..................................... 5
5. Operation ......................................................................................... 9
5.1 Turn it on ................................................................................ 9
5.2 Password, user right and main menu ...................................... 9
5.3 Test ....................................................................................... 10
5.3.1 Sample information ................................................... 11
5.3.2 Rapid Test ................................................................. 12
5.3.3 Single Test................................................................. 15
5.3.4 Multi -Test ................................................................. 20
5.4 Report ................................................................................... 24
5.5 Edit ........................................................................................ 27
5.6 System .................................................................................. 33
6. Daily maintenance ............................................................................. 37
Page 1 of 41
Microplate Reader Operation Manual
1. Introduction
1.1 Summary
1.1.1 Profile
Please read the manual carefully before using it.
a) Product Name: Microplate Reader (hereby called reader)
b) Product Code: PKL PPC142
c) Safety class: Anti-electric shock Class Ⅰ, overvoltag e protection Class, contamination Class
Ⅰ
d) Management class: IVD 6840, Management class Ⅱ
g) Standard related to our device: YZB/0040-2010 Microplate Reader.
h) Manufacturer:PARAMEDICAL srl
i) Via Antonio Amato 24/26 (Z.I.) 84131 Salerno- ITALY
j) Tel:+39 089.385027 Fax:+39 089.3854479
k) service@paramedical.it
Web:www.paramedical.it
After-sale service should be done by our company or our authorized distributor.
1.1.2 Features
a). Indication stability: ± 0.002A;
b). Wavelength indication error: ± 3nm;
c). Wavelength repeatability: ± 1.5nm;
d). Half-width: ≤ 12nm;
e). Absorbance indication error: ± 0.015A;
f). Optical density Repeatability: <0.1%;
g). Sensitivity: ≥ 0.01 mg / L;
h). Channel difference: ≤ 0.01A;
i). Reading speed: 3 seconds / 96 well (continued way into the board);
j). Dimensions: 525mm × 420mm × 260mm;
k). Net Weight: 9 Kg.
1.1.3 Product composition
Microplate Reader is composed of light source system, monochromator system, sample
chamber, detector and other components.
1.1.4 Application scope
PKL PPC 142 can read microplates used in Enzyme-Linked Immunology Assay (ELISA) for the
qualitative and quantitative analysis, which is used widely for immunoassay in clinic lab, blood
center, hospital, university, institute or other academic organization.
Page 2 of 41
Microplate Reader Operation Manual
1.2.1 Sign
Sign Note
1.2.2 Labels
Label Note
Manufacturing Date
Page 3 of 41
Microplate Reader Operation Manual
1.4 Training
To get a good performance for our Microplate Reader, free training course will be available for
you in our factory. For training details and schedule, please contact with us.
2.2 Safety
a. Biological contamination: All test samples, control materials, calibration materials, etc.,
should be regarded as contagious, and any components contacted with sample shall be
regarded as contagious as well, you should wear gloves when contact ;
b. High temperature: You can change the lamp after you turn off the device more than 10
minutes when lamp becomes normal temperature.
c. Physical injury: Don’t put hands to the moving mechanism when PKL PPC 142 is working.
2.3 Notes
a. If the device does not work in the way specified by the manufacturer, it may damage the
protective function in device itself;
b. Do not place the device where it is difficult to turn it off.
3. Installation
Page 4 of 41
Microplate Reader Operation Manual
3.5 Installation
Open box and take out device from the box, and keep it away from water and unstable desk;
Open box and take out the accessory kit and power adapter cable. Connect one end of the
power adapter back into the instrument's power jack and the other end with the power cord;
Check whether parts in device are loose or not;
Connect the power cable into plug.
Page 5 of 41
Microplate Reader Operation Manual
1 2 3 4 3 5 6 7 8
9 10
11
1. Light source 2. Optical filter 3. Optical lens 4. Object 5. Convert optical signals to
electrical signals 6. Bottom position industrial control panel 7. Upper position industrial control
panel 8. LCD 9. Optical filter movement mechanism 10. Microplate and sample
movement mechanism 11.Microplate and sample inlet
Page 6 of 41
Microplate Reader Operation Manual
2 .5
7
3
8
Page 7 of 41
Microplate Reader Operation Manual
Note: Model PKL PPC 142 is equipped with embedded PCB (include power supply board).
5. Operation
Page 8 of 41
Microplate Reader Operation Manual
5.1 Turn it on
Press the red button in the front panel, which will make the indicator lamp light.
After Built-in computer starts, it will enter windows XP and show following interface.
Double-click lnk HSIMM as below to start data analysis software for microplate reader.
Software had already been installed in model PKL PPC 142 in factory.
Instrument will automatically do self-test to light source, filter and movement mechanism
respectively. If there is anything unusual, screen will show that there is an error. The instrument
would be used better after start up 15 minutes.
Page 9 of 41
Microplate Reader Operation Manual
b) User right: Admin is the manager who can create, delete other user. Admin can operate
some system operation as below shows:
c) Main menu:
After starting HSIMM software,it will show the main menu as follows:
Title :It shows the name of software is “Micro-plate system(v2.1)”.
Menu: Pull-down menu to execute the main function
5.3 Test
Click Test button to input/edit sample information, measuring sample in different ways: Rapid
test, Single Test, Multi-test.
Sample information can be inputted, modified, checked and deleted. After you click sample
information, it shows as below:
Page 10 of 41
Microplate Reader Operation Manual
When you open Sample Information window, it only displays today’s sample information by
system default. Same sample No. is not allowed in the same day, if date is different, same
sample No. is allowed, which means you can’t name same sample No. in the same day. All
samples information in the same day are displayed on the right side, while each sample details
are displayed on the left side.
1) Save
Input sample information (such as name, Sex, Age, specimen type, blood type, ID No. and
submission doctor, etc.) in the left. You can also click drop-down menu in the chosen item, and
then press ENTER to complete input. After inputted, click Save button. Sample number will be
automatically added, and current sample information will be automatically displayed on the
right.
2)Delete
If need to delete sample information, you can click Delete after the sample No. had been
selected in the right list. Prompt will show as below:
Page 11 of 41
Microplate Reader Operation Manual
Click yes to delete what you have selected, and click No to cancel deletion.
3)Modify:
Select the sample you want to modify in the left column or the right sample information list.
When you single click modify, sample number and reading date become gray which indicates
that it can’t be operated since then. But all other functions can be modified. In this case, click
save S can save modified information. sample number and reading date will be restored.
4) Check(Find)
Choose starting date and ending date. Click Find, it can show all sample information during
selected time. When sample No.or name is shown, click check, all information you selected will
be shown.
Page 12 of 41
Microplate Reader Operation Manual
2) Plate No.:It can be set by users according to date and project. The plate No. will be
increased automatically.
3)Plate style: Plate style is the positional relation of each well in the microplate. In the right
drop-down list, style can be chosen. If it changes, you can modify it through the read kinds
setting (details refers to sample type setting in the manual), in order to get the plate type you
want. Regular plate style can be saved for future usage.
4) Choose wavelength: When you choose single wavelength, WL1 should be selected. When
you choose double wavelength, both WL1 and WL2 should be selected, and they can not be
the same wavelength.
5)Judgment formula:According to Cutoff calculation formula in reagent manual, input blank
value 、negative control , positive control and their calculation relationship with Cutoff value.
To choose Positive value or Cutoff value by single clicking < or > . Proportions between
Little Positive and CUTOFF can be set by inputting percentage parameters.
Page 13 of 41
Microplate Reader Operation Manual
If drag the mouse vertically, the sample number will be arranged vertically. After complete one
drag, it will show next number.
Click above NC (negative), PC (positive), HP(high positive), LP(little positive), QC( quality
control) and BLK (blank). Select one you want, and then drag it into main interface. In this
process, it should be corresponding with the real position of reagent in microplate.
7) Click Test, device start testing according to project parameters. Later measurement data is
immediately displayed in the data display. If you click Close Window before start measuring, it
will cancel measurement and return to the main interface.
8)Click Save to store the results, Export, Print can be used only after results had been saved
successfully.
9)If you need to test another new plate, please click Test new plate to enter initial Rapid test
interface, repeat same steps as above.
10) Go back main interface: After Rapid test, please click <close window>to go back the main
interface.
★Note:Rapid test can’t do quantitative testing, whose testing results can only be
printed out or exported.
Page 14 of 41
Microplate Reader Operation Manual
In this case, you should click Edit in the main menu, select Project Parameters. In the
parameter setting interface, set the project parameters ( refers to 5.5 Edit on page 27). Click
single test and it will show following interface.
1)【Test Item】
a. Plate No: Use system default number.
b. Plate Style: in the right drop-down list, style can be chosen. If it changes, you can modify it
Page 15 of 41
Microplate Reader Operation Manual
through the read kinds setting (details refers to sample type setting in the manual), in order to
get the plate type you want. Regular plate style can be saved for future usage.
c. Single click to select the project horizontally and vertically, if select the project horizontally,
the colors corresponding to the line A-H. If vertical row is selected, the corresponding item
on the color shown in column 1-12.
d. Project selection: project module is selected by clicking, in the data list. It is displayed by the
color module.
2)【Test Parameters】
a. Click < Test Parameters >, enter test parameters interface.
Page 16 of 41
Microplate Reader Operation Manual
b. Choose wavelength: All should be set in the Edit menu instead of here. When you select
double wavelength, WL1 and WL2 should be selected separately. When you select single
wavelength, WL1 should be selected.
c. Cutoff, positive judgment can only be displayed but can’t be modified here. If you need
change it, you can return the Edit menu, to change the project parameters.
d. Testing type settings: Select sample, input Start No. Same project can’t use same sample
number in different projects. When the panel has multi-well blank or and multi-well negative
control, software will automatically calculate the average value.
Drag the mouse to select the test area, if horizontally drag the mouse, the sample number will
be horizontally arranged as below:
If you vertically drag the mouse, sample number will be vertically arranged.
Click NC (negative), PC (positive), HP(high positive), LP(little positive), QC( quality control)
and BLK (blank). Check their boxes below to add what you want. Click the appropriate location;
you can set the corresponding negative, positive, strong positive, little positive, quality control,
Page 17 of 41
Microplate Reader Operation Manual
blank holes. In this process, it need be corresponding with the microplate status.
Note: Blank and the control of projects involved in the calculation (including Positive,
Negative, Standard, QC) should be set completely.
3)Click Test, instrument panel executes on ELISA optical density test. And then current
measurement data will be calculated, together with optical density result of holes in the data
display. If click "Close Window" before you start measuring, it returns the main interface,
and it will cancel the measurement.
Note: If setting is incomplete, please click Test, instrument will prompt parameter is
set incompletely. You can recall historical records or re-enter the missing holes
in the OD value of control. The method is same as “Recalculate."
4)Click Recalculate. If it is qualitative test, it will show following Standard
Page 18 of 41
Microplate Reader Operation Manual
Historical value in the same item is shown in the left list, including BLK, NC, PC, LP, HP, etc. In
the right side will show the current value in the control wells. Users can click some historical
control data to modify, or control thoroughly the data in the optical density of a well. Click Select
to use modified control data, to recalculate the current board test results on the optical density.
Current plate result together with revised control value can be saved.
5) Click Recalculate. If it is quantitative test, it shows following standard curve as follows:
Page 19 of 41
Microplate Reader Operation Manual
Users can change the curve type or the corresponding standard optical density to re-fit the
curve. You can also select the historical curve to modify current project type and standard
optical density. And then click Select to recalculate the optical density in current test results.
Results are kept together with the modified curve.
6)Click Save, test data can be saved, click Export (Note: it can export to EXCEL ), can be
saved in the data into EXCEL, and also you can directly click Print. Once it is completed,
click Close Window to return the main interface.
In this case, you should click Edit in the main menu, select Project Parameters. In the
parameter setting interface, set the project parameters up there as 5.5 Edit in page 27 says.
Page 20 of 41
Microplate Reader Operation Manual
1)【Test Item】
a. Plate No: Use system default number.
b. Plate Style: in the right drop-down list, style can be chosen. If it changes, you can modify it
through the read kinds setting (details refers to sample type setting in the manual), in order to get
the plate type you want. Regular plate style can be saved for future usage.
c. Single click to select the project horizontally and vertically, if select the project horizontally, the
colors corresponding to the line A-H. If vertical row is selected, the corresponding item on the
color shown in column 1-12.
d. Project selection: project module is selected by clicking, in the data list. It is displayed by color.
Page 21 of 41
Microplate Reader Operation Manual
2)【Test Parameters】
a. Click Test Parameters, enter test parameters interface
b. Choose wavelength: All should be set in the Edit menu instead of here. When you select double
wavelength, WL1 and WL2 should be selected separately. When you select single wavelength, WL1
should be selected.
c. Cutoff, positive judgment can only be displayed but can’t be modified here. If you need change it,
you can return the Edit menu, to change the project parameters.
d. Testing type settings: Select sample, input Start No. Same project can’t use same sample
number in different projects. When the panel has multi-well blank or and multi-well negative control,
software will automatically calculate the average value.
Drag the mouse to select the test area, if horizontally drag the mouse, the sample number will be
horizontally arranged as below:
Page 22 of 41
Microplate Reader Operation Manual
Click NC (negative), PC (positive), HP(high positive), LP(little positive), QC( quality control)
and BLK (blank). Check their boxes below to add what you want. Click the appropriate location;
you can set the corresponding negative, positive, strong positive, little positive, quality control,
blank holes. In this process, it need be corresponding with the microplate status.
3)Click Test, instrument can identify the wavelength of all items in the current microplate. And
it execute absorbance detection at all wavelengths, same wavelength detects only one time,
whose test result data and the absorbance data will be shown at the data area. If click Close
Window before you start it, it return the main interface, which means it can’t be measured.
4)Click Save, test data can be saved, click Export (Note: it can export to EXCEL only when
you save), can be saved in the data into EXCEL, and also you can directly click Print .Once it is
completed, click Close Window to the main interface.
Page 23 of 41
Microplate Reader Operation Manual
5.4 Report
In the main interface, click drop-down menu Report:
Report by plates is a uniform reporting of all specimens for single rapid test absorbance,
multi-tests for microplate, negative and positive, quantitative value and the S / CO values, etc
on the same microplate, the specific formats as follows,
Page 24 of 41
Microplate Reader Operation Manual
Page 25 of 41
Microplate Reader Operation Manual
Page 26 of 41
Microplate Reader Operation Manual
5.5 Edit
Click drop-down menu Edit, select Project Parameters to enter the program parameter setting
interface.
1) Qualitative measurement
Click Project Parameters as below:
Page 28 of 41
Microplate Reader Operation Manual
Page 29 of 41
Microplate Reader Operation Manual
b. Quadratic
The quadratic function fits the best parabola to the data. The parabola is a curved line based on the
equation:
y = (A * x^2) + (B * x) + C, where C is the intercept, B is the slope of the curve at the intercept, and A
is the measure of the curvature of the parabola.
The quadratic fit is most appropriate when the standard curve has a tendency to curve up or down.
c. Semi-Log
The semi-log function fits the best straight line to a set of data for log(X) plotted against Y. The
resulting curve displayed will be a straight line with the X axis drawn in logarithmic scale. The
equation for a semi-log fit is:
Y = B + [A* log10(X)], where B is the y-intercept of the line and A is the slope (note: log in this
equation is the common or base-10 logarithm).
d. Log-Log
The log-log function fits the best straight line to the set of data which consists of the logarithm of the
readings on the Y-axis (the response) and the logarithm of the dose on the X-axis. The resulting
display will be a straight line with both axes drawn in logarithmic scale. The equation for the log-log
fit is:
log10(Y) = B + [A * log10(X)], where B is the log10Y-intercept of the line for log10(X) = 0 and A is the
slope (note: log in this equation is the common or base-10 logarithm).
Page 30 of 41
Microplate Reader Operation Manual
e. Point-to-Point
A linear equation is fit to each pair of data points. Fit parameters are not given. The point-to-point
curve fit is a linear fit composed of pieces that assume a linear relationship between each pair of
data points. The line segment defined by each pair is used to interpolate data between those points.
Since there are multiple line segments, fit parameters are not shown on the graph.
Page 31 of 41
Microplate Reader Operation Manual
f. Four-Parametric
The Four-Parametric function fits the best straight line to the data. The equation for this fit has
the form of:
Y = (A-D)/(1+(X/C)^B) + D,
Most of reagent reaction should use Four-Parametric function according to some researches.
Page 32 of 41
Microplate Reader Operation Manual
5.6 System
1)Click the "System"
2) Click "Initialization", reset the instrument components, including light bulbs, filters and return
the initial tray position.
3)Filter: Click the "System" drop-down menu, select "Optical Filter" and click into the filter
screen, showing the four filters microplate reader corresponding wavelength are 405nm,
450nm, 492nm, 630nm, according to user’s requirements, different wavelength filters can
be added, you can fill in the 5,6,7,8 positions to add the wavelengths, click the "Save" to
save settings, click "Close" the filter interface, and then back the main interface.
Page 33 of 41
Microplate Reader Operation Manual
4) Absorbance correction: Click the "System" drop-down menu, select "Absorbance correction"
and click into the absorption correction interface, where the manufacturers engineers need to
adjust the password to enter the program, mainly for calibration when there are large difference
between the wavelengths of light density values and the standard value of standard substance.
This function is not available to users. New equipment is not to be calibrated when it is send
out.
5)Operating parameters: Click "System" drop-down menu, select <operating parameters> and
click to enter the operating parameters of interface
Operating parameters include the setting name of the units, connecting port settings and
sampling parameters. Name of the users’ organization will be printed on the results of the report;
connecting port is parameters for two-way communication between external control computer
equipment and the instrument. Sampling parameters is amplification needed to process the
original signals received by the software
WARNING: For changing, it requires setting parameters of the manufacturer’s engineers
Page 34 of 41
Microplate Reader Operation Manual
Select time period for records need to be cleared, when the project list is not marked with"√",
the record, can’t be deleted. Click "Confirm", the system pops up the following dialog box.
Click "Confirm", the system pops up, that record deleted successfully.
7)Backup Database: In order to ensure data security, you can regularly back up the system
database to other disks of the computer
Page 35 of 41
Microplate Reader Operation Manual
Click the button to select the backup to backup directory and its name, click "confirm (S)"
instrument tips Backup successfully.
8)Recovery Database: When users need to query the historical records, just back up and
restore the database, the system software can query, print historical records.
Note: When restoring the database, the system will coverage the data currently being
used, so make sure to back up the database before restoring.
Click the button to select the database to be recovered , click "Sure(S)", the system will prompt
Page 36 of 41
Microplate Reader Operation Manual
9)Commissioning Panel: Click the "System” in drop-down menu, select " Commissioning Panel
" and click to enter the external pass test interface, mainly for engineers in the
maintenance of equipment after the equipment used in debugging. This function is not opened
to the user, only used by the commissioning engineer debug.
6. Daily Maintenance
6.1 Maintenance
1) In order to ensure continued stability and accuracy of PKL PPC 142 Microplate Reader, It
should be avoided interference any part of the optical system. Optical path of adjustment error
can affect measurement results.
2) Maintaining the optical system clean to ensure the accuracy of the results of normal function
and should avoid liquid flow into the equipment. Prevent dust completely from endogenous
substances , do not touch the lens surfaces, filters, photoelectric detector with fingers.
detector with hands, clean upper lens with the above methods.
Note: Do not use allethrolone to clean plastic lens, softly clean.
3) After clean the clean upper lens, replace the bracket of photoelectric detector.
Note: Must replace the pillar sensor into the slot.
4) Replace the fixed frame with a gasket of the bracket of photoelectric detector, check the
alignment of edge of the fixed frame, ensure the fixed frame of the bracket of photoelectric
detector well replaced.
Note: Do not fold the wires of circuit board.
5) Put the bottom wires back in place. Turn the cover screws well. Screws connect cover and
the bottom of device to cut off the circuit interference.
6) Use lint-free cloth or lens paper to clean the filters, note that when you wipe the filters, do not
use liquid.
6.4 Replace spare parts
Equipment replacement and purchase of spare parts are required to provide and change
by manufacturers or their agents, the following details of the replacement parts:
Drive mechanism of a. Poor transmission a. Check the belt of tray loose or the belt
6 microplate tray does mechanism pulley stuck
not wove b. Motor damaged b. Check if the motor damaged
Filter wheels do not a. Poor movement a. Check the motor rotor is damaged or
7 move b. Motor damaged stuck by friction
b. Check if the motor damaged
Big movement noise a. Poor lubrication a. Lubricating
b. Not tighten the transmission b. Check the motor lock and the motor
mechanism pulley is loose
8
c. Drive belt tightness is c. Adjust pulley position
inappropriate d. Contact service provider
d. Mechanical deformation
Block microplate a.Microplate placed in a. Reset microplate
unproper place b.Choose the appropriate type of microplate
b.Special specification c. Contact service provider
9
microplate
c.Mechanical structure
re-placed loose or deformed
Mouse / Keyboard a. Bad contact a. Check the external socket and
failure b. The mouse / keyboard motherboard socket
10 damaged b. Replace the mouse / keyboard
c. The mouse / keyboard c. Contact service provider
interface is damaged
Page 39 of 41
Microplate Reader Operation Manual
d. motherboard mouse /
keyboard interface damage
Page 40 of 41