Technical Data
Short description
The M100 DR DIN rail transmitter is a 2-wire transmitter with HART® communication capability for analytical
measurements. The M100 DR is a single-channel, multi-parameter transmitter compatible with ISM sensors
for measuring pH/ORP, pH/pNa, oxygen and conductivity. Thanks to its compact design the M100 DR
requires only a small installation space in the plant.
Featuring Plug and Measure functionality the M100 DR reduces maintenance efforts and allows simplified
sensor installations up to 80 m (260 ft) from the transmitter, resulting in significant cost savings. The
transmitter configuration and integration of sensor diagnostics into asset management tool is possible
thanks to the integrated HART® protocol. The support of all major asset management tools ensures maximum
compatibility and easy integration of sensor diagnostics.
Features
––DIN Rail mounting, suitable for 35 mm wide DIN rail systems
––Compact housing, 22.5 mm width
––Displayless
––Multi-parameter transmitter (pH/ORP, pH/pNa, oxygen, conductivity)
––1 Analog output (4 to 20 mA with HART®)
––HART® communication as standard
––Configuration via HART handheld or other HART Asset Management tools
––ISM functionality
––Sensor diagnostics DLI, ACT and TTM via HART
Specifications 3
Ordering information 6
2
Dimension drawings
99,0 mm 22,5 mm
3.90" 0.89"
Dimensions and installation drawings
5 6 7 8 4 3 2 1
TB2 TB1
5 6 7 8 4 3 2 1
ISM- ISM- Aout- Aout+
A In+ A In- D In- D In+
Data GND HART- HART+
4.5"
114,5 mm
Mettler-Toledo AG
CH-8902 Urdorf, Switzerland
M100/2H DR
Feeding Voltage: 14 to 30 V DC
Ambient Temp: –20 to 60 °C !
Made by METTLER TOLEDO in China
P/N: 30127720 S/N: P0000000003
M100 DIN rail transmitter
Specifications M100 DIN rail transmitter
Amperometric oxygen
Measurement parameters Dissolved oxygen: Saturation or concentration and temperature
Oxygen measuring ranges ––Saturation: 0 to 500 % air, 0 to 200 % O2
––Concentration: 0 ppb (µg/L) to 50.00 ppm (mg/L)
Polarization voltage – 550 mV or – 674 mV (configurable)
Temperature input NTC 22 kΩ, Pt1000, Pt100
Temperature compensation Automatic
Temperature measuring range – 10 to + 80 °C (+ 14 to + 176 °F)
Max. sensor cable length 80 m (260 ft)
Calibration ––Configuration tool: Process and 1-point
––iSense software: 1-point
Conductivity 4-e
Measurement parameters Conductivity/resistivity and temperature
Conductivity ranges 0.01 to 650 mS /cm (1.54 Ω x cm to 0.1 MΩ x cm)
Chemical concentration curves NaCl: 0 – 26 % @ 0 °C to 0 – 28 % @ +100 °C
NaOH: 0 – 12 % @ 0 °C to 0 – 16 % @ + 40 °C to 0 – 6 % @ +100 °C
HCl: 0 – 18 % @ – 20 °C to 0 – 18 % @ 0 °C to 0 – 5 % @ +50 °C
HNO3: 0 – 30 % @ – 20 °C to 0 – 30 % @ 0 °C to 0 – 8 % @ +50 °C
H2SO4: 0 – 26 % @ –12 °C to 0 – 26 % @ + 5 °C to 0 – 9 % @ +100 °C
H3PO4: 0 – 35 % @ + 5 °C to + 80 °C
TDS ranges NaCl, CaCO3
Temperature input Pt1000
Temperature measuring range – 40 to + 200 °C (– 40 to + 392 °F)
Max. sensor cable length 80 m (260 ft)
Calibration ––Configuration tool: Process, 1-point and 2-point
––iSense software: 1-point and 2-point
3
Specifications M100 DIN rail transmitter
Environmental specifications
Storage temperature – 40 to + 70 °C (– 40 to + 158 °F)
Ambient temperature – 20 to + 60 °C (– 4 to + 140 °F)
operating range
Relative humidity 0 to 95 % non-condensing
EMC According to EN 61326-1 (general requirements)
Emission: Class B, Immunity: Class A
CE mark The measuring system is in conformity with the statutory requirements
of the EC Directives. METTLER TOLEDO confirms successful testing of
the device by affixing to it the CE mark.
Mechanical specifications
Dimensions See page 2.
Suitable DIN rail systems 35 mm wide
Weight 0.5 kg
Material PA-FR
Enclosure rating IP 20
4
Connector terminal definition, LED functionality M100 DIN rail transmitter
5 6 7 8 1 2 4 3 2 1
TB2 TB1
5 6 7 8 4 3 2 1
ISM- ISM- Aout- Aout+
Data GND
A In+ A In- D In- D In+
HART- HART+
1 Terminal block TB2
2 Terminal block TB1
Terminal Description
TB2 5 ISM-Data Sensor input, see table “M100 DR parameter fit guide” on page 6.
6 ISM-GND
7 Ain+ Analog input 4 to 20 mA (for pressure compensation)
8
Mettler-Toledo
Ain–
AG
CH-8902 Urdorf, Switzerland
LED functionality
2 3
5
Ordering information M100 DIN rail transmitter
Ordering information
Transmitter
Description Order no.
M100/2H DR, 1-channel multi-parameter 30 127 720
Accessories
Description Order no.
iSense 2.0 full version 30 130 614
iSense 2.0 lite version Available for free
iSense 2.0 mobile version Available for free
iLink cable for iSense 52 300 383
M100/2H DR
ISM
pH/ORP •
pH/pNa •
Conductivity 4-e •
Amp. DO ppm/ppb/trace • 1)/• 2)/• 2)
1) INGOLD and THORNTON sensors
2) INGOLD sensors
6
Notes M100 DIN rail transmitter
7
METTLER TOLEDO Market Organizations
ISO ISO
9001
ISO 14001
ISO
certified certified
9001 14001
certified certified