FID-Tower NM PLUS
Operation Manual
This Operation Manual is applicable for the following Models:
Table of Contents
2. Introduction .......................................................................................... 5
2.1 Symbols Definition ....................................................................................................... 5
3. Safety .................................................................................................. 6
3.1 Personal safety instructions .......................................................................................... 6
4. Specifications ........................................................................................ 7
4.1 The most common options ............................................................................................ 8
5.7 SchemaTICs.............................................................................................................. 16
7. Operation ........................................................................................... 26
7.1 Hydrogen Part ........................................................................................................... 26
9. Maintenance ........................................................................................ 31
9.1 Routine maintenance ................................................................................................. 31
1. VERSION HISTORY
2. INTRODUCTION
The FID-T NM PLUS is a Hydrogen generator producing pure Hydrogen for analytical and industrial
applications. As an option, a Zero Air generator can be incorporated.
Purity and flow ranges are optimized for scientific and high-tech applications.
Symbol Description
Fuse.
3. SAFETY
Read all instructions and warnings. Not respecting the following instructions and warnings could
induce electrical shock, fire and/or serious injury to people.
Normal precautions as for any Hydrogen supply should be taken when using the
generator. DO NOT use in sealed or unventilated rooms.
DO NOT use near open flames or other sources of ignition.
Keep the power cable away from hot surfaces or oil. Damaged or entangled power
cables could increase the risk of electrical shock
Unplug the power cable only when the generator is turned off. The power cable
connector is the disconnecting device in case of an emergency, it must be easily
accessible at any time.
Only staff qualified by VICI DBS s.r.l. should service or repair the generator.
Always unplug the unit from the mains power supply before accessing the internal
components for replacement.
If the generator is used in a way not specified by VICI DBS s.r.l., the protection
provided by the equipment could be compromised.
The FID-T NM PLUS is designed to produce Hydrogen for laboratory and industrial applications.
The unit must only be operated for this purpose, according to the specifications and instructions
provided in this manual. The following warnings must be observed at all times. Critical damages
on the device could occur if the following warnings are neglected.
4. SPECIFICATIONS
General
Operating Temperature +15°C to +35°C
Operating relative Humidity 0-80%, non-condensing
Casing Dimension (WxHxD) 140x490x580 mm
Input Voltage 110-120 / 220-240V AC (Use voltage selector and adapt fuses) 50-60 Hz
Transient overvoltage until overvoltage category II (Accept transient
Electrical Protection overvoltage occurring on the network)
General earth connection is done by the power cable. Fuse F2:
5x20mm, 4.0 or 6.3 A T (see fuses and voltage setting)
Hydrogen Part
Hydrogen Flowrate FID-T NM-100 PLUS: 100 ml/min
FID-T NM-160 PLUS: 160 ml/min
FID-T NM-250 PLUS: 250 ml/min
Purity 99.99996% :FID-T NM-300 PLUS: 300 ml/min
FID-T NM-500 PLUS: 500 ml/min
100, 160, 250, 300, 500, 600,ml/min
FID-T NM-600 PLUS: 600
FID-T 1350
1000, NM-1000 PLUS: 1000 ml/min
FID-T NM-1350 PLUS: 1350 ml/min
Max Outlet Pressure 11.0 bar (160 psi) Standard version DB-xxxx
9.0 bar (130 psi) Japanese version DB-xxxx-JP
Power Consumption Max 500W
Pin Description
3+4 Standby
Cascading
4.1.2. Cascading
Gas Line
Pressure controler
RS-485 port 1 RS-485 port 2 RS-485 port 1 RS-485 port 2 RS-485 port 1 RS-485 port 2
For part-numbers for the cascading option please consult the list above
The cascading function will only work properly if the gas outlets of all the generators
are connected to the same gas line.
Ver. 1.02.0000 - 10 -
JASCO ANALITICA SPAIN, S.L.
Rafael Bergamin 16
28043 Madrid – Spain
Tel : 34 91 2342454
info@jasco-spain.com
www. jasco-spain.com
Ver. 1.02.0000 - 11 -
JASCO ANALITICA SPAIN, S.L.
Rafael Bergamin 16
28043 Madrid – Spain
Tel : 34 91 2342454
info@jasco-spain.com
www. jasco-spain.com
5.4 WEIGHT
Ver. 1.02.0000 - 12 -
JASCO ANALITICA SPAIN, S.L.
Rafael Bergamin 16
28043 Madrid – Spain
Tel : 34 91 2342454
info@jasco-spain.com
www. jasco-spain.com
Do a water purge before moving the generator. Do not handle the generator as long as
it is still full of water.
Ver. 1.02.0000 - 13 -
JASCO ANALITICA SPAIN, S.L.
Rafael Bergamin 16
28043 Madrid – Spain
Tel : 34 91 2342454
info@jasco-spain.com
www. jasco-spain.com
Check the setting of the voltage selector at the rear of the unit. The set voltage is indicated in the
white window.
To change the voltage or fuse, proceed as follows:
• Use a small flat screwdriver to push the snap clip of the voltage selector insert to remove
it.
• To change the voltage, spin the white holder. To change the fuse value, just replace it. Use
5x20mm fuses.
• Put the voltage selector insert back and check that the white window shows the chosen
voltage.
• Use a screwdriver to unscrew the cap from fuse holder F3. Replace the fuse and screw
the cap back in its original position.
F3
Ver. 1.02.0000 - 14 -
JASCO ANALITICA SPAIN, S.L.
Rafael Bergamin 16
28043 Madrid – Spain
Tel : 34 91 2342454
info@jasco-spain.com
www. jasco-spain.com
Ver. 1.02.0000 - 15 -
JASCO ANALITICA SPAIN, S.L.
Rafael Bergamin 16
28043 Madrid – Spain
Tel : 34 91 2342454
info@jasco-spain.com
www. jasco-spain.com
5.7 SCHEMATICS
5.7.1. Fluidics
The FID-T NM PLUS generator produces pure Hydrogen (and Oxygen as a by-product) through
the electrolysis of water. The key element of the generator is an electrochemical cell assembly
which contains a solid polymer electrolyte. No free acids or alkaline are used. Deionized or pure
distilled water is the only liquid which may come into contact with the cell. As this is consumed it
must be refilled from time to time as required. The generated Hydrogen gas is accumulated in the
Hydrogen/water separator and the desiccant housing. The internal pressure is controlled by a
pressure transducer. The outlet pressure is controlled by a proportional valve. The Hydrogen is
dried by passing through the automatic dryer. The Hydrogen then passes through the outlet port
at the rear.
Ver. 1.02.0000 - 16 -
JASCO ANALITICA SPAIN, S.L.
Rafael Bergamin 16
28043 Madrid – Spain
Tel : 34 91 2342454
info@jasco-spain.com
www. jasco-spain.com
5.7.2. Electronics
1. Display Connection
2. USB Connection
3. I/O Board Connection
4. Water Level/Quality Connection
5. Water Pump Connection
6. Fans (NM) Fans+Outlet Valve (PG) Connection
7. Dryer (NM) Connection
8. Power Switching Connection
9. Negative PEM Cell Pole Connection
10. Positive PEM Cell Pole Connection
Ver. 1.02.0000 - 17 -
JASCO ANALITICA SPAIN, S.L.
Rafael Bergamin 16
28043 Madrid – Spain
Tel : 34 91 2342454
info@jasco-spain.com
www. jasco-spain.com
All units have been carefully inspected before they were shipped. Visual checks for damage and
functional tests should be performed upon receipt. Any damage must immediately be noted and
reported. The generator must only be returned according to the shipping instructions provided.
Ver. 1.02.0000 - 18 -
JASCO ANALITICA SPAIN, S.L.
Rafael Bergamin 16
28043 Madrid – Spain
Tel : 34 91 2342454
info@jasco-spain.com
www. jasco-spain.com
1 Hydrogen Generator
1 Deionizer Bag
1 Power Cable
Ver. 1.02.0000 - 19 -
JASCO ANALITICA SPAIN, S.L.
Rafael Bergamin 16
28043 Madrid – Spain
Tel : 34 91 2342454
info@jasco-spain.com
www. jasco-spain.com
The FID-T NM PLUS generator must be placed on a flat, level, vibration- and shock-free surface.
Do not place the generator near a heat source, as this may cause the generator to overheat. The
unit should not be in contact with any other objects on any side, and the Air inlet must not be
blocked. Leave at least 30 cm of free space at the rear for ventilation. Do not operate the generator
in a sealed or unventilated room, near open flames or other sources of ignition. Do not operate
the generator at temperatures below zero degrees C. Operation is guaranteed at temperatures
between +15 and +35°C.
Normal precautions as for any Hydrogen supply should be taken when using the
generator.
DO NOT use in sealed or unventilated rooms.
DO NOT use near open flames or other sources of ignition.
Pure dry Hydrogen at regulated pressure is available at the Hydrogen outlet port at the rear of
the generator. This port must be connected to 1/8” tubing using a stainless-steel or copper
Swagelok connector. Teflon connectors are not suitable. The pressure at this port is adjusted and
shown on the display. The Hydrogen release port at the rear of the unit can be connected to an
exhaust hood or other venting system.
Plug the oil-free compressed Air source to the inlet (AIR IN) of the generator. Use the Air filter
provided in the package.
Zero Air is available at the 1/8” rapid fitting (AIR OUT)
The line from the release port should never be connected in such a way that back
pressure can develop.
CAUTION!
Ver. 1.02.0000 - 20 -
JASCO ANALITICA SPAIN, S.L.
Rafael Bergamin 16
28043 Madrid – Spain
Tel : 34 91 2342454
info@jasco-spain.com
www. jasco-spain.com
Check the setting of the voltage selector at the rear of the unit.
The set voltage is indicated in the white window on the inlet power filter. To change the voltage,
proceed as follows:
Using a small screwdriver, remove the voltage selector insert.
Replace the voltage selector insert so that the white arrow points to the correct voltage.
The FID-T NM PLUS has an internal water tank included in its chassis. The tank must be refilled
with deionized or distilled water with a conductivity below 2µS.
Follow the steps in order to fill the automatic water tank system correctly.
• Connect the tubing of the external tank to the refill fitting at the front of the FID-T NM
Plus
• The external tank must be placed above the generator. (The water will refill the tank
system by gravity).
• Disconnect the external tank when the water level has reached the MAX mark on the
indicator.
Ver. 1.02.0000 - 21 -
JASCO ANALITICA SPAIN, S.L.
Rafael Bergamin 16
28043 Madrid – Spain
Tel : 34 91 2342454
info@jasco-spain.com
www. jasco-spain.com
Ver. 1.02.0000 - 22 -
JASCO ANALITICA SPAIN, S.L.
Rafael Bergamin 16
28043 Madrid – Spain
Tel : 34 91 2342454
info@jasco-spain.com
www. jasco-spain.com
In order to protect the cell from bad water quality, it is necessary to use a deionization bag.
Inspect the bag thoroughly for holes or tears, indicated by loose deionizer beads on the outer
surface. If the bag is damaged in any way, discard and replace it with a new one. Only use original
parts. Wash the deionizer bag with deionized water before proceeding.
Insert the free end of the “T” fastener through one of the small holes at the edge of the
cover/holder (red circle), until it is securely fastened. The bag should not block the outlet at the
bottom of the tank. Once in place, the bag should not be allowed to dry out.
The internal tank is located behind the front panel:
Ver. 1.02.0000 - 23 -
JASCO ANALITICA SPAIN, S.L.
Rafael Bergamin 16
28043 Madrid – Spain
Tel : 34 91 2342454
info@jasco-spain.com
www. jasco-spain.com
Use the arrows “up” and “down” to select the desired outlet
pressure.
Then press “Exit”
Press “Start”
The generator will first build up the internal pressure. Wait until
it reaches 100% before going on.
At this point, make sure that you have connected the Hydrogen
outlet to your GC-system. At least make sure you have a
restriction and that the outlet is not fully open.
Ver. 1.02.0000 - 24 -
JASCO ANALITICA SPAIN, S.L.
Rafael Bergamin 16
28043 Madrid – Spain
Tel : 34 91 2342454
info@jasco-spain.com
www. jasco-spain.com
Press the “H2 ON/OFF” button to turn the Hydrogen part off.
Press the “AIR ON/OFF” button to turn the Zero Air part off.
Ver. 1.02.0000 - 25 -
JASCO ANALITICA SPAIN, S.L.
Rafael Bergamin 16
28043 Madrid – Spain
Tel : 34 91 2342454
info@jasco-spain.com
www. jasco-spain.com
7. OPERATION
The actual status of the generator is shown on the main screen of the graphic display.
The main screen has three options at the bottom, corresponding to the three buttons on the unit,
which are used to run the various functions and access the configuration and diagnostics of the
generator, following the tree structure shown in the figure below.
Please note: Read the Specifications before you change the pressure settings. The allowed
pressure
maximum is depending on the generator model.
Set Clock
Autorefill Low Pressure Error
Autorefill Pressure rise
Autorefill Timeout Pressure drop delay
Auto start
Scroll Select Exit Beeper
Cascading mode
N. Gen H2 in cascading
ID number
Remote start /stop mode
Autorefill
Autorefill timeout Remote relay mode
Autorefill
No 60s Pre alarms in alarm list
Pressure units
Exit Temp units
Exit Volume units
Lock keyboard
Set default values
Ver. 1.02.0000 - 26 -
JASCO ANALITICA SPAIN, S.L.
Rafael Bergamin 16
28043 Madrid – Spain
Tel : 34 91 2342454
info@jasco-spain.com
www. jasco-spain.com
*BE CAREFUL: If set at 100%, the generator will never go to External Pressure Alarm.
Ver. 1.02.0000 - 27 -
JASCO ANALITICA SPAIN, S.L.
Rafael Bergamin 16
28043 Madrid – Spain
Tel : 34 91 2342454
info@jasco-spain.com
www. jasco-spain.com
Item Description
Production Tot. Total production of Hydrogen
Operating Time (h) Total number of hours the unit is operational
Wat. Quality (S) Actual water conductivity
Int. Press.: Actual internal pressure of the unit
Cell Voltage (V) Actual cell voltage
Cell Current (A) Actual cell current
PS.Temp. Actual temperature of the power supply
Cell Voltage Peak (V) The maximum voltage the cell ever reached
PS. Temp. Peak The maximum temperature the power supply reached
Pump Flow (l/min) Actual flow of water
Item Description
Dryer Status Shows the temperatures of the drying columns and the
position of the cycle
System Test This offers several test functions - see below
Internal Leak Test The outlet valve is closed, the pressure is set to the max.,
when the pressure reaches the max., production is stopped
and the pressure drop is measured during 2 minutes. The
pressure drop should be below 0.7 bar or 10 PSI
Important: The leak test will only work if the generator is
in standby mode.
External Leak Test Works similar to the internal leak test except that the outlet
valve is open and the test pressure is equal to the set
pressure
This function can be used to test the external gas lines
Important: The external leak test will only work if the
generator is in standby mode.
Flow Test This function sets the outlet valve to provide a certain flow
adjustable from 0 to 100% of maximum flow.
Important: This function will only work if the generator is
in standby mode. It will take approx. 2 minutes to get a
stable flow!
Rinse Flow Test This test checks the amount of gas flow, for the dryer
regeneration
Complete System Test Combination of all tests above
Hardware Options Not available for user (password protected)
Ver. 1.02.0000 - 28 -
JASCO ANALITICA SPAIN, S.L.
Rafael Bergamin 16
28043 Madrid – Spain
Tel : 34 91 2342454
info@jasco-spain.com
www. jasco-spain.com
The FID-T NM PLUS Zero Air generator option utilizes a 3-stage process to purify ambient Air into
Zero grade Air. All main components are manufactured with high grade stainless steel, oriented
neatly in the cabinet for easy access and service.
Stage 1: Prefiltration and Pressure Regulation
Air is initially passed through a highly efficient combination of coalescing/particulate pre-filters to
ensure that only quality Air enters the system. The pre-filter removes bulk moisture and larger
particulates greater than 5 microns in size. The filter is supplied with an automatic float drain to
remove periodic moisture build up.
The Air pressure is then regulated with an adjustable forward pressure regulator, pre-set to 7 bar.
The minimum allowable pressure is about 3 bar; the maximum 7.5 bar.
Stage 2: Hydrocarbon and CO Removal
Air is next channelled into a stainless steel catalyst chamber where any background hydrocarbons
and carbon monoxide are oxidized. This chamber is filled with a highly efficient platinum catalyst,
heated to a very high temperature to ensure hydrocarbon removal to less than 0.1 ppm.
Stage 3: Final Filtration
A final high-grade filter is used to remove 99.99% of particulates greater than 0.5 microns in size.
This will ensure that no harmful particles will be able to enter your GC-system.
Shutdown Procedure
Before performing any service or maintenance to the Zero Air generator, be sure to complete the
following steps:
• Turn the power off using the main switch of the unit.
• Disconnect the outlet tubing, and freely run Air through the unit for several minutes to cool
the catalyst. Allow a minimum of 2 hours to cool the catalyst before replacing it.
• Shut the Air supply off.
• De-pressurize the unit and disconnect the power source before servicing.
Ver. 1.02.0000 - 29 -
JASCO ANALITICA SPAIN, S.L.
Rafael Bergamin 16
28043 Madrid – Spain
Tel : 34 91 2342454
info@jasco-spain.com
www. jasco-spain.com
Please contact your service support and report the S/N of the unit and the number of the error.
Ver. 1.02.0000 - 30 -
JASCO ANALITICA SPAIN, S.L.
Rafael Bergamin 16
28043 Madrid – Spain
Tel : 34 91 2342454
info@jasco-spain.com
www. jasco-spain.com
In the event of any faults or damage, first notify the agent or distributor who supplied the unit.
Please also provide full details of the problem, including the model and serial number. Instructions
will then be provided for the service or the return of the unit. If the fault is due to misuse of the
unit, all repair and shipping costs are to be paid by the customer. All other costs are borne by the
customer, except as otherwise expressly agreed upon.
WARNING
If the unit needs to be transported, make sure that the water tank is completely
empty, and place the plug (supplied with the unit) on the Oxygen vent at the rear of
the unit. Close the water tank with the cap. Use the ORIGINAL packaging. The unit
should be transported in an upright position; this warning should be marked on the
outside of the packaging.
9. MAINTENANCE
With proper care and maintenance, your generator should provide you with years of trouble-free
operation. There are no adjustments to be made to the generator. The only routine service
operations are those described below.
Nonetheless, the generator should be inspected approximately every 2 years. Contact your
supplier.
The following section describes the maintenance operations required for the correct operation of
the Hydrogen generator.
The routine maintenance consists of:
• Fill the 7 lt internal water tank when needed
• Replace the deionization bag every 6 months
• Replace the water filter every 6 months
• Replace the filtering element of the dust filter every 6 months
Ver. 1.02.0000 - 31 -
JASCO ANALITICA SPAIN, S.L.
Rafael Bergamin 16
28043 Madrid – Spain
Tel : 34 91 2342454
info@jasco-spain.com
www. jasco-spain.com
IMPORTANT!
The manufacturer reserves the right to change or modify its products without prior
notice.
Ver. 1.02.0000 - 32 -
JASCO ANALITICA SPAIN, S.L.
Rafael Bergamin 16
28043 Madrid – Spain
Tel : 34 91 2342454
info@jasco-spain.com
www. jasco-spain.com
Part-Number Description
Ver. 1.02.0000 - 33 -
JASCO ANALITICA SPAIN, S.L.
Rafael Bergamin 16
28043 Madrid – Spain
Tel : 34 91 2342454
info@jasco-spain.com
www. jasco-spain.com
Part-Number Description
Ver. 1.02.0000 - 34 -
JASCO ANALITICA SPAIN, S.L.
Rafael Bergamin 16
28043 Madrid – Spain
Tel : 34 91 2342454
info@jasco-spain.com
www. jasco-spain.com
Part-Number Description
DB-NGC004-016 Main board for Zero Air part v.2 (from S/N: NGC130115441)
DB-RH200-019 In/Out air filters AF20 with Swagelok fitting for RACK/FID-S/FID-T
Ver. 1.02.0000 - 35 -
JASCO ANALITICA SPAIN, S.L.
Rafael Bergamin 16
28043 Madrid – Spain
Tel : 34 91 2342454
info@jasco-spain.com
www. jasco-spain.com
Ver. 1.02.0000 - 36 -