PUMP MODULE
CHANNEL
SELECT
8100 SERIES
PUMP
PAUSE
CHANNEL
OFF
RESTART
DIRECTIONS FOR USE
TABLE OF CONTENTS
INTRODUCTION
ABOUT THE SYSTEM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
INTRODUCTION
FEATURES AND DEFINITIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
SYMBOLS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
GETTING STARTED
WARNINGS AND CAUTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
CONTROLS AND INDICATORS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Unpacking Pump Module ................................................................................ 15
ATTACHING AND DETACHING CHANNELS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
GETTING STARTED
DISPLAYS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Main Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Dynamic Pressure Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
START-UP ...................................................................................................... 17
Powering On System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Responding to Maintenance Reminder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Selecting New Patient and Profile Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Entering Patient ID . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Modifying Patient ID . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
PREPARING INFUSION ....................................................................................... 17
ALARMS, ERRORS,
Flo-Stop® Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Priming Primary Administration Set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
MESSAGES
Loading and Removing Primary Administration Set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Setting Up Primary Administration Set for Gravity Infusion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
PRIMARY MODE - BASIC INFUSION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Starting Rate/Volume Infusion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Starting Volume/Duration Infusion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Pausing Infusion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Restarting Infusion Following Infusion Complete . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Changing Rate or VTBI During Infusion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Stopping Infusion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
MAINTENANCE
Selecting Pressure Mode - Pump / Selectable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Viewing and Clearing Volume Infused . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
AutoRestart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
SECONDARY MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Changing Primary Infusion Parameter During Secondary Infusion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Stopping Secondary Infusion and Returning to Primary Infusion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
CHANGING PRIMARY SOLUTION CONTAINER . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
CHANNEL LABELS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Selecting Channel Label . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Removing Channel Label . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
POWERING OFF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Powering Off System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
APPENDIX
TABLE OF CONTENTS i
GETTING STARTED (Continued)
SETTING UP DRUG CALCULATION ......................................................................... 39
Drug Calculation Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Using Guardrails® Drug Library . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Using Non-Library Drug . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
PROGRAMMING BOLUS DOSE ............................................................................. 49
Using Guardrails® Drug Library Calculation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Using Non-Library Drug Calculation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Stopping Bolus Dose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Restoring Bolus Dose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
ANESTHESIA MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Enabling Anesthesia Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Disabling Anesthesia Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
DELAY OPTIONS .............................................................................................. 64
Delaying Infusion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Scheduling a Callback . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Pausing Infusion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
MULTIDOSE MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Programming with Volume/Duration Enabled . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Programming with Volume/Duration Disabled . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
REVIEWING SERIAL NUMBER . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
REVIEWING SOFTWARE VERSION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
ALARMS, ERRORS, MESSAGES
DEFINITIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
AUDIO CHARACTERISTICS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
ALARMS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
ERRORS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
MESSAGES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
MAINTENANCE
SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
CONFIGURABLE SETTINGS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
System Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
Shared Infusion Settings (Pump Module and Syringe Module) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
Pump Module Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
CLEANING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
INSPECTION REQUIREMENTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
SERVICE INFORMATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
WARRANTY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
APPENDIX
TRUMPET AND START-UP CURVES ......................................................................... 95
ii TABLE OF CONTENTS
GENERAL CONTACT INFORMATION
Customer Advocacy
For clinical and technical questions, feedback, and troubleshooting assistance.
Phone, toll-free, within the United States and Canada: (800) 854-7128, Ext. 7812
E-Mail: CustomerFeedback@alarismed.com
Technical Support
For technical information related to maintenance procedures and service manual support.
For more detailed information, refer to the “Service Information” section of this document.
T H I S PA G E
I N T E N T I O N A L LY
LEFT BLANK
INTRODUCTION
INTRODUCTION
and monitoring system intended for use in today’s growing
professional healthcare environment, for use in adult, pediatric
and neonatal care.
The Medley™ Medication Safety System consists of the
Programming Module (8000 Series), the Guardrails® Safety
Software, and up to four detachable modules (or “channels”),
which provide infusion or monitoring capabilities.
INTRODUCTION 1
About the System (Continued)
This document provides directions for use for the Medley™
Pump Module. Read all instructions, for both the Pump Module
and Programming Module, before using the Medley™ System.
The Medley™ Pump Module uses a wide variety of Medley™
System/ Gemini Administration Sets. The sets are designed for
use with the Pump Module as well as for gravity-flow, stand-
alone use. For specific administration set instructions, refer to
the directions for use provided with the set. For set priming and
loading instructions, refer to the “Start-Up” section in the
“Getting Started” chapter of this document.
Contraindications: None known.
2 INTRODUCTION
Features and Definitions
Refer to the “Alarms, Errors, Messages” chapter of this Directions for Use for the definitions of various
alerts. Refer to the Medley™ Programming Module Directions for Use for system features and definitions.
INTRODUCTION
AutoRestart The AutoRestart feature is part of the Medley™ System’s advanced
Downstream Occlusion Detection system. If enabled, the AutoRestart feature
minimizes nuisance patient-side occlusion alarms caused by momentary
kinking of tubing, IV pushes, etc.
Bolus Dose The Bolus Dose mode allows a bolus infusion to be programmed using either
the Guardrails® Drug Library or the drug calculation feature. The bolus
infusion can be programmed with or without a continuous infusion following
the bolus.
Channel Labels The Channel Labels feature is available when the Profiles feature is enabled. It
provides a hospital-defined list of labels, displayed in the Channel Message
Display, and identifying the channel with the solution being infused, the
catheter location, or other helpful information.
Delay Options The Delay Options feature allows the system to be programmed to delay the
start of an infusion a) for up to 120 minutes or b) for a specific time up to
23 hours 59 minutes. A callback for a programmed delay can be scheduled to
give an alert Before an infusion is to be initiated, After an infusion is
completed, Before and After an infusion, or no alert (None).
Dynamic Pressure Display The Dynamic Pressure Display appears on the Main Display. If enabled, it
graphically displays the current patient-side occlusion pressure set point and
the current patient-side operating pressure for that module. (Reference
“Displays” section in “Getting Started” chapter for additional “Dynamic
Pressure Display” information.)
Free Flow Protection All Medley™ System/Gemini administration sets utilize a unique clamping
device, the Flo-Stop® Device, to prevent inadvertent free flow when the
administration set is removed from the instrument.
Guardrails® Clinical Advisory A Guardrails® Clinical Advisory is a visual message that appears when a
designated drug is selected, to remind a clinician of specific hospital standards
of practice when programming an IV medication. A specific clinical advisory
can be associated with a selected drug within any of the patient care profiles.
Guardrails® Drug Library The Guardrails® Drug Library feature is a drug calculation mode available when
the Profiles feature is enabled. It provides a hospital-defined list of drugs and
concentrations appropriate for use in as many as 10 profiles. Using the Drug
Library automates programming steps, including the drug name, drug amount
and diluent volume, and activates the hospital-established best-practice
Guardrails® Limits.
INTRODUCTION 3
Features and Definitions (Continued)
KVO Rate Adjust The KVO Rate Adjust option is used to select the KVO (Keep Vein Open) rate
(0.1 to 20 mL/h allowed). This determines the rate of fluid flow after an
“Infusion Complete” occurs. The KVO rate will never exceed the infusion rate.
Multidose Mode The Multidose Mode option allows 2 - 24 doses to be programmed at equally
spaced intervals on the same Pump Module over a 24-hour period. This mode
is designed to allow delivery of multiple, equal doses from the same IV
container at regularly scheduled intervals.
Restore To simplify programming, the Restore feature can be used to recall previous
rate and volume settings for the same patient. This option is only available if
the patient is not new and the system is powered up within 8 hours of last
usage.
Secondary Infusions Dual rate sequential piggyback (Secondary) infusions may be infused at
delivery rates and volumes independent of the primary infusion parameters.
Automatic changeover occurs to the primary infusion parameters when the
secondary infusion is complete if a Medley™ System/Gemini Check Valve
Administration Set is used.
4 INTRODUCTION
Symbols
Canadian and U.S. Certification Mark: Products bearing this mark have been tested and
certified in accordance with applicable U.S. and Canadian electrical safety and performance
+ 75
standards (CSA C22.2 No. 601.1, UL 2601-1 and IEC 60601–2–24).
INTRODUCTION
Electrical Shock Protection Rating: Type CF, Defibrillation-proof
Manufacturing Date: Number adjacent to symbol indicates the month and year of
MM-YYYY manufacture.
Only CAUTION: Federal (U.S.A.) law restricts this device to sale by or on the order of a physician.
Single-Use
Product DOES NOT contain a particular element; such as, LATEX = administration set is
latex-free.
XX Drops per milliliter specification for product will be identified on drop symbol.
Product incorporates SmartSite® Needle-Free Valve Ports and should not be accessed by a
needle.
Expiration date for product will be identified near hour glass symbol.
INTRODUCTION 5
T H I S PA G E
I N T E N T I O N A L LY
LEFT BLANK
6 INTRODUCTION
GETTING STARTED
WARNING
A warning is an alert to potential serious outcomes (death,
injury or serious adverse events) to the patient or user.
CAUTION
GETTING STARTED
A caution is an alert to take special care for the safe and
effective use of the device.
WARNING
This instrument is designed to stop fluid flow under alarm
conditions. Periodic patient monitoring must be performed to
ensure the infusion is proceeding as expected.
GETTING STARTED 7
Warnings and Cautions (Continued)
WARNING
The Guardrails® Safety Software incorporates dosing limits and
instrument configuration parameters based on hospital
protocol. The software adds a test of reasonableness to drug
programming based on the limits defined by the hospital.
Qualified personnel must ensure the appropriateness of the drug
dosing limits, the compatibility of the drugs, and the
performance of each instrument, as part of the overall infusion.
Potential hazards include drug interactions, and inappropriate
delivery rates and pressure alarms.
WARNING
When loading a data set with the Guardrails® Safety Software,
ensure the correct profile (for patient care area) is selected prior
to starting an infusion. Failure to use the appropriate profile
could cause serious consequences.
WARNING
This Medley™ Pump Module is a positive displacement delivery
system, capable of developing positive fluid pressures to
overcome widely varying resistances to flow encountered in
practice, including resistances to flow imposed by small gauge
catheters, filters and intra-arterial infusion. It is neither designed
nor intended to detect infiltrations and will not alarm under
infiltration conditions.
WARNING
Hospital/facility personnel must ensure the compatibility of the
drugs as well as the performance of each module as part of the
overall infusion. Potential hazards include drug interactions,
inaccurate delivery rates, inaccurate pressure alarms and
nuisance alarms.
WARNING
Do not use the Medley™ System in close proximity of Magnetic
Resonance Imaging (MRI).
WARNING
Use only Medley™ System/Gemini Series Administration Sets.
The use of any other set may cause improper instrument
operation, resulting in an inaccurate fluid delivery or other
potential hazard. For a list of compatible sets, refer to the Set
Compatibility Card (provided separately).
8 GETTING STARTED
Warnings and Cautions (Continued)
WARNING
To prevent a potential free-flow condition, ensure no extraneous
object (for example, bedding, tubing, glove) is enclosed or
caught in the Medley™ Pump Module door.
WARNING
The use of positive displacement infusion devices ported
together with gravity flow infusion systems into a common IV
site may impede the flow of common “gravity only” systems,
affecting their performance. Hospital/facility personnel must
ensure the performance of the common IV site is satisfactory
under these circumstances.
GETTING STARTED
WARNING
References in this document to specific drugs and drug doses
are for illustration purposes only. Refer to specific drug product
labeling for information concerning appropriate administration
techniques and dosages.
Parallel Infusions
There are no contraindications regarding the use of the
Medley™ System with any other positive displacement
infusion device when ported together into a common IV site
location.
User Precautions
GETTING STARTED 9
Warnings and Cautions (Continued)
NOTES:
• If applicable, attach syringe to SmartSite® Needle-Free Valve Port
and aspirate minute air bubbles.
• In an emergency, SmartSite® Valve may be accessed by a needle
and will leak if punctured. To access port with needle without
causing leakage, attach a “PRN” adapter of sufficient length to
SmartSite® Needle-Free Valve Port.
10 GETTING STARTED
Warnings and Cautions (Continued)
Artifacts
It is normal for an infusion device to produce nonhazardous
currents when infusing electrolytes. These currents vary
proportional to the infusion device flow rate. When the ECG
monitoring system is not functioning under optimal
conditions, these currents may appear as artifacts, simulating
actual ECG readings. To determine if ECG abnormalities are
caused by patient condition or the ECG equipment, place the
infusion device on hold. If the ECG readings become normal,
the ECG equipment requires attention. Proper setup of the
GETTING STARTED
ECG equipment should eliminate these artifacts. Reference
the appropriate ECG monitoring system documentation for
instructions on setup and maintenance.
Dropping/Jarring
Should an instrument be dropped or severely jarred, it should
be immediately taken out of use and inspected by qualified
service personnel, to ensure its proper function prior to reuse.
GETTING STARTED 11
Warnings and Cautions (Continued)
12 GETTING STARTED
Controls and Indicators
Rate Display
GETTING STARTED
STANDBY
Channel Message Display
INFUSE
RM
ALA
Channel Identification
NEL
CHAN T
SELEC
Door Handle
Channel Release Latch
GETTING STARTED 13
Controls and Indicators (Continued)
Platen
IUI Connector, Right
Upper Occluder
Flo-Stop® Recess
Tubing Keeper
Air-in-Line Sensor
14 GETTING STARTED
Installation
GETTING STARTED
not cut or torn.
GETTING STARTED 15
Displays
Main Display
16 GETTING STARTED
Displays (Continued)
Midtown Hospital
Adult ICU
VTBI = 250.0 mL
GETTING STARTED
CAUTION
Although the dynamic pressure
display bars for the Medley™ Syringe
Module and Pump Module both use
the full width of the screen for
display, they each represent different
ranges. The Pump Module’s range is
50 mmHg to 525 mmHg.
Start-Up
Preparing Infusion
The primary administration set must be primed before use (see WARNING
“Priming Primary Administration Set” section). It can be loaded Use only Medley™ System/Gemini
into the Pump Module to deliver a large volume infusion (see Series Administration Sets. The use of
“Loading and Removing Primary Administration Set” section) or any other set may cause improper
it can be set up to deliver a gravity infusion (see “Setting Up instrument operation, resulting in an
Primary Administration Set for Gravity Infusion” section). inaccurate fluid delivery or other
potential hazard. For a list of
The Flo-Stop® Device is a tubing fitment that is part of all compatible sets, refer to the Set
Medley™ System/Gemini administrations sets (see “Flo-Stop® Compatibility Card (provided
Device” section). separately).
GETTING STARTED 17
Preparing Infusion (Continued)
Flo-Stop® Device
6. To prime tubing and clear air from injection sites and tubing
fitments, slowly open roller clamp.
18 GETTING STARTED
Preparing Infusion (Continued)
GETTING STARTED
4. Holding tubing below pumping segment, insert Flo–Stop® CAUTION
Fitment into recess below mechanism, with arrow pointing When reloading an administration
into Pump Module. (See figure at end of this section.) set, leave the Flo-Stop® Fitment in the
closed position (see “Flo-Stop® Device
section).
GETTING STARTED 19
Preparing Infusion (Continued)
20 GETTING STARTED
Preparing Infusion (Continued)
GETTING STARTED
Primary Mode - Basic Infusion
NOTES:
• The illustrations in this section assume the following:
♦ Drug Calculation, Dynamic Pressure Display, Profiles, and
Volume Duration configurable settings are enabled.
♦ Delay Options configurable setting is disabled.
• If Delay Options is enabled, the PAUSE soft key becomes DELAY
OPTIONS.
• The RESTORE soft key appears only if a previous infusion was
programmed for the same patient.
GETTING STARTED 21
Primary Mode - Basic Infusion (Continued)
RESTORE EXIT
1. To enter flow rate, press RATE soft key and use numeric data A Infusion Setup
entry keys. RATE _ _ _ mL/h
VTBI _ _ _ _ mL
2. To enter VTBI, press VTBI soft key and use numeric data entry A Infusion Setup
keys. RATE 40 mL/h
VTBI mL
3. Attach administration set to patient’s vascular access device.
>Select VTBI
VOLUME
DURATION
4. Verify correct infusion parameter entry and press START soft A Infusion Setup
key. RATE 40 mL/h
VTBI 250 mL
NOTE: The infusion may be paused by pressing the PAUSE
soft key. Refer to “Pausing Infusion” section.
>Press START
VOLUME SECOND-
PAUSE DURATION
START
ARY
22 GETTING STARTED
Primary Mode - Basic Infusion (Continued)
GETTING STARTED
Channel Rate Display changes to KVO rate.
INFUSION COMPLETE–KVO scrolls in Channel Message
Display.
KVO appears on Main Display.
♦ If infusion was delayed, COMPLETE scrolls in Channel
Message Display and appears on Main Display.
VTBI _ _ _ _ mL
2. To enter VTBI, press VTBI soft key and use numeric data entry A Infusion Setup
keys. RATE = mL/h
VTBI _ _ _ _ mL
DURATION h min
>Select VTBI
RESTORE RATE
VOLUME
GETTING STARTED 23
Primary Mode - Basic Infusion (Continued)
3. To enter volume duration, press DURATION soft key and use A Infusion Setup
numeric data entry keys. RATE = mL/h
VTBI 1000 mL
• Rate is automatically calculated. DURATION h min
>Select DURATION
RATE
VOLUME
5. Verify correct infusion parameter entry and press START soft A Infusion Setup
key. RATE = 125 mL/h
VTBI 1000 mL
• During infusion: DURATION 8:00 hh:mm
♦ Green Infusing Status Indicator illuminates.
♦ Rate appears in Channel Rate Display.
♦ VTBI counts down on Main Display. >Press START
RATE SECOND-
START
PAUSE VOLUME ARY
NOTE: To view infusion Time Left, press CHANNEL SELECT A Infusion Setup
key. To return to previous screen, press START soft key. RATE 125 mL/h
VTBI 875 mL
• At completion of infusion: Time Left: 07 h 00 min
♦ If infusion ends in KVO:
An audio prompt sounds.
Red Alarm Status Indicator flashes. >Press START
VOLUME AUDIO
INFUSED ADJUST
24 GETTING STARTED
Primary Mode - Basic Infusion (Continued)
Pausing Infusion
GETTING STARTED
• PAUSED appears on Main Display. >Press START
OR
VOLUME AUDIO
INFUSED ADJUST
• Press CHANNEL SELECT key and then press START soft key A Infusion Setup
(on Programming Module). RATE 40 mL/h
VTBI 500 mL
>Press START
VOLUME SECOND-
PAUSE DURATION ARY START
GETTING STARTED 25
Primary Mode - Basic Infusion (Continued)
RESTORE EXIT
4. Verify parameters are valid and press START soft key. A Infusion Setup
RATE 40 mL/h
NOTE: To change a restored parameter: VTBI 500 mL
a. Press applicable soft key, VTBI or RATE.
b. Enter desired parameter using Up/Down Arrows for rate
titration, or numeric data entry keys.
c. Press START soft key. >Press START
VOLUME SECOND-
PAUSE DURATION ARY START
26 GETTING STARTED
Primary Mode - Basic Infusion (Continued)
>Press START
GETTING STARTED
PAUSE VOLUME SECOND-
DURATION ARY START
3. To enter desired parameter, use Up/Down Arrows for rate Infusion Setup
titration or use numeric data entry keys. RATE _50 mL/h
VTBI 240 mL
4. Verify correct infusion parameter entry and press START soft
key.
>Press START
Stopping Infusion
NOTES:
• If no other channel is active, the system powers down when the
CHANNEL OFF key is released.
• To interrupt the power down sequence, quickly press any one of
the numeric keys on the Programming Module.
GETTING STARTED 27
Primary Mode - Basic Infusion (Continued)
Multidose
Pressure Limit - P
Channel Labels
EXIT
EXIT
CONFIRM
28 GETTING STARTED
Primary Mode - Basic Infusion (Continued)
1. To view volume infused, press VOLUME INFUSED soft key. Midtown Hospital
Adult ICU
PAUSED
• Total volume infused (primary + secondary), and time VTBI = 497.0 mL
and date volume infused was last cleared, display for VTBI = 57.0 mL
each channel. VTBI = 249.0 mL
GETTING STARTED
2. To view primary and secondary volume(s) infused, press Volume Infused
PRI/SEC VOLUME soft key. TOTAL VOLUME (mL) LAST
CLEARED
08:00
401.1 2002-03-10
07:30
42.5 2002-03-11
08:00
478.1 2002-03-10
12:00
789.1 2002-03-10
0.0 0.0
0.0 0.0
0.0 0.0
0.0 0.0
>Select Channels to Clear
or Press CLEAR ALL
PRI/SEC CLEAR MAIN
VOLUME ALL SCREEN
GETTING STARTED 29
Primary Mode - Basic Infusion (Continued)
AutoRestart
30 GETTING STARTED
Secondary Mode
GETTING STARTED
NOTE: Prepare the secondary container and administration set.
Lower the primary container using the hanger included with the
Underestimating the volume will
secondary set. cause the remaining secondary
solution to be infused at the
primary rate; overestimating will
result in the primary solution
being infused at the secondary
rate. Multiple doses from a single
container are not possible.
• The clamp on the secondary
administration set must be
opened. If the clamp is not
opened, the fluid will be delivered
from the primary container.
• The secondary administration set
must be primed prior to beginning
the secondary infusion.
GETTING STARTED 31
Secondary Mode (Continued)
8. Press SECONDARY soft key and continue with next step. A Infusion Setup
RATE 40 mL/h
VTBI 240 mL
OR
>Press START
SECOND-
PAUSE ARY
START
9. To enter secondary infusion rate, press RATE soft key and A Infusion Setup
use numeric data entry keys. SECONDARY
RATE _ _ _ mL/h
VTBI mL
>Select RATE
32 GETTING STARTED
Secondary Mode (Continued)
>Select VTBI
GETTING STARTED
11. Open clamp on secondary administration set. A Infusion Setup
SECONDARY
12. Verify correct infusion parameters and press START soft key. RATE 100 mL/h
VTBI 55 mL
>Press START
GETTING STARTED 33
Secondary Mode (Continued)
>Press START
>Press START
SECOND-
PAUSE ARY START
>Press START
SECOND-
PAUSE ARY START
>Press START
34 GETTING STARTED
Secondary Mode (Continued)
GETTING STARTED
>Press START
>Press START
NOTE: The SEC to PRI alert does NOT sound when the
>Select Yes or No
infusion is manually ended and returned to primary.
GETTING STARTED 35
Changing Primary Solution Container
6. To enter VTBI, press VTBI soft key and use numeric data entry Infusion Setup
keys. RATE 40 mL/h
VTBI 240 mL
7. Open roller clamp.
>Press START
Channel Labels
Multidose
Pressure Limit - P
Channel Labels
EXIT
36 GETTING STARTED
Channel Labels (Continued)
NOTE: To view additional labels, press a soft key next to a 3% NaCl F-J
letter group to navigate through alphabet, and/or PAGE UP CVVHDF Dialysate K-O
and PAGE DOWN soft keys. D5 1/2 NS P-T
PAGE
EXIT
DOWN
GETTING STARTED
A Channel Label Display
Adult ICU
Peripherial Art Line P
Replacement Solution Q
Swan R
Triple Lumen S
T
PAGE PAGE
UP BACK DOWN
>Press START
GETTING STARTED 37
Channel Labels (Continued)
Multidose
Pressure Limit - P
Channel Labels
EXIT
D5 1/2 NS P-T
Epidural U-Z
CLEAR PAGE
EXIT
LABEL DOWN
>Press START
38 GETTING STARTED
Powering Off
The drug calculation can be set up for a drug stored in the WARNINGS
Guardrails® Drug Library or for a non-library drug, as described • The Drug Calculation feature is to
in the following sections. To access the drug library, a hospital- be used only by personnel properly
GETTING STARTED
defined best-practice data set must be uploaded, using the trained in the administration of
Guardrails® Safety Software, and the Profiles feature must be continuously infused medications.
enabled. Extreme caution should be
exercised to ensure the correct
Drug Calculation Parameters entry of the drug calculation
infusion parameters.
The Medley™ System uses the following parameters, entered
• References in this document to
during the drug calculation setup procedure:
specific drugs and drug doses are
• Bolus dose duration: Time period over which bolus dose is for illustration purposes only.
to be administered. Refer to specific drug product
• Bolus dose units: Units used in calculating bolus dose. labeling for information
Bolus dose units are selected from alternatives provided. concerning appropriate
administration techniques and
• Diluent volume: Volume of fluid used as diluent for drug
dosages.
(mL).
• Dosing units: Units used to calculate continuous infusion
drug dose. Dosing Units are selected from alternatives
provided.
• Drug amount: Amount of drug in IV container (gram, mg,
mcg, mEq, or units).
• Patient weight: Weight of patient (kg); this is an optional
parameter that is not needed unless drug dose is normalized
for patient weight.
• Time units: Time base for all calculations (minute, hour, or
day).
GETTING STARTED 39
Setting Up Drug Calculation (Continued)
The bolus dose, drug dose, and flow rate parameters are
calculated using the above parameters, as follows:
• Bolus dose = bolus dose x patient weight (if used).
• Bolus dose administration rate (INFUSE AT:):
When duration is entered = total dose / duration in minutes.
When Max Rate is used = Max Rate / 60 x concentration.
• Bolus dose duration = bolus VTBI / bolus rate.
• Bolus dose VTBI = bolus dose / drug concentration.
• Bolus rate = bolus VTBI / duration.
• Continuous drug dose = flow rate x drug concentration
(normalized for patient weight if specified by entering a
patient weight).
• Continuous flow rate = drug dose / drug concentration
(normalized for patient weight if specified by entering a
patient weight).
• Drug concentration = drug amount / diluent volume.
• Total bolus dose:
Bolus dose not weight-based = bolus dose entered.
Bolus dose weight-based = bolus dose x patient weight.
40 GETTING STARTED
Setting Up Drug Calculation (Continued)
GETTING STARTED
>Select an Option or EXIT
RESTORE EXIT
next to a letter group to navigate through alphabet, and/or Dopamine 400mg/250mL P-T
PAGE UP and PAGE DOWN soft keys. Dopamine 800mg/250mL U-Z
>Press Yes or No
• If Yes was selected and facility has defined a Clinical A Guardrails Drug Library
Lidocaine
information has been noted and continue programming, Watch IV site carefully for
signs of extravasation.
press CONFIRM soft key.
CONFIRM
GETTING STARTED 41
Setting Up Drug Calculation (Continued)
TIME
Not Used
NEXT
confirm.
[Conc]: 8 mg/mL
>Select Rate or Dose
SETUP BOLUS
42 GETTING STARTED
Setting Up Drug Calculation (Continued)
VTBI is entered.
GETTING STARTED
• In the Drug Calculation mode, the system infuses at the
calculated rate rounded to the nearest one-hundredth of a
mL per hour (as displayed on the Channel Programming
screen). The rate shown in the Channel Rate Display will
be rounded to the nearest one-tenth of a mL per hour.
10. Verify parameters are correct and press START soft key. A Guardrails Drug Library
Lidocaine
CONTINUOUS INFUSION
RATE
67.5 mL/h
VTBI 250 mL
DOSE 9 mg/min
[Conc]: 8 mg/mL
>Press START
NOTE: If the programmed continuous dose infusion is outside A Guardrails Drug Library
Lidocaine
the Guardrails® Soft Limit for that care area, a prompt appears Dose exceeds Yes
Guardrail limit of
before programming can continue. If the Yes soft key is 8 mg/min. No
pressed, programming continues; if the No soft key is pressed, Proceed?
the infusion needs to be reprogrammed.
>Press Yes or No
the Guardrails® Hard Limit for that care area, a prompt Dose exceeds Reprogram
Guardrails hard
appears before programming can continue. The infusion limit of 8 mg/min.
needs to be reprogrammed.
>Press REPROGRAM
GETTING STARTED 43
Setting Up Drug Calculation (Continued)
44 GETTING STARTED
Setting Up Drug Calculation (Continued)
GETTING STARTED
>Select an Option or EXIT
RESTORE EXIT
>Select Drug/Concentration
DRUG
LIBRARY
GETTING STARTED 45
Setting Up Drug Calculation (Continued)
TIME
UNITS
DOSING
UNITS
DRUG
LIBRARY
TIME
UNITS
DOSING
UNITS
DOSING
UNITS
DRUG
LIBRARY
46 GETTING STARTED
Setting Up Drug Calculation (Continued)
DOSING
UNITS
DRUG
LIBRARY
GETTING STARTED
13. To select time base for drug calculation, press either Min, A
A Drug Calculation
Hour, or Day soft key. DRUG
AMOUNT 250 mg
DILUENT
VOLUME 250 mL
PATIENT
WEIGHT 70 kg Min
TIME
UNITS >>>> Hour
DOSING
UNITS Day
[Conc]: 1000 mcg/mL
>Select Time Units
DRUG
LIBRARY
DOSING
UNITS >> mg/kg/min
[Conc]: 1000 mcg/mL
>Select the Desired
Dosing Units
DRUG
LIBRARY
DRUG
LIBRARY NEXT
GETTING STARTED 47
Setting Up Drug Calculation (Continued)
16. To make a rate or dose entry, press applicable soft key, RATE A
A Drug Calculation
or DOSE, and use numeric data entry keys (other value is CONTINUOUS INFUSION
SETUP BOLUS
17. To enter volume to be infused, press VTBI soft key and use A
A Drug Calculation
numeric data entry keys. CONTINUOUS INFUSION
RATE 21 mL/h
NOTES: VTBI _ _ _ _ mL
• At rates less than 10 mL/h, the rate is displayed to two DOSE _ _ _ _5 mcg/kg/min
decimal places, and the VTBI can be entered and is
displayed to two decimal places. [Conc]: 1000 mcg/mL
>Select VTBI
• The BOLUS soft key appears only if Bolus Dose is enabled
within the selected profile and a VTBI is entered. SETUP BOLUS
18. Verify parameters are correct and press START soft key. A
A Drug Calculation
CONTINUOUS INFUSION
• During infusion:
RATE 21 mL/h
♦ Green Infusing Status Indicator illuminates. VTBI 250 mL
♦ Rate appears in Channel Rate Display. DOSE 5 mcg/kg/min
48 GETTING STARTED
Setting Up Drug Calculation (Continued)
GETTING STARTED
♦ If infusion was delayed, COMPLETE scrolls in Channel
Message Display and appears on Main Display.
NOTES:
• If the Bolus Dose feature is enabled, the BOLUS soft key appears
in the Continuous Infusion screen and becomes active when a
VTBI is entered.
• The bolus VTBI cannot exceed the programmed continuous
infusion VTBI.
• Programming and starting a bolus dose deletes any programmed
delay.
• If no continuous rate is entered, the infusion will end when the
bolus has been delivered. No KVO infusion will follow.
GETTING STARTED 49
Programming Bolus Dose (Continued)
[Conc]: 8 mg/mL
>Press START
[Conc]: 10 units/mL
>Press START
• DOSE is highlighted.
NOTES:
• If the programmed continuous dose infusion is outside the
Guardrails® Soft Limit for that care area, a prompt appears
before programming can continue. If the Yes soft key is
pressed, programming continues; if the No soft key is
pressed, the infusion needs to be reprogrammed.
• If the programmed continuous dose infusion is outside the
Guardrails® Hard Limit for that care area, a prompt appears
before programming can continue. The infusion needs to
be reprogrammed.
50 GETTING STARTED
Programming Bolus Dose (Continued)
NOTE: After a bolus dose and weight (if used) are entered,
bolus VTBI and concentration [conc] alternate in the Main
Display.
• If no weight has previously been programmed in system A
A Guardrails Drug Library
Lidocaine
and bolus dose is weight-based, weight entry is empty. BOLUS DOSE
DOSE _ _ _ _ mg/kg
PATIENT
WEIGHT
GETTING STARTED
DURATION
[Conc]: 8 mg/mL
>Enter Bolus Dose
SETUP
[Conc]: 10 units/mL
>Enter Bolus Dose
SETUP
[Conc]: 8 mg/mL
>Select PATIENT WEIGHT
SETUP
GETTING STARTED 51
Programming Bolus Dose (Continued)
[Conc]: 8 mg/mL
>Select DURATION
SETUP
-- OR --
SETUP
[Conc]: 10 units/mL
>Press NEXT to Confirm
DRUG
LIBRARY NEXT
c. To change patient weight, use numeric data entry A Guardrails Drug Library
Heparin
keys. DRUG
AMOUNT 5000 units
d. Press NEXT soft key.
DILUENT
VOLUME 500 mL
PATIENT
WEIGHT _ _ 11 kg
TIME
UNITS Hour
DOSING units/kg/h
UNITS
[Conc]: 10 units/mL
>Press NEXT to Confirm
DRUG
LIBRARY NEXT
52 GETTING STARTED
Programming Bolus Dose (Continued)
Yes
No
>Select Yes or No
GETTING STARTED
e. Press BOLUS soft key. A
A Guardrails Drug Library
Heparin
CONTINUOUS INFUSION
RATE 20.0 mL/h
VTBI 500 mL
DOSE 18.18 unit/kg/h
[Conc]: 10 units/mL
>Press START
[Conc]: 10 units/mL
>Enter Bolus Dose
SETUP
SETUP
GETTING STARTED 53
Programming Bolus Dose (Continued)
profile and setup, and automatically calculate bolus DURATION _ _ min Max
Rate
duration, press Max Rate soft key. TOTAL DOSE = 355 mg
BOLUS VTBI = 44.4 mL
• TOTAL DOSE alternates with INFUSE AT rate. >Enter Duration
SETUP
BOLUS DOSE
• During infusion:
DOSE 5 mcg/kg
mg/kg
♦ Green Infusing Status Indicator illuminates. PATIENT
PATIENT
WEIGHT
WEIGHT 71 kg
kg (999 mL/h)
Max
♦ Rate appears in Channel Rate Display. DURATION 2 minmin
<1 Rate
INFUSE
BOLUS AT: 133.2
VTBI mg/min
= 50.0 mL
♦ Dose and drug name scroll in Channel Message [Conc]: 8
800
mg/mL
mcg/mL
Display. >Press START
Start to Begin
to Begin
Infusing Bolus Dose
CONT-
NOTE: If a continuous dose outside of the Guardrails® PAUSE SETUP INUOUS START
VOLUME AUDIO
INFUSED ADJUST
Midtown Hospital
Adult ICU
A Lidocaine
5 mg/kg Bolus
VOLUME AUDIO
INFUSED ADJUST
54 GETTING STARTED
Programming Bolus Dose (Continued)
GETTING STARTED
♦ System beeps twice.
♦ BOLUS DOSE COMPLETE scrolls in Channel Message
Display.
• DOSE is highlighted.
CONTINUOUS INFUSION
RATE 21 mL/h
VTBI 250 mL
DOSE 5 mcg/kg/min
GETTING STARTED 55
Programming Bolus Dose (Continued)
NOTE: After a bolus dose and weight (if used) are entered, DOSE _ _ 500 mcg/kg
bolus VTBI and concentration [conc] alternate in the Main PATIENT
WEIGHT mg
Display. DURATION mg/kg
4. Press soft key for appropriate unit of measure for dose. [Conc]: 1000 mcg/mL
>Select the Desired Dosing
NOTE: If mcg or mg is selected as the dosing unit, a PATIENT Units
profile and setup, and automatically calculate bolus DURATION _ _ min Max
Rate
duration, press Max Rate soft key. TOTAL DOSE = 500 mcg
[Conc]: 1000 mcg/mL
• TOTAL DOSE alternates with INFUSE AT rate. >Enter Duration
SETUP
Midtown Hospital
Adult ICU
A BOLUS VTBI = 0.5 mL
VOLUME AUDIO
INFUSED ADJUST
56 GETTING STARTED
Programming Bolus Dose (Continued)
NOTE: To see details during the bolus infusion, press the A Drug Calculation
GETTING STARTED
• At completion of bolus infusion:
♦ System beeps twice.
♦ BOLUS DOSE COMPLETE scrolls in Channel Message
Display.
GETTING STARTED 57
Programming Bolus Dose (Continued)
3. To stop bolus and start continuous infusion, press Yes soft A Guardrails Drug Library
Lidocaine
key. Stop Bolus and Start Yes
Continuous infusion?
No
>Press Yes or No
4. To stop continuous infusion, press and hold CHANNEL OFF Midtown Hospital
Adult ICU
key until a beep is heard (approximately 1.5 seconds). A VTBI = 250 mL
VOLUME AUDIO
INFUSED ADJUST
58 GETTING STARTED
Programming Bolus Dose (Continued)
GETTING STARTED
[Conc]: 8 mg/mL
>Press START
[Conc]: 8 mg/mL
>Enter Bolus Dose
RESTORE SETUP
BOLUS DOSE
DOSE 5 mcg/kg
mg/kg
PATIENT
PATIENT
WEIGHT
WEIGHT 71 kg
kg (999 mL/h)
Max
DURATION <1minmin
2 Rate
INFUSE
BOLUS AT: 133.2
VTBI mg/min
= 50.0 mL
[Conc]: 8
800
mg/mL
mcg/mL
>Press START
Start to Begin
to Begin
Infusing Bolus Dose
CONT-
PAUSE SETUP INUOUS START
RESTORE EXIT
GETTING STARTED 59
Programming Bolus Dose (Continued)
3. Verify parameters and press NEXT soft key. A Guardrails Drug Library
Lidocaine
DRUG
AMOUNT 2 gram
DILUENT
VOLUME 250 mL
PATIENT
WEIGHT Not Used
TIME
UNITS min
DOSING mg/min
UNITS
[Conc]: 8 mg/mL
>Press NEXT to Confirm
NEXT
4. Verify infusion parameters and press BOLUS soft key. A Guardrails Drug Library
Lidocaine
CONTINUOUS INFUSION
RATE
67.5 mL/h
VTBI 250 mL
DOSE 9 mg/min
[Conc]: 8 mg/mL
>Press START
[Conc]: 8 mg/mL
>Enter Bolus Dose
RESTORE SETUP
BOLUS DOSE
DOSE 5 mcg/kg
mg/kg
PATIENT
PATIENT
WEIGHT
WEIGHT 71 kg
kg (999 mL/h)
Max
DURATION 2 minmin
<1 Rate
INFUSE
BOLUS AT: 133.2
VTBI mg/min
= 50.0 mL
[Conc]: 8
800
mg/mL
mcg/mL
>Press START
Start to Begin
to Begin
Infusing Bolus Dose
CONT-
PAUSE SETUP INUOUS START
60 GETTING STARTED
Anesthesia Mode
GETTING STARTED
• Dose checking mode is set to “Smart”.
• Key-press audio is turned off.
• Tamper Resist Mode (panel locked) is not available.
• All Guardrails® Drug Library entries are available for selection.
• Bolus dose is automatically available for:
♦ drugs in Guardrails Drug Library that have bolus dose
®
>Select an Option or
EXIT
PAGE
EXIT DOWN
GETTING STARTED 61
Anesthesia Mode (Continued)
Anesthesia Mode
VOLUME AUDIO
INFUSED ADJUST
>Select an Option or
EXIT
PAGE
EXIT DOWN
62 GETTING STARTED
Anesthesia Mode (Continued)
GETTING STARTED
CONFIRM
CONFIRM
>Select Yes or No
GETTING STARTED 63
Delay Options
Delaying Infusion
Specifying by Minutes
>Press START
DELAY SECOND-
OPTIONS ARY START
BACK CONFIRM
64 GETTING STARTED
Delay Options (Continued)
GETTING STARTED
BACK CONFIRM
A Start in 20 min
VOLUME AUDIO
INFUSED ADJUST
VOLUME AUDIO
INFUSED ADJUST
GETTING STARTED 65
Delay Options (Continued)
>Press START
DELAY SECOND-
OPTIONS ARY START
BACK CONFIRM
3. If Current time displayed is correct, press CONFIRM soft key; A Delayed Options
otherwise, press Change Time and enter correct time. Time of Day
CONFIRM
BACK CONFIRM
66 GETTING STARTED
Delay Options (Continued)
GETTING STARTED
BACK CONFIRM
VOLUME AUDIO
INFUSED ADJUST
VOLUME AUDIO
INFUSED ADJUST
GETTING STARTED 67
Delay Options (Continued)
Scheduling a Callback
1. Prior to pressing CONFIRM soft key to initiate delay during A Delay Options 08:00
Delay for or Delay until programming process, press CALL Pause
BACK soft key. Delay for
Delay until _ 9:00
CALL
BACK None
2. Press soft key corresponding to desired callback option. A Delay Options 08:00
Pause
• Scheduled callback appears on Main Display.
Delay for Before
Before
Delay until 09:00 And After
CALL >>>>>> After
BACK
None
BACK CONFIRM
68 GETTING STARTED
Delay Options (Continued)
OR
VOLUME AUDIO
INFUSED ADJUST
GETTING STARTED
• If Delay for programming, delay period counts down on Midtown Hospital
Adult ICU 08:00
VOLUME AUDIO
INFUSED ADJUST
ended.
♦ Yellow Standby Status Indicator flashes.
♦ DELAY COMPLETE scrolls in Channel Message Display
and appears on Main Display.
VOLUME AUDIO
INFUSED ADJUST
GETTING STARTED 69
Delay Options (Continued)
4. To respond to a callback:
• Before callback
Press CHANNEL SELECT key and then START soft key.
OR
Press RESTART key.
• After callback
Press CONFIRM soft key.
• Before and After callback
Respond as indicated above for both Before and After.
Pausing Infusion
>Press START
DELAY SECOND-
OPTIONS ARY START
• Using the Pause function in the Delay Options screen is the Delay until
BACK CONFIRM
70 GETTING STARTED
Delay Options (Continued)
>Press START
DELAY SECOND-
OPTIONS ARY START
GETTING STARTED
Multidose Mode
NOTES: WARNINGS
• Since, by definition, a multidose infusion will not be infusing for • The Multidose feature is to be
a programmed period of time, it is assumed that another used only by personnel properly
infusing IV line will keep the vein open until the beginning of trained in using multidose
the first dose and between subsequent doses. There is no keep infusions.
vein open (KVO) infusion at the completion of a programmed
Delay until infusion. • Caution labels, which clearly
differentiate single dose and
• The Delay Options function for multidose infusions is similar to multidose containers, must be
Delay Options for continuous drug infusions, with the following utilized.
differences:
♦ Delay for (when scheduling a callback) option is not available • Single dose piggybacking systems
in Multidose Mode. employing check valve sets are not
designed for use with multidose
♦ Maximum allowable delay on a multidose infusion is 8 hours.
containers.
GETTING STARTED 71
Multidose Mode (Continued)
RESTORE EXIT
VTBI _ _ _ _ mL
Multidose
Pressure Limit - P
Channel Labels
EXIT
CONFIRM
72 GETTING STARTED
Multidose Mode (Continued)
DOSE
INTERVAL
# OF
DOSES
>Enter Rate
VOLUME
DURATION
GETTING STARTED
9. To enter volume to be infused for each dose, use numeric A Multidose 13:00
data entry keys. RATE = _ _ _ mL/h
VOLUME/
DOSE _ _ _ _ mL
DURATION
DOSE
INTERVAL
# OF
DOSES
>Enter Volume/Dose
RATE
VOLUME
10. To enter duration for each dose, press DURATION soft key A Multidose 13:00
and use numeric data entry keys. RATE = _ _ _ mL/h
VOLUME/
DOSE 50 mL
NOTE: RATE is calculated with each keystroke for DURATION. DURATION _ _ h _ _ min
DOSE
INTERVAL
# OF
DOSES
>Enter Duration
RATE
VOLUME
11. To enter time interval (1 to 24 hours) between doses, press A Multidose 13:00
DOSE INTERVAL soft key and use numeric data entry keys. RATE = 100 mL/h
VOLUME/
DOSE 50 mL
DURATION 00 h 30 min
DOSE
INTERVAL every _ _ h
# OF
DOSES
RATE
VOLUME
GETTING STARTED 73
Multidose Mode (Continued)
12. To enter number of doses, press #OF DOSES soft key and use A Multidose 13:00
DURATION 00 h 30 min
appears. DOSE
INTERVAL every _6 h
# OF
program.
13. To begin multidose infusion, press START soft key. A Multidose 13:00
RATE = 100 mL/h
VOLUME/
DOSE 50 mL
DURATION 00 h 30 min
DOSE
INTERVAL every 06 h
# OF
DOSES _4 doses
>Press START
• Main Display shows remaining VTBI for that dose. Midtown Hospital
Adult ICU 13:00
A VTBI = 50.0 mL
VOLUME AUDIO
INFUSED ADJUST
VOLUME AUDIO
INFUSED ADJUST
74 GETTING STARTED
Multidose Mode (Continued)
14. To see detail screen during or between infusions, press A Multidose 12:30
PAUSE START
GETTING STARTED
• Between infusions: A Multidose 12:00
♦ Number of doses completed and when next dose Rate = 100 mL/h
Volume/ = 50 mL
starts display. Dose
Duration = 00 h 30 min
♦ Yellow Standby Status Indicator illuminates. every 06 h x 04 doses
Doses completed = 1
Dose 2 Starts 19:00
>Press START
PAUSE START
GETTING STARTED 75
Multidose Mode (Continued)
RESTORE EXIT
VTBI _ _ _ _ mL
Multidose
Pressure Limit - P
Channel Labels
EXIT
DOSE
INTERVAL
# OF
DOSES
>Enter Rate
76 GETTING STARTED
Multidose Mode (Continued)
DOSE
INTERVAL
# OF
DOSES
>Enter Volume/Dose
GETTING STARTED
9. To enter time interval (1 to 24 hours) between doses, press A Multidose 13:00
DOSE INTERVAL soft key and use numeric data entry keys. RATE 100 mL/h
VOLUME/
DOSE 50 mL
DOSE
INTERVAL every _ _h
# OF
DOSES
10. To enter number of doses, press #OF DOSES soft key and use A Multidose 13:00
numeric data entry keys. RATE 100 mL/h
VOLUME/
50 mL
• If Delay Options is enabled, DELAY OPTIONS soft key
DOSE
appears. DOSE
INTERVAL every _ 6h
# OF
NOTE: Reference “Delay Options” section to program an DOSES
11. To begin multidose infusion, press START soft key. A Multidose 13:00
RATE 100 mL/h
VOLUME/
DOSE 50 mL
DOSE
INTERVAL every 06 h
# OF
DOSES _4 doses
>Press START
PAUSE START
GETTING STARTED 77
Multidose Mode (Continued)
• Main Display shows remaining VTBI for that dose. Midtown Hospital
Adult ICU 13:00
A VTBI = 50.0 mL
VOLUME AUDIO
INFUSED ADJUST
VOLUME AUDIO
INFUSED ADJUST
>Press START
PAUSE START
♦ Number of doses completed and when next dose Rate = 100 mL/h
Volume/ = 50 mL
starts displays. Dose
every 06 h x 04 doses
♦ Yellow Standby Status Indicator illuminates. Doses completed = 1
Dose 2 Starts 19:00
>Press START
PAUSE START
78 GETTING STARTED
Reviewing Serial Number
GETTING STARTED
GETTING STARTED 79
T H I S PA G E
I N T E N T I O N A L LY
LEFT BLANK
80 GETTING STARTED
ALARMS, ADVISORIES, ERRORS
To enhance safety and ease of operation, the Medley™ System provides a full range of audio and visual
alarms, errors, and messages.
Definitions
Advisory A sequence of audio and/or visual signals indicating the operating status
of the Medley™ Medication Safety System. The audio may be silenced
for approximately two minutes by pressing the SILENCE key.
Alarm An audio and visual signal that a potentially unsafe condition is present.
Immediate action is required. The audio may be silenced for
approximately two minutes by pressing the SILENCE key.
Error An audio and/or visual signal that a failure has been detected.
Immediate action is required.
Guardrails® Clinical Advisory A visual message when a designated drug is selected, to remind clinician
of specific hospital standards of practice when programming an IV
ALARMS, ERRORS,
medication. A specific clinical advisory can be associated with a selected
MESSAGES
drug within any of the patient care profiles.
Maintenance Reminder A visual message that, when enabled, appears at module startup when
scheduled preventive maintenance is due/overdue for any part of the
Medley™ System (Programming Module or attached module).
NOTE: This feature is only available with Pump Module Versions 1.11.22.5
and greater.
Prompt A visual message, appearing on the bottom line of the Main Display or in
the Channel Message Display. The message may be accompanied by an
audio signal that can be silenced for twelve seconds by pressing the
SILENCE key.
The Programming Module and Main Display provide various types of alert information. The characteristics
of the accompanying audio sounds are as follows:
Advisory One short beep every two seconds Variable volume; can be silenced for
two minutes.
Alarm Choice of three alarm audio profiles, Variable volume; can be silenced for
selectable in System Configuration two minutes.
Error (Hardware Detected) Pairs of long beeps Fixed maximum decibel volume;
cannot be silenced.
Error (Software Detected) Pairs of long beeps Fixed maximum decibel volume; can
be silenced for two minutes.
Illegal Key Press Two short beeps Variable volume; cannot be silenced.
Prompt One short beep every two seconds Variable volume; can be silenced.
Switchover Six short beeps: secondary switching Variable volume; can be silenced and
to primary. Two short beeps: bolus disabled in System Configuration.
switching to continuous.
Accumulated Air-in-Line A large number of air bubbles Clear air from line. To continue
smaller than current air-in-line limit infusion, press RESET soft key and
has recently passed detector. then RESTART key.
Air-in-Line Air has been detected in administration Ensure tubing is properly installed in
set during an infusion. Infusion Air-in-Line Detector. If air is present,
stops on affected channel. clear air from administration set.
Press RESTART key, or press CHANNEL
SELECT key and then START soft key.
Channel Disconnected Channel(s) disconnected while in To silence alarm and clear message
operation or has a communication from screen, press CONFIRM soft key.
problem. Reattach module, if desired,
ensuring it is securely “clicked” into
place at Channel Release Latch. If
alarm is still present, replace channel
with an operational instrument.
Check IV Set Administration set is not properly Close roller clamp, remove and
installed. Infusion stops on affected reinstall administration set, close
channel. door, open roller clamp, and then
press RESTART key.
ALARMS, ERRORS,
Close Door Door opened during an infusion. Close door. Press RESTART key, or
MESSAGES
Infusion stops on affected channel. press CHANNEL SELECT key and then
START soft key.
Flo-Stop Open - Close Door Flo-Stop® Device is in open position Close roller clamp on administration
while door is open. set or close door.
Occluded - Fluid Side/Empty Indicates either upstream occlusion Clear occlusion on fluid side of
Container or empty container. Infusion stops instrument. If necessary, refill drip
on affected channel. chamber. Press RESTART key, or
press CHANNEL SELECT key and then
START soft key.
Occluded - Patient Side Increased back pressure sensed while Clear occlusion. Press RESTART key,
infusing in pump delivery mode. or press CHANNEL SELECT key and
Infusion stops on affected channel. then START soft key.
Partial Occlusion - Patient Side Partial occlusion of patient side of IV Clear occlusion. Press RESTART key,
line detected by AutoRestart feature. or press CHANNEL SELECT key and
then START soft key.
Pump Chamber Occluded Blocked pump chamber detected. Open door and inspect pump
chamber. To open blockage, as
required, massage tubing. To
continue infusion, press RESET soft
key and then RESTART key.
Restart Channel Door opened and closed during an ¤ Close door. Press RESTART key, or
infusion. Infusion stops on affected press CHANNEL SELECT key and then
channel. START soft key.
Channel paused for 2 minutes. ¤ Press RESTART key, or press CHANNEL
SELECT key and then START soft key
Errors
Channel Error Error detected. Operation stops on To silence alarm and continue
affected channel. operation of unaffected channels,
press CONFIRM soft key . Replace
channel with an operational
instrument, as required. Service by
qualified personnel is required.
Messages
Anesthesia Mode Anesthesia Mode discontinued when Press CONFIRM soft key.
disconnected from AC.
Delay Complete Delay time completed. Press RESTART key, or press CHANNEL
SELECT key and then START soft key.
Panel Locked Tamper Resist feature is active and a If appropriate, deactivate Tamper
key was pressed. Resist feature using Tamper Resist
Control on back of Programming
Module.
Start time for next dose has passed. Start of next dose passed. Press CONFIRM soft key.
ALARMS, ERRORS,
MESSAGES
Specifications
Accumulated Air Window: Single Bolus Volume Window % Air that Causes
Setting (mL) Alarm
50 2.8 10%
75 8.0 20%
250 8.0 30%
*500 12.0 30%
Critical Volume: The maximum over-infusion which can occur in the event of a single fault condition is
0.6 mL.
MAINTENANCE
Atmospheric Pressure: 525 to 4560 mmHg 375 to 760 mmHg
(700 to 6080 hPa) (500 to 1013 hPa)
Equipment Orientation: To ensure proper operation, the system must remain in an upright position.
MAINTENANCE 87
Specifications (Continued)
Infusion Pressure,
Maximum: 654 mmHg (Maximum Occlusion Alarm Threshold plus tolerance)
KVO Selection Range: KVO rate can be set in System Configuration from 0.1-20 mL/h in 0.1 mL/h increments.
Rate Accuracy: Rate accuracy of the Medley™ Medication Safety System is ±5% at rates between 1 and
999 mL/h and ±5.5% at rates <1 mL/h, 95% of the time with 95% confidence, under the
conditions listed below.
Infusion Rate Range: 0.1 to 999 mL/h
Ambient Temperature: 68 ±4°F (20 ±2°C)
Source Container Height: 20 inches above the top of the Pump Module
Test Solution: Distilled Water
Distal Back pressure: 0 mmHg (0 kPa)
Needle: 18 gauge
Administration Set Model 2210
WARNING
Variations of head height, back pressure or any combination of these may affect
rate accuracy. Factors that can influence head height and back pressure are:
Administration set configuration, IV solution viscosity and IV solution temperature.
Back pressure may also be affected by type of catheter. Refer to the “Trumpet
and Start-Up Curves” section in “Appendix” chapter for data on how these
factors influence rate accuracy.
88 MAINTENANCE
Specifications (Continued)
Configurable Settings
System Settings
MAINTENANCE
Refer to the Medley™ Programming Module Directions for Use.
MAINTENANCE 89
Configurable Settings (Continued)
90 MAINTENANCE
Cleaning
Inspection Requirements
REGULAR INSPECTIONS
PROCEDURE FREQUENCY
INSPECT FOR DAMAGE:
Exterior Surfaces Each usage
Keypad Each usage
Seal Each usage
Mechanical Parts Each usage
CLEANING As required
START-UP Each usage
MAINTENANCE
• Labels and markings are legible.
• No tape or other foreign material is on sides of case;
anything of this nature could prevent proper latching of
channels.
• IUI Connectors have not been damaged.
2. Keypad
Check membrane switches for damage.
MAINTENANCE 91
Inspection Requirements (Continued)
Service Information
92 MAINTENANCE
WARRANTY
ALARIS Medical Systems, Inc., (hereinafter referred to as “ALARIS Medical Systems”) warrants that:
A. Each new ALARIS Medical Systems® Medley™ Pump Module is free from defects in material and
workmanship under normal use and service for a period of one (1) year from the date of delivery
by ALARIS Medical Systems to the original purchaser.
B. Each new accessory is free from defects in material and workmanship under normal use and
service for a period of ninety (90) days from the date of delivery by ALARIS Medical Systems to the
original purchaser.
If any product requires service during the applicable warranty period, the purchaser should
communicate directly with the relevant account representative to determine the appropriate repair
facility. Except as provided otherwise in this warranty, repair or replacement will be carried out at
ALARIS Medical Systems’ expense. The product requiring service should be returned promptly,
properly packaged and postage prepaid by purchaser. Loss or damage in return shipment to the
repair facility shall be at purchaser’s risk.
In no event shall ALARIS Medical Systems be liable for any incidental, indirect or consequential
damages in connection with the purchase or use of any ALARIS Medical Systems® Product. This
warranty shall apply solely to the original purchaser. This warranty shall not apply to any subsequent
owner or holder of the product. Furthermore, this warranty shall not apply to, and ALARIS Medical
Systems shall not be responsible for, any loss or damage arising in connection with the purchase or
use of any ALARIS Medical Systems® Product which has been:
(a) repaired by anyone other than an authorized ALARIS Medical Systems Service Representative;
(b) altered in any way so as to affect, in ALARIS Medical Systems’ judgment, the product’s stability or
reliability;
(c) subjected to misuse or negligence or accident, or which has had the product’s serial or lot number
altered, effaced or removed;
or
(d) improperly maintained or used in any manner other than in accordance with the written
instructions furnished by ALARIS Medical Systems.
MAINTENANCE
This warranty is in lieu of all other warranties, express or implied, and of all other obligations or
liabilities of ALARIS Medical Systems, and ALARIS Medical Systems does not give or grant, directly or
indirectly, the authority to any representative or other person to assume on behalf of ALARIS Medical
Systems any other liability in connection with the sale or use of ALARIS Medical Systems® Products.
ALARIS Medical Systems DISCLAIMS ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING ANY WARRANTY
OF MERCHANTABILITY OR OF FITNESS FOR A PARTICULAR PURPOSE OR APPLICATION.
See packing inserts for international warranty, if applicable.
MAINTENANCE 93
T H I S PA G E
I N T E N T I O N A L LY
LEFT BLANK
94 MAINTENANCE
APPENDIX
APPENDIX 95
Trumpet and Start-Up Curves (Continued)
Trumpet Curve
Trumpet Curve at
at 0.1
0.1mL/h
mL/h(initial)
(initial) Trumpet Curve at 1 mL/h (initial)
Trumpet
TrumpetCurve
Curve at
at 0.1 mL/h (72
0.1 mL/h (72hrs)
hrs) Trumpet Curve at 1 mL/h (72 hrs)
Flow Rate Error (%)
Flow Rate Error (%)
Legend:
NOTE: The plot range has been P Maximum rate error
increased to ±100%, to allow Overall rate error
[
96 APPENDIX
Trumpet and Start-Up Curves (Continued)
Legend:
P Maximum rate error
Overall rate error
[
APPENDIX 97
T H I S PA G E
I N T E N T I O N A L LY
LEFT BLANK
98 APPENDIX
ALARIS Medical Systems, Inc.
10221 Wateridge Circle
San Diego, California 92121 U.S.A.
Mail:
P.O. Box 85335
San Diego, California 92186-5335 U.S.A.
ALARIS®, ALARIS Medical Systems®, Flo-Stop®, Guardrails®, Medley™, and SmartSite® are trademarks and registered trademarks of ALARIS Medical
Systems, Inc.
All other trademarks belong to their respective owners.
US Patent Nos. 4,764,166; 4,690,673; 5,165,873; 5,601,445; 5,713,856; 5,800,387; 5,836,910; 5,941,846; AU Patent Nos. 596,552; 634,811; 693,662;
703,178; 703,203; 719,254; 728,366; AT 0,225,158; BE Brevet Belge No. 0,225,158; CA Patented/Breveté 1,258,212; 1,296,092; 2,026,518;
CH +0,225,158; DE D.B.P. Nos. 0,422,855; P3686558.3; FR Brevet Nos. 0,225,158; 0,422,855; GB Patent Nos. 0,225,158; 0,422,855;
IT Brevettato Vol. R.A. No. 0,225,158; JP Patent Nos. 1,816,872 ; 2,594,604 ; NL Ned. Octrooi No. 0,225,158;
SE Sv. pat. nr 0,225,158; TW Patent No. NI–107963. Other Patents Pending
147029-006 Copyright© 2003 ALARIS Medical Systems, Inc. All Rights Reserved