Professional Documents
Culture Documents
Operation Manual RS-100
Operation Manual RS-100
FOR
SHIMADZU REMOTE CONTROLLED
X-RAY DIAGNOSTIC TABLE
MODEL RS-100
Si-t1rAAOZU CORPORATION
KYOTO �APAN
CM506-828B
OPERATION MANUAL
FOR
SHIMADZU REMOTE CONTROLLED
X-RA Y DIAGNOSTIC TABLE
MODEL RS-100
Contents
Page
1. General ................................................................................................................ 1
1.1 Application ...................................................................................................... 1
1.2 Use conditions .................................................................................................. 1
1.3 Constitution .................................................................................................... 2
1.4 Specifications .................................................................................................... 3
2. Operation.............................................................................................................. 5
2.1 Table tilting .................................................................................................... 5
2.2 Tabletop longitudinal/lateral sliding.................................................................... 6
2.3 F/R unit sliding ............................... ................................................................. 7
2.4 FFD selection ................................................................................................ 8
2.5 Oblique beam projection ................................................................................ 8
2.6 Tabletop vertical sliding .................................................................................... 9
2.7 Compression cone ............................................................................................ 9
2.8 Collimator .......................................... . ........................................................ ..... 10
2.9 Loading of cassette .................................................................... . ..................... 12
2.10 Unloading of cassette ......................................................................................... 13
2.11 Selection and changing of exposuse program .................................................... 14
2.12 Fluoroscopy..................................................................................................... . 16
2.13 Radiography .................................................................................................... 17
2.14 Shifting of grid ........................... ;...................................................................... 24
2.15 Selection of rapid serial exposure...................................................................... 24
2.16 Finish and interruption of radiography ............................................................. 25
2.17 Procedure of emergency.................................................................................... 25
2.18 Setting of accessaries ...................................................-..................................... 27
3. Mai ntenance Service an d Check ............................................................................ 29
4. Er ror Code............................................................................................................ 30
Remote Control Board Drawing ...................................... Dwg. No. 501-09656
Local Control Board Drawing ............................................ Dwg. No. 501-09657
Equipment Drawing ............................................................ Dwg. No. 501-09655
1. General
1.1 Application
The RS-100 applies to the following examination techniques.
I.I. -TV fluoroscopy
Spot filming
Bucky radiography
Planigraphy
Indirect radiography
Enlargement radiography
Oblique projections
Film changer serial radiography
1.2 Use conditions
To make the most of functions of this equipment, use it under the following conditions.
(1) Cassette used CM or INCH size cassette based on IEC standard
(2) Ceiling height More than 2850 mm
(3) Power requirements 3</> AC200V, 3.0kVA, 50/60Hz
lq, ACl00V, 0.SkVA, 50/60Hz
(4) Environmental conditions
+10 ~ 40 C (based on IEC's General Requirements on
°
Ambient tern perature
Safety of Medical Electrical Equipment).
Relative humidity 30 ~ 75% (based on IEC's General Requirements on
Safety of Medical Electrical Equipment).
Atmospheric pressure 700 ~ 1,060 mbar (based on IEC's General Requirements
on Safety of Medical Electrical Equipment).
Atmosphere To be conformed to JIS Z 4703 on medical X-ray
mechanical equipment.
( 5) Grounding Grounding to the ground resistance of less than 100 n
with annealed copper wire of 1.6 mm diameter for
grounding wire.
-1-
s ti tu tion
1.3 C o n
0 1 -0 9 6 55
ns in g No. 5
D im e n s io S e e D ra w
ht 5 0 0 kg
T o ta l w e ig Approx. 1
'-----
13
a ss e m b ly @
CD Grips @ T a b le to p Base
G) @ tr o l b o a rd
0 Tube sup
p o rt F o o t re st Local c o n
m
device su p p o rt s @) y
0 C o m p re ss
io n c o n e S h o u ld e r T a b le b o d
a rd
u p h o ld e r @ o n tr o l b o
0 C o ll im a to
r
B a ri u m c R e m o te c
n so le
@ device @ o n tr o l c o
0 G ri p s S p o t ftlm R e m o te c
-2-
1.4 Specifications
° ° °
Table tilting 90 /-30 (Trendelenburg), speed 90 / 17 s
Table top height 93.5 cm to 115.5 cm, speed 1.1 cm/s(60Hz), 0.9cm/s(50Hz)
Table top dimensions 67.5cm X 210cm
Table top longitudinal sliding headwards 95cm/foot wards 55cm, speed 4.5cm/s
Table top lateral sliding ±12.5 cm, speed 3 emfs (60Hz), 2.5 emfs (50Hz)
Central-ray height 105 cm to 182 cm (above floor with unit in vertical
position)
F/R unit longitudinal sliding 77 cm, speed 6.0cm/s
° °
Oblique beam projection max. 30 caudo-cranial/35 cranio-caudal, with parallax
COJnpensation of the I.I., point of rotation from 15 mm to
235 mm height adjustable (above table top)
°
Tomography linear tomography (Planigraphy), FFD 110 cm, 40 /2s
° °
20 /ls, 8 /0.4 s
Layer height 15 mm to 235 mm (above table top), speed 1.1 cm/s (60Hz),
0.9 cm/s (50Hz)
FFD 110 cm, 120 cm, 140 cm, speed 2 emfs (60Hz),
1.7 emfs (S0Hz)
Table top to film distance 6 cm to 28 cm
Compression force less than 10 kgf.
-3-
Spotfilm device pushbu tton-co ntrolled automa tic program selectio n ,
rapid serial devided exposu res
CM INCH
IJ []]
14 X 17
35 X 43
IIJ ITI•
3 0 X 40
35 X 35
1
■ IJ I]
IJ I] • *
14 X 14
■ IJ I]
◊
'---/
■• 11] 1]
11 X 14
24 X 30
1 I] [] 10 X 12
1 I] II]
■ IJ [IJ ~ ~ ■ IJ [] ~ ~
~ ~
~ ■ IJ I]
9½ X 9½
~
■ IJ (IJ
24 X 24
18 X 24
■ • IJ II] 8 X 10
■ ■ IJ ITI
-4 -
2. Operation
To switch in the power again after cutting off the power, wait 5 seconds before switching-in.
2 1 Table tilting
Remote control board Local control board
/ r190º
1
I'
· ct!J
1 1
-30º~0°
When the switch is released while the table is tilting, the table stops at any position.
Depending on the tablet<?P longitudinal sliding position and the table tilting angle, the automatic
pulling-back function of tabletop will be worked to avoid the interference of the tabletop an d
the floor or ceiling. The relation of the table top longiturtinal sliding and table tilting angle is
as follows .
90°
45cm
10cm
Fig. 1
-5-
2.2 Tabletop longitudina l/lateral sliding
C)
t
.-@ [~: 1
¿ 1
¾,
Refer to Fig. 1 about the upper limit
C) @
¾J
~
~ - - - 7
.
___ J
:
t
r--- - -- ---7 ~
C) .. @
¾,
,~,
+-- - --t
+-C ) @
¾, +-*-+~
1L ..J
_____ _ __ _J1
, ' <:;?
[:70
C) '@ ,._
¿ el_)
¾, <'-;::i
t? <'¿:¡ t? (;)
1 1
e~::,_
When the switch is relased while the tabletop is sliding, the tabletop stops at any position.
The longitudina l limit of the tabletop sliding depends on the table tilting angle (refer to Fig. 1)
-6-
2.3 F /R unit sliding
......
--ª-.,
= Refer to Fig. 2 about the u pper limit.
--ª-.,
= Refer to Fig. 2 about the lowest limit.
When the switch is released while the F /R unit is sliding, the F /R unit stops at any position.
The limit of the sliding of F /R unit depends on the table tilting and the oblique of the
tube support.
____ _ __ _
(72._ ____, No limitation
S)cm
i------ --+---
(S_S_.S.:__
)c_m_;-- - ---------------<
At the table Trendelenburg position
Fig. 2
-7-
----
2.4 FFD selection
When push the switch of desired FFD, the tube support auto-
matically telescopes to the selected FFD.
The illumination of the pushed switch is turned on and off
untill the selected FFD is set.
The relation of the selectable FFD and oblique beam projection is as follows.
o o
o o
.--< N
.--< .--<
Fig. 3
* Note: In the case of planigraphy, angiography (mode selection switch " 1 :l, I ", " 1 is' I "
The limit of oblique angle depends on the table tilting, FFD and the kind of I. I.
35°
No limitation
~ - - At FFD 1400 mm
11
At the table Trendelenburg position and 12
~----i three-channel l. l. attached
Fig. 4
-8-
2.6 Table top vertical sliding
2
e = =----=-i:-=- -=--- ---::i] , h--
1
L
~
' - -- -- -- -- -~
...
~
o
( ~____,) § -- -- o
o
o
~
M
7 7 7 I 77 / I 7/77 1
/ ..... -
\
_.,,.
,,.,....,,,..-
___ _ _ _ _
-..... .....
_ ',
__
I'\
1
_ •.,/_L___ ___
M
M o
r....
~
or.... r....
o r....
co
N
8
o
'°
At 10cm above table top,
min. magnification factor is 1.13
max. magnification factor is 1.53
When the switch is released while the tabletop is sliding, the tabl etop stops at any position.
2. 7 Compression cone
Two other cone shapes are usable, flat type and projected type, which are interchangeable.
The compression cone comes into the X-ray field and compress
with a compression force tilla maximum of 10 kgf.
For safety , if the table top or F/R unit is slided, or the tube support is obliqued during
compression, the compression cone retreats. And if the tub e support is obliqued beyond
±4° , the compression cone does not work.
-9-
2.8 Collimator
A full automatic collimator is equipped in this equipment to control the X-ray field
automatically.
However, for purpose of reducing the patient's X-ray exposure, practice the positive X-ray
field regulation respectively in fluoroscopy and radiography by operating the collimetor handle.
The field is freely controlled by operating the collimeter handle to the following
directions in so far as the field size does not exceed the above sizes.
*,--íl--, +!: - -
#
clase
i
open
Field control in vertical
qp
:# t==f -#
close
-
~
open
Field control in horizontal
- 10 -
.....
.,.,. ¡¡:
.......
-rr
~
:tt -
close
-
ope n
Field control In vertical
qp
* F. -~
close
-
A=
open
Field control in horizontal
[@]
FULL-AUTO
X-ray field is automatic ally acljusted to the selec ted
film formar.
~
MANUAL
fl uo roscopic fielcl size.
NOTE
Keep the key switch of the collimator at the auto U position.
lf the key switch is at the manual ( ~ · ) position, the fluoroscopy
does not irradiate. The illumination of the fluo-spot filming selection
2 - 11
(2) In case bucky radiography
11 11
Set the collimator mode selection switch on the remate control panel FULL-AUTO •
Toe X-ray field is automatically adjusted to the selected film format. If you want to undersize ,
adjust the size of the light field by the knob of the collimator at key switch auto ( +t ) position.
1 NOTE 1
2 - 12
2.9.2 Shift the cassette to the stand-by position
After loading the cassette, push the cassette IN/OUT switch.
The carriage is shifted to the stand-by position, and the illumination
of the switch is turned off when the carriage reaches to the stand-by
position .
2.10.2 Pull up the jaw, and the cassette catches are depressed and the cassette comes out.
Fig. 7
- 13 -
2.11 Selection and changing of exposure program
2.11.1 Selection of exposure program
CM INCH
35 X 43
I] II] 14 X 17
I] II]
30 X 40
35 X 35
1 iJ I]
IJ (]]
- *
14 X 14
IJ (]]
◊ 11 X 14
• IJITI
24 X 30
1 IJ I]] 10 X 12
1 IJ [I]
- IJ IJJ ~ ~ ■ IJ I] ~ ~
■ IJ [I] ~ ~ ■ IJ I] ~ ~
24 X 24 9½ X 9½
18 X 24
■ • IJ IJJ 8 X 10
1 ■ IJ (I]
- 14 -
2.11.2 Changing of exposure program
The exposure program can be changed to the another selectable exposure program
freely whenever the first exposure has not been taken.
In case of the program change on the way of the subdivisional exposure, the following
program changes are possible.
CM 24X30
CM 24X24
□ ~
After first exposure of
EB
INCH 10X12
□
2-div., change 4-div. ex-
posure program DE
INCH 91/:zX 9½
~ EEE
After fust exposure of
3-div., change 6-div. ex-
posure program rnrn 5
- 15 -
2.12 Fluoroscopy
2.12.1 Preparation of fluoroscopy
\ 1
Change over the TECHNIQUE selector of the X-ray control at the position of" ~ ' ' ~-
When the fluoroscopy preparation switch is pushed, the preparation of
fluoroscopy is set and the illumination of the switch is turned on.
When the switch is pushed again, the preparation of fluoroscopy is
released and the illumination is turned off.
Note: l. When the mode selection switch "1 {IIll} I","1 c.c:!• l" or "ii'c' I" is selected,
this switch does not work.
2. When the cassette carriage is at cassette loading position, the switch
does not work.
3. When a fluoroscopy is not to be performed or is terminated, be sure to set
this switch in the fluoroscopic OFF state.
is steped.
2.12.3 Selection of 1.1. field size
In case of combination with the multi-field 1.1., the fluoroscopic field format can be changed
l~~~I
/ 1 /
Large Middle Small
- 16 -
2.13 Radiography
2.13.1 Selection of the switches
Select the TECHNIQUE selector of the X-ray control and the mode selection switch
of the remate control board in accordance with radiographic technique. By pushing the mode
selection switch, the preparatory motion to each radiographic technique starts and the illumi-
nation of the switch is turned on and off. When the preparation is finished, the illumination
of the switch is turned on.
~
' \ I /
Fluo-spot filming ~ Remate control board
Bucky radiography
Q
r=7 [jf] X-ray control
Planigraphy :¡_
~ X-ray control
Indirect radiography
(with spot-camera)
Dl
~ Remate control board
~
Serial radiography o Film changer
(with film changer) D
º'~
*Note: l. When radiography, do not use combinations shown in the above table.
Using unavailable combination may cause misexposure.
2. If the exposure switch of the remate control board is mis-pushed, the X-ray does not
irradiate and " ¡:::,, " is shown on the exposure sequence indicator of the remate
control board.
3. The symbol of the TECHNIQUE selector may be different depended on the
difference of the type of the X-ray control.
- 17 -
(1) Fluo-spot filming .
Push the mode selection switch "1 ; ,,,.
\ \ I /
F or fl uoroscopy and radiography, set the TECHNIQUE selector of the X-ray con trl " <@> ".
* Note: Set the key switch of the collimator to the auto ( g) position.
(refer to paragraph 2.8.2)
( 2) Bucky radiography
- 18 -
*Note: After turn on LINE switch, slide clown the table top vertically to the
lowest limit once. Otherwise, the real layer height can not be indicated.
o Adjustment and check of the layer height can be made also directly with a light pointer.
By pushing the layer height projector control switch, the light closs line is
00
~ projected onto the patient for about 30 seconds and it can be viewed on the
mirror. The center of the closs line shows the layer height.
o The layer height (center of rotation) above the film is 295mm fixed and FFD is auto-
matically set 1100mm. Therefore , the magnification factor in planigraphy is 1.37 ·
irrespective of the layer height above the table top.
For planigraphy, set the TECHNIQUE selector of the X-ray control"~'- By changing
,,,, fl
the TECHNIQUE se1ector o f the X-ray contra1to " <@)> ", uoroscopy is possible.
- 19 -
Slide the F /R unit headwards to the upper limit.
Position a cutfilm changer mounted on the mobile stand at the head end of the table.
o Allign the X-ray beam with the film changer by the light field of the collimator.
Position the radiographi c region of the patient in the X-ray field by the table top
longitudinal sliding with light field of the collimator.
o
Far serial radiography , set the TECHNIQU E selector of the X-ray control " D ''
To return the fluoroscopi c possible condition, push the mode selection switch "' ! i"
o The X-ray tube assembly swievels to the direction of the I.I ..
\ \ r I
By changing the T ECHNIQUE selector of the X-ray control " <@l> ", fluoroscopy is
possible.
0 Shift the cassette Carriage to fluoroscopi c position by the cassette IN/OUT switch.
To return the tube support to the mid-positio n , keep pushing "\➔Y~ \:'
o The F /R unit slides foot wards and then the tube support returns to the mid-positio n.
Fig. 8
board.
- 20 -
Push the mode selection switch ' 1
=a
f, 1".
-j__..___________
Coupling
/
Fig. 9
For serial radiography, set the TECHNIQUE selector of the control " □
_Q_ "
In the case of peripheral angiography, the time chart of the table top stepwise
movement and exposure is as follows.
_Jl.___ _ _ _ _ _ _
Exp osures
nL-_ _../(.______n_
))
2 sec.
Table step
command
Table drive
motor
Delay l. 7 sec
0.2 sec
If the table top stepwise movement does not programmed, abdominal angiography
is possible.
- 21 -
To return the fluoroscopic possible condition, push the mode selection switch "1! \"·
o The X-ray tube assembly swievels to the direction of the I.I. . ·
By changing the TECHNIQUE selector of the X-ray control " <!§J> ", fluoroscopy is
possible.
o Shift the cassette carriage to the fluoroscopic position by the cassette IN/OUT switch.
To return the tube support to the mid-possition, keep pushing "l ➔t~ I"·
/~--
\
e:_ __ _ "
Fig. 10
~
During the exposure , the X-ray irradiation indicator "X-RA Y "
1 X-RAY 1
lights up except in the case of film changer serial radiography and
general radiography.
- 22 -
*Note: l. In the case of fluo-spot filming or indirect radiography, push the exposure
switch under the fluoroscopic condition.
If the fluoroscopy preparation switch is not pushed, the fluo-spot filming
can not be done and " A " is shown on the exposure sequence indicator
of the remote control board when the exposure switch is pushed.
2. In case of planigraphy, set the exposure time 3 sec.
When a radiography is over, release your hand from the exposure switch. Then,
the exposure sequence indicator shows the number of remaining exposures by
one exposure.
To conduct fluoroscop y, turn off the fluoloscopic preparation switch "I<@> I" and
turn it on again.
2. In case of pre-transport technique
Press the exposure switch for a moment first.
The cassette carriage is transported to the exposure position but the X-ray does not
irradia te.
At the optional timing, push the exposu re switch again, then the X-ray irradiates.
- 23 -
2.14 Shifting of grid
The moving grid is provided to give images free of grid lines.
The grid can be retracted from the X-ray field , too.
When the illumination of the switch is turned on, the grid is retracted
from the X-ray field.
When the switch is pushed,. the illumination of the switch is turned off and
the grid is shifted to the X-ray field and stops there.
[j] When the switch is pushed, the rapid serial exposure is selected and the
illumination of the switch is turned on.
When the switch is pushed, again, the rapid serial exposure is reset.
Note: The rapid serial exposure can be set, only when both the TECHNIQUE selector
" '<!§1> " and the mode selection switch "1 ! I" is set.
- 24 -
2.16 Finish and interruption of radiography
2.16.1 Finish of radiography
When the exposure program is finished, the exposure sequence indicator shows the number
" L7 ", and the illumination of the exposure program selection switch is turned off.
The exposed cassette is shifted to the fluoroscopic position. To take up the cassette,
operate the following procedures as same as 2.10.
(1) Press the cassette IN/OUT switch.
Shift the cassette to the casette loading position.
(2) Take up the casette
Note: It is possible to shift the exposed cassette to the cassette loading position· auto-
matically instead of the fluoroscopic position after the exposure program is finished.
(Refer to the installation manual about setting)
2.16.2 Interruption of radiography
To take up the cassette on the way of the subdivisional exposure, operate the same
procedures.
(1) Press the cassette IN/OUT switch.
(2) Take up the cassette.
When the LINE switch is turned off, all functions stop. When the emergency switch is pushed,
the following movements stop.
o Table tilting
0 Oblique beam projection (including planigraphic movement)
0 Table top sliding
o F/R unit sliding
0 Tube support telescope
o X-ray tube assembly swievel
o Compression cone movement
After dissolution of emergency, turn on "LINE" switch or turn off emergency switch by
turning it clockwise.
- 25 -
By the way, if the following condition should happen,
° °
0 Table tilting over 92 , -32 (Trendelenburg)
o Table lon�tudinal over headwards 460mm, footwards 110mm in table
sliding tilting over 40°
° °
0 Oblique beam projection over 32 caudo-cranial, 37 cranio-caudal
the same movements when the emergency switch is pushed will stop to keep safety of the patient
and avoid the collision with floor and ceiling.
In this cae, return the table to the safety zone by the corresponding switch with pushing the
push button on the backside of the spotfilm device. (Fig. 11)
I
But, for safety, please get the patient to hold
Spotfilm d,m<
the grips in his hands.
Fig. 12
- 26 -
2.18 Setting of accessaries
2.18.1 Foot rest
(1) Mounting and dismounting of the foot rest must be done in the state of the table
horizontal.
(2) Gripping the levers at both right and left bottoms of the foot-rest, insert the guides
into groove rail on both sides of the tabletop.
(3) The adjustment of height is to be carried out bygripping the levers. After adjustment,
the foot rest can be completely locked by removing hand from levers.
Fig. 13
2.18.2 Grips
Fig. 14
- 27 -
2.18.3 Shoulder supports
For safety, never forget to set the shoulder supports in the table trendelenburg position
and fix firmly.
Fig. 15
Fig.16
- 28 -
3. Maintenance Service and Check
3.1 If a fuse is blown, replace the fuse, which is provided on the side of the supporting
base cover.
• For motor circuit (U. V. W) Cartridge fuse 1 SA
• For control circuit (L0, L100) Cartridge fuse SA
3.2 Make sure that all sliding surfaces are kept clean. If dust is allowed to collect on the sliding
surface, it may damage them and give troublesome operation.
3.3 Before and after use, make sure that all parts are in good condition and in working order..
- 29 -
4. Error Code
In case of the operation miss, the following display is shown on the exposure sequence indicator
of the remote control board
,,-
_ in the carriage, unusable cassette load the usable cassette again.
has loaded.
In the case that the table is not Tilt the table to the possible position
horizontal at serial radiography for each mode.
I
u mode, or the table is Trendelenburg
position at planigraphy mode.
In the case of peripheral angio- Select the other mode or select peripheral
graphy, the table top beyond angiography mode again.
I the safety zone or the encoder
is broken.
- 30 -
1 4
11
6 20 16 14
(J) t
L er he ht indicator @) E�osure seq_uence indicator
�a•·�fi-� ■ •17-}o/�-'./•�-
� lm�t i�-
® ;R;ti:to
Table tiltinj, slidi� and F /R unit sliding handle
� Cf� *�ili/J f'°F/' :/ r·· Jv
@) R d serial divided exposure selection switch
j!I! a J'RA1 •:;f-
® Table Trendelenburg switch
;R;ti�fflf.:J.ffl A1 •:;f-
® Gri� IN /OUT switch
1· 1) •:; I-' IN/OUT A1 •:;f-
@ Oblillue beam projection switch @ Emergency switch
Xt'i! 'iaiftj#,:j.A1 •:;f- �:w.W.Jl.fflfll!;f?':.;
@ FFD selection switch Foot switch for fluorosciy ( not illustrated)
Xt'i!'i1iflllH&lli/JA1 •:;f- :i!tJ! 7 - 1- A1 •:1 1- c 1 7A 1- )
@ F /R unit sliding swith
�{l��lli/JA1 •:;f-
E] DJ E]
+
m
RS-100 ifrtl1¥f'F��im@
501-09657