Professional Documents
Culture Documents
DXD 400
DXD 400
User Manual
Technical Publication
0232F EN 20191213
DX-D 400
User Manual
REVISION HISTORY
D OCT 15, 2018 New logo, new system label, Applied parts
warning, IEC fouth Edition, table Emergency
Stop images, new pedal overlay images, Tilting
Wall Stand Images.o.
This Document is the English original version, edited and supplied by the manufacturer.
The Revision state of this Document is indicated in the code number shown at the bottom of this page.
0232F EN 20191213
DX-D 400
User Manual
ADVISORY SYMBOLS
The following advisory symbols will be used throughout this manual. Their
application and meaning are described below.
0232F EN 20191213
DX-D 400
User Manual
TABLE OF CONTENTS
Section Page
1 INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
2.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2 Responsibilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
0232F EN 20191213 i
DX-D 400
User Manual
Section Page
4 OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
ii 0232F EN 20191213
DX-D 400
User Manual
Section Page
Section Page
iv 0232F EN 20191213
DX-D 400
User Manual
Section Page
0232F EN 20191213 v
DX-D 400
User Manual
vi 0232F EN 20191213
DX-D 400
User Manual
SECTION 1 INTRODUCTION
This manual contains all the necessary information to understand and operate
the DX-D 400 X-ray System. It provides a general description, safety
information, operating instructions and specifications concerning the
equipment. This manual is not intended to teach radiology or to make any type
of clinical diagnosis. It includes the different configuration possibilities;
Elevating Table or Fixed Height Table and Control Panel.
The Tube Support Column, the RAD Table and the RAD Wall Stand are
associated equipment to the X-ray Generator Unit.
The Control Panel of the Column is ergonomically built, equipped with controls
and indicators logically arranged and easily accessible. A soft thumb pressure
on the control buttons allows the linear and rotation movements.
Illustration 1-1
DX-D 400 X-ray System
0232F EN 20191213 1
DX-D 400
User Manual
The Generator Cabinet comprises the Power Module (which contains the
power and control components) and the High Voltage Transformer.
The operator controls and displays for radiographic operations are shown on
one of the following Consoles: the DX-D Overlay Console, the DX-D Overlay
Console with APR, the DX-D Touch Screen Console or the Software Console
of the NX Workstation. All functions, displays and controls are logically
arranged, easily accessible and identified to prevent confusion. Technique
factors and functions are selected by pushing the corresponding buttons or by
touching directly on the screen, as applicable.
The High Frequency X-ray Generator provides all the advantages of high
frequency waveform Generators including lower patient dose, shorter exposure
times and greater accuracy and consistency.
2 0232F EN 20191213
DX-D 400
User Manual
RAD Table:
0232F EN 20191213 3
DX-D 400
User Manual
X-Ray Generator:
G Three point control by selecting kVp, mA and ms, two point control
by selecting kVp and mAs (no AEC), or one point control by
selecting kVp with AEC operations.
G Equipped with closed loop control of X-ray Tube current, kVp and
filaments, which minimizes potential errors and the need for
readjustments.
4 0232F EN 20191213
DX-D 400
User Manual
Product.
Model.
Volts (V), Line Phases, Frequency (Hz), and Power (kVA, kW).
Date of manufacture.
Serial number.
Reference.
Manufacturer.
Place of manufacture.
Certification.
Illustration 1-2
Identification Labels
0232F EN 20191213 5
DX-D 400
User Manual
Images can be obtained with the patient in the sitting, standing or lying position.
Examinations can be performed to any kind of patient group. Patients may be
physically abled, disabled, immobilized or shocked.
This DX-D 400 contributes to the metrics of imaging performance ensuring the
efficient use of radiation.
As example of X-ray image receptors types that can be used: Cassette with
Film, CR (Computed Radiography) or Digital Detector.
The Normal Use of this equipment is defined as the Intended Use plus the
Maintenance and Service tasks.
1.3.3 CONTRAINDICATIONS
Do not use the equipment for any purposes other than those for which it is
intended. Operation of the equipment for unintended purposes could lead to
fatal or other serious injury.
6 0232F EN 20191213
DX-D 400
User Manual
Applied Parts refer to parts of the medical equipment that in Normal Use
necessarily comes into physical contact with the patient for the medical
equipment to perform its function. This RAD equipment includes the following
Applied Parts:
Other accessories.
0232F EN 20191213 7
DX-D 400
User Manual
8 0232F EN 20191213
DX-D 400
User Manual
Regulatory information and symbols used in the equipment are detailed in this
section to operate it safely.
2.1 GENERAL
0232F EN 20191213 9
DX-D 400
User Manual
10 0232F EN 20191213
DX-D 400
User Manual
Total of all layers composing the front panel of cassette holder 1.2 1.2
Total of all layers composing the front panel of FILM CHANGER 1.2 1.2
Note 1.-- Devices such as RADIATION DETECTORS are not included in the item listed in this table.
Note 2.-- Requirements concerning the ATTENUATION properties of RADIOGRAPHIC CASSETTES and of INTENSIFYING
SCREENS are given in ISO 4090 [3], for ANTI--SCATTER GRIDS in IEC 60627[1].
Note 3.-- ATTENUATION caused by table mattresses and similar accessories is not included in the maximum ATTENUATION
EQUIVALENT for PATIENT SUPPORT.
Note 4.-- Maximum ATTENUATION EQUIVALENT mm Al is only applied to the corresponding item. If several items given in this table
are located in the path of the X-RAY BEAM between the PATIENT and the X-RAY IMAGE RECEPTOR, each corresponding maximum
ATTENUATION EQUIVALENT mm Al is separately applied to each item.
0232F EN 20191213 11
DX-D 400
User Manual
2.2 RESPONSIBILITIES
12 0232F EN 20191213
DX-D 400
User Manual
0232F EN 20191213 13
DX-D 400
User Manual
Because exposure to X-ray radiation can be damaging to the health, use great
care to ensure protection against exposure to the primary beam. Some of the
effects of X-ray radiation are cumulative and may extend over a period of
months or years. The best safety rule for an X-ray operator is “Avoid exposure
to the primary beam at all times”.
Any object in the path of the primary beam produces secondary (scattered)
radiation. The intensity of secondary radiation depends on the energy and
intensity of the primary beam and the atomic number of the object material
struck by the primary beam. Secondary radiation may be of greater intensity
than that of the radiation reaching the receptor. Take protective measures to
safeguard against it.
14 0232F EN 20191213
DX-D 400
User Manual
- Use the smallest X-ray field collimation. Make sure that the
area of interest will be completely exposed and the X-ray field
does not exceed the area of interest.
0232F EN 20191213 15
DX-D 400
User Manual
16 0232F EN 20191213
DX-D 400
User Manual
Manufacturer.
Date of Manufacture.
Medical Device.
Serial Number.
Model Configuration.
0232F EN 20191213 17
DX-D 400
User Manual
Ionizing radiation.
Dangerous voltage.
18 0232F EN 20191213
DX-D 400
User Manual
Warning: Electricity.
No pushing.
No sitting.
No stepping on surface.
Do not handle.
0232F EN 20191213 19
DX-D 400
User Manual
Emergency stop.
“Stand-by” power.
(Only applies to IEC 60601-1:2005 and IEC 60601-1:2005+AMD1:2012)
“ON” power.
“OFF” power.
Alternating current.
20 0232F EN 20191213
DX-D 400
User Manual
Direct current.
Earth (Ground).
0232F EN 20191213 21
DX-D 400
User Manual
2.7.1 CERTIFICATIONS
22 0232F EN 20191213
DX-D 400
User Manual
0232F EN 20191213 23
DX-D 400
User Manual
X-ray equipment specified for examinations that do not need the operator or
staff to be close to the patient during normal use shall be provided with means
to allow the following control functions from a “Protected Area” (refer to
illustration below):
Generator
PATIENT ENVIRONMENT
EXAM ROOM
OPERATOR CONTROL
PROTECTED AREA
NO RADIATION AREA
24 0232F EN 20191213
DX-D 400
User Manual
X-Ray equipment specified for any radiological examination that requires the
operator or staff to be close to the patient during normal use (a.e. some pediatric
examinations or other types of examinations for patients that may require
assistance), shall have at least one “Significant Zone of Occupancy” for the use
of the operator and staff, designated as follows:
Illustration 2-1
Radiographic Examination on the Chest Unit or Front Panel
CHEST UNIT
X-RAY TUBE
Focal Spot
S
140 cm
Focal Spot
S
Protective Device
Wall
Dosimeter d Patient Support
X--Ray Receptor
CHEST UNIT
Focal Spot
X-RAY TUBE
0232F EN 20191213 25
DX-D 400
User Manual
Illustration 2-2
Radiographic Examination on any Patient Support or any Table
X-RAY TUBE
Focal Spot
SID 100 cm
Phantom
X--Ray Receptor
Patient Support
RAD TABLE
Focal Spot
S
SIGNIFICANT ZONE OF OCCUPANCY
AT THE LEFT SIDE OF THE RAD TABLE
(CATHODE)
Dosimeter d
S4
SIGNIFICANT ZONE OF OCCUPANCY
AT FRONT SIDE OF THE RAD TABLE
RAD TABLE
S3
X-RAY TUBE
Focal Spot
S5
26 0232F EN 20191213
DX-D 400
User Manual
0232F EN 20191213 27
DX-D 400
User Manual
Illustration 2-3
Distribution of Stray Radiation with the Receptor in Vertical Position.
200
180
160
140
HEIGHT ABOVE FLOOR (cm)
120
100
80
60
40
20
0
0 1 2 3 4 5 6 7 8 9 10
STRAY RADIATION (μGy/h)
S1
S2
Protective Device
Wall
Patient Support
X--Ray Receptor
CHEST UNIT
Focal Spot
X-RAY TUBE
28 0232F EN 20191213
DX-D 400
User Manual
Illustration 2-4
Distribution of Stray Radiation with the Receptor in Horizontal Position.
200
180
160
140
HEIGHT ABOVE FLOOR (cm)
120
100
80
60
40
20
0
S3 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
S5
S4
SIGNIFICANT ZONE OF OCCUPANCY
AT FRONT SIDE OF THE RAD TABLE
RAD TABLE
d
S3 d
X-RAY TUBE
d
Focal Spot
S5
0232F EN 20191213 29
DX-D 400
User Manual
This equipment generates, uses, and can radiate radio frequency energy.
increase the separation between the equipment and the affected device,
power the equipment from a source different from that of the affected
device,
30 0232F EN 20191213
DX-D 400
User Manual
The DX--D400 radiographic room is intended for use in the electromagnetic environment specified below.
The customer or the user of The DX--D400 radiographic room should assure that it is used in such an environment.
RF emissions
Class A
CISPR 11
The DX--D400 radiographic room is suitable for
use in all establishments other than domestic
Harmonic emissions
Class A and those directly connected to the public
IEC 61000--3--2
low voltage power supply network that supplies
low--voltage
buildings used for domestic purposes.
Voltage fluctuations/flicker emissions
Complies
IEC 61000--3--3
NOTE - In accordance with Standard IEC 61601-1-2:2014, the emissions characteristics of this equipment make it suitable for use in industrial areas
and hospitals (CISPR 11 Class A). If it is used in a residential environment (for which CISPR 11 Class B is normally required) this equipment might not
offer adequate protection to radio-frequency communication services. The user might need to take mitigation measures, such as relocating or
re-orientating the equipment.
0232F EN 20191213 31
DX-D 400
User Manual
The DX--D400 radiographic room is intended for use in the electromagnetic environment specified below.
The customer or the user of The DX--D400 radiographic room should assure that it is used in such an environment.
40% UT
60%
(60% dip in UT)
for 5 periods
for 5 cycles
70% UT
30%
(30% dip in UT)
for 25 periods
for 25 cycles
Mains power quality should be that of a typical
Voltage dips, short
commercial or hospital environment. If the user
interruptions and voltage >95%
< 5% UT of the Radiographic Room requires continued
variations on power supply for 250 periods
(>95% dip in UT) operation during power mains interruptions, it is
input lines.
for 250 cycles recommended that the Radiographic Room be
powered from an uninterruptible power supply
IEC 61000--4--11
or a battery.
>95%
< 5% UT
for 0.5 periods
(>95% dip in UT)
for 0.5 cycle
>95%
< 5% UT
for 1 period
(>95% dip in UT)
for 1 cycle
NOTE -- UT is the a.c. mains voltage prior to application of the test level.
32 0232F EN 20191213
DX-D 400
User Manual
The DX--D400 radiographic room is intended for use in the electromagnetic environment specified below.
The customer or the user of The DX--D400 radiographic room should assure that it is used in such an environment.
IEC 60601-1-2:2007
Immunity test Compliance level Electromagnetic environment -- guidance
Test Level
NOTE 2 -- These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which The DX--D400 radiographic room is used exceeds the applicable RF compliance level above, The DX--D400
radiographic room should be observed to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as re--orienting or relocating The DX--D400 radiographic room.
b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
0232F EN 20191213 33
DX-D 400
User Manual
The DX--D400 radiographic room is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of The DX--D400 radiographic room can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and The DX--D400
radiographic room as recommended below, according to the maximum output power of the communications equipment.
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance ’d’ in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where ’P’ is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 -- At 80MHz and 800MHz, the separation distance for the higher frequency range applies.
NOTE 2 -- These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
34 0232F EN 20191213
DX-D 400
User Manual
The DX--D400 radiographic room is intended for use in the electromagnetic environment specified below.
The customer or Operator of The DX--D400 radiographic room should assure that it is used in such an environment.
NOTE - UT is the a.c. mains voltage prior to application of the test level.
0232F EN 20191213 35
DX-D 400
User Manual
The DX--D400 radiographic room is intended for use in an electromagnetic environment specified below.
The customer or Operator of thisRadiographic Room should assure that it is used in such an environment.
Proximity fields from RF Refer to next table “IMMUNITY Refer to next table “IMMUNITY
wireless Communications REQUIREMENTS FOR RF REQUIREMENTS FOR RF Portable RF communications equipment
equipment WIRELESS COMMUNICATIONS WIRELESS COMMUNICATIONS (including peripherals such as antenna
IEC 61000-4-3 EQUIPMENT” EQUIPMENT” cables and external antennas) should be
used no closer than 30 cm to any part of the
equipment, including cables specified by
manufacturer. Otherwise, degradation of the
performance of this equipment could result.
Conducted disturbances 3 Vrms 3 Vrms
induced by RF fields from 150 kHz to 80 Mhz from 150 kHz to 80 Mhz
IEC 61000-4-6
6 Vrms in ISM bands 6 Vrms in ISM bands
from 150 kHz to 80 MHz from 150 kHz to 80 MHz
NOTE -- The ISM (industrial, scientific and medical) bands between 0.15 MHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz;
26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
The amateur radio bands between 0.15 MHz and 80 MHz are 1.8 MHz to 2.0 MHz; 3.5 MHz to 4.0 MHz; 5.3 MHz to 5.4 MHz; 7 MHz to 7.3 MHz; 10.1 MHz
to 10.15 MHz; 14 MHz to 14.2 MHz; 18.07 MHz to 18.17 MHz; 21.0 MHz to 21.4 MHz; 24.89 MHz to 24.99 MHz; 28.0 MHz to 29.7 MHz; and 50.0 MHz
to 54.0 MHz.
36 0232F EN 20191213
DX-D 400
User Manual
The DX--D400 radiographic room is intended for use in an electromagnetic environment specified below.
The customer or Operator of The DX--D400 radiographic room should assure that it is used in such an environment.
Pulse modulation b)
380 -- 390 27
18 Hz
FM c)
430 -- 470 ¦5 kHz deviation 28
1 kHz sine
Pulse modulation b)
704 -- 787 9
217Hz
Pulse modulation b)
800 -- 960 28
18Hz 0.3
Pulse modulation b)
1700 -- 1990 28
217Hz
Pulse modulation b)
2400 -- 2570 28
217Hz
Pulse modulation b)
5100 -- 5800 9
217Hz
0232F EN 20191213 37
DX-D 400
User Manual
Equipment:
Instrumentation used:
Dosimeter: Vacudap
Dosimeter: Unfors
Test Details:
38 0232F EN 20191213
DX-D 400
User Manual
Quantitative Information
Source--Image Distance
Output Dose
Input Dose
Focal Spot
Collimator
SID Dose
Selection
(μGy*m2)
Phantom
Phantom
Phantom
Phantom
opening
Time (s)
(mGy/h)
(mmAl)
(mGy)
(mGy)
(Gy/h)
(μGy)
mAs
HVL
(cm)
(cm)
KVp
SID
mA
0232F EN 20191213 39
DX-D 400
User Manual
Quantitative Information
Source--Image Distance
Output Dose
Input Dose
Focal Spot
Collimator
SID Dose
Selection
(μGy*m2)
Phantom
Phantom
Phantom
Phantom
opening
Time (s)
(mGy/h)
(mmAl)
(mGy)
(mGy)
(Gy/h)
(μGy)
mAs
HVL
(cm)
(cm)
KVp
SID
mA
40 0232F EN 20191213
DX-D 400
User Manual
Quantitative Information
Source--Image Distance
Output Dose
Input Dose
Focal Spot
Collimator
SID Dose
Selection
(μGy*m2)
Phantom
Phantom
Phantom
Phantom
opening
Time (s)
(mGy/h)
(mmAl)
(mGy)
(mGy)
(Gy/h)
(μGy)
mAs
HVL
(cm)
(cm)
KVp
SID
mA
0232F EN 20191213 41
DX-D 400
User Manual
Quantitative Information
Source--Image Distance
Output Dose
Input Dose
Focal Spot
Collimator
SID Dose
Selection
(μGy*m2)
Phantom
Phantom
Phantom
Phantom
opening
Time (s)
(mGy/h)
(mmAl)
(mGy)
(mGy)
(Gy/h)
(μGy)
mAs
HVL
(cm)
(cm)
KVp
SID
mA
42 0232F EN 20191213
DX-D 400
User Manual
Quantitative Information
Source--Image Distance
Output Dose
Input Dose
Focal Spot
Collimator
SID Dose
Selection
(μGy*m2)
Phantom
Phantom
Phantom
Phantom
opening
Time (s)
(mGy/h)
(mmAl)
(mGy)
(mGy)
(Gy/h)
(μGy)
mAs
HVL
(cm)
(cm)
KVp
SID
mA
0232F EN 20191213 43
DX-D 400
User Manual
Quantitative Information
Source--Image Distance
Output Dose
Input Dose
Focal Spot
Collimator
SID Dose
Selection
(μGy*m2)
Phantom
Phantom
Phantom
Phantom
opening
Time (s)
(mGy/h)
(mmAl)
(mGy)
(mGy)
(Gy/h)
(μGy)
mAs
HVL
(cm)
(cm)
KVp
SID
mA
44 0232F EN 20191213
DX-D 400
User Manual
Deterministic effects may occur when the Radiation dose to a certain organ or
tissue exceeds a specific threshold. Particular organs or tissues of such
concern in diagnostic Radiology are the skin and the eye lens. The numerical
value of the threshold dose is in the range between 1 Gy and 3 Gy.
Any health care professional (for example a customer or a user) who has any
complaints or has experienced any dissatisfaction with the quality, durability,
reliability, safety, effectiveness, or performance of this product must notify Agfa.
Agfa Service Support - local support addresses and phone numbers are listed
on www.agfa.com
0232F EN 20191213 45
DX-D 400
User Manual
46 0232F EN 20191213
DX-D 400
User Manual
The System should be powered by the same Room Electrical Cabinet where
the X-ray Generator is connected, that is, the whole System will be powered
from the same Electrical Cabinet.
2. Check that the Emergency OFF Switch of the Table is not depressed and
the protective Circuit Breaker located at the lower back of the Table, right
side of the lower cover is in ON position to allow Table operation.
0232F EN 20191213 47
DX-D 400
User Manual
3. Press the X-Ray Generator Control ON button, the Green Lamp turns
ON.
Illustration 3-1
Start Up sequence
1
Room
Electrical
Cabinet
NX Workstation
2 3
X-Ray
Generator
Control
48 0232F EN 20191213
DX-D 400
User Manual
2. Press the X-Ray Generator Control OFF button, the Green Lamp turns
OFF.
3. Turn OFF the Room Electrical Cabinet Switch (with the Emergency
Switch in OFF position).
Illustration 3-2
Shutdown sequence
1 3
NX Workstation Room
Electrical
Cabinet
X-Ray
Generator
Control
0232F EN 20191213 49
DX-D 400
User Manual
50 0232F EN 20191213
DX-D 400
User Manual
SECTION 4 OPERATION
Operation controls for the Floor Mounted Tube Stand are located at the Control
Panel and on the Semicircular Base of the Column.
The Column rotation must be manually positioned by the Operator. Step on the
Pedal located at the Base of the Column and then rotate the Column to the
desired angle. Release the Pedal to lock the Column in position.
This motion can reach ±180o with detents at 0o, 90o and --90o.
0232F EN 20191213 51
DX-D 400
User Manual
Illustration 4-1
DX-D 400 Analog Control Panel
52 0232F EN 20191213
DX-D 400
User Manual
VERTICAL MOVEMENT: Press and hold this button to allow vertical movement
of the Tube-Collimator Assembly in order to achieve the desired SID with
respect to the Horizontal Receptor. This control is also used to vertically position
the Tube-Collimator Assembly with respect to the Vertical Receptor. Release
this button to lock the Tube-Collimator Assembly in vertical position.
ROTATION: Press and hold this button to allow rotation of the Tube-Collimator
Assembly to the desired angle. Release this button to lock in position.
This movement has detents at 0o, 90o and --90o which are shown on the Angle
Display. Press this button again to continue with movement.
FREE MOVEMENTS: Press and hold this button to allow Horizontal, Vertical
and Rotation movements at the same time. Release button to lock in position.
LIGHT INDICATORS: An indicator will illuminate at the Control Panel when one
of the following positions is reached:
Not applicable.
0232F EN 20191213 53
DX-D 400
User Manual
DX-D 400 with Analog Control Panel and Fixed height Table includes a Vertical
SID reference label to provide the operator a quick SID reference.
Illustration 4-2
Vertical SID Reference Label
54 0232F EN 20191213
DX-D 400
User Manual
Illustration 4-3
DX-D 400 Digital Control Panel
After the System is turned ON, a start up routine begins and the Control Panel
shows for three seconds the software version (a.e. V01 R0.0 = Vers.01 R0.0).
Then, the Display shows the position data of the Tube-Collimator Assembly.
(Angle and SID).
SID DISPLAY: Indicates the Source-Image Distance (SID). The SID Display
shows “------” when the SID value is not defined, that is, when the
Tube-Collimator Assembly is not within the range of ±45o with reference to its
vertical axis or when it is not perpendicular to the RAD Wall Stand Receptor.
When the Tube-Collimator Assembly is pointing the RAD Wall Stand Receptor,
the Display shows the SID values configured during installation at the time each
distance is reached (a.e 110-150-180).
0232F EN 20191213 55
DX-D 400
User Manual
This movement is provided with detents related to the distance from the RAD
Wall Stand Receptor (that can be modified during the configuration of the
equipment). The detent point is shown at SID Display when the distance is
reached with respect to the RAD Wall Stand Receptor. The Display only shows
configured positions (a.e 110-150-180). Press this button again to continue
movement.
This movement is provided with a detent at 100 cm (40”) (shown at SID Display)
when the distance is adjusted with respect to the RAD Table Receptor. Press
this button again to continue movement.
This movement has detents at 0o, 90o and --90o shown at Angle Display. Press
this button again to continue with movement.
56 0232F EN 20191213
DX-D 400
User Manual
Table 4-1
Error Codes
0232F EN 20191213 57
DX-D 400
User Manual
Collimator controls consist of a button to switch on the Collimator lamp and two
knobs to open or close the internal blades of the Collimator.
When pressing the Collimator Lamp push-button, the Collimator light and an
optional Laser light turn on. They remain lighting for 30 seconds before they
switch Off automatically (lighting time can be configured).
Exposure field on the Receptor is adjusted by setting the two knobs. The table
on the Front Panel shows the number to set with the knobs to open the blades
according to the SID and X-ray field to be used.
Illustration 4-4
Collimator Controls
Additional variable filtration with manual settings may be optionally added to the
minimum collimator filtration in the form of an aluminium disk, 1mm thickness,
controlled manually from the front panel.
The disk features a hole for the passage of X-Rays and accommodates the
following three filters:
( ): no filtration.
58 0232F EN 20191213
DX-D 400
User Manual
Collimator controls:
1. Collimator Display
2. Manual Blade Control
3. Automatic Mode Indicator (Green)
4. System not ready Indicator (Red)
5. Manual Mode Indicator (Yellow)
6. Change of Filter
7. Collimator Lamp Control (Led “ON”)
8. Retractable Metric Tape
9. Laser Pointer Window
10. Laser Pointer On/Off Button
1 1
6 2
7
10 9 8
When pressing the Collimator Lamp push-button, the Collimator light and an
optional Laser light turn on. They remain lighting for 30 seconds before they
switch Off automatically (lighting time can be configured).
The Collimator can rotate ±90o on its vertical axis while the Tube remains in the
same position. This movement is performed by manually turning the Collimator
and has detents every 90o.
0232F EN 20191213 59
DX-D 400
User Manual
The optional Dosemeter device is related to the Collimator installed in the Tube
Stand. It is used for simultaneously determining the dose area product, dose
area product rate and irradiation time. The usual compatible Dosemeter
devices are VacuDAP 2004 Series.
Follow operation instructions for “RAD Table -- Fixed Height Table” (refer to
Section 4.3.1) or for “RAD Table -- Elevating Table” (refer to Section 4.3.2)
depending on the type of table installed at site.
Emergency-Off Switch
60 0232F EN 20191213
DX-D 400
User Manual
ON / OFF / EMERGENCY OFF SWITCH: The Table and the Floor Mounted
Tube Stand are ON whenever the Generator Console is ON and the Table
Emergency OFF Switch is not depressed. The Table and the Floor Mounted
Tube Stand are OFF when the Generator Console is OFF or when the
Emergency OFF Switch is pressed.
0232F EN 20191213 61
DX-D 400
User Manual
Press and hold Pedal in order to release longitudinal and transverse brakes of
Table-Top. Release Pedal to apply brakes.
The transverse motion is blocked when the Table-Top is centered with respect
to its horizontal axis. Step again to continue motion.
The Collimator Lamp turns on automatically when this Pedal is pressed for ease
in patient positioning.
The RAD Table with its variable height has been designed for easy access from
beds, wheelchairs or stretchers with patient care and comfort in mind.
The Table-Top travels are controlled with the Pedals located at the Table Base.
Also the Table incorporates two safety devices and a control for the RAD Table
positioning.
Emergency-Off Switch
Telescopic Covers
Up Pedal
Table-Top Lock Switch
Down Pedal
Table-Top Motion Pedal
62 0232F EN 20191213
DX-D 400
User Manual
ON / OFF / EMERGENCY OFF SWITCH: The Table and the Floor Mounted
Tube Stand are ON whenever the Generator Console is ON and the Table
Emergency OFF Switch is not depressed. The Table and the Floor Mounted
Tube Stand are OFF when the Generator Console is OFF or when the
Emergency OFF Switch is pressed.
0232F EN 20191213 63
DX-D 400
User Manual
Illustration 4-5
Elevating Table Pedals
TABLE-TOP MOVEMENT PEDALS: Step on any of the two pedals (at both
ends of the Pedal row) in order to release longitudinal and transverse brakes,
this allows free movement of Table-Top for better positioning of patient.
Release Pedal to block movement.
The Service Engineer can modify the configuration of the pedals and change
from the single step to release longitudinal and transverse brakes of the
Table-Top to a double step.
Note . A Height Detent in the Up and Down Travel of the Table-top can
be set by the Service Engineer to stop the Table-Top automatically
at the operator desired height.
64 0232F EN 20191213
DX-D 400
User Manual
The Hand Grips are used by patients and operators to keep their hands away
from the Tabletop edges and make the patient feel safe while the table-top is
being positioned.
Insert the Hand Grip in the rails and fix it at the desired position by turning
clockwise the Grip.
0232F EN 20191213 65
DX-D 400
User Manual
Illustration 4-6
Compression Band Operation
Compression Band as delivered Insert the Band Support and the opposite Support in the Table-top
Rails
Lock Lever
Extend the Band, hook one end and obtain the required tension with the Lock Lever
66 0232F EN 20191213
DX-D 400
User Manual
The Lateral Detector Holder is used for Table lateral work, including knee,
shoulder, skull, etc.
This Lateral Detector Holder is installed directly on the Tabletop rails. Insert the
Hand Grip in the rail and fix it at the desired position by turning clockwise the
Grip.
Illustration 4-7
Lateral Detector Holder placement
0232F EN 20191213 67
DX-D 400
User Manual
The Lateral Detector Holder is used for Table lateral work, including knee,
shoulder, skull, etc.
Illustration 4-8
Lateral Detector Holder placement
68 0232F EN 20191213
DX-D 400
User Manual
Illustration 4-9
Lateral Detector Holder 35x43 with Trolley
Illustration 4-10
Detector Installation
Note . Make sure that the Vertical Lock is blocked when mounting the
Detector to avoid unexpected falls down and damages of the
Detector and Support.
0232F EN 20191213 69
DX-D 400
User Manual
The Holder is adjustable for height, the vertical travel of the Detector is 750 mm
(29.5”). To move up/down the Detector Support:
1. Loosen CCW the Vertical Lock Lever To get free the Detector Support.
Hold the Detector Support during this procedure to avoid unexpected
falls down.
Illustration 4-11
Vertical Lock Lever
3. Tighten CW the Lever to fix again the Support at its new position.
The Holder is also mobile, it is provided with four wheels and each with its own
brake pedal. To lock the wheel step the brake pedal.
Illustration 4-12
Wheel Brake Pedal
70 0232F EN 20191213
DX-D 400
User Manual
To move the Holder in order to place it in its working position or to store it,
proceed as indicated below:
2. Push the Holder from its Vertical Bar and carry it where corresponds.
ASSEMBLY PROCEDURE
Before its first use, the Holder must be mounted in the field as it is shipped
splitted. Refer to the image below for graphical information about its assembly
procedure.
2. Mount the Detector Support and lock it with the Vertical Lock Lever.
Illustration 4-13
Detector Holder Assembly Procedure
0232F EN 20191213 71
DX-D 400
User Manual
Chin Rest.
Front Panel with AEC Detector Areas and very low absorption level.
Front Panel
Removable Grid
AEC Areas
72 0232F EN 20191213
DX-D 400
User Manual
Column
Vertical Carriage
Table-Top
0232F EN 20191213 73
DX-D 400
User Manual
Chin Rest.
74 0232F EN 20191213
DX-D 400
User Manual
The Receptor Assembly allows a tilting range from --20o to a horizontal position
of 90o (0o is the vertical position). A mechanical detent is provided to lock the
Receptor Assembly at 0o (vertical position).
Release the Mechanical Detent. Press and hold the Tilting Lock Button to
release the electrical lock and manually tilt the Receptor Assembly. Place the
hand at the lower part of the Receptor Assembly and push it slightly in opposite
direction of the Tilting movement and hold the Receptor Assembly until it
reaches the desired Angle. The inclination is indicated on both lateral
goniometers.
0232F EN 20191213 75
DX-D 400
User Manual
Push the Lever Lock located at the right side of the Receptor Assembly and
manually rotate the Receptor Assembly to the desired position. The rotation
angle is indicated in the goniometer located at the back side. Once rotated, lock
the Lever for Rotation by placing it in its upper position.
76 0232F EN 20191213
DX-D 400
User Manual
As well, make sure that the patient extremities are inside the
tabletop or accessories limits during operation: movement of
parts may cause serious damages to patient.
0232F EN 20191213 77
DX-D 400
User Manual
The Wall Stand can be prepared to hold an Arm Support. In that case, the side
of the Column Carriage includes an Arm Support holder with a knob.
Insert the Arm Support in the holder and screw the knob.
The Wall Stand may include Hand Supports. In that case, the Assembly will be
provided with the corresponding Hand Supports (x2) on two sides of it.
78 0232F EN 20191213
DX-D 400
User Manual
The Receptor Assembly may also house other optional parts such as the Grid
(Fixed or Removable) and the Ion Chamber, used for AEC exposures.
Fixed Receptor Assembly in a Rad Table Portable Receptor Assembly in a Wall Stand
Illustration 4-14
Travel distances of the Receptor Assembly
1500 mm
255 255
0232F EN 20191213 79
DX-D 400
User Manual
The Receptor can be moved beneath the Tabletop to the desired position. Press
and hold the brake button to unlock the Receptor and move it manually.
Illustration 4-15
Receptor Brake Buttons
The Receptor is assembled to a carriage that slides vertically along the Column
Stand. Hold down the Carriage Handle button in order to perform vertical
positioning with respect to the patient. Release button to stop motion. The
Vertical movement is locked when the equipment is turned OFF.
80 0232F EN 20191213
DX-D 400
User Manual
4.6.2 RECEPTORS
The Table and Wall Stand are compatible with a wide range of Receptors:
Type of Receptors
0232F EN 20191213 81
DX-D 400
User Manual
RECEPTOR TRAY (Cassette / CR): Pull out the Tray to load a Receptor
according to its size and orientation. Place manual clamps at corresponding
numbered notch, open the automatic clamp and insert the Cassette. This Tray
accepts all standard Cassette/CR sizes.
Fully open the Tray Set the manual clamps at the corresponding number
Open the Automatic Clamp, Insert the Cassette and Lock the Cassette and insert the Tray
adjust it to the Cassette
82 0232F EN 20191213
DX-D 400
User Manual
It includes the Detector, Detector Cabinet with fixed or removable Grid and Ion
Chamber Housing.
Rad Tables include the Brake Button to move horizontally the Assembly.
Brake Button
Fixed Detector in a Table
Wall Stands include the Brake Button in the Carriage to move Vertically the
Assembly.
Brake Button
0232F EN 20191213 83
DX-D 400
User Manual
Illustration 4-16
Portable Receptor Assembly in the Rad Table
Tabletop
Grid
Digital Detector
Tray
Handle Bar
Assembly Lock Button
Illustration 4-17
Portable Receptor Assembly with Rotating Tray in the Wall Stand
Tray
Chin Rest
Handle Bar
Tabletop
AEC Areas
Digital Detector
Grid
84 0232F EN 20191213
DX-D 400
User Manual
The Portable Receptor Assembly with Rotating Tray has been designed to
conveniently house a Receptor, an Ion Chamber and a Grid. It can also provide
the system with information about the position and status of the Grid and
Detector. The Digital Detector is always loaded in Landscape position and can
be manually rotated to Portrait position.
The Handle of the Assembly includes a Bar for extracting the Tray. In the RAD
Tables, the Brake Button also allows the horizontal movement of the Receptor
Assembly.
The Rotating Tray includes two main detents in its horizontal movement:
0232F EN 20191213 85
DX-D 400
User Manual
1. Grab the Handle and pull the Tray until it is completely out.
2. Then place the Detector, centered on the Tray and insert it in the tray.
Push slightly the end-stops with the Detector end until it is fitted in the
frame of the tray.
Illustration 4-18
Tray Loading
4. Push the Detector towards the end Stop and carefully lift and remove the
Detector with both hands.
Illustration 4-19
Tray unloading
1
1
86 0232F EN 20191213
DX-D 400
User Manual
1. Pull out the Receptor Tray to the end of its travel (Rotating Position).
4. Guide the Receptor Cable through the Cable Clips and insert the Tray.
The Handle of the Assembly includes a Brake Bar for extracting the Tray.
In Rad Tables, the Handle also contains a Brake Button to move horizontally the
Receptor Assembly.
0232F EN 20191213 87
DX-D 400
User Manual
Illustration 4-20
Portable Receptor Assembly with Non-Rotating Tray in Rad Table and Wall Stand
Brake Bar
1. Grab the Handle and pull the Tray until it is completely out.
2. Then place the Detector centered in the Tray and push slightly the
end-stops with the Detector until it is fitted in the four stops of the tray.
Illustration 4-21
Detector Loading
3. Guide outside the Detector Cable (if applicable) to avoid cable jams. Use
the Cable Clips to guide the cable out safely and fully insert the Tray.
4. To unload the Detector, grab the Handle and pull the Tray until it is
completely out.
88 0232F EN 20191213
DX-D 400
User Manual
5. Push the Detector towards the end Stop and carefully remove the
Detector .
In Systems with a Single panel for Wall Stand and Table or Systems with
external connection of the Detector, it is recommended to use a Extensible
Cable Hanger (not included in the System) located in a convenient place of the
ceiling, between the Table and the Wall Stand (or any other type of positioners),
to avoid damages to the Detector Cable.
Illustration 4-22
Cable to the Horizontal and Vertical Positioner
Extensible Cable Hanger Extensible Cable Hanger
Digital Detector Cable
Digital Detector Cable
0232F EN 20191213 89
DX-D 400
User Manual
4.6.6 GRIDS
The Wall Stand and the Table may hold a Removable Grid. The Grids are
labelled 100 cm (40”), 150 cm (59”) or 180 cm (70”). Use the corresponding Grid
according to the SID (Source to Image Distance).
When inserting the Grid in the Grid Slot, pay special attention to the type of
focalization distance of each Grid.
Check the correct insertion of the Grid in the slot with the label side facing the
Tube. A click sound means that the Grid is in place.
Handle the Grid with care. Dropping the Grid could cause
damage and reduced image quality.
Removable Grid
SID Label
Grid Handle
SID Label
Table
Wall Stand
90 0232F EN 20191213
DX-D 400
User Manual
Before Exposure, check the equipment daily and confirm that it works properly.
During exposure, do not use the Detector near devices generating a strong
magnetic field.
For Wireless Detectors, do not cover the IR Data Port with hands or other parts
of the body and do not use the selected frequency channel (2.4 GHz band) for
other wireless devices.
After every examination, wipe with a cloth slightly damped the patient contact
surfaces as well as the handle and Grid with disinfectants such as ethanol. For
cleaning, wipe with a cloth damped in neutral detergent.
0232F EN 20191213 91
DX-D 400
User Manual
2. Center the Collimator light, which corresponds to the X-Ray beam, with
respect to receptor. For that, use the Collimator Light centering marks
and the laser line on the receptor handle if applicable.
4. Turn ON the Collimator Lamp and adjust the field size with the Collimator
controls.
Illustration 4-23
Patient Positioning
Collimator Knobs
92 0232F EN 20191213
DX-D 400
User Manual
0232F EN 20191213 93
DX-D 400
User Manual
94 0232F EN 20191213
DX-D 400
User Manual
All controls, indicators and displays located on the DX-D Touch Screen Console
are functionally grouped. Also the Console shows different menus (screens)
according to the selected operations.
Note . Use the operating controls as described in this manual. Any other
non-indicated combination may result in the incorrect operation of
the equipment.
Illustration 5-1
General Controls
2 1
3 5
12
4
6 7
9
10
11 12 13
0232F EN 20191213 95
DX-D 400
User Manual
ON: The Generator and the Console are turned ON by pressing and holding for
two seconds the “ON / OFF Switch” of the Console Housing. This starts the
application and the last parameters and workstation used are automatically
selected.
OFF: The Generator and Console are turned OFF by touching this button on the
Touch Screen.
For Generators equipped with “High Speed Rotor Controller”, if the X-ray Tube
is rotating when touching the “OFF” push-button, the Generator will stop the
anode immediately and then the unit will be turned off (approx. 3 seconds). The
equipment only turns off if “Preparation” is not activated.
Note . The proper way for switching OFF the Generator and Console
must be by touching on the “OFF” button on the Touch Screen,
never use the “ON / OFF Switch” of the Console.
96 0232F EN 20191213
DX-D 400
User Manual
Write down the configuration of the Workstation assigned to each button in the
table.
Note.-- Workstation data such as Icon, X-ray Tube, Receptor, Ion Chambers, etc... must be registered.
0232F EN 20191213 97
DX-D 400
User Manual
A Focal Spot indicator shows the selected Focal Spot of the X-ray Tube: “Small”
or “Large”.
The Focal Spot is changed by touching this indicator. It keeps kVp and constant
mAs, whenever it is possible. The mA value available is set according to
maximum power, instantaneous power, space charge, etc.
When a Focal Spot is selected, it sets the highest mA value available for the
selected Focal Spot and the respective Exposure Time in order to keep constant
mAs. If the highest mA value available coincides with the maximum mA station
of the Generator, it sets one mA station below of the maximum mA station of the
Generator.
Note . The Focal Spot change can be done whenever the present
conditions of the X-ray Tube allow it. The mA station set for the
Focal Spot change is configured by the field engineer during the
installation.
98 0232F EN 20191213
DX-D 400
User Manual
RADIOGRAPHIC DISPLAYS: They are divided in the kVp, mA, mAs and
Time (s) Displays where the following data are shown.
kVp DISPLAY shows the radiographic kVp value selected for the technique.
When an exposure is made with AEC, it shows the actual mAs at the end
of the exposure, whenever the “Prep” button has not been released.
When an exposure is made with AEC, it shows the back-up Time during
the exposure and the actual Time at the end of the exposure whenever
the “Prep” button has not been released.
the values (blinking) of the actual Time, the calculated mAs, and the
selected kVp and mA radiographic parameters of the last exposure, after
touching the “Reset Error” button.
0232F EN 20191213 99
DX-D 400
User Manual
Also, the value can be directly selected by clicking on its position on the bar.
When this indicator is positioned over a value not allowed, it comes back to the
previous position and the parameter value does not change.
mA: Selects the X-ray Tube current. The Focal Spot selection can be
changed when the mA value is selected with the “Increase” or “Decrease”
buttons. The slider of the bar can only set the mA values of the selected
Focal Spot.
The following table indicates the different Information Messages and Display
flashing that may appear in the Console when increasing or decreasing one of
the exposure parameters.
DISPLAY FLASHING
INFORMATION MESSAGE DESCRIPTION
kVp mA mAs s
X Minimum mA
Min mA
(Generator limit configured for each Focal Spot)
X Maximum mA
Max mA
(Generator limit configured for each Focal Spot)
Max mAs X Maximum mAs = 500 mAs (Regulatory limit for AEC)
The AEC mode is activated by touching any of the three AEC Field buttons. The
AEC mode is deactivated by touching all the selected AEC Field buttons until
none of them is selected.
In AEC mode the back-up time (or back-up mAs) MUST BE SET MANUALLY
by the operator using the Console controls.
Note . The value of the back-up time (or mAs) must be set at a greater
value than the previously considered for the exposure time (or
mAs). A value above 50% of the considered value is the
recommended. Very extreme values of back-up time (or mAs)
should be avoided to prevent patient from excessive exposure
when a control error is produced.
FIELD SELECTION: Each button indicates its related physical location of the
selected field in the AEC Exposure Detector, and it may be selected or
deselected by touching it. Any combination of fields can be selected and the
color of buttons change (highlighted) when active.
DENSITY: These buttons are used to adjust the radiographic image density.
The selected value is shown on the Density Display. Normal image density is
N (default value).
AEC RESET: If the exposure is aborted by the AEC back-up timer, the “AEC
Reset” button blinks accompanied by an audible alarm and the message “Not
Enough Dose” is shown on the Console. Next exposure is inhibited until the
AEC function is reset by touching the “AEC Reset” button. When the Generator
is in “Prep” mode, the AEC function can not be reset.
Before the exposure, if the message “Wrong AEC Selection” is shown on the
Console, it means that the selected kVp value, AEC Density and/or Film/Screen
Combination set a technique that is out of the operative range with AEC and the
next exposure will be inhibited. Change any parameter (kVp value, AEC Density
or Film / Screen Combination) in order to obtain a technique enabled for AEC.
This safety device terminates the X-ray exposure when no radiation is detected
in the Ion Chamber or the selected parameters (short backup time/mAs) are not
appropriate for an exposure with AEC.
AEC Rapid Termination compares the AEC ramp with a 25% of the final value
at the 30% of the Backup Time. It is activated after 30% of the exposure backup
time and after 10 ms of exposure, both conditions have to be fulfilled.
The APR techniques are intended only as a guide line, they are only starting
points that can be replaced by more specific protocols developed by the
operator. Accurate exposure factors are dependent among other things on grid
factors, table top absorption, screen film combinations, film processing or
detector features.
APR is activated when one of the three Patient Size (small, medium or large
size) is selected (highlighted) and it is deactivated when all of them are
deselected.
APR DISPLAY: It shows the different Body Regions and Anatomical Views
available for each APR technique and the final APR selection. As the area of
the APR Display is limited in length, the name of some Regions, Views or
Projections is abbreviated.
When a Body Region has been selected, its indication is locked on the screen
and the APR Display show directly all its respective Anatomical Views and then
its respective Projections (if the Anatomical View has different Projections).
(Refer to Appendix B for the APR Regions / Views / Projections factory
pre-programmed).
When the APR selection is finished, the Console shows the final selection (APR
Display), the Workstation, its respective parameters (RAD Displays) and the
AEC controls (if applicable) related to the APR technique.
To go back to the previous level, touch on the right-lower APR selector (“Up”).
PATIENT SIZE: APR is activated when one of the three Patient Size (small,
medium or large size) is selected. These buttons are used to adapt the APR
technique chosen according to patient size. Six patient sizes are available.
The three right-hand buttons select Small, Medium and Large adults sizes (only
one selected at the same time). The left-hand “Pediatric” button changes the
function of the right-hand three buttons from adult patient size to Pediatric
patient size. (In this mode, the “Pediatric” button and one of the other three
buttons may be selected at the same time).
APR SELECTORS: Each button is related to the nearest area of the APR
Display and they are used to select one of the displayed Body Regions,
Anatomical Views or Projections. The selected Region, View or Projection
appears in the middle of the APR Display.
2. Verify that all factors of the technique are at the required values.
3. Touch the “Store” button to store the new technique. The selected
technique is now stored in the “APR Working File” and can be recalled
for future examinations.
Note . This procedure only changes the technique values of the selected
Patient Size, if necessary repeat the procedure for other Patient
Sizes.
Since “Store” modifies the data stored in the “APR Working File”, (according to
the configuration of the Unit), the Console requires the Operator Password to
allow the storage of the new data in the “APR Working File”.
The default Operator Password is “1111”, but it can be replaced by a new one
with the “Change Password” option in the “Edit” menu (refer to “Edit”).
Enter the Operator Password and touch “OK” , then confirm or cancel the
operation.
OK
MODIFY: After touching this button, the Console shows a Password pop--up
screen, enter the password required to access to the “Modify” options.
The default Operator Password is “1111”, but it can be changed by a new one
with the “Change Password” option in the “Edit” menu (refer to “Edit”).
Enter the Operator Password and touch “OK”. The APR module will show the
following options: Restore Default, Edit and Exit.
EXIT: Touch this button to exit from the “Modify” options and go back to the Main
APR screen.
RESTORE DEFAULT: Touch this button to delete all the modifications made
by the operator in all the APR techniques and restore all of them with the values
and selections stored in the “APR Backup File”.
Note . “Restore Default” replaces the data stored in the “APR Working
File” with the data stored in the “APR Backup File”.
EDIT: After touching this button, the Console shows the “Edit” menu with the
different controls to create, modify or delete the APR techniques, or change the
operator password.
Note . Exposures are inhibited when the “Edit” menu is shown on the
Console.
Body
Regions
Pediatric
Anatomical
Views Delete a Label
Projections
Modify a
Exit from Label name
“Edit” menu
This warning message is shown until the parameters / selections Save changes before exiting
for the new created APR Technique are set from “Edit” menu
QUIT: Touch this button to exit from the “Edit” options and go back to the
previous screen (Restore Default / Edit / Exit).
CHANGE PASSWORD: After touching this button, the Console shows a Panel
for entering the new Operator Password.
Follow the “Change Password” procedure and enter twice the new Password
(four digits) then touch “OK” to close this option. Now, the default Password
“1111” is replaced by the new Password.
PATIENT SIZE: These buttons are used to select the Patient Size (small,
medium or large size) for the APR technique chosen. Six patient sizes are
available.
The three upper buttons select “Small”, “Medium” and “Large” adult sizes (only
one selected at the same time). The lower button selects “Pediatric” , this button
is selected together with one of the “Small”, “Medium” and “Large” sizes.
When a Body Region has been selected, its label is highlighted on the screen
and the View level shows all its respective Anatomical Views.
When an Anatomical View has been selected, its label is highlighted on the
screen and the Projection level shows all its respective Projections (if the
Anatomical View has different Projections).
When the APR selection is finished, the Console shows the Workstation, its
respective parameters (RAD Displays) and the AEC controls (if applicable)
related to the APR technique.
Note . The maximum number of labels for each level is ten (10).
LABEL POSITIONERS: Select a label and change its position inside the level
by touching any of these buttons.
NEW: To create a new Region, View or Projection, first click on the lateral
indicator of the Region, View or Projection or on any Region, View or Projection
label (the lateral indicator of the selected level appears highlighted), then touch
on the “New” button.
A virtual keyboard appears on the screen. Write the name of the new Region,
View or Projection and click on “Enter”. The new label of the Region, View or
Projection is located in the last position of the list.
Now, if what it was created was a Region, the operator will have to create
(assign) Views and Projections to that new Region, for that, the program
requires to create at least one View for that Region and after that, create the
Projections for that View (if required). Repeat the process to create and name
the new Views and Projections.
The warning message “APR technique not defined ” is displayed on the “Edit”
menu. Set the Patient Size, Radiographic Parameters and Selections
(Workstation and AEC if available) related to that Region / View / Projection and
touch the button “Set Technique” .
SET TECHNIQUE: Touch this button to assign the selected Patient Size,
Radiographic Parameters and Selections (Workstation and AEC if available) to
the new or existing Region / View / Projection. A pop message requires a
confirmation.
MODIFY: Select a label (Region, View or Projection), then touch this button to
modify the label name using the virtual keyboard that appears on the screen.
Write the new name and click on “Enter”.
DELETE: To delete a Region, View or Projection select the respective label and
touch this button. When a Region is deleted, all the labels and APR techniques
(Views and Projections) related to this Region are deleted. When a View is
deleted, all the labels and APR techniques (Projections) related to this View are
deleted.
SAVE: Touch this button to save all the changes performed in the “Edit” menu
before exiting.
Illustration 5-2
Creating a New APR Technique
2
2. Touch on “Edit” button.
New Region
3
3. Select any Region , press on “New” and write
the name of the new Region. Repeat this procedure
in order to assign Views and Projections to the new
created Region. New Region
New View
New Projection
Illustration 5-3
Modifying an existing APR Technique
OK
OFF Radiographic exposures from the Console are made with the “Prep”
Prep (preparation) and “Expose” buttons on the Handswitch. The status of the
Exp exposure is indicated by the “Ready” and “X-ray On” indicators for the duration
of the exposure.
PREP: Press the Handswitch half-way (“Prep”) to prepare the selected X-ray
Tube for exposure. The “Ready” indicator on the Console appears in yellow
when the Tube is prepared and there are no interlock failures or system faults.
Anode rotation.
Note . Press “Prep” only when the technique is selected on the Console
and the patient is ready for the exposure. High Speed Generators
can be configured so that the anode remains running for
approximately one minute when “Prep” is pressed three
consecutive times in intervals shorter than 30 seconds.
EXP: When the “Ready” indicator is ON, completely pressing the Handswitch
commences the exposition. Releasing the button before the Generator
completes the selected time or AEC time will cause the exposure to prematurely
terminate. The Console will then show the actual mAs and Exposure Time
values. The message “Exp aborted by the operator” appears until the “Reset
Error” button is pressed to reset the error condition.
READY: When highlighted, this button indicates that the technique selected is
properly set, there are no interlock failures or system faults, the anode is rotating
and the X-ray Tube is ready for exposure.
X-RAY ON: When highlighted, this button indicates that the X-ray exposure is
in progress. When radiographic exposures are performed, an audible signal
sounds.
DOOR DOOR OPEN: Indicates the X-ray room door is open when the X-ray equipment
is in use.
TUBE OVERLOAD: Indicates that either the technique selected is beyond the
TUBE OVERLOAD
X-ray Tube ratings or present conditions of the X-ray Tube inhibit the exposure
(anode overheated). Parameters for next exposure may be temporally limited
by the Generator (change the exposure values or wait for the X-ray Tube to
cool).
Check that heat units available are lower than the calculated for the next
exposure (heat units close to zero). Reduce exposure factors or wait for the
X-ray Tube to cool. (To display the Heat Units refer to Section 5.6).
ROTOR ERROR: Indicates that the X-ray Tube anode is not rotating while
ROTOR ERROR
“Prep” is active, then exposures are inhibited.
HEAT: Indicates that the X-ray Tube thermostat / pressurestat is open due to
HEAT
overheating of the Tube housing (housing is too hot, wait for the housing to cool)
or a thermostat / pressurestat mal-function (housing is cool). Heat units may
raise to any value.
if a failure on the Automatic Collimator has been detected (blades are full
open or in movement during exposure, etc.). In this case the indicator
lights continuously.
The “Heat Units” Display shows the percentage of the thermal capacity of the
Tube that is used. For example, a display of “0%” would indicate that all the Heat
Units capacity of the X-ray Tube remains.
The Dose Display shows the accumulated radiation value of the X-ray Tube
related to the selected Workstation. The radiation value of each exposure is
added to the accumulated radiation value for the selected X-ray Tube. The
accumulated radiation value is reset by touching the “Dose Reset” button.
Radiation measure is read as DAP value (Dose Area Product) in mGy*cm2 (for
example: 64.50.5 mGy*cm 2).
Error codes indicate the potential cause of a system failure. Error codes are
shown on the Console, at the same time an alarm sounds. Correct the error
cause and keep touched the “Reset” button until the Console indication
disappears. (Refer to Table 5-1).
All these error codes are preceded by the letter “E” (i.e., E01) and they will
enable the operator to indirectly convey the possible source of error to service
personnel. This may prevent the need for a service call or enable service
personnel to anticipate corrective actions prior to arriving on site.
Table 5-1
Error Codes
Turn the Generator OFF, check the proper external cable connections
E01, E02 Communication error. and then turn the Generator ON. If the equipment remains inoperative,
turn it OFF and call Field Service.
E12 No mA during exposure or mA value is out of range. Touch the “Reset” button.
Repeat with same technique values, If the error code persists try with
another combinations of kV and mA values.
E13 No kV during exposure or kV value is out of range. If the equipment remains inoperative, turn it OFF and call Field Service.
It is not necessary to touch the “Reset” button, the error indication will
Rotor error. The X-ray tube anode is not rotating while
disappear automatically.
E18 “Prep” is active, then exposures are inhibited, or the X-ray
If the error code persists, turn the Generator OFF and ON.
tube anode is rotating without console command.
If the equipment remains inoperative, turn it OFF and call Field Service.
Door Open error. The X-ray room door is open when the
E35
X-ray equipment is in use.
It is not necessary to touch the “Reset” button
button, the error indication will
Heat Units error. The X-ray Tube thermostat / pressurestat disappear automatically.
is open due to the Tube housing being overheated (housing If the error code persists, turn the Generator OFF and ON.
E36 is too hot, wait for the housing to cool) or a thermostat / If the equipment remains inoperative, turn it OFF and call Field Service.
pressurestat mal-function (housing is cool). Heat units may
raise to any value.
Touch the “Reset” button and call Field Service. This error does not
E98 Service Mode Active.
inhibit normal operation.
All controls, indicators and displays located on the Control Console are
positioned depending upon their functions.
Note . Use the operating controls as described in this manual, any other
non-indicated combination may cause an incorrect operation of
the equipment.
Illustration 6-1
Rad Console
2 1
7 8
9
11
4
10 10
5 6
12
For Generators equipped with “High Speed Rotor Controller”, if the X-ray Tube
is rotating when pressing the “OFF” push-button, the Generator will stop the
anode immediately and then the unit will be turned off (approx. 3 seconds). The
equipment only turns off if “Preparation” is not activated.
Note in the following table the configuration of the Workstation assigned for
each push-button.
+ 1
1
2
1 +
1 +
+ 1
2
2
2 +
2 +
Note.-- Workstation data such as X-ray Tube, Device, Ion Chambers, etc... must be registered.
LARGE FOCAL SPOT: Indicates that the “Large Focal Spot” of the X-ray Tube
has been selected.
SMALL FOCAL SPOT: Indicates that the “Small Focal Spot” of the X-ray Tube
has been selected.
FOCAL SPOT SELECTORS: Each push-button selects the related Focal Spot
of the X-ray Tube, keeping kVp and mAs constant (maximum mA available and
minimum Exposure Time).
Note . The Focal Spot change can be done whenever the present
conditions of the X-ray Tube allow it.
Note . The mA station set for the Focal Spot change is configured by the
field engineer during the installation.
The actual X-ray Tube heat unit value after pressing the “On” push-button
(Refer to Section 6.6).
The error messages during a system fault, preceded by the letter “E”
(a.e., E02) (Refer to Section 6.9).
When an exposure is made with AEC, it shows the actual mAs at the end
of the exposure whenever the “Prep” push-button has not been released.
The Time value (in seconds) selected for the radiographic technique.
When an exposure is made with AEC, it shows the back-up time during
the exposure and the actual Time at the end of the exposure whenever
the “Prep” push-button has not been released.
The actual Time, the calculated mAs, and the selected kVp and mA
radiographic parameters of the last exposure, with or without AEC, after
pressing the “AEC Reset” push-button (values flashing).
The AEC mode is selected by pressing any of the three AEC Field push-buttons.
To exit the AEC mode, press all the illuminated AEC Field push-buttons until
none are lit.
In AEC mode the back-up time (or back-up mAs) MUST BE SET MANUALLY
by the operator using the Console controls.
Note . The value of the back-up time (or mAs) must be set at a greater
value than the previously considered for the exposure time (or
mAs). A value above 50% of the considered value is the
recommended. Very extreme values of back-up time (or mAs)
should be avoided to prevent patient from excessive exposure
when a control error is produced.
DENSITY: These push-buttons are used to adjust the radiographic film density.
Normal film density (0) is the automatic default value when AEC is selected.
AEC RESET: If the exposure is aborted by the AEC back-up timer, the indicator
on the “AEC Reset” push-button blinks accompanied of an audible alarm. Next
exposure is inhibited until the AEC function is reset by pressing the “AEC Reset”
push-button. When the Generator is in “Prep” mode, the AEC function can not
be reset.
The “AEC Reset” push-button may blink when the kVp value, AEC Density and
Film / Screen Combination select a technique that is out of the operative range
with AEC, it inhibits the next exposure. Change any parameter (kVp value, AEC
Density or Film / Screen Combination) in order to obtain a technique enabled
for AEC.
This safety device terminates the X-ray exposure when no radiation is detected
in the Ion Chamber or the selected parameters (short backup time/mAs) are not
appropriate for an exposure with AEC.
AEC Rapid Termination compares the AEC ramp with a 25% of the final value
at the 30% of the Backup Time. It is activated after 30% of the exposure backup
time and after 10 ms of exposure, both conditions have to be fulfilled.
The APR techniques are intended only as a guide line, they are only starting
points that can be replaced by more specific protocols developed by the
operator. Accurate exposure factors are dependent among other things on grid
factors, table top absorption, screen film combinations, film processing or
detector features.
The APR is activated when one of the three Patient Size (small, medium or large
size) is selected (push-button lighted), and it is deactivated when all of them are
deselected.
APR DISPLAY: Shows the different Body Regions and Anatomical Views
available for each APR technique, and the final APR selection.
Because each area of the APR Display is limited in length to eight characters,
some regions and views are abbreviated.
PATIENT SIZE: These push-buttons are used to adapt the APR technique
chosen according to patient size. Six patient sizes are available. The three
right-hand push-buttons select Small, Medium and Large adult sizes (one only
illuminated when selected). The left-hand pediatric push-button changes the
function of the right-hand three push-buttons from adult patient size to pediatric
patient size. (In this mode, the pediatric push-button and one of the other three
buttons may be illuminated at the same time).
The APR display shows the Body Regions, Organs and Projections. When a
Body Region has been selected its indication is locked on the screen and the
APR Display shows directly all its respective Organs. In some cases, an Organ
may show several Projections.
When the APR selection is finished, the Console shows the final selection (APR
Display), its respective parameters (RAD Display), the Workstation and AEC
controls related to the technique. (Refer to Appendix B for the APR Regions /
Organs / Projections factory pre-programmed).
2. Verify that all factors of the technique are at the required values.
3. Store the new technique by first pressing the selected Patient Size and
without releasing it press the Projection selected. (Refer to
Illustration 6-2.)
Note . This procedure only changes the technique values of the selected
patient size, repeat the procedure for other patient sizes.
Illustration 6-2
APR Storage Push-buttons
AP/PA SPECIAL
SKULL
LATERAL
SKULL
SEMIAX.
LATERAL
AXIAL
Example : press push-buttons: 1 st Medium Patient Size and Lateral Projection, to store the APR technique selected.
Radiographic exposures from the Control Console are made with the “Prep”
(preparation) and “Expose” (X-ray exposure) push-buttons or with the
Handswitch. The status of the exposure is indicated by the “Ready” and “X-ray
On” indicators for the duration of the exposure.
PREP: Press the ”Prep” push-button to prepare the selected X-ray Tube for
exposure. The “Ready” indicator on the Console will light when the X-ray Tube
is prepared and there are no interlock failure or system faults.
Anode rotation.
Note . Press “Prep” only when the technique is selected on the Console
and the patient is ready for the exposure. High Speed Generators
can be configured so that the anode remains running for
approximately one minute when “Prep” is pressed three
consecutive times in intervals shorter than 30 seconds.
EXP: After the “Ready” indicator is illuminated, press this push-button to start
a X-ray exposure. If the push-button is released before the Generator
completes the selected time or the AEC time, the exposure will be prematurely
terminated and the actual mAs and Exposure Time will be displayed.
The “X-ray On” indicator remains illuminated during the length of exposure.
READY: Indicates that the technique selected is properly set, there are no
interlock failures or system faults, the anode is rotating and the X-ray Tube is
ready for exposure.
X-RAY ON: Indicates that the X-ray exposure is in progress. At the same time
that radiographic exposures are being made, an audible signal sounds.
OFF Radiographic exposures can also be initiated with the X-ray Handswitch which
Prep is connected to the Control Console.
Exp
The X-ray Handswitch button has three positions: “Off”, “Preparation”, and
“X-ray Exposure”, which operate in the same way that “Prep” and “Exp”
push-buttons on the Control Console.
Press the Handswitch half-way for “Prep” and fully for “Exp”.
To view the remaining Heat Units, press the “On” push-button. The kVp Display
shows the percentage of Heat Units that remain preceded by the letter “H”. For
example, a display of “H75” would indicate that 75% of Heat Units capacity of
the X-ray Tube remains. “H -- -- ” indicates that all the capacity remains. The kVp
Display reverts to its normal function after releasing the “On” push-button.
The operator can read the number of exposures made by the Generator, as
indicated below:
= 123.456 exposures
kV mAs
DOOR OPEN: Indicates the X-ray room door is open when the X-ray equipment
DOOR
is in use.
It can be also shown making a high power and long exposure with the X-ray
Tube cool (X-ray Tube has not been warmed-up).
TUBE OVERLOAD: Indicates that either the technique selected is beyond the
T.OVL
X-ray Tube ratings or the present conditions of the X-ray Tube inhibit the
exposure (anode overheated). Parameters for next exposure may be
temporally limited by the Generator (change the exposure values or wait for the
X-ray Tube to cool).
Check that heat units available are lower than the calculated for the next
exposure (heat units close to zero). Reduce exposure factors or wait for the
X-ray Tube to cool. (To display the Heat Units refer to Section 6.6).
ROTOR ERROR: Indicates that the X-ray Tube anode is not rotating while
ROTOR
“Prep” is active, then exposures are inhibited.
HEAT: Indicates that the X-ray Tube thermostat / pressurestat is open due to
HEAT
overheating of the Tube housing (housing is too hot, wait for the housing to cool)
or to a thermostat / pressurestat mal-function (housing is cool). Heat units may
raise to any value.
if a failure on the Automatic Collimator has been detected (blades are full
open or in movement during exposure, etc.). In this case the indicator
lights continuously.
Error codes indicate the potential cause of a system failure. Error codes are
shown on the kVp Display at the same time an audio signal is emitted. Correct
the error cause and keep pressed the “AEC Reset” push-button till the Console
indication disappears. (Refer to Table 6-1).
All these error codes are preceded by the letter “E” (i.e., E01) and they will
enable the operator to indirectly convey the possible source of error to service
personnel. This may prevent the need for a service call or enable service
personnel to anticipate corrective actions prior to arriving on site.
Table 6-1
Error Codes
Turn the Generator OFF, check the proper external cable connections
E01, E02 Communication error. and then turn the Generator ON. If the equipment remains inoperative,
turn it OFF and call Field Service.
E12 No mA during exposure or mA value is out of range. Press the “AEC Reset” push-button.
Repeat with same technique values, If the error code persists try with
another combinations of kV and mA values.
E13 No kV during exposure or kV value is out of range. If the equipment remains inoperative, turn it OFF and call Field Service.
This error does not require to press the “AEC Reset” push-button, its
Rotor error. The X-ray tube anode is not rotating while
indication disappears automatically.
E18 “Prep” is active, then exposures are inhibited, or the X-ray
If the error code persists, turn the Generator OFF and ON.
tube anode is rotating without console command.
If the equipment remains inoperative, turn it OFF and call Field Service.
Door Open error. The X-ray room door is open when the
E35
X-ray equipment is in use.
These errors do not require to press the “AEC Reset” push-button,
push button theirs
Heat Units error. The X-ray Tube thermostat / pressurestat indications disappear automatically.
is open due to the tube housing is overheated (housing is If the error code persists, turn the Generator OFF and ON.
E36 too hot, wait for the housing to cool) or a thermostat / If the equipment remains inoperative, turn it OFF and call Field Service.
pressurestat mal-function (housing is cool). Heat units may
raise to any value.
Press the “AEC Reset” push-button and call Field Service. This error
E98 Service Mode Active.
does not inhibit normal operation.
It is recommended that the following procedure will be performed for X-ray Tube
warm-up, at the start of each day and when the Tube selected has not been in
use for approximately one hour.
4. Instruct patient to maintain the required position. Prepare the X-ray Tube
by pressing the handswitch button to the “Prep” position and maintain it
until the “Ready” indicator is illuminated.
5. Instruct patient to remain still and to hold their breath as required, then
make the X-ray exposure by pressing the handswitch button fully to the
“Exp” position and maintain it throughout the exposure. The “X-ray On”
indicator will light and an alarm will sound during the exposure.
3. Select the “workstation” and enter in AEC mode by selecting at least one
Area Detector “Field” on the NX Station or on the DX-D Console.
5. Select the technique parameters (back-up time / mAs) using the RAD
controls on the Console.
6. Continue with the radiographic operation. (Refer to Section 7.3 -- step 4.)
4. Collimate the X-Ray beam so that it completely covers all three Ion
Chambers (Left, Center and Right).
6. Set a technique, for example: 70 kVp, 250 mA, 1.0 second back-up time.
Make a RAD exposure and note the exposure mAs and time. For a
proper functioning of the AEC, the exposure must not be aborted by the
AEC back-up timer, that is, the “AEC Reset” push-button is not flashing.
Make a RAD exposure and note the exposure mAs and time. For a
proper functioning of the AEC, the exposure must not be aborted by the
AEC back-up timer, that is, the “AEC Reset” push-button is not flashing.
Make a RAD exposure and note the exposure mAs and time. For a
proper functioning of the AEC, the exposure must not be aborted by the
AEC back-up timer, that is, the “AEC Reset” push-button is not flashing.
10. The noted Exposure mAs and time have to be equal ¦10% between
all three Ion Chambers. If not, contact Service.
11. Repeat the above steps changing the Density and/or the homogeneous
phantom (e.g. a bucket with 5 cm of water).
Compare the Exposure mAs and time between each Ion Chamber and
between the values noted before (for a lower density or less water, lower
mAs and a shorter time; for half of density or half of water, half of mAs
/ time). If not, contact Service.
12. Finally, check the proper functioning of the AEC back-up timer by making
a RAD exposure with the selections indicated in step 6., but with the
Collimator blades fully closed.
The exposure must be finished by the AEC back-up timer, that is, the
exposure length is 1.0 second and the “AEC Reset” Push-button is
flashing. If not, contact Service.
If any of these situations occur and the “Preparation” button on the Handswitch
is pressed, one of the following messages appears on the Console
accompanied by an error code (depending on the type of Console used in the
System):
There are two levels of maintenance, the first consists of tasks which are
performed by the user/operator, and the second are those tasks to be
performed by qualified X-ray service personnel.
Heavy duty installations (more than125 patients per day) require a regular six
(6) month maintenance.
The manufacturer undertakes to have available spare parts for this equipment
at least for ten (10) years after the unit manufacturing.
The tasks of this periodic maintenance shall include the following items:
Do not apply directly any liquid on the surfaces, nor use cleaners
containing bleach, ammonia or any other abrasive or solvent liquid, it
could cause damage to the equipment.
WORKING
700 hPa 1060 hPa 30% 75% 10 oC (50 oF) 35 oC (95 oF)
500 hPa 1060 hPa 10% 90% -20 oC (--4 oF) 70 oC (158 oF)
MINIMUM INPUT
POWER BRAKE POWER
EQUIPMENT VOLTAGE FREQUENCY POWER
OUTPUT SUPPLY
REQUIRED
RAD WALL STAND * See Note 0.5 kVA 0.10 kW (4A) 24 VDC, 2A
* NOTE: Power Line for the RAD Wall Stand is supplied by the X-ray Generator (110 -- 230 VAC, Single Phase, 50 / 60 Hz)
Dimensions
Height . . . . . . . . . . . . . . . . . . . . . . . . . . . 2370 mm
Width . . . . . . . . . . . . . . . . . . . . . . . . . . . 2755 mm
Length . . . . . . . . . . . . . . . . . . . . . . . . . . 1267 mm
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 345 kg
Dimensions
Height . . . . . . . . . . . . . . . . . . . . . . . . . . . 750 mm
Width . . . . . . . . . . . . . . . . . . . . . . . . . . . 2200 mm
Length . . . . . . . . . . . . . . . . . . . . . . . . . . 800 mm
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190 kg
Dimensions
Maximum Height . . . . . . . . . . . . . . . . . . 900 mm
Minimum Height . . . . . . . . . . . . . . . . . . 500 mm
Width . . . . . . . . . . . . . . . . . . . . . . . . . . . 2200 mm
Length . . . . . . . . . . . . . . . . . . . . . . . . . . 800 mm
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 280 kg
Dimensions
Height . . . . . . . . . . . . . . . . . . . . . . . . . . . 2235 mm
Width . . . . . . . . . . . . . . . . . . . . . . . . . . . 657 mm
Length . . . . . . . . . . . . . . . . . . . . . . . . . . 381 mm
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145 kg
Dimensions
Height . . . . . . . . . . . . . . . . . . . . . . . . . . . 2235 mm
Width . . . . . . . . . . . . . . . . . . . . . . . . . . . 657 mm
Length . . . . . . . . . . . . . . . . . . . . . . . . . . 890 mm
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 175,5 kg
GRIDS
Illustration 9-1
Dimensions of the Equipments in the X-ray System (RAD Table -- Elevating Table).
1626
64”
1106.5
43.6”
o
180
o
135
50”
1267
8.9”
225
Illustration 9-2
Dimensions of the Equipments in the X-ray System (RAD Table -- Fixed Height Table).
1626
64”
1106.5
43.6”
o
180
o
135
50”
1267
8.9”
225
Illustration 9-3
Dimensions of the RAD Wall Stand
381 (15”)
2235 (88”)
1860 (73.2”)
1500 (59”)
200 (7.8”)
360 (14.1”)
657 (25.8”)
485 (19”)
559 (22”)
Illustration 9-4
IDimensions of the RAD Wall Stand - Manual Tilting
1500 (59”)
1500 (59”)
2155 (84.8”)
1860 (73.2”)
399 (15.7”)
512 (20.1”)
655 (25.7”)
360 (14.1”)
90º
570 (22.4”)
20º
2235 (88”)
657 (25.8”)
Receptor Assembly
485 (19”)
559 (22”)
Table-top
9.3.1 FACTORS
Maximum Power kW 32 kW 50 kW 64 kW 80 kW
Maximum kVp 150 kVp 150 kVp 150 kVp 150 kVp
A C D
230 / 240 VAC, Single-Phase, 50 / 60 Hz 400 / 415 / 440 VAC, Three-Phase, 50 / 60 Hz 480 VAC, Three-Phase, 50 / 60 Hz
NOTES: -- For Generators operating with lines at 208 VAC or below an Auxiliary Boost Transformer is required to adequate the
line voltage to 230 / 240 VAC.
-- For 80 kW Generators operating with lines at 400 / 415 / 440 VAC an auxiliary boost transformer is required to adequate
the line voltage to 480 VAC.
PARAMETER RANGE
mAs Product of mA x Time values from 0.1 mAs to 500 mAs (640 mAs under request).
The Generator duty cycle is continuous, but limits should be set during
installation depending on the capacity of the X-ray Tube.
DIMENSIONS
COMPONENT WEIGHT
LENGTH WIDTH HEIGHT
Illustration 9-5
Dimensions
Cable Access
Cable Access
568
GENERATOR GENERATOR
(1 TUBE, HSS) (1 TUBE, LSS)
690
TABLE SUPPORT
PEDESTAL
WALL SUPPORT
9.5 COLLIMATORS
Please review the following link and reduce pediatric technique factors
accordingly: http://www.pedrad.org/associations/5364/ig/
As the use of Grids require higher doses, never use Grids in pediatric
exams. Remove the Grid from the receptor assembly and select the
lower possible doses. If the Grid can not be detached, pediatric exams
can not be performed using this device.
Appendix A A-1
DX-D 400
User Manual
Pediatric patients are not as likely as adults to understand the need to remain
still during the procedure. Therefore it makes sense to provide aids to
maintaining stable positioning. It is strongly recommended the use of
immobilizing devices such as bean bags and restraint systems (foam
wedges, adhesive tapes, etc.) to avoid the need of repeating exposures due to
the movement of the pediatric patients. Whenever possible use techniques
based on the lowest exposure times.
Shielding:
Technique factors:
You should take steps to reduce technique factors to the lowest possible levels
consistent with good image acquisition.
For example if your adult abdomen settings are: 70--85 kVp, 200--400 mA,
15--80 mAs, consider starting at 65--75 kVp, 100--160 mA, 2.5--10 mAs for a
pediatric patient. Whenever possible use high kVp techniques and large SID
(Source Image Distance).
Summary:
Use the lowest amount of radiation for adequate imaging based on size
of the child (reducing tube output -- kVp and mAs).
Try to use always short exposure times, large SID values and
immobilizing devices.
A-2 Appendix A
DX-D 400
User Manual
Note . For DX-D Touch Screen Console, the language of the factory
pre-programmed APR techniques is configured according to the
customer order and it can not be changed by the operator (refer
to Section B.1).
Note . For DX-D Overlay Console with APR, the language pack of the
factory pre-programmed APR techniques is configured according
to the customer order and it can not be changed by the operator
(refer to Section B.1).
Appendix B B-1
DX-D 400
User Manual
LATERAL
AP
SKULL SEMIAXIAL THORAX ODONTOID AP
LATERAL
AXIAL
SPECIAL
AP
AP/PA
ORBITA -- UPPER RIBS C-SPINE LATERAL
OBLIQUE
OBLIQUE
AP
PETROUS AP/PA
BONES -- LOWER RIBS T-SPINE LATERAL
SKULL / STENVERS THORAX OBLIQUE SPINE
OBLIQUE
AP
LATERAL
MASTOID -- STERNUM L-SPINE LATERAL
OBLIQUE
OBLIQUE
AP/PA
AP
PARANASAL
-- SCAPULA LATERAL L5-- S1
SINUSES
LATERAL
OBLIQUE
AP
SELLA LATERAL CLAVICLE AP/PA SACRUM
LATERAL
NOSE LATERAL
AP/PA
AP
LATERAL AP
ABDOMEN FEMUR SHOULDER SEMIAXIAL
DECUBITUS LATERAL
OBLIQUE
STANDING
AP AP
KIDNEY AP KNEE HUMERUS
LATERAL LATERAL
LATERAL PA
KIDNEY
AP PATELLA ELBOW
/ URETER
AXIAL LATERAL
AP AP PA
ABDOMEN / GALL- LOWER UPPER
LOWER LEG FOREARM
PELVIS BLADDER EXTREMITIES EXTREMITIES
OBLIQUE LATERAL LATERAL
AP AP
AP
PELVIS ANKLE LATERAL WRIST LATERAL
LATERAL
OBLIQUE OBLIQUE
AP AP AP
AP
TOES AP/LAT/OBLIQ FINGER
LATERAL
B-2 Appendix B
DX-D 400
User Manual
AP/PA
LATERAL
AP
CRANE SEMIAXIAL THORAX ATLAS AP
LATERAL
AXIAL
SPECIAL
AP
AP/PA
ORBITAE -- COTES SUP. COL. CERV LATERAL
OBLIQUE
OBLIQUE
AP
AP/PA
M. PEDRO -- COTES INF. COL. DORS LATERAL
CRANE THORAX OBLIQUE COLONNE
OBLIQUE
AP
LATERAL
MASTOID -- STERNUM COL. LOMB LATERAL
OBLIQUE
OBLIQUE
AP/PA
AP
SINUS -- OMOPLATE LATERAL L5-- S1
LATERAL
OBLIQUE
AP
SELLA LATERAL CLAVICLE AP/PA SACRUM
LATERAL
NEZ LATERAL
AP/PA
AP
LATERAL AP
ABDOMEN FEMUR EPAULE SEMIAXIAL
DECUBITUS LATERAL
OBLIQUE
DEBOUT
AP AP
REINS AP GENOU HUMERUS
LATERAL LATERAL
LATERAL PA
URETERE AP ROTULE COUDE
AXIAL LATERAL
AP AP PA
ABDOMEN / EXTREMITES EXTREMITES
VES. BIL. JAMBE INF. BRAS. INF.
BASSIN INFERIEUR SUPERIEUR
OBLIQUE LATERAL LATERAL
AP AP
AP
PELVIS CHEVILLE LATERAL POIGNET LATERAL
LATERAL
OBLIQUE OBLIQUE
AP AP AP
AP
ORTEIL AP/LAT/OBLIQ DOIGTS
LATERAL
Appendix B B-3
DX-D 400
User Manual
AP/PA
LATERAL
AP
CRANEO SEMIAXIAL TORAX ODONTAL AP
LATERAL
AXIAL
ESPECIAL
AP
AP/PA
COSTILLAS COLUMNA
ORBITA -- LATERAL
SUPERIORES CERVICAL
OBLICUO
OBLICUO
AP
AP/PA
COSTILLAS COLUMNA
PEÑASCO -- LATERAL
INFERIORES DORSAL
CRANEO TORAX OBLICUO COLUMNA
OBLICUO
AP
LATERAL
COLUMNA
MASTOID -- ESTERNON LATERAL
LUMBAR
OBLICUO
OBLICUO
AP/PA
AP
SEN. PAR. -- ESCAPULA LATERAL L5-- S1
LATERAL
OBLICUO
AP
SILLA TURCA LATERAL CLAVICULA AP/PA SACRO
LATERAL
NARIZ LATERAL
AP/PA
AP
LATERAL AP
ABDOMEN FEMUR HOMBRO SEMIAXIAL
DECUBIT0 LATERAL
OBLICUO
CARGA
AP AP
RIÑON AP RODILLA HUMERO
LATERAL LATERAL
LATERAL PA
RIÑON /
AP ROTULA CODO
URETRA
AXIAL LATERAL
AP AP PA
ABDOMEN / VESICULA EXTREMIDADES PIERNA EXTREMIDADES
ANTEBRAZO
PELVIS BILIAR INFERIORES INFERIOR SUPERIORES
OBLICUO LATERAL LATERAL
AP AP
AP
PELVIS TOBILLO LATERAL MUÑECA LATERAL
LATERAL
OBLICUO OBLICUO
AP AP AP
AP
DEDOS DEDOS DE
AP/LAT/OBL
DEL PIE LA MANO
LATERAL
B-4 Appendix B
DX-D 400
User Manual
AP/PA
LATERAL
AP
SCHÄEDEL HALBAX. THORAX DENS AP
LATERAL
AXIAL
SPEZIAL
AP
AP/PA
OBERER
ORBITA -- HWS LATERAL
RIPPEN
OBLIQUE
OBLIQUE
AP
AP/PA
UNTERER
STENVERS -- BWS LATERAL
RIPPEN
SCHÄEDEL THORAX OBLIQUE W. SAEULE
OBLIQUE
AP
LATERAL
MASTOID -- STERNUM LWS LATERAL
OBLIQUE
OBLIQUE
AP/PA
AP
NNH -- SCAPULA LATERAL L5-- S1
LATERAL
OBLIQUE
AP
SELLA LATERAL CLAVICUL AP/PA SAKRUM
LATERAL
NASE LATERAL
AP/PA
AP
LATERAL AP
ABDOMEN O. SCHENKEL SCHULTER HALBAX.
DECUBIT. LATERAL
OBLIQUE
STEHEND
AP AP
NIERE AP KNIE OBERARM
LATERAL LATERAL
LATERAL PA
NIERE / UR AP PATELLA ELLENBOGEN
AXIAL LATERAL
AP AP PA
ABDOMEN / GALLEN- UNTERER OBERER
U. SCHENKEL UNTERARM
BECKEN BLASE EXTREM. EXTREM.
OBLIQUE LATERAL LATERAL
AP AP
AP
BECKEN SPRGEL. LATERAL HANDGELENK LATERAL
LATERAL
OBLIQUE OBLIQUE
AP AP AP
AP
ZEHEN AP/LAT/OBLIQ FINGER
LATERAL
Appendix B B-5
DX-D 400
User Manual
AP/PA
BOCZNE
AP
CZASZKA POLOSIOWE KLATKA P OBROTNIK AP
BOCZNE
OSIOWE
SPECIAL
AP
AP/PA
OCZODOLY -- ZEBRA G. KR. SZYJ. BOCZNE
SKOS
SKOS
AP
AP/PA
USZY SR. -- ZEBRA D. KR. PIER. BOCZNE
CZASZKA KLATKA P SKOS KREGOSL.
SKOS
AP
BOCZNE
PODSTAWA -- MOSTEK KR. LEDZ. BOCZNE
SKOS
SKOS
AP/PA
AP
ZATOKI -- LOPATKA BOCZNE C. L5-- S1
BOCZNE
SKOS
AP
SIODELKO BOCZNE OBOJCZYK AP/PA K. KRZYZ.
BOCZNE
NOS BOCZNE
AP/PA
AP
BOCZNE AP
J. BRZUSZ UDO BARK POLOSIOWE
BOK-LEZ BOCZNE
SKOS
BOK-STOJ
AP AP
NERKI AP KOLANO RAMIE
BOCZNE BOCZNE
BOCZNE PA
P. MOCZ. AP RZEPKA LOKIEC
OSIOWE BOCZNE
AP AP PA
J. BRZUSZ P. ZOLCI. KONCZ D. GOLEN KONCZ G. PRZEDR.
SKOS BOCZNE BOCZNE
AP AP
AP
MIEDNICA STAW ZK. BOCZNE NADGARS. BOCZNE
BOCZNE
SKOS SKOS
AP AP AP
AP
PALCE AP/BOK PALCE
BOCZNE
B-6 Appendix B
DX-D 400
User Manual
AP/PA
LATERAL
AP
KAFA SEMIAKSIYAL TORAKS ODONTOID AP
LATERAL
AKSIYAL
ÖZEL
AP
AP/PA
GÖZ ÇUK -- Ü. KABUR. C. OMUR LATERAL
OBLIK
OBLIK
AP
AP/PA
STENVERS -- A. KABUR. T. OMUR LATERAL
KAFA TORAKS OBLIK OMURGA
OBLIK
AP
LATERAL
MOSTOID -- STERNUM L. OMUR LATERAL
OBLIK
OBLIK
AP/PA
AP
SIN.PAR. -- SKAPULA LATERAL L5-- S1
LATERAL
OBLIK
AP
SELLA LATERAL KLAVICU. AP/PA SAKRUM
LATERAL
BURUN LATERAL
AP/PA
AP
LATERAL AP
BATIN FEMUR OMUZ SEMIAKS
DEKUBIT LATERAL
OBLIK
AYAKTA
AP AP
BÖBREK AP DIZ KOL
LATERAL LATERAL
LATERAL PA
ÜRINER AP PATELLA DIRSEK
AKSIYAL LATERAL
AP AP PA
A. Ü.
BATIN/PELVIS SAFRA K. CRURAL ÖN KOL
EXTREM EXTREM
OBLIK LATERAL LATERAL
AP AP
AP
PELVIS A. BILEGI LATERAL BILEK LATERAL
LATERAL
OBLIK OBLIK
AP AP AP
AP
A. PARMAK AP/LAT/OBLIK PARMAK
LATERAL
Appendix B B-7
DX-D 400
User Manual
B-8 Appendix B
DX-D 400
User Manual
AP/PA
LATERAL
SKULL SEMIAXIAL
AXIAL
SPECIAL
ORBITA
MASTOID
PARANASAL
SELLA LATERAL
NOSE LATERAL
AP/PA
LATERAL
SKULL SEMIAXIAL
AXIAL
SPECIAL
ORBITA
MASTOID
PARANASAL
SELLA LATERAL
NOSE LATERAL
Appendix B B-9
DX-D 400
User Manual
AP
THORAX
LATERAL
AP/PA
UPPER RIBS
OBLIQUE
AP/PA
LOWER RIBS
OBLIQUE
LATERAL
STERNUM
OBLIQUE
AP/PA
SCAPULA LATERAL
OBLIQUE
CLAVICLE AP/PA
AP
THORAX
LATERAL
AP/PA
UPPER RIBS
OBLIQUE
AP/PA
LOWER RIBS
OBLIQUE
LATERAL
STERNUM
OBLIQUE
AP/PA
SCAPULA LATERAL
OBLIQUE
CLAVICLE AP/PA
B-10 Appendix B
DX-D 400
User Manual
ODONTOID AP
AP
CSSPINE LATERAL
OBLIQUE
AP
T-SPINE LATERAL
OBLIQUE
AP
L-SPINE LATERAL
OBLIQUE
AP
L5 S1
L5-S1
LATERAL
AP
SACRUM
LATERAL
ODONTOID AP
AP
CSSPINE LATERAL
OBLIQUE
AP
T-SPINE LATERAL
OBLIQUE
AP
L-SPINE LATERAL
OBLIQUE
AP
L5 S1
L5-S1
LATERAL
AP
SACRUM
LATERAL
Appendix B B-11
DX-D 400
User Manual
AP / PA
LATERAL
ABDOMEN
DECUBITUS
STANDING
KIDNEY AP
KIDNEY / URETER AP
AP
GALLBLADDER
OBLIQUE
AP
PELVIS
LATERAL
AP
AXIAL
AP / PA
LATERAL
ABDOMEN
DECUBITUS
STANDING
KIDNEY AP
KIDNEY / URETER AP
AP
GALLBLADDER
OBLIQUE
AP
PELVIS
LATERAL
AP
AXIAL
B-12 Appendix B
DX-D 400
User Manual
AP
FEMUR
LATERAL
AP
KNEE
LATERAL
LATERAL
PATELLA
AXIAL
AP
LOWER LEG
LATERAL
AP
ANKLE LATERAL
OBLIQUE
AP
FOOT LATERAL
OBLIQUE
TOES AP/LAT/OBLIQ
AP
FEMUR
LATERAL
AP
KNEE
LATERAL
LATERAL
PATELLA
AXIAL
AP
LOWER LEG
LATERAL
AP
ANKLE LATERAL
OBLIQUE
AP
FOOT LATERAL
OBLIQUE
TOES AP/LAT/OBLIQ
Appendix B B-13
DX-D 400
User Manual
AP
SHOULDER SEMIAXIAL
OBLIQUE
AP
HUMERUS
LATERAL
PA
ELBOW
LATERAL
PA
FOREARM
LATERAL
AP
WRIST LATERAL
OBLIQUE
AP
HAND LATERAL
OBLIQUE
AP
FINGER
LATERAL
AP
SHOULDER SEMIAXIAL
OBLIQUE
AP
HUMERUS
LATERAL
PA
ELBOW
LATERAL
PA
FOREARM
LATERAL
AP
WRIST LATERAL
OBLIQUE
AP
HAND LATERAL
OBLIQUE
AP
FINGER
LATERAL
B-14 Appendix B
Manufacturer: Agfa NV,
Septestraat 27, B--2640 Mortsel -- Belgium