APPLICATIONS
Central Sterilizing Supply Department (CSSD)
Operating Theater
Out Patient Clinic
Rev. 06-08-2009 Tuttnauer - Your Sterilization & Infection Control Partners Page 1 of 10
TECHNICAL SPECIFICATION
66 Mid Range Sterilizer Series
PRODUCT SPECIFICATION
CHAMBER VOLUME & SIZE VOLTAGE SUPPLY
Three phase or one phase electricity supply is available.
VOLUME 415V 3 Phase with Neutral
MODEL CHAMBER DIMENSIONS (mm)
(liter) 380V / 400V 3 Phase with Neutral
Width Height Depth 230V 3 Phase with Neutral
208V 3 Phase without Neutral
6690 915 340
220V / 230V 1 Phase *
610 610
66120 1215 450 110V / 115V 1 Phase *
* Must have building steam supply and vacuum by ejector
6671130 1295 610
Frequency: 50 Hz or 60 Hz
660 710
6671162 1620 760 STEAM GENERATOR
MODEL
POWER (kW)
6690
DOOR SELECTION 66120 36
Single door 6671130
Double door (pass through)
6671162 63
Right hinged (control panel on left, automatic locking) Customized electrical specifications available (additional cost).
Left hinged (control panel on right, automatic locking)
Standard Configuration Materials: Built-in Steam Generator Safety: A water level control
Chamber + Door: 316L stainless steel system maintains a constant water level that ensures safe
operation of the heaters. Furthermore, a sensor ensures that
Jacket: 304L stainless steel the heater does not operate without a minimum water intake.
Optional Materials (Stainless Steel): Emergency Shut-Off: Easily accessible emergency switches
Chamber + Door: 316Ti for immediate cycle shut-off.
Jacket: 316L or 316Ti
Double Independent Monitoring: The combined digital and
mechanical monitoring systems provide a cross-reference and
Chamber Design
guarantee accurate results. The operator has two means of
The chamber is fully jacketed. The chamber floor is slightly
monitoring temperature and pressure.
sloped towards an internal drain to facilitate drainage. A
stainless steel mesh strainer protects the drain port from
blockage by debris. The chamber is mounted on a steel
framework which is height adjustable. WATER RING VACUUM PUMP
The vacuum pump effectively removes more than 99% air from
Door Gasket the chamber. The pump is mounted on a damping mechanism
Manual hinged door gasket: a silicone gasket is permanently to minimize vibration.
fixed in the door.
During vacuum action the flow path of the fluids is through the
Automatic sliding door gasket: a silicone gasket is fixed in a condenser (heat exchanger) and the vacuum pump/ejector to
groove in the door frame and the gasket is sealed against the drain. This protects the vacuum system from excessive
door by means of steam pressure. temperatures.
Surface Treatment
A passivation layer is applied to the internal surface through an PIPES & COMPONENTS
electro-chemical treatment which results in a smooth and shiny The piping system of the autoclave consists of air-operated ball
stainless steel. The resultant surface is polished to less than valves, which control the condensate and steam flow in and out
0.8 µm which is highly protected against corrosion. The of the chamber, operates the vacuum, and the air inlet valve.
internal corners are rounded to aid cleaning.
Standard configuration:
Chamber Heating Piping and fittings: 304 threaded stainless steel
The fully jacketed chamber is pre-heated by steam in the Components*: Brass
jacket.
Optional configuration (stainless steel):
Insulation
The sterilizer jacket and door are completely insulated with a 1" Primary piping & fittings **: 316 + triclamp fittings
chloride free glass wool thereby keeping the autoclave cool on Primary components: 316 triclamps or 316 threaded
the outside. Customized thicker insulation is optional. The
insulation is completely enclosed within a stainless steel cover. * Components – Manual valve, non-return valve,
pressure regulator, pneumatic valves, steam trap, etc.
The insulation reduces the energy consumption by keeping the
heat inside the jacket and chamber. ** Primary piping system that connects to the chamber
up to and including the first valve.
SAFETY FEATURES
Door Safety Systems: PRESSURE GAUGES
The single door configuration has four pressure gauges
A pneumatic safety component (pressure switch) is
that indicate the pressure in the chamber, jacket, door
installed in the autoclave which prevents opening of the
gasket, and the steam generator line.
doors until pressure in the chamber reaches room
pressure.
These pressure gauges are located on the autoclave's front
Door chamber cannot be opened when chamber is panel (loading side).
pressurized.
Steam is not allowed into the chamber when the door is The double door configuration has two additional pressure
open. gauges on the unloading side that indicate the pressure in
A cycle cannot start if the door is open or not properly the chamber and the pressure in the gasket of the second
locked. door.
The sliding door movement will stop immediately if an
object is detected in front of the closing door.
Double door safety is implemented through interlocks which
prevent both doors from being opened simultaneously.
The touch screen control panel supports 23 languages. Temperature & pressure sensor failure
Phase time-outs
Twelve password access levels (0-12) are provided to control Door(s) not properly closed
access/operation of the machine preventing unauthorized Power failure
access. These access levels are customizable. Access No water in the water reservoir
control can be applied to the following functions operated from Optional utility alarms: no water / no steam / no air
the control panel:
running test cycles
sensor calibration, service and maintenance CYCLE DOCUMENTATION - PRINTER
cycle selection, cycle start, customizing cycle parameters The autoclave is equipped with a printer which prints a detailed
18 preset sterilization programs and 2 preset test programs history of each cycle performed by the instrument. Various
door control types of printers are listed in the Special Options and
testing inputs and outputs Accessories section towards the end of this document.
With the standard factory configuration, calibration of the The printing format is 24 characters per line. The following
temperature circuits and calibration of the pressure circuits information and set parameters are printed when the
require an access code. sterilization cycle begins:
The data is printed from the bottom up, beginning with the date
and ending with “O.K.” for a complete cycle or “FAIL” for an
aborted cycle.
TESTING CYCLES
STERILIZATION CYCLES Program 11: Bowie & Dick Test
The autoclave is supplied with 10 preprogrammed sterilization This program is intended to test the efficiency of air removal
cycles and 2 preprogrammed test cycles. An authorized user from within the chamber.
can change the parameters of each cycle. The number and
type of vacuum pulses, the atmospheric pressure value, and Test parameters:
the pressure profile are also programmable. There are 6 Sterilization temperature 134 ºC
selectable exhaust modes. The cycle programs are Sterilization time 3.5 minutes
summarized as follows: Fast exhaust
Drying time of 2 minutes
Ventilation
10 air replacements per hour
Note: Upon request Tuttnauer can provide special configurations so that the autoclaves will fit through narrow passage-ways / elevators
during transportation in and out of buildings.
Notes:
Par 3, 4 Residual air is displaced by 4 vacuum pulses (down to 25 kPa) and 3 steam pulses up
Atmospheric
to 160 kPa. Drying is performed during a "low-pressure" phase. When the pressure
Pressure
reaches atmospheric pressure the vacuum pump is activated until the pressure drops
to 25 kPa. Only the number and type of vacuum pulses can be modified.
Program 12 Air Leakage/Vacuum Test Air leakage test to chamber through door and other seals
Parameters Units Range Factory Setting
Atmospheric 1. Stabilizing Time min - 5
Pressure 2. Test Time min - 10
Par 1, 2
Notes:
Chamber vacuum is brought to 10 kPa. Thereafter all valves and motors are closed for
5 minutes enabling pressure stabilization. The acceptable pressure change over the
5 min 10 min
following 10 minutes is 1.3 kPa.