Professional Documents
Culture Documents
Total Artificial Hearts Past, Present, Future
Total Artificial Hearts Past, Present, Future
the risk of infection was clearly seen to parts that are most likely to fail because the right side and 15 mmHg on the left.14
increase with the duration of support. In of cyclic fatigue are outside the body, All these TAHs were powered by trans-
1983, approval of the immunosuppressant where they can be closely monitored and cutaneous energy transfer systems based
drug ciclosporin led to dramatic improve- promptly exchanged. At present, exter- on inductive coupling of radiofrequency
ments in long-term survival after cardiac nal drivers are exchanged pre-emptively alternating current through the intact
transplantation and to a resurgence in every 6–12 weeks. The flexible components skin without the need for a percutaneous
interest in this therapy. As a result, the implanted in the chest—the pneumatically driveline. An implanted battery provided
focus shifted from developing permanent actuated diaphragms—also have limited continuous operation in the event of tran-
replacements for failing hearts to bridg- durability, because they must flex between sient disconnection of the transcutaneous
ing patients with biventricular failure to 30 million and 50 million times per year. energy transfer system. In addition, TAHs
heart transplantation. The transcutaneous air hoses are an impor- were developed in parallel at the National
The Jarvik 7 heart trials were temporar- tant physiological liability, because of the Cerebral and Cardiovascular Center, Suita,
ily suspended by the FDA in 1990, 8 but increased incidence of ascending driveline Osaka Prefecture, Japan,15 and the Vacord
resumed after extensive technical refine- infections in patients undergoing long-term Bioengineering Research Company, Brno,
ments were made to the device. The device support. Although device infections have Czech Republic.16,17
has been renamed several times to the caused few deaths in patients awaiting trans-
Symbion Artificial Heart, the CardioWest plantation, the mean duration of SynCardia Sarns-3M TAH
TAH, and (for the past 13 years) the support is 15–90 days at different centres. In The Sarns‑3M TAH (Figure 2) was designed
SynCardia temp orary TAH (SynCardia a large series of 171 patients, 60% were sup- by pioneering innovator Dick Sarns and
Systems Inc., USA). 9 Refinements have ported by the SynCardia device for <2 weeks, used a reciprocating brushless direct-
included reduced size of the pumping cham- and the average duration of support was current motor and translating roller screw
bers from 100 cm3 to 70 cm3, integration of a 24 days.13 A total of 37% of patients experi to alternately actuate left and right pusher
specifically designed modified Medtronic– enced severe infectious complications plates, which alternately compressed sys-
Hall carbon-pyrolytic tilting disc valve that necessitating urgent transplantation. temic and pulmonary blood sacs made
is more suitable for the high closing forces from seamless polyether polyurethane urea
than the previous valves (because of rigid Totally implantable TAH technology to provide stroke volumes of up to 90 ml.18,19
mounting), and improvements in manu- Although the results of heart transplanta- While one side of the TAH ejected, the other
facturing techniques. The device received tion continued to improve after the intro- side filled. A total of 4.5 rotations of the
CE mark approval in 1999 and was the first duction of better immunosuppressive motor caused the screw to travel 19 mm.
TAH approved by the FDA, in 2004, for use agents, the supply of donor hearts remained The pump displacement was 64 cm3, and the
as a bridge to heart transplantation.10 a limiting factor. The National Heart, Lung, maximum output was 8 l/min at 125 bpm.
Various iterations of the SynCardia and Blood Institute increased funding in Motor direction was reversed when the left
temp orary TAH have subsequently been 1988 to develop a completely implantable pump filled, as detected by three Hall-effect
implanted in >1,300 gravely ill patients, TAH with the aim of producing a practi- sensors on the left pump membrane and
around 80% of whom have been successfully cal long-term heart-replacement device. by monitoring of the motor current. The
bridged to heart transplantation (1‑year sur- Many ambitious attempts were made to pusher plates were not attached directly to
vival is 70%).11 A total of 241 patients have develop durable, self-contained, volume- the blood sacs which could, therefore, fill
survived for >6 months, 92 for >1 year, 16 for displacement TAHs with internal actuation passively. Early left pump filling was evi-
>2 years, and three for >3 years; the longest mechanisms that eliminated the need for dence of high atrial pressure, which resulted
duration of bridging so far is 1,374 days.10 an external driver and percutaneous drive- in an autonomous decrease in right pump
A further intended development is to lines or pneumatic hoses. The resulting stroke volume to restore balance between
make a 50 cm3 device, which could substan- devices included the Sarns‑3M TAH (3M the systemic and pulmonary circulations.
tially improve access to this technology for Health Care, USA, in conjunction with the The space between the blood sacs was con-
children and small adults. A portable mini Pennsylvania State University, University nected to a gas-filled compliance chamber,
ature compressor that weighs 6.1 kg and can Park , PA, U S A) , t he Ni mbus TAH which required percutaneous injection of
function for 3 h on fully charged batteries has (Nimbus, USA, in conjunction with the additional air every 6 weeks. As in other
improved quality of life by enabling patients Cleveland Clinic, Cleveland, OH, USA), TAHs, each chamber was fitted with Björk–
to participate in a wide range of activities, the AbioCor®TAH (ABIOMED, USA, in Shiley inlet and outlet valves. The inter-
such as golf or bike riding, that were not conjunction with Texas Heart Institute nal battery provided 45 min operation in
possible with earlier compressors.12 and the Jewish Hospital in Louisville, the event of transient disconnection of the
The use of external drivers for the KY, USA), and the electrohydraulic TAH transcutaneous energy transfer system.
SynCardia temporary TAH has several (developed at the University of Utah, In mock circulation loop testing, the
advantages over internal drivers. By inte- Salt Lake City, UT, USA). Each of these device could accommodate a wide range of
grating the electronic components and most devices met the National Heart, Lung, preload pressures without suction events,
of the mechanical complexity in the para- and Blood Institute criteria for success, autonomously increased pump output if
corporeal pneumatic compressor, the com- which included the capacity to pump left atrial pressure rose (1 l/min/mmHg),
ponents that need to be implanted could 8 l/min against a mean systemic blood pres- and showed relative insensitivity to after-
be decreased in size and mechanistically sure of 110 mmHg and a mean pulmonary load. Flow balancing was achieved by esti-
simplified, which facilitated implanta- artery pressure of 25 mmHg while main- mating the left pump’s end-diastolic volume
tion in smaller patients. In addition, those taining a filling pressure of ≤10 mmHg on on the basis of motor speed and voltage,
Nimbus TAH
The Nimbus TAH (Figure 3) used a brush-
less direct-current motor and gear pump
to generate hydraulic pressure, which actu-
ated a hydraulic piston. 14 A spool valve
redirected hydraulic fluid to reverse the
piston direction. On either end of the recip-
rocating piston were flat plates that alter-
nately compressed systemic and pulmonary
blood sacs, but were not attached to them
in order to allow passive filling, as in the
Sarns‑3M TAH. The hydraulic mechanism
was positioned in the 21 mm space between
Figure 2 | Sarns‑3M positive-displacement
the blood sacs, which was vented to a gas-
Nature Reviews | Cardiology
filled intrathoracic compliance chamber. The Figure 4 | AbioCor® Nature Reviews | Cardiology
positive-displacement
total artificial heart (3M Health Care, USA,
blood-contacting surfaces of the pumps were total artificial heart (ABIOMED, USA).
and Pennsylvania State University, University
Permission obtained from ABIOMED, USA.
Park, PA, USA). Reprinted with permission lined with a seamless coating of glutaralde-
from Rosenberg, G. et al. A roller screw drive hyde cross-linked gelatin, which was deemed
for implantable blood pumps. ASAIO J. 28 (1), biocompatible and was intended to eliminate one side of the TAH to the other, so that one
123–126 (1982). the need for systemic anticoagulation. Four blood pump was filling while the other was
bovine pericardial tissue valves (or, in some ejecting. The inlet and outlet ports were
iterations, human dura mater valves) were positioned on a cylinder that continuously
integrated into the two blood sacs. The outer rotated around the hydraulic pump and cycli-
shell of the device was constructed from cally changed the direction of the silicone oil
epoxy reinforced with carbon fibre. without necessitating a change in the rota-
The hydraulic cylinder stroke length tional direction of the motor. This design was
was 13.2 mm, and the maximum stroke believed to improve mechanical durability.
volume was 64 cm3, but the normal operat- The blood-contacting surfaces of the
ing stroke volume was 53 cm3. Maximum AbioCor®were constructed from proprietary
output was 9.6 l/min at 150 bpm. As with multisegmented polyurethane. The TAH was
the Sarns‑3M device, in mock circulation manufactured in such a way that the entire
loop testing the Nimbus TAH maintained blood path was seamless, including the
systemic–pulmonarybalance and autono- points of attachment of the four trileaflet
mously increased pump output if left atrial valves, which were also fabricated from the
pressure rose (0.5 l/min/mmHg).14 Iterations same material. The blood-contacting com-
of the device were implanted in 12 calves, ponent was housed in a polycarbonate shell.
which survived for an average of 32 days The TAH had a stroke volume of 80 cm3 and
(longest survival 120 days).22 Several animals could generate up to 9.6 l/min of blood flow
Figure 3 | Nimbus Nature Reviews | Cardiology
positive-displacement could exercise on a motorized treadmill for at an ejection rate of 120 bpm. The device
total artificial heart (Nimbus, USA, and an average of 22 min. Despite these encour- was fairly large, measuring 100 mm in width
Cleveland Clinic, Cleveland, OH, USA). aging results, many experiments had to be and 85 mm in diameter.
Reprinted from McCarthy, P. M. et al. The terminated prematurely because of mechan- The AbioCor®had a unique mechanism
Cleveland Clinic-Nimbus total artifical heart. ical failures and, ultimately, the risk was to equalize pulmonary and systemic flow.
In vivo hemodynamic performance in calves
deemed prohibitive to proceeding to human A small reservoir connected to the right
and preclinical studies. J. Thorac. Cardiovasc.
Surg. 108 (3), 420–428 © (1994), with implantation. hydraulic space and separated from the
permission from Elsevier. left atrium by a thin flexible membrane
AbioCor® TAH (2001) contained silicone oil. When left atrial
The AbioCor®TAH (Figure 4) was the only pressure became excessively high, the oil
and by adjusting the right pump’s dia fully implantable self-contained TAH from was displaced from the reservoir into the
stolic time.20 Devices were implanted in 14 this era to be used in humans. Again, the right hydraulic space and decreased end-
calves by 1993, and the longest survival was device consisted of two pumping chambers diastolic filling of the right blood pump.
150 days.21 Causes of premature termina- with flexible membranes alternately com- As a result, the right pump stroke volume
tion of the experiment included respiratory pressed by an internal mechanism.23 The would be decreased for the next heartbeat,
failure, infection, and device failure. Formal actuation mechanism, however, was unique resulting in decreased return of blood to
review of design readiness began in 1999. in that there were no pusher plates. A centrif- the left atrium and an autonomous reduc-
Despite moderate success, performance ugal pump was used to pressurize hydraulic tion in left atrial pressure. This technique
was never considered adequate for human fluid, which alternately compressed the flex- was refined to produce a sensitivity to
implantation, especially because of the ible medial aspects of the two blood pumps preload of 0.4 l/min/mmHg to keep the
device’s large size, poor system reliability, directly. The centrifugal pump, which rotated total volume of blood and hydraulic fluid
and limited durability. at 4,000–8,000 rpm, shuttled silicone oil from within the pump constant over time.24 As a
increase autonomously during exercise by continuous-flow TAH, the patient devel- SmartHeart
means of a similar mechanism.43 oped haemoptysis, and a biopsy revealed The SmartHeart TAH, being developed by
Different pumps showed different degrees amyloid infiltration of the lungs. Hepatic Cleveland Heart and Cleveland Clinic, has
of pressure sensitivity because of differing encephalopathy progressed and support was a cylindrical rotating element supported
impeller designs. A hypothesis was formed discontinued after 5 weeks.46 Although no by a fluid-film hydrodynamic bearing.51,52
that an impeller could be designed spe- successful long-term outcome was seen in The length of the cylinder is roughly twice
cifically to optimize this functionality and, this case, it demonstrated for the first time its diameter, and the cylinder is capped at
therefore, improve autonomous pulmonary– that a continuous-flow TAH could be used both ends by centrifugal flow impellers.
systemic balance.44 Nevertheless, to accom- in a human and suggested that assessment When the systemic–pulmonary circulation
modate the somewhat reduced pressure was warranted. becomes imbalanced because of a transient
sensitivity compared with that of the natural Pirk and colleagues performed a similar perturbation in physiology, the hypofunc-
heart, slightly elevated filling pressures operation on a patient (aged 38 years) with tioning side of the TAH is exposed to pro-
would be required to keep the atria from infiltrating cardiac fibrosarcoma.47 The pro- gressively increasing atrial pressure, which
transient collapse. The introduction of feed- cedure was initially successful, but the patient creates a hydrodynamic force on the spin-
back control of pump speed was considered died from a fulminant Aspergillus spp. infec- ning element and shifts the axis of rotation
as a means of mitigating this problem and tion after 6.5 months.48 Other groups have towards the side with lower atrial pressure.
improving the capacity of these pumps to championed the use of dual HeartWare® Each impeller is positioned in such a way
balance flows under all conditions.45 HVADs, either as biventricular assist devices with respect to the volute in which it is spin-
The thrombotic complications that occa- or as cardiac replacement devices after exci- ning that alignment is improved or wors-
sionally occurred in the right pump were sion of the ventricles.49 The clinical expe- ened depending on whether the rotor shifts
thought to be caused by venous thrombi rience with continuous-flow pumps for to the left or right. This change adjusts the
that were liberated from peripheral veins biventricular replacement suggests a possible relative outputs of the left and right sides
and became entangled in the pump mecha- role for specialized continuous-flow TAHs of the device and continuously balances
nism. This hypothesis was supported by that leverage the small size, mechanistic pumping efficiency without the need for
the observation that the left pump was simplicity, and improved durability of rotary sensors or feedback algorithms. Chronic
unaffected, presumably because it pumped blood pumps. testing of the device in animals is ongoing,
blood filtered by the lung microcirculation. but 6‑week survival has been reported in
Modulation of the right pump speed—by Single-device rotary TAHs one experiment.53
transiently decreasing speed by as much In 1987, Qian and colleagues attempted to
as 30% for 0.20 s every 2 s—reduced right create a single-device rotary TAH inten BiVACOR
pump complications, perhaps by improv- ded for long-term cardiac replacement, The BiVACOR rotary TAH (Figure 6b)50
ing pump washout, decreasing the areas of but could not adequately balance the left originated in Australia and is being devel-
stasis, and facilitating the passage of small and right circulation. 38 Howe ver, after oped by BiVACOR and a multinational
thromboemboli arising in the periphery the success of supporting patients with consortium of research centres. The device
through the right pump. HF using rotary blood pumps, BiVACOR uses a different mechanism from the
In 2012, a patient (aged 55 years) with (Houston, TX, USA)50 and Cleveland Heart SmartHeart TAH to achieve pulmonary–
systemic amyloidosis involving the heart, (Cleveland, OH, USA)51,52 each developed systemic balance. The rotating element in
liver, and kidneys was placed on para a specialized rotary TAH with intrinsic the BiVACOR TAH is more of a disc than
corporeal left-heart bypass. The patient flow-balancing mechanisms. These TAHs a cylinder, with its diameter being consid-
was deemed too ill to survive a heart trans- contain single rotating elements with sys- erably greater than its length. The spinning
plantation. Cardiac amyloidosis is generally temic and pulmonary impellers on oppos- element is suspended by an active magnetic
a contraindication for the SynCardia tem- ing faces. The rotational speed of the left and bearing system, in which stable levitation is
porary TAH because of technical issues, right impellers is the same, and the differ- maintained by variations in magnetic forces
and the patient’s ventricles were too small ent flow characteristics of the systemic and generated by electromagnets in response
to accommodate an LVAD inflow cannula. pulmonary circulation are achieved with to position feedback sensors.54 As in the
Frazier and Cohn suggested implant- differences in impeller size and geometry. SmartHeart, small shifts in the position
ing two HeartMate II® devices to create Similarly, small changes in the position of of the rotating element along the axis of
a continuous-flow TAH.46 Custom cuffs the rotating element along the axis of rota- rotation substantially change the relative
were fashioned from polypropylene hernia tion result in a substantial change in relative pumping efficiencies of the two pumps,
mesh, Dacron®cardiovascular patches, and efficiency of the two opposing pumps and while the rotor position in the BiVACOR
medical silicone. The inflow cannulas of the provide autonomous systemic–pulmonary device is determined by the differential
two HeartMate II®LVADs were removed, flow balance. Although the mechanisms pressures (forces) acting on the rotor. Spe
and the pumps were attached to the cuffs for flow balance differ between the two cifically, the surrounding pressures cause
by a previously developed technique. 46 devices, they both offer the advantages of the magnetic system to pull the rotor auto-
The implantation was uneventful, and the centrifugal pumps. Specifically, neither has matically towards the underperforming side,
patient was extubated on day 2 after surgery any mechanical bearings or other sources which decreases the efficiency of the oppo-
in a haemodynamically stable condition. of mechanical wear and no flexible com- site pump so that the hydrodynamic forces
The patient, however, remained anuric on ponents or valves, and both have only one that arise from differences in the left and
haemodialysis and liver failure continued to moving part and are extremely compact and right atrial pressure are precisely counter
progress. After 4 weeks of support with the energy efficient. balanced. 55 During in vitro testing, the
device demonstrated a maximum capacity of four pericardial tissue valves; the blood- Future challenges include gaining addi-
16 l/min at 100 mmHg and, when operated contacting surfaces of the blood sacs are lined tional insights into the effects of reduced or
with speed modulation, can produce flow with a microporous biocompatible material absent pulsatile blood flow. Devices using
pulses of 0–18 l/min and pulse pressures of aimed at obviating the need for anticoagula- positive-displacement technology continue
120/80 mmHg at 60 bpm. tion. In addition, multiple sensors embedded to be developed and used clinically. However,
An early series of nine acute ovine studies in the device provide autonomous regulation given their limitations in terms of durabil-
demonstrated the hydraulic performance of the pump rate and output in response to ity, they are unlikely ever to be considered a
and balancing capacity of the single-rotor activity level and physiological factors. The true alternative to transplantation. Instead,
concept.56 Later chronic studies in calves microprocessor that controls the device is initial successful experiments leveraging two
involved seven implantations, with the integrated into the pump housing, where it LVADs as a TAH suggest that a rotary TAH
longest duration being 30 days in a 71 kg is cooled by silicone oil hydraulic fluid. The has potential. The reduced size, improved
calf. The outflow for this calf varied from device is quite large and heavy (1 kg), but durability, and reduced power consumption
7 l/min to 15 l/min and averaged 12 l/min at computer-aided anatomical fitting studies inherent to rotary blood pumps might lead
100 mmHg at a motor power consumption of have predicted that it will fit 85% of men and to rotary TAHs succeeding where previous
13 W. The animal could exercise on a tread- 65% of all patients. A percutaneous power attempts have failed to provide a practical
mill for 30 min, and autonomously balanced lead enters through the posterior scalp and is replacement for the human failing heart.
pulmonary–systemic flow was maintained affixed to a skull-mounted pedestal, as previ-
despite changes in posture and activity, gen- ously described by Jarvik,58 in an attempt to Cardiovascular Surgical Research Laboratory
and the Center for Cardiac Support, Texas
erating flows of 16 l/min without evidence decrease the incidence of driveline infection.
Heart Institute, MC 2‑114A, PO Box 20345,
of haemolysis. A CARMAT®TAH was implanted into Houston, TX 77225, USA (W.E.C., D.L.T., O.H.F.).
a patient aged 75 years, who was supported Correspondence to: W.E.C.
Positive-displacement TAHs and haemodynamically stable for 74 days, wcohn@texasheart.org
Despite the potential benefits of continuous- at which point the device suddenly stopped.
1. Go, A. S. et al. Heart disease and stroke
flow TAHs, whether the physiological After an in-depth analysis and subsequent
statistics—2014 update: a report from the
limitations imposed by nonpulsatile or regulatory approval, a second implantation American Heart Association. Circulation 129,
reduced-pulsatile perfusion will prove to has been performed, and device support e28–e292 (2014).
be prohibitive remains unknown. For this was ongoing at the time of writing. 2. Colvin-Adams, M. et al. OPTN/SRTR 2012
annual data report: heart. Am. J. Transplant. 14
reason, SynCardia is developing a positive- (Suppl. 1), 113–138 (2014).
displacement TAH with a 50 cm3 blood- ReinHeart TAH 3. Akutsu, T. & Kolff, W. J. Permanent substitutes
pumping chamber to enable use in a wider ReinHeart (Bad Oeynhausen, Germany) for valves and hearts. ASAIO J. 4, 230–234
(1958).
range of patients. Additionally, two European and the Helmholtz Institute, RWTH-Aachen 4. Cooley, D. A. et al. Orthotopic cardiac
groups have readdressed the challenges in University (Aachen, Germany) are develop- prosthesis for two-staged cardiac replacement.
developing a self-contained, totally implant- ing the ReinHeart TAH. This device has a Am. J. Cardiol. 24, 723–730 (1969).
able, durable volume-displacement TAH by reciprocating piston to alternately com- 5. Frazier, O. H., Akutsu, T. & Cooley, D. A. Total
artificial heart (TAH) utilization in man. ASAIO J.
applying new technology. press two polyurethane sacs as the left and 28, 534–538 (1982).
right ventricles. In an attempt to improve 6. Kolff, W. J. Artificial organs—forty years and
SynCardia durability, a voice-coil and linear actuator beyond. Trans. Am. Soc. Artif. Intern. Organs 29,
6–24 (1983).
Owing to limited use of the SynCardia combination is used instead of a screw and 7. DeVries, W. C. et al. Clinical use of the total
temporary TAH device in paediatric and roller bearings. The stroke volume of the artificial heart. N. Engl. J. Med. 310, 273–278
young-adult patients (7% of 1,091 total ReinHeart TAH is only 50 ml and, there- (1984).
8. Leprince, P. et al. Bridge to transplantation with
implantations between 1985 and 2012), the fore, the size is 84 × 90 mm and the weight
the Jarvik‑7 (CardioWest) total artificial heart:
company has been developing a smaller is 923 g. 59,60 An implantable compliance a single-center 15-year experience. J. Heart Lung
version, called the SynCardia 50/50. 57 chamber allows passive filling of the ventric- Transplant. 22, 1296–1303 (2003).
The device is intended to be implanted in ular sacs, which provides a degree of inher- 9. Slepian, M. J. The SynCardia Temporary Total
Artificial Heart—evolving clinical role and future
patients with a body surface area <1.85 m2. ent passive flow balancing. A maximum status. US Cardiol. 8, 39–46 (2011).
In an attempt to meet the longer-term needs output of 7 l/min has been demonstrated in 10. Torregrossa, G. et al. Results with SynCardia
of patients with HF, the device has also been calves; the longest survival was 50 h.61 total artificial heart beyond 1 year. ASAIO J. 60,
626–634 (2014).
approved by the FDA for use as destina- 11. Copeland, J. G. et al. Cardiac replacement
tion therapy under the Humanitarian Use Conclusions with a total artificial heart as a bridge to
Device programme.10 Over the past 50 years, researchers have transplantation. N. Engl. J. Med. 351, 859–867
focused on developing a safe, durable, and (2004).
12. Friedline, K. & Hassinger, P. Total artificial heart
CARMAT practical TAH to meet the clinical demands freedom driver in a patient with end-stage
The CARMAT®TAH (CARMAT, France; created by end-stage HF. Early devices aimed biventricular heart failure. AANA J. 80, 105–112
Figure 6c), which is under development at replicating the natural physiological pulse (2012).
13. Johnson, K. E., Prieto, M., Joyce, L. D.,
by Carpentier and colleagues, uses two were large, heavy, and prone to mechani- Pritzker, M. & Emery, R. W. Summary of the
reciprocating rotary gear pumps to alter- cal failure, dampening enthusiasm for clinical use of the Symbion total artificial heart:
nately shuttle hydraulic fluid between two this technology. A shift in focus to LVADs a registry report. J. Heart Lung Transplant. 11,
103–116 (1992).
blood sacs with compressible diaphragms. and the success with rotary blood pumps
14. Massiello, A. et al. The Cleveland Clinic-Nimbus
Although superficially similar to the renewed interest and paved the way for total artificial heart. J. Thorac. Cardiovasc. Surg.
AbioCor®TAH, the CARMAT®TAH has new approaches to total heart replacement. 108, 412–419 (1994).
15. Tatsumi, E. et al. In vivo evaluation of the National 33. Frazier, O. H. & Leon, P. J. Small pumps for 48. Netuka, I. et al. Novel treatment of an infiltrating
Cardiovascular Center electrohydraulic total ventricular assistance: progress in mechanical cardiac fibrosarcoma. Tex. Heart Inst. J. 41,
artificial heart. Artif. Organs 23, 242–248 (1999). circulatory support. Cardiol. Clin. 25, 553–564 248–249 (2014).
16. Vasku, J. et al. 150-day survival of a calf with (2007). 49. Strueber, M., Schmitto, J. D., Kutschka, I.
a polymethylmethacrylate total artificial heart: 34. Frazier, O. H. et al. First clinical use of the & Haverich, A. Placement of 2 implantable
TNS‑BRNO‑II. Artif. Organs 5, 388–400 (1981). redesigned HeartMate II Left Ventricular Assist centrifugal pumps to serve as a total artificial
17. Vasku, J. & Urbanek, P. Constructional and System in the United States. Tex. Heart Inst. J. heart after cardiectomy. J. Thorac. Cardiovasc.
functional characteristics of recent total 31, 157–159 (2004). Surg. 143, 507–509 (2012).
artificial heart models TNS Brno VII, VIII, and IX. 35. Kirklin, J. K. et al. Interagency Registry for 50. Timms, D. et al. The BiVACOR rotary
Artif. Organs 19, 535–543 (1995). Mechanically Assisted Circulatory Support biventricular assist device: concept and in vitro
18. Rosenberg, G. et al. A roller screw drive for (INTERMACS) analysis of pump thrombosis in investigation. Artif. Organs 32, 816–819
implantable blood pumps. Trans. Am. Soc. Artif. the HeartMate II left ventricular assist device. (2008).
Intern. Organs 28, 123–126 (1982). J. Heart Lung Transplant. 33, 12–22 (2014). 51. Fukamachi, K. et al. An innovative, sensorless,
19. Snyder, A. J. et al. An electrically powered total 36. Salamonsen, R. F., Mason, D. G. & Ayre, P. J. pulsatile, continuous-flow total artificial heart:
artificial heart: over 1 year survival in the calf. Response of rotary blood pumps to changes device design and initial in-vitro study. J. Heart
ASAIO J. 38, M707–M712 (1992). in preload and afterload at a fixed speed Lung Transplant. 29, 13–20 (2010).
20. Weiss, W. J. et al. Steady state hemodynamic setting are unphysiological when compared 52. Fumoto, H. et al. In vivo acute performance of
and energetic characterization of the Penn with the natural heart. Artif. Organs 35, E47–E53 the Cleveland Clinic self-regulating, continuous-
State/3M Health Care Total Artificial Heart. (2011). flow total artificial heart. J. Heart Lung Transplant.
ASAIO J. 45, 189–193 (1999). 37. Golding, L. R. et al. Chronic nonpulsatile blood 29, 21–26 (2010).
21. Snyder, A. J. et al. In vivo testing of a completely flow in an alive, awake animal 34-day survival. 53. Karimov, J. H. et al. In vivo evaluation of the
implanted total artificial heart system. ASAIO J. Trans. Am. Soc. Artif. Intern. Organs 26, 251–255 Cleveland Clinic continuous-flow total artificial
39, M177–M184 (1993). (1980). heart in calves [abstract]. J. Heart Lung
22. Harasaki, H. et al. Progress in Cleveland 38. Qian, K. X., Pi, K. D., Wang, Y. P. & Zhao, M. J. Transplant. 33 (Suppl.), S165–S166 (2014).
Clinic‑Nimbus total artificial heart development. Toward an implantable impeller total heart. 54. Greatrex, N., Timms, D., Kurita, N., Palmer, E.
ASAIO J. 40, M494–M498 (1994). ASAIO Trans. 33, 704–707 (1987). & Masuzawa, T. Axial magnetic bearing
23. Kung, R. T. V. et al. Progress in the development 39. Slaughter, M. S. et al. Advanced heart failure development for the BiVACOR Rotary BiVAD/
of the ABIOMED Total Artificial Heart. ASAIO J. treated with continuous-flow left ventricular TAH. IEEE Trans. Biomed. Eng. 57, 714–721
41, M245–M248 (1995). assist device. N. Engl. J. Med. 361, 2241–2251 (2010).
24. Kung, R. T. V., Yu, L.‑S., Ochs, B. & Frazier, O. H. (2009). 55. Timms, D., Kurita, N., Greatrex, N.
An atrial hydraulic shunt in a total artificial 40. Frazier, O. H., Tuzun, E., Cohn, W., Tamez, D. & Masuzawa, T. BiVACOR—a magnetically
heart: a balance mechanism for the bronchial & Kadipasaoglu, K. A. Total heart replacement levitated biventricular artificial heart [abstract].
shunt. ASAIO J. 39, M213–M217 (1993). with dual centrifugal ventricular assist devices. Presented at the 20th MAGDA Conference.
25. Dowling, R. D. et al. Initial experience with the ASAIO J. 51, 224–229 (2005). 56. Timms, D. et al. Initial acute in-vivo animal
AbioCor implantable replacement heart system. 41. Frazier, O. H., Tuzun, E., Cohn, W. E., Conger, J. L. experience with the BiVACOR Rotary Bi-
J. Thorac. Cardiovasc. Surg. 127, 131–141 & Kadipasaoglu, K. A. Total heart replacement Ventricular Assist Device [abstract].
(2004). using dual intracorporeal continuous-flow Artif. Organs 31, A67 (2007).
26. Nose, Y. FDA approval of totally implantable pumps in a chronic bovine model: a feasibility 57. Ryan, T. D., Jefferies, J. L., Zafar, F., Lorts, A.
permanent total artificial heart for humanitarian study. ASAIO J. 52, 145–149 (2006). & Morales, D. L. S. The evolving role of the total
use. Artif. Organs 31, 1–3 (2007). 42. Cohn, W. E. et al. Eight-year experience with a artificial heart in the management of end-stage
27. Frazier, O. H. et al. Multicenter clinical continuous-flow total artificial heart in calves. congenital heart disease and adolescents.
evaluation of the HeartMate 1000 IP left ASAIO J. 60, 25–30 (2014). ASAIO J. 61, 8–14 (2015).
ventricular assist device. Ann. Thorac. Surg. 53, 43. Frazier, O. H., Cohn, W. E., Tuzun, E., 58. Jarvik, R., Westaby, S., Katsumata, T., Pigott, D.
1080–1090 (1992). Winkler, J. A. & Gregoric, I. D. Continuous-flow & Evans, R. D. LVAD power delivery:
28. Frazier, O. H. Chronic left ventricular support total artificial heart supports long-term survival a percutaneous approach to avoid infection.
with a vented electric assist device. of a calf. Tex. Heart Inst. J. 36, 568–574 (2009). Ann. Thorac. Surg. 65, 470–473 (1998).
Ann. Thorac. Surg. 55, 273–275 (1993). 44. Frazier, O. H., Khalil, H. A., Benkowski, R. J. 59. Fritschi, A. et al. Image based evaluation of
29. Rose, E. et al. The REMATCH trial: rationale, & Cohn, W. E. Optimization of axial-pump mediastinal constraints for the development
design and end points: Randomized Evaluation pressure sensitivity for a continuous-flow of a pulsatile total artificial heart. BioMed. Eng.
of Mechanical Assistance for the Treatment of total artificial heart. J. Heart Lung Transplant. OnLine 12, 81 (2013).
Congestive Heart Failure. Ann. Thorac. Surg. 67, 29, 687–691 (2010). 60. Pohlmann, A. et al. Experimental validation
723–730 (1999). 45. Khalil, H. A. et al. Continuous flow total artificial of the linear drive train for a total artificial
30. Frazier, O. H. et al. Initial clinical experience heart: modeling and feedback control in a heart system. Mechatronics 23, 222–226
with the Jarvik 2000 implantable axial-flow mock circulatory system. ASAIO J. 54, 249–255 (2013).
left ventricular assist system. Circulation 105, (2008). 61. Laumen, M. et al. A novel total artificial heart
2855–2860 (2002). 46. Frazier, O. H. & Cohn, W. Continuous-flow for destination therapy: in-vitro and in-vivo
31. Wampler, R. K., Moise, J. C., Frazier, O. H. total heart replacement device implanted in study. Biomed. Tech. http://dx.doi.org/
& Olsen, D. B. In vivo evaluation of a a 55-year-old man with end-stage heart failure 10.1515/bmt-2013-4373.
peripheral vascular access axial flow blood and severe amyloidosis. Tex. Heart Inst. J. 39,
pump. ASAIO Trans. 34, 450–454 (1988). 542–546 (2012). Author contributions
32. Frazier, O. H. et al. First human use of the 47. Pirk, J. et al. Total artificial heart support with W.E.C. and D.L.T. researched data for the article.
Hemopump, a catheter-mounted ventricular two continuous-flow ventricular assist devices All the authors discussed the content of the
assist device. Ann. Thorac. Surg. 49, 299–304 in a patient with an infiltrating cardiac sarcoma. article, wrote the manuscript, and reviewed drafts
(1990). ASAIO J. 59, 178–180 (2013). before submission.