Download as pdf or txt
Download as pdf or txt
You are on page 1of 9

PERSPECTIVES

OPINION septal defects, previously implanted pros-


thetic valves, or complex native valve disease.
Total artificial hearts: past, present, These patients would benefit from develop-
ment of self‑contained, energy-efficient­,
and future durable TAHs.

William E. Cohn, Daniel L. Timms and O. H. Frazier TAH device requirements


Multiple technical challenges are associated
Abstract | A practical artificial heart has been sought for >50 years. An increasing with the development of a self-contained,
number of people succumb to heart disease each year, but the number of hearts implantable TAH for use as destination
available for transplantation remains small. Early total artificial hearts mimicked therapy. The device must fit within the
the pumping action of the native heart. These positive-displacement pumps could mediastinum of a high percentage of patients
provide adequate haemodynamic support and maintain the human circulation for with HF and must be configured to facilitate
short periods, but large size and limited durability adversely affected recipients’ suturing to the left and right atrial remnants
quality of life. Subsequent research into left ventricular assist devices led to the use and the great vessels after the diseased native
of continuous-flow blood pumps with rotating impellers. Researchers have attempted heart is excised. The TAH must generate
to integrate this technology into modern total artificial hearts with moderate clinical adequate cardiac output to accommodate
success. The importance of pulsatile circulation remains unclear. Future research is, the physiological needs of the patient and
some level of exertion, and the device must
therefore, needed into positive-displacement and rotary total artificial hearts.
do so without injuring the formed elements
Cohn, W. E. et al. Nat. Rev. Cardiol. 12, 609–617 (2015); published online 2 June 2015; of the blood or generating thromboemboli.
doi:10.1038/nrcardio.2015.79 Devices must be biocompatible, must not
generate excessive heat, and must last for
Introduction exclusively as a bridge to transplantation. ≥5 years—all without consuming exces-
The need for a practical mechanical replace- The search for a permanent (destination sive amounts of power. If a transcutaneous
ment for the failing human heart has moti- therapy), totally implantable artificial heart energy transfer system is used, an internal
vated scientists and clinicians for more than continues. Although newer self-contained battery must have the capacity to power
half a century. The deceptive simplicity of volume-displacement TAHs might succeed the TAH long enough to allow reposition-
the heart and the magnitude of the unmet where earlier devices failed, interest is also ing of the external component if necessary.
clinical need associated with advanced being shown in leveraging continuous-flow Perhaps most challenging is the need for
heart failure (HF) have attracted a diverse rotary blood pump technology. the TAH to balance the systemic and pul-
group of innovators, including Charles Heart disease remains the leading cause of monary circulations despite a wide variety
Lindbergh, Paul Winchell, Alexis Carrel, death in the USA and most of the developed of h­aemodynamic perturbations.
Willem Kolff, and Robert Jarvik. Most world. Each year, almost 400,000 people die The entire output of the right ventricle is
efforts have focused on creating total arti- from HF-related causes in the USA alone,1 pumped through the pulmonary circulation
ficial hearts (TAHs) with pulsatile output, and this number is likely to increase. Cardiac to the left atrium, but the same is not true of
composed of two volume-displacement transplantation is an effective treatment, but the left ventricular output. Bronchial arter­
pumps, each with unidirectional inlet and limited donor-organ availability has resulted ial branches arising from the descending
outlet valves. TAHs with self-contained in a modest epi­demiological effect. For the thoracic aorta drain through the pulmon­
actuation mechanisms have been devised, past 20 years, only 2,000–2,500 hearts have ary veins. A substantial proportion of the
but all have been limited by large size and been transplanted annually in the USA and, left ventricular output, therefore, returns to
poor durability. Some TAHs have performed therefore, a very small proportion of patients the left rather than the right atrium. Con­
well in chronic animal studies and even in have benefitted from them.2 Many patients sequently, the output of the left ven­tricle
small clinical pilot studies. Encouraging who are not selected for transplantation might be 10–15% greater than that of the
results notwithstanding, the only artificial and who have HF that primarily affects only right ventricle. If this balance is not prop-
heart that has been implanted with any the left ventricle can be treated with a per- erly maintained, the lesser-functioning­side
frequency employs an external pneumatic manent left ventricular assist device (LVAD), will rapidly develop atrial hyper­tension. Left
actuation mechanism that is connected to and some have had implanted pumps for atrial hypertension results in pulmonary
the device via a pair of transcutaneous air >1 decade. Nevertheless, many patients oedema and respiratory failure. Right atrial
hoses. As a result, this device is used almost have severe biventricular failure or other hypertension results in anasarca, ascites, and
conditions that would make LVAD implant­ renal and hepatic insufficiency. Efforts to
ation difficult or unwise, such as refractory decrease excessive atrial pressure in a TAH
Competing interests
D.L.T. is the founder of BiVACOR, Inc. The other HF and large recent infarctions, intract­able recipient who develops pulmonary–systemic
authors declare no competing interests. arrhythmias, postinfarction ventricular imbalance can result in suction events on

NATURE REVIEWS | CARDIOLOGY VOLUME 12 | OCTOBER 2015 | 609


© 2015 Macmillan Publishers Limited. All rights reserved
PERSPECTIVES

a b c the flocked Dacron® lining of the inner


surface. At post-mortem examination, the
surfaces of the device that came into contact
with blood were observed to be lined with
a smooth neointima of platelets and fibrin.
Therefore, although implantation of the
Liotta–Cooley TAH was a momentous clini-
cal and engineering feat, ultimately it was
not considered a success.

Akutsu III TAH (1981)


On the basis of this initial clinical experi-
ence with a TAH, Akutsu and colleagues
Figure 1 | Three early TAHs. a | Liotta–Cooley TAH. b | Akutsu III TAH. cNature
| Jarvik 7 TAH. | Cardiology
Reviews working in the Texas Heart Institute Cullen
Abbreviation: TAH, total artificial heart. All images © Texas Heart Institute. Cardiovascular Research Laborator y
developed the Akutsu III artificial heart
(Figure 1b). As with the Liotta–Cooley
the low-pressure side, which might lower Wada–Cutter hingeless valves by Cooley heart, this device used compressed air and
cardiac output. A mechanism that auton­ immediately before clinical implementa- inflatable air sacs to displace blood, but
omously assesses and maintains proper tion). Dacron®skirts attached to the inlet the blood-contacting surfaces were con-
balance is essential. valves were sutured to the left and right atrial structed from smooth, multisegmented
remnants after the native heart was excised. polyurethane, and the device contained
Early beginnings Knitted Dacron® tube grafts attached to four Björk–Shiley carbon-pyrolytic tilting
One of the first descriptions of an artificial the outlet valves were sutured end to end disc valves. In 1981, after numerous success-
heart was made by Leonardo da Vinci in the to the pulmonary artery and aorta. When ful implantations in animals, the Akutsu III
15th century, but it was not until 1957 that 70–80 cm3 of compressed air from the exter- TAH was used as a bridge to transplantation
a TAH was completely implanted into an nal drive console was used to inflate the in a patient aged 36 years who had devel-
animal by Akutsu and Kolff.3 This milestone pneumatic side of the flexible diaphragm, oped refractory ventricular fibrillation after
sparked a boom in the research of TAHs a similar volume of blood was displaced undergoing CABG surgery.5 The TAH pro-
that culminated in Michael DeBakey’s suc- from the opposite side of the rigid chamber vided adequate haemodynamic support, but
cessful request to US President Lyndon B. through the outlet valves into the aorta and after 24 h, the patient developed profound
Johnson for financial support for a TAH pulmon­ary artery. When the air was rapidly hypoxia requiring venovenous extracor-
programme, which was established in 1964. withdrawn, the left and right chambers poreal membrane oxygenation, probably
The challenge was set to produce a fully refilled with blood from the pulmonary and because of pulmonary venous obstruc-
functional artificial heart by the time that systemic veins, respectively. By repeating tion by the device. The TAH supported
humans set foot on the moon. this cycle 70 times per minute, an output of the patient for 55 h until transplantation.
5 l/min could be achieved. The patient died 1 week later from multi­
Liotta–Cooley TAH (1969) On 4 April 1969, 3 months before the organ failure. No haemolysis was seen
In 1961, Domingo Liotta, who had already first lunar landing, Denton A. Cooley, of during implantation, presumably because
been working on a TAH at the University the Texas Heart Institute, Houston, TX, of improved materials.
of Cordoba, Argentina, was recruited to USA, became the first surgeon to implant
work in DeBakey’s laboratory at Baylor a TAH in a human.4 Cooley, an associate Jarvik 7 TAH (1982)
College of Medicine in Houston, TX, USA. of DeBakey at the time, offered the experi- Meanwhile, Kolff and Jarvik had been devel-
Liotta performed numerous experiments in mental therapy to a patient aged 47 years oping the Jarvik 7 TAH (Figure 1c).6 As with
calves with limited success. The TAH used as a bridge to transplantation if resection the Akutsu III TAH, the Jarvik 7 was con-
an external compressor to cyclically deliver and repair of his extensive left ventricular structed from multisegmented polyurethane
and evacuate compressed air to and from aneurysm were impossible. The patient and had four Björk–Shiley tilting valves.
two 100 ml rigid, flocked chambers lined could not be weaned from cardiopulmon­ After extensive validation in large animals,
with Dacron®(Du Pont de Nemours, USA), ary bypass and, therefore, the Liotta– approval was obtained from the FDA for
each partitioned into two halves—one for Cooley device (Figure 1a) was implanted. a small series of clinical implantations as
blood and the other for compressed air— The patient remained haemodynamically destination therapy. The first implantation
by a flexible silicone-elastic and Dacron® stable with the device for 64 h, until trans- was performed by DeVries in a patient with
diaphragm. One of the rigid chambers plantation was performed. He was extu- end-stage ischaemic HF.7 The patient lived
served as the systemic pump, the other as bated on the first day after surgery, but died for 112 days with the device. Over the next
the pulmonary pump. Two pairs of unidi- 32 h after transplantation, partly owing to year, the Jarvik 7 TAH was implanted in
rectional DeBakey ball valves at the inlets sepsis. Notably, haemolysis and progressive four further patients, one of whom lived for
and outlets of the blood-filled half of each renal impairment were seen in the early 620 days. Among the five patients, two had
of the two chambers ensured that blood period after implantation. These effects debilitating strokes. One died from haem-
flow was unidirectional (the DeBakey ball were attributed to the poor function of the orrhagic complications, and the other four
valves were subsequently replaced with hingeless valves used in the device and to died from septic complications. Therefore,

610 | OCTOBER 2015 | VOLUME 12  www.nature.com/nrcardio


© 2015 Macmillan Publishers Limited. All rights reserved
PERSPECTIVES

the risk of infection was clearly seen to parts that are most likely to fail because the right side and 15 mmHg on the left.14
increase with the duration of support. In of cyclic fatigue are outside the body, All these TAHs were powered by trans-
1983, approval of the immunosuppressant where they can be closely monitored and cutaneous energy transfer systems based
drug ciclosporin led to dramatic improve- promptly exchanged. At present, exter- on inductive coupling of radiofrequency
ments in long-term survival after cardiac nal drivers are exchanged pre-emptively alternating current through the intact
transplantation and to a resurgence in every 6–12 weeks. The flexible components skin without the need for a percutaneous
interest in this therapy. As a result, the implanted in the chest—the pneumatically driveline. An implanted battery provided
focus shifted from developing permanent actuated diaphragms—also have limited continuous operation in the event of tran-
replacements for failing hearts to bridg- durability, because they must flex between sient disconnection of the transcutaneous
ing patients with biventricular failure to 30 million and 50 million times per year. energy transfer system. In addition, TAHs
heart transplantation. The transcutaneous air hoses are an impor- were developed in parallel at the National
The Jarvik 7 heart trials were temporar- tant physiological liability, because of the Cerebral and Cardiovascular Center, Suita,
ily suspended by the FDA in 1990, 8 but increased incidence of ascending driveline Osaka Prefecture, Japan,15 and the Vacord
resumed after extensive technical refine- infections in patients undergoing long-term Bioengineering Research Company, Brno,
ments were made to the device. The device support. Although device infections have Czech Republic.16,17
has been renamed several times to the caused few deaths in patients awaiting trans-
Symbion Artificial Heart, the CardioWest plantation, the mean duration of SynCardia Sarns-3M TAH
TAH, and (for the past 13 years) the support is 15–90 days at different centres. In The Sarns‑3M TAH (Figure 2) was designed
SynCardia tem­p orary TAH (SynCardia a large series of 171 patients, 60% were sup- by pioneering innovator Dick Sarns and
Systems Inc., USA). 9 Refinements have ported by the SynCardia device for <2 weeks, used a reciprocating brushless direct-
included reduced size of the pumping cham- and the average duration of support was curren­t motor and translating roller screw
bers from 100 cm3 to 70 cm3, integration of a 24 days.13 A total of 37% of patients experi­ to alternately actuate left and right pusher
specifically designed modified Medtronic– enced severe infectious complications plates, which alternately compressed sys-
Hall carbon-pyrolytic tilting disc valve that ­necessitating urgent transplantation. temic and pulmonary blood sacs made
is more suitable for the high closing forces from seamless polyether polyurethane urea
than the previous valves (because of rigid Totally implantable TAH technology to provide stroke volumes of up to 90 ml.18,19
mounting), and improvements in manu- Although the results of heart transplanta- While one side of the TAH ejected, the other
facturing techniques. The device received tion continued to improve after the intro- side filled. A total of 4.5 rotations of the
CE mark approval in 1999 and was the first duction of better immunosuppressive motor caused the screw to travel 19 mm.
TAH approved by the FDA, in 2004, for use agents, the supply of donor hearts remained The pump displacement was 64 cm3, and the
as a bridge to heart transplantation.10 a limiting factor. The National Heart, Lung, maximum output was 8 l/min at 125 bpm.
Various iterations of the SynCardia and Blood Institute increased funding in Motor direction was reversed when the left
tem­p orary TAH have subsequently been 1988 to develop a completely implantable pump filled, as detected by three Hall-effect
implanted in >1,300 gravely ill patients, TAH with the aim of producing a practi- sensors on the left pump membrane and
around 80% of whom have been successfully cal long-term heart-replacement device. by monitoring of the motor current. The
bridged to heart transplantation (1‑year sur- Many ambitious attempts were made to pusher plates were not attached directly to
vival is 70%).11 A total of 241 patients have develop durable, self-contained, volume- the blood sacs which could, therefore, fill
survived for >6 months, 92 for >1 year, 16 for displacement TAHs with internal actuation passively. Early left pump filling was evi-
>2 years, and three for >3 years; the longest mechanisms that eliminated the need for dence of high atrial pressure, which resulted
duration of bridging so far is 1,374 days.10 an external driver and percutaneous drive- in an autonomous decrease in right pump
A further intended development is to lines or pneumatic hoses. The resulting stroke volume to restore balance between
make a 50 cm3 device, which could substan- devices included the Sarns‑3M TAH (3M the systemic and pulmonary circulations.
tially improve access to this technology for Health Care, USA, in conjunction with the The space between the blood sacs was con-
children and small adults. A portable mini­ Pennsylvania State University, University nected to a gas-filled compliance chamber,
ature compressor that weighs 6.1 kg and can Park , PA, U S A) , t he Ni mbus TAH which required percutaneous injection of
function for 3 h on fully charged batteries has (Nimbus, USA, in conjunction with the additional air every 6 weeks. As in other
improved quality of life by enabling patients Cleveland Clinic, Cleveland, OH, USA), TAHs, each chamber was fitted with Björk–
to participate in a wide range of activities, the AbioCor®TAH (ABIOMED, USA, in Shiley inlet and outlet valves. The inter-
such as golf or bike riding, that were not conjunction with Texas Heart Institute nal battery provided 45 min operation in
­possible with earlier compressors.12 and the Jewish Hospital in Louisville, the event of transient disconnection of the
The use of external drivers for the KY, USA), and the electrohydraulic TAH transcutaneous energy transfer system.
SynCardia temporary TAH has several (developed at the University of Utah, In mock circulation loop testing, the
advan­tages over internal drivers. By inte- Salt Lake City, UT, USA). Each of these device could accommodate a wide range of
grating the electronic components and most devices met the National Heart, Lung, preload pressures without suction events,
of the mechanical complexity in the para- and Blood Institute criteria for success, autonomously increased pump output if
corporeal pneumatic compressor, the com- which included the capacity to pump left atrial pressure rose (1 l/min/mmHg),
ponents that need to be implanted could 8 l/min against a mean systemic blood pres- and showed relative insensitivity to after-
be decreased in size and mechan­istically sure of 110 mmHg and a mean pulmonary load. Flow balancing was achieved by esti-
simplified, which facilitated implanta- artery pressure of 25 mmHg while main- mating the left pump’s end-diastolic volume
tion in smaller patients. In addition, those taining a filling pressure of ≤10 mmHg on on the basis of motor speed and voltage,

NATURE REVIEWS | CARDIOLOGY VOLUME 12 | OCTOBER 2015 | 611


© 2015 Macmillan Publishers Limited. All rights reserved
PERSPECTIVES

Nimbus TAH
The Nimbus TAH (Figure 3) used a brush-
less direct-current motor and gear pump
to generate hydraulic pressure, which actu-
ated a hydraulic piston. 14 A spool valve
re­directed hydraulic fluid to reverse the
piston direction. On either end of the recip-
rocating piston were flat plates that alter-
nately compressed systemic and pulmonary
blood sacs, but were not attached to them
in order to allow passive filling, as in the
Sarns‑3M TAH. The hydraulic mechanism
was positioned in the 21 mm space between
Figure 2 | Sarns‑3M positive-displacement
the blood sacs, which was vented to a gas-
Nature Reviews | Cardiology
filled intrathoracic compliance chamber. The Figure 4 | AbioCor® Nature Reviews | Cardiology
positive-displacement
total artificial heart (3M Health Care, USA,
blood-contacting surfaces of the pumps were total artificial heart (ABIOMED, USA).
and Pennsylvania State University, University
Permission obtained from ABIOMED, USA.
Park, PA, USA). Reprinted with permission lined with a seamless coating of glutaralde-
from Rosenberg, G. et al. A roller screw drive hyde cross-linked gelatin, which was deemed
for implantable blood pumps. ASAIO J. 28 (1), biocompatible and was intended to eliminate one side of the TAH to the other, so that one
123–126 (1982). the need for systemic anticoagulation. Four blood pump was filling while the other was
bovine pericardial tissue valves (or, in some ejecting. The inlet and outlet ports were
iterations, human dura mater valves) were positioned on a cylinder that continuously
integrated into the two blood sacs. The outer rotated around the hydraulic pump and cycli-
shell of the device was constructed from cally changed the direction of the silicone oil
epoxy reinforced with carbon fibre. without necessitating a change in the rota-
The hydraulic cylinder stroke length tional direction of the motor. This design was
was 13.2 mm, and the maximum stroke believed to improve mechanical durability.
volume was 64 cm3, but the normal operat- The blood-contacting surfaces of the
ing stroke volume was 53 cm3. Maximum AbioCor®were constructed from proprietary
output was 9.6 l/min at 150 bpm. As with multisegmented polyurethane. The TAH was
the Sarns‑3M device, in mock circulation manufactured in such a way that the entire
loop testing the Nimbus TAH maintained blood path was seamless, including the
systemic–pulmonary­balance and autono- points of attachment of the four tri­leaflet
mously increased pump output if left atrial valves, which were also fabricated from the
pressure rose (0.5 l/min/mmHg).14 Iterations same material. The blood-contactin­g com-
of the device were implanted in 12 calves, ponent was housed in a polycarbonate shell.
which survived for an average of 32 days The TAH had a stroke volume of 80 cm3 and
(longest survival 120 days).22 Several animals could generate up to 9.6 l/min of blood flow
Figure 3 | Nimbus Nature Reviews | Cardiology
positive-displacement could exercise on a motorized treadmill for at an ejection rate of 120 bpm. The device
total artificial heart (Nimbus, USA, and an average of 22 min. Despite these encour- was fairly large, measuring 100 mm in width
Cleveland Clinic, Cleveland, OH, USA). aging results, many experiments had to be and 85 mm in diameter.
Reprinted from McCarthy, P. M. et al. The terminated prematurely because of mechan- The AbioCor®had a unique mechanism
Cleveland Clinic-Nimbus total artifical heart. ical failures and, ultimately, the risk was to equalize pulmonary and systemic flow.
In vivo hemodynamic performance in calves
deemed p­rohibitive to proceeding to human A small reservoir connected to the right
and preclinical studies. J. Thorac. Cardiovasc.
Surg. 108 (3), 420–428 © (1994), with implantation. hydraulic space and separated from the
permission from Elsevier. left atrium by a thin flexible membrane
AbioCor® TAH (2001) contained silicone oil. When left atrial
The AbioCor®TAH (Figure 4) was the only pressure became excessively high, the oil
and by adjusting the right pump’s dia­ fully implantable self-contained TAH from was displaced from the reservoir into the
stolic time.20 Devices were implanted in 14 this era to be used in humans. Again, the right hydraulic space and decreased end-
calves by 1993, and the longest survival was device consisted of two pumping chambers diastolic filling of the right blood pump.
150 days.21 Causes of premature termina- with flexible membranes alternately com- As a result, the right pump stroke volume
tion of the experiment included respiratory pressed by an internal mechanism.23 The would be decreased for the next heartbeat,
failure, infection, and device failure. Formal actuation mechanism, however, was unique resulting in decreased return of blood to
review of design readiness began in 1999. in that there were no pusher plates. A centrif- the left atrium and an autonomous reduc-
Despite moderate success, performance ugal pump was used to pressurize hydraulic tion in left atrial pressure. This technique
was never considered adequate for human fluid, which alternately compressed the flex- was refined to produce a sensitivity to
implantation, especially because of the ible medial aspects of the two blood pumps preload of 0.4 l/min/mmHg to keep the
device’s large size, poor system re­liability, directly. The centrifugal pump, which rotated total volume of blood and hydraulic fluid
and limited durability. at 4,000–8,000 rpm, shuttled silicone oil from within the pump constant over time.24 As a

612 | OCTOBER 2015 | VOLUME 12  www.nature.com/nrcardio


© 2015 Macmillan Publishers Limited. All rights reserved
PERSPECTIVES

result, the AbioCor®was the only positive-


a b c
displacemen­t device that did not require a
gas-filled intrathoracic compliance chamber.
Over a 12‑year period, researchers im­plan­
ted AbioCor®devices into 120 calves; the last
57 of these devices used the electro­hydraulic
converter and autonomous pressure equali-
zation system described. Several of the
animals survived beyond the 90‑day term of
the study and could exercise on a motorized
treadmill. Formal design readiness testing
was begun in 1999, and >200 of the devices
Figure 5 | Transition of positive-displacement pumps to rotary blood pumps
Nature(left ventricular
Reviews | Cardiology
were manufactured. assist devices). a | HeartMate®XVE (TherMedics, USA, and later Thoratec Corporation, USA).
In 2001, Gray and Dowling implanted an Permission obtained from Thoratec Corporation, USA. b | Hemopump (Nimbus, USA). Reprinted
AbioCor®TAH in a human.25 Over the next from Frazier, O. H. et al. First human use of the Hemopump, a catheter-mounted ventricular
2 years, 13 additional TAHs were implanted assist device. Ann. Thorac. Surg. 49 (2), 299–304 © (1990), with permission from Elsevier.
at four US centres. All the patients were c | HeartWare®HVAD (HeartWare, USA). Image © Texas Heart Institute.
men who weighed 84–120 kg and, there-
fore, could accommodate the large size of
the device. Of these patients, three died The implantable HeartMate® pneumatic Despite these limitations, many research-
in the perioperative period from haemor- pump (TherMedics, USA, and later Thoratec ers believed that the HeartMate® XVE
rhage (n = 2) or air embolism (n = 1), and six Corp­­oration, USA) was first implanted in a pump performed well enough to justify a
patients died from multiorgan failure within human by Frazier in 1986.27 Although the multicentre, prospective, randomized study
the first 9 months. The remaining five original intention in developing implantable to compare outcomes with those of optimal
patients survived for 9–15 months, but died LVAD systems in the 1970s was to provide medical therapy. In the REMATCH trial,29
from complications related to stroke, infec- a permanent, completely implantable pump conducted in 129 patients with NYHA
tion, or organ failure, except the longest- for patients with left-sided HF, the intro- class IV HF, 1‑year survival was 52% in
surviving patient, in whom the device failed duction of ciclosporin and the success of patients who received the LVAD, compared
because an internal flexible membrane rup- heart transplantation resulted in a shift with 25% in those who received medical
tured. No serious complications related to of focus to bridging transplantation-eligible management only, and 2‑year survival was
the transcutaneous energy transfer system patients until a suitable donor heart could be 24% versus 8%, respectively. LVAD technol-
were reported. On the basis of these results, obtained. Development of a pneumatic pump ogy was, therefore, deemed to prolong life
study enrolment was suspended. was a comparatively cost-effective and easily in appropriately selected patients with end-
ABIOMED subsequently refined the attainable goal for this intermediate-term stage HF primarily affecting the left ventri-
device to decrease the overall size and sought application. As use of the HeartMate®pump cle. Nevertheless, a smaller, more-durable
approval for 60 additional implantations. The increased, the merit of mechanical circula- pump was required.
FDA approved the refined device for use in tory assistance of the left ventricle in trans-
humans, but the project was deemed com- plantation candidates became apparent, but Rotary blood pumps
mercially unviable and prohibitively difficult, its application was restricted by the scar- Rotary blood pumps were proposed to solve
and was discontinued in 2007.26 The prospect city of donor hearts and limitations of the the size and durability challenges posed by
of a totally implantable, permanent mechani- p­neumatic device. positive-displacement pumps, but were
cal heart‑replacement­device still seemed In an attempt to address these limitations, initially viewed with scepticism relating
very far off. Thoratec redirected its efforts towards devel- to the absence of a physiological pulse. In
oping the HeartMate®XVE (Figure 5a). This 1986, Jarvik and Wampler independently
Development of LVADs device was first implanted in a human by approached Frazier about starting animal
Although intense TAH research continued, Frazier in 1991.28 It was similar to the pneu- studies with implantable rotary blood
other groups focused on developing LVAD matic pump, but used a self-contained high- pumps. These efforts resulted in the first
technology because of the decreased tech- torque motor and rotating follower and cam two clinically used continuous-flow pumps:
nical demands associated with assisting a rather than pressurized air to actuate the the intraventricular Jarvik 2000® (Jarvik
weakened heart rather than replacing it. pusher plate; therefore, the device needed Heart, USA) and the temporary Hemopump
no external drive console, although a drive- (Nimbus, USA; Figure 5b).30–32 Development
Positive-displacement pumps line and pneumatic vent were still required of these devices showed compellingly that
Early short-term paracorporeal LVADs were to attach to an external battery pack. The continuous, nonpulsatile blood flow was
intended for use in patients who could not be device was large, which made it unsuitable physiologically tolerated, that rapidly spin-
weaned from cardiopulmonary bypass after for small patients and made the implant­ ning blood pumps could be designed that
heart surgery and relied on percutaneous ation procedure invasive and associated with did not cause haemolysis or liberate par-
air hoses similar to those used in the initial early morbidity and mortality. The greatest ticulate emboli, and that rapidly spinning
TAH systems. Development of a pneumatic shortcoming, however, was probably limited rotors in the blood could be supported by
pump for this intermediate-term applica- durability, with few pumps lasting beyond bearings, if the bearings were positioned so
tion was cost-effective and easily attainable. 24 months. that they were continually washed by pump

NATURE REVIEWS | CARDIOLOGY VOLUME 12 | OCTOBER 2015 | 613


© 2015 Macmillan Publishers Limited. All rights reserved
PERSPECTIVES

a b c in the first decade of the 21st century,39 the


concept of using this technology for TAHs
gained traction.

Dual rotary LVADs as TAHs


Researchers at the Texas Heart Institute
began to investigate leveraging the prom-
ising aspects of continuous-flow LVADs
in TAH design. In 2005, they carried out
Figure 6 | The future of total artificial heart technology. a | Dual HeartMate
NatureII® left ventricular
Reviews | Cardiology a series of experiments in which the heart
assist devices (Thoratec Corporation, USA). Image © Texas Heart Institute. b | BiVACOR total
of a Corriente-cross calf was excised and
artificial heart. Image © Texas Heart Institute. c | CARMAT®total artificial heart. Permission
obtained from CARMAT, France.
replaced with two continuous-flow LVADs,
thereby creating a continuous-flow TAH.40,41
Over 8 years, a wide variety of LVAD pumps,
flow. The early clinical experience with for >8 years.35 Results with the HeartWare® including the HeartAssist 5®, HeartWare®
the Jarvik 2000® also demonstrated con- HVAD have been similarly encouraging. HVAD, HeartMate II®, HeartMate III®,
vincingly that rotary blood pumps could Therefore, continuous-flow LVADs are seen and Jarvik FlowMaker, were implanted in
be made much smaller, quieter, and more as important tools in the management of 65 calves.42 Two calves survived to 90 days,
durable than volume-displacemen­t pumps. end-stage left ventricular HF. and 29 lived for ≥1 week. Eight calves could
One of the first recipients of this device sur- All rotary blood pumps have some degree exercise on treadmills. Common reasons
vived for >7 years with the pump and even- of inflow and outflow pressure sensitivity.36 for premature termination of the experi-
tually died from complications arising from As a result, when continuous-flow LVAD ment included infection in the device and
a reaction to transfusion.33 Post-mortem inflow pressure increases, as occurs with thrombus involving the bearings of the right
assessment showed almost no wear to each cardiac systole, instantaneous flow pump. Use of dual continuous-flow LVADs
the bearings. through the pump also increases. The in this configuration required the fabri-
In the 1990s, most teams working on pump flow rate, therefore, varies with each cation of customized atrial cuffs to allow
mechanical circulatory assistance concen- ventricular contraction, even when the the inflow cannulas to be attached to the
trated on the development of continuous- device is operated at a constant speed. Even atrial remnants. When dual centrifugal
flow LVADs, including the Jarvik FlowMaker when all cardiac output is going through the pumps were implanted, modified LVAD
(Jarvik Heart), the HeartMate II®(Thoratec pump and the aortic valve remains closed sewing rings were used as cuffs. When dual
Corporation), the HeartWare® HVAD throughout the cycle, variation occurs in HeartMate II®(Figure 6a) or HeartAssist 5®
(HeartWare, USA; Figure 5c), and the arterial pressure. Many patients supported pumps were implanted, however, the devices
HeartAssist 5®(ReliantHeart, USA; formerly by continuous-flow LVADs have marked were modified by removing the inflow and
MicroMed, USA). Each device is unique, attenuation of pulse pressure, which is outflow cannulas and having customized
but they share similarities. All devices use associated with new physiological chal- cuffs fitted that would allow the cannulas to
a rapidly spinning impeller to add energy lenges, such as gastrointestinal haemor- be implanted in an acceptable configuration.
to the blood, resulting in pressure and rhage (perhaps resulting from destruction The pressure sensitivity intrinsic to all
flow. Several devices use an axially oriented of von Willebrand factor) and progressive rotary blood pumps was hypothesized to
Archimedes’ screw supported at each end aortic valve insufficiency. Autonomic regu­ be advantageous to TAH design because
by blood-washed bearings and are classi- lation of afterload and control of blood it would provide autonomous balance
fied as axial-flow pumps. Powerful rare- pressure are challenging in some patients, between the systemic and pulmonary sides
earth magnets integrated into the screw-like perhaps because of an alteration in baro­ of the device. Maintaining this balance was a
impeller cause it to rotate in response to receptor function. Nevertheless, because of design challenge with positive-displacement
a microprocessor-commutated electro­ the small size, energy efficiency, and dramat- TAHs, frequently requiring sensors, compli-
magnetic field produced by the stator sur- ically improved durability of c­ontinuous- ance chambers, and modulation of stroke
rounding the screw. Rotation of the impeller flow LVADs, these devices have changed the volume from beat to beat. By contrast, with
at 8,000–15,000 rpm results in flow parallel field of cardiac support. a c­ontinuous-flow TAH, if a physiologi-
to the axis of rotation. Other devices use cal change resulted in hypofunction of the
a magnetically or hydrodynamically sus- The future of TAH technology right pump, such as is encountered in tran-
pended spinning disc-like structure with fins Rotary TAHs sient pulmonary hypertension seen with
or grooves, and are referred to as ce­ntrifugal- Contrary to the belief that a physiologi- intense coughing, the corresponding gradual
flow pumps. They have axisymmetrical cal pulse is essential for TAH application, increase in right atrial pressure would
inflow, but the outflow arises tangentially Golding and colleagues showed in the 1980s autonomously increase the right pump
from the circumference of the spinning disc. that mammalian physiology could be sup- output without any change in pump speed.
The first human implantation of the ported long-term by nonpulsatile flow.37 The This effect was proven in numerous animal
HeartMate II®LVAD was performed in Nov­ first concepts for implantable rotary TAHs studies in which right-side pump speed alone
em­ber 2003 by Frazier;34 since then, >19,000 were introduced shortly afterwards.38 After was increased, resulting in increased left
HeartMate II®pumps have been implanted multiple attempts, coupled with the ascend- atrial pressure and moderately, but notice-
in >185 countries. Overall, 1‑year survival ance of rotary blood pump technology as a ably, increased left pump flow.43 In addition,
is >80%, and several patients have survived form of mechanical circulatory assistance the total continuous-flow TAH output would

614 | OCTOBER 2015 | VOLUME 12  www.nature.com/nrcardio


© 2015 Macmillan Publishers Limited. All rights reserved
PERSPECTIVES

increase autonomously during exercise by continuous-flow TAH, the patient devel- SmartHeart
means of a similar mechanism.43 oped haemoptysis, and a biopsy revealed The SmartHeart TAH, being developed by
Different pumps showed different degrees amyloid infiltration of the lungs. Hepatic Cleveland Heart and Cleveland Clinic, has
of pressure sensitivity because of differing encephalopathy progressed and support was a cylindrical rotating element supported
impeller designs. A hypothesis was formed discontinued after 5 weeks.46 Although no by a fluid-film hydrodynamic bearing.51,52
that an impeller could be designed spe- successful long-term outcome was seen in The length of the cylinder is roughly twice
cifically to optimize this functionality and, this case, it demonstrated for the first time its diameter, and the cylinder is capped at
therefore, improve autonomous p­ulmonary– that a continuous-flow TAH could be used both ends by centrifugal flow impellers.
systemic balance.44 Nevertheless, to accom- in a human and suggested that assessment When the systemic–pulmonary circulation
modate the somewhat reduced pressure was warranted. becomes imbalanced because of a transient
sensitivity compared with that of the natural Pirk and colleagues performed a similar perturbation in physiology, the hypofunc-
heart, slightly elevated filling pressures operation on a patient (aged 38 years) with tioning side of the TAH is exposed to pro-
would be required to keep the atria from infiltrating cardiac fibrosarcoma.47 The pro- gressively increasing atrial pressure, which
transient collapse. The introduction of feed- cedure was initially successful, but the patient creates a hydrodynamic force on the spin-
back control of pump speed was considered died from a fulminant Aspergillus spp. infec- ning element and shifts the axis of rotation
as a means of mitigating this problem and tion after 6.5 months.48 Other groups have towards the side with lower atrial pressure.
improving the capacity of these pumps to championed the use of dual HeartWare® Each impeller is positioned in such a way
balance flows under all conditions.45 HVADs, either as biventricular assist devices with respect to the volute in which it is spin-
The thrombotic complications that occa- or as cardiac replacement devices after exci- ning that alignment is improved or wors-
sionally occurred in the right pump were sion of the ventricles.49 The clinical expe- ened depending on whether the rotor shifts
thought to be caused by venous thrombi rience with continuous-flow pumps for to the left or right. This change adjusts the
that were liberated from peripheral veins biventricular replacement suggests a possible relative outputs of the left and right sides
and became entangled in the pump mecha- role for specialized c­ontinuous-flow TAHs of the device and continuously balances
nism. This hypothesis was supported by that leverage the small size, mechanistic pumping efficiency without the need for
the observation that the left pump was simplicity, and improved durability of rotary sensors or feedback algorithms. Chronic
unaffected, presumably because it pumped blood pumps. testing of the device in animals is ongoing,
blood filtered by the lung microcirculation. but 6‑week survival has been reported in
Modulation of the right pump speed—by Single-device rotary TAHs one experiment.53
transiently decreasing speed by as much In 1987, Qian and colleagues attempted to
as 30% for 0.20 s every 2 s—reduced right create a single-device rotary TAH inten­ BiVACOR
pump complications, perhaps by improv- ded for long-term cardiac replacement, The BiVACOR rotary TAH (Figure 6b)50
ing pump washout, decreasing the areas of but could not adequately balance the left originated in Australia and is being devel-
stasis, and facilitating the passage of small and right circulation. 38 How­e ver, after oped by BiVACOR and a multinational
thromboemboli arising in the periphery the success of supporting patients with consortium of research centres. The device
through the right pump. HF using rotary blood pumps, BiVACOR uses a different mechanism from the
In 2012, a patient (aged 55 years) with (Houston, TX, USA)50 and Cleveland Heart SmartHeart TAH to achieve pulmonary–
systemic amyloidosis involving the heart, (Cleveland, OH, USA)51,52 each developed systemic balance. The rotating element in
liver, and kidneys was placed on para­ a specialized rotary TAH with intrinsic the BiVACOR TAH is more of a disc than
corporeal left-heart bypass. The patient flow-balancing mechanisms. These TAHs a cylinder, with its diameter being consid-
was deemed too ill to survive a heart trans- contain single rotating elements with sys- erably greater than its length. The spinning
plantation. Cardiac amyloidosis is generally temic and pulmonary impellers on oppos- element is suspended by an active magnetic
a contraindication for the SynCardia tem- ing faces. The rotational speed of the left and bearing system, in which stable levitation is
porary TAH because of technical issues, right impellers is the same, and the differ- maintained by variations in magnetic forces
and the patient’s ventricles were too small ent flow characteristics of the systemic and generated by electromagnets in response
to accommodate an LVAD inflow cannula. pulmonary circulation are achieved with to position feedback sensors.54 As in the
Frazier and Cohn suggested implant- differences in impeller size and geometry. SmartHeart, small shifts in the position
ing two HeartMate II® devices to create Similarly, small changes in the position of of the rotating element along the axis of
a continuous-flow TAH.46 Custom cuffs the rotating element along the axis of rota- rotation substantially change the relative
were fashioned from polypropylene hernia tion result in a substantial change in relative pumping efficiencies of the two pumps,
mesh, Dacron®cardiovascular patches, and efficiency of the two opposing pumps and while the rotor position in the BiVACOR
medical silicone. The inflow cannulas of the provide autonomous systemic–pulmonar­y device is determined by the differential
two HeartMate II®LVADs were removed, flow balance. Although the mechanisms pressures (forces) acting on the rotor. Spe­
and the pumps were attached to the cuffs for flow balance differ between the two cifically, the surrounding pressures cause
by a previously developed technique. 46 devices, they both offer the advantages of the magnetic system to pull the rotor auto-
The implantation was uneventful, and the centrifugal pumps. Specifically, neither has matically towards the underperforming side,
patient was extubated on day 2 after surgery any mechanical bearings or other sources which decreases the efficiency of the oppo-
in a haemodynamically stable condition. of mechanical wear and no flexible com- site pump so that the hydrodynamic forces
The patient, however, remained anuric on ponents or valves, and both have only one that arise from differences in the left and
haemodialysis and liver failure continued to moving part and are extremely compact and right atrial pressure are precisely counter­
progress. After 4 weeks of support with the energy efficient. balanced. 55 During in vitro testing, the

NATURE REVIEWS | CARDIOLOGY VOLUME 12 | OCTOBER 2015 | 615


© 2015 Macmillan Publishers Limited. All rights reserved
PERSPECTIVES

device demonstrated a maximum capacity of four pericardial tissue valves; the blood- Future challenges include gaining addi-
16 l/min at 100 mmHg and, when operated contactin­g surfaces of the blood sacs are lined tional insights into the effects of reduced or
with speed modulation, can produce flow with a microporous biocompatible material absent pulsatile blood flow. Devices using
pulses of 0–18 l/min and pulse pressures of aimed at obviating the need for anticoagula- positive-displacement technology continue
120/80 mmHg at 60 bpm. tion. In addition, multiple sensors embedded to be developed and used clinically. However,
An early series of nine acute ovine studies in the device provide autonomous regulation given their limitations in terms of durabil-
demonstrated the hydraulic performance of the pump rate and output in response to ity, they are unlikely ever to be considered a
and balancing capacity of the single-rotor activity level and physiological factors. The true alternative to transplantation. Instead,
concept.56 Later chronic studies in calves microprocessor that controls the device is initial successful experiments leveraging two
involved seven implantations, with the integrated into the pump housing, where it LVADs as a TAH suggest that a rotary TAH
longest duration being 30 days in a 71 kg is cooled by silicone oil hydraulic fluid. The has potential. The reduced size, improved
calf. The outflow for this calf varied from device is quite large and heavy (1 kg), but durability, and reduced power consumption
7 l/min to 15 l/min and averaged 12 l/min at computer-aided anatomical fitting studies inherent to rotary blood pumps might lead
100 mmHg at a motor power consumption of have predicted that it will fit 85% of men and to rotary TAHs succeeding where previous
13 W. The animal could exercise on a tread- 65% of all patients. A percutaneous power attempts have failed to provide a p­ractical
mill for 30 min, and autonomously balanced lead enters through the posterior scalp and is r­eplacement for the human failing heart.
pulmonary–systemic flow was maintained affixed to a skull-mounted pedestal, as previ-
despite changes in posture and activity, gen- ously described by Jarvik,58 in an attempt to Cardiovascular Surgical Research Laboratory
and the Center for Cardiac Support, Texas
erating flows of 16 l/min without evidence decrease the incidence of driveline infection.
Heart Institute, MC 2‑114A, PO Box 20345,
of haemolysis. A CARMAT®TAH was implanted into Houston, TX 77225, USA (W.E.C., D.L.T., O.H.F.).
a patient aged 75 years, who was supported Correspondence to: W.E.C.
Positive-displacement TAHs and haemodynamically stable for 74 days, wcohn@texasheart.org
Despite the potential benefits of c­ontinuous- at which point the device suddenly stopped.
1. Go, A. S. et al. Heart disease and stroke
flow TAHs, whether the physiological After an in-depth analysis and subsequent
statistics—2014 update: a report from the
limitations imposed by nonpulsatile or regulatory approval, a second implantation American Heart Association. Circulation 129,
reduced-pulsatile perfusion will prove to has been performed, and device support e28–e292 (2014).
be prohibitive remains unknown. For this was ongoing at the time of writing. 2. Colvin-Adams, M. et al. OPTN/SRTR 2012
annual data report: heart. Am. J. Transplant. 14
reason, SynCardia is developing a positive- (Suppl. 1), 113–138 (2014).
displacement TAH with a 50 cm3 blood- ReinHeart TAH 3. Akutsu, T. & Kolff, W. J. Permanent substitutes
pumping chamber to enable use in a wider ReinHeart (Bad Oeynhausen, Germany) for valves and hearts. ASAIO J. 4, 230–234
(1958).
range of patients. Additionally, two European and the Helmholtz Institute, RWTH-Aachen 4. Cooley, D. A. et al. Orthotopic cardiac
groups have readdressed the challenges in University (Aachen, Germany) are develop- prosthesis for two-staged cardiac replacement.
developing a self-contained, totally implant- ing the ReinHeart TAH. This device has a Am. J. Cardiol. 24, 723–730 (1969).
able, durable volume-displacement TAH by reciprocating piston to alternately com- 5. Frazier, O. H., Akutsu, T. & Cooley, D. A. Total
artificial heart (TAH) utilization in man. ASAIO J.
applying new technology. press two polyurethane sacs as the left and 28, 534–538 (1982).
right ventricles. In an attempt to improve 6. Kolff, W. J. Artificial organs—forty years and
SynCardia durability, a voice-coil and linear actuator beyond. Trans. Am. Soc. Artif. Intern. Organs 29,
6–24 (1983).
Owing to limited use of the SynCardia combination is used instead of a screw and 7. DeVries, W. C. et al. Clinical use of the total
temporary TAH device in paediatric and roller bearings. The stroke volume of the artificial heart. N. Engl. J. Med. 310, 273–278
young-adult patients (7% of 1,091 total ReinHeart TAH is only 50 ml and, there- (1984).
8. Leprince, P. et al. Bridge to transplantation with
implantations between 1985 and 2012), the fore, the size is 84 × 90 mm and the weight
the Jarvik‑7 (CardioWest) total artificial heart:
company has been developing a smaller is 923 g. 59,60 An implantable compliance a single-center 15-year experience. J. Heart Lung
version, called the SynCardia 50/50. 57 chamber allows passive filling of the ventric- Transplant. 22, 1296–1303 (2003).
The device is intended to be implanted in ular sacs, which provides a degree of inher- 9. Slepian, M. J. The SynCardia Temporary Total
Artificial Heart—evolving clinical role and future
patients with a body surface area <1.85 m2. ent passive flow balancing. A maximum status. US Cardiol. 8, 39–46 (2011).
In an attempt to meet the longer-term needs output of 7 l/min has been demonstrated in 10. Torregrossa, G. et al. Results with SynCardia
of patients with HF, the device has also been calves; the longest survival was 50 h.61 total artificial heart beyond 1 year. ASAIO J. 60,
626–634 (2014).
approved by the FDA for use as destina- 11. Copeland, J. G. et al. Cardiac replacement
tion therapy under the Humanitarian Use Conclusions with a total artificial heart as a bridge to
Device programme.10 Over the past 50 years, researchers have transplantation. N. Engl. J. Med. 351, 859–867
focused on developing a safe, durable, and (2004).
12. Friedline, K. & Hassinger, P. Total artificial heart
CARMAT practical TAH to meet the clinical demands freedom driver in a patient with end-stage
The CARMAT®TAH (CARMAT, France; created by end-stage HF. Early devices aimed biventricular heart failure. AANA J. 80, 105–112
Figure 6c), which is under development at replicating the natural physiological pulse (2012).
13. Johnson, K. E., Prieto, M., Joyce, L. D.,
by Carpentier and colleagues, uses two were large, heavy, and prone to mechani- Pritzker, M. & Emery, R. W. Summary of the
reciprocating rotary gear pumps to alter- cal failure, dampening enthusiasm for clinical use of the Symbion total artificial heart:
nately shuttle hydraulic fluid between two this technology. A shift in focus to LVADs a registry report. J. Heart Lung Transplant. 11,
103–116 (1992).
blood sacs with compressible diaphragms. and the success with rotary blood pumps
14. Massiello, A. et al. The Cleveland Clinic-Nimbus
Although superficially similar to the renewed interest and paved the way for total artificial heart. J. Thorac. Cardiovasc. Surg.
AbioCor®TAH, the CARMAT®TAH has new approaches to total heart replacement. 108, 412–419 (1994).

616 | OCTOBER 2015 | VOLUME 12  www.nature.com/nrcardio


© 2015 Macmillan Publishers Limited. All rights reserved
PERSPECTIVES

15. Tatsumi, E. et al. In vivo evaluation of the National 33. Frazier, O. H. & Leon, P. J. Small pumps for 48. Netuka, I. et al. Novel treatment of an infiltrating
Cardiovascular Center electrohydraulic total ventricular assistance: progress in mechanical cardiac fibrosarcoma. Tex. Heart Inst. J. 41,
artificial heart. Artif. Organs 23, 242–248 (1999). circulatory support. Cardiol. Clin. 25, 553–564 248–249 (2014).
16. Vasku, J. et al. 150-day survival of a calf with (2007). 49. Strueber, M., Schmitto, J. D., Kutschka, I.
a polymethylmethacrylate total artificial heart: 34. Frazier, O. H. et al. First clinical use of the & Haverich, A. Placement of 2 implantable
TNS‑BRNO‑II. Artif. Organs 5, 388–400 (1981). redesigned HeartMate II Left Ventricular Assist centrifugal pumps to serve as a total artificial
17. Vasku, J. & Urbanek, P. Constructional and System in the United States. Tex. Heart Inst. J. heart after cardiectomy. J. Thorac. Cardiovasc.
functional characteristics of recent total 31, 157–159 (2004). Surg. 143, 507–509 (2012).
artificial heart models TNS Brno VII, VIII, and IX. 35. Kirklin, J. K. et al. Interagency Registry for 50. Timms, D. et al. The BiVACOR rotary
Artif. Organs 19, 535–543 (1995). Mechanically Assisted Circulatory Support biventricular assist device: concept and in vitro
18. Rosenberg, G. et al. A roller screw drive for (INTERMACS) analysis of pump thrombosis in investigation. Artif. Organs 32, 816–819
implantable blood pumps. Trans. Am. Soc. Artif. the HeartMate II left ventricular assist device. (2008).
Intern. Organs 28, 123–126 (1982). J. Heart Lung Transplant. 33, 12–22 (2014). 51. Fukamachi, K. et al. An innovative, sensorless,
19. Snyder, A. J. et al. An electrically powered total 36. Salamonsen, R. F., Mason, D. G. & Ayre, P. J. pulsatile, continuous-flow total artificial heart:
artificial heart: over 1 year survival in the calf. Response of rotary blood pumps to changes device design and initial in-vitro study. J. Heart
ASAIO J. 38, M707–M712 (1992). in preload and afterload at a fixed speed Lung Transplant. 29, 13–20 (2010).
20. Weiss, W. J. et al. Steady state hemodynamic setting are unphysiological when compared 52. Fumoto, H. et al. In vivo acute performance of
and energetic characterization of the Penn with the natural heart. Artif. Organs 35, E47–E53 the Cleveland Clinic self-regulating, continuous-
State/3M Health Care Total Artificial Heart. (2011). flow total artificial heart. J. Heart Lung Transplant.
ASAIO J. 45, 189–193 (1999). 37. Golding, L. R. et al. Chronic nonpulsatile blood 29, 21–26 (2010).
21. Snyder, A. J. et al. In vivo testing of a completely flow in an alive, awake animal 34-day survival. 53. Karimov, J. H. et al. In vivo evaluation of the
implanted total artificial heart system. ASAIO J. Trans. Am. Soc. Artif. Intern. Organs 26, 251–255 Cleveland Clinic continuous-flow total artificial
39, M177–M184 (1993). (1980). heart in calves [abstract]. J. Heart Lung
22. Harasaki, H. et al. Progress in Cleveland 38. Qian, K. X., Pi, K. D., Wang, Y. P. & Zhao, M. J. Transplant. 33 (Suppl.), S165–S166 (2014).
Clinic‑Nimbus total artificial heart development. Toward an implantable impeller total heart. 54. Greatrex, N., Timms, D., Kurita, N., Palmer, E.
ASAIO J. 40, M494–M498 (1994). ASAIO Trans. 33, 704–707 (1987). & Masuzawa, T. Axial magnetic bearing
23. Kung, R. T. V. et al. Progress in the development 39. Slaughter, M. S. et al. Advanced heart failure development for the BiVACOR Rotary BiVAD/
of the ABIOMED Total Artificial Heart. ASAIO J. treated with continuous-flow left ventricular TAH. IEEE Trans. Biomed. Eng. 57, 714–721
41, M245–M248 (1995). assist device. N. Engl. J. Med. 361, 2241–2251 (2010).
24. Kung, R. T. V., Yu, L.‑S., Ochs, B. & Frazier, O. H. (2009). 55. Timms, D., Kurita, N., Greatrex, N.
An atrial hydraulic shunt in a total artificial 40. Frazier, O. H., Tuzun, E., Cohn, W., Tamez, D. & Masuzawa, T. BiVACOR—a magnetically
heart: a balance mechanism for the bronchial & Kadipasaoglu, K. A. Total heart replacement levitated biventricular artificial heart [abstract].
shunt. ASAIO J. 39, M213–M217 (1993). with dual centrifugal ventricular assist devices. Presented at the 20th MAGDA Conference.
25. Dowling, R. D. et al. Initial experience with the ASAIO J. 51, 224–229 (2005). 56. Timms, D. et al. Initial acute in-vivo animal
AbioCor implantable replacement heart system. 41. Frazier, O. H., Tuzun, E., Cohn, W. E., Conger, J. L. experience with the BiVACOR Rotary Bi-
J. Thorac. Cardiovasc. Surg. 127, 131–141 & Kadipasaoglu, K. A. Total heart replacement Ventricular Assist Device [abstract].
(2004). using dual intracorporeal continuous-flow Artif. Organs 31, A67 (2007).
26. Nose, Y. FDA approval of totally implantable pumps in a chronic bovine model: a feasibility 57. Ryan, T. D., Jefferies, J. L., Zafar, F., Lorts, A.
permanent total artificial heart for humanitarian study. ASAIO J. 52, 145–149 (2006). & Morales, D. L. S. The evolving role of the total
use. Artif. Organs 31, 1–3 (2007). 42. Cohn, W. E. et al. Eight-year experience with a artificial heart in the management of end-stage
27. Frazier, O. H. et al. Multicenter clinical continuous-flow total artificial heart in calves. congenital heart disease and adolescents.
evaluation of the HeartMate 1000 IP left ASAIO J. 60, 25–30 (2014). ASAIO J. 61, 8–14 (2015).
ventricular assist device. Ann. Thorac. Surg. 53, 43. Frazier, O. H., Cohn, W. E., Tuzun, E., 58. Jarvik, R., Westaby, S., Katsumata, T., Pigott, D.
1080–1090 (1992). Winkler, J. A. & Gregoric, I. D. Continuous-flow & Evans, R. D. LVAD power delivery:
28. Frazier, O. H. Chronic left ventricular support total artificial heart supports long-term survival a percutaneous approach to avoid infection.
with a vented electric assist device. of a calf. Tex. Heart Inst. J. 36, 568–574 (2009). Ann. Thorac. Surg. 65, 470–473 (1998).
Ann. Thorac. Surg. 55, 273–275 (1993). 44. Frazier, O. H., Khalil, H. A., Benkowski, R. J. 59. Fritschi, A. et al. Image based evaluation of
29. Rose, E. et al. The REMATCH trial: rationale, & Cohn, W. E. Optimization of axial-pump mediastinal constraints for the development
design and end points: Randomized Evaluation pressure sensitivity for a continuous-flow of a pulsatile total artificial heart. BioMed. Eng.
of Mechanical Assistance for the Treatment of total artificial heart. J. Heart Lung Transplant. OnLine 12, 81 (2013).
Congestive Heart Failure. Ann. Thorac. Surg. 67, 29, 687–691 (2010). 60. Pohlmann, A. et al. Experimental validation
723–730 (1999). 45. Khalil, H. A. et al. Continuous flow total artificial of the linear drive train for a total artificial
30. Frazier, O. H. et al. Initial clinical experience heart: modeling and feedback control in a heart system. Mechatronics 23, 222–226
with the Jarvik 2000 implantable axial-flow mock circulatory system. ASAIO J. 54, 249–255 (2013).
left ventricular assist system. Circulation 105, (2008). 61. Laumen, M. et al. A novel total artificial heart
2855–2860 (2002). 46. Frazier, O. H. & Cohn, W. Continuous-flow for destination therapy: in-vitro and in-vivo
31. Wampler, R. K., Moise, J. C., Frazier, O. H. total heart replacement device implanted in study. Biomed. Tech. http://dx.doi.org/
& Olsen, D. B. In vivo evaluation of a a 55-year-old man with end-stage heart failure 10.1515/bmt-2013-4373.
peripheral vascular access axial flow blood and severe amyloidosis. Tex. Heart Inst. J. 39,
pump. ASAIO Trans. 34, 450–454 (1988). 542–546 (2012). Author contributions
32. Frazier, O. H. et al. First human use of the 47. Pirk, J. et al. Total artificial heart support with W.E.C. and D.L.T. researched data for the article.
Hemopump, a catheter-mounted ventricular two continuous-flow ventricular assist devices All the authors discussed the content of the
assist device. Ann. Thorac. Surg. 49, 299–304 in a patient with an infiltrating cardiac sarcoma. article, wrote the manuscript, and reviewed drafts
(1990). ASAIO J. 59, 178–180 (2013). before submission.

NATURE REVIEWS | CARDIOLOGY VOLUME 12 | OCTOBER 2015 | 617


© 2015 Macmillan Publishers Limited. All rights reserved

You might also like