No DESCRIPTION Unit Amount

10. BIOLOGICAL SAFETY CABINET CLASS II TYPE B2:


Type B2 cabinets are sometimes referred to as “total exhaust”. This type of unit.
Maintains minimum average inflow velocity of 100 fpm through the work area access
openings.
 Has HEPA-filtered down flow air drawn from the laboratory or outside air (i.e.,
down flow air is not recirculated from the cabinet exhaust air).
 Exhaust all inflow and down flow air to the atmosphere after filtration through
HEPA filter without recirculation in the cabinet or return to the laboratory room
air.
 Has all contaminated ducts and plenums under negative pressure, or surrounded
by directly exhausted negative pressure ducts and plenums (not recirculated
through the work area).
 Type B2 cabinets may be used with biological agents treated with toxic chemicals
and radio nuclides required as an adjunct to microbiological studies.

STANDARD FEATURES

 Pre-Filter Assemblage for Multi-Dispose particle predominantly in the size range of 5-


20 microns
 HEPA Filter 0.3 microns
 Airflow: Designed for 0.45m/s to 0.65 m/s when measure across the entire filters space
during the rated life cycle of filters.
 Cleanliness level – class 100
 Fluorescent lamp
 Flow Ratio: Class II Type B2 : 100% Exhaust
 UV- light
 Front Door: 8MM thickness toughened glass, vertically sliding, balanced with counter
weight.
 Noise level : 65 +/- 5 db
 Vibration level : less than 2.5 um (0.0001)
 Power supply : 230 V Ac
 Virus tight extract HEPA filter
 Stainless steel work table
 Gas cock
 Extra power outlet switch
 Color half white
EVERFLOW SCIENTIFIC INSTRUMENTS Continue Sheet .15.

S.No DESCRIPTION Unit Amount


TECHNICAL FEATURES:

AIR FLOW:
 Designed for 0.45m/s to 0.65 m/s.

PRE-FILTERS:
Size : 900 x 150 x 50 mm
Type : Flange type
Media : Synthetic, Non-oven Polyester fiber, Washable type
Casing : M.S. Powder coated
Gasket : Neoprene
Retention :10 Micron
Efficiency : 90%
Pressure drop: 6 to 8 mm

SUPPLY HEPA FILTER:


Size : 1220 x 457 x 75 mm
Type :Box type, Minipleat
Media :Ultra clean glass fiber paper-imported
Casing :M.S. Powder coated
Gasket :Neoprene
Retention :0.1 Micron
Efficiency :99.99%
Pressure drop:25 mm of W.c.

EXHAUST HEPA FILTER:


Size : 1220 x 150 x 305 mm
Type : Box type, Minipleat
Media : Ultra clean glass fiber paper-imported
Casing : M.S. Powder coated
Gasket : Neoprene
Retention : 0.1 Micron
Efficiency : 99.99%
Pressure drop: 25 mm of W.c.

IMPELLERS:
High performance noise abated statically and dynamically balanced.
EVERFLOW SCIENTIFIC INSTRUMENTS Continue Sheet .16.

S.No DESCRIPTION Unit Amount


MOTOR –BLOWERS:-
Volts : 230 V
Hz : 50
HP : 1/3
Amps : 3.5
Watts : 250
RPM : 1440
Qty : 1 no for supply and 1 no for Exhaust

CLEANLINESS LEVEL: Class 100


ILLUMINATION : 40 watts Fluorescent tubes
NOISE LEVEL : 65 +/- 5 db
VIBRATION LEVEL : less than 2.5 um (0.0001)
POWER SUPPLY : 230v AC, 50Hz

OFFERED AS UNDER
 Cabinet made of GI powder coated with inner skin
Stainless Steel 304
Inner Dimension: 1220 x 610 x 610 mm (W x D x H)
Overall Dimension: 1350 x 850 x 2300 mm (W x D x H)