Professional Documents
Culture Documents
BS EN ISO 10079-3 2014 Medical Suction Equipment
BS EN ISO 10079-3 2014 Medical Suction Equipment
BS EN ISO 10079-3 2014 Medical Suction Equipment
'.:
Medi ment
Suctio
..;
{:
* Part 3: wered from a
vacuum or po as source (lSO
10079-3:2014)
f,i
{.
'9.
lr
R-+io7p61".gl .l|*#}If)
{
t* R-e** Nswo4$
,.
t;
i;.
:>"
3
r
)
t::
t:
\
rh I r i,
{' ...making excellence a na Dt[;
=:
"ii
lii '
.:1:
:.1:;
EUROPEANI STANDARD EN rSO 10079-3
i',
ii
NORME EUROPEENNE
L:.
'..:
EUROPAISCHE NORM May2074
t,
lcs 11.040.10 Supersedes EN lS@ trO079€:2.#S!
English Vension
AbsawgefAte (l
pression (lSO I 0079-3:2014)
. .']
::'
is
'/^
;):
IY
@ S014 CEN All rights of exploitation in any form ard by any .neans reseri,ed Ref, No. EH ISO l00n&gr2s1#fi
uorldwide for CEN national Members.
BS EN ISO {0029€:2di4
EN lSO r0029-3:2014 (E)
Foreword
(
This document
)has rnittee tso/Tc 121
and respiratory borati vb' r' r v -2lS
SEN/TC I! "Res
anaesthetb equi of wh {
..'
This European standlard has been prepared under a mandate given to cEN
and the European Frce Trade Association to provide a means of confonn
Directive 93l42lEEC.
once this standard is cited in the official Joumal of the European union u- that Directive and has
implerner*ed as a national standard in at least one Member State, the normative c{auses
this standard given in Table ZA.1 confers, within the lirnits of the standard, a presumption
conformity with the relevant Essential Requirements of that Directive.
NOTE lMen an Essential Requirement does not appear in TabtezA.l, ii stat it is not addrssed by
European Standard.
4.1,4.4,12t) covered
are no requirements for rn:rterials apart from
to perforna a dsk assessrnent and
disclose the presence of latex.
As these devices are only for errtracting body
toxicity and biological cornpatib,ililty is not
a risk.
rt"z f$
13"3 0
12 b) 1$.4
12.
13.6a) q, fuilarS,W
13.3 a)" b),
d)" o'g)' h)' j)'
ff,!|'
,l2w
13.6 W
13.64
12 b)" c$. r|}, h), j), v) 13.6 4
12 i| 13.6 h9
t24 13.6 i}
'r24 rs.6 q!
k
*^.*^ll9$_:1"_01*".tnrsand
trte smpe,of frlis
onrsB.t mqybe produr*(s! fdSWS
standard.
&
ii
;',' : ";
Contents
Foneword...
t Scope a
2 Normative references
3 Termsand <lefinitions
4
4.L Riskrnanagement..,.........
4.2 Usability......
10
11
11. of symbols
[L.2
ffi{ssHB!#-}w
BS EN l5O 10079-3:201$:
Iso 10o79-3:20la(E|
Foreword
For an explanation on
_the meaning of ISO sp expressions related to conform
assessment, as well as information
4boutlSO,s adl principles in the Technical Ba
to Trade [TBT) see thc'following U$L: Foreword
information
The committee responsible for this docu
L2l, Anaesthetic and respiratory
Subcommittee SC B, Snction devicesfor hos
cane ase.
This third edition cancels and replacBs the
revised. 0S0 10CI79-3:1999), which has been techn
IS0 10079 consists of the followi the general Eide Medical suction equipment:
Part 1 : Electrically, pow@d
Annex B contains statemeqts for some ofthe requirernents of this part oflso 10079. The clau
and subclauses with an astenisk {*) after their number have correiponding rationale conta
in Anne- B. i addl
recom
uirements qfill not only facilitate the properapplication ofthis partof ISo
but t re\tisions.
fi
,{r
;,$
Part 3:
q^-
Suction
Crrnlinn
equipment
lrr.,L,llrr|' cy
r, powere
ur l,rltgtl'L pl,,
^^rt!--as* wtul. s
pressure ga$ source
L Scope s
This part of ISO 10079 specifies safety and performance requ ical suction equiprnerg
powered frorn a vacuum or positive pressure gas soufioe nffri suction. tt apeffus tci
equipment connected to medical gas pipeline sJrstems or crilmd and uri attachments.
:ll'.^G-^+^- .
illustrates rL^
tuurLrdLcb rL-^^
Lue three
the parts
paf
L[r.ee ---!- -f t^^
rs oI
of
lAAd^
IJU IUUr'y
ISO L00Tgby
I
Dy provlol
providi
The equipment can be stand-alone or part ofan
not
This part of IS0 L0071, does not a
,t!
e) s scaven4_in
tion; @
$
o fi
t
;
;]
k) breastp I
I
l)l
m)u i;
n) plume tion, systems.
2 Normative references
The following documents, in whole or in part, are nor.mfit{ydy d in ttais docurment and atqr
indispensable for its application. For dated referenoe$i, wlty t*ue esiEion cited applies. Fbn undateS
references, the latest r-.dition of the referenced doeurnent ug any arnendrnents) applies,
,,
td
BS EN fSO 10079-3:2014
ISo 10O79-3:zO1a@) 1!'
-
ISO 5356-1, /4 naesthetic and respiratory equipment Conical cownectsts Part 7: Canes and' sockffi
- - t
IS0 15223-L , Medical device:s Symbols to be used with medical device hahels"
be supplied Part 7: General-requirements
-
ISO 80369 (all parts), Small'-bore connectors for liquids and gases fn
EC 6 0 60 1-1 : 2 0 0 5 +A L2, Me d[cal ctrical equipment Faft n : forsafery
I 1:2 0 e Ie
-
IEC 60601-1-6, Medical elect:rical equipment- PartT-6: Generul sofety and
p e rformance- Collateral stun dard : U sability
IEC 6L672-L, Electroacoustir:s Sound level meters Part L:
- -
IEC 62366, Medical devices -- Application of usability
3.1
collection container
container in which liquids and solid p lected
2'' &'
ner qccarmhlw
nor asselmbly
er and its closu for suction
3.3
drainage
removal of liquid, solid particles or 6m body cavityonwound
3.4
end-piece
that part of the applied to the patient ryvhich laegins at the site where tu
drawn in and ends connection
Note 1to uspd end-pieces are a Yankausr sucker and a suctlon satheter.
3.5
exhaust port
opening through w xhaust gas is discharged
1l The n anhical symbol collections of trSO 7000, ISO 7001 and IS(} P&LS are aNso auailable on the Ool'ine
Platro rm-http ://wvvrv.iso.org,/obp.
WEfi$l{StS t, -3:20t{'i
ffi sdrsr.{-G&
3.6
field use i
use of suction eguiprnent in situations outside of the health rm sqfu nffin:,mff &fr
emergencres
3.7
filter
'device for retenfion o{'particulate matter
3.8
free airflowrate
rate of unrestricted fftrw of air through a designated inlet
3.9
highflowrate
free air flowrate of 20 l/min or rnone
3.10
high vacuum
vacuum level of50 kPa or rnore below atrnospheric pnessure
intermediate tubing
tubing between the colllection container and tt
3.13
intermittent vacrrunn
type of suction in which the negative pressu; tothe edrd@flmak' affid lll
returned to atrnospheric pressure
3.L4
ii
,l:
g
3,
T
{
;
3.16
medical gas pipeligr
;fystem l
complete system clh cornprises a supply systern, a riom ,and awd a
,j
tiCIfi
system with ter rnits at the points where medical gams,or trrE i
3.19
overfill protection device
device intended to preventliquid on solid particles from the
3.20
single fault condition
condition in which a single tneans for protectiora against a safety hazard in equipmentis r#tyfeu&itr* {m.s
single external abnormal condition is presewt
3.2L
suction
application ofvacuum to rernove liquid, solid parti
3.22
suction tubing
tubing for conduction of liquid, solid particles orgas
9.23
thoracic drainage
drainage of liquids and gas From the thoracic cavi
the patient
Note 1 to entry: For the purposes of this partof lS0 10S79, alt thoraci,cdrainqge is considered to be sfti1€.
3.24
transport use
use during patient transport outside ofa health cane
3.25
vacuumlevel
pressure less than atmospheric pressure
Note 1 to entry: In this part oflS0 10029, vacuunN a difrflerence frorn atrnospherk gtxrmae.
3.26
vacuum level indicator
device for displaying the vacuum I
3.27
vacuum source
component ofdevice for generati
3.28 ,o..
vacuum regulator
device for controlling the applied-
4 General
Suction equipment nts controlNed by electr,ical means, e"g" electronic timing, s&a[t wsr#t
the relevant reouir 60 50 1-1 :2005 +41-: 20 1 2"
4.L Risk
with suction bquipment po*,:red frorn a vacuurn or positive gas Jo*noe" R* blished riskrna
process shall be applied to the design of the device. The. risk nramagement process shall dap
following elements:'
risk analysis;
risk evaluation;
f,
risk control;
-
production and p ost-production infornnation.
lt
fl
EXAMPLE IS014971. tj
4.L.4 The manufacturer mayuse Brpe tests detailed within fhisparrofffi 1007
if an equivalent degrere of safety is obtained" methods shall be rralidated the
methods specified in r{nnexA of this part
4.2 Usabiliry
4.3 Clinicalinvesrtigil
Where
is claimed and in the risk management file. The clinical investigation shalfl mmn@ywdth
requirements of 55.
publish-ed and/or urnpriblished reports on other clinical experience of either the device in queshima ora
device for which equivalence to the device in question c"n be de*onstrated-
,.
@ ISO 2014 - All rights resewed
'u
lft
BS EN l5O 10079-3:2014
ISO 10079-3:20La(E)
Parts intended for re-use shall meet the requirements of Clauses Z af,Hr
those components have been submitted to 30 cycles of cleanireg and sterilization as
recommended by the manufacturer.
6 Design Fequirements
6.L Collectioncontainer
6.1.L General
6
O ISO 2014 - All rights reservnd
.,."1fr,'-':'^',1.'I;iii
6.2 Connections;
The connectors for the suction tubing and the intermediate tubing shall be designed to ."r#
assembly or clearly marked to indicate correct asse5nblylvhen all parts
".. -"I*d4 t
Check compliance b5r functional tesEingand inspection. i!
I
NOTE Incorrect connections have frequently been a cause ofspill orzer into rod n bssy$
suction. 'l
,.t
6.2.2 Inletport
The inside diameter of the suction tubing connector finlet port of
least 6mrn and the i,nside diameter of ttle suction tubing conneqg
Iarger than the inside diameter of the largest tubing size as
,,}
The inlet shall not be compatible with arqr conical connector s ISO S3S6-L or snmg-borE
-
connectors specified in ISO 80369 fall parts].
-$
S
by,functional
Check compliance
6.3.1 shal
Suction tubing meter of not less than G mm. j;
The degree of collapr;e of thfs hall be less than 0,5 throughout its entire i:
Check compliance by the using the tubing specified by the manufaswrer nf$rc c,uctiq[3
equlpment. '
9.3,2!*) Suction t lied or recommended by the manufacturer" shatt foalre a mim,iirnu$f
length of 1,3 m. 'l
NOTE Suctio Formance may be markedly affected by the length and diarneter sftAe lnetnreen
collection contain f,end-piece, See,Annex C.
6.4 va
Sucti an operator-adiustable vacuum regulator; shall have a rmeamo sf g
atmospheric pressure at the patient end when attached to a s&Eti@m r,sd$e{:er
6.4,L The full scale of analog vacuum level indicators shall be not more than 20$ 9{r of Sse mtamii
vacuum level below artmospheric pressune as specified bythe manufacturer.
6.4.2 Analog displays shall have gradr.lations not less than 2 rnm apartt each grad
not more than 5 %o of'the full-scale value.
6-4.3 Digital displays shall display vacuum level below aunospheric pressure
at intenrrals of ao*
than 5 of the full-scale value.
a/o
6.4.6 Low vacuum equipn:nt shall be fitted with a vacuum level indicaton
and collection container.
7 Operational requireme
7.2 Dismantling
Suction equipment in
dismantled by fo shall be desi
correct reassemblv indicate corre
accordance vre+r trii
y- antlingand
rer's instructions, tne sucuon equipment shail rnit the requirennenffi,ffi
Clause 9 as
7.3 Mechan
equipm_ent intended for field and
*::t:l
Deing dropped from a height
use shall enrs of Ctause g af&sr
o:[ 1 m onto or in the
If the suction equipment can be operated
equipment shall'be drop-tested -"Uoiu rrying case, individual parts of tfte sucbfim:
d. TG reassernbled suction equr.pment sh#l
meet the requirements given "r 9. as"
in Clause
Check compliance
by the tests given in AJ.
I
@ l5o 2014 - All rights reser*n4"nr
85 E$J l5O t0079.3:20tfi
ffiO X.0O79-3:21O1
7.4 Stability
Su mentintendel f-,:ldl"d/o1t menrsgirren [n
as e, when placed l"j
on a surface of (
'11:
If a device to lirnit the maxi# is fitted, the dlall nnt exceed the set vacuura lr*4
by more than 10 o/0.
!t
.!
Check compliance by*Un ,y
#
a ptxitfue pren$ure at dae patientend clf more
7.6
iX
7.7 Airleakage it
The maximum leakage into the collection container assernbly shail{ wcft 't&r*
collection container is intenrCed for use with suction equipment har,{nwa &.ee msre tha.s
1 l/min, the pressure increarse shall be less than 3,3 lipaTV in f,O q, me of Se
collection container in litres.
collection containers intended for re-use shall be tested after 3o and disinfectioneir
sterilization as recommende d by the manufacturer.
Check compliance by the tesits given in,A.9.1.
8.2 Mass
I
The mass of s ntended for field use, comple&e w,itle irs rarrytns case or frame awd
accessories,
kg.
I
NOTE
ofilen combined with resuscitation equiprmcmq r&ich may rnal<e l[
s
to define a equipment alone. In these circumstances thiisisum wary not,apply" uut au eq
intended for field usi
be as lightweight as possible.
Check compliance by functional
testing.
10
& nm jZ&'[+ - Alt nights.rcsurrm$
tsS ENl5O l@I$-3:Zstffi
ISg 10$79-fi:eflfa{Sg
I'l
I
,i-1
il
h
9.1 High vacuum/high flowrate equipnrent G
ix:
lli
;'f
f*ll.::tliancebyt,hetestgivenin.4.10,amhemaximumandminimumsupprypreosuree
by the manufacturer.
marked "thoracic drainage" shal! be adjustable to a static vacuum level p'f Z kpa
lfflo*S*. Such equipme^1t sfrall produce a free air flowrate of ar leasr t5 n n-.-
be
fl:'j::lffIt:"q1:ss,ur9. LJ and
A[t[t s
Sn
be capable ora"vetooing, within
4,5 I total capacity.
5 s, 95 % of t'he servacuum r;t i;h;;;-";.;;lfi;;^ s,/srem
ll:liol.uguipment intended for thoracic drainage Shall not develop a pressure in excess f
patient tr flrpa at
inlet with a freer air flowrate of 10 l/min.
Suction equipment for intermittent vacuum- shall prodrrce a vacuurn le$e[ e t& 16 of rhe vacrrurm
hry#
specified by the manufacturer or t 10 0/o of the rnid-range.level if ttm Mnu T&H
cycling fre.quency shall be 'ithin 10 0/o of the specified frequency or ltl th if&s
range is adjustable.
Suction equ
transportuseshallmeettherr,equirernents tmChusesfl
and I as ap to temperatures of -18 "C and +50 "ff"
Check complianbe b n inA.L7.Z.I and A_LT.2J,
10.2 Storage
11 Marking
Yitll*-t"d informationto
wnere appropriate symbols
be supplied by the manufacl.urer shall comp[y wirh EM amd
speiiried in ISo 7000 or Iso 1szz3-1.
as
12 @ffiCIp0$#-,swnffikf
4l
r1;
85 EN l5O 10079-3:2014,
ISO 1OO79-3:zO1a(E!
,+
f
The following information shall be permanently and legibly rnarked on the suction equiprnent,
or o$
parts of it, where aPPlicable:
an indication that the device, or parts thereof, are for single acturer's irdication of
use shall be consistent);
all equipmentgenerating suction shall be with words indicating suction. This marking
be visible in the normal working position;
')
k) an indication thatthe equip for pharyngeal suction onlY;
n) the maximum vacuum which the equipment is designed shall be marked prominently
the display immed adjacent to it;
either with the vacuum levr:l which can be attained or with wo ls indicating "low vacuum level";
s) if a progressive variation in the degree of vacuum level is available, the direction of adiustment tc
increase vacuum shall be clearly and permanently marked;
for suction equipment powered from a positive pressure gas source, which can be detactlbd frorm
---
the power source, the recommended maximum and minimum supplypr"rru*q
NOTE For medir:al g1s prpeline systems complying with ISo 7396, the positive press upplyisset
between 270kPa and550 kPa.
L4
@ ISO 2014 - All rights reserved
* ''1'.
Try'-rF'w-
-_. _,. -.:,;ffi:;.:_::;*. :4
i
85 EN fSO 10079-3:2014;
IS0 10079-3:20!.a(E$
n) details of the operation of arry overfill protection device fitted to olle container
and the usa.ble capacity of the collection container in all the r# ended inclined planes
operation;
NOTE In sorne cases, this rnay require servicing by turer or his authorized agenL
q) if applicable,, a statement thert suction ceases
r)
s)
Yacuum tevel;
w) fault-finding and
x) whether or not is suitable for use in an MRI environment;
y) any warnings and/or s to take;
AnnexA
[normative)
Test methods
A.l General
The apparatus and test methods specified in this annex
are not intended to
measuring devices or me^th"o^d_,s^which yierd resurtr ;;;;;;;cy"equat
to or sr* than specifid-
In case of dispute, the methods given in this parroilio
"f roozg ;ha, be the
If an inline ; .:o]!,Oi?l
mended,
container) in_ a protective encl'osu re, at20.C to 25 "C.
outlet port. A
attach the filter forihe test. Attach
tion container and accessories (if present) under
,ou.."u lo
","iur* ure
Lv qtt
manufactu test to Lzi % of tIrc
atmospheric TlIjTllT,"."cuum level or to a u"",r,r'n levei not ex.;;dl;;
is less. Hold the vacuum level for s min,
tng bels$ssiip,
procedure ana tri"n ."r-u"ll. Ii"p"rt tl*
possibte flyingdebris. be hazardous. Proper care should be taken
,rt#lt"l?:t-can to protect perconnel from,r
Key
L vacuum source
2 vacuumlevelindicator
3 protective enclosure (loose fitting,
not seal
4 closedto atrnosphere
5 collection container unde,rb$
Figure A.l
- for testing collection containen strength
@ IsO 2014
- All rights reserved
L'f,
ias rN lso 1m79-328T4
;tSO fOOZg-322&
o
o
a) Eefore test
Key
1 vacuumsourqe
2 vacuumlevelindicatm
3 collectioncontainer
4 suction tubing
5 plug
@mrpfl4-dhm[&h.
Dimensionsirn
I
x
alt
E
o
o
Key
I urce wit}r a flowna@ of 10l/min
2
3 suctiontube
4 droracicdrainagesystem
- Ilpical
Figure,A.3 apparatus fortesting
q ln thorac{cdnainage
@ ISO 2CI1{
- AH rirghts resened
1$
i eS eN FO 10079-3:20't4
!tso roozg-s:20!.a@)
ri
Dimensions in mil
Key
1 inlet
2 container (capacity 1 lto 2 l)
3 drivirg gas at manufacturer's recommended pressure
4 device under test
5 seal above exhaust port
6 exhaust enclosure, at least 100 mm clear ofventuri
7 onr/offvalve
8 outlet (atleast 100 % venturi exhaust cro
&
Figure A.4 Tlpical apparatus fo{
-
Noise test
Place the microphone of
-61672_11
specified in tEC
nd-level meter compl the s for a type I instrumemtr
rough the
position of maximum ssu horizontal plane passing
I suction equipment at a radius of 1 m" The rneasured sound pressure
lev, el sha[ not value.
For this tes
rent shall be operated
with the inlet port open to atrnosphere and with
the inlet
recom
its normal working range of florivrate, includingthe maxirnum flowrate
manufacturer. Measurements shall be taken using the frequency-weightdng
characteristic A
t time-weighting characteristic S on the sound-levefmeter. The rneasurernens
be taken in a free
over a reflecting plane as specified in ISO 3744.
background level of extraneous noise shall be atleast tr{} dB below thatmeasured dulrir4g
le$-weighted
:,
NOTE collection codainers wlll usually have a pneumatic compliance of mvlKlla per nttF
volume.
Key
1
2
3
30 kPa and 50 NrPa elory ricpressure
4
5 closed to atmosphere
6 test collection conEiner
U
FigureA.S
- Tlpical testing leakage of collection fuEgeneral use
j i Dirmensions in ml"llimeaes
t
$
1 rracuum source
vacuum regulator
v:rcuurn level indicatol accurat€ to 2,5 o/o maxirnum scale value
ony'offvalve
component or system under test
closed to atnosphere
suction tube (6 mm inside diarnete[ square cut)
waterbottle
Figure A.6 Typical apparatus for testing leak lectionr €o$tainer for thoracic drainage
-
'li-
R
rii
q
{,
ci
,t
i(
t(ey x
1 flow measuring device
2 connecting piece (inside diameter 10 mrn to 20 rnrn and less tfoan I fth)
3 ctramber (100 ml)
4 -- length
diame@r 10 mm, length
ide diame@,r tubing as recommended bythe
All vacuurn level readings shall be e, ssed as the occluded {no-ftrow) value as shown on a rracartlm
indicator scaled 0 kPa to 100 kPa Elo spheric pnessure.
&
A.llTest for maximum v fuel and fuee air flowrate of low vacuum equip-
ment '
With the collection coi pty, switch on the,suction equipment with the vacuurn
adiusted to give the maxiil um. Occlude the inlet port to the collection container and nste
port and attach a low-resistance flowmeter ts it.
s are reached fsee Figure A.7).
ic drainage
t port of the equipment to a collecHon containerfs) to bring the totail
evacuated to 4,5 I r 0,1 l.
Note the time taken for the reading on the vacuum level indicator to increase frorn zero to 95 96 of
set vacuum level. Note the final vacuum level.
5 EN fSO 1O079-3:2014
ISO 10079-3:2OM(E) Il
fr"
:;i
:S"ffi::,"il1H".F.,j:,:,??:jfHil:*.immerriatery
t Dimensjsns iu
I(ev
1 low-resistance flowmeter (<0,L kpa
at 25 l/min)
2 tubing of length Z m
3 equipment under test
414"2 proc,
,ressure.
or. lncrea
poftand
Repett the above
test three times. Report
the widest percentage deviation fnorn the fixed
setting.
ES EN ISO I0O79-S:ZO1#
ISO 10079-3:2CI1
Key
1 on,/offvalve
2 vacuumlevelindicator
3 vacuumregulatorundertest
atnospheric pres$rrq 50 [/ant$
A.16Test for ph
4"16.1.1 vornitus
Prepare the
adding 100
NOT
benzoic acid may be added as a preserrnatirye-
A:17.2 Procedure
PL17A.4 High