Professional Documents
Culture Documents
Magnetic Flux Leakage Reliability Experiment
Magnetic Flux Leakage Reliability Experiment
Magnetic Flux Leakage Reliability Experiment
Summary:
Basic principle of Magnetic Flux Leakage (MFL) is that a powerful magnet is used to magnetize
the steel. At areas where there is corrosion or missing metal, the magnetic field "leaks" from the
steel. Analysts interpret the chart recording of the leakage field to identify damaged areas and to
estimate the depth of metal loss.
1. Advantage:
3) Effectively inspection;
4) The output from the detector can be amplified, filtered, digitized, etc and stored to produce
It is a very suitable tool for detecting bad spots in the plates. The fast MFL inspection is done only
the “suspicious” areas of the surface will be quantified by the slower but more accurate UT
technique.
2. Disadvantage:
Calibration by sample plates: 10mm and 12mm, test results please see below table 1 and 2:
Defects No. 1# 2# 3# 4#
Defects Depth 2mm/20% 4mm/40% 6mm/60% 8mm/80%
Results ╳ ● ● ●
Note:●Can be detected,╳Can’t be detected。
Table 2 12mm sample plate Test results
Defects No. 1# 2# 3# 4#
Defects Depth 3mm/25% 5mm/42% 7mm/58% 10mm/83%
Results ╳ ● ● ●
Note:●Can be detected,╳Can’t be detected。
It can’t detect the lowest holes at two different thickness sample plates.
(If increase sensitivity, the lowest hole can be detected, but the error alarm rate increases, so it is
not reliable)
5. Experiment at site:
6. Conclusion:
6.1 Normal wall thickness: 12mm; Detected defect depth: 6mm; Wall thickness loss: 50%.
Experiments show that MFL HandScan is valid for natural defect detection, suitable for large
6.2 During the experiment, it was found that the Ferro internals of equipment may cause the alarm
when the thickness was increased (such as internal sacrificial anode, support, etc.).
6.3 The heat exchanger diameter is less than 1500mm, probe is difficult to maintain balance when
scanning, have to keep the probe balance manually.
The effectively advantage of MFL is not distinct at this experiment, so it is not suitable for the
device that diameter less than 1500mm.