Professional Documents
Culture Documents
05 Validation of The Mindray VS9 Vital Signs Monitor in A Combined Adult and Pediatric Population According To ISO Standard 81060-22018
05 Validation of The Mindray VS9 Vital Signs Monitor in A Combined Adult and Pediatric Population According To ISO Standard 81060-22018
05 Validation of The Mindray VS9 Vital Signs Monitor in A Combined Adult and Pediatric Population According To ISO Standard 81060-22018
We aimed to validate the accuracy of the Mindray 0.3 ± 6.9 mmHg (systolic/diastolic) when the inflation
VS9 Vital Signs Monitor, which features the Mindray algorithm was used. For validation Criterion 2, the SD
TrueBP inflation algorithm for oscillometric blood of the averaged BP differences between the test device
pressure (BP) measurement, to check if it complies and the reference BP per subject was 5.35/6.33 mmHg
with the International Organization for Standardization (systolic/diastolic) when the deflation algorithm was used,
Standard (ISO 81060-2:2018) in a combined adult and and 5.17/5.75 mmHg (systolic/diastolic) when the inflation
pediatric population. A total of 86 participants, including algorithm was used. The VS9 Vital Signs Monitor fulfilled
both adult and pediatric subjects, were recruited. The all the criteria in the ISO Standard. Moreover, the inflation
distribution of their ages, gender, BPs and limb sizes algorithm had a shorter Measure Time (by 7–21 s) and
all complied with the requirement of the ISO standard. lower maximum inflation pressure (by 9.7–22 mmHg). The
The inflation and deflation algorithms were validated VS9 Vital Signs Monitor fulfilled all the requirements of
independently using the same-arm sequential BP the ISO Standard (ISO 81060-2:2018) in a combined adult
measurement method. For each subject, the BP was and pediatric population and is recommended for clinical
first determined by two independent observers using use. Blood Press Monit XXX: XXXX–XXXX Copyright ©
a mercury sphygmomanometer (R1). The BP of the 2024 Wolters Kluwer Health, Inc. All rights reserved.
subject was then determined by the third observer Blood Pressure Monitoring XXX, XXX:XXXX–XXXX
using the test equipment (T1). Then, using a mercury
Keywords: inflation algorithm, non-invasive blood pressure, oscillometry
sphygmomanometer, two independent observers were
asked to determine the subject’s BP (R2) again. R1-T1-R2 a
Department of Anesthesiology, The Third Affiliated Hospital of Sun Yat-sen
were considered a valid pair of data. This cycle continued University, Guangzhou and bShenzhen Mindray Bio-medicical Electronics Co.,
LTD., Shenzhen, Guangdong, China
until 3 pairs of valid data were achieved. We collected 258
pairs of valid BP data for the validation of the inflation Correspondence to Shangrong Li, MD, PhD, Department of Anesthesiology,
The Third Affiliated Hospital of Sun Yat-sen University, Guangzhou, Guangdong
and deflation algorithms respectively. For validation 510630, China
Criterion 1, the mean ± SD of the differences between Tel: +86 020 85253132; fax: +86 020 87580867;
e-mail: lishangr@mail.sysu.edu.cn
the readings obtained from the test device and reference
BP was 0.0 ± 6.6/−1.8 ± 7.1 mmHg (systolic/diastolic) *Na Guo and Yihan Zhang contributed equally as first authors of this work.
when the deflation algorithm was used, and 2.4 ± 6.3/ Received 29 February 2024 Accepted 4 March 2024.
1359-5237 Copyright © 2024 Wolters Kluwer Health, Inc. All rights reserved. DOI: 10.1097/MBP.0000000000000704
2 Blood Pressure Monitoring XXX, Vol XXX No XXX
Methods and feet placed flat on the floor and have the measured
Test device arm supported at the heart level. If the lower limb was
The VS9 Vital Signs monitor is an automated oscillo- measured, the subject would lie down on a bench and
metric BP monitor intended for use in adults and chil- maintain the measured limb at the same level as their
dren. The device is equipped with 4 sizes of cuffs for heart. During measurement, subjects did not speak or do
limb circumferences 18–26 cm, 25–35 cm, 33–47 cm, and anything that might cause BP fluctuation. Each subject
46–66 cm. The test devices were operated according to participated in two independent rounds of trial, being
Downloaded from http://journals.lww.com/bpmonitoring by BhDMf5ePHKav1zEoum1tQfN4a+kJLhEZgbsIHo4XMi0h
China; it also complies with ICH E6(R3) Good Clinical presented in Table 3. A normalized Bland-Altman scat-
Practice guidelines. ter plot of the difference in test-reference BP relative to
their mean is shown in Fig. 1.
Results In addition to verifying the test data, we also compared
Eighty-six participants were enrolled. Their composition how long the measurement took (Measure Time), and
met all the requirements for age, gender, BP, and cuff the maximum inflation pressure for each BP reading
distribution in the ISO standard. The participants’ char-
Downloaded from http://journals.lww.com/bpmonitoring by BhDMf5ePHKav1zEoum1tQfN4a+kJLhEZgbsIHo4XMi0h
Fig. 1
Downloaded from http://journals.lww.com/bpmonitoring by BhDMf5ePHKav1zEoum1tQfN4a+kJLhEZgbsIHo4XMi0h
CywCX1AWnYQp/IlQrHD3i3D0OdRyi7TvSFl4Cf3VC4/OAVpDDa8K2+Ya6H515kE= on 04/10/2024
Standardized Bland-Altman scatterplots of the Test-Reference BP differences against their average. (a) SBPs with deflation algorithm; (b) DBP with
deflation algorithm; (c) SBPs with inflation algorithm; (d) DBP with inflation algorithm.
Table 4 Comparison of the measure time and maximum inflation pressure between deflation and inflation algorithms
Average measure time (s) Maximum Inflation Pressure (mmHg)
Systolic pressure Deflation Inflation Deviation of Deflation Inflation Deviation of maximum
distribution (mmHg) measurement measurement measure time (s) measurement measurement inflation pressure (mmHg)
cause inaccurate results. However, we did not observe and the limb length is short relative to the circumference
any mode transitions during this study. [10]. Tronco-conically shaped cuffs may better fit the arm
shape of these people, but are not available with our refer-
There are several points we would like to address. First, ence sphygmomanometer. The intra-arterial measurement
the bladder size of the maximum reference cuff is could be considered as the gold standard for reference BP
20.5 cm×38 cm. This cuff is mainly used when measuring measurement in this population, but this method has its
BP in the thighs, and it could not meet the width require- own methodological pitfalls in comparison with cuff-based
ment (37–50%) stated in the ISO standard. The ausculta- BP measurements [10]. In addition, the majority of the sub-
tory readings can be inaccurate when using this maximum jects in this study were healthy volunteers, so arterial cath-
cuff. This is a common issue for manufacturers worldwide eterization without therapeutic purposes remains to be an
since people with large limb circumference are often obese ethical concern. Therefore, we did not choose this method.
Validation of the Mindray VS9 Vital Signs Monitor Guo et al. 5
Secondly, no children in this study fell in the hypertension 2 Alpert BS. Validation of the Welch Allyn SureBP (inflation) and StepBP
(deflation) algorithms by AAMI standard testing and BHS data analysis.
categories ≥140 mmHg and ≥160 mmHg, posing a potential Blood Press Monit 2011; 16:96–98.
risk of data imbalance. However since there is no specific 3 Alpert BS. Validation of the Welch Allyn Home blood pressure monitor
requirement in the ISO standard for hypertensive pediatric with professional SureBP algorithm with a special feature of accuracy
during involuntary (tremor) patient movement. Blood Press Monit 2019;
cases, we did not intentionally recruit such subjects. And 24:89–92.
we did not come across any hypertensive children during 4 John O, Campbell NRC, Brady TM, Farrell M, Varghese C, Velazquez
recruitment. Oscillometric devices have been demon- Berumen A, et al. The 2020 ‘WHO technical specifications for automated
Downloaded from http://journals.lww.com/bpmonitoring by BhDMf5ePHKav1zEoum1tQfN4a+kJLhEZgbsIHo4XMi0h