M506-E361G

Dec. 2016

X-ray R/F System

Installation Manual
This manual is for professional service engineers.
Do NOT use this manual unless you have received proper training.
Categories of information

In this manual, safety or utilizable information is categorized as follows.

Symbol Significance

Indicates critically hazardous situation that, if not avoided, may


result in serious injury or death.

Indicates indirectly or potentially hazardous situation that, if not


avoided, may result in serious injury or death.

Indicates hazardous situation that, if not avoided, may result in


minor or moderate injury, damage to the product, or fire.

Indicates information for proper use of the product.

Indicates information for better performance of the product.


Table of Contents

Chapter 1 Installation
1.1 Installation ............................................................................................... 1-2
1.2 Install Optional Features ........................................................................ 1-17
1.3 In case of connecting with CH-200 (Without Bucky Table) ..................... 1-27
1.4 In case of connecting with CH-200 (With Bucky Table) .......................... 1-31

Chapter 2 Appendix
2.1 Troubleshooting ....................................................................................... 2-2

M506-E361G SONIALVISION G4 Installation Manual 1


Revision History

Describe the revision history of this document in the following table.

REVISION DATE CHANGES

Original Mar. 2013 First released.


A Oct. 2013 Added description of color marking on inter-unit cables to “Assigning
Installation”.
Corrected clerical errors.
Added Chapter 2 "Appendix".
B Feb. 2014 Chapter 1
Modified the description of NOTE in ‘Adjusting Image Quality’.
Added ‘1.2 Install Optional Features’.
Added ‘1.3 In case of connecting with CH-200’.
Chapter 2
Modified the description in ‘No.11 (PHV-DR affection)’.
Added ‘No.24, 25, and 26 (Error during FPD calibration)’.
C Dec. 2014 Chapter 1
Added the IQ-AM document in ‘1.1 Installation’.
Modified the title of the reference of DR in ‘Input Standard Data’.
D May 2015 Chapter 1
Added the setting SIM-Link communication with Side Station i3 in ‘1.1
Installation’.
Modified the description in ‘1.2.2 SLOT Radiography’ and ‘1.2.3
TOMOS Radiography’.
Added ‘1.2.4 Iteration Reconstruction’, ‘1.2.5 Side Station’ and ‘1.4 In
case of connecting with CH-200 (with Bucky Table)’.
E Feb. 2016 Chapter 1
Added the meaning of tape color on the cardboard in ‘1.1 Installation’.
Changed the startup location of the UD Maintenance Tool to REF PC.
Chapter 2
Changed the description in No.23 and 24.
F Oct. 2016 Chapter 1
Added Dip SW settings of TERMINAL-2002 PCB in ‘1.4 In case of
connecting with CH-200 (With Bucky Table)’.
G Dec. 2016 Supported SFD-1717BF type FPD.

2 SONIALVISION G4 Installation Manual M506-E361G


Chapter 1
Installation
Chapter Contents

1.1 Installation

1.2 Install Optional Features

1.3 In case of connecting with CH-200 (Without Bucky Table)

1.4 In case of connecting with CH-200 (With Bucky Table)

M506-E361G SONIALVISION G4 Installation Manual


Chapter 1 Installation

1.1 Installation

Installation procedure is described in this chapter.


The following installation manuals are mentioned in this chapter.

Abbreviation Document No. Document name

ZS-IM M506-E362 ZS-200 Installation Manual


UD-IM M501-E401 D150BC-40S/VC-40S Installation Manual

DR-IM M517-E647 DR-300 Installation Manual

IQ-AM *1 M517-E648 SONIALVISION G4 Image Quality


Adjustment Manual

M517-E744 SONIALVISION G4 Image Quality


Adjustment Manual for EU

Type R-300 M526-E320 COLLIMATOR TYPE R-300 Installation


Installation Manual Manual

SONIALVISION G4 M517-E209 X-ray R/F System SONIALVISION G4


Instruction Manual Instruction Manual

SONIALVISION G4 M517-E260 SONIALVISION G4 Procedure Preset List


Procedure Preset
List *1 M517-E279 SONIALVISION G4 Procedure Preset List
for EU

DR-300 Instruction M517-E210 Digital Radiography System DR-300


Manual (IA) Instruction Manual (Image Acquisition)

SONIALVISION G4 M517-E684 X-ray R/F System SONIALVISION G4


Release Notice Release Notice

CH-IM M514-E350 CH-200/CHU-200 Installation Manual


BR-IM M511-E368 BR-120/120T Installation Manual

SS-IM M517-E753 Side Station i3 Installation Manual

*1: The default setting has the 3 types below. Select an appropriate one according to the region.

(a) Standard
The maximum skin dose for fluoroscopy is set as 5.75 R/min (= 50 mGy/min). It is for a region
where subjects tend to be thinner (coverage of 30 cm or less) and image quality has high priority.
Use this for the region where the maximum skin-dose of 10 R/min is not permitted.

(b) Bariatric

(c) The maximum skin dose for fluoroscopy is set as 10 R/min (= 86 mGy/min). Use this for the region
where the maximum skin-dose of 10 R/min is permitted and subjects tend to be thicker (coverage of
more than 30 cm).EU
It is based on DIN 6868-150. Use this for the region where dose regulation equivalent to that in
Germany is enforced.

When you use “(a) Standard” or “(b) Bariatric”, refer to M517-E648 in IQ-AM and M517-E260 in
SONIALVISION G4 Procedure Preset List. When you use “(c) EU”, refer to M517-E744 in IQ-AM and
M517-E279 in SONIALVISION G4 Procedure Preset List.

1-2 SONIALVISION G4 Installation Manual M506-E361G


1.1 Installation

Focus on the underlined description and install.

Introduction

Check that no hollows on the cover and scratches on the table.


Be sure to check the parts when unpacking cardboard.
(Take out all the packing materials once to check nothing is left.)
Unpacking
Color of tapes on the cardboard indicates the follows;
Yellow : Parts for operation room are packed
Brown or Clear : Parts for examination room are packed.

According to the layout to install the table, ZS cabinet,


high-voltage generator cabinet, DR-EXAM cabinet, DR-OPE
cabinet and console.
Installing
More than 1050x1900 mm is required for carry-in entrance.
(Refer to “Appendix 4 Installation dimensional drawing” in
ZS-IM.)

Attach main frame ASSY to base ASSY.


Assembling Table Attach post ASSY.
Remove the tabletop from main frame.
(Refer to “2.2 Sub-assembly” in ZS-IM “PART1: Installation”)

To reduce the installation period, separate in 3 groups of table,


control cabinets and console for wiring. Wiring descriptions for
each group are shown as follows:
#1 Wiring table
#2 Wiring control cabinet
#3 Wiring console
Assigning Installation
Wire #1-1, #2-1~#2-3 and #3-1 first for enabling the operation of
table.

Each inter-unit cable is equipped with color marking on both


ends in accordance with its destination as below. Wire the
cables in reference to the marking.
ZS-200 main body: Red
ZS cabinet: Yellow
UD cabinet: Orange
DR EXAM cabinet: Blue
Remote console and DR OPE cabinet: Gray
Local console and Monitor cart: Brown

M506-E361G SONIALVISION G4 Installation Manual 1-3


Chapter 1 Installation

#1-1. Wiring for enabling table operation


(Refer to “2.3 Wiring for main body” in ZS-IM “PART1: Installation”)
 2.3.1 Wiring of main frame
 2.3.2 Wiring for control cabinet
 2.3.3 Wiring for control panel
Check that procedures from #2-1 to #2-3 for wiring control cabinet
and #3-1 for wiring console are completed.

Working on the following section before power distribution.


 2.3.4 Setting of jumper (NEX-SH PCB)
2.3.7 Setting DIP Switch (NEX-SH PCB)

Be sure to connect and disconnect the cable when the power is


OFF. A case is reported that ZSCONT-09 PCB got out of order
due to the connection or disconnection of ZS-REMOTE cable
during the power distribution.
And working on the following section after power distribution.
 2.3.8 Initializing AC servo motor (FPD carriage and post ASSY)
 2.3.9 Registering lever potentiometer (Remote/ Local console )

Wiring post ASSY in standing position after the table can be tilted.
 2.3.10 Wiring Post ASSY
#1 Wiring Table
1
After wiring post ASSY, mount the X-ray tube and then wire the
low-voltage cable A.
To prevent short circuit, be sure to connect the terminal of
low-voltage cable A to the X-ray tube first, and then distribute the
power.
 2.3.11 Initializing AC servo monitor (Tube Support)

#1-2. Wiring X-ray tube


(Refer to “2.4 Mounting the X-ray tube” in ZS-IM “PART1:
Installation”)
Check that Al filter is 0.3mm+1.2mm (=1.5mm).
Set the table to standing position (or Trendelenburg position) and
rotate the tube rotation at 90° to clockwise (or counterclockwise).
Attach trunnion ring and X-ray tube cover mounting bracket.
Pay attention to the attaching place of trunnion ring.
Mount and fix the X-ray tube.
Connect the X-ray tube side of X-ray low-voltage cable.

(“2.5 Mounting collimator” in ZS-IM “PART1: Installation”)


Mounting collimator to X-ray tube.
Remove collimator cover and cable tie of the leaf.
Wire the collimator cable and connect to the collimator.
Replace the collimator cover.

1-4 SONIALVISION G4 Installation Manual M506-E361G


1.1 Installation

(Refer to “3.4.1.8 Maximum Field Size Adjust” in ZS-IM


“PART2: Maintenance”)
If collimator has no polygonal collimator (C-leaves) and
single-acting H mask (L/R mask), set field size to 476×476
mm (SID = 1 m) for the maximum open of H/V leaves.

(Refer to “3.1 Wiring of X-ray tube” in ZS-IM “PART1:


Installation”)
Bind the high-voltage cable, X-ray tube oil hose and
low-voltage cable on the tube side and fix with a plate.
Have an extra length of high-voltage cable for wiring to plug-in
or out it upon maintenance.
Connect the connector for low-voltage cable of tube support,
work on the cable clamp, fix the high-voltage cable, oil hose
and collimator cable to the upper area of the tube support with
a plate, adjust the length and then bind with a spiral tube.
Fix the cable from tube rotation with a cable tie.
Return the tabletop to horizontal position.
Adjust the length of cables in the back of column and bind with
a spiral tube. At this time, consider the order of cables on the
side of column.
Bind the cables on the side of column and fix with brackets.
Wire the cables such as high-voltage cable to the runner flex.

#1 Wiring Table #1-3 Mounting FPD MOUNT UNIT and Wiring FPD
2 (Refer to “2.6 Attaching FPD” in ZS-IM “Section 1:
INSTALLATION”)

(Refer to “3.2 Wiring of FPD” in ZS-IM “Section 1:


INSTALLATION”)

#1-4 Wiring main frame mounting base


(Refer to “3.6 Wiring of the microphone cable” in ZS-IM
“Section 1: INSTALLATION”)
Wire the cable 2 and 3 of table microphone cable and pull out
the cable 3 into the back of main frame mounting base.

M506-E361G SONIALVISION G4 Installation Manual 1-5


Chapter 1 Installation

(Refer to “3.3 Wiring on back of main body fixing base” in ZS-IM


“Section 1: INSTALLATION”)
Pull out the cable from FPD carriage and main frame ASSY foot side
to the exit of main frame mounting base.
Fix the cables from post ASSY runner flex with fixing brackets and
pass through the cables in order of 2 high-voltage cables, 2 oil hose,
X-ray tube low-voltage cable and other cables to the duct holder
without crossing the cables.
Pass through the cables from FPD carriage and main frame ASSY
foot side to the duct holder.
Attach the duct cover with safety sensor. Connect the safety sensor
to CN-MAT 3.

#1-5 Wiring on the back of base ASSY


(Refer to “3.4 Wiring on the back of Base” in ZS-IM “Section 1:
INSTALLATION”)
Pass the cables from the duct holder of main frame mounting base to
the cable BOX of back of base ASSY, and adjust and bind the cables.
If space between back of the table and the wall is narrow for
convenience of layout of control room, remove the plate and terminal
box from the cable BOX once and fix the cable to plate and terminal
box with aluminum clamp and connect the earth cable. And then
replace to the cable BOX.
Pass the cable from main frame mounting base left (seeing from the
table front) through the cable BOX of back of base ASSY, and adjust
the length and bind.
#1 Wiring Table To place the temperature and humidity data logger on the cable BOX,
3 pull out the edge of MODULER CABLE from the cable BOX and pull
out the opposite side of this cable from the table as well as the other
cables.

#1-6 Mounting heat exchanger


(Refer to “2.7 Installing the heat exchanger” and “2.8 Installing
the heat exchanger (For Narrow Backside)” in ZS-IM “Section 1:
INSTALLATION”)
Mounting X-ray tube heat exchanger. Fix the extra oil hose and also
pull out the cable from heat exchanger from the base ASSY.
If there is insufficient clearance space between back of the table and
the wall, remove the mount from base ASSY and mount heat
exchanger. Then fix the mount to base ASSY.

#1-7 Mounting tabletop

#1-8 Attaching cover with safety sensor switch


(Refer to “4 Attaching the Covers” in ZS-IM “Section 1:
INSTALLATION”)
When operating the table in the initial setting mode, safety sensor will
be ignored. To operate in the normal mode, mounting safety sensor is
required.
Attach stand cover L/R and connect to CN-MAT 5/6.
Connect the safety sensor of main frame mounting base to
CN-MAT 4.
Attach duct cover and connect to CN-MAT 3.
Attach base cover L/R and connect to CN-MAT 1/2.

1-6 SONIALVISION G4 Installation Manual M506-E361G


1.1 Installation

#2-1 Wiring around high-voltage generator cabinet


Measure the voltage of 3-phase power supply of distribution
board and change the entrance cable connection of T1-40
transformer and T2-CE transformer in high-voltage generator
cabinet.
(Refer to “3.5 Changing T1 and T2 Connections” in UD-IM
“Installation”)
Change the connection of XAT-2H tap according to the
voltage of 3-phase power supply.
Connect 3-phase power cable and shielded wire between the
distribution board and high-voltage generator cabinet.
Connect the earth cable between the distribution board and
high-voltage generator cabinet.

Pull in the cable from other unit to the high-voltage generator


cabinet.
From the table:
High-voltage cable (HV transformer, SGC)
(Refer to “3.7 Connecting the High-Voltage Cable” in
UD-IM “Installation”)
Phototimer cable (PHTAFU-2009 J22 [DET1])
(Refer to “3.9 Connecting the Direct Phototimer
Cable” in UD-IM “Installation”)
From ZS cabinet:
ZS power supply cable
#2 Wiring Control Cabinet
Cable,UD-MECA(COMDAR-2009 J13 CN MECA)
1
Connect ZS power supply cable just before the power
distribution.
From DR-EXAM cabinet:
IBS signal (COMDAR-2009 TV7,TV2),
Cable,UD-DR(COMDAR-2009 J34 DAR)
(Refer to “3.10 Connecting the System Cable” in UD-IM
“Installation”)
For combination of the second X-ray tube, refer to “3.11
Connecting to External Devices” in UD-IM “Installation”
for connection.

#2-2 Checking the setting of high-voltage generator cabinet


switch and setting of temporary operation
(Refer to “3.13 Board Switch Settings and Jumber Pins
Settings” in UD-IM “Installation”)
Check DIP SW and Jumper settings of UD CONT-2002,
TERMINAL-2002, COMDAR-2009, PHTAFU-2009, GRID
CONT-2000 boards.
Until the adjustment after X-ray exposure, set the high-voltage
generator cabinet to temporary operation for safety.
(Refer to “1.8 Preparing for Test Operation” in UD-IM
“Maintenance”)

M506-E361G SONIALVISION G4 Installation Manual 1-7


Chapter 1 Installation

#2-3 Wiring for enabling table operation


Connect ZS power supply cable (with earth) to ZS cabinet from the
high-voltage generator cabinet.
(Refer to “3.10 Connecting the System Cable” in UD-IM
“Installation”)

Change the connection of ZAT tap according to 3-phase power


supply voltage if necessary.
(Refer to “2.3.5 Changing Auto Transformer ZAT-1 (Optional)
Connection in ZS-IM “Section 1: INSTALLATION”)

Change the connection of Starter SA-61S T2 Transformer. SA-61S is


placed in ZS cabinet.
(Refer to “3.6 Connecting the SA-64S Starter” in UD-IM
“Installation”)

Pay attention not forge to change the connection of SA-61S T2


transformer when wiring 400V series. The system may be broken
if make a mistake in connection change.

Shorten 2-3 of ZS CONT-09 PCB JP4 to bypass the stop switch.


(Normally shorten 1-2)
Check that procedures #1-1 for Wiring table is completed.
And then enable to distribute 3-phase power supply to ZS cabinet.
#2 Wiring Control Cabinet
2 #2-4 Wiring around DR-EXAM cabinet
Attach the cover on the upper area of DR-EXAM cabinet.
Put non-slip zipper on MPC and CP2 and fix.
(Refer to “3.4 Fastening the Cabinet” in DR-IM)

Wire the following cables to CP2 / PowerBox.


CABLE, MAINS 110V CP2 POWER – S4
LAN CABLE,CP2 CP2 ETHERNET – HUB CH2
ISYS-CP2 LVDS CABLE CP2 16bBIT DIGITAL – I-SYS CH1
ISYS-CP2 I/O CABLE CP2 EXTERNAL SYNC – I-SYS CN20
SERIAL CABLE,4343CB CP2 D-sub – MPC COM5
(Refer to “3.6.3 Connecting CPLC, 4343CB (CP2)” in DR-IM or
“3.6.4 Connecting POWERBOX, AS-10(PowerBox)” in DR-IM)

Measure the power supply voltage of single-phase 200VAC of


distribution board and change the connection of P-200 cable to
transformer if necessary.
(Refer to “3.5 Supplying Voltage Through a Connection” in
DR-IM)

Connect the power supply and earth cables between distribution


board and power supply ASSY.
(Refer to “3.6.2 Connecting the DR-EXAM Cabinet” in DR-IM)

1-8 SONIALVISION G4 Installation Manual M506-E361G


1.1 Installation

To connect to DR-OPE cabinet, connect and bind the


following cables and draw them into the console.
POWER CABLE, OPE – RELAY - ASSY J105
CABLE,DR-REMOTE - ACQ PC COM1 & ACQ PC AUDIO
DVI PS CABLE - DVI J31
LAN CABLE REF – HUB CH5
LAN DVI CABLE4 - DVI MAIN J13,J21,J23,J25
Each cable connected to DVI MAIN J13,J21,J23,J25, will be
connected DR OPE cabinet DVI OUT CN3,CN1,CN2, TOUCH
PANEL LCD LANCN2 so mark on the cable.
(Refer to “3.6.2 Connecting the DR-EXAM Cabinet” in
DR-IM)

To connect to the high-voltage generator cabinet, connect and


draw the following cables into the high-voltage generator
cabinet.
CABLE,UD-DR – ISYS CN23, J501
IBS CABLE FOR EVOLVE - I-SYS CN15
(Refer to “3.6.2 Connecting the DR-EXAM Cabinet” in
DR-IM)

Draw the following cables for the table into DR-EXAM Cabinet
and wire.
FAN CABLE,FPD – RELAY ASSY J107
FPD Optical Cable (CABLE,20M OPTIC) – CP2/PowerBox
FIBER
#2 Wiring Control Cabinet
FPD Power Supply Cable (CABLE,20M POWER) –
3
CP2/PowerBox
(Refer to “3.6.2 Connecting the DR-EXAM Cabinet” in
DR-IM)

Connect the cable P-CABLE, MONITOR CART for local


monitor cart to RELAY ASSY J104.
If there is a large local console option, connect the LAN DVI
cable from it to DVI MAIN J24 and CALBE,DR-REMOTE to
ACQ PC COM2 and ACQ PC AUDIO.
(Refer to “3.6.2 Connecting the DR-EXAM Cabinet” in
DR-IM)

Check the switch setting


Check the switch setting of I-SYS BOARD,DVI MAIN BOARD.
(Refer to “3.10 Checking the DIP Switch Settings” in
DR-IM)

#2-5 Wiring around ZS cabinet


Connect the cable “Cable,UD-MECA” from high-voltage
generator to ZS CONT-09 PCB CN3.

Check the switch setting of MOTHER-SA60,NEX-SH for


SA61S.
(Refer to “3.13 Board Switch Settings and Jumber Pin
Settings” in UD-IM “Installation”)

M506-E361G SONIALVISION G4 Installation Manual 1-9


Chapter 1 Installation

Check that procedures #1-1 for wiring table and #2-1~#2-3 for wiring
control cabinets are completed.

#3-1 Wiring around control console


Remove covers for DR-OPE cabinet shipping and etc. and ready for
wiring.
(Refer to “3.3.2 Installing the DR-OPE Cabinet” in DR-IM)

Connect the following cables from DR-EXAM cabinet.


POWER CABLE, OPE – DVI SUB CN5
CABLE,DR-REMOTE - TOUCH Panel LCD CN8 D-sub & SONIAL
REMOTE 2009 JPC
DVI PS CABLE – DVI SUB CN5
LAN CABLE REF – REF PC LAN
LAN DVI CABLE4 – DVI SUB CN3,CN1,CN2 & TOUCH Panel LCD
LAN CN2
(Refer to “3.6.6 Connecting the DR-OPE Cabinet” in DR-IM)

Connect remote monitor to DVI OUT CN3,CN5.


Connect MONITOR PW CABLE OA and MONITOR PW CABLE OR
#3 Wiring Console
to TB3[E,R0,R100].
Connect keyboard and mouse to REF PC.
Connect EARTH CABLE,OPE from DR-OPE cabinet earth bracket to
distribution board and EARTH CABLE,#10 to remote console.
Connect the following cables related to intercom to
SonialRemote-2009 PCB.
Table Microphone Cable3 – JPAT1 PAT MIC
Microphone Cable from Local Monitor Cart – JPAT2 PAT2 MIC
Microphone for Control Console – JPC DOC MIC
Speaker for control room – Red TB1-3 DOCTOR(+),Black TB1-4
DOCTOR(-)
Speaker for examination room - RedTB1-1 PATIENT(+),Black
TB1-2 PATIENT (-)
FOOT SW Cable – JFT FOOT SW
Connect hand switch to J9
(Refer to “1.2.2 Connection” in UD-IM “Appendix”)

Set high-voltage generator cabinet as a temporary operation (refer to


#2 Wiring Control Cabinet 1) and connect only cable JS1 to
UDCONT-2002 PCB here.

Checking Power After wiring the table, control cabinet and console, distribute the power
Distribution and check that there is no startup error and unusual smell.
Check the voltage at the following point:
HIGH-VOLTAGE GENERATOR CABINET TERMINAL-2002 PCB
L0,L100: 100VAC±5V
DR-EXAM CABINET TRANS ASSY 0V,100V: 100VAC±5V

1-10 SONIALVISION G4 Installation Manual M506-E361G


1.1 Installation

Apply “Revision UP” if Revision of SW/FW is less than the


Checking SW/FW standard setting of each unit in “Table 1-1 Revisions and Items
Revision to be checked before adjustment” in “1. Introduction” in
IQ-AM

Logon to REF-PC with “shimadzu” for user name and “service”


for password.
Input software license key.
(Refer to “5 Entering the Software License Key” in DR-IM)
Set date and time.
(Refer to “6 Setting the Data and Time” in DR-IM)
Set FPD type.
(Refer to “40 [Appendix 28] Change FPD Type” in DR-IM)
Set Local Area Connection 2 setting for network in hospital.
(Refer to “7 Setting the Network” in DR-IM)
Set environment of installation.
(Refer to from “8.1 Starting Up the Installation
Environment Setting Tools” to “8.4 Study List
Setting” in DR-IM)
Set network such as MWM/MPPS, STORAGE and PRINT.
Check the connection and operation of network in hospital and
remote maintenance network.
(Refer to from “8.5 Setting the MWM/MPPS (Option)” to
“8.12 The Checking the Connection ([Ping] and [Echo]
Buttons) in DR-IM)
Initial Setting of DR-300
Set data saving to CD-R/DVD-R media.
(Refer to “8.13 Setting the Media Storage” in DR-IM)
Set SIM-Link communication with Side Station i3.
(Refer to “8.14 Setting the SIM-Link” in DR-IM)
Set FPD calibration setting.
(Refer to “8.15 Setting the FPD Calibration” in DR-IM)
Set acquisition PC.
(Refer to “8.17 Setting the Acquisition PC” in DR-IM)
Set the setting of touch panel.
(Refer to “8.18 Setting the Touch Panel” in DR-IM)
Set the setting of console.
(Refer to “8.19 Setting the Console” in DR-IM)
Adjust the density of printer.
(Refer to “9 Setting the Print Density” in DR-IM)
Register FPD Calibration Data.
(Refer to “18 [Appendix 6]The Initial Setting After
Installing or Replacing the FPD” in DR-IM)
Set the setting of FPD sensitivity correction coefficient.
(Refer to “19.5 Changing the System Parameter” in
DR-IM)

Connect PC to NEX-SH and set the initial setting.


(Refer to “5 Settings After Assembly” in ZS-IM “Section 1:
INSTALLATION”)
Initial Setting of ZS-200 Set the initial setting of date, time and tilting counter.
Register the origin of servo motors and set the setting of DIP SW
on NEX-SH PCB. Check that not error occurs in each operation
of ZS-200.

M506-E361G SONIALVISION G4 Installation Manual 1-11


Chapter 1 Installation

Restore standard data from the master media 565-10180(SONIAL G4


SYS SW).
・ UD
Upload the standard adjustment data in accordance with X-ray tube
on site and image quality adjustment.
(Refer to “3.3.8.2 Restoring from Backup Data” in UD-IM
“Appendix”)
Startup REF PC and upload with UD maintenance tool.
Set HD mode of Pulsed Fluoroscopy according to the region.
(Refer to “5.2.2 Setting DC Fluoroscopy and Pulsed
Fluoroscopy Modes” in IQ-AM)
・ DR
Upload the procedure in accordance with image quality adjustment.
(Refer to “17 [Appendix 5] Changing to the Bariatric (Low
Input Standard Data Dose) / EU Setting Mode” in DR-IM)
When using Bariatric procedure for United States, add the
procedure separately.
(Refer to “10.7 Register System Procedure Tool” in DR-IM)
The settings of procedures are described in detail in
“SONIALVISION G4 Procedure Preset List”.
・ ISYS
Upload the image quality adjustment data in accordance with the
FPD type and image quality adjustment.

Input data will be different depending on the FPD type and the
image quality adjustment. Refer to “4.7 Setting Image Quality
Adjustment Data” in IQ-AM for which data to be used.

・ Return the mode of UD from temporary operation mode to normal


operation mode.
・ Use maintenance tool to initialize the exposure counter and input
the system serial number.
(Refer to “4.11 Entering Serial Numbers and Initializing X-Ray
Tube Usage Log [Information] – [Report] in UD-IM
“Installation”)
・ Go to FVR mode, set to the 1st tube and input 4.00A for FVR value.
Operation Check of X-ray (Refer to “4.15 Adjusting the Radiography Tube Current
Tube [Adjustment] – [FVR] in UD-IM “Installation”)
Warming-up Tube ・ Use a hand switch for Ready operation to check that Ready Up is
Adjusting Tube Current operated correctly. And also check there is no error on the rotating
1 sound.
・ If the 2nd tube is connected, change the tube. And check that
change by listening to the sound.
As same as the 1st tube, input 4.00A for FVR value. Use a hand
switch for Ready operation to check that Ready Up is operated
correctly. And check that the 2nd tube starts correctly by listening to
the rotation sound of X-ray tube, and also check there is no error on
the rotating sound.

1-12 SONIALVISION G4 Installation Manual M506-E361G


1.1 Installation

・ Adjust radiography X-ray tube current FVR while closing the


collimator.
(Refer to “4.15 Adjusting the Radiography Tube Current
[Adjustment] – [FVR] in UD-IM “Installation”)
・ Adjust acquisition X-ray tube current FVF while closing the
collimator.
(Refer to “4.18 Adjusting the Fluoroscopy Tube Current
Operation Check of X-ray [Adjustment] – [FVF] in UD-IM “Installation”)
Tube
Warming-up Tube
Adjusting Tube Current Set the adjustment mode to MANUAL and adjust when DC
2 fluoroscopy is selected with low-voltage current, and se to
AUTO and adjust when pulsed fluoroscopy is selected.

・ Adjust flash calibration.


(Refer to “4.20 Adjusting the Flash for 0.5-Second Spot
Filming [Adjustment] – [Flash]” in UD-IM “Installation”)

・ The adjustment is also the X-ray tube warming-up.

Use alignment adjustment jig for adjusting X-ray tube alignment


and collimator leaf.
Draw the center of FPD with software.
(Refer to “15 [Appendix 3] Showing the cross cursor on
the ACQ-PC monitor” in DR-IM)

Adjusting X-ray Tube


Alignment Pay attention not to open the collimator leaf more than
requires.
・ On SONIALVISION G4 system, virtual collimation is
installed so the acquisition of leaf is not acceptable.
Disable to acquire the accuracy of virtual collimation if
open the collimator leaf more than requires.
・ Open the collimator leaf more than requires may give
excess exposure.

Check the monitor setting.


Checking Monitor Setting
(Refer to “4.7 Monitor Adjustment” in IQ-AM)

M506-E361G SONIALVISION G4 Installation Manual 1-13


Chapter 1 Installation

According to the Image Quality Adjustment Manual, adjust and check


the following things:
Check the characteristics of fluoroscopy tube current
Check fluoroscopy dose
FPD calibration
Check fluoroscopy IBS balance tube voltage
Check pixel value of fluoroscopy
Input memory shot F-T curve
(Refer to “5 Adjusting Fluoroscopy Mode” in IQ-AM)

Adjusting Image Quality Adjust phototimer of DR radiography mode


Check pixel value of DR radiography
(Refer to “6 Adjusting DR Radiography Mode” in IQ-AM)

Exit the UD Maintenance Tool after the adjustment of phototimer


(PHV-DR). The adjusted value will take effect after exiting the UD
Maintenance Tool.
Check preset parameters
(Refer to “Appendix C Mode Information” in IQ-AM)

After completing all the image quality adjustment, adjust the dose
display.
Adjusting Dose Display
(Refer to “4.24 Adjusting the Dose Display” in UD-IM
“Installation”)

1-14 SONIALVISION G4 Installation Manual M506-E361G


1.1 Installation

Set the customized keys of remote control, local console and


R-300 control panel.
Check that factory settings are set as follows:
Remote Control Panel
<SW1> FLUO Store ,<SW2> Relative Mode
<SW3> V Flip ,<SW4> H Flip
<SW5> FLUO HD ON/OFF ,<SW6> FOV 6inch
<SW7> FOV 9inch (Roll Step CW *1)
<SW8> FOV 12inch (Roll Step CCW *1)
<SW9> FLUO RATE Down ,<SW10> SDA Toggle
Large local Console
<SW1> FLUO Store ,<SW2> Relative Mode
<SW3> V Flip ,<SW4> H Flip
<SW5> FLUO HD ON/OFF ,<SW6> FOV 6inch
<SW7> FOV 9inch (Roll Step CW *1)
<SW8> FOV 12inch (Roll Step CCW *1)
<SW9> FLUO RATE Down ,<SW10> SDA Toggle
Mini local Console
<SW1> Previous Protocol ,<SW2> FLUO RATE Down
<SW3> Collimator Lamp (Roll Step CW *1)
Setting Customize Keys <SW4> SDA Toggle (Roll Step CCW *1)
<SW5> Next Protocol ,<SW6> Tube Obli HEAD
<SW7> Tube Obli CENTER ,<SW8> Tube Obli FOOT
*1: Set if there is foot rest.
R-300
<SW1> Tube Rot (CW) , <SW2> FR-Slide(Head)
<SW3> Tube Rot (CCW) , <SW4> FR-Slide(Foot)
Consult about the selection of assigned functions with hospital.
Put switch symbols on the operation key corresponding to
assigned functions.
Setting Procedure:
・ ZS, R-300
(Refer to “3.2.9 SW Customize” and “3.2.10 R-300 SW
Customize” in ZS-IM “Section 2: MAINTENANCE”)
・ DR
(Refer to “8.17.1 Console Buttons” in DR-IM)

Pay attention not to assign the function of both ZS


and DR to the same key on the console. Or it may not
operate correctly.

Fixing ZS-200 Fix the table to the floor. Also fix all cabinets to the wall.
Attaching Covers and etc. Attach covers of the table, cable duct, zipper tube and etc.

M506-E361G SONIALVISION G4 Installation Manual 1-15


Chapter 1 Installation

Making setting by reference to ‘1.2.3 TOMOS Radiography’


Setting for TOMO
in this book.

Making setting by reference to ‘1.2 Install Optional Features’


in this book if the following optional features are included.
DSA Option
Setting Optional Features SLOT Radiography Option
TOMOS Radiography Option
IR (Iteration Reconstruction) Option
Side Station Option

Backup the result of adjustment of each component.


・ ZS-200
(Refer to “3.2.11 Data Download/Upload” in ZS-IM
“Section 2: MAINTENANCE)
・ R-300
(Refer to “3.4.1.6 Data Download/Upload” in ZS-IM
Backup Adjustment Data “Section 2: MAINTENANCE”)
・ UD
(Refer to “3.3.8.1 Creating Backup Data” in UD-IM
“Appendix”)
・ DR
(Refer to “10.1 Backup of the Environmental Setting
Table” in DR-IM)

Clean

Put the tape on the floor to indicate the operating range of


Operating Range of ZS-200 ZS-200.

End

1-16 SONIALVISION G4 Installation Manual M506-E361G


1.2 Install Optional Features

1.2 Install Optional Features

Installation procedure of the following optional features is described in this section. Please
refer to the installation manual of each component for other optional features than below.

Optional Features Refer to

DSA 1.2.1
SLOT Radiography 1.2.2

TOMOS Radiography 1.2.3

IR (Iteration Reconstruction) 1.2.4


Side Station 1.2.5

Since TOMO is included by default as Standard feature, please follow the install steps
described in ‘1.2.3 TOMOS Radiography’ even TOMOS radiography option is not
installed.

1.2.1 DSA

(1) Add the following procedure for DSA. If they are already registered in the
system, then it’s not required registering same procedures again.
Refer to “10.7 Register System Procedure Tool” in DR-IM (Rev.F or later)

Procedure Name

DSA

DSA(Abd.)
DSA(XA)

DSA(Shunt)

RSM-DSA(Leg)

M506-E361G SONIALVISION G4 Installation Manual 1-17


Chapter 1 Installation

1.2.2 SLOT Radiography

.The SLOT radiography is supported in Rev.01.02.00 or later. When the


SONIALVISION G4 system is older than Rev.01.02.00, update it to Rev.01.02.00 or
later before starting work in this section.

The Variable SLOT radiography is supported in Rev.01.05.00 or later. When the system
having the SLOT radiography option is updated to Rev.01.05.00 or later, add the
radiography conditions for the Variable SLOT according to the steps below:

(1) Peel the blindfold seal from the button for the slot radiography on the X-ray
diagnostic table front control panel.
Refer to “2.2.2 X-ray Diagnostic Table Front Control Panel” in
SONIALVISION G4 IM

(2) Add the following X-Ray condition for SLOT Radiography. If they are
already registered in the system, then it’s not required registering same
conditions again. Registered conditions are added in ‘Tab3’, then rename
‘Tab3’ to ‘SLOT’.
Refer to “3.4 Radiography Preset Editor” in UD-IM “Appendix” (Rev.E or
later)

No. Name Remarks

300 SLOT HS F

301 SLOT HS L
302 SLOT HQ F

303 SLOT HQ L

306 SLOT HS F (no Grid)

307 SLOT HS L (no Grid)

308 SLOT HQ F (no Grid)

309 SLOT HQ L (no Grid)


342 VD SLOT HS F Added in Rev.01.05.00

343 VD SLOT HQ F Added in Rev.01.05.00

348 VD SLOT HS F (no Grid) Added in Rev.01.05.00


349 VD SLOT HQ F (no Grid) Added in Rev.01.05.00

1-18 SONIALVISION G4 Installation Manual M506-E361G


1.2 Install Optional Features

(3) Add the following procedure for SLOT Radiography. If they are already
registered in the system, then it’s not required registering same procedures
again.
Refer to “10.7 Register System Procedure Tool” in DR-IM (Rev.F or later)

Procedure Name

SLOT Spine

SLOT Leg

SLOT Spine (nonGrid)


SLOT Leg (nonGrid)

SLOT Long

The procedure for the Variable SLOT radiography is not provided as standard
procedure. Set up procedures for the variable SLOT radiography.
Refer to 12.2.7 "Radiography Tab" in DR-300 Instruction Manual (IA) (Rev.J or
later).

(4) Check to see the appropriate collimator open area.


Refer to “27 [Appendix 15] Adjusting Collimator open area for Slot
Radiography” in DR-IM

M506-E361G SONIALVISION G4 Installation Manual 1-19


Chapter 1 Installation

1.2.3 TOMOS Radiography

The Tomosynthesis is supported in Rev.01.03.00 or later. When the SONIALVISION


G4 system is older than Rev.01.03.00, update it to Rev.01.03.00 or later before starting
work in this section.

The low dose and high definition Tomosynthesis is supported in Rev.01.05.00 or later.
When the system having the Tomosynthesis option is updated to Rev.01.05.00 or later,
add the radiography conditions for the low dose and high definition Tomosynthesis
according to the steps in this section:

(1) [TOMO and TOMOS]


Add the following X-Ray condition for TOMO and TOMOS Radiography. If
they are already registered in the system, then it’s not required registering
same conditions again. Registered conditions are added in ‘Tab4’, then
rename ‘Tab4’ to ‘TOMO’.
Refer to “3.4 Radiography Preset Editor” in UD-IM “Appendix” (Rev.E or
later)

No. Name Remarks

400 TOMO Head F

401 TOMO Head L

402 TOMO C-Spine F


403 TOMO C-Spine L

404 TOMO L-Spine F

405 TOMO L-Spine L


406 TOMO Shoulder

407 TOMO Hand F

408 TOMO Hand L

409 TOMO Elbow

410 TOMO Knee

411 TOMO Hip Joint


412 TOMO Chest

413 TOMO Abdomen

430 LD TOMO Hand F Added in Rev.01.05.00


431 LD TOMO Knee Added in Rev.01.05.00

1-20 SONIALVISION G4 Installation Manual M506-E361G


1.2 Install Optional Features

432 LD TOMO Hip Joint Added in Rev.01.05.00

448 HD TOMO Hand F Added in Rev.01.05.00

449 HD TOMO Knee Added in Rev.01.05.00

M506-E361G SONIALVISION G4 Installation Manual 1-21


Chapter 1 Installation

(2) [TOMO and TOMOS]


Add the following procedure for TOMO and TOMOS Radiography. If they
are already registered in the system, then it’s not required registering same
procedures again.
If TOMOS Radiography is not installed in the system, do not add Procedures
under ‘Category TOMOS’.
Refer to “10.7 Register System Procedure Tool” in DR-IM (Rev.F or later)

Category Procedure Name

TOMO TOMO

TOMOS TOMOS Head


TOMOS C-Spine

TOMOS L-Spine

TOMOS Should Elb


TOMOS Hand

TOMOS Hip Joint

TOMOS Knee

TOMOS Chest

TOMOS Abdomen

The procedure for the low dose and high definition Tomosynthesis is not provided as
the standard procedure. Set up procedures for the low dose and high definition
Tomosynthesis.
Refer to 12.2.7 "Radiography Tab" in DR-300 Instruction Manual (IA) (Rev.J or later).

(3) [TOMO and TOMOS]


Perform TOMOS calibration.
Refer to 7.5 "Calibration" in DR-300 Instruction Manual (IA) (Rev.J or
later).

(4) [TOMO and TOMOS]


Check image pixel value.
Refer to “29 [Appendix 17] Checking image pixel value for TOMO /
TOMOS Radiography” in DR-IM (Rev.F or later).

(5) [TOMO and TOMOS]


Adjust TOMO and TOMOS Height of RF-Table.
Refer to “3.2.13 Adjust Tomo Height” in ZS-IM “PART2: MAINTENANCE”
(Rev.C or later).

1-22 SONIALVISION G4 Installation Manual M506-E361G


1.2 Install Optional Features

1.2.4 Iteration Reconstruction

The Iteration Reconstruction is supported in Rev.01.05.00 or later. When the


SONIALVISION G4 system is older than Rev.01.05.00, update it to Rev.01.05.00 or
later before starting work in this section.

The IR is executed in the Side Station. Install the Side Station before starting the work
in this section
Refer to 1.2.5 “Side Station” in this book.

(1) Procedure Edit


Edit settings of the procedure for the Iteration Reconstruction (IR).
Refer to 12.2.7 "Radiography Tab" in DR-300 Instruction Manual (IA)
Rev. J or later

a) Select a procedure for the IR and click [Edit]. The figure below
shows the case where the IR is applied to "TOMOS Hip Joint".

M506-E361G SONIALVISION G4 Installation Manual 1-23


Chapter 1 Installation

b) Set up the condition for IR.

Item Value
Edge E0
SIM-Link Select the checkbox.
IR Preset Select an appropriate reconstruction preset
according to the setup protocol.
*1: When "SIM-Link" checkbox is cleared,
the IR Preset is invalid.
*2: Presets starting with "IR" is for IR.

1-24 SONIALVISION G4 Installation Manual M506-E361G


1.2 Install Optional Features

1.2.5 Side Station

(1) Install the Side Station.


Refer to “2 Installation/Environmental Conditions” in SS-IM.

When the EXAM cabinet does not support the Side Station, perform the work in
M517-E754 (DR-300 Service Manual (Rewiring KIT Installation Section), and then
connect the Side Station.

(2) Configure the Side Station.

SS-IM Contents Remarks


3 Basic Configuration on Required
Side Station i3 PC
4.2 Setting the Installation Required
Site Information
4.3 Setting the System Required as appropriate
Configuration
4.4 Setting the Image List Required as appropriate
4.5 Setting the Image Required when the system has
Transmission the option below:
Destination ・DICOM transfer Option
4.6 Setting the Image Not required
Receiving
4.7 Setting the Printing Required when the system has
Destination the option below:
・DICOM Print Option
4.8 Checking the Required as appropriate
Connection
4.9 Setting the Media Required as appropriate
Storage
5 Overlay Edit Tool Required as appropriate
6 Setting the Print Required when the system has
Density the option below:
・DICOM Print Option
7 Other Tools Required as appropriate
8 Operation Check Required after installation
completed
9 Installing Speed-Up Required when the system has
Hardware the option below:
・IR Option

(3) Set up The SIM-Link with DR-300.


Refer to “8.14 Setting the SIM-Link” in DR-IM (Rev. J or later).

M506-E361G SONIALVISION G4 Installation Manual 1-25


Chapter 1 Installation

Images between the DR-300 and the Side Station is sent through the SIM-Link
high-speed image transfer, not the DICOM transfer. The DICOM transfer set up is
unnecessary.

1-26 SONIALVISION G4 Installation Manual M506-E361G


1.3 In case of connecting with CH-200 (Without Bucky Table)

1.3 In case of connecting with CH-200 (Without


Bucky Table)

Installation procedure of the SONIALVISION G4 system when connecting with Ceiling Tube
Support CH-200 and Bucky stand is described in this section.

In this connection, the emergency stop switches of ZS, CH, and Bucky Stand work
together.

Refer to Section 1.4 for installation in combination with interlocking Bucky Table.

1.3.1 Installation of CH-200

Install CH-200 by reference to CH-IM. Wiring between units is shown below;

E L0-L100
UD
CH +
A0-A200 (1) GSC-2002L
J17
power distribution

L0-L100
Signal *
(2)

ZS Buck Stand

Cable list attached to ‘565-10264 (CH-G4 Connective Kit)’


No. P/N Name Length Connection
(1) 563-75789-26 DAR ARCNET CABLE 10 MT [CH] CHC-200 [UD] COMDAR-2009
J4 J3 or J4
(2) 565-10337 ZS-CH SIGNAL CABLE 10 MT [CH] CHC-200 [ZS] ZS CONT-09
J16, J19 CN13, CN34

* Signal cable between CH and Bucky Stand


When the Bucky Stand BR-120 or BR-120T has the electric option (Foot Switch), in
addition to the cable above, the Arcnet cable should be connected.
Connect with the cable between the BRC-BASE board CN5 and the CHC200 board
J3 according to the BR MEM. The combination with manual type Bucky Stand, such
as BR-120M, does not require this cable.

M506-E361G SONIALVISION G4 Installation Manual 1-27


Chapter 1 Installation

Use cables only attached to ‘565-10264 (CH-G4 Connective Kit)’.

* The radiography signal from CH-200 has to be connected to ZS CONT-09, instead of


TERMINAL-2002. Use ‘565-10337 (ZS-CH SIGNAL CABLE)’.

* Use ‘563-75789-26 (DAR ARCNET CABLE)’ for the Arcnet communication. Arcnet
communication error will be happen when using the cable attached to CH-200.

1-28 SONIALVISION G4 Installation Manual M506-E361G


1.3 In case of connecting with CH-200 (Without Bucky Table)

1.3.2 Configuration of DIP SW and Jumper SW

(a) CH-200
CHC200 Board
JP3 : 2-3 USE DR
Dip SW 1-8 : ON
Dip SW 3-3 : ON
Dip SW 3-4 : OFF (When BR-120/BR-120T has an electric option, ON)

PERIP-SH / PERIP-SH2 Board


Dip SW 2-1: ON

Configure each board based on the Bucky Stand connected as shown below;
Stand
CH-200
BR-120 / BR-120T Other
Configuration BRC120 BRC-BASE BRC120
CHC200 Board
Board Board Board
JP4 SW6 JP1 JP5 JP1
No Bucky Stand 1-2 OFF
BR-120 / BR-120T 2-3 OFF 2-3 2-3
Stand by others 2-3 OFF 2-3

Select ‘No Bucky Stand’ configuration in the table above when connecting with manual
type stand, such as BR-120M.

(b) ZS-200
ZS CONT-09 Board
JP6 [CH(BR)] : 1-2 VALID

M506-E361G SONIALVISION G4 Installation Manual 1-29


Chapter 1 Installation

1.3.3 Setting

(a) CH-200
Confirm system configuration of Bucky Table in CH-200 that ‘Table Type’ is set as
‘OFF’.
Refer to “11.4.2 Setting of the radiography table” in CH-IM “Section 1:
INSTALLATION”

Do not set the ZS-200 for the Table type.

Confirm system configuration of Bucky Stand in CH-200 initial setting window.


Set up it according to the type of stand connected.
Refer to “11.4.3 Setting of the radiography stand” in CH-IM “Section 1:
INSTALLATION”

(b) D150BC-40S
Set CH-200 #2 as ‘Used’ for the arcnet communication with CH-200.
Refer to “4.8 Communication Settings [Setup] – [Communication]” in UD-IM
“Part1: Installation”

1-30 SONIALVISION G4 Installation Manual M506-E361G


1.4 In case of connecting with CH-200 (With Bucky Table)

1.4 In case of connecting with CH-200 (With Bucky


Table)

Installation procedure of the SONIALVISION G4 system when connecting with Ceiling Tube
Support CH-200, Buck table, and Bucky stand is described in this section.
When connecting with manual type stand, such as BR-120M, refer to section 1.3.

In this connection, the emergency stop switches of CH, Bucky Stand and Bucky Table
work together. The emergency stop switch of ZS does not interlocked with that of CH,
Bucky Stand and Bucky Table.

M506-E361G SONIALVISION G4 Installation Manual 1-31


Chapter 1 Installation

1.4.1 Installation of CH-200

Install CH-200 by reference to CH-IM. Wiring between units is shown below;

E Bucky Table

Signal *
L0-L100
power distribution

J16
L0-L100 UD
E Signal
CH J25 +
A0-A200
(1) GSC-2002L
J17

L0-L100
Signal ※
ZS

E Bucky Stand

Cable list attached to ‘565-10264 (CH-G4 Connective Kit)’


No. P/N Name Length Connection
(1) 563-75789-26 DAR ARCNET CABLE 10 MT [CH] CHC200 [UD] COMDAR-2009
J4 J3 or J4
(2) 565-10337 ZS-CH SIGNAL CABLE 10 MT Not Used

* Signal cable between CH and Bucky Stand


When the Bucky Stand, BR-120 or BR-120T, has the electric option (Foot Switch),
in addition to the cable above, the Arcnet cable should be connected.
Connect with the cable between the BRC-BASE board CN5 and the CHC200 board
J3 according to the BR Installation Manual. The combination with manual type
Bucky Stand, such as BR-120M, does not require this cable.

* The radiography signal from CH-200 has to be connected to TERMINAL-2002 board


J6 CN9 using ‘X-RAY I/F Cable B’ attached to the CH-200.

* Use ‘563-75789-26 (DAR ARCNET CABLE)’ included in ‘565-10264 (CH-G4


Connective Kit)” for the arcnet communication. Arcnet communication error will be
happen when using the cable attached to CH-200.

1-32 SONIALVISION G4 Installation Manual M506-E361G


1.4 In case of connecting with CH-200 (With Bucky Table)

1.4.2 Configuration of DIP SW and Jumper SW

(a) CH-200
CHC200 Board
JP3 : 2-3 USE DR
Dip SW 1-8 : ON
Dip SW 3-3 : ON
Dip SW 3-4 : OFF (When BR-120/BR-120T has an electric option, ON)

PERIP-SH / PERIP-SH2 Board


Dip SW 2-1 : ON

Configure each board based on the Bucky Stand and Table connected as shown
below;
Stand Table
CH-200
BR-120 / BR-120T Other BK-120 / BK-200 Other
BRC-
Configuration BRC120 BRC120 BRC120
CHC200 Board BASE BKC 300 Board
Board Board Board
Board
JP4 SW6 JP1 JP5 JP1 JP1 JP2 SW7-1 SW7-2 JP1
BR-120 / BR-120T +
2-3 OFF 2-3 2-3 / 1-2 1-2 ON ON /
BK-120 / BK-200
BR-120 / BR-120T
2-3 ON 2-3 2-3 / / / / / 1-2
Table by others
Stand by others +
2-3 OFF / / 2-3 1-2 1-2 ON ON /
BK-120 / BK-200
Stand by others +
2-3 ON / / 2-3 / / / / 1-2
Table by others
BK-120 / BK-200 only 1-2 OFF / / / 1-2 1-2 ON ON /
Table by others only 1-2 ON / / / / / / / 1-2

(b) ZS-200
ZS CONT-09 Board
JP6 [CH(BR)]: 2-3 INVALID

(c) D150BC-40S
TERMINAL-2002 Board
Dip SW1 (all switches) : OFF
Dip SW2 (all switches) : ON

M506-E361G SONIALVISION G4 Installation Manual 1-33


Chapter 1 Installation

1.4.3 Setting

(a) CH-200

Confirm system configuration of Bucky Table in CH-200 initial setting window.


Set up it according to the type of table connected.
Refer to “11.4.2 Setting of the radiography table” in CH-IM “Section 1: Installation”

Do not select [ZS-200] for the Table type.

Confirm system configuration of Bucky Stand in CH-200 initial setting window.


Set up it according to the type of stand connected.

Refer to “11.4.3 Setting of the radiography stand” in CH-IM “Section 1:


Installation”

(b) D150BC-40S
Set CH-200 #2 as ‘Used’ for the arcnet communication between CH-200 with the
UD maintenance tool.
Refer to “4.8 Communication Settings [Setup] – [Communication]” in UD-IM
“Part1: Installation”

1-34 SONIALVISION G4 Installation Manual M506-E361G


Chapter 2
Appendix

Chapter Contents

2.1 Troubleshooting

M506-E361G SONIALVISION G4 Installation Manual


Chapter 2 Appendix

2.1 Troubleshooting

This section describes solutions for expected installation problems.


No. Problem Solutions
1 Error E1-34 (Collimator Communication error between UD, ZS, and R-300.
Communication Error) occurs at Check points:
the system startup. ・ ZS
- Setting DIP SW on NEX SH PCB
"4.1 About communication" in ZS-IM "Section 2 :
MAINTENANCE"
・ UD
2 The collimator cannot be - Setting DIP SW on COMDAR PCB
operated from the remote "3.13 Board Switch Settings and Jumper Pin
console. Settings" in UD-IM "Installation"
- Archnet communication setting
"4.8 Communication Settings [Setup] -
[Communication]" in UD-IM "Installation"

3 The origin point of the AC servo The information on AC servo origin point is held by the battery
motor disappeared by attaching connected to the ZS CONT-09 PCB. Removing the connector between
or detaching the ZS connector. the ZS CONT-09 PCB and each motor will cause the information lost.
Register the origin point again. Replace the battery while the system is
turned ON
"2.5.2 Encoder setting for position detection" in ZS-IM
"Section 2 : MAINTENANCE"
4 The X-ray irradiation clearance Press the fluoroscopy switch or the Exposure switch with holding down
button is not lit when it is the X-ray irradiation clearance button for the X-ray irradiation in
pressed in the installation mode the installation mode.
of ZS.
5 How to execute the FPD Not available.
calibration before attaching the In order to perform the FPD calibration, communication between ZS
collimator. and R-300 must be established.
6 When the breaker in the DR When the DR EXAM cabinet is turned off, the calibration data inside
EXAM cabinet is turned ON, the FPD will be lost.
fluoroscopy and radiography When the system is turned on, calibration data will be written in FPD
fails. from REF PC. During this period, fluoroscopy and radiography cannot
<SFD-1717AF type FPD only> be performed. The ACQ monitor shown "Detector: Warming up" and
the REF monitor shows "Starting study is impossible" on the upper
right of the screen. It takes from 15 minutes to 20 minutes.
7 The leaf of virtual collimation is The collimation will be normal when the gap between them is smaller
displayed inside that of the than about 7 mm on the monitor.
actual collimator. When the gap is larger than it, measure SID and adjust again.
"2.5.1 Potentiometer calibration" in ZS-IM "Section 2 :
MAINTENANCE"
8 The imaging unit going down Check points:
with the stand or reverse ・ Check backlash of the table head side rail
position makes abnormal ・ Check installation of the column cover
vibration noise.
9 Fluoroscopy fails after closing The fluoroscopy table in ACQ PC may not match with that inside UD.
the UD maintenance tool and The two tables are checked at the time of the system start. If any
restarting the system. abnormalities is detected, "Fluo Table Check: FAIL" is outputted to UD

2-2 SONIALVISION G4 Installation Manual M506-E361G


2.1 Troubleshooting

No. Problem Solutions


10 The S/W of the touch panel Terminal.
crashed at the time of system Execute "Adjustment data file sending program" again with REF PC to
restarting after closing the UD match these tables. If the matching keeps failing, copy the file of the
maintenance tool. fluoroscopy table to ACQ PC from REF PC using the USB memory.
"3.5 Adjusted Data File Sending Program" in UD-IM
"Appendix"
11 Changing the value of PHV-DR The radiography conditions including acquisition mode, resolution,
in adjustment of the photo timer acquisition rate, and FOV have the same setting between the DR side
does not affect the exposure and the UD maintenance tool.
time. Check that the condition change by the DR side (REF, Touch panel,
Remote console) accompanies that with the UD maintenance tool, and
then start adjustment.
"4.21.5 Adjusting PHV-DR [Adjustment] - [PHV-DR]" in
UD-IM "Installation"
12 Error E5-39 (Dosimeter Check points:
Communication Error) occurs at ・ Connection between the Dosimeter and the CLC12 PCB
the system startup. "6.6 Area Dosimeter Adapter (For VACUTEC)" in
DR-300-IM
・ R-300 option setting
"2.7.2 Option registration" in ZS-IM "Section 2 :
MAINTENANCE"
13. Error E5-38 (System This error is displayed when communication between UD and DR-300
Communication Defect) is TOUCH has been lost 30 seconds or more.
displayed on the system Since the shutdown of DR-300 precedes that of UD/ZS in the system
shutdown. shutdown, there is no problem even though this error is displayed.
* This problem is corrected in Revision.01.02.00 or later.
14 Communication fails even when Since the serial port of UD CONT-2002 PCB has been connected to
the serial cable is connected to COM1 of MPC, the connector J15 cannot be used. Use MPC for
the D-SUB connector J15 of the adjustment.
UD CONT-2002 PCB. "2.1 Connection" in UD-IM "Maintenance"
15 Neither the touch-panel Check points:
operation sound nor fluoroscopy ・ Connection of the audio cable
sound is not herd from the "3.6.2 Connecting the DR-EXAM Cabinet" in DR-IM
remote console (large local ・ Rotational resistance of the SonialRemote-2009 PCB
console). Remote console: VR2
Large local console: VR1
・ Sound-volume setting of DR Maintenance Tool
"8.17.1 Sound" in DR-IM
(Revision A or later. SYSTEM S/W Revision.01.01.00 or
later)
16 How to output the fluoroscopy The "AUX balance" of the adjusting tool of the remote console allows
sound and the touch-panel the setting of output sound volume to the speaker in the examination
operation sound to the speaker room.
in the examination room? "1.3.7 Adjusting the Sound" in UD-IM "Appendix"
17 The screen of the ACQ monitor Set up the screen on ACQ PC.
is replaced with that of the touch "[Appendix 2] Troubleshooting for Monitor Mismatch of
panel. ACQ Application and Touch Panel Application" in DR-IM "Section
18 Operating the touch panel 14"
moves the cursor on the ACQ
monitor.

M506-E361G SONIALVISION G4 Installation Manual 2-3


Chapter 2 Appendix

No. Problem Solutions


19 The touch panel of the large Check points:
local console shows no ・ Connection of the touch-panel cable
response. "3.6.2 Connecting the DR-EXAM Cabinet" in DR-IM
・ Setting of the touch panel
"23 [Appendix 11] How to Set Touch Panel with Large
Local Console Connected" in DR-IM
・ Calibration of the touch panel
"22 [Appendix 10] How to Set the Driver for Touch
Panel" in DR-IM
20 How to display the cross cursor "15 [Appendix 3] Showing the Cross Cursor on the ACQ-PC
on the fluoroscopy image for the Monitor" in DR-IM
alignment of the X-ray tube and
the collimator?
21 How to display the SMPTE "16 [Appendix 4] Showing the SMPTE Pattern" in DR-IM
pattern on the ACQ monitor?
22 The maintenance tool of ZS and The Microsoft .NET Framework 2.0 SP2 needs to be installed.
UD does not start with the "Appendix 2 Installation and setting of maintenance tool"
notebook PC. in ZS-IM "Section 2 : MAINTENANCE"
"3.3 UD Maintenance Tool" in UD-IM "Appendix"
23 How to edit radiography X-ray The user-editable X-ray condition is provided in Rev. 01.06.00 or later.
condition preset by users. "4.15 Editing Radiography Program" in SONIALVISION G4
Instruction Manual
24 FPD calibration failed with error Make the changes based on error code as follows;
0x0402 or 0x0403. In case of 0x0402:
This error indicates there might be some object within FOV or
X-ray Tube Voltage might be low. If this error happens again after
checking no object within FOV, then increase X-ray Tube Voltage
by 3kV and retry FPD calibration.
In case of 0x0403:
This error indicates X-ray Tube Voltage might be high. Decrease
X-ray Tube Voltage by 3kV and retry FPD calibration.
"8.14.3 Setting the Calibration Mode" in DR-IM
25 FPD calibration failed with error This error indicates there might be something wrong with tabletop
0x3003. control by RF Table. Additional error information will be displayed on
Remote Console, then please make a proper action based on the
message.
"5.3.1 Message Displayed on Remote Console" in
SONIALVISION G4 Instruction Manual
26 FPD calibration failed with error This error indicates the examination room door was opened during
0x3004. FPD calibration. If this error happens even the door is closed, then
please check the following setting of UD.
UD Maintenance Tool [Setup] – [Signal] – [DAR I/F OUT-2]
Set “10” to No.8
"4.10 Signal Input/Output Settings [Setup] – [Signal]" in
UD-IM

2-4 SONIALVISION G4 Installation Manual M506-E361G