CI-B5 BOD TESTER

Applications
This instrument is to test BOD value in surface water, sewage, industrial wastewater especially
used to measure waste loadings at wastewater treatment plants and to examine the efficiency of
wastewater treatment.
Principle
Designed by the respirometric method, CI-B5 BOD Tester simulates the degradation process of
organics in the nature. Viz. the air in the bottle above the sample contains 21% oxygen and
replenishes the dissolved oxygen used by the bacteria. Carbon dioxide is made when
microorganisms oxidize organic matter in the sample. The carbon dioxide must be removed by
the potassium hydroxide from the system so it does not interfere with the measurement. Pressure
sensors monitor air pressure in the sample bottles. The pressure drop correlates directly to BOD
and BOD is correspondence with the quantity of oxygen consumed by bacteria. The instrument
will deal with this correlation and then the BOD value will be displayed on the screen.
Technical Specifications
1. Accuracy: ±5% (full range)
2. Sample range: liquid sample, including dissoluble organic sample, e.g. released sewage and
river water
3. Measurement range: 0 to 35, 0 to 70, 0 to 350, 0 to 700mg/L (no need to dilute)
4. 700 to 4000 mg/L (need to dilute)
5. Sample quantity: ≤6
6. Volume: 500 ml
7. Repeatability: ≤5%
8. Resolution ratio: 0.1mg/L (0 to 35, 0 to 70mg/L)
1mg/L (0 to 350, 0 to 700mg/L)
9. Display: LCD
10. Power source: AC 220, 50Hz
11. Dimensions: 35×40×35cm

Key Features
1. Conforms to GB7488-87
2. Display the trend curve
3. Printer output
4. Link to the compute through USB interface
5. Mercury-free devices to ensure the accuracy and safety of the BOD measurement
6. Real-time clock display
7. Automatic collection and storage of data
8. Depict BOD curves for the user to view
9. Link to the computer to print curves
10. No parameter and data loss after suddenly electricity cut-off
11. LCD humanization interface display, easy to operation
12. 1-30days of the test time, setting by the keys
13. Store data and update the result, observe the change rate of BOD in the LCD screen during
any testing time, store 240 data points and 20 groups data of each sample.