User’s Manual
ordinary equipment, which is Class I Type B and protection against electric shock.
The manufacturer accepts the responsibility for security, reliability and performance
authorized personnel;
Electric utilities of the machine room must be complied with related regulations;
graphic symbols".
The packaged products should be adapted to transport by air, rail, ship and tertiary
highways.
• Articles weighs more that 30kg should be not allow storage on the top of
packing case.
Dear customer:
Firstly, we would express our gratitude to you and your company to select our
products!
Be sure to read the Product Manual carefully before operating the equipment,
especially the following safety precautions, in order to operate the equipment
correctly for you.
● Please read the following precautions carefully, in order to operate the equipment
correctly for you.
● Be sure to get fully known about the performance and operation of the equipment
before operating the equipment.
● Pay attention to the following matters for setup and storage location of the
equipment:
Equipment not suitable for use or store in presence of inflammable and
explosives;
Equipment not suitable for use or store in presence of temperature, humidity and
atmospheric pressure exceeding permitted ranges;
Equipment not suitable for use or store in presence of ingress of water;
Avoid tilt, shake and shock (including transportation);
Pay attention to power supply frequency and voltage as well as allowable
current;
Keep ventilation and avoid direct sunlight.
● Pay attention to the following matters before using the equipment:
Electromagnetic emission and electromagnetic immunity of the equipment
should meet requirements specified in YY0505-2012(idt IEC 60601).The
equipment should be operated in typical health care environment specified in
YY0505-2012(idt IEC 60601);
Confirm power supply connection with the equipment and check whether the
equipment is operated normally;
Confirm ground connection of the equipment well;
Reconfirm the power supply voltage.
In case of equipment failure, instead of disassembling, please paste a note
written with “Break down!” on the equipment, and the Company will assign
authorized personnel to eliminate the failure.
User’s Manual V2.11.1E
The following warning symbols used in this manual. Their application and indication
are described below.
The following security identity symbols are used in the equipment. Their indications
are described below.
Symbol Description
Ionizing radiation
Dangerous voltage
Earth-fault protection
Table of Contents
CHAPTER1 OVERVIEW .................................................................................. 1
CHAPTER1 OVERVIEW
Used to radiography patients and obtain single image for clinical diagnosis.
Contraindications: incompatible patients, pregnant women and lactation women,
children, infants, and critically ill patients are not suitable.
Figure 1-1
1
User’s Manual V2.11.1E
Main features:
1. Diagnostic Table
Visual display man-machine interface and touch screen control makes
man-machine dialogue more direct-viewing, convenient and easy to understand.
Programmable control device technology with high control precision,
good stability and excellent anti-interference performance.
Frequency converter control technology, the table operates with soft
startup and soft shutdown, which is smooth without noise and the position, is
more accurate.
Reliable angular displacement sensor technology with excellent
precision, little dynamic noise and long mechanical service life.
Tilt range: 90° ~ 0° ~ -25°, it can carry out conventional examinations
about alimentary organs, and it also can be used for a variety of special
examinations.
Humanized design of diagnostic table side switch, which is more
convenient for positioning when operating local the console.
FPD and Toshiba high-quality image intensifier, which further improves
image clarity and contrast ratio, and makes it much easier and clearer to identify
the nidus.
The moving range of spot-film device is wide, so we adopt the operating
mode with machine moving but people unmoving, which thus simplifies the
implementation of a series of inspections from the throat and esophagus to the
lower abdomen.
2. Generator
Generator with X-ray radiography and fluoroscopy functions and its
maximum output power is 65.5 kW.
Output voltage is 150kV.
Smaller and lighter block design
During exposure, you can adjust kV and mA buttons to achieve
constant X-ray output.
Large screen LCD flat panel display, which is used to display APR
conditions and practical conventional items.
User-friendly operating system configuration.
2
User’s Manual V2.11.1E
3
User’s Manual V2.11.1E
4
User’s Manual V2.11.1E
5
User’s Manual V2.11.1E
Radiography System
YY/T0742-2021 Particular Specifications for Gastrointestinal
Diagnostic X-ray Equipment
YY/T0743-2009 Particular Specifications for Gastrointestinal
Diagnostic Table
YY 0505-2012 Medical Electrical Equipment
part 1-2: General Requirements for Safety-
parallel standards:
electromagnetic compatibility requirements and
tests
strictly according to provisions of the Manual and cannot be used for other
purpose except medical radiography. This equipment can only be used by trained
personnel with needed radiation knowledge. Special attention will be paid on the
following matters during using:
7
User’s Manual V2.11.1E
8
User’s Manual V2.11.1E
When the equipment is retired, the X-ray tube assembly and transformer oil in high
voltage generator shall be reclaimed by specialized industrial waste and waster oil
collection station; while other steel and cooper materials shall be reclaimed by
corresponding department. In particular, the equipment contains the following materials
and/or components:
“Reducers”: lubricating oil, steel, aluminium
Motor: iron, copper, hard plastic material casings
Electric cabinet: iron, aluminium, copper, non-biodegradable plastics,
vetronite for printed circuits.
The equipment life-cycle is 10 years under the condition of correct using, and the
X-ray tube life-cycle is about 100,000 times, the damaged X-ray tube can be sented to our
company for replacement. If the equipment is damaged and it is beyond repair after
correct using for 10 years, the equipment shall be retired.
Symbol Description:
Dangerous Ionizing
Radiography
voltage radiation
Small Focal Type-B Protective
spot symbol equipment grounding
Talkback Emergency
Fluoroscopy
conversion stop
9
User’s Manual V2.11.1E
Figure 1-4
Figure 1-5
Figure 1-6
10
User’s Manual V2.11.1E
Figure 1-7
Figure 1-8
6.Nameplate of X-ray tube(see the picture below):
Figure 1-9
11
User’s Manual V2.11.1E
Figure 1-10
Figure 1-11
Figure 1-12
12
User’s Manual V2.11.1E
13
User’s Manual V2.11.1E
Attention:
● X-ray tube assembly equipped to this equipment (hereinafter referred to
as assembly) only applies to medical Radiography.
● Pay attention to that load exceeding the stipulated condition is not
allowed, because overload operation may result in damage to tube sleeve and
X-ray tube.
● The assembly is equipped with normally-closed temperature switch,
which will start when wall temperature of the assembly reaches 75°C.
● The assembly will generate X-ray while working under high voltage;
therefore attention shall be paid to protect it during operation and adjustment.
● Using the assembly in the environment with inflammable or poisonous
gas is not allowed.
● The operation environment temperature is 10℃ ~ 40℃, relative
humidity is 30%~85% and the atmospheric pressure is (70~106)kPa.
● As for transportation and storage environment conditions, the
14
User’s Manual V2.11.1E
temperature range is -20℃ ~ 70℃, the relative humidity is 20%~90% and the
atmospheric pressure is (50~106)kPa.
● Inside the assembly, there’s glass X-ray tube mounted, hence it shall be
handled gently during transportation and installation to avoid damage to X-ray
tube.
● As per the national regulations, installation, adjustment and using of the
assembly shall be carried out by personnel with professional qualification
certificates.
1. X-ray tube assembly type: E7252X
2. X-ray tube component level: Class I Type B
3. X-ray tube parameters:
Model of X-ray tube: XH-106V
Target construction: Rhenium-tungsten faced molybdenum
The reference axis involved in the target Angle and focus characteristics of the X ray
tube: a straight line perpendicular to the axis line of the X ray tube and passing through
the window marker
Anode angle: 12º
Nominal focal spot value: Large focus: 1.2mm Small focus: 0.6mm
Nominal X-ray tube voltage: Fluoroscopic: 125kV Radiographic: 150kV
Nominal anode input power(at 0.1s): Large focus: 75kW Small focus: 27kW
Anode heat content: 210KJ (300kHU)
Filament voltage:
Large focus(5.5A)12.7 ~ 17.1V
Small focus(5.2A)7.0 ~ 9.4V
Filament frequency limits: 0~25 kHz
Permanent filtration: 0.9mm Al/75kV
Anode speed of X-ray tube: 180Hz: Minimum 9700 min-1
Maximum anode heat dissipation: 475W(667HU/s)
15
User’s Manual V2.11.1E
4. Motor Ratings:
Project Starting Running
Driven frequency (Hz) 180 180
Input power (W) 2300 300
Voltage (V) 460 130
Current (A) 5.4 2.0
Min. Speed Up (s) 1.0 —
Capacitor (μF) 3 3
Min. Braking (s) 1.5/90V(DC)
16
User’s Manual V2.11.1E
17
User’s Manual V2.11.1E
18
User’s Manual V2.11.1E
20. See figure 2-5 for the wiring of the anode plate of the component
C :Common
L :Large focus
S :Small focus
19
User’s Manual V2.11.1E
2.4 Collimator
1.Exposure field:
a.shape of irradiation field: rectangle
b.minimum irradiation field: 0×0 (when the blades coincide)
c.maximum radiation field: 430 mm×430 mm (at SID 1000 mm)
2.Visible light irradiation brightness : Average central illumination
brightness:>100Lux
3.Lamp:LED,d.c. 3.6V,5W
4.Lighting time:5~45s,5s at each step
5.Filtration:≥1.0mmAl;Additional filter:3.5 mmAl
6.Blade driving mode:Remote button operation and setting closed-loop control,
local forward operation button electric
7.Overall dimension:225mm×262mm×210mm(length×width×height)
8.Weight:about 9.0kg
9.Power supply: DC24V
20
User’s Manual V2.11.1E
2、Tablet parameters:
A / D conversion: 16 bit
0 lp/mm ≥62%
0.5 lp/mm ≥56%
1.0 lp/mm ≥50%
1.5 lp/mm ≥45%
2.0 lp/mm ≥40%
5/2µGy 2.5 lp/mm ≥35%
3.0 lp/mm ≥30%
3.5 lp/mm ≥25%
4.0 lp/mm ≥21%
4.5 lp/mm ≥16%
5.0 lp/mm ≥11%
21
User’s Manual V2.11.1E
Frame rate:
≥2fps (1x1binning, 3072 x3072, 43cmx43cm
FOV)
≥6fps (1x1 binning, 1536x1536, 23cmx23cm
FOV)
≥13fps (2x2 binning, 1408x1408, 43cmx43cm
FOV)
≥20fps (1x1 binning, 1024x1024,
15.6cmx15.6cm FOV)
≥30fps (2x2 binning, 768x768,
23.7cmx23.7cm FOV)
Modulation transfer function:
1lp/mm ≥65%
2lp/mm ≥35%
3lp/mm ≥20%
4lp/mm ≥10%
5lp/mm ≥7%
Power: DC 24V
Weight: 6 Kg
4、Environment conditions
temperature: 10℃ ~ 40 ℃
Relative humidity: 30 % ~ 75 % RH
22
User’s Manual V2.11.1E
5、Storage conditions
temperature: -20 ℃ ~ 55 ℃
Focus mA kV s
10、12.5、16、20、
150 10
25、32
150 8
40
119 10
150 4
50
109 10
Small focus
150 1.6
63
89 10
150 0.63
80
69 10
150 0.2
100
59 10
150 2
125
69 8
150 1
160
79 5
150 0.63
200
59 5
150 0.2
250
59 3.2
Large focus
320 150 0.4
24
User’s Manual V2.11.1E
69 2.5
150 0.25
400
59 2.5
131 0.16
500
49 2
104 0.125
630
49 1.25
81 0.16
800
49 0.8
25
User’s Manual V2.11.1E
CHAPTER3 PRE-INSTALLATION
The foundation shall adopt cement with grade no less than 400"
and casting thickness no less than 200mm.
f) Cable trench and cable trench cover shall be installed flush with the ground;
g) Room should be equipped with special grounding line, grounding resistance is
not greater than 4 Ω;
h)The X-ray protection ability of the machine room shall meet the requirements
of national regulations.
26
User’s Manual V2.11.1E
1. Power condition:
a) Power supply voltage: a.c.380V±38V, 50/60Hz±1Hz;
b) The power resistance should be no greater than 0.17 Ω;
c) Power supply capacity shall not be less than 105kVA.
2. The power distribution box shall be equipped with a three-phase air switch with
rated voltage of 500V and rated current of 130A as the disconnection switch with
the power supply network. The power distribution box shall be firmly fixed on the
wall with a floor level of not less than 1.5m away from the room, and a power
source with a power capacity of not less than 105kVA and an independent ground
wire (not more than 4 Ω).
27
User’s Manual V2.11.1E
28
User’s Manual V2.11.1E
4.1 Overview
The equipment must be installed and adjusted by professional personnel of our
Company or authorized by our Company.
Before opening the packing case, it shall be stored in the rain-proof and dust-free
environment; and board of box shall be opened reasonably while opening the case.
While handling the parts, attention shall be paid to avoid scratching parts surface;
check according to the packing list after opening the case and installation shall only
be conducted after confirming the machine is defect and omission; any problem shall
be informed to the Company for us to dispose in time.
29
User’s Manual V2.11.1E
1) Install the diagnostic table, high voltage generator and console panel
according to the layout requirements of the machine room, and connect the
wiring according to the circuit diagram.
2) Put through power supply, the table body tilt to 90 °, with X-ray tube column.
3) Installation of the Tube-Collimator combination
30
User’s Manual V2.11.1E
the fixed collar connection plate with four hexagon socket head screw M6X25;
2. Nsert the tube connection plate into the mounting hole of the cross arm of the
tube, and fix it with three M10x25 inner hexagon screws. Pay attention to the
direction of installing the tube (the direction shown in the figure, with the cathode
facing up).
3. Stack the subplate of collimator, tube mounting plate and connecting plate
together, and connect to the tube with cross recessed countersunk head screw
M6x25
4. After the collimator adapting piece is installed,back-out thecollimator with
Hexagon socket head screwM5×40,not fall off..Install collimator ,rotate the
collimator,align the notch with the notch of the connecting plate, then rotate the
collimator to the appropriate position and tighten the M5x40 socket hex screw.
5. Install the fan on the tube housing and secure the fan to the tube housing using
four m4x10 hex socket screws.
6. Install the outer cover of the tube, insert the T-shaped slot of the outer cover into
the notch of the mounting plate of the tube, and connect the outer cover of the
tube with the transverse arm of the tube by using two inner hexagon screws
M4x15
32
User’s Manual V2.11.1E
Wall gate wiring, must ensure that the switch is in the open
state, unless the installation wiring has been completed,
otherwise do not allow closing.
2.USB Cable:
Computer CAN-USB converter
USB Cable
Plug Plug
USB —— —— USB
34
User’s Manual V2.11.1E
4. Stator cable B:
35
User’s Manual V2.11.1E
36
User’s Manual V2.11.1E
37
User’s Manual V2.11.1E
6. AEC Cable B:
Generator AEC Cable B Ion Chamber
Interface board (XJZ50.84) plug
PT8 —— —— plug
38
User’s Manual V2.11.1E
The base of the diagnostic table is fixed, the X-ray tube Collimator is installed
in the center, the image intensifier is fixed, the image intensifier TV system is
installed accurately, and the cable connection of the equipment is confirmed to be
accurate.
39
User’s Manual V2.11.1E
Description:
1:Manual and automatic switch button of the collimator
2:compressor tracking button
3:Table negative Angle button
4:tilt down button
5:tilt up button
6:column up button
7:column down button
8、9:Subfunction switch button(Only used in debug mode)
1.Debugging Need to Know
Limit Debugging
Enter working state,cover the corresponding optocoupler with a visor or hold the
corresponding stroke switch limit by hand ,at the same time, check whether the
corresponding LED lamp on the PCB board is on or off normally(General cover limit
or press the travel switch limit corresponding to the LED off, release the visor or
release the hand LED light ).Be sure to ensure that each limit is working properly
before the next step of debugging.
40
User’s Manual V2.11.1E
Potentiometer Debugging
During debugging, the potentiometer is released from the gear;The AD value
increases when the table up,The AD value decreases when the table down.The AD
value increases when the column up,The AD value decreases when the column
down.The potentiometer AD value can be measured with a multimeter,The voltage of
4%~96% VCC is normal or the range of 32~990 displayed on the control panel is
normal. If the range is beyond this, the potentiometer is abnormal and the parameters
cannot be saved. In normal mode, the corresponding error code of the potentiometer is
displayed.
box,then press button to starting up the machine,If there is no error, the digital
tube shows C01, indicating that the debugging function has been entered,press
button switch between C01 and C19.If there are sub-functions in the main
function, the operation handle will switch back to the sub-function switch key, and the
sub-function code will be switched from 01 to 99 (according to the actual need) .
41
User’s Manual V2.11.1E
The description of main function code and sub-function code is shown in Table 2:
Table 2
Main function Sub-function Function button Description of function button
code code
table down
table up
SID increase
C02
01 Max SID,AD value is saved
Column down
Column up
C04
01 Save the maximum AD value
when the column up
02 Save the minimum AD value
when the column down
03 Column protection AD value
save
Enter the show
42
User’s Manual V2.11.1E
Tabletop relocation
Compressor up
Pedal corotation
43
User’s Manual V2.11.1E
Note:before the table tilt up and down debugging,make sure the emergency
stop works properly.In the debugging state, the table has no automatic function.
When the table rotates, be sure to moves the column to the middle position to
prevent collision during the table movesment.
1.Table up and down travel limit debugging:The photoelectric coupler is used as
the limit protection device for thetable, and the potentiometer is used as the limit
device. It is installed on the optical coupling control in the base assembly, as shown in
Figure 5-2.
,LED14, LED17 bright,LED16 off,table down , The handle of the control box
3.table up and down travel limit debugging:The table is inverted using potentiometer
as a limiting device,as shown in Figure 5-2. The potentiometer must be released from
gear before debugging. Make sure the potentiometer is installed after adjustment.
value; Switch to "02", adjust the table to 0° level, press the button to save the
corresponding current AD value; Switch to "04", set the table to -25°, press the
button to save the corresponding current AD value; Switch to the negative protection
position (about -7°) of table 03, and press the button to save the corresponding
current AD value.
After the completion of debugging, the test of the inverted function should
be carried out: special attention, before testing the inverted action, first confirm
that the inverted emergency stop can work normally.In the debugging state,
there is no automatic function. When the table rotates, pay attention to prevent
collision during the table movesment.
45
User’s Manual V2.11.1E
press the table up and down button on the near panel or ,control the table
up and down, clockwise can reach +90°, rotating to the limit or let go immediately
stop; Counterclockwise direction can reach -25°, limit or release immediately stop, if
from above the level of counterclockwise rotation, to the level will stop, to continue to
the negative angle of the table , there are two operations : One is controlled by the
control box, the other is controlled by the buttons on the near control board. When
operating through the control box, it is necessary to release the operation handle first,
then press and hold the table negative angle function button on the control box,
again operate the operation handle on the control box to reverse , the table
movess to negative angle, release the button or operation handle, or the table
reaches the negative angle limit, the table negative angle motion stops. When using
the near control board, be sure to release the operation button first, press the near
control board button again, the table movess to the negative angle, release the
button or the table to the negative angle limit, the negative angle table movesment
stops. If the tablestarts to rotate clockwise at a negative angle and stops when it is
horizontal, if you want to continue to moves to +90°, you must release the operation
button or handle first, and then start the movesment operation, you can continue to
rotate to +90°, release the operation button or handle or to +90° limit, and the
tablestarts to stop.
Finally, adjust the position of the photoelectric coupler of each angle of the table
to meet the needs of the table rotation angle.
1.Debugging of the travel limit of the x-ray tube column : The longitudinal
movesment of the tube column uses a travel switch as a limit protection device,As a
limiting device, the potentiometer is installed on the travel switch and potentiometer
in the table assembly, as shown in Figure 5-3, 5-4 .
In the debugging mode "C04", hold the tube column down travel switch limit,
and then check whether the corresponding LED6 light on the PCB is off. When you
46
User’s Manual V2.11.1E
release the hand, LED6 light on the PCB is on. If yes, it is normal; otherwise, check
the corresponding line and the corresponding travel switch until normal. Similarly,
test the limit of the column upward travel switch and check LED8 lamp.
operation handle of the control box on the operation console moves to , LED15
and LED17 are on, LED16 is off, the column movess down, the operation handle
moves to , LED15 and LED16 are on, LED17 is off, the column movess up. If
the tube column movess in the opposite direction, it is necessary to adjust any two
input drive lines of the motor, and confirm that the limit collision block is installed.
3. Tube column moving potentiometer debugging
1-Tube-column potentiometer
Figure 5-4 Tube-column longitudinal shift potentiometer
Before the potentiometer is debuged, loosen and disengage the gear; Install it
after debugging.
After the debugging of limit and motor direction is completed, switch "01", "02"
47
User’s Manual V2.11.1E
and "03" respectively by the reverse sub-function switch key of the control box
operation handle on the operation control. moves the tube column to the middle
position of the table, adjust the tube column moving potentiometer voltage to 2.5V or
view the console display "512", install the potentiometer . The handle of the control
box on the control console moves to ,the column up,AD value increase.When the
column stops moving to the upper limit, switch to 01 and press the button on the
control box, save the maximum AD value of the tube column.The handle of the
control box on the control console moves to ,the column down,AD value
decrease,when moving to the lower limit, the column stops moving. Switch to 02 and
press the button on the control box , save the minimum AD value of the tube
column.The handle of the control box on the control console moves to ,moves
the column down to the lower limit and press the button on the control box to
switch,enter"C01”,The handle of the control box on the control console moves
to ,swing the table to a negative angle,press the button on the control box
to switch,return “C04”,The handle of the control box on the control console moves
to ,moves the column up slowly until the column housing is close to the ground.
Switch to 03 and press the button on the control box to save the column
protection AD value.
After the debugging is completed, the longitudinal movesment function of the
tube column should be tested:
Keep the table level, then operate the handle to reverse or , or press the
column moves button or on the near control panel to control the movesment
of the column of the tube. When the column of the tube is pressed to the limit of the
travel switch, the column of the tube should stop moving immediately, and the limit
will be effective. Control the operation handle or press the column moving button on
the near control panel, you can control the movesment of the tube rack, release the
handle or release the button to stop immediately, the column moving stroke is 970 ㎜,
48
User’s Manual V2.11.1E
1) Table top surface lateral movement direction debugging: click the Table top
surface access button in the local control panel or (ensure that the lateral
movement of the Table top surface is not in the limit state). If the sliding direction is
the same as the direction marked by the button, the lateral movement direction of the
table surface is correct. If it is opposite, the input drive lines of two motors at the
non-public end should be adjusted to invert the motor.
2) table side movement travel limit debugging: table side movement USES
photoelectric coupler as limit device, which is installed in table assembly.
or , release the handle or release the button to stop immediately, and stop
immediately when reaching the limit.
Finally, the position of the photoelectric coupler of each travel limit is adjusted to
meet the designed table surface travel.
49
User’s Manual V2.11.1E
1.Compressor travel limit debugging:A travel switch is used to limit the up and
down of the compressor, and is installed in the assembly of the compressor, as shown
in Figure 5-5.
In the debugging mode "C07", press the ascending stroke switch by hand, LED5
off, let go of light, the limit is normal. The operate handle upside down , LED49
bright, LED48 off, the compressor up.Pressing the upper limit switch by hand will
the handle of the control box on the operation console moves to , the compressor
is down, moves to the compressor is up. If the reverse should be done, the input
drive line of the compressor motor should be adjusted, and the non-common two
motor drive lines of 2407, 2408 and 2409 at the lower end of the terminal block can
be reversed to make the compressor motor.
Finally, adjust the position of the limit stroke switch of the compressor to meet
the needs of the advance and retreat stroke of the compressor.
SID limit Debugging:The travel limit of the tube uses a travel switch as the
limiting device, as shown in Figure 5-6.
movess to , LED46 is on, LED47 is off, the tube rises; The operating handle
moves to , LED47 is on, LED46 is off, and the tube drops. Press the
corresponding travel switch by hand, if the tube stops moving immediately, then the
travel switch limit is normal, otherwise, check the corresponding line and the
corresponding limit switch.
Finally, adjust the position of the tube lift limit switch to meet the needs of SID.
51
User’s Manual V2.11.1E
The rotating pedal uses a DC motor,under the debugging state "C08", the
operation handle moves to ,LED42 on, LED43 off, and the pedal is turning
forward; The operating handle moves to ,to LED43 on and LED42 off, and the
pedal is reversed. If the direction is opposite, replace the two motor drive lines 2415
and 2416 at the lower end of the terminal block.
In the mode of "C09", the handle is inverted to the subfunction switch key to
enter the subfunction. At this time, SID nix tube will display the setting value of
overlap width of mosaic image, and the third nix tube will display the saving value of
ranges from 2 to 15; Press button save set value, save after the pendulum third
digital tube update display saved value.
In "C19" mode, the operation handle is reversed to the subfunction switch key to
enter the subfunction, SID nixie tube shows the set value of table type, and the third
nixie tube shows the saved value of table type. The operation handle moves to
or , select the current actual table type set value, the table type of the
equipment should be set to "20F"; Press to save the set table type, after saving
the third digital tube update display saved value.
When all debugging work is complete and the test is normal in normal mode,
machine aging tests can be performed if necessary. Enter the debugging mode "C10"
and press start aging test. At this time, the second nixie tube displays the value
of aging step, and the third nixie tube displays the value of aging waiting
52
User’s Manual V2.11.1E
time; During the aging process, press again to exit the aging test and return to
the debugging mode.
For the official version, user needs to start the installation from the CD-ROM, if
it isn’t automatically start, double click “DFOC_XXXX.exe”; for the demo version,
mostly the user will get an installation package, and then just double click
“DFOC_XXXX _DEMO.exe”.
Double click to start the installation, the following steps will display:
Select language
If it is to upgrade the software, there would pop up the following interface:
Uninstall option
53
User’s Manual V2.11.1E
Select folder
Confirm installation
54
User’s Manual V2.11.1E
Installation process
Install the granddog driver. (For official version only)
Start installation
Finish installation
After successful installation, there would display the following interface:
.
Finish installation
55
User’s Manual V2.11.1E
After the installation steps above, the program will create shortcut of
“DRConsole.exe” and “DRDongle.exe”. User can double click the “DRConsole.exe”
to start the program, or double click the “DRDongle.exe” to register the software.
[Warning] if the official version is unregistered, the program will show message
in Figure 2.8, then the program will automatically shut down
Unauthorized messages
The official version of the software will come with a USB hardware dongle.
After software installation; the user inserts the dongle into the computer's USB port to
run dongle "111". Click the " " button. If the dongle driver is installed
correctly and the USB dongle has been successfully identified by the computer, the
dongle serial number and authorization information will appear simultaneously.
56
User’s Manual V2.11.1E
After the operating system starts, the login window displays. Enter the correct
user name and password on the system login interface to enter:
In the user login window, enter the correct user number and password and press
" " to enter the loading interface. If you do not want to enter the system, click
Click " " button from the navigation bar to enter the management options
and select hardware configuration:
Click the " " button to enter the hardware configuration check box,
including Syncbox, high voltage, detector, collimator, DAP.
57
User’s Manual V2.11.1E
5.3.2.1 Syncbox
High voltage generator: RuiYing (set according to the actual high voltage model)
Log level: normal mode
58
User’s Manual V2.11.1E
5.3.2.3 Detector
Under WIN10 system, open the control panel of the network connection, find
local address workstations with plate connection, double-click to open and click
59
User’s Manual V2.11.1E
"properties" find "Internet protocol version 4 (TCP/IP)" and click "properties", open
the dialog box as shown, select "use the following IP address", as shown in the figure
below set the IP address and subnet mask:
The name of each element on the self-test interface and its role:
1.Self-test process information display area
(A) High voltage generator works normally and is turned on for the first time:
it shows the process of self-test of the system.
(B) High voltage generator is working normally, not the first time it is turned
on: it shows the current system status.
60
User’s Manual V2.11.1E
(C) High voltage not connected: only "CAN open successfully" is displayed.
(D) Open the software repeatedly: display the following information.
After pressing the "OK" button, the software enters the self-test interface, and the
prompt appears again, indicating that the last software is running. At this time, please
close all running HVSDK.exe, and then open the software to run normally.
Note:
(A) If the software does not open repeatedly, when the following prompt appears
when opening the self-test interface, please press the software element 2 buttons
(close button), then unplug the CAN controller communication line and plug it in
again to open the software. Normally enter the self-test process.
61
User’s Manual V2.11.1E
ready:
The status of this light when the second brake (exposure) is pressed:
62
User’s Manual V2.11.1E
(3) Automatic X-ray tube calibration switch button: there are two states,
default and pressed. When this button is pressed, the sub-interface is switched to the
automatic X-ray tube calibration interface.
(4) Configuration button: there are two states, default and pressed. When this
button is pressed, the sub-interface is switched to the configuration interface.
(5) System date and time display
(6) Display area of X-ray tube heat capacity percentage: after one or more
exposures, the current X-ray tube heat capacity percentage will be displayed.
(7) Filament adjustment (subtraction) button: click this button, the filament
value is reduced by 1. The filament value will be further decreased when long press
this button. There are three states: default and mouse stay and being pressed.
(8) Display area of filament value: display current filament parameter value
with the decimal point at 0.When adjusting the KV or MA value, the filament
parameter value will change accordingly.
(9) Filament adjustment (add) button: click this button, the filament value is
added by 1. The filament value will be further added when long press this button.
There are three states: default and mouse stay and being pressed.
(10) KV adjustment (subtraction) button: click this button, the KV value is
reduced by 1. The KV value will be further decreased when long press this button.
There are three states: default and mouse stay and being pressed.
(11) Display area of KV value: display current KV parameter value with the
decimal point at 0, and the range is [40,150].
(12) KV adjustment (add) button: click this button, the KV value is added by 1.
The KV value will be further added when long press this button. There are three states:
default and mouse stay and being pressed.
(13) Display area of KV average value: display the average KV sampling data
with the decimal point at 2.
Note: "Error" will be displayed here when the simulated sampling Error value
exceeds 40%.
(14) Display area of KV error percentage value: After receiving exposure
once, the system will sample KV data, calculate the average value of the data and
analyze the error.
Display value algorithm for this region:
63
User’s Manual V2.11.1E
((|KV average display value - set value |)/ set value) *100% with 1 decimal point.
Note: "Error" will be displayed here when the simulated sampling Error value
exceeds 40%.
(15) MS adjustment (subtraction) button: click this button, the MS value is
reduced by 1. The MS value will be further decreased when long press this button.
There are three states: default and mouse stay and being pressed.
(16) Display area of MS value: display current MS parameter value with the
decimal point at 2.
(17) MS adjustment (add) button: click this button, the MS value is added by 1.
The MS value will be further added when long press this button. There are three states:
default and mouse stay and being pressed.
(18) MA adjustment (subtraction) button: click this button, the MA value is
reduced by 1. The MA value will be further decreased when long press this button.
There are three states: default and mouse stay and being pressed.
(19) Display area of MA value: display current MA parameter value with the
decimal point at 1.
(20) MA adjustment (add) button: click this button, the MA value is added by
1. The MA value will be further added when long press this button. There are three
states: default and mouse stay and being pressed.
(21) Display area of MA average value: display the average MA sampling data
with the decimal point at 2.
Note: "Error" will be displayed here when the simulated sampling Error value
exceeds 40%.
(22) Display area of MA error percentage value: After receiving exposure
once, the system will sample MA data, calculate the average value of the data and
analyze the error.
Display value algorithm for this region:
((|MA average display value - set value |)/ set value) *100% with 1 decimal
point.
Note: "Error" will be displayed here when the simulated sampling Error value
exceeds 40%.
(23) Display area of KV analog waveform: after a manual exposure, this area
64
User’s Manual V2.11.1E
will show the relation between sampling KV and exposure time. The ordinate is
sampling KV, and the abscissa is exposure time. Click the waveform and pull it to the
right. The waveform becomes larger and the mouse is restored to the left waveform.
(24) Display area of MA analog waveform: after a manual exposure, this area
will show the relation between sampling MA and exposure time. The ordinate is
sampling MA, and the abscissa is exposure time. Click the waveform and pull it to the
right. The waveform becomes larger and the mouse is restored to the left waveform.
(25) System log button: when this button is pressed, the sub-interface is
switched to the system log interface.
(26) Display area of system communication information:
Display when the system is working normally:
65
User’s Manual V2.11.1E
then press the second gear of the hand brake and do not release it. At this time, enter
the automatic calibration process. Starting small focus, for example, high voltage
generator will continuous exposure, since 50 kV and adjusting filament value, so that
the filament current direct or close to 10 MA, the elements of 40 corresponding gear
(50 kV, 10 MA) region shows "tick", said the end of the calibration, after the system
to continue a (50 kV, 12.5 MA) filament calibration, until the 120 kV, 100 MA
calibration is completed. If the calibration is completed, the system will pop up the
completion prompt dialog box and press ok. If the hand brake is released during the
process, the software will pop up the inquiry dialog box and ask whether to continue.
If you press ok, you will stay in the current calibration status. If you press cancel, you
will exit the automatic calibration interface and return to the manual calibration
interface.
(38) Current selected tube model
Configuration interface
67
User’s Manual V2.11.1E
(41) Filter selection drop-down box: click this drop-down box and the filter
selection box will pop up. Click a certain type.
(42) AEC curve select drop-down box: click this drop-down box, will produce
11 AEC curves, and then click a certain type.
68
User’s Manual V2.11.1E
(44) Select Drop Down List Box of Treadmill mode:Press Select Drop Down
List Box,there will appear 7 kinds of Treadmill modes as followed:
71
User’s Manual V2.11.1E
The display of the dose indication value is obtained by calculating the loading
factors of the high voltage generator, so the output accuracy of the high voltage
generator must be controlled to ensure the accuracy of the dose indication value.
Placed in the center of the image intensifier 100 x100 square iron box (coarse 5
x5), adjusting the limit of the beam splitter to X and Y to speak, in the light of wild
slightly larger than the iron box (5 mm or so), by looking through the method of the
limit of optical field box beam and the place of iron frame around the gap is consistent,
otherwise adjust the limit of internal view the location of the light beam splitter, lead
to the agreement clearance (error is about 2 mm).
, 3 Seconds later, four "8s" appear in the window indicating the real-time
opening size and enter the debugging state (the power on the new Collimator without
72
User’s Manual V2.11.1E
debugging parameters will automatically enter the debugging state). At this time, L1
and L2 in the real-time opening size indication will be off, and four "0's" will be
displayed on the digital tube.
b. At this point, click or , you can select the function, and the first
under "C4".Click , Display 4 "d", that is, the opening size of the Collimator has
been adjusted. The opening size of Collimator can be adjusted by "X opening" and "Y
opening". "X opening" corresponds to A value, and "Y opening" corresponds to B
displayed by the first group of nixie tubes. Click or , the opening size
can be adjusted, and the second set of digital tube display its opening real-time AD
value. When the corresponding value is selected, press Set values can be
saved. The comparison table of relevant parameters is as follows:
73
User’s Manual V2.11.1E
Table 5-9
X to open Y to open
A1=24mm b1=30mm
b2=54mm
A2=28mm
(perspective)
A3=30mm b3=48mm
A4=33mm b4=60mm
A5=36mm b5=66mm
A6=42mm b6=72mm
A7=48mm b7=84mm
A8=60mm b8=101mm
A9=66mm b9=Max
AA=72mm
Ab=84mm
AC=54mm
(perspective)
Ad=Max
2) Setting description:
In the normal working state, the delay button of the point light turns on the visual
field light, and the control button of the point related lead blade reduces the lead blade
in two directions to zero and enters the debugging state. The real-time AD value is
saved under the function of "C1", the real-time AD value is saved under the function
of "C2", and the corresponding delay time is saved under the function of "C3"."C4"
function is set in the next menu level, as function description 3.Among them, "C4" is
also divided into whole piece setting, vertical setting and horizontal setting.C5
"function is X, Y axis automatically open and close test, press Entering the
test, facing the Collimator operation panel, the 2-bit nixie tube on the left shows the
value of AD on the X-axis, and the 2-bit nixie tube on the right shows the value of AD
on the Y-axis (the remaining 4 keys are useless in the test);then press Close the
test and display "C500".
Full film setting: turn on the Collimator light, turn the vertical and horizontal
74
User’s Manual V2.11.1E
lights to the maximum limit, then shrink back a little and press down Save the
Settings.
Setting of vertical division, vertical dichotomy only requires the cooperation of
large light bar board, turn on the Collimator lamp, reduce the longitudinal direction to
the point where the light falls on the lead sheet of large light bar, and press down
light shining on the lead sheet of the small light bar, and then press down
Save the Settings.
3) exit setting description:
To exit the debugging state, you must first ensure that the function display is L1
and L2 off, digital tube display 4 "0", function display key can make the function
display back to this state, press and hold 3 seconds after the display of 4 "8"
to quit. One from the "C4" next level menu back to "C4", click or
75
User’s Manual V2.11.1E
(2) In the "DRMode" menu, select the DRS3072 mode (the green is selected), select
the maximum frame rate 5 on the right, and click the "Calibration" button. At this
time, the "Offset" button will be switched from unselected to selected, as shown
in the picture below:
76
User’s Manual V2.11.1E
(4) Click the "Gain" button and adjust the fluoroscopic dose for fluoroscopic
(Suggested KV parameter 50KV). The gray Value of the bright field map after the
Current loaded offset calibration will be displayed in "Current Value". Compare
this gray Value with "Expected Value 2000" to make the "Current Value" close to
the "Expected Value" with a deviation of no more than 5%. When the two values
meet the above requirements, keep the foot brake on and click "Select"
button. The gain calibration starts. The value next to "Process" displays the
current calibration status. When "Process" displays "19/19", the current
calibration step has been completed.
77
User’s Manual V2.11.1E
Note: The Select button can only be selected if the handbrake is pressed all
the time.
(5) When the dialog box "Stop Exposure and Start to acquire dark Image" pops up on
the interface, release the foot brake to collect dark field images.
78
User’s Manual V2.11.1E
(6) Then enter the second step of calibration, and the stage state becomes "2/3". Click
the "Gain" button;
(7) Adjust the fluoroscopy dose to perform fluoroscopy (e.g., Suggested fluoroscopy
KV parameter 110KV). The gray Value of the bright field map after the Current
loaded offset calibration will be displayed in "Current Value". Compare this gray
Value with "Expected Value 12000". Keep the "Current Value" close to the
"Expected Value" and the deviation within 5%. When the two values meet the
above requirements, keep the foot brake on and click the "Select" button to start
gain calibration. The value after "Process" will display the current calibration
status. When "Process" displays "19/19", the current calibration step has been
completed.
(8) When the dialog box "Stop Exposure and Start to acquire dark Image" pops up on
the interface, release the foot brake to collect dark field images.
After dark field map is collected, the state is changed from non-optional to
optional. Click "Accept" button to generate the second step correction file.
79
User’s Manual V2.11.1E
(9) Then enter the third step of calibration, and the stage state becomes "3/3". Click
the "Gain" button;
(10)Adjust the fluoroscopy dose for fluoroscopy (e.g. Suggested KV, 70KV), and the
gray Value of the bright field map after the Current loaded offset calibration will
be displayed in "Current Value". Compare this gray Value with "Expected Value
2000". Keep the "Current Value" close to the "Expected Value" and the deviation
within 5%. When the two values meet the above requirements, keep the foot
brake on and click the "Select" button to start gain calibration. The value after
"Process" will display the current calibration status. When "Process" displays
"19/19", the current calibration step has been completed.
(11) When the dialog box "Stop Exposure and Start to acquire dark Image" pops up on
the interface, release the foot brake to collect dark field images.
After dark field map is collected, the state is changed from non-optional to
optional. Click "Accept" button to generate the third step correction file.
80
User’s Manual V2.11.1E
3) Click the "Gain" button and adjust the fluoroscopic dose for fluoroscopic
(Suggested KV parameter 50KV). The gray Value of the bright field map after the
Current loaded offset calibration will be displayed in "Current Value". Compare
this gray Value with "Expected Value 2000" to make the "Current Value" close to
the "Expected Value" with a deviation of no more than 5%. When the two values
meet the above requirements, keep the foot brake on and click "Select"
button. The gain calibration starts. The value next to "Process" displays the
current calibration status. When "Process" displays "19/19", the current
calibration step has been completed.
81
User’s Manual V2.11.1E
4) When the dialog box "Stop Exposure and Start to acquire Dark Image" pops up
on the interface, release the foot brake to collect dark field image, as shown in the
figure.
5) Then enter the second step of calibration, and the stage state becomes "2/3". Click
the "Gain" button;
6) Adjust the fluoroscopy dose to perform fluoroscopy (e.g., Suggested fluoroscopy
KV parameter 110KV). The gray Value of the bright field map after the Current
loaded offset calibration will be displayed in "Current Value". Compare this gray
Value with "Expected Value 12000". Keep the "Current Value" close to the
"Expected Value" and the deviation within 5%. When the two values meet the
above requirements, keep the foot brake on and click the "Select" button to start
gain calibration. The value after "Process" will display the current calibration
status . When "Process" displays "19/19", the current calibration step has been
completed.
82
User’s Manual V2.11.1E
7) When the dialog box "Stop Exposure and Start to acquire Dark Image" pops up
on the interface, release the foot brake to collect dark field image, as shown in the
figure.
After dark field map is collected, the state is changed from non-optional
to optional. Click "Accept" button to generate the first-step correction file.
8) Then enter the third step of calibration, and the stage state becomes "3/3". Click
the "Gain" button;
9) Then adjust the fluoroscopy dose for fluoroscopy (e.g. Suggested KV, 70KV),
and the gray Value of the bright field map after the Current loaded offset
calibration will be displayed in "Current Value". Compare this gray Value with
"Expected Value 2000". Keep the "Current Value" close to the "Expected Value"
and the deviation within 5%. When the two values meet the above requirements,
keep the foot brake on and click the "Select" button to start gain calibration. The
value after "Process" will display the current calibration status . When "Process"
displays "19/19", the current calibration step has been completed.
83
User’s Manual V2.11.1E
10) When the dialog box "Stop Exposure and Start to acquire Dark Image" pops up
on the interface, release the foot brake to collect dark field image, as shown in the
figure.
11) After the dark field map is collected, the non-optional state is changed to optional
state. Click "Accept" button to generate the third step correction file, and then the
perspective mode calibration is officially completed.
84
User’s Manual V2.11.1E
1. Power on button.
Power on:1.Make sure the emergency stop switch unscrew,press the start
Tilt button
86
User’s Manual V2.11.1E
Fluoroscopy instructions
Power on button
87
User’s Manual V2.11.1E
1) Table tilt: Operating on the remote control operating handle or near the
tilt buttons on control board, can control the table tilt up down motion,
close control, operation can once again, the console operation, At the same
time, press this button and the inverted table handle button to rotate the
table from the horizontal to the negative angle direction. When the table
moves up, the tube should be full of a certain height from the ground. If not,
the tube column will be automatically adjusted to avoid falling and touching
the ground
control panel. Release the handle or release the button and stop immediately.
3) Press cup advance and retreat:Press the press into button on the remote
control table to control the downward action of the pressure cup, press the
press back button on the remote control table to control the upward action of
the pressure cup. In the compressor tracking key indicator light and
and bed panel has a relative displacement will make the relief of
movement handle or press the near control tabletop lateral movement key
can control the tabletop lateral movement, reflected as the perspective image
on the TV to move left and right, release the handle or release the key
hand brake (preparation), ready indicator in the upper right corner of the ready
button lights up.
By pressing the exposure button or the second gear of hand brake (exposure)
controlled by the generator, the exposure indicator lights up and a buzzer goes
out through the process through which the X-ray is emitted.
You can use the first gear of hand brake or (ready) and the second gear
90
User’s Manual V2.11.1E
Press the first gear of hand brake or (ready), when ready, press the
Ready: press the first gear of hand brake or , prepare to select the X-ray
tube for exposure. When the X-ray tube is ready and there is no interlock failure or
system failure, the preparation indicator light on the generator control box will be lit.
When the foot brake switch is pressed for X-ray fluoroscopy, The
perspective indicator lights up.
1. Basic structure
The shape of the collimator is shown in the figure below. The main internal parts
are transmission device, detection device, transverse variable lead blade, longitudinal
variable lead blade and limit mechanism. The outside is mainly composed of lead
blade opening and closing adjusting buttons, collimator lighting buttons and interfaces
connected with X-ray tube components.
91
User’s Manual V2.11.1E
2. Instructions
a. The size of the radiation field of view is shown by the size of the irradiated
area of the collimator. The size of the field of view is adjusted by clicking the
corresponding button of the lead plate opening and closing indicator icon on the panel.
The operating table also has corresponding buttons to control the size of the visual
field.
b. In the fluoroscopy state, when it is switched to the manual state, the vertical
and horizontal lead blades can be remotely adjusted to change the visual field size. In
the automatic state, the Pedals only tracks to the preset value when it is pressed. The
same is true in radiography.
The user can speak to the patient by stepping on the left-foot switch.
92
User’s Manual V2.11.1E
6.1.5.1 Login
After the operating system starts, the login window displays. Enter the correct
user name and password on the system login interface to enter:
In the user login window, enter the correct user number and password and press
" " to enter the loading interface. If you do not want to enter the system, click
6.1.5.2 Exit
93
User’s Manual V2.11.1E
Enter the login password ,Click ,After the system saves the image, exit
the console software.Click ,cancel the exit operation and return to the software
operation interface.
1.Work list
Work list
94
User’s Manual V2.11.1E
Search Items:
1. Patient ID: user need to input the patient’s id in the textbox, and define the correct
date time.
2. Accession: user need to input the patient’s accession in the textbox, and define
the correct date time.
3. Name: user need to input the patient’s name in the textbox, and define the correct
date time.
4. Today: query the patients’ data in today.
5. Two days: query the patients’ data in two days.
6. Three days: query the patients’ data in three days.
7. One week: query the patients’ data in a week.
8. Custom: query the patients’ data in user defined period. Once the condition set
down, the work list will query data frequently, it can also refresh immediately
after the button “Refresh” clicked.
Delete:
Delete current selected patient.
Begin to study:
Begin to acquiring image of current patient.
Emergency:
Register a patient in emergency with auto generated information.
95
User’s Manual V2.11.1E
2.Registration
Registration
The register is used to record information of a new patient. User can select study
protocols through protocol or Item mode.
Accession*: input the accession number of patient. It will be automatically
generated and usually suggest user don’t change it. It must unique in the system.
PatientID*: input the ID of a patient. It will be automatically generated and
usually suggest
user don’t change it. Input PatientID and click , system will automatically
fill the patient’s information in the text box.
96
User’s Manual V2.11.1E
Study Bodypart:Select the oval gray button on the body as needed, and click the
button, andthe gray turns green to indicate that the site is selected. The list on the right
has a corresponding check position.
Study Protocols and Items:
97
User’s Manual V2.11.1E
Click button
Switching Rad/
Fluoroscopy.
Study list
Study list
It is used to query the information of the patients that finished the study. Same
with the work list, the result can be displayed under different condition.
can query the specified patient information and related photographic image records.
a) List Information:
b) Information search:
Study management provides the ability to search on a certain basis. Enter the
ranges can be quickly selected via buttons, including "Today," "Two Days," "Three
Days," "One Week and Custom" ranges. After the retrieval criteria are selected, click
the "Confirm" button to check and place eligible patients for examination in the list.
[Note] If the patient number, the study number, the patient name is empty,
indicates any condition.
c) Delete records:
Select a checked patient in the list to be examined, click the button and
If you confirm that delete the patient record have selected, click OK, otherwise
Cancel.
d) Browse images:
Select a patient in the study list and double-click it to enter the image browsing
interface.
e) Re Examine:
Click button enter the image acquisition interface, you can add a
100
User’s Manual V2.11.1E
Print: Print images with connected print devices, Click the button . See
chapter 8.3
Selection All: All patient information in the list is selected, and all
Lock: Select one or more patient records and click the button to pop
up the check box, This record is locked after click OK , Lock bar marked as “1”
Storage queues: When the system is not connected to the PACS server,
click the button to add the selected item to the upload queue and wait for the
communication to connect to upload. The Storage Queue check box can be refreshed,
retried, removed, and canceled.
Print Queue: when the system is not connected to the PACS server, click the
button to add the selected items to the network print queue and print after the
101
User’s Manual V2.11.1E
communication is connected. The Print Queue check box can be refreshed, retried,
removed, and canceled.
102
User’s Manual V2.11.1E
Note: Interface functions and operations vary with hardware and system
configurations.
1.Thumbnail navigation area
2.Exposure condition and position selection area.
3.System status.
4.Image preview.
5.Image processing tool.
6.Dynamic image playback control area, image save area.
7.Functional control area.
8.Other Information.
a)Thumbnail navigation area
103
User’s Manual V2.11.1E
104
User’s Manual V2.11.1E
There are the body position diagrams selected for patient information registration.
You can see the specific study body part and the standard position for intuitive
reference.
For emergency patients this region is blank. See Chapter 4.2 for steps to add a
body position.
Before Image acquisition, the parameters of generator should be property
configured, and the trigger mode of detector must be correct. For more information,
please read the manual of these devices.
Exposure signal lamp and Reset error:
Press first gear of the hand brake, then the signal lamp
will light up. The yellow light will keep lighting up in the
exposure process.
105
User’s Manual V2.11.1E
Exposure position
Table Wall Reserve
106
User’s Manual V2.11.1E
Focus
Small focus Big focus
Exposure mode
mA/ms mode: KV mA and mS parameters can be
adjusted. mAs mode: Only KV and mAs can be adjusted.
KVp: Fluoroscopy KV
voltage
mA:Fluoroscopy mA
[Note]This region show the default parameters of
the current item. You can adjust these parameter
by click the arrow button.
107
User’s Manual V2.11.1E
108
User’s Manual V2.11.1E
Cycle again.
Play.
109
User’s Manual V2.11.1E
Play forward.
Play back.
f)Functional control area
Hide exposure conditions and position selection area.
The printing interface pops up. See Chapter 8.3 for details.
Click the button to pause the current check, and the page
returns to the work list interface.
Click the button to save the image and enter the studylist
interface.
g)Other Information
110
User’s Manual V2.11.1E
4.Image browse
Image preview
After acquired an image from detector, the processed image will be displayed.
There are many powerful tools for advanced user to take a better view of the image.
a) Tools
Tool button Description
Page Pre/Next.
Zoom In/Out:
Histogram WW/WL
111
User’s Manual V2.11.1E
Invert Color
Zoom out
Zoom in
Screen Size
Text Edit
Rotate 90
112
User’s Manual V2.11.1E
Flip Vertical
Image hint
Magnifier:
Copy:
The function of image contrast has been optimized. Click the
“Copy” button in the preview interface of patient A, then switch to
the preview interface of patient B, select a blank box and click
the “Paste” button to add the copied image. Be sure to click
the “Save” button before exiting if you want to save the image.
113
User’s Manual V2.11.1E
Email:
Save as:
Its role is similar to the “copy” / “paste” tool. All three of them
are used to compare an image with itself. The “copy” / “paste” tool
can be used among different patients, while the “Save as” tool
can only be work with the images of one patient (the “Save as” tool
has the same function with a
combination of the “copy” and “paste” tool)
114
User’s Manual V2.11.1E
Inverse Compare:
When you click on the button, the Software will show another
inverted image as guidance for doctor observing and diagnosis.
Notes: the “Image color inversing” tool can be only used in
the image viewing interface of the DR Software.
Window/Level
ROI Window/Level
Auto WW/WL
Full Size
Mark Left
Mark Anterior
Mark Right
Clip Tool
Rotate -90
Flip Horizontal
115
User’s Manual V2.11.1E
Distance Measure:
Image stitching.
ROI Magnifier:
Arrow
Paste:
The function of image contrast has been optimized. Click the
“Copy” button in the preview interface of patient A, then switch to
the preview interface of patient B, select a blank box and click the
“Paste” button to add the copied image.
Be sure to click the “Save” button before exiting if you want to
save the image.
Free Rotate:
Click the “Free Rotation” tool, then pop-up a box for rotation
angle. Enter any
angle value you want and click the “OK” button to rotating
image. Rotate to the left by default. Click the “Undo” button the
image will turn back.
Cobb angle:
116
User’s Manual V2.11.1E
Compare:
When you click on the button, the Software will show a pop-up
box for image selection. Then you should click on the “Add” button
and select two or more images, finally click on the “Open” button to
enter the Image contrast interface. Note here that if the Software is in
the image Acquiring interface, please click on the “Save” button
before image selection.
Notes: the “Image contrast” tool can be only used in the image
viewing
interface of the DR Software.
117
User’s Manual V2.11.1E
Clip Options
Polygon measurement
118
User’s Manual V2.11.1E
Rectangular measurement
[Note] Hold the ctrl key and pull it into a square
Elliptical measurement
[Note] Hold the ctrl key to pull it into a perfect circle
Finish &Save:
End current study and save all images to local drive as DICOM files. If
successfully, then user can find the study in the studylist.
b) Image stitching
in the image
preview interface. Before to enter the image stitching, please make sure the
current study has at least two saved images.
119
User’s Manual V2.11.1E
Move the mouse to the image need to be printed, then there will display a white
120
User’s Manual V2.11.1E
check box at the lower right corner of the image. Check the box then the box to add a
black “√” and the there will appear a yellow number at the lower left corner. Click
again to cancel the selected image.
【2】 Diagnostic content and conclusion edit box
You can edit the diagnostic content and conclusion in the edit box, click the
content and description template on the right side if you need to add.
【3】 Select the template for printing
1Standard300.rvf : One image with the height of 300
2Standard300.rvf : Two images with the height of 300
4Standard300.rvf : Four images with the height of 300
6Standard300.rvf : Six images with the height of 300
121
User’s Manual V2.11.1E
122
User’s Manual V2.11.1E
123
User’s Manual V2.11.1E
You can add and update the doctor information in the interface.
Button Function Describe
When there are more than one monitor,
Multi-screen display the doctor can write a report while viewing
the image.
b) Report Printing
124
User’s Manual V2.11.1E
This chapter will concentrate on image archive, export and printing which used
in the study list.
Study list
a) Image archive
In archive process, the images will be sent to PACS server follow the DICOM
3.0 standard.
Before the use of image archive, the parameters of DICOM storage must be
properly configured (reference to Configuration manual).
125
User’s Manual V2.11.1E
If there is a failed message, then user should check the communication with the
PACS.
b) Image export
The system provides two ways of export, and support customized export. User
can export all or selected patient’s data from the study list.
i. Burning CD
Select the patients that need to be imported or burned. Hold down Ctrl to select
multiple patients
CD2GO
Enter or select the appropriate information. FileSetID:The name of a burning or
producing ISO file. File compression: Standard, Compression.
Structure: Selecting DCMIMG means that only DICOM images will be output,
and the system will automatically generate by patient name and check well
File’s name: Select DCMDIR to output the patient's information and DICOM
image as a CDViewer. Check BMP or JPG to output the image in either format
Save Settings: Generate ISO files or burn directly to CD. ISO files can be
126
User’s Manual V2.11.1E
DCMDIR export
Running the CDViewer:
CDViewer
127
User’s Manual V2.11.1E
Double click the select patient or click preview , user can preview the
image of current patient.
128
User’s Manual V2.11.1E
Print preview
Click the , user can configure the property of select printer. All the property
is referenced to DICOM 3.0 standard.
Configuration of printer
Choose layout of images:
Film size: the film size should consistent with the printer.
File orientation: “LANDSCAPE” for horizon, “PORTRAIT” for vertical. Print
129
User’s Manual V2.11.1E
Start printing
all images
6.Exit
130
User’s Manual V2.11.1E
Enter the login password, click the system saves the image and exits
,
the console software; click , Cancel the exit operation and return to the
software operation interface.
6.2.1 Fluoroscopy
When you need to fluoroscopy, you need to enter the image acquisition
workstation first, set up the image acquisition options, set the image acquisition
fluoroscopy, step the fluoroscopy foot switch under the console. The system will
display the image in real time. When the foot switch is loosened, the final image will
be kept and displayed automatically.
6.2.2 Radiography
Enter the image acquisition workstation first, set the image acquisition options, set
the image acquisition perspective options and photography conditions, step on the
perspective foot switch under the console, the system will display the image in real
time, move to the area of interest, press the exposure hand brake for photography.
When the device is in use and the power supply is suddenly interrupted and the
compressor is in the position of pressing the patient, in order not to cause injury to the
patient due to the long duration of the pressure, please press the button marked
"release button" in the figure below to lift the pressure cup of the compressor upward
and relieve the pressure on the patient.
131
User’s Manual V2.11.1E
Before pressing the button, press the pressure cup in the direction of the patient,
and then press and hold the button to eliminate the pressure cup in the direction of the
patient, and the pressure cup will automatically spring up.
The initial run-up of the equipment after it has first been installed, by the
professional installation personnel in accordance with the adjustment description of
the high voltage generator in the specification, until the initial run-up is completed.
132
User’s Manual V2.11.1E
CHAPTER7 MAINTENANCE
1. General Rules
This equipment needs regular maintenance; following suggestions will be good for
usage and safety.
● Keep room clean, dry, and with good ventilation.
● Mechanical safety will be degraded due to long time usage. In order to reduce
potential damage to both user and patient, it is necessary to check mechanism if any
loose and if brake is good.
● Correct electrical calibration can well perform machine’s performance.
Calibration work should be completed by professional staff authorized by our
company.
● Machine should be in charged by designed person and establish file management
system.
● Any changes to machine are forbidden to ensure both patient and operator’s
safety, if any changes needed, please contact with manufacturer.
● Adjustment for contact gap among moving parts with orbital bearing assembly
should be done once a year.
● Regular big repair should be done, for details pls contact with manufacturer.
● If product has malfunction, contact with professional staff approved by
manufacturer.
● Operator should be well trained and can complete normal inspection.
133
User’s Manual V2.11.1E
134
User’s Manual V2.11.1E
135
User’s Manual V2.11.1E
low, mutation and other anomalies, machine should be closed immediately and
check out the problem, it can only be used after problem solved.
5.Do not press heavily on all buttons of controller to avoid dislocation, injury or poor
contact, even damaged.
6.Check frequently protective earth of each components is in good condition
(grounding resistance is very small).
7.Check power supply on a regular basis, and check power supply resistance value (or
the power supply voltage value is changed), to make sure it meet machine’s
requirement.
8.After using the controller, it should be cut off the power in time.
9.Avoid striking of x-ray tube components during operation.
10.When X-ray tube components are used for the first time or idled for a longer time,
it must be progressive loading before X-ray tube used, and do the sophisticated
test to avoid damaging to the X-ray tube by instantaneous load.
11.When machine works continuously, please pay attention to the X-ray tube
assembly to allow Thermal capacity, or check surface temperature of X-ray tube
assembly , it can’t exceed 75 ℃.
136
User’s Manual V2.11.1E
137
User’s Manual V2.11.1E
7.4.1 HV Generator
Fault
Information Adjustment method
Code
ER00 Communication fails 1. Check if the CANBUS communication
wire of the upper computer is of good
contact.
ER1 First gear of hand break 1.Firstly, turn off high voltage generator ,
plugs in then using avo meter to check if the 1st gear
during self-checking. of handbrake is right. If the handbrake is
abnormal, change it and then check again. If
difficult is settled, problem will be closed
or,otherwise, proceed the next procedure
2.check the wire from mini control box to
the plug of power panel PP1
3.change interface CPU board and turn on
high voltage generator to check , if fault
disappears, problem will be closed;
4.if trouble is still there after above
procedures, you can shut down, change
main cpu board, and then check again. If
trouble disappears, it can be ascribed to
handbrake problem and problem will be
closed.
138
User’s Manual V2.11.1E
ER2 Second gear of hand break 1. Firstly, turn off high voltage generator,
plugs in using avo meter to check if 1st gear of band
during self-checking. brake signal is right and, if abnormal,
change handbrake and go on your
examination. If trouble is settled, it can be
credited to handbrake problem and problem
will be closed. Otherwise, proceed the next
procedure.
2.check the wire from mini control box to
the plug of power panel PP1
3. Change main control board and turn on
high voltage generator to check. If fault
disappears, it can be attributed to main
control board problem and problem will be
closed;
4. If trouble remains after above procedures,
you can shut down and change interface
board to check. If fault disappears, it can be
139
User’s Manual V2.11.1E
ER4 Tested out excess feedback 1. Check if the network electricity is in the
value of bus voltage AD range of generator standard limitation, if
during soft powering beyond, turn of the machine;
2. Malfunction of power board; change
power board ;
ER7 Tested out excess feedback 1. Check if the network electricity is in the
value of bus voltage AD range of generator standard limitation, if
after main beyond, turn off the machine;
contactor plugged in. 2. Malfunction of power distribution board;
change power distribution board;
ER9 SLINK is high level during 1. Check if the ground wire of oil tank is
self-checking. It is possible rightly linked
that oil tank is not linked or 2. Check if the wire between main control
didn`t attach ground wire. board and oil tank is rightly linked
140
User’s Manual V2.11.1E
lamp filament
malfunctioned
during self-checking.
ER15 PHVOK
Please contact technical support
signal malfunctioned.
malfunctioned.
141
User’s Manual V2.11.1E
handbrake to ordinary
shooting.
sudden heating
XR1
142
User’s Manual V2.11.1E
ER47 KV feedback
Please contact technical support
malfunctioned
143
User’s Manual V2.11.1E
correction
correction
correction
HEVALUE
144
User’s Manual V2.11.1E
145
User’s Manual V2.11.1E