PLD6000

Medical Diagnostic X-ray System

User’s Manual

Nanjing Perlove Medical Equipment Co., Ltd.


User’s Manual V2.11.1E

Electrical equipment security type: This product is a permanent installation of

ordinary equipment, which is Class I Type B and protection against electric shock.

The mode of operation is continuous operation with intermittent loading.

The manufacturer accepts the responsibility for security, reliability and performance

of the equipment, as show as the following items:

 Assembling, adjustment, operation and maintenance must be carried out by

authorized personnel;

 Electric utilities of the machine room must be complied with related regulations;

 The equipment must be operated in accordance with user’s manual.

Packaging, storage and transportation requirements:

The packaging, storage and transportation of this product should be in accordance

with the provisions of YY/T 1099-2007 "packaging, transportation and storage of

medical X - ray equipment" and "GB/T 191-2008" "packaging and transportation

graphic symbols".

The packaged products should be adapted to transport by air, rail, ship and tertiary

highways.

Transport and Storage:

• Ambient temperature range: -20℃ to 65℃

• Relative humidity: ≤93%

• Atmospheric pressure range: 50kPa to 106kPa

• Indoor environment without corrosive gas

• Articles weighs more that 30kg should be not allow storage on the top of
packing case.

Production date: see product nameplate


User’s Manual V2.11.1E

Dear customer:
Firstly, we would express our gratitude to you and your company to select our
products!
Be sure to read the Product Manual carefully before operating the equipment,
especially the following safety precautions, in order to operate the equipment
correctly for you.
● Please read the following precautions carefully, in order to operate the equipment
correctly for you.
● Be sure to get fully known about the performance and operation of the equipment
before operating the equipment.
● Pay attention to the following matters for setup and storage location of the
equipment:
 Equipment not suitable for use or store in presence of inflammable and
explosives;
 Equipment not suitable for use or store in presence of temperature, humidity and
atmospheric pressure exceeding permitted ranges;
 Equipment not suitable for use or store in presence of ingress of water;
 Avoid tilt, shake and shock (including transportation);
 Pay attention to power supply frequency and voltage as well as allowable
current;
 Keep ventilation and avoid direct sunlight.
● Pay attention to the following matters before using the equipment:
 Electromagnetic emission and electromagnetic immunity of the equipment
should meet requirements specified in YY0505-2012(idt IEC 60601).The
equipment should be operated in typical health care environment specified in
YY0505-2012(idt IEC 60601);
 Confirm power supply connection with the equipment and check whether the
equipment is operated normally;
 Confirm ground connection of the equipment well;
 Reconfirm the power supply voltage.
 In case of equipment failure, instead of disassembling, please paste a note
written with “Break down!” on the equipment, and the Company will assign
authorized personnel to eliminate the failure.
User’s Manual V2.11.1E

● No refitting the equipment.


● Maintenance and inspection
● Inspect the equipment and spare parts regularly;
● If there is a long time to do not operate the equipment, be sure to confirm whether
the equipment can be operated normally and securely before operating it.

The following warning symbols used in this manual. Their application and indication
are described below.

DANGERS ADVISE OF CONDITIONS OR SITUATIONS


THAT IF NOT HEEDED WILL CAUSE DEATH OR
SERIOUS PERSONAL INJURY, AND COULD CAUSE
PART OR ALL OF DAMAGE OF THE EQUIPMENT.

ADVISE OF CONDITIONS OR SITUATIONS THAT IF


NOTHEEDED COULD CAUSE PERSONAL INJURY,OR
DAMAGEOF EQUIPMENT OR DATA.

ADVICE OF CONDITIONS OR SITUATIONS THAT IF


NOT HEEDED COULD CAUSE MISOPERATION AND
ERROR ACTION
User’s Manual V2.11.1E

The following security identity symbols are used in the equipment. Their indications
are described below.
Symbol Description

NOTE:consult the accompanying documentation

Ionizing radiation

Equipment with Type B applied parts

Dangerous voltage

Earth-fault protection

Some components contain material and liquid


that, at the end of the equipment life-cycle, must
be disposed of at the recycling centers appointed
by the Local Health Units.
User’s Manual V2.11.1E

Table of Contents
CHAPTER1 OVERVIEW .................................................................................. 1

1.1 APPLICATION OF THE EQUIPMENT ................................1


1.2 MAIN STRUCTURE AND FEATURE OF THE EQUIPMENT ................ 1
1.3 DAILY PREHEATING PROCESS OF X-RAY TUBE ...................... 4
1.4 DUTY CYCLE LIMIT OF GENERATOR ..............................4
1.5 STANDARDS AND SPECIFICATIONS ............................... 4
1.6 MANUFACTURER’S RESPONSIBILITIES .............................6
1.7 SAFETY PRECAUTIONS ........................................ 6
1.8 USE LOGO IDENTITY ..........................................9
1.9 TRANSPORTATION AND STORAGE ENVIRONMENT CONSTRAINTS ....... 13

CHAPTER2 TECHNICAL PARAME ................................................................ 14

2.1 X-RAY HIGH VOLTAGE GENERATOR ............................. 14


2.2 X-RAY TUBE ASSEMBLY ...................................... 14
2.3 DIAGNOSTIC TABLE ......................................... 20
2.4 COLLIMATOR ............................................... 20
2.5 FLAT PANEL DETECTOR ...................................... 21
2.6 ELECTRICAL OUTPUT DATA ................................... 23
2.7 COMBINATION OF DIAGNOSTIC TABLE LOADING FACTORS ........... 24
2.8 EARTH LEAKAGE CURRENT ................................... 25

CHAPTER3 PRE-INSTALLATION ................................................................... 26

3.1 THE ROOM REQUIREMENTS AND LAYOUTS ....................... 26


3.2 ELECTRICAL REGULATIONS ....................................27
3.3 ENVIRONMENTAL CONDITIONS ................................. 28

CHAPTER4 INSTALLATION OF THE EQUIPMENT ..................................... 29

4.1 OVERVIEW .................................................29


4.2 INSTALLATION OF THE DIAGNOSTIC TABLE ....................... 29
User’s Manual V2.11.1E

4.3 EQUIPMENT CABLE CONNECTION ...............................33


4.4 FINAL VERIFICATION ......................................... 39

CHAPTER5 CALIBRATION OF THE EQUIPMENT .......................................40

5.1 DEBUGGING OF DIAGNOSTIC TABLE .............................40


5.2 SOFTWARE INSTALLATION OF IMAGE ACQUISITION WORKSTATION ..... 53
5.3 SOFTWARE SETTING OF IMAGE ACQUISITION WORKSTATION .......... 57
5.4 CALIBRATION OF GENERATOR ..................................60
5.5 X-RAY AND IMAGE CALIBRATION ............................... 71

CHAPTER6 OPERATION OF THE EQUIPMENT ........................................... 85

6.1 SYMBOL DESCRIPTIONON EQUIPMENT ........................... 85


6.2 USE OF EQUIPMENT - DIGITAL RADIOGRAPHY MODE .............. 131

CHAPTER7 MAINTENANCE .........................................................................133

7.1 MAINTENANCE BY OPERATOR .................................133


7.2 MAINTENANCE BY THE AUTHORIZED TECHNICIAN ................ 135
7.3 THE CYCLE OF MAINTENANCE AND REPAIR ......................137
7.4 FAULT CODE AND TROUBLESHOOTING .......................... 138

CHAPTER8 ATTACHMENT ...........................................................................146

8.1 APPENDIXES ATTACHED WITH FOLLOWING CIRCUITS .............. 146


User’s Manual V2.11.1E

CHAPTER1 OVERVIEW

1.1 Application of the Equipment

Used to radiography patients and obtain single image for clinical diagnosis.
Contraindications: incompatible patients, pregnant women and lactation women,
children, infants, and critically ill patients are not suitable.

1.2 Main Structure and Feature of the Equipment

This equipment mainly consists of System console, high-voltage generator ,


Collimator , X-ray tube assembly , Flat panel detector , DRF acquisition workstation,
diagnostic table see Figure 1-1.

1 -X-ray tube assembly 2 -Collimator 3 -diagnostic table 4 -Flat panel detector


5 -high-voltage generator 6-DRF acquisition workstation 7-System console

Figure 1-1
1
User’s Manual V2.11.1E

Main features:
1. Diagnostic Table
 Visual display man-machine interface and touch screen control makes
man-machine dialogue more direct-viewing, convenient and easy to understand.
 Programmable control device technology with high control precision,
good stability and excellent anti-interference performance.
 Frequency converter control technology, the table operates with soft
startup and soft shutdown, which is smooth without noise and the position, is
more accurate.
 Reliable angular displacement sensor technology with excellent
precision, little dynamic noise and long mechanical service life.
 Tilt range: 90° ~ 0° ~ -25°, it can carry out conventional examinations
about alimentary organs, and it also can be used for a variety of special
examinations.
 Humanized design of diagnostic table side switch, which is more
convenient for positioning when operating local the console.
 FPD and Toshiba high-quality image intensifier, which further improves
image clarity and contrast ratio, and makes it much easier and clearer to identify
the nidus.
 The moving range of spot-film device is wide, so we adopt the operating
mode with machine moving but people unmoving, which thus simplifies the
implementation of a series of inspections from the throat and esophagus to the
lower abdomen.
2. Generator
 Generator with X-ray radiography and fluoroscopy functions and its
maximum output power is 65.5 kW.
 Output voltage is 150kV.
 Smaller and lighter block design
 During exposure, you can adjust kV and mA buttons to achieve
constant X-ray output.
 Large screen LCD flat panel display, which is used to display APR
conditions and practical conventional items.
 User-friendly operating system configuration.
2
User’s Manual V2.11.1E

 A variety of diagnostic procedures with prompt information for the


operator.
3. Flat Panel Detector
 The high-tech FPD flat panel detector converts X-rays directly into
electrical signals for better clarity and sensitivity
 17×17-inch large-sized flat detector, which can completely cover the
full belly and full chest, allows the shooting without any
restrictions.
 Fast direct imaging speed, perfect combination with DFOC
acquisition software, greatly improving system efficiency.
 Amorphous silicon has stable performance and low requirement of
environmental temperature. No irreversible damage caused by
crystallization in high-temperature environment.

4. Powerful Image System


 One button setting exposure conditions and acquisition protocol,
greatly improving the patient pass rate and work efficiency.
 Parallel image processing and symphonic multi-channel filtering
equalization technology can shorten imaging time and improve
imaging quality.
 Open collection and background curve to meet the personalized
diagnostic needs of different specialized hospitals.
 Rich and complete image processing and measurement system, and
provide abundant clinical diagnostic tools.
 Comprehensive DICOM support and network connection processing
functions to meet the needs of digital expansion of the system.
 Fully comply with the latest DICOM3.0 standards, compatible with
DICOM-WORKLIST, and realize all applications such as patient
management/image
management/transmission/display/archive/print/CD burning, etc.
 Provide static and dynamic image output based on WINDOWS
format for easy browsing.

3
User’s Manual V2.11.1E

1.3 Daily Preheating Process of X-ray Tube


This process should be entered into when opening the generator for the first
time or it is restarted after a long time out of service for hours. Exposure at
medium power should be firstly conducted in this process before start of exposure
at the max. mA or kV, in order to reduce the damage of anode and high-voltage
components. No testing setting is required. Please select the following technical
options at startup in order to acquire the optimal stability and reliability: Large
focal, 80 kV and 200 mAs. 1-3 exposures at the interval of 30s.

1.4 Duty Cycle Limit of Generator


Much heat will be generated in the generator and lead to major damage if it is
not fully cooled between exposures. The generator can send out alextremity when
detecting excessive heat relying on its function of monitoring internal duty cycle.

1.5 Standards and Specifications

GB/T 191—2008 Packaging-pictorial Marking for Handling of


Goods
GB 9706.1—2007 Medical Electrical Equipment
Part 1: General requirements for safety (idt
IEC 60601-1)
GB 9706.3—2000 Medical Electrical Equipment
Part 2: Particular requirements for the safety of
high-voltage
generators of diagnostic X-ray generators (idt
IEC 60601-2-7)
GB 9706.11—1997 Medical Electrical Equipment
Part 2: Particular requirements for the safety of
X-ray source
assemblies and X-ray tube assemblies for
medical diagnosis
(idt IEC 60601-2-28)
GB 9706.12—1997 Medical Electrical Equipment

4
User’s Manual V2.11.1E

Part 1: General requirements for safety 3.


Collateral
standard: General requirements for radiation
protection in
diagnostic X-ray equipment (idt IEC 601-1-3)
GB 9706.14—1997 Medical Electrical Equipment
Part 2: Particular requirements for the safety of
associated
equipment requirements of X-ray equipment
(idt IEC 601-2-32)
GB 9706.15—2008 Medical Electrical Equipment
Part 1: General Requirements for Safety 1.
Collateral
standard: Safety requirements for medical
electrical systems
(idt IEC 60601-1-1)
GB/T 19042.1—2003 Evaluation and Routine Testing in Medical
Imaging
departments - Part 3-1: Acceptance tests -
Imaging
performance of X-ray equipment for
radiographic and
radioscopic systems (idt IEC 61223-3-1)
YY 0076—1992 Coating Classifications for Metal Product
Technical Conditions
YY/T 0291—2016 Environmental Requirements and Test Methods
for Medical X-ray Equipment
YY/ T1099-2007 Packaging, Transportation and Storage for
Medical X-ray Equipment
GB/T10151-2008 Medical X-ray Equipment - Specification for
High Voltage
Cable Plug and Socket Connections
YY/T 0741-2018 General Specifications for Digital Medical X-ray

5
User’s Manual V2.11.1E

Radiography System
YY/T0742-2021 Particular Specifications for Gastrointestinal
Diagnostic X-ray Equipment
YY/T0743-2009 Particular Specifications for Gastrointestinal
Diagnostic Table
YY 0505-2012 Medical Electrical Equipment
part 1-2: General Requirements for Safety-
parallel standards:
electromagnetic compatibility requirements and
tests

1.6 Manufacturer’s Responsibilities

The manufacturers equipment in compliance with safety requirements;moreover,


it provides all necessary information for an appropriate use and warnings about the
dangers associated with X-ray generators.
The manufacturers does not take any responsibility for:
 The use of the equipment different from the one it was originally desigened for.
 Damages to the equipment, to the operator, to the patient caused by both wrong
installation and maintenance not in conformity with the procedures contained in
the equipment's user and maintenance manuals, and by incorrect operating
techniques.
 Mechanical and/or electrical modifications performed during or after installation.
Only authorized personnel can provide technical assistance for the
system.
Only authorised personnel may remove the protections of the electrical
cabinet and of the equipment, and access the parts under tension and the moving
parts.

1.7 Safety Precautions

1.7.1 General Rules

The purpose of the Manual is guiding certified fluorographer to operate the


equipment safely. As radioactive equipment, this equipment will be operated
6
User’s Manual V2.11.1E

strictly according to provisions of the Manual and cannot be used for other
purpose except medical radiography. This equipment can only be used by trained
personnel with needed radiation knowledge. Special attention will be paid on the
following matters during using:

1. It is not allowed to use the equipment if it has any electric or mechanical


fault.
2. It is not allowed to removes protective circuit or protective device to make
the equipment working under power on.
3. It is not allowed to use the equipment under the environment of moist,
inflammable and explosive.
4. Consent of our Company or maintenance personnel authorized by the
Company shall be obtained from any modification to the equipment.
5. It shall be ensured that replacing components of the equipment will not
endanger any patient or operator; where necessary, maintenance personnel
of our company may be contacted.
6. Improper operation may result in damage to the equipment; hence regular
inspection and maintenance are necessary.
7. The company will be liable for any personal and property losses caused by
the equipment, excluding that due to non-compliance to the Manual or
modifying the system without authorization.

1.7.2 Electrical Safety

1.Ensure network voltage and current meet requirements of the equipment


on power.
2.The equipment shall be put in well-ventilated room without flammable
and combustible gas leakage.
3.Cleaning, disinfection and sterilization shall be conducted when the
complete machine is power off.
4.If the floor where the high voltage generator locates is not made of wood,
the generator shall be put on wood base.
5. Shut off the machine and switch off the power after using.

7
User’s Manual V2.11.1E

1.7.3 X-ray Safety Protection

Every time before exposure, it shall be ensured to adopt measures to prevent


X-ray and keep in mind the following matters:
1. Wear X-ray protective clothing and goggles for tableside work;
2. Operate in compartment to the greatest extent to be far away from X-ray
source;
3. Low dose exposure to the greatest extent under permitted projection condition.

1.7.4 The Effective Occupation

Figure 1-2 Effective occupation area

Figure 1-3 Effective occupation area

8
User’s Manual V2.11.1E

1.7.5 Environmental Risks and Treatment

When the equipment is retired, the X-ray tube assembly and transformer oil in high
voltage generator shall be reclaimed by specialized industrial waste and waster oil
collection station; while other steel and cooper materials shall be reclaimed by
corresponding department. In particular, the equipment contains the following materials
and/or components:
 “Reducers”: lubricating oil, steel, aluminium
 Motor: iron, copper, hard plastic material casings
 Electric cabinet: iron, aluminium, copper, non-biodegradable plastics,
vetronite for printed circuits.

1.7.6 Retirement of Equipment

The equipment life-cycle is 10 years under the condition of correct using, and the
X-ray tube life-cycle is about 100,000 times, the damaged X-ray tube can be sented to our
company for replacement. If the equipment is damaged and it is beyond repair after
correct using for 10 years, the equipment shall be retired.

1.8 Use Logo Identity

Symbol Description:
Dangerous Ionizing
Radiography
voltage radiation
Small Focal Type-B Protective
spot symbol equipment grounding

Large focal “Speaking” “Listening”


spot symbol device device

Talkback Emergency
Fluoroscopy
conversion stop

9
User’s Manual V2.11.1E

1.8.1 External Label

1. Nameplate of the equipment (see the picture below):

Figure 1-4

2. Nameplate of high voltage generator (see the picture below):

Figure 1-5

3.Nameplate of console (see the picture below):

Figure 1-6

10
User’s Manual V2.11.1E

4.Nameplate of diagnostic table (see the picture below):

Figure 1-7

5. Nameplate of Collimator (see the picture below):

Figure 1-8
6.Nameplate of X-ray tube(see the picture below):

Figure 1-9

11
User’s Manual V2.11.1E

7.Nameplate of DR Acquisition Workstation(see the picture below):

Figure 1-10

8.Nameplate of Flat panel detector(see the picture below):

Figure 1-11

1.8.2 Equipment Nameplate Location

The equipment nameplate location is as shown in Figure 1-12.

Figure 1-12
12
User’s Manual V2.11.1E

1.9 Transportation and Storage Environment Constraints

1. The product must be fixed in the process of transport measures, prevent


wobbling in the boxcar. Transport by gondola car, it must be cover well with
felt cloth. If the product is found damage, fault or any other problem, please
contact with our company and show the following situations:
a) Product model;
b) Production serial number or date;
c) Purchase date;
d) Specific content of the problem to be contacted, such as the fault condition,
the fault point or damage.
2. Transport and storage
a)Environmental temperature:-20°C~55°C.
b)Relative humidity: ≤90%.
c)Atmospheric pressure: 70kPa~106kPa.
d)Indoor environment without corrosive gas.
e)The packaging box should not be stacked with heavy objects over 30 kg.

13
User’s Manual V2.11.1E

CHAPTER2 TECHNICAL PARAME

2.1 X-ray High Voltage Generator

1、 Maximum power: 65.5kW


2、 kV range: Radiographic:40~150kV Fluoroscopic:40~125kV
3、 mA range: Radiographic:10~800 mA
Fluoroscopic:0.5~10 mA(Continuous fluoroscopy)
5~20 mA(Palsed fluoroscopy)
4、 mAs range: 0.1~1000 mAs
5、 Time range:: Radiographic:1.0~10000ms Fluoroscopic:0 to 10min

2.2 X-ray Tube Assembly

Attention:
● X-ray tube assembly equipped to this equipment (hereinafter referred to
as assembly) only applies to medical Radiography.
● Pay attention to that load exceeding the stipulated condition is not
allowed, because overload operation may result in damage to tube sleeve and
X-ray tube.
● The assembly is equipped with normally-closed temperature switch,
which will start when wall temperature of the assembly reaches 75°C.
● The assembly will generate X-ray while working under high voltage;
therefore attention shall be paid to protect it during operation and adjustment.
● Using the assembly in the environment with inflammable or poisonous
gas is not allowed.
● The operation environment temperature is 10℃ ~ 40℃, relative
humidity is 30%~85% and the atmospheric pressure is (70~106)kPa.
● As for transportation and storage environment conditions, the
14
User’s Manual V2.11.1E

temperature range is -20℃ ~ 70℃, the relative humidity is 20%~90% and the
atmospheric pressure is (50~106)kPa.
● Inside the assembly, there’s glass X-ray tube mounted, hence it shall be
handled gently during transportation and installation to avoid damage to X-ray
tube.
● As per the national regulations, installation, adjustment and using of the
assembly shall be carried out by personnel with professional qualification
certificates.
1. X-ray tube assembly type: E7252X
2. X-ray tube component level: Class I Type B
3. X-ray tube parameters:
Model of X-ray tube: XH-106V
Target construction: Rhenium-tungsten faced molybdenum
The reference axis involved in the target Angle and focus characteristics of the X ray
tube: a straight line perpendicular to the axis line of the X ray tube and passing through
the window marker
Anode angle: 12º
Nominal focal spot value: Large focus: 1.2mm Small focus: 0.6mm
Nominal X-ray tube voltage: Fluoroscopic: 125kV Radiographic: 150kV
Nominal anode input power(at 0.1s): Large focus: 75kW Small focus: 27kW
Anode heat content: 210KJ (300kHU)
Filament voltage:
Large focus(5.5A)12.7 ~ 17.1V
Small focus(5.2A)7.0 ~ 9.4V
Filament frequency limits: 0~25 kHz
Permanent filtration: 0.9mm Al/75kV
Anode speed of X-ray tube: 180Hz: Minimum 9700 min-1
Maximum anode heat dissipation: 475W(667HU/s)

15
User’s Manual V2.11.1E

4. Motor Ratings:
Project Starting Running
Driven frequency (Hz) 180 180
Input power (W) 2300 300
Voltage (V) 460 130
Current (A) 5.4 2.0
Min. Speed Up (s) 1.0 —
Capacitor (μF) 3 3
Min. Braking (s) 1.5/90V(DC)

5. Resistance between housing and low voltage terminals: 2MΩ(Min)


6. Available additional filter combination (0.4 - 1.5 mm): ≥1.9mm Al/75kV
7. Radiation protection of the assembly shall meet GB9706.12 III. General
Requirements on Radiation Protection of X-ray Diagnosis Equipment of Part I:
General Safety Requirements of Medical Electric Device (idt IEC601-1-3), i.e.
150kV, 3.4mA.
8. X-ray coverage: 345 ×345mm at SID 835 mm
9. Assembly weight (approximate): 18kg
10. High voltage receptacle: To meet the requirements of IEC 60526
11. Cooling medium: transformer oil with temperature≤75℃
12. Cooling method of assembly: Without Air-circulator
13. X-ray tube assembly heat content: 900kJ (1250kHU)
14. Nominal continuous input power (Without Air-circulator): 200W
(16HU/min)

16
User’s Manual V2.11.1E

15. Emission & filament characteristics as shown in figure 2- 1

Figure 2-1 Emission & Filament Characteristics


16. Maximum rating charts as shown in figure 2-2.

Figure 2-2 Maximum Rating Charts (three-phase 180Hz)

17
User’s Manual V2.11.1E

17. Anode thermal characteristics as show in figure 2-3.

Figure 2-3 Anode thermal characteristics of X-ray tube

19. Cooling characteristics of the component are shown in figure 2-4.

Figure 2-4 thermal characteristics of X-ray tube assemblies

18
User’s Manual V2.11.1E

20. See figure 2-5 for the wiring of the anode plate of the component

Figure 2-5 wiring of the anode


21. See figure 2-6 for the overall dimensions of the components

C :Common
L :Large focus
S :Small focus

Figure 2-6 Outline Drawing

19
User’s Manual V2.11.1E

2.3 Diagnostic Table

1.The table body tilt range:+90°~ 0°~ -25°


2.ateral movesment range of tabletop:220 mm
3.Minimum distance between the compressor and the tabletop surface: no more
than 150mm
4.Compression force of the compressor:80N~130N
5.Bearing capacity of tabletop:200 kg
6.Tabletop X- ray attenuation equivalent: should be less than 1.2mmAl
7.Longitudinal movesment range of spot film device:1000 mm
8. Lift the X-ray tube:
Manual lifting: increase the SID to 1800mm ± 20mm ,Reduce the SID to
1000mm ± 20mm

2.4 Collimator

1.Exposure field:
a.shape of irradiation field: rectangle
b.minimum irradiation field: 0×0 (when the blades coincide)
c.maximum radiation field: 430 mm×430 mm (at SID 1000 mm)
2.Visible light irradiation brightness : Average central illumination
brightness:>100Lux
3.Lamp:LED,d.c. 3.6V,5W
4.Lighting time:5~45s,5s at each step
5.Filtration:≥1.0mmAl;Additional filter:3.5 mmAl
6.Blade driving mode:Remote button operation and setting closed-loop control,
local forward operation button electric
7.Overall dimension:225mm×262mm×210mm(length×width×height)
8.Weight:about 9.0kg
9.Power supply: DC24V

20
User’s Manual V2.11.1E

2.5 Flat Panel Detector

1、Plate type: DRF-1717A

2、Tablet parameters:

Active Area: 434 mm × 434 mm

Pixel Array: 3072 ×3072 pixels

Pixel Pitch: 139 μm

Limiting resolution: ≥3.7 lp/mm typ

Energy range: 40 - 150kvp

A / D conversion: 16 bit

Power input: DC 24V

Quantum detection efficiency:

Radiation RQA/(including Gy) Spatial frequency /mm-1 DQE(Horizontal/vertical)

0 lp/mm ≥62%
0.5 lp/mm ≥56%
1.0 lp/mm ≥50%
1.5 lp/mm ≥45%
2.0 lp/mm ≥40%
5/2µGy 2.5 lp/mm ≥35%
3.0 lp/mm ≥30%
3.5 lp/mm ≥25%
4.0 lp/mm ≥21%
4.5 lp/mm ≥16%
5.0 lp/mm ≥11%

21
User’s Manual V2.11.1E

Frame rate:
≥2fps (1x1binning, 3072 x3072, 43cmx43cm
FOV)
≥6fps (1x1 binning, 1536x1536, 23cmx23cm
FOV)
≥13fps (2x2 binning, 1408x1408, 43cmx43cm
FOV)
≥20fps (1x1 binning, 1024x1024,
15.6cmx15.6cm FOV)
≥30fps (2x2 binning, 768x768,
23.7cmx23.7cm FOV)
Modulation transfer function:
1lp/mm ≥65%
2lp/mm ≥35%
3lp/mm ≥20%
4lp/mm ≥10%
5lp/mm ≥7%

3、Components and characteristics

Cooling: Natural air cooling

Power: DC 24V

Size: 460mm (l) × 460mm (w) × 28mm (h)

Weight: 6 Kg

4、Environment conditions

temperature: 10℃ ~ 40 ℃

Relative humidity: 30 % ~ 75 % RH

atmospheric pressure: 700~1060hPa

22
User’s Manual V2.11.1E

5、Storage conditions

temperature: -20 ℃ ~ 55 ℃

Relative humidity: 10% ~ 90 % RH

atmospheric pressure: 700~1060hPa

2.6 Electrical Output Data


Fluoroscopy:
● Tube voltage: 40kV~125kV, continuous adjustment
● Tube current: 0.5mA-10mA (continuous fluoroscopy)
5mA~20mA(palsed fluoroscopy)
Maximum power output: 80kV×10mA=0.8 kW (continuous fluoroscopy)
Continuous X-ray tube maximum voltage: 125kV
Corresponding maximum tube current: 6.3mA
Continuous X-ray tube current: 10mA
Corresponding maximum tube voltage: 80kV
Maximum power output: 125kV×20mA=2.5kW (palsed fluoroscopy)
palsed fluoroscopy X-ray tube maximum voltage: 125kV
Corresponding maximum tube current: 20mA
palsed fluoroscopy X-ray tube maximum current: 20mA
Corresponding maximum tube voltage: 125kV
Radiography:
● Tube voltage: 40kV ~ 150kV, continuously adjust
● Tube current: 10mA ~ 800mA, each value is: (selected by R '10 coefficient
value)
● Exposure time: 0.001s-10s. The value of each step is: (selected by R '10
coefficient value)
● Nominal X-ray tube voltage: 150kV corresponding maximum tube current:
400mA
● Nominal X-ray tube current: 800mA Corresponding maximum tube voltage:
79kV
● Maximum output power: 104kV× 630mA=65.5kW (104kV, 630mA)
23
User’s Manual V2.11.1E

● Nominal electric power: 100kV× 630mA=63kW (100kV、630mA、0.1s)


●Technical parameters of automatic exposure control:
Automatic exposure control shortest irradiation time: 1ms
Minimum irradiation conditions for automatic exposure control:
Nominal X-ray tube voltage: 150kV corresponding maximum tube current:
400mA
Nominal X-ray tube current: 800mA corresponding maximum tube voltage: 79kV

2.7 Combination of Diagnostic Table Loading Factors

Focus mA kV s

10、12.5、16、20、
150 10
25、32

150 8
40
119 10
150 4
50
109 10
Small focus
150 1.6
63
89 10
150 0.63
80
69 10
150 0.2
100
59 10
150 2
125
69 8
150 1
160
79 5
150 0.63
200
59 5
150 0.2
250
59 3.2
Large focus
320 150 0.4

24
User’s Manual V2.11.1E

69 2.5
150 0.25
400
59 2.5
131 0.16
500
49 2
104 0.125
630
49 1.25
81 0.16
800
49 0.8

2.8 Earth Leakage Current

Normal status:≤5mA;Single failure status: ≤10mA

25
User’s Manual V2.11.1E

CHAPTER3 PRE-INSTALLATION

3.1 The Room Requirements and Layouts

1. Machine Room Requirements


a)The area of the X-ray room shall not be less than 20m2;
b)The control room covers an area of 5000×2500mm, which should be connected
to the machine room, and the observation window should be equipped with lead glass;
c)The height of the machine room with a clearance of at least 2905mm;
d)Door of the machine room, 1500mm wide and 2300mm high;
e ) The diagnostic table should be fixed on the ground directly, based on the
foundation should be level and can carry 1500 kg load;

The foundation shall adopt cement with grade no less than 400"
and casting thickness no less than 200mm.

f) Cable trench and cable trench cover shall be installed flush with the ground;
g) Room should be equipped with special grounding line, grounding resistance is
not greater than 4 Ω;
h)The X-ray protection ability of the machine room shall meet the requirements
of national regulations.

26
User’s Manual V2.11.1E

2. Layout of Machine Room

Figure 3-1 machine room layout

3.2 Electrical Regulations

1. Power condition:
a) Power supply voltage: a.c.380V±38V, 50/60Hz±1Hz;
b) The power resistance should be no greater than 0.17 Ω;
c) Power supply capacity shall not be less than 105kVA.
2. The power distribution box shall be equipped with a three-phase air switch with
rated voltage of 500V and rated current of 130A as the disconnection switch with
the power supply network. The power distribution box shall be firmly fixed on the
wall with a floor level of not less than 1.5m away from the room, and a power
source with a power capacity of not less than 105kVA and an independent ground
wire (not more than 4 Ω).

27
User’s Manual V2.11.1E

3.3 Environmental Conditions

Ambient temperature: 10℃~40℃


Relative humidity: 30%~75%
Atmospheric pressure: 70kPa~106kPa

28
User’s Manual V2.11.1E

CHAPTER4 INSTALLATION OF THE EQUIPMENT

4.1 Overview
The equipment must be installed and adjusted by professional personnel of our
Company or authorized by our Company.
Before opening the packing case, it shall be stored in the rain-proof and dust-free
environment; and board of box shall be opened reasonably while opening the case.
While handling the parts, attention shall be paid to avoid scratching parts surface;
check according to the packing list after opening the case and installation shall only
be conducted after confirming the machine is defect and omission; any problem shall
be informed to the Company for us to dispose in time.

4.2 Installation of the Diagnostic Table


4.2.1 Diagnostic Table Base Positioning

Figure 4-1 The installation of the base


1.According to the figure 4-1, the diagnostic table installation position is
measured, the table is placed well, the diagnostic table is fixed on the ground base
directly, the base level can bear the load of 1500Kg, the cement mark of the
foundation is not less than 400, and the pouring thickness is not less than 200mm.
2.The table base according to the requirements put good, as shown with
percussion drill in the bolt hole on the ground dozen Ø 14 deep 60 mm hole

29
User’s Manual V2.11.1E

3.Fix the base with four M10×100mm expansion bolts.

4.2.2 Installation of the Tube-Collimator Combination

1) Install the diagnostic table, high voltage generator and console panel
according to the layout requirements of the machine room, and connect the
wiring according to the circuit diagram.
2) Put through power supply, the table body tilt to 90 °, with X-ray tube column.
3) Installation of the Tube-Collimator combination

1-tube cross arm 2-Hexagon socket head screw M10X25


3-Hexagon socket head screw M6x25 4-tube connection plate
5-Collimator 6-cross recessed countersunk head screw M6x25
7-Subplate of collimator 8-X ray tube housing mounting plate
9-Hexagon socket head screw M6X20 10-fixed collar connection plate
11-X-ray tube assembly 12-fixed collar assembly
13-Hexagon socket head screw M5x20 14-Hexagon socket head screw M4X10
15-fan 16-X-ray tube housing 17-Hexagon socket head screw M4×15
Figure 4-2 The installation diagram of the tube-collimator combination
1. Fixed the fixed collar assembly on the tube with four Hexagon socket head screw
M5x20,then fixed the fixed collar connection plate on the fixed collar assembly
with four hexagon socket head screw M6X20,fixed the tube connection plate on

30
User’s Manual V2.11.1E

the fixed collar connection plate with four hexagon socket head screw M6X25;
2. Nsert the tube connection plate into the mounting hole of the cross arm of the
tube, and fix it with three M10x25 inner hexagon screws. Pay attention to the
direction of installing the tube (the direction shown in the figure, with the cathode
facing up).
3. Stack the subplate of collimator, tube mounting plate and connecting plate
together, and connect to the tube with cross recessed countersunk head screw
M6x25
4. After the collimator adapting piece is installed,back-out thecollimator with
Hexagon socket head screwM5×40,not fall off..Install collimator ,rotate the
collimator,align the notch with the notch of the connecting plate, then rotate the
collimator to the appropriate position and tighten the M5x40 socket hex screw.
5. Install the fan on the tube housing and secure the fan to the tube housing using
four m4x10 hex socket screws.
6. Install the outer cover of the tube, insert the T-shaped slot of the outer cover into
the notch of the mounting plate of the tube, and connect the outer cover of the
tube with the transverse arm of the tube by using two inner hexagon screws
M4x15

4.2.3 Installation of the Flat Panel Dectetor

1-film holder 2-spout 3-spout support plate


4-hexagon socket cap screws M4×16
5-cross recess pan head screw M5×8 6-Ionization chamber 7-grid
31
User’s Manual V2.11.1E

8-Dynamic flat panel detector 9-flat baffle


10-cross recess pan head screw M4×8 11-Protection of lead
12-Flat mounting plate 13-hexagon socket cap screws M5×12
14-film holder front cover 15-cross recess pan head screw M4x12
Figure 4-3 The installation diagram of the Flat panel dectetor
1. Fixed the spout support plate(part 3 )on both sides of the film holder(part 1 ) with
six cross recess pan head screws M5×12.
2. Installation of the Ionization chamber:put the Ionization chamber (part 6 )on
spout support plate(part 3 ),press the ionization chamber with spout (part 2) on
both sides respectively,fixed the spout and spout support plate with six
hexagon socket cap screws M4×16.
3. Installation of the flat panel detector: put the dynamic flat panel detector (part 8)
on flat mounting plate (part 12),put flat gasket and spring washer ,and connect
with four cross recess pan head screws M4×8(part 10),Secure the plate
mounting plate to the plate box (part 1) with six hexagon socket screws M5x12
(piece 13), flat washers, and spring washers. Ensure that the screws are consistent
with the center line of the tube.
1.Flat baffle installation:First, stick the protective lead (part 11) on the plate
baffle plate with AB adhesive, and then install the plate baffle plate (part 10) on
the plate box (part 1) with the cross recut pan head screw (part 10), flat washer
and spring washer.
4. Installation of front cover sealing plate of film holder: fix the front cover sealing
plate of plate box (part 14) on the film holder (part 1) with M4x12 cross-recessed
countersunk head screws (part 15).
5. Installation of grid:Insert the grid (part 8) into the left and right chutes in the film
holder, and the assembly is completed.

32
User’s Manual V2.11.1E

4.3 Equipment Cable Connection

4.3.1 Power Cables and Protective Grounding Wires

1. Main power cable D:

High voltage generator Main power cable D Main Power Supply


X1 Terminal (XJZ10.04) Terminal
U L1
V L2
W —— —— L3
N N

2.Diagnostic Table Power Cable A:

Diagnostic Table Diagnostic Table Power Cable A High voltage generator


X1 Terminal (XJZ20.01) X2 Terminal
X1-1 A U
X1-2 B V
X1-3 C W
X1-4 N N

3.Console Panel Power Cable:

Console Console Panel Power Cable Diagnostic Table


PCB60.32 (XJZ20.10) PCB20.20
J1-1 2125 J33-7
J1-2 2124 J33-8

4.Grounding Wires: BRV 4mm2,yellow-green line,length is 8 m,connect from


the console ground point to the ground point of the diagnostic table base
electrical mounting plate
5.Other protective earthing wires: console protective earthing wire with a length
of 5m is connected from the console base protective earthing terminal to the
33
User’s Manual V2.11.1E

independent protective earthing wire of the wall gate; Detector protective


ground wire, length of 5m, connected from the protective ground terminal of
the diagnostic table base to the protective ground of the plate detector; The
x-ray tube protective earthing wire is 12m in length, which is connected by the
protective earthing terminal of the high voltage generator tank to the
protective earthing point E of the x-ray tube.

Protective earthing line to ensure the safety of equipment


and staff, the user must provide independent meet the
national standards for this equipment is the protection of the ground line, the
earthing resistance must not be greater than 4 Ω.

Wall gate wiring, must ensure that the switch is in the open
state, unless the installation wiring has been completed,
otherwise do not allow closing.

4.3.2 High Voltage Generator Cable Connection

1.CAN BUS Communication Cable:

High voltage generator CAN BUS Communication Cable CAN-USB converter


Interface board (XJZ10.80) Plug
PT4 —— —— CAN1

2.USB Cable:
Computer CAN-USB converter
USB Cable
Plug Plug
USB —— —— USB

34
User’s Manual V2.11.1E

3.Generator Control Cable O:

Control box Generator Control Cable O HV Generator


PCB20.18 (XJZ50.04) Terminal
2885 X3-4
2553 X3-5
2883 X3-6
2870 X3-10
J7 2872 X3-11
2869 X3-28
2871 X3-29
2875 X3-30
2873 X3-31

4. Stator cable B:

High voltage generator Stator cable B X-ray tube


PD3 Terminal (XJZ10.51) Terminal
1 1 1
2 3 3
3 2 2
5 4 5
6 5 6
FG —— —— FG

5. High Voltage Cable:

High Voltage Generator X-ray Tube


High Voltage Cable
Interface Interface

-CATHODE- CATHODE -CATHODE-


+ANODE+ ANODE +ANODE+

35
User’s Manual V2.11.1E

High voltage cable bending should not be less than 10 times


of its diameter, as far as possible to avoid bending. Excessive
bending will cause high voltage cable breakdown,
endangering the safety of equipment and personnel.

4.3.3 Diagnostic Table Cable Connection

1.Collimator Power Cable A:

Diagnostic Table Collimator Power Cable A Collimator


PCB20.20 (XJZ10.42) plug
J26-1 2502
2P Aviation Socket
J26-2 2511

2.Collimator Communication Composise Cable B:

Diagnostic Table Collimator Communication Collimator


Composise Cable B
PCB20.20 plug
(XJZ10.46)
J10 —— —— 6P Aviation Socket

3.Diagnostic Table Control and Communication Composise Cable B:

Diagnostic Table Diagnostic Table Control and Control Box


Communication Composise Cable B
plug plug
(XJZ20.24)
J11 —— —— J2

4.Machine Room Speaker Cable :


Intercom System Machine Room Speaker cable Speaker
PCB40.02 (XJZ40.23) plug
J4 —— —— +/-

36
User’s Manual V2.11.1E

5.Machine Room Microphone cable:


Intercom System Machine Room Microphone cable Diagnostic Table
PCB40.02 (XJZ40.24) plug
J3 —— —— X6

6.Console Panel Speaker cable:


Intercom System Console Panel Speaker cable Speaker
PCB40.02 (XJZ40.22) plug
J5 —— —— +/-

7.Generator Emergency Stop Cable J:


Diagnostic Table Generator Emergency Stop Cable J High voltage generator
PCB20.20 (XJZ50.54) plug
J13-1 2595 PP8B-2
J13-2 2596 PP8B-3

4.3.4 Cable Connection of Image System

1.Gigabit Ethernet cable G:


Detector Computer
Gigabit Ethernet Cable G (XJZ30.37)
plug plug
Internet —— —— LAN

2.Serial Port Cable:


Computer Serial Port Cable Control Box
plug (DRA80.24) plug
COM —— —— DB9/M

37
User’s Manual V2.11.1E

3.Generator SYNC Cable:


Generator Generator SYNC Cable Control Box
plug (DRA80.25) plug
X3-1 3852
X3-7 3850
X3-8 3851
DB37/F
X3-9 3001
X3-26 3854
X3-27 3855

4.Detector SYNC Cable A:


Detector Detector SYNC Cable A Control Box
plug (DRA80.20) plug
DB15/M —— —— DB37/M

5.Detector Power Cable A:


Switching Power Supply Detector Power Cable A Detector
plug (DRA80.21) plug
+24V 3004
POWER
GND 3001

6. AEC Cable B:
Generator AEC Cable B Ion Chamber
Interface board (XJZ50.84) plug
PT8 —— —— plug

7.Detector Exposure Control Cable A:


Control Box Detector Exposure Control Cable A Control Box
PCB20.18 (KZT70.41) plug
J7
DB9/F
J9 —— ——

38
User’s Manual V2.11.1E

4.4 Final Verification

The base of the diagnostic table is fixed, the X-ray tube Collimator is installed
in the center, the image intensifier is fixed, the image intensifier TV system is
installed accurately, and the cable connection of the equipment is confirmed to be
accurate.

39
User’s Manual V2.11.1E

CHAPTER5 CALIBRATION OF THE EQUIPMENT

5.1 Debugging of Diagnostic Table

Figure 5-1 Symbols of buttons and handles in debugging

Description:
1:Manual and automatic switch button of the collimator
2:compressor tracking button
3:Table negative Angle button
4:tilt down button
5:tilt up button
6:column up button
7:column down button
8、9:Subfunction switch button(Only used in debug mode)
1.Debugging Need to Know

Limit Debugging
Enter working state,cover the corresponding optocoupler with a visor or hold the
corresponding stroke switch limit by hand ,at the same time, check whether the
corresponding LED lamp on the PCB board is on or off normally(General cover limit
or press the travel switch limit corresponding to the LED off, release the visor or
release the hand LED light ).Be sure to ensure that each limit is working properly
before the next step of debugging.

40
User’s Manual V2.11.1E

Motor Direction Debugging


On the basis of the completion of limit debugging, the direction of each motor is
debugged:
Enter the debugging state,debug the direction of up and down motor,column
moving motor,SID motor ,tabletop transverse motor,compressor motor,Rotary foot
pedal motor(optional),ensure that the motor direction are accurate,If the direction of
the motor is opposite, replace the two wiring of the motor. The common end drive
line of the compressor is unchanged.

Potentiometer Debugging
During debugging, the potentiometer is released from the gear;The AD value
increases when the table up,The AD value decreases when the table down.The AD
value increases when the column up,The AD value decreases when the column
down.The potentiometer AD value can be measured with a multimeter,The voltage of
4%~96% VCC is normal or the range of 32~990 displayed on the control panel is
normal. If the range is beyond this, the potentiometer is abnormal and the parameters
cannot be saved. In normal mode, the corresponding error code of the potentiometer is
displayed.

2.Enter debugging state

Ensure the emergency stop switch is unscrewed,press button on control

box,then press button to starting up the machine,If there is no error, the digital
tube shows C01, indicating that the debugging function has been entered,press

button switch between C01 and C19.If there are sub-functions in the main
function, the operation handle will switch back to the sub-function switch key, and the
sub-function code will be switched from 01 to 99 (according to the actual need) .

41
User’s Manual V2.11.1E

The description of main function code and sub-function code is shown in Table 2:
Table 2
Main function Sub-function Function button Description of function button
code code
table down

table up

01 table 90° The current AD


value is saved
C01 02 table 0° The current AD value
is saved
03 table negative protection The
current AD value is saved
04 table -15° The current AD
value is saved
SID decrease

SID increase
C02
01 Max SID,AD value is saved

02 Min SID,AD value is saved

Column down

Column up
C04
01 Save the maximum AD value
when the column up
02 Save the minimum AD value
when the column down
03 Column protection AD value
save
Enter the show

Tabletop movesment direction


set value switch plus

42
User’s Manual V2.11.1E

Tabletop movesment direction


set value switch minus
C06 Tabletop ingression

Tabletop relocation

01 Save the set value of tabletop


movesment directioN
C07 Compressor down

Compressor up

C08 Pedal reversal

Pedal corotation

Subfunction Enter the show


switch button
Mosaic image overlap width
C09 increased
Mosaic image overlap width
decreased
01 Mosaic image overlap width
saved
C10 Start aging action
Stop aging action
C19 Subfunction Enter the show
switch button

Table size switch plus

Table size switch minus

01 Save the table size

43
User’s Manual V2.11.1E

5.1.1 Diagnostic Table Tilt Up and Down Debugging

Note:before the table tilt up and down debugging,make sure the emergency
stop works properly.In the debugging state, the table has no automatic function.
When the table rotates, be sure to moves the column to the middle position to
prevent collision during the table movesment.
1.Table up and down travel limit debugging:The photoelectric coupler is used as
the limit protection device for thetable, and the potentiometer is used as the limit
device. It is installed on the optical coupling control in the base assembly, as shown in
Figure 5-2.

1-table 90° limit optocoupler 2-table negative angle protection limit


optocoupler 3-table -25° limit optocoupler 4-table 0° limit optocoupler
5-table scram protection
Figure 5-2 Tilt Up and Down limit installation mechanism

At debugging mode “C01”,first,insert a block of suitable size into the infrared


transmission slot of the photoelectric coupler in the direction of +90° from the table,
and then check whether the LED4 light on the corresponding PCB is off. When
removesd, LED4 light on the PCB is on, if it is normal, otherwise check the
corresponding line and the corresponding photoelectric coupler until
normal. Similarly, the -25° photoelectric coupling of the table was tested. Check the
LED1 on the PCB, and finally confirm that the visor is installed.
2.Action debugging :The handle of the control box on the control console moves to
44
User’s Manual V2.11.1E

,LED14, LED17 bright,LED16 off,table down , The handle of the control box

on the control console moves to ,LED14 、 LED16 bright,LED17 off, table


up.Make sure that the table is in the horizontal state, and the starting and reversing
direction is the same as the direction indicated by the control box mask, then the
table's starting and reversing direction is correct. If it is opposite, adjust the motor
input drive line, adjust any two motor drive lines marked 2401, 2402 and 2403 to
make the motor reverse phase.

3.table up and down travel limit debugging:The table is inverted using potentiometer
as a limiting device,as shown in Figure 5-2. The potentiometer must be released from
gear before debugging. Make sure the potentiometer is installed after adjustment.

Table up and down AD value debugging


Control box on the operation console operating handle backwards respectively
subfunction switch key switch "01", "02", "03", "04" .First , the table inverted angle
to 37.5°, the table inverted potentiometer value adjustment is about 2.5V or view the
console control box display "512", install the potentiometer. Switch to "01", the table

turns to 90°, press the button to save the corresponding current AD

value; Switch to "02", adjust the table to 0° level, press the button to save the

corresponding current AD value; Switch to "04", set the table to -25°, press the
button to save the corresponding current AD value; Switch to the negative protection

position (about -7°) of table 03, and press the button to save the corresponding
current AD value.

After the completion of debugging, the test of the inverted function should
be carried out: special attention, before testing the inverted action, first confirm
that the inverted emergency stop can work normally.In the debugging state,
there is no automatic function. When the table rotates, pay attention to prevent
collision during the table movesment.

45
User’s Manual V2.11.1E

The handle of the control box on the control console moves to or , or

press the table up and down button on the near panel or ,control the table
up and down, clockwise can reach +90°, rotating to the limit or let go immediately
stop; Counterclockwise direction can reach -25°, limit or release immediately stop, if
from above the level of counterclockwise rotation, to the level will stop, to continue to
the negative angle of the table , there are two operations : One is controlled by the
control box, the other is controlled by the buttons on the near control board. When
operating through the control box, it is necessary to release the operation handle first,

then press and hold the table negative angle function button on the control box,

again operate the operation handle on the control box to reverse , the table

movess to negative angle, release the button or operation handle, or the table
reaches the negative angle limit, the table negative angle motion stops. When using

the near control board, be sure to release the operation button first, press the near

control board button again, the table movess to the negative angle, release the

button or the table to the negative angle limit, the negative angle table movesment
stops. If the tablestarts to rotate clockwise at a negative angle and stops when it is
horizontal, if you want to continue to moves to +90°, you must release the operation
button or handle first, and then start the movesment operation, you can continue to
rotate to +90°, release the operation button or handle or to +90° limit, and the
tablestarts to stop.
Finally, adjust the position of the photoelectric coupler of each angle of the table
to meet the needs of the table rotation angle.

5.1.2 Debugging of Longitudinal movesment of X-ray Tube Column

1.Debugging of the travel limit of the x-ray tube column : The longitudinal
movesment of the tube column uses a travel switch as a limit protection device,As a
limiting device, the potentiometer is installed on the travel switch and potentiometer
in the table assembly, as shown in Figure 5-3, 5-4 .
In the debugging mode "C04", hold the tube column down travel switch limit,
and then check whether the corresponding LED6 light on the PCB is off. When you
46
User’s Manual V2.11.1E

release the hand, LED6 light on the PCB is on. If yes, it is normal; otherwise, check
the corresponding line and the corresponding travel switch until normal. Similarly,
test the limit of the column upward travel switch and check LED8 lamp.

1-column move down limit optocoupler 2-column move up limit optocoupler


Figure 5-3 Sliding Limit of Vertical Column
2. Debugging of column direction:Under the debugging mode "C04", the

operation handle of the control box on the operation console moves to , LED15
and LED17 are on, LED16 is off, the column movess down, the operation handle

moves to , LED15 and LED16 are on, LED17 is off, the column movess up. If
the tube column movess in the opposite direction, it is necessary to adjust any two
input drive lines of the motor, and confirm that the limit collision block is installed.
3. Tube column moving potentiometer debugging

1-Tube-column potentiometer
Figure 5-4 Tube-column longitudinal shift potentiometer
Before the potentiometer is debuged, loosen and disengage the gear; Install it
after debugging.
After the debugging of limit and motor direction is completed, switch "01", "02"
47
User’s Manual V2.11.1E

and "03" respectively by the reverse sub-function switch key of the control box
operation handle on the operation control. moves the tube column to the middle
position of the table, adjust the tube column moving potentiometer voltage to 2.5V or
view the console display "512", install the potentiometer . The handle of the control

box on the control console moves to ,the column up,AD value increase.When the

column stops moving to the upper limit, switch to 01 and press the button on the
control box, save the maximum AD value of the tube column.The handle of the

control box on the control console moves to ,the column down,AD value
decrease,when moving to the lower limit, the column stops moving. Switch to 02 and

press the button on the control box , save the minimum AD value of the tube

column.The handle of the control box on the control console moves to ,moves

the column down to the lower limit and press the button on the control box to
switch,enter"C01”,The handle of the control box on the control console moves

to ,swing the table to a negative angle,press the button on the control box
to switch,return “C04”,The handle of the control box on the control console moves

to ,moves the column up slowly until the column housing is close to the ground.

Switch to 03 and press the button on the control box to save the column
protection AD value.
After the debugging is completed, the longitudinal movesment function of the
tube column should be tested:

Keep the table level, then operate the handle to reverse or , or press the

column moves button or on the near control panel to control the movesment
of the column of the tube. When the column of the tube is pressed to the limit of the
travel switch, the column of the tube should stop moving immediately, and the limit
will be effective. Control the operation handle or press the column moving button on
the near control panel, you can control the movesment of the tube rack, release the
handle or release the button to stop immediately, the column moving stroke is 970 ㎜,

48
User’s Manual V2.11.1E

0 ㎜, that is, the column moving total stroke is 970 ㎜.


Finally, adjust the limit stroke switch of column movesment to meet the need of
longitudinal stroke of column movesment of tube.

5.1.3 Lateral Movesment Debugging of the Tabletop

1) Table top surface lateral movement direction debugging: click the Table top

surface access button in the local control panel or (ensure that the lateral
movement of the Table top surface is not in the limit state). If the sliding direction is
the same as the direction marked by the button, the lateral movement direction of the
table surface is correct. If it is opposite, the input drive lines of two motors at the
non-public end should be adjusted to invert the motor.
2) table side movement travel limit debugging: table side movement USES
photoelectric coupler as limit device, which is installed in table assembly.

1—Table internal displacement limit photocoupler


2—Table face outer-limiting photocoupler
Figure 5-3 Lateral Movement Limits of Table Surface
Control the operating handle or press the table surface movement button on the
local control panel to control the internal and external movement of the table surface

or , release the handle or release the button to stop immediately, and stop
immediately when reaching the limit.
Finally, the position of the photoelectric coupler of each travel limit is adjusted to
meet the designed table surface travel.

49
User’s Manual V2.11.1E

5.1.4 Debugging of the Advance and Backward Movesment of the


Compressor

1.Compressor travel limit debugging:A travel switch is used to limit the up and
down of the compressor, and is installed in the assembly of the compressor, as shown
in Figure 5-5.
In the debugging mode "C07", press the ascending stroke switch by hand, LED5

off, let go of light, the limit is normal. The operate handle upside down , LED49
bright, LED48 off, the compressor up.Pressing the upper limit switch by hand will

not continue to rise,it is normal.The operate handle upside down ,LED48


bright, LED49 off,the compressor down.it is normal.When the compressor reaches
the limit of the travel switch, it should be able to stop the movesment. Confirm that
the limit block is installed.
2.Debugging of the motion direction of the compressor:Under normal conditions,

the handle of the control box on the operation console moves to , the compressor

is down, moves to the compressor is up. If the reverse should be done, the input
drive line of the compressor motor should be adjusted, and the non-common two
motor drive lines of 2407, 2408 and 2409 at the lower end of the terminal block can
be reversed to make the compressor motor.

1—compressor move up limit


Figure 5-5 Compressor Stroke Switch
50
User’s Manual V2.11.1E

Finally, adjust the position of the limit stroke switch of the compressor to meet
the needs of the advance and retreat stroke of the compressor.

5.1.5 Debugging of tube lifting

SID limit Debugging:The travel limit of the tube uses a travel switch as the
limiting device, as shown in Figure 5-6.

1—The tube move down limit 2—The tube move up limit


Figure 5-6 The limit of tube lifting
First,switch SW2 2, 4 to the ON side that on PCB20.20 B5.
Under the debugging state "C02", keep the table 90°, the operation handle

movess to , LED46 is on, LED47 is off, the tube rises; The operating handle

moves to , LED47 is on, LED46 is off, and the tube drops. Press the
corresponding travel switch by hand, if the tube stops moving immediately, then the
travel switch limit is normal, otherwise, check the corresponding line and the
corresponding limit switch.
Finally, adjust the position of the tube lift limit switch to meet the needs of SID.

51
User’s Manual V2.11.1E

5.1.6 Debugging of Pedal Rotation

The rotating pedal uses a DC motor,under the debugging state "C08", the

operation handle moves to ,LED42 on, LED43 off, and the pedal is turning

forward; The operating handle moves to ,to LED43 on and LED42 off, and the
pedal is reversed. If the direction is opposite, replace the two motor drive lines 2415
and 2416 at the lower end of the terminal block.

5.1.7 Mosaic image overlap width Setting

In the mode of "C09", the handle is inverted to the subfunction switch key to
enter the subfunction. At this time, SID nix tube will display the setting value of
overlap width of mosaic image, and the third nix tube will display the saving value of

overlap width of mosaic image. Operate the handle moves to or ,select


the desired set value of the overlap width of the mosaic image, and the set value

ranges from 2 to 15; Press button save set value, save after the pendulum third
digital tube update display saved value.

5.1.8 Table type Setting

In "C19" mode, the operation handle is reversed to the subfunction switch key to
enter the subfunction, SID nixie tube shows the set value of table type, and the third
nixie tube shows the saved value of table type. The operation handle moves to

or , select the current actual table type set value, the table type of the

equipment should be set to "20F"; Press to save the set table type, after saving
the third digital tube update display saved value.

5.1.9 Aging test

When all debugging work is complete and the test is normal in normal mode,
machine aging tests can be performed if necessary. Enter the debugging mode "C10"

and press start aging test. At this time, the second nixie tube displays the value
of aging step, and the third nixie tube displays the value of aging waiting
52
User’s Manual V2.11.1E

time; During the aging process, press again to exit the aging test and return to
the debugging mode.

5.2 Software Installation of image Acquisition Workstation

5.2.1 DFOC Software Installation

For the official version, user needs to start the installation from the CD-ROM, if
it isn’t automatically start, double click “DFOC_XXXX.exe”; for the demo version,
mostly the user will get an installation package, and then just double click
“DFOC_XXXX _DEMO.exe”.
Double click to start the installation, the following steps will display:

Select language
If it is to upgrade the software, there would pop up the following interface:

Uninstall option

53
User’s Manual V2.11.1E

Uninstall Existing Version

Select folder

Confirm installation

54
User’s Manual V2.11.1E

Installation process
Install the granddog driver. (For official version only)

Start installation

Finish installation
After successful installation, there would display the following interface:

.
Finish installation
55
User’s Manual V2.11.1E

After the installation steps above, the program will create shortcut of
“DRConsole.exe” and “DRDongle.exe”. User can double click the “DRConsole.exe”
to start the program, or double click the “DRDongle.exe” to register the software.
[Warning] if the official version is unregistered, the program will show message
in Figure 2.8, then the program will automatically shut down

Unauthorized messages

5.2.2 Software Registration

The official version of the software will come with a USB hardware dongle.
After software installation; the user inserts the dongle into the computer's USB port to

run dongle "111". Click the " " button. If the dongle driver is installed
correctly and the USB dongle has been successfully identified by the computer, the
dongle serial number and authorization information will appear simultaneously.

If no list of authorization information appears, check to see if the USB dongle


driver was installed successfully or if the USB dongle has been read in the computer
device manager.

56
User’s Manual V2.11.1E

5.3 Software Setting of Image Acquisition Workstation


5.3.1 Log on

After the operating system starts, the login window displays. Enter the correct
user name and password on the system login interface to enter:

In the user login window, enter the correct user number and password and press

" " to enter the loading interface. If you do not want to enter the system, click

" " to exit.

5.3.2 Hardware Setting

Click " " button from the navigation bar to enter the management options
and select hardware configuration:

Click the " " button to enter the hardware configuration check box,
including Syncbox, high voltage, detector, collimator, DAP.

57
User’s Manual V2.11.1E

5.3.2.1 Syncbox

Settings for Syncbox include:


Port: 2 (set according to the actual sync box interface)
SyncboxType: Single FPD (set according to the actual number of tablets)
Log: normal mode
High voltage: RuiYing (set according to the actual high voltage model)
Detector: CareRAY (set according to the actual tablet model)
5.3.2.2 High-voltage generator

High voltage generator: RuiYing (set according to the actual high voltage model)
Log level: normal mode

58
User’s Manual V2.11.1E

5.3.2.3 Detector

Detector: CareRay (set according to the actual flat detector model);


Log level: normal
5.3.2.4 Collimator

Collimator:Type1 (set according to the actual collimator);


Port:6 (set according to the actual sync box interface).

5.3.3 Workstation IP Settings

Under WIN10 system, open the control panel of the network connection, find
local address workstations with plate connection, double-click to open and click

59
User’s Manual V2.11.1E

"properties" find "Internet protocol version 4 (TCP/IP)" and click "properties", open
the dialog box as shown, select "use the following IP address", as shown in the figure
below set the IP address and subnet mask:

Click “OK” to exit.

5.4 Calibration of Generator

5.4.1 Software Interface

The name of each element on the self-test interface and its role:
1.Self-test process information display area
(A) High voltage generator works normally and is turned on for the first time:
it shows the process of self-test of the system.
(B) High voltage generator is working normally, not the first time it is turned
on: it shows the current system status.
60
User’s Manual V2.11.1E

(C) High voltage not connected: only "CAN open successfully" is displayed.
(D) Open the software repeatedly: display the following information.

After pressing the "OK" button, the software enters the self-test interface, and the
prompt appears again, indicating that the last software is running. At this time, please
close all running HVSDK.exe, and then open the software to run normally.
Note:
(A) If the software does not open repeatedly, when the following prompt appears
when opening the self-test interface, please press the software element 2 buttons
(close button), then unplug the CAN controller communication line and plug it in
again to open the software. Normally enter the self-test process.

(B) If the software is opened, the following information is displayed on the


self-test interface, and there is no change, indicating that the CAN controller is
connected normally, but the high voltage communication fails, which may be caused
by high voltage failure and high voltage abnormality.

2. Close software button


Press this button to exit the software while the software is running.
Note:
(1) Turn on the high voltage. If the high voltage is reported incorrectly, the

61
User’s Manual V2.11.1E

program will pop up the error code prompt box.


( 2 ) The software will automatically enter the SDK interface only after the
high-voltage self-test ends normally.

5.4.2 SDK Service Mode Interface

Manual X-ray tube calibration interface


The left side of the service mode is the menu bar and the right side is the
sub-interface. The service mode defaults to manual X-ray tube calibration.

Description of each element on the interface


(1) Preparation and exposure lamp:

The status of this light during standby:


Press the first brake (prepared), the state of the light when the generator is

ready:

The status of this light when the second brake (exposure) is pressed:

The state of the light after the exposure is over:


(2) Manual X-ray tube calibration switch button: there are two states, default
and pressed. When this button is pressed, the sub-interface is switched to the manual
X-ray tube calibration interface.

62
User’s Manual V2.11.1E

(3) Automatic X-ray tube calibration switch button: there are two states,
default and pressed. When this button is pressed, the sub-interface is switched to the
automatic X-ray tube calibration interface.
(4) Configuration button: there are two states, default and pressed. When this
button is pressed, the sub-interface is switched to the configuration interface.
(5) System date and time display
(6) Display area of X-ray tube heat capacity percentage: after one or more
exposures, the current X-ray tube heat capacity percentage will be displayed.
(7) Filament adjustment (subtraction) button: click this button, the filament
value is reduced by 1. The filament value will be further decreased when long press
this button. There are three states: default and mouse stay and being pressed.
(8) Display area of filament value: display current filament parameter value
with the decimal point at 0.When adjusting the KV or MA value, the filament
parameter value will change accordingly.
(9) Filament adjustment (add) button: click this button, the filament value is
added by 1. The filament value will be further added when long press this button.
There are three states: default and mouse stay and being pressed.
(10) KV adjustment (subtraction) button: click this button, the KV value is
reduced by 1. The KV value will be further decreased when long press this button.
There are three states: default and mouse stay and being pressed.
(11) Display area of KV value: display current KV parameter value with the
decimal point at 0, and the range is [40,150].
(12) KV adjustment (add) button: click this button, the KV value is added by 1.
The KV value will be further added when long press this button. There are three states:
default and mouse stay and being pressed.
(13) Display area of KV average value: display the average KV sampling data
with the decimal point at 2.
Note: "Error" will be displayed here when the simulated sampling Error value

exceeds 40%.
(14) Display area of KV error percentage value: After receiving exposure
once, the system will sample KV data, calculate the average value of the data and
analyze the error.
Display value algorithm for this region:
63
User’s Manual V2.11.1E

((|KV average display value - set value |)/ set value) *100% with 1 decimal point.
Note: "Error" will be displayed here when the simulated sampling Error value

exceeds 40%.
(15) MS adjustment (subtraction) button: click this button, the MS value is
reduced by 1. The MS value will be further decreased when long press this button.
There are three states: default and mouse stay and being pressed.
(16) Display area of MS value: display current MS parameter value with the
decimal point at 2.
(17) MS adjustment (add) button: click this button, the MS value is added by 1.
The MS value will be further added when long press this button. There are three states:
default and mouse stay and being pressed.
(18) MA adjustment (subtraction) button: click this button, the MA value is
reduced by 1. The MA value will be further decreased when long press this button.
There are three states: default and mouse stay and being pressed.
(19) Display area of MA value: display current MA parameter value with the
decimal point at 1.
(20) MA adjustment (add) button: click this button, the MA value is added by
1. The MA value will be further added when long press this button. There are three
states: default and mouse stay and being pressed.
(21) Display area of MA average value: display the average MA sampling data
with the decimal point at 2.
Note: "Error" will be displayed here when the simulated sampling Error value

exceeds 40%.
(22) Display area of MA error percentage value: After receiving exposure
once, the system will sample MA data, calculate the average value of the data and
analyze the error.
Display value algorithm for this region:
((|MA average display value - set value |)/ set value) *100% with 1 decimal
point.
Note: "Error" will be displayed here when the simulated sampling Error value

exceeds 40%.
(23) Display area of KV analog waveform: after a manual exposure, this area
64
User’s Manual V2.11.1E

will show the relation between sampling KV and exposure time. The ordinate is
sampling KV, and the abscissa is exposure time. Click the waveform and pull it to the
right. The waveform becomes larger and the mouse is restored to the left waveform.
(24) Display area of MA analog waveform: after a manual exposure, this area
will show the relation between sampling MA and exposure time. The ordinate is
sampling MA, and the abscissa is exposure time. Click the waveform and pull it to the
right. The waveform becomes larger and the mouse is restored to the left waveform.
(25) System log button: when this button is pressed, the sub-interface is
switched to the system log interface.
(26) Display area of system communication information:
Display when the system is working normally:

"High pressure disconnected!" is displayed when high-pressure is closed”


"Communication line is disconnected!" is displayed when the communication
line of CAN controller is disconnected.
(27) Minimize software button: when press this button, the software will be
minimized to the operating system taskbar and the pallet bar in the lower right corner
of the system. Click the taskbar to restore the software.
(28) Exit the program button: when press this button and the program will exit
completely.
(29) Error reset button: only reset errors at the beginning of EL
(30) Return to the SDK button: when this button is pressed, the service mode
is exited and the software enters the SDK interface.

65
User’s Manual V2.11.1E

Manual X-ray tube calibration interface

(31) KV gear currently being calibrated: represents the KV gear currently


being calibrated, and this area cannot be manually adjusted
(32) Display area of current filament current value: this area cannot be
manually adjusted to display filament current value
(33) Display area of current exposure time value: this area cannot be manually
adjusted to display exposure time
(34) Display area of current X-ray tube filament current value: this area
cannot be manually adjusted to display filament current value
The algorithm is as follows: filament analog value = filament value
*2/4096.0*5.0
(35/36) Small / Large focus selection box.
When and only one of them can be selected, the selection of small focus
indicates that the MA of the calibration is 100MA or below, and the selection of large
focus indicates that the MA of the calibration is 125MA or above.
(37) Prompt for the success of the automatic calibration of the
corresponding shift position: the starting KV of the automatic calibration is fixed at
50KV. After setting the focus of the size, press the first gear of the hand brake, and
66
User’s Manual V2.11.1E

then press the second gear of the hand brake and do not release it. At this time, enter
the automatic calibration process. Starting small focus, for example, high voltage
generator will continuous exposure, since 50 kV and adjusting filament value, so that
the filament current direct or close to 10 MA, the elements of 40 corresponding gear
(50 kV, 10 MA) region shows "tick", said the end of the calibration, after the system
to continue a (50 kV, 12.5 MA) filament calibration, until the 120 kV, 100 MA
calibration is completed. If the calibration is completed, the system will pop up the
completion prompt dialog box and press ok. If the hand brake is released during the
process, the software will pop up the inquiry dialog box and ask whether to continue.
If you press ok, you will stay in the current calibration status. If you press cancel, you
will exit the automatic calibration interface and return to the manual calibration
interface.
(38) Current selected tube model
Configuration interface

67
User’s Manual V2.11.1E

Description of each element on the interface


(39) X-ray tube selection box: click the arrow to the right of this box and the
X-ray tube selection table will pop out (element 48).
(40) Select the drop-down box for ionization chamber type: click this
drop-down box, the following two types of ionization chamber types will pop up:
AID/ Claymound, Simense, and the selection box after clicking a certain type shows
the currently selected ionization chamber type.

(41) Filter selection drop-down box: click this drop-down box and the filter
selection box will pop up. Click a certain type.

(42) AEC curve select drop-down box: click this drop-down box, will produce
11 AEC curves, and then click a certain type.

(43)Select system language, currently English by default

68
User’s Manual V2.11.1E

(44) Select Drop Down List Box of Treadmill mode:Press Select Drop Down
List Box,there will appear 7 kinds of Treadmill modes as followed:

(45) X-ray tube selection button


Note: The parameters list of each type of X-ray tube is as follows:
Minimum Maximum Maximum Maximum Maximum
Number X-ray tube
(KV) (KV) (mA) (mAs) power(KW)
1 HD1020 40 125 320 320 20
2 HD2040 40 125 400 400 32
3 HD3050 40 125 640 640 50
4 XD3050 40 150 640 640 50
5 E7242 40 125 640 640 47
6 e7252-h 50 150 1000 1000 77
7 e7252-l 50 150 640 640 47
8 ray12 40 150 640 640 50
9 ray14_h 40 150 1000 1000 77
10 ray14_l 40 150 640 640 50
11 raD14 40 150 1000 1000 77
12 raD12 40 150 640 640 50
13 A192 40 150 1000 1000 80
14 RAD-8 40 150 640 640 47
15 A132 40 150 800 800 63
16 E7239 40 125 640 640 47
17 E7240 40 150 320 320 20
18 E7254 40 150 1000 1000 80
19 E7255 40 150 1000 1000 80
69
User’s Manual V2.11.1E

20 e7299 40 150 500 500 40


21 E7813 40 150 1000 1000 80
22 E7823 40 150 1000 1000 80
23 E7846 50 130 400 400 32
24 LQ16-2040/150 40 125 500 500 40
25 LQ16-2040/150 40 150 500 500 40
26 H1074 40 125 500 500 40
27 H1075 40 150 500 500 40
28 H1076 40 125 640 640 50
29 H1077 40 150 400 400 32
30 H1080 40 150 640 640 50
31 H1083 40 125 400 400 32
32 H1086 40 150 640 640 50
33 XD57-2050/150 40 150 640 640 50
34 XD52-3050/125 40 125 640 640 50
35 XD52-3050/150 40 150 640 640 50
(46)X-ray tube model
System log interface

Description of each element on the interface


(47) System error log table: this table displays all error information stored in
the error file (errlog.log) of the system, including error time, information, KV, MA,
70
User’s Manual V2.11.1E

MS, exposure mode and focus.


(48) Refresh button: press this button and the software will read the current
latest error information.
(49) Total times of high pressure exposure: show total times of high pressure
exposure.
(50) Refresh button: press this button and the software will read the current
total number of exposures.
(51) Return button: press this button to push the system log interface back to
the service mode interface.

5.4.3 Final Inspection

High voltage generator adjustment of the equipment must be approved by the


factory or factory technician, adjustment method and related technical questions,
please contact the supplier.

5.5 X-ray and Image Calibration

5.5.1 X - ray Beam Collimation Adjustment

Adjust the table in a horizontal position, will collimator opening to maximum,


the collimating tube placed in the center of the image intensifier is.Open the
acquisition workstation software and manually set the photography conditions:
50kV,5mAs, observed by the method of perspective collimating tube round imaging
the edge of the monitor, if the edge size uniform (projection edge thickness is the
difference between the maximum and minimum 1 mm or less) suggests that X-ray
focus is perpendicular to the wire grid. If uneven, adjust the tube to the thicker side of
the edge until even.

5.5.2 Adjustment of Dosimetry Indication Values

Accuracy of dosimetry indication values:


1. When the specific kinetic energy rate of air and cumulative specific kinetic
energy of air exceed 6 mGy/min and 100 mGy, the deviation from their displayed
values should not be more than ± 35% ;
2. The deviation from the cumulative display value of the dose area above
5uGy•m2 should not be greater than ±35 % ;

71
User’s Manual V2.11.1E

The display of the dose indication value is obtained by calculating the loading
factors of the high voltage generator, so the output accuracy of the high voltage
generator must be controlled to ensure the accuracy of the dose indication value.

5.5.3 Collimator X - ray Field - light Field Coincidence Calibration

Placed in the center of the image intensifier 100 x100 square iron box (coarse 5
x5), adjusting the limit of the beam splitter to X and Y to speak, in the light of wild
slightly larger than the iron box (5 mm or so), by looking through the method of the
limit of optical field box beam and the place of iron frame around the gap is consistent,
otherwise adjust the limit of internal view the location of the light beam splitter, lead
to the agreement clearance (error is about 2 mm).

5.5.4 Collimator Adjustment

1-Real-time opening size indication 2- Horizontal (X direction) lead vane button


3-Light delay button 4-Longitudinal (Y - direction) lead leaf button
5-SID indicator lightL1、L2
Figure 5-5
1) Function description:
a. The valid button for Collimator debugging is as shown in the operation panel
diagram of the Collimator. After the Collimator is powered on, press and hold it

, 3 Seconds later, four "8s" appear in the window indicating the real-time
opening size and enter the debugging state (the power on the new Collimator without

72
User’s Manual V2.11.1E

debugging parameters will automatically enter the debugging state). At this time, L1
and L2 in the real-time opening size indication will be off, and four "0's" will be
displayed on the digital tube.

b. At this point, click or , you can select the function, and the first

group of digital tube display, click or , can be adjusted parameters, and


the second set of digital tube display.
The digital tube display functions are as follows: the first group of digital tube
display "C1" is the setting function of horizontal motor AD, the second group is the
real-time AD value, and the "longitudinal lead adjustment button" can control the
motor action; The first group of digital tube display "C2" for longitudinal motor AD
setting function, the second group for real-time AD value, "longitudinal lead
adjustment button" can control the motor action; The first group of digital tube
display "C3" for the lamp delay setting, the second group for the time setting display
5 to 45 seconds, 5 seconds per step, "longitudinal lead adjustment button" to select the
corresponding time; The first set of digital tube display "C4" for the Collimator
opening size debugging, debugging the next level menu. Click on the

or , c1-c5 can be selected in a cycle. When the corresponding value is

selected, press Set values can be saved.


c. Adjust the opening size of the Collimator. Debug it in the next level menu

under "C4".Click , Display 4 "d", that is, the opening size of the Collimator has
been adjusted. The opening size of Collimator can be adjusted by "X opening" and "Y
opening". "X opening" corresponds to A value, and "Y opening" corresponds to B

value. Click or A1-ad and b1-b9 can be selected circularly and

displayed by the first group of nixie tubes. Click or , the opening size
can be adjusted, and the second set of digital tube display its opening real-time AD

value. When the corresponding value is selected, press Set values can be
saved. The comparison table of relevant parameters is as follows:

73
User’s Manual V2.11.1E

Table 5-9
X to open Y to open
A1=24mm b1=30mm
b2=54mm
A2=28mm
(perspective)
A3=30mm b3=48mm
A4=33mm b4=60mm
A5=36mm b5=66mm
A6=42mm b6=72mm
A7=48mm b7=84mm
A8=60mm b8=101mm
A9=66mm b9=Max
AA=72mm
Ab=84mm
AC=54mm
(perspective)
Ad=Max
2) Setting description:
In the normal working state, the delay button of the point light turns on the visual
field light, and the control button of the point related lead blade reduces the lead blade
in two directions to zero and enters the debugging state. The real-time AD value is
saved under the function of "C1", the real-time AD value is saved under the function
of "C2", and the corresponding delay time is saved under the function of "C3"."C4"
function is set in the next menu level, as function description 3.Among them, "C4" is
also divided into whole piece setting, vertical setting and horizontal setting.C5

"function is X, Y axis automatically open and close test, press Entering the
test, facing the Collimator operation panel, the 2-bit nixie tube on the left shows the
value of AD on the X-axis, and the 2-bit nixie tube on the right shows the value of AD

on the Y-axis (the remaining 4 keys are useless in the test);then press Close the
test and display "C500".
Full film setting: turn on the Collimator light, turn the vertical and horizontal

74
User’s Manual V2.11.1E

lights to the maximum limit, then shrink back a little and press down Save the
Settings.
Setting of vertical division, vertical dichotomy only requires the cooperation of
large light bar board, turn on the Collimator lamp, reduce the longitudinal direction to
the point where the light falls on the lead sheet of large light bar, and press down

Save the Settings.


Setting of horizontal division: the horizontal division only needs the small light
bar to cooperate with each other. First, move the small light bar to the required
opening size, turn on the Collimator lamp, and reduce the horizontal division to the

light shining on the lead sheet of the small light bar, and then press down
Save the Settings.
3) exit setting description:
To exit the debugging state, you must first ensure that the function display is L1
and L2 off, digital tube display 4 "0", function display key can make the function

display back to this state, press and hold 3 seconds after the display of 4 "8"

to quit. One from the "C4" next level menu back to "C4", click or

make digital tube display 4 "d", click Can return.


Description of real-time opening size: SID110 light indicates that the value of
digital tube is SID=1100 light field size; SID150 indicates that the value of the nixie
tube is SID=1500 light field size.
4) PCB jumping function
JP2 and JP3 in PCB board (PCB10.31) do not need short connection

75
User’s Manual V2.11.1E

5.5.4 Flat Panel Detector Adjustment

(1) Close the DFOC, open D:\XRayDR\Xray\XRayRF\DFOC\,


find and open Calibration Tools.exe,as show below:

(2) In the "DRMode" menu, select the DRS3072 mode (the green is selected), select
the maximum frame rate 5 on the right, and click the "Calibration" button. At this
time, the "Offset" button will be switched from unselected to selected, as shown
in the picture below:

76
User’s Manual V2.11.1E

Note; Calibration is divided into point-slice, continuous perspective and


pulse perspective according to the use mode. The following is the point-slice
calibration process. Because DR3072 and DRS3072 use the same calibration data,
DR3072 and DR1536 do not need to be calibrated, only for DRS3072 and
DRS1536.
(3) Select the "Offset" button in the figure above to start the collection of dark field
image of the flat panel. The gray value of the currently obtained image will be
displayed in "Current Value", and the "Process" on the right will indicate the
current obtaining status, as shown in the figure on the left below. If the current
acquisition status is displayed as "32/32", it means that 32 dark field diagrams
need to be collected in the current mode, and the collection has been completed.
In this case, "Gain" button is optional, and "Offset calibration completes
successfully" is displayed in the following information box. As shown on the
right below:

(4) Click the "Gain" button and adjust the fluoroscopic dose for fluoroscopic
(Suggested KV parameter 50KV). The gray Value of the bright field map after the
Current loaded offset calibration will be displayed in "Current Value". Compare
this gray Value with "Expected Value 2000" to make the "Current Value" close to
the "Expected Value" with a deviation of no more than 5%. When the two values
meet the above requirements, keep the foot brake on and click "Select"
button. The gain calibration starts. The value next to "Process" displays the
current calibration status. When "Process" displays "19/19", the current
calibration step has been completed.

77
User’s Manual V2.11.1E

Note: The Select button can only be selected if the handbrake is pressed all
the time.
(5) When the dialog box "Stop Exposure and Start to acquire dark Image" pops up on
the interface, release the foot brake to collect dark field images.

After dark field map is collected, the state is changed from


non-optional to optional. Click "Accept" button to generate the first-step correction
file.

78
User’s Manual V2.11.1E

(6) Then enter the second step of calibration, and the stage state becomes "2/3". Click
the "Gain" button;
(7) Adjust the fluoroscopy dose to perform fluoroscopy (e.g., Suggested fluoroscopy
KV parameter 110KV). The gray Value of the bright field map after the Current
loaded offset calibration will be displayed in "Current Value". Compare this gray
Value with "Expected Value 12000". Keep the "Current Value" close to the
"Expected Value" and the deviation within 5%. When the two values meet the
above requirements, keep the foot brake on and click the "Select" button to start
gain calibration. The value after "Process" will display the current calibration
status. When "Process" displays "19/19", the current calibration step has been
completed.

(8) When the dialog box "Stop Exposure and Start to acquire dark Image" pops up on
the interface, release the foot brake to collect dark field images.

After dark field map is collected, the state is changed from non-optional to
optional. Click "Accept" button to generate the second step correction file.

79
User’s Manual V2.11.1E

(9) Then enter the third step of calibration, and the stage state becomes "3/3". Click
the "Gain" button;
(10)Adjust the fluoroscopy dose for fluoroscopy (e.g. Suggested KV, 70KV), and the
gray Value of the bright field map after the Current loaded offset calibration will
be displayed in "Current Value". Compare this gray Value with "Expected Value
2000". Keep the "Current Value" close to the "Expected Value" and the deviation
within 5%. When the two values meet the above requirements, keep the foot
brake on and click the "Select" button to start gain calibration. The value after
"Process" will display the current calibration status. When "Process" displays
"19/19", the current calibration step has been completed.

(11) When the dialog box "Stop Exposure and Start to acquire dark Image" pops up on
the interface, release the foot brake to collect dark field images.

After dark field map is collected, the state is changed from non-optional to
optional. Click "Accept" button to generate the third step correction file.

80
User’s Manual V2.11.1E

Continuous perspective mode calibration


Note: The modes are CF1536, CFZoom, and CF1024 respectively.
1) Select "CF1536" mode in "RFMode", and the current maximum frame rate will
be displayed by default. Confirm that the left perspective mode is continuous
perspective mode.
2) Click the "Calibration" button in the upper right corner to enter the Calibration
state. At this time, the Offset of the current mode is optional. After selection, the
dark field map will be collected, and the background gray Value of the current
dark field will be displayed after "Current Value", and the number of dark field
images currently obtained will be displayed in "Process" progress.

3) Click the "Gain" button and adjust the fluoroscopic dose for fluoroscopic
(Suggested KV parameter 50KV). The gray Value of the bright field map after the
Current loaded offset calibration will be displayed in "Current Value". Compare
this gray Value with "Expected Value 2000" to make the "Current Value" close to
the "Expected Value" with a deviation of no more than 5%. When the two values
meet the above requirements, keep the foot brake on and click "Select"
button. The gain calibration starts. The value next to "Process" displays the
current calibration status. When "Process" displays "19/19", the current
calibration step has been completed.

81
User’s Manual V2.11.1E

4) When the dialog box "Stop Exposure and Start to acquire Dark Image" pops up
on the interface, release the foot brake to collect dark field image, as shown in the
figure.

After dark field map is collected, the state is changed from


non-optional to optional. Click "Accept" button to generate the first-step
correction file.

5) Then enter the second step of calibration, and the stage state becomes "2/3". Click
the "Gain" button;
6) Adjust the fluoroscopy dose to perform fluoroscopy (e.g., Suggested fluoroscopy
KV parameter 110KV). The gray Value of the bright field map after the Current
loaded offset calibration will be displayed in "Current Value". Compare this gray
Value with "Expected Value 12000". Keep the "Current Value" close to the
"Expected Value" and the deviation within 5%. When the two values meet the
above requirements, keep the foot brake on and click the "Select" button to start
gain calibration. The value after "Process" will display the current calibration
status . When "Process" displays "19/19", the current calibration step has been
completed.

82
User’s Manual V2.11.1E

7) When the dialog box "Stop Exposure and Start to acquire Dark Image" pops up
on the interface, release the foot brake to collect dark field image, as shown in the
figure.

After dark field map is collected, the state is changed from non-optional
to optional. Click "Accept" button to generate the first-step correction file.

8) Then enter the third step of calibration, and the stage state becomes "3/3". Click
the "Gain" button;
9) Then adjust the fluoroscopy dose for fluoroscopy (e.g. Suggested KV, 70KV),
and the gray Value of the bright field map after the Current loaded offset
calibration will be displayed in "Current Value". Compare this gray Value with
"Expected Value 2000". Keep the "Current Value" close to the "Expected Value"
and the deviation within 5%. When the two values meet the above requirements,
keep the foot brake on and click the "Select" button to start gain calibration. The
value after "Process" will display the current calibration status . When "Process"
displays "19/19", the current calibration step has been completed.

83
User’s Manual V2.11.1E

10) When the dialog box "Stop Exposure and Start to acquire Dark Image" pops up
on the interface, release the foot brake to collect dark field image, as shown in the
figure.
11) After the dark field map is collected, the non-optional state is changed to optional
state. Click "Accept" button to generate the third step correction file, and then the
perspective mode calibration is officially completed.

Note:Pulse perspective calibration data adopts continuous perspective


calibration data, so after continuous perspective calibration is completed, pulse
perspective does not need to be calibrated.

5.5.5 Final Confirmation Inspection

The high-voltage generator of the equipment must be debugged by the technical


personnel approved by the factory or the factory. Please contact the technical
personnel approved by the factory or the factory for debugging method and training
management and related technical problems.

84
User’s Manual V2.11.1E

CHAPTER6 OPERATION OF THE EQUIPMENT

6.1 Symbol Descriptionon Equipment

The on/off button of the device is on the control box,

1. Power on button.

2. Power off button.

Power on:1.Make sure the emergency stop switch unscrew,press the start

button on the generator control box ,power on the system.


2.Press the PC button,after the PC has finished booting, open the
image workstation.
Power off: Close the image acquisition workstation computer manually, close the

PC , and press button on the generator control box , the power


can be turned off.

6.1.1 Table Exercises for Interface and Symbols

Figure 6-1 DR box remote control panel


85
User’s Manual V2.11.1E

Figure 6-2 local control panel


6.1.1.1 Description of Table Control Symbol
Emergency stop button:
Emergency stop diagnostic table
Tube-column move up button

Tube-column move down button

Tabletop lateral movement button


Using this button control the tabletop
lateral movement

Tilt button

Forward button of compressor:


Using this button control the compressor
forward move close to the patient's body
Backward button of compressor:
Using this button control the compressor
backward move far away from the patient's
body
Protection button of compressor:
When this button be pressed, under the
compressor pressure while banning Tube
stand column longitudinal movement,
Tabletop transversal movement and Foot
stand rotate.
Pedal rotation button

86
User’s Manual V2.11.1E

Vertical open /close button of the


collimator

Horizontal open/close button of the


collimator
Auto mode button of the collimator:
Using this button collimator can be
switch to automatic tracking state, in spot
radiography while collimator automatically
adjust to the size according to the film size
and subdivision format
Negative Angle start button:
Press this button and the Table tilt down
button at the same time, the Table body is tilt
down from horizontal position to the negative
angle direction
SID adjustment button:
Press to set the diagnostic table SID to
1000mm
Automatic reduction to specified distance
SID adjustment button:
Press to set the diagnostic table SID to
1800mm
Automatic reduction to specified distance
Exposure preparation buttons and instructions

Exposure buttons and instructions

Fluoroscopy instructions

Power on button

Power off button

87
User’s Manual V2.11.1E

6.1.1.2 Operation of Table Control

1) Table tilt: Operating on the remote control operating handle or near the

tilt buttons on control board, can control the table tilt up down motion,

loosen the button to immediately stop or limit to immediately stop

automatically, is the direction to + 90 °, arrived in the opposite direction -

25 °, you have automatic stop function level 0 °, to continue to exercise

close control, operation can once again, the console operation, At the same

time, press this button and the inverted table handle button to rotate the

table from the horizontal to the negative angle direction. When the table

moves up, the tube should be full of a certain height from the ground. If not,

the tube column will be automatically adjusted to avoid falling and touching

the ground

Do not place foreign matter in the rotation range of the


table , otherwise it will cause damage to the table.

2) Vertical movement of x-ray tube column: the vertical movement of x-ray

tube column can be controlled by operating handle on the operating table or

by pressing the vertical movement of X-ray tube column on the local

control panel. Release the handle or release the button and stop immediately.

The upward movement distance is 1000mm and the downward movement is

0mm, that is, the total distance is 1000mm.

3) Press cup advance and retreat:Press the press into button on the remote

control table to control the downward action of the pressure cup, press the

press back button on the remote control table to control the upward action of

the pressure cup. In the compressor tracking key indicator light and

compression state, the compressor in the direction of non-compression force

and bed panel has a relative displacement will make the relief of

compression force failure, and accompanied by an alarm sound, to remind


88
User’s Manual V2.11.1E

the operator to pay attention.

4) Tabletop movement:With the control table lateral tabletop lateral

movement handle or press the near control tabletop lateral movement key

can control the tabletop lateral movement, reflected as the perspective image

on the TV to move left and right, release the handle or release the key

immediately stop, the total movement distance is 220mm.

5) SID:SID adjustment the SID changes from 1000mm to 1800mm

6) Pedal rotation:The human body can be rotated in either direction by

rotating the handle.

6.1.2 Generator Operation Instructions

6.1.2.1 Generator iIndicator Light and Operation


High voltage generator control box and hand brake switch are shown below.

1. Perspective indicator light.

2. Exposure indicator light.

3. Or the first gear of hand brake: prepare button.

4. Or the second gear of hand brake: exposure button.


89
User’s Manual V2.11.1E

Note: in case of emergency, please force the emergency stop


switch (usually red mushroom button) to cut off the
connection between the high voltage generator and the main
power supply.

Preparation, X-ray exposure, and exposure indicators


Press the preparation button of the generator control box or the first gear of

hand brake (preparation), ready indicator in the upper right corner of the ready
button lights up.
By pressing the exposure button or the second gear of hand brake (exposure)

controlled by the generator, the exposure indicator lights up and a buzzer goes
out through the process through which the X-ray is emitted.

You can use the first gear of hand brake or (ready) and the second gear

of hand brake or (X-ray exposure) photographic exposure with two hand


brakes.
Use the dot film controller on the corresponding dot film device for dot film
exposure.
During the exposure process, the preparation and exposure indicator light
indicate the exposure status.
X-ray hand brake button has three positions: hand brake release, the first gear of
hand brake and the second gear of hand brake.

90
User’s Manual V2.11.1E

Press the first gear of hand brake or (ready), when ready, press the

second gear of hand brake or (exposure).

Ready: press the first gear of hand brake or , prepare to select the X-ray
tube for exposure. When the X-ray tube is ready and there is no interlock failure or
system failure, the preparation indicator light on the generator control box will be lit.

After pressing this button, the following functions will be


activated:

 The anode rotates.


 Filament current switches from standby to selected mA.

When the foot brake switch is pressed for X-ray fluoroscopy, The
perspective indicator lights up.

6.1.2.2 High Voltage Generator Parameter Adjustment


Parameter adjustment of high voltage generator is described in 5.4, see 5.4.1 for
details.

6.1.3 Operation Instruction of Collimator

1. Basic structure
The shape of the collimator is shown in the figure below. The main internal parts
are transmission device, detection device, transverse variable lead blade, longitudinal
variable lead blade and limit mechanism. The outside is mainly composed of lead
blade opening and closing adjusting buttons, collimator lighting buttons and interfaces
connected with X-ray tube components.

91
User’s Manual V2.11.1E

Shape diagram of Collimator


1- Light button 2-Real-time opening size indication 3-Set screw
4-Additional filter mounting port 5-Horizontal lead adjustment button
6-Longitudinal lead adjustment button
Figure 6-3

2. Instructions
a. The size of the radiation field of view is shown by the size of the irradiated
area of the collimator. The size of the field of view is adjusted by clicking the
corresponding button of the lead plate opening and closing indicator icon on the panel.
The operating table also has corresponding buttons to control the size of the visual
field.
b. In the fluoroscopy state, when it is switched to the manual state, the vertical
and horizontal lead blades can be remotely adjusted to change the visual field size. In
the automatic state, the Pedals only tracks to the preset value when it is pressed. The
same is true in radiography.

6.1.4 Intercom System Operation Instructions

The user can speak to the patient by stepping on the left-foot switch.

92
User’s Manual V2.11.1E

6.1.5 Image Acquisition Workstation Software Settings

6.1.5.1 Login

After the operating system starts, the login window displays. Enter the correct
user name and password on the system login interface to enter:
In the user login window, enter the correct user number and password and press

" " to enter the loading interface. If you do not want to enter the system, click

" " to exit.

6.1.5.2 Exit

Click , the exit the confirmation box pops up:

93
User’s Manual V2.11.1E

Enter the login password ,Click ,After the system saves the image, exit

the console software.Click ,cancel the exit operation and return to the software
operation interface.

6.1.5.3 Study Management

1.Work list

Work list

94
User’s Manual V2.11.1E

Search Items:

1. Patient ID: user need to input the patient’s id in the textbox, and define the correct
date time.
2. Accession: user need to input the patient’s accession in the textbox, and define
the correct date time.
3. Name: user need to input the patient’s name in the textbox, and define the correct
date time.
4. Today: query the patients’ data in today.
5. Two days: query the patients’ data in two days.
6. Three days: query the patients’ data in three days.
7. One week: query the patients’ data in a week.
8. Custom: query the patients’ data in user defined period. Once the condition set
down, the work list will query data frequently, it can also refresh immediately
after the button “Refresh” clicked.

Register new Patient:


Input the patient data and register a new patient for study.

Delete:
Delete current selected patient.

Begin to study:
Begin to acquiring image of current patient.

Emergency:
Register a patient in emergency with auto generated information.

95
User’s Manual V2.11.1E

2.Registration

Registration
The register is used to record information of a new patient. User can select study
protocols through protocol or Item mode.
Accession*: input the accession number of patient. It will be automatically
generated and usually suggest user don’t change it. It must unique in the system.
PatientID*: input the ID of a patient. It will be automatically generated and
usually suggest

user don’t change it. Input PatientID and click , system will automatically
fill the patient’s information in the text box.

IDCardNO*: input the CardNO of a patient. Input CardNO or click ,


system will automatically fill the patient’s information in the text box, the premise is
that the computer is connected to the ID card recognizer.

Name: input the name of a patient.


Gender: select gender of a patient. (Male, Female, Other (unknown)) Age: input
age of a patient at the age unit of year, month or day.
Height(cm): input height of a patient. Weight(kg): input weight of a patient.
Birthday: select birthday of a patient. Description: input more comments of a patient.
[Note]The items with “*” means this item must be filled. If not, there will
be a warning message:

96
User’s Manual V2.11.1E

Study Bodypart:Select the oval gray button on the body as needed, and click the
button, andthe gray turns green to indicate that the site is selected. The list on the right
has a corresponding check position.
Study Protocols and Items:

Click button => Study Protocols


interface

Double-click or click the button after


checking the name of the agreement you want in the
protocol bar,
The Protocol appears in the selected Protocol
Tube stand.

Double-click or click the button after


selecting an agreement name in the selected
agreement bar and the agreement name is removed
from the selected Tube stand.

97
User’s Manual V2.11.1E

Click button => Study Protocols interface

Click button
Switching Rad/
Fluoroscopy.

Double-click or click the button


after checking the desired item name in the item bar
and the item will appear in the selected item bar.

Double-click or click the button after


selecting an item name in the selected item bar, and
the item name is removed from the selected Tube
stand.

Study list

Study list
It is used to query the information of the patients that finished the study. Same
with the work list, the result can be displayed under different condition.

Click the button to enter the Studylist interface, where you


98
User’s Manual V2.11.1E

can query the specified patient information and related photographic image records.

a) List Information:

Click button , The list includes patient information that has


been examined: patient number, name, gender, age, height, weight, date
of birth, check number, check site, description, number of images,
archive identification, printed identification, etc. These lists can be
configured for display and according to hospital requirements for easy
display.

b) Information search:
Study management provides the ability to search on a certain basis. Enter the

conditions thatneed to be retrieved, only click button .


The search conditions include the registration time range of the patient to be
examined, the patient number, examination number, name query. Registration time
99
User’s Manual V2.11.1E

ranges can be quickly selected via buttons, including "Today," "Two Days," "Three
Days," "One Week and Custom" ranges. After the retrieval criteria are selected, click
the "Confirm" button to check and place eligible patients for examination in the list.

[Note] If the patient number, the study number, the patient name is empty,
indicates any condition.
c) Delete records:

Select a checked patient in the list to be examined, click the button and

the software will pop up the delete confirmation box.

If you confirm that delete the patient record have selected, click OK, otherwise
Cancel.
d) Browse images:
Select a patient in the study list and double-click it to enter the image browsing
interface.
e) Re Examine:

Click button enter the image acquisition interface, you can add a

study protocol or item to restart the examine.


f) Image archiving and burning:

Burning: Select one or more records and click the button to

start the disc burningoperation, see chapter 8.2.

Archiving: Upload images to the network. Click the button to

100
User’s Manual V2.11.1E

upload the selectedimage to the set storage address.

Print: Print images with connected print devices, Click the button . See

chapter 8.3

Selection All: All patient information in the list is selected, and all

the information in the list is marked gray.

Lock: Select one or more patient records and click the button to pop

up the check box, This record is locked after click OK , Lock bar marked as “1”

, If click Cancel. Record will be unlocked. Delete operation cannot be


performed after patient record is locked.

Storage queues: When the system is not connected to the PACS server,

click the button to add the selected item to the upload queue and wait for the
communication to connect to upload. The Storage Queue check box can be refreshed,
retried, removed, and canceled.

Print Queue: when the system is not connected to the PACS server, click the
button to add the selected items to the network print queue and print after the
101
User’s Manual V2.11.1E

communication is connected. The Print Queue check box can be refreshed, retried,
removed, and canceled.

6.1.4 Image Acquisition


There are several paths can enter the acquisition interface, as follows:
Select a patient record in the work list double click the record or click the button

Click the emergency button in the work list or register interface.

Double-click on a record in the worklist to enter the acquisition interface.


The acquisition interface includes the body size of patient, the exposure position,
the study item, the study protocol, generator exposure parameter setting and the
diagram of body position. The following are detailed introduction.

102
User’s Manual V2.11.1E

Note: Interface functions and operations vary with hardware and system
configurations.
1.Thumbnail navigation area
2.Exposure condition and position selection area.
3.System status.
4.Image preview.
5.Image processing tool.
6.Dynamic image playback control area, image save area.
7.Functional control area.
8.Other Information.
a)Thumbnail navigation area

103
User’s Manual V2.11.1E

 Thumbnails are arranged in a tree,separately


open for each individual,In each image tree, the
top appears the current position name.
 The exposure type of the current image (DR/RF)
is displayed in the upper right corner of each
thumbnail.
 Each thumbnail shows the total number of
current images in the lower left corner. The
number of DR images is shown as 1, or RF
images, as the actual total.

104
User’s Manual V2.11.1E

b)Exposure condition and position selection area

Left-click switch bodypart and images


right-click Add Items -> Study Protocols interface
Image Archive ->Send Image to Pacs.
Print->
Delete->Delete current image
StoreDir->Open store directory

There are the body position diagrams selected for patient information registration.
You can see the specific study body part and the standard position for intuitive
reference.
For emergency patients this region is blank. See Chapter 4.2 for steps to add a
body position.
Before Image acquisition, the parameters of generator should be property
configured, and the trigger mode of detector must be correct. For more information,
please read the manual of these devices.
Exposure signal lamp and Reset error:

This button will light up when there is a system error, and


there will show the error code following the button. Click this
button to reset the error and if it does not work, please contact
customer service engineer
This lamp will rotate during exposure.

Press first gear of the hand brake, then the signal lamp

will light up. Green light means being ready;


Continue to press the hand brake to second gear, the signal lamp

will light up. The yellow light will keep lighting up in the

exposure process.

105
User’s Manual V2.11.1E

There is default parameter for every body part. You can


adjust these parameters as needed. Click this button after you
change these parameter. If don’t click the save button, the adjusted
parameters is only used for the current exposure.
Patient body size:
Fat adults Medium adults

Thin adults Baby

Generator exposure parameter setting


Left-click to switch single-frame acquisition
and sequence acquisition, and you can select the
frame rate of the sequence acquisition in the
drop-down box on the right.
[Note]Different detector have different frame
rates and modes.

KVp:KV voltage mA: mA


mS: Duration
DEN(Reserve): Switch density in AEC mode
This region show the default parameters of
the current item. You can adjust these
parameter by click the arrow button.

Exposure position
Table Wall Reserve

106
User’s Manual V2.11.1E

Focus
Small focus Big focus

Exposure mode
mA/ms mode: KV mA and mS parameters can be
adjusted. mAs mode: Only KV and mAs can be adjusted.

AEC mode: on this mode, can adjust density and field.


Software support 3 fields mode

Generator fluoroscopy exposure parameter setting

Left-click to switch Continuous fluoroscopy


and Pulse fluoroscopy, and you can select the
frame rate of these mode in the drop-down box
on the right.
[Note]Different detector have different
frame rates and modes.

KVp: Fluoroscopy KV
voltage
mA:Fluoroscopy mA
[Note]This region show the default parameters of
the current item. You can adjust these parameter
by click the arrow button.

107
User’s Manual V2.11.1E

Other setting on fluoroscopy exposure


LIH: Only display fluoroscopy
images not save.
F-Store: Store all frame images.

Fluoroscopy cumulative time.

Click button can zeroing time.

In ABS mode, the fluoroscopy parameters


are automatically adjusted to the appropriate
value based on the grayscale value of the image.
For different generator, this feature will be
different, the software supports software ABS
and hardware ABS, More specific applications
can contact generator
manufacturers and software manufacturers.
c) System status
This area shows the status of each part of the software system.

The components of the system are working


properly.

Detector parts are not working properly


may be dropped or other circumstances, need to
check that the part is powered up or that the
correct model is selected. If it still can't be
resolved. need to contact the software
manufacturer or detector manufacturer.
Generator parts are not working properly,
may be dropped or other circumstances, need to
check that the part is powered up or that the
correct model is selected. If it still can't be
resolved. need to contact the software
manufacturer or generator manufacturer.

108
User’s Manual V2.11.1E

The sync box parts are not working


properly, May the equipment does not power
up, does not connect the wire, does not
configure serial port; If it still can't be resolved.
Need to contact the software manufacturer.
d)Image preview
After the exposure trigger, the image collected by the system will be displayed in
the image preview area as follows:

The same position can be photographed/perspective multiple times.


Click the left mouse button to move the image; click the right button to drag can
adjust the WW/WL.
[Note] The information display in the four corners of the image can be modified
in the configuration tool. See the configuration manual for details.
e)Dynamic image playback control area

Play speed adjustment you can drag the


slider to adjust.
Play progress control.

Cycle again.

The last frame.

Play.

The next frame.

109
User’s Manual V2.11.1E

Play forward.

Play back.
f)Functional control area
Hide exposure conditions and position selection area.

The report interface pops up. See Chapter 6 for details.

Save the current series of images.

The printing interface pops up. See Chapter 8.3 for details.

Click the button to pause the current check, and the page
returns to the work list interface.

Click the button to save the image and enter the studylist
interface.
g)Other Information

Description of system operating status


information.
Image storage hard drive remaining capacity.

System current time.


Ondispaly current patient information .On
worklist interface and

studylist interface, Click ico will


pop up modifying the patient information interface.

Click the button to save the image and enter


the studylist interface.

110
User’s Manual V2.11.1E

4.Image browse

Image preview
After acquired an image from detector, the processed image will be displayed.
There are many powerful tools for advanced user to take a better view of the image.
a) Tools
Tool button Description
Page Pre/Next.

Clicking this button will exit the multi-layout selection box.


Default 1x1 layout.

Delete selected tool.

Zoom In/Out:

Histogram WW/WL

111
User’s Manual V2.11.1E

Invert Color

Zoom out

Zoom in

Screen Size

Text Edit

Image advanced process

Rotate 90

112
User’s Manual V2.11.1E

Flip Vertical

Measure the angle:

Image hint

Point gray value of the image:

Reset: reset the image.

Magnifier:

Copy:
The function of image contrast has been optimized. Click the
“Copy” button in the preview interface of patient A, then switch to
the preview interface of patient B, select a blank box and click
the “Paste” button to add the copied image. Be sure to click
the “Save” button before exiting if you want to save the image.

113
User’s Manual V2.11.1E

CTR: cardio-thoracic ratio


Make the following marks on the
chest radiograph: Line 1: Positive
midline;
Line 1: Maximum transverse diameter of thorax;
Line 3+4: maximal transverse
diameter of the heart; CTR = (length
(Line 3+ Line 4))/length (Line 2)

Email:

Click the button , input the mail address,


subject, content in the appropriate edit box of the pop-up box and
select current image by clicking “Add” button if there were exposed
images. The DR Software would automatically call the Outlook to
send the mail after you click the “Send” button.

Save as:
Its role is similar to the “copy” / “paste” tool. All three of them
are used to compare an image with itself. The “copy” / “paste” tool
can be used among different patients, while the “Save as” tool
can only be work with the images of one patient (the “Save as” tool
has the same function with a
combination of the “copy” and “paste” tool)

114
User’s Manual V2.11.1E

Inverse Compare:
When you click on the button, the Software will show another
inverted image as guidance for doctor observing and diagnosis.
Notes: the “Image color inversing” tool can be only used in
the image viewing interface of the DR Software.

Window/Level

ROI Window/Level

Auto WW/WL

Full Size

Mark Left

Mark Anterior

Mark Right

Clip Tool

Rotate -90

Flip Horizontal

115
User’s Manual V2.11.1E

Distance Measure:

Image stitching.

Ellipse clip tool

ROI Magnifier:

Arrow

Paste:
The function of image contrast has been optimized. Click the
“Copy” button in the preview interface of patient A, then switch to
the preview interface of patient B, select a blank box and click the
“Paste” button to add the copied image.
Be sure to click the “Save” button before exiting if you want to
save the image.
Free Rotate:
Click the “Free Rotation” tool, then pop-up a box for rotation
angle. Enter any
angle value you want and click the “OK” button to rotating
image. Rotate to the left by default. Click the “Undo” button the
image will turn back.
Cobb angle:

1. The patient stands while a front view x-ray of the spine is


taken.

116
User’s Manual V2.11.1E

2. The doctor uses the x-ray to locate the apex vertebra,


which is at the deepest part of the scoliosis curve, as well as the
most-tilted vertebra above the apex and most-tilted vertebra
below the apex.

3. A perpendicular line extending from the most-tilted


vertebra above the apex is drawn. The same is then done for the
most-tilted vertebra below the apex.

4. Where the two lines extending from the most-tilted


vertebra above the apex and most-tilted vertebra below the apex
join together gives the Cobb angle

Compare:
When you click on the button, the Software will show a pop-up
box for image selection. Then you should click on the “Add” button
and select two or more images, finally click on the “Open” button to
enter the Image contrast interface. Note here that if the Software is in
the image Acquiring interface, please click on the “Save” button
before image selection.
Notes: the “Image contrast” tool can be only used in the image
viewing
interface of the DR Software.

117
User’s Manual V2.11.1E

Clip Options

Polygon measurement

118
User’s Manual V2.11.1E

Rectangular measurement
[Note] Hold the ctrl key and pull it into a square

Elliptical measurement
[Note] Hold the ctrl key to pull it into a perfect circle

End of the browse

Finish &Save:
End current study and save all images to local drive as DICOM files. If
successfully, then user can find the study in the studylist.
b) Image stitching

User can enter the stitching interface by clicking “Image Stitching”

in the image
preview interface. Before to enter the image stitching, please make sure the
current study has at least two saved images.

119
User’s Manual V2.11.1E

: Exit the stitching interface.


5.Image and Text Report

Click the【 】button to enter the report interface:


a) Report Edit
Report editing module includes image selection, patient information editing,
reporting content and inspection results editing, print template selection, etc. are
presented below.

【1】 Select image

Move the mouse to the image need to be printed, then there will display a white
120
User’s Manual V2.11.1E

check box at the lower right corner of the image. Check the box then the box to add a
black “√” and the there will appear a yellow number at the lower left corner. Click
again to cancel the selected image.
【2】 Diagnostic content and conclusion edit box
You can edit the diagnostic content and conclusion in the edit box, click the
content and description template on the right side if you need to add.
【3】 Select the template for printing
1Standard300.rvf : One image with the height of 300
2Standard300.rvf : Two images with the height of 300
4Standard300.rvf : Four images with the height of 300
6Standard300.rvf : Six images with the height of 300

【4】 Report content template selection

Button Function Describe


Select diagnostic template of the body
Add diagnostic describe part on the right hand of the drop-down
list, click the “+Desc.” button to add
diagnostic describe.
Select diagnostic template of the body
Add diagnostic part on the right hand of the drop-down
conclusion list, click the “+Diag.” button to add
diagnostic describe.
Add diagnostic and Click this button to add diagnostic
describe conclusion describe and conclusion.

Click this button there will pop up the


Edit diagnostic template box of Knowledge bose editor. You can
edit, add and
delete the diagnostic template

121
User’s Manual V2.11.1E

Button Function Describe


Add: Enter a new name in the edit box of
Add/Delete class Class Name, click the ” button to add the
new class.
Delete: select a class you want to delete
from the drop down list on the left. Click the
“ ” button to delete it. (The deleted class
will be moved to recycle bin).
Edit the template
Click the button to Edit the template
corresponding to
corresponding to current report.
current report
a) Enter a new name in the edit box
of the “Item name”.
Add new item b) Enter the describe information in
the edit box of “Desc.”
c) Enter the conclusion information
in the edit box of “Diag.”.
d) Click this button to add the new
item.
Update template Select the template you want to edit.
information Click the update button to save the change
after editing.

Select the item you want to delete. Click


Delete Item this button to delete the the template. (The
deleted item will be moved to recycle bin).

122
User’s Manual V2.11.1E

Click this button to enter the recycle bin


Recycle bin interface to find deleted templates.

Button Function Describe


Recover the deleted Select the items or class you want to recover
class or item and click this button. Then the selected items or
class will recover to the knowledge base
Completely delete the
Select the items or class you want to delete
deleted items or class in
completely and click this button. Then the selected
the recycle bin
items or class will delete completely and it can be
recovered forever.
【5】 Tools and Setting

Click button to enter the Dictionary Setting interface:

123
User’s Manual V2.11.1E

You can add and update the doctor information in the interface.
Button Function Describe
When there are more than one monitor,
Multi-screen display the doctor can write a report while viewing
the image.

Save Click this to save the report. And the


report will enter read only mode and looks as
follows.

Print Click to enter the print report interface.


See chapter 7.1

Exit Click to quit the report interface and


back to the study list interface.

b) Report Printing

Click the print button to enter the print report interface:

Button Function Describe


Quick Print Print current report with the default printer.

Print Select installed printer to print report.

124
User’s Manual V2.11.1E

5.Image archive, export and printing

This chapter will concentrate on image archive, export and printing which used
in the study list.

Study list

a) Image archive
In archive process, the images will be sent to PACS server follow the DICOM
3.0 standard.
Before the use of image archive, the parameters of DICOM storage must be
properly configured (reference to Configuration manual).

User can click button “image archive” to execute


the image archive process. But it is unnecessary if configured to archive automatically
after end of a study.

125
User’s Manual V2.11.1E

If there is a failed message, then user should check the communication with the
PACS.
b) Image export
The system provides two ways of export, and support customized export. User
can export all or selected patient’s data from the study list.
i. Burning CD
Select the patients that need to be imported or burned. Hold down Ctrl to select
multiple patients

Click on Burn export button popup window

CD2GO
Enter or select the appropriate information. FileSetID:The name of a burning or
producing ISO file. File compression: Standard, Compression.
Structure: Selecting DCMIMG means that only DICOM images will be output,
and the system will automatically generate by patient name and check well
File’s name: Select DCMDIR to output the patient's information and DICOM
image as a CDViewer. Check BMP or JPG to output the image in either format
Save Settings: Generate ISO files or burn directly to CD. ISO files can be

126
User’s Manual V2.11.1E

selected to save the file directory, burning CD to choose burning device


Execute after the configuration is completed

Click on the , The schedule displays the progress of the execution,


and a checkbox prompts if it succeeds
ii. Export to selected folder
Besides burning CD, user can also export patient’s data to a selected folder. First
user should select a folder for item “ExportStudy”. And then there are two items:
DCMIMGA: Export only DICOM images.
The export will create a fold named with patient name and accession. For
example: Emergency41-000041. All DICOM image of current patient will save to the
folder.
DCMDIR: Export patient’s information, DICOM images with a viewer program.

DCMDIR export
Running the CDViewer:

CDViewer

127
User’s Manual V2.11.1E

Double click the select patient or click preview , user can preview the
image of current patient.

Image preview of CDViewer


The CDViewer also support to upload patient images to PACS server after
configuration.

Image uploading of CDViewer


c) Image print
When accept the print operations, system will send images to printers that
support DICOM print services protocols, and user can defined the parameters of these
printers.

In the image preview interface, click button “Print”:

128
User’s Manual V2.11.1E

Print preview

The preview interface provides several items:

Click the , user can configure the property of select printer. All the property
is referenced to DICOM 3.0 standard.

Configuration of printer
Choose layout of images:

Film size: the film size should consistent with the printer.
File orientation: “LANDSCAPE” for horizon, “PORTRAIT” for vertical. Print

129
User’s Manual V2.11.1E

Num: the copy of current image.


Print Range: Print all the pages or only current page. Zoom Type: Print at ratio of
1:1 or auto ratio.
Button Describe
Switch between dicom printing and local printing.

Select installed printer to print report.

Start printing

Delete the currently selected image

Refresh image Delete

all images

Add other images

6.Exit

Click , Pop up the exit confirmation box:

130
User’s Manual V2.11.1E

Enter the login password, click the system saves the image and exits
,

the console software; click , Cancel the exit operation and return to the
software operation interface.

6.2 Use of Equipment - Digital Radiography Mode

6.2.1 Fluoroscopy

When you need to fluoroscopy, you need to enter the image acquisition
workstation first, set up the image acquisition options, set the image acquisition
fluoroscopy, step the fluoroscopy foot switch under the console. The system will
display the image in real time. When the foot switch is loosened, the final image will
be kept and displayed automatically.

6.2.2 Radiography

When photography is needed, it is necessary to enter the image acquisition


workstation first, set the photography parameters, press the exposure handbrake I, the
bulb anode starts, the filament warms, wait for the READY light to turn on, press the
exposure handbrake II, the generator exposes, and the exposure indicator lights up and
makes a "beep" sound. Release the handbrake when photography is over.

6.2.3 Fluoroscopy Spot-film

Enter the image acquisition workstation first, set the image acquisition options, set
the image acquisition perspective options and photography conditions, step on the
perspective foot switch under the console, the system will display the image in real
time, move to the area of interest, press the exposure hand brake for photography.

6.2.4 Emergency Relief of Compressor Pressure

When the device is in use and the power supply is suddenly interrupted and the
compressor is in the position of pressing the patient, in order not to cause injury to the
patient due to the long duration of the pressure, please press the button marked
"release button" in the figure below to lift the pressure cup of the compressor upward
and relieve the pressure on the patient.

131
User’s Manual V2.11.1E

Before pressing the button, press the pressure cup in the direction of the patient,
and then press and hold the button to eliminate the pressure cup in the direction of the
patient, and the pressure cup will automatically spring up.

Pressure relief indication

6.2.5 Initial RUN-UP

The initial run-up of the equipment after it has first been installed, by the
professional installation personnel in accordance with the adjustment description of
the high voltage generator in the specification, until the initial run-up is completed.

132
User’s Manual V2.11.1E

CHAPTER7 MAINTENANCE

7.1 Maintenance by Operator

1. General Rules
This equipment needs regular maintenance; following suggestions will be good for
usage and safety.
● Keep room clean, dry, and with good ventilation.
● Mechanical safety will be degraded due to long time usage. In order to reduce
potential damage to both user and patient, it is necessary to check mechanism if any
loose and if brake is good.
● Correct electrical calibration can well perform machine’s performance.
Calibration work should be completed by professional staff authorized by our
company.
● Machine should be in charged by designed person and establish file management
system.
● Any changes to machine are forbidden to ensure both patient and operator’s
safety, if any changes needed, please contact with manufacturer.
● Adjustment for contact gap among moving parts with orbital bearing assembly
should be done once a year.
● Regular big repair should be done, for details pls contact with manufacturer.
● If product has malfunction, contact with professional staff approved by
manufacturer.
● Operator should be well trained and can complete normal inspection.

2. Maintenance and Repair for unit


(1) Cleaning the surface of the unit,can use soft cloth with a little anhydrous alcohol
to scrub.
Organic solution can't be used to scrub the surface of the machine parts absolutely.
(2) Anti-bacterial requested machine surface touched by patients, such as sterilizing
the table panels, can use the ethanol for disinfection, but mercurochrome, iodine are
not allowed. It is suggested that Ultraviolet radiation etc. is configured on a regular
disinfection in the machine room.

133
User’s Manual V2.11.1E

For the disinfection work, unit power must be shut down!


And disinfection liquid should be paid attention that it
can’t flow into the unit interior, lest the phenomenon of
leakage and short circuit is occurred, which will lead to all kinds of malfunction,
even cause injuries.

3. Maintenance and Repair for collimator


(1) Collimator should be kept clean, dry, transmission parts and components shall
not have dust.
(2)After period of time usage, check whether the bar lead-leaf get loose, and fix it
in time (check it every three to six months).Few thin lubricant can be added on
internal mechanical parts.
(3)When any malfunction is happened, it should be done by professional person or
contact factory in time for maintenance.
4. Maintenance and repair for X-ray tube assembly
(1)After continuous exposure, it should pay attention to X-ray tube temperature
lift. When the temperature of outer surface components exceed 75 ℃,X-ray generator
must stop loading automatically.
(2)It should ensure that high-voltage cable plugs, sockets have good contact so
X-ray tube filament is always on, and it can avoid X-ray tube to withstand cold
high-voltage, thus extended x-ray tube life.
(3)When X-ray tube components are used for the first time or idled for a longer
time, it must be progressive loading before X-ray tube used, and do the sophisticated
test to avoid damaging to the X-ray tube by instantaneous load.
5. Maintenance and repair for high-voltage generator
(1)It should ensure that high-voltage cable plugs, sockets have good contact so
X-ray tube filament is always on, and it can avoid X-ray tube to withstand cold
high-voltage, thus extended x-ray tube life.
(2)After removal of high voltage cable sockets, it should pay attention to the
residual charge, do not touch electrode by hand to avoid electric shocks. If needed,
discharge it first.

Attention: Clean dust on top of high-voltage generator.

134
User’s Manual V2.11.1E

6.Maintenance and repair for tables


(1)All surfaces of Radiography table should be kept clean, dry , and variety of
residues can be scrubbed with a soft cloth dipped with few absolute alcohol
(2)Regular check if bearing or movement orbit is short of oil .After cleaning, it
should coat with lubricant (grease) in time.
(3)No flammable liquids, gases near the equipment to prevent explosion.
(4)Protect grounding facilities, and inspections should be done every half year.
(5)Diagnostic table’s sterilization methods are same as the unit’s.

7.2 Maintenance by the Authorized Technician

PLD6000 Medical Diagnostic X-ray System is sophisticated, valuable medical


diagnostic equipment. It is necessary that user’s knowledge about machine’s technical
performance, structure principle, operating procedure and daily maintenance to give
full play to the efficiency of the machine, prolong the useful life of the machine,
ensure the normal use of equipment, ensure that the safety of operators and personnel
inspected.
1.Properly operating in the process of using X-ray machine, make its normal
operation, and what’s more important is that it should pay special attention to the
body of X-ray protection, make X-ray operator and the patient's damage down to a
minimum. For example, machine room should have protective equipment and
measures (e.g., lead plate), the operator must use protective appliances (such as
lead glasses) and wear protective clothing (lead cap lead, lead aprons, gloves),
operator should be occupied area in effective operation when the patients with
radiation inspection.
2.This machine should be placed in air circulation, clean and dry work condition. It
cannot be placed in the humidity, heat and sunlight exposure area.
3.Control cabinet should be opened front and back cover regularly (depending on the
environment, such as the six months to one year), and removed internal dust. For
damaged parts, purchase the same model of spare parts from factory for
replacement. During examination, pay attention whether there is any loose
terminals and displacement phenomena, and fasten them.
4.In area where the power supply and voltage has a high fluctuation, always pay
attention to whether power supply and voltage is in permitted range. If it has high,

135
User’s Manual V2.11.1E

low, mutation and other anomalies, machine should be closed immediately and
check out the problem, it can only be used after problem solved.
5.Do not press heavily on all buttons of controller to avoid dislocation, injury or poor
contact, even damaged.
6.Check frequently protective earth of each components is in good condition
(grounding resistance is very small).
7.Check power supply on a regular basis, and check power supply resistance value (or
the power supply voltage value is changed), to make sure it meet machine’s
requirement.
8.After using the controller, it should be cut off the power in time.
9.Avoid striking of x-ray tube components during operation.
10.When X-ray tube components are used for the first time or idled for a longer time,
it must be progressive loading before X-ray tube used, and do the sophisticated
test to avoid damaging to the X-ray tube by instantaneous load.
11.When machine works continuously, please pay attention to the X-ray tube
assembly to allow Thermal capacity, or check surface temperature of X-ray tube
assembly , it can’t exceed 75 ℃.

The personnel for the above inspection, maintenance and


processing must be accredited by professionals; It’s not
allowed for other people to carry out such inspection,
maintenance and processing.

136
User’s Manual V2.11.1E

7.3 The Cycle of Maintenance and Repair

Time Interval Contents of Maintenance and Inspection

 Check if the signal, display and indicator are normal.

 Check if warming and danger SYMBOLS are good.

 Check if X-ray generator assembly has abnormal noise


Daily
during exposure.
Inspection
 Check if all grounding is good.

 Check if there is any loose, stuck on the button of the


Console.

 Check if any oil leakage of the X-ray generator component.

 Check flexibility of table and column bearing, observe if


any friction is too large, and coat bearing orbit with
Weekly
Inspection lubricating oil.

 Check if any loose and shedding of fastener parts between


components, if any need, it should be fixed them in time.

 Check machine grounding resistance (less than 0.2Ω).

 Check if any filler come out of head of the HV cable


because of heating.

 Clear silicone oil on high-voltage outlet, scrub High


Check once six voltage plugs and sockets with ether or carbon
month
tetrachloride, and then smear some new silicone grease,
tighten the fixing ring of plug.

 Check the center of X-ray generator assembly and


collimator.

137
User’s Manual V2.11.1E

7.4 Fault Code and Troubleshooting

7.4.1 HV Generator

Fault
Information Adjustment method
Code
ER00 Communication fails 1. Check if the CANBUS communication
wire of the upper computer is of good
contact.
ER1 First gear of hand break 1.Firstly, turn off high voltage generator ,
plugs in then using avo meter to check if the 1st gear
during self-checking. of handbrake is right. If the handbrake is
abnormal, change it and then check again. If
difficult is settled, problem will be closed
or,otherwise, proceed the next procedure
2.check the wire from mini control box to
the plug of power panel PP1
3.change interface CPU board and turn on
high voltage generator to check , if fault
disappears, problem will be closed;
4.if trouble is still there after above
procedures, you can shut down, change
main cpu board, and then check again. If
trouble disappears, it can be ascribed to
handbrake problem and problem will be
closed.

138
User’s Manual V2.11.1E

ER2 Second gear of hand break 1. Firstly, turn off high voltage generator,
plugs in using avo meter to check if 1st gear of band
during self-checking. brake signal is right and, if abnormal,
change handbrake and go on your
examination. If trouble is settled, it can be
credited to handbrake problem and problem
will be closed. Otherwise, proceed the next
procedure.
2.check the wire from mini control box to
the plug of power panel PP1
3. Change main control board and turn on
high voltage generator to check. If fault
disappears, it can be attributed to main
control board problem and problem will be
closed;
4. If trouble remains after above procedures,
you can shut down and change interface
board to check. If fault disappears, it can be

attributed to interface board problem and


problem will be closed;

ER03 Not received PHVOK 1. Voltage feedback wire in bus is of bad


signal during soft powering contact. Please close the generator and
check if wires are rightly connected.
2.malfunction of power distribution board;
change power distribution board;

139
User’s Manual V2.11.1E

ER4 Tested out excess feedback 1. Check if the network electricity is in the
value of bus voltage AD range of generator standard limitation, if
during soft powering beyond, turn of the machine;
2. Malfunction of power board; change
power board ;

ER5 Tested out too 1. Check if the network electricity is in the


low feedback value of bus range of generator standard limitation, if

voltage AD beyond, turn off the machine;


during soft powering 2. Malfunction of power board; change
power board ;

ER6 SAMFAULT signal is low 1. Change the main control board


level during self-checking

ER7 Tested out excess feedback 1. Check if the network electricity is in the
value of bus voltage AD range of generator standard limitation, if
after main beyond, turn off the machine;
contactor plugged in. 2. Malfunction of power distribution board;
change power distribution board;

ER8 Tested out too 1. Check if the network electricity is in the


low feedback value of bus range of generator standard limitation, if
voltage AD after main beyond, turn off the machine;
contactor plugged in. 2. Malfunction of power distribution board;
change power distribution board;

ER9 SLINK is high level during 1. Check if the ground wire of oil tank is
self-checking. It is possible rightly linked
that oil tank is not linked or 2. Check if the wire between main control
didn`t attach ground wire. board and oil tank is rightly linked

ER10 STLOW output is low level 1. Change anode starter board


during self-checking.

140
User’s Manual V2.11.1E

ER11 Hard signal to filament 1. Change filament board


board FLIAFAULT is low 2. Change the main control board
level even when
the filament has yet to
be heated
during self-checking.

ER12 Feedback to small 1. Change filament board

lamp filament
malfunctioned
during self-checking.

ER13 Hard signal to filament 1. Change filament board


FLIAFAULT is low level 2. Change the main control board
during filament heating in
the course of self-checking.

ER14 Feedback to big 1. Change filament board 2.check the link of


lamp filament cathode high voltage cable
malfunctioned
during self-checking.

ER15 PHVOK
Please contact technical support
signal malfunctioned.

ER16 Excess bus voltage Please contact technical support

ER17 Bus voltage deficit Please contact technical support

ER18 Temperature control signal


in lamp tube Please contact technical support

malfunctioned.

ER19 Not connected oil tank Please contact technical support

ER20 Read-write 2864 failure Please contact technical support

141
User’s Manual V2.11.1E

ER21 Pressed the 1st gear of


Please contact technical support
hand brake as configuration

ER22 Filament feedback deficit Please contact technical support

ER23 Excess filament feedback Please contact technical support

ER24 FILAFAULT induced


Please contact technical support
break-off.

ER25 SAMFAULT induced


Please contact technical support
break-off.

ER26 Other signal induced


Please contact technical support
break-off

EL27 Detect that door switch


signal is high level when
pressed 1st gear of Please contact technical support

handbrake to ordinary
shooting.

ER28 Detect that FILIAFAULT


signal is low level during Please contact technical support

sudden heating

ER29 STLOW is Low level. Please contact technical support

ER30 Correspondence failure


Please contact technical support
during preparation

ER31 Not received hand 2 Please contact technical support

ER32 Door switch is high level Please contact technical support

ER33 Not detected signal from


Please contact technical support
2ed gear of handbrake

ER34 Not received return signal


within 1s after sending out Please contact technical support

XR1

142
User’s Manual V2.11.1E

ER35 KVAD feedback


Please contact technical support
malfunction

ER37 Braked in advance Please contact technical support

ER38 Hand 2 not loosed Please contact technical support

ER39 Hand 1 not loosed Please contact technical support

ER40 Signal malfunction at


Please contact technical support
AECEND

ER41 Door switch malfunctioned Please contact technical support

ER42 Filament malfunction Please contact technical support

ER43 Rotating anode STLOW


Please contact technical support
malfunctioned

ER44 Correspondence failure


Please contact technical support
during preparation

ER45 Not received hand 2 Please contact technical support

EL46 Door switch malfunctioned Please contact technical support

ER47 KV feedback
Please contact technical support
malfunctioned

ER48 2ed gear of handbrake


Please contact technical support
loosed in advance

ER49 2ed gear of handbrake


Please contact technical support
malfunctioned

ER50 Door switch malfunctioned Please contact technical support

ER51 FILAFAULT failure during


Please contact technical support
auto correction

ER52 STLOW failure during auto


Please contact technical support
correction

143
User’s Manual V2.11.1E

ER53 Correspondence failure


during preparation for
Please contact technical support
exposure in the course of
auto correction

ER54 Not received signal from


handbrake 2 during auto Please contact technical support

correction

ER55 Handbrake 2 loosed in


advance during auto Please contact technical support

correction

ER56 Correspondence failure


during exposure in the Please contact technical support

course of auto correction

ER57 KV feedback failure during


Please contact technical support
auto correction

ER58 Handbrake loosed in


advance during auto Please contact technical support

correction

ER59 Excess inverter current Please contact technical support

ER60 Error-reporting due to


excess thermal capacitance Please contact technical support

HEVALUE

ER61 Maximum value of Large


focus filament beyond Please contact technical support

2400 during auto correction

ER62 Value of small focus


filament beyond 2060 Please contact technical support

during auto correction

144
User’s Manual V2.11.1E

ER63 self-tube-training failure Please contact technical support

ER70 PHVOKMA malfunction


induced breakoff during Please contact technical support
self-checking

ER86 2rd gear of hand break 5s


not loosed after exposure Please contact technical support
during AEC shooting

ER20 Communication failure


Please contact technical support
7 during exposure

7.4.2 Diagnostic Table Fault Code

Fault code Adjustment method


Diagnostic E01 Communication fault of table, check 485 chip and
table communication line.
E02 Communication fault of Film, check 485 chip and
communication line.
E03 Communication fault of collimator, check 485 chip and
communication line.

145
User’s Manual V2.11.1E

CHAPTER8 ATTACHMENT

8.1 Appendixes Attached with Following Circuits

1.Medical Diagnostic X-ray System:PLD6000 JX


2.The Connection Diagram of System Console
3.The connection diagram of diagnostic table:ZDC20.00JX (1)(2)(3)

146