Professional Documents
Culture Documents
Scanpro Moisture Sensor
Scanpro Moisture Sensor
704228-C
Rev Date Pages
All pages.
C November 1, 2002 Describe changes to screens and revised
measurement processing.
13-22
B April 13, 2001
Describe new screens and buttons.
NDC INFRARED ENGINEERING
11/01/02
SCANPRO MOISTURE SENSOR
INTRODUCTION
The Scanpro single (Figure 1) or dual head1 (Figure 2) microwave sensor is used in combination
with a basis weight sensor, such as the NDC beta or gamma sensor, to measure the moisture
content of a product such as paper or paperboard.
The Scanpro Dual Head sensor has three different operating modes:
Start mode - activated when power is turned on to the system. The LED on the sensor
remains lit in Start mode.
Garage mode - activated during the standardization routine to calculate corrections for
deviations in the gap. The LED on the sensor remains lit in Garage mode.
Normal mode - normal operating mode for collecting measurements. The LED on the sensor
flashes in Normal mode.
The Scanpro Single Head sensor has only one operating mode and no LED.
1
For the dual head sensor, the measurement gap should be between 7-11 mm, ± 1.5 mm.
· Data Collection
· Standardization
· Curve Fit
· Temperature Compensation
q Data Collection
· The first input is a 0 – 10 V signal that is linear with the weight of water. This input
connects to the Scanpro gauge output.
q Standardization
The Scanpro sensor signal is standardized against a zero reading taken when the
standardization routine runs. Readings are taken during the standardize period when the
shutter is open. During standardization, the average value for the Scanpro signal is computed.
This value is subtracted from the raw input signal during normal measurements.
q Curve Fit
where
A0, A1, A2 and A3 are coefficients entered on SCANPRO CALIB SET-UP menu Page 1
The output is basis weight compensated, based on mass readings from the basis weight
sensor. A predetermined basis weight setpoint is used if the system does not include a basis
weight sensor. If the basis weight measurement is not in grams per square meter, it is
converted to that unit.
This section describes the setup procedure performed at the NDC factory, and on-site by the
NDC Field Service engineer.
Step 4 - Create at least two separate Grid Setups and assign the moisture grid to the
moisture sensor (optional)
2. Touch NUM. SCANNERS and set the number of scanners in the system.
3. Touch NUM. GAGES and set the number of gauges in the system. This number will be
larger than NUM. SCANNERS because of the basis weight and moisture gauge on the same
scanner.
4. Check that the SCANNER and GAGE REAL/SIM toggles are set to REAL.
SCANNER GAGE
1 REAL SIM 1 REAL SIM NUM. SCANNERS. . . . 1
NUM LINESPEEDS. . . . 1
PAGE 1 OF 7
1. Touch ->>- from the BASIC CONFIGURATION SETUP page to move to the GAGE
CONFIGURATION SET-UP page (Figure 4).
2. Scanners must be manually assigned as there are two gauges mounted on a scanner and the
NUM. GAGES entry (on the BASIC CONFIGURATION SETUP page) is greater than
NUM. SCANNERS.
To assign a scanner, touch the blue SCANNER button and select a scanner for the gauge
shown on the left. See Section 14.3 of the 8000 manual.
3. For each gauge, touch GAGE TYPE and select the type of sensor. Select SCANPRO for the
Scanpro sensor.
See Section 14.4 of the 8000 manual for a discussion of gauge types.
7. Touch ->>- > ->>- > ->>- > ->>- to move to the PORT MAP page.
8. Touch PRINT PORT MAP to print a list of port allocations for gauges and system options.
Verify that the ports are correctly allocated.
The DRY TOTAL toggle (on the CONFIGURATION page: SET-UP > SITE MENU > CONFIG
> ->>-) can be enabled for systems with a moisture sensor. If this toggle is ON, the Net is
(100% - Moisture)
calculated as NET = ´ Mass
100
2. If there is one Scanpro sensor in the system, set the label for the Scanpro sensor to SPRO.
If there is more than one Scanpro sensor in the system, set the labels for the Scanpro sensors
to SPRO1 and SPRO2.
LABELS
GAGE1 GAGE1
GAGE2 GAGE2
GAGE3 SPRO1
GAGE4 SPRO2
PAGE 2 OF 3
UNITS
GAGE1 . . . . G/M2
GAGE2 . . . . G/M2
SPRO1 . . . . PERCENT
SPRO2 . . . . PERCENT
PAGE 2 OF 2
2. Assign units for the gauges, choosing PERCENT for the Scanpro sensor(s).
2 MOIST
PAGE 3 OF 3
2. Touch GRID SET-UPS and select at least 2 grid setups, allowing one for the moisture
sensor and as many others as required by your installation.
A maximum of 4 setups is allowed.
3. Touch GRID SET-UP LABELS and enter a descriptive name to identify each grid.
(E.g., If there are two grid set-ups, the grid labels could be “MASS” and “MOIST.”)
5. Assign a grid to each gauge, choosing the moisture grid for the moisture sensor.
LABELS GRID
PAGE 2 OF 3
Note: This step is not optional. If only one grid is used, the UNITS for that grid must be set
to #.
X-AXIS Y-AXIS
SELECT
MIN . . . . . 0 HI MED LO
MASS
MAX . . . . . 25 MAX . . . . . 5.0 15.0 25.0
MOIST
STEP . . . . 5 1.0 3.0 5.0
STEP . . . .
REDRAW
TOP UNITS % #
Parameters relating to Scanpro setup are found in the SCANPRO CALIB SET-UP menu, and on
the PROCESS page.
Unless otherwise noted, the settings in this menu are assigned at the factory, and should not
be changed at any time.
Move to the SCANPRO CALIB SET-UP menu (SET-UP > SITE MENU > GAGE MENU >
SCANPRO CALIB). There are three pages in this menu. Use the arrow button on the top right
corner to scroll the pages.
If there is more than one Scanpro sensor in the system, there is a separate set of setup pages for
each Scanpro sensor. The screen will show the parameters for the currently selected Scanpro
sensor. The label at the top of each page shows the current Scanpro sensor (see Figure 10). Use
the arrow buttons (-<< and ->>-) on each side of the label to select a different Scanpro sensor.
A3. . . 0.0
PAGE 1 OF 3
Buttons on SCANPRO CALIB SET-UP menu Page 1 (Figure 10) are described below.
1 2 Selects a different Scanpro sensor and displays the parameters for that sensor.
q TEMP COMP
q BASIS WEIGHT
If this toggle is set to GAGE_, readings from the basis weight gauge will be used in
calculations of percent moisture. If SETPT is selected, the BW SETPOINT value on the
PROCESS page will be used in calculations of percent moisture. See Figure 13.
PAGE 2 OF 3
Figure 11 shows SCANPRO CALIB SET-UP menu Page 2. Buttons on this page include:
A multiplying factor for the temperature delta (difference between the current temperature
and the temperature at standardization), expressed in gsm/deg C. In practice, this is the
negated temperature coefficient of the gauge.
q DIV BY BW
Sets whether division by basis weight is used in the Scanpro algorithm. Set this toggle to
OFF if the Scanpro sensor measurement is in weight per unit area of water. Set this toggle to
ON if the Scanpro sensor measurement is a percentage of basis weight.
Full scale temperature. Set this value to 100. This applies only to single-sided Scanpro
sensors.
bw (G/M2) 80.0
PAGE 3 OF 3
The SCANPRO DEBUG page (Figure 12) is used primarily by NDC personnel when testing the
gauge at the factory, and during subsequent troubleshooting.
This page includes the intermediate variables used in the calculation of percent moisture. In the
case of a suspected measurement problem, you may be asked to print and fax this page to
NDC Infrared Engineering.
Touch the PRINT button to print a copy of the parameters on this page. The REFRESH button
updates all the data values on this page. Invalid values are reverse printed on orange.
q S-pro Volts
q Volts - Stdz
q Linearized
q - Tempcomp
q / Basis Wt
Calculated value based on temperature delta (difference between temp now and
temp at std) and Temp coeff.
Current temperature for the temperature sensor in the Scanpro sensor. If the system has a
beta gauge, the temp now value is obtained from the beta gauge gap temperature sensor.
Temperature at standardization.
q basis weight
Mass reading used by the Scanpro sensor in calculations. This reading is obtained from the
basis weight gauge or the BW SETPT value on the PROCESS page, depending on whether
the BASIS WEIGHT selection is a gauge label or SETPT (on SCANPRO CALIB SET-UP
menu Page 1).
q Temp coeff
Result of temp at std (C) x TEMP COEFF gsm/degc (entered on SCANRPO CALIB
SET-UP menu Page 2).
q bw (G/M2)
q std pos
PAGE 1 OF 2
The BW SETPT button on the PROCESS page (Figure 13) is only visible if the
BASIS WEIGHT selection is SETPT (on SCANPRO CALIB SET-UP menu Page 1).
BW SETPT sets the mass reading used by the Scanpro sensor in percent moisture calculations.
This is used to provide the mass reading only for systems without a basis weight gauge.