TEC-8352K Saving Data

4-2.11
While the “Turning power off : Saving data” message is displayed on the screen, do not
disconnect the power cord from the AC/DC module or AC outlet and do not remove the
battery pack and/or AC/DC module from the defibrillator. Otherwise, the internal data may
be damaged.

4-4.13
The following waveforms are saved in memory.
· Waveform of 10 seconds before and 12 seconds after discharge
· Waveform after the CHARGE/AED button is pressed and before analysis is complete

4-4.14
Waveforms after the CHARGE/AED button is pressed can be saved together with the analysis
result as a report. Up to 1,000 seconds of waveforms can be saved in memory. The oldest
waveforms are deleted when you save waveforms of 1001 seconds or more. For details,
refer to Section 7 “Recording and Reports”.

4.5
When there is an alarm, * is dsplayed in front of the data and the data is displayed in red.

7.28
· If you turn the defibrillator off within 30 seconds after turning the defibrillator on, data is
not saved.
· When the defibrillator automatically turns its power off due to battery expiration, the last
30 seconds data may be lost.

7.30
For details on operation on the QP-551VK Defibrillator Report Viewer Software, refer to the
oprator’s manual of the QP-551VK Defibrillator Report Viewer Software.

8.28
Alarm report, Event report, Defib report and VF analysis reports are automatically saved as
waveform reports when an alarm or event occurs.

SD Card

Problem Possible Cause


Upgrade Kit is inserted instead of an SD card.
The card is not inserted.
The card is not inserted properly.
A card other and specified card is inserted.
Data cannot be saved Memory is full.
in the SD Card. The card is write-protected.
The card is not formatted.
The SD card ware removed while the defibrillator power
was on and the defibrillator was affected by static
electricity.