TEC-8352K Saving Data
TEC-8352K Saving Data
4-2.11
While the “Turning power off : Saving data” message is displayed on the screen, do not
disconnect the power cord from the AC/DC module or AC outlet and do not remove the
battery pack and/or AC/DC module from the defibrillator. Otherwise, the internal data may
be damaged.
4-4.13
The following waveforms are saved in memory.
· Waveform of 10 seconds before and 12 seconds after discharge
· Waveform after the CHARGE/AED button is pressed and before analysis is complete
4-4.14
Waveforms after the CHARGE/AED button is pressed can be saved together with the analysis
result as a report. Up to 1,000 seconds of waveforms can be saved in memory. The oldest
waveforms are deleted when you save waveforms of 1001 seconds or more. For details,
refer to Section 7 “Recording and Reports”.
4.5
When there is an alarm, * is dsplayed in front of the data and the data is displayed in red.
7.28
· If you turn the defibrillator off within 30 seconds after turning the defibrillator on, data is
not saved.
· When the defibrillator automatically turns its power off due to battery expiration, the last
30 seconds data may be lost.
7.30
For details on operation on the QP-551VK Defibrillator Report Viewer Software, refer to the
oprator’s manual of the QP-551VK Defibrillator Report Viewer Software.
8.28
Alarm report, Event report, Defib report and VF analysis reports are automatically saved as
waveform reports when an alarm or event occurs.
SD Card