Atellica Solution

Adjustment
Atellica Chemistry (CH) Analyzer
IMT System
2 Document Version / Disclaimer / Copyright

Document Version
Siemens Healthcare GmbH reserves the right to change its products and services at any
time.
In addition, manuals are subject to change without notice. The hardcopy documents cor‐
respond to the version at the time of system delivery and/or printout. Versions to hard‐
copy documentation are not automatically distributed.
Please contact your local Siemens Healthineers office to order a current version or refer to
our website http://www.siemens-healthineers.com.

Disclaimer
Siemens Healthcare GmbH provides this documentation “as is” without the assumption of
any liability under any theory of law.
The content described herein shall be used by qualified personnel who are employed by
Siemens Healthcare GmbH or one of its affiliates or who are otherwise authorized by Sie‐
mens Healthcare GmbH or its affiliates to use such documents.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens Healthcare GmbH or one of its affiliates are not entitled to
use this documentation without prior written authority.

Copyright
“© Siemens Healthineers, 2018” refers to the copyright of a Siemens Healthineers entity
such as:
Siemens Healthcare GmbH - Germany
Siemens Shenzhen Magnetic Resonance Ltd. - China
Siemens Shanghai Medical Equipment Ltd. - China
Siemens Healthcare Private Ltd. - India
Siemens Medical Solutions USA Inc. - USA
Siemens Healthcare Diagnostics Inc. - USA and/or
Siemens Healthcare Diagnostics Products GmbH - Germany

Atellica Solution LDAT-010.842.03.06.02 Page 2 of 20 © Siemens Healthineers, 2018


06.20 Restricted
Table of Contents 3

1 Preparation 4

1.1 General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4


1.2 Material Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.2.1 Parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.2.2 Tools and Supplies. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

2 Align IMT 6

3 Align IMT Sample Port to Dilution Arm 9

4 Changes to Previous Version 14

4.1 Version 06 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
4.2 Version 05 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.3 Version 04 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.4 Version 03 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.5 Version 02 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
5 List of Hazard IDs 19

© Siemens Healthineers, 2018 LDAT-010.842.03.06.02 Page 3 of 20 Atellica Solution


Restricted 06.20
4 1 Preparation

1.1 General Information


Read this procedure in its entirety before starting the alignment.
For general safety information, refer to the Safety section of the Atellica Solution CB-DOC.

 WARNING
Carriers on the Atellica Magline Transport generate strong magnetic fields that
F can interfere with cardiac medical implants.
Use extreme caution while interacting with sample carriers installed on Atellica
M Magline Transport.
 Do not position pacemakers or other implanted medical devices closer than
7.5 cm (3 inches) from the Atellica Magline Transport.

Atellica Solution LDAT-010.842.03.06.02 Page 4 of 20 © Siemens Healthineers, 2018


06.20 Restricted
Preparation 1 5

1.2 Material Information

1.2.1 Parts
n.a

1.2.2 Tools and Supplies


n.a.

© Siemens Healthineers, 2018 LDAT-010.842.03.06.02 Page 5 of 20 Atellica Solution


Restricted 06.20
6 2 Align IMT

 If the vacuum pump control loop is not checked from the Instrument tab, run an Au‐
toCheck sub-system for the vacuum pump.

1. Verify that the peristaltic pump tubing is installed and the peristaltic pump latch is
closed on the tubing.
2. Verify that the IMT multisensor is installed and that the door is closed.
3. Click the Subsystem tab, then click IMT on the left side of the screen.
4. Click Auto Align Subsystem.

If the IMT rotary valve fails to auto align, fill the sample port with 0.5 mL DI water and
 click the Cycle Targets several times until the sample port empties. Then redo Auto Align
IMT RV.

Fig. 1: Auto Align Subsystem

5. Click the Sequences tab.


6. Select IMT on the left side of the screen.

Atellica Solution LDAT-010.842.03.06.02 Page 6 of 20 © Siemens Healthineers, 2018


06.20 Restricted
Align IMT 2 7

7. Click Prime All Fluids.


Fig. 2: Prime All Fluids

8. Click Align Peristaltic Pump.


 The typical range is 50-70 µL/revolution.
Fig. 3: Align Peristaltic Pump

9. Click Std A Cycle.

© Siemens Healthineers, 2018 LDAT-010.842.03.06.02 Page 7 of 20 Atellica Solution


Restricted 06.20
8 2 Align IMT

10. Click Std B Cycle.


Fig. 4: Standard A and B Cycle

The result should be black numbers in both Air and Liquid boxes for both Std A and
Std B; if the numbers are not black, repeat the prime by clicking Prime All Fluids (step
6).

Atellica Solution LDAT-010.842.03.06.02 Page 8 of 20 © Siemens Healthineers, 2018


06.20 Restricted
Align IMT Sample Port to Dilution Arm 3 9

The IMT Alignment tool helps to ensure that the manifold sample port position is in cor‐
rect radial alignment to the dilution arm. Once this alignment is set, the Dilution Arm au‐
to-alignment will position the probe in the sample port in the optimal off-center position
for performance.

1. Enter Diagnostic Mode:


 If the primary display is locked, log on with the CSE logon.
a) Click the System tab.
b) Click the Status tab.
c) Click the CH analyzer icon.
d) From the options on the left, click the Diagnostics button (if the analyzer is not in
the 'Stopped' state).

 The Stopped state will also allow the user to enter Diagnostics.

2. Connect to the analyzer diagnostics via Remote Module Access ( Remote Module Ac‐
cess (RMA) / LDAT-000.840.03).
3. Unlock the front lid:
a) In Diagnostics, click the Instrument tab.
b) Click Unlock Front Lid.
4. Open the front lid.
Fig. 5: Opening the Front Lid

(1) Front Lid

5. Verify that the Probe Radial Alignment ( Probe Radial Alignment / LDAT-010.842.06)
and the Align Upper Arm to Home ( Align Upper Arm to Home / LDAT-010.842.06)
have been completed successfully.
6. Auto-align the dilution arm if adjustments were made in step 5. Otherwise, continue
to the next step.
a) Click the Auto Align tab.

© Siemens Healthineers, 2018 LDAT-010.842.03.06.02 Page 9 of 20 Atellica Solution


Restricted 06.20
10 3 Align IMT Sample Port to Dilution Arm

b) Check the DilutionArm checkbox.


c) Click Auto Align All Selected.
7. Insert the IMT service tool ( Atellica CH Analyzer Special Tools / LDAT-000.810.01)
into the sample port of the IMT Detector Manifold.
Fig. 6: Inserting the IMT Alignment Tool

(1) IMT Alignment Tool


(2) Manifold

8. Move the dilution arm to the IMT sample port:


a) In Diagnostics, click the Subsystem tab.
b) On the left side of the screen, select DilutionArm.
c) Under the angular motor alignment points, click IMTSamplePort.
9. Step the arm down until just above the oval slot in the IMT alignment tool by using
the large down arrow under the vertical motor section. The default step value should
be 192.

Atellica Solution LDAT-010.842.03.06.02 Page 10 of 20 © Siemens Healthineers, 2018


06.20 Restricted
Align IMT Sample Port to Dilution Arm 3 11

10. Visually verify that the probe will enter the oval slot in the IMT alignment tool. If the
probe is within the oval slot then this procedure is completed. Otherwise, continue to
the next step.
Fig. 7: Centering the Dilution Probe

(1) Dilution Probe


(2) IMT Alignment Tool

© Siemens Healthineers, 2018 LDAT-010.842.03.06.02 Page 11 of 20 Atellica Solution


Restricted 06.20
12 3 Align IMT Sample Port to Dilution Arm

11. Using a long T-30, loosen the three screws that mount the manifold mounting bracket
to the IMT base mount.
Fig. 8: Loosening the Manifold Mounting Bracket Screws

(1) Mounting Screws


(2) IMT Base

12. Move the IMT Detector Manifold and bracket front to back until the slot in the IMT
alignment tool and the probe line up.
13. Once the alignment is complete maintain this position with one hand while tighten‐
ing the bracket mounting screws with the other.
14. Initialize the Dilution Arm by clicking the Initialize Subsystem button on the bottom
left of the screen.
15. Remove the IMT alignment tool from the IMT sample port.
16. Auto align the dilution arm:
a) Click the Auto Align tab.
b) Click the DilutionArm checkbox.

Atellica Solution LDAT-010.842.03.06.02 Page 12 of 20 © Siemens Healthineers, 2018


06.20 Restricted
Align IMT Sample Port to Dilution Arm 3 13

c) Click Auto Align All Selected.


Fig. 9: Dilution Probe Dispense Position in IMT Sample Port

(1) IMT Subsystem


(2) Dispense Position
(3) Target Area

© Siemens Healthineers, 2018 LDAT-010.842.03.06.02 Page 13 of 20 Atellica Solution


Restricted 06.20
14 4 Changes to Previous Version

4.1 Version 06
Change No. Changes
LFT0012656 2. Align IMT
n ( Align IMT / Page 6): added a typical range to step 8.

Atellica Solution LDAT-010.842.03.06.02 Page 14 of 20 © Siemens Healthineers, 2018


06.20 Restricted
Changes to Previous Version 4 15

4.2 Version 05
Change No. Changes
LFT0012527 3. Align IMT Sample Port to Dilution Arm
n ( Align IMT Sample Port to Dilution Arm / Page 9): added a 3D animation to the
procedure.

© Siemens Healthineers, 2018 LDAT-010.842.03.06.02 Page 15 of 20 Atellica Solution


Restricted 06.20
16 4 Changes to Previous Version

4.3 Version 04
n Added ( Fig. 9 Page 13) and added a new step 6 to the “Align IMT Sample Port to
Dilution Arm” procedure ( Align IMT Sample Port to Dilution Arm / Page 9).

Atellica Solution LDAT-010.842.03.06.02 Page 16 of 20 © Siemens Healthineers, 2018


06.20 Restricted
Changes to Previous Version 4 17

4.4 Version 03
n Added a new step 2 to the “Align IMT“ procedure ( Align IMT / Page 6).
n Added the “Align IMT Sample Port to Dilution Arm” procedure ( Align IMT Sample
Port to Dilution Arm / Page 9).

© Siemens Healthineers, 2018 LDAT-010.842.03.06.02 Page 17 of 20 Atellica Solution


Restricted 06.20
18 4 Changes to Previous Version

4.5 Version 02
n Initial publication to CB-DOC.

Atellica Solution LDAT-010.842.03.06.02 Page 18 of 20 © Siemens Healthineers, 2018


06.20 Restricted
List of Hazard IDs 5 19

There are no Hazard IDs in this document.

© Siemens Healthineers, 2018 LDAT-010.842.03.06.02 Page 19 of 20 Atellica Solution


Restricted 06.20
- Restricted - All documents may
only be used by authorized person‐
nel for rendering services on Sie‐
mens Healthcare Products. Any
document in electronic form may
be printed once. Copy and distribu‐
tion of electronic documents and
hardcopies is prohibited. Offenders
will be liable for damages. All other
rights are reserved.

siemens-healthineers.com

Siemens Healthineers Headquarters


Siemens Healthcare GmbH
Henkestr. 127
91052 Erlangen
Germany
Telephone: +49 9131 84-0
siemens-healthineers.com

Print No.: LDAT-010.842.03.06.02 | Replaces: LDAT-010.842.03.05.02


Doc. Gen. Date: 06.20 | Language: English