160 APPENDIX B: GC/MS TROUBLESHOOTING QUICK REFERENCE
Problem 10: High vacuum pump would not turn on. No pressure reading.
Cause a: Filament is burned out. No filament voltage.
Treatment: Shutdown vacuum. Replace filament.
Cause b: System leak. Rough pump cannot reach starting vacuum.
B.4 MS SOURCE AND CALIBRATION PROBLEMS 163
Problem 12: With repeller at maximum, cal gas 502 cannot be found.
Cause a: Source is dirty.
Treatment: Shut down vacuum. Clean ion source and lens.
Cause b: Mass axis is badly out of calibration.
Treatment: Autotune. If 502 is still off scale, recalibrate mass 131
against 69, then 264 to 131. Cal gas 502 should be on
scale. Calibrate 502. Recheck the 131 and 69 masses.
Cause c: Detector showing loss of sensitivity/burn out.
Treatment: Increase EM voltage to 3500 V. If still no 502, replace
the detector.
164 APPENDIX B: GC/MS TROUBLESHOOTING QUICK REFERENCE