Specification Sheet: Specifications TC20B

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 1

Tamson Instruments

Specification sheet
Specifications TC20B
Tamson Circulator 20 litres

Item Unit TC20B


P/N
00T0160
230V/50~60Hz
P/N
00T0165
115V/60Hz
Power [W] 1450
Used materials
Stainless steel
inside bath
ambient .. +200C
Range
ambient.. +392F
Reading Standard C, F on request
Setting [°] 0.1
Stability ± [°C] Better than 0.02
Heating [W] 1400 (1 heater)
Bath volume [L] 20
Pump pressure [mBar] 300 max
Pump cap. [L/min] 7.5 max
Opening [mm] 300 x 320 (round edge)
Depth bath [mm] 150 (140 effective)
Length [mm] 555
Completely stainless steel
Width [mm] 375
Compact, small footprint
Height [mm] 425
Auto tuning, high precision
Fluid level detection Weight [Kg] 13

Easy to operate CE Confirms to CE regulations


TC20B

General Safety
The TC20B can be used as a circulator or bath. The The bath conforms to CE regulation. It further is
bath fluid is heated via a microprocessor controlled equipped with multiple safety features in order to
heating element. When working below ambient guarantee safe around the clock operation. A
temperature cooling fluid can be circulated through the mechanical safety thermostat protects the bath against
cooling coil to lower the bath temperature. high temperatures and will automatically reset itself
after switching the bath off and on again. The safety
The bath is equipped with fluid level detection. The thermostat can be set between ambient and +300°C. In
microprocessor controlled heating with auto tune case of over temperature electronics and motor are
function guarantees precise temperature control. switched off. The pump motor is protected against
mechanical overload via a resettable fuse.
Span
The bath temperature can be set to a maximum of Temperature readout
200C (392F) in steps of 0.1 degree. The overall Standard available in C, on request in F.
system accuracy is  0.02C.
Pump
Accessories A pump circulates the bath content or provides external
Top lid is included. circulation. The pump pressure is 300 mBar and the
maximum flow is 7.5 litres per minute with no counter
Options pressure.
P/N 03T0031 Levelling platform
P/N 02T3025 RS232 communication

Free TamCom® program, please see


www.tamson.com