Quick Reference Guide SLA-UNI H-Type

H-type: For use with Haag-Streit type slit lamps User Protection Filter Adjustment
Mount the user protection filter back to
User Protection Filter Adjustment the SLA main arm and adjust it so it will
not touch the front lens of the slit lamp
Untighten the two mounting screws of and stand free of the moving deflection
the user protection filter holding arm to mirror of the slit lamp illumination.
the SLA main arm and remove the user Untighten the holding screws of the filter
protection filter device. frame and andjust the filter in its height
until it is in the centre of the observation
path of the slit lamp. Affix the holding
screws.

Adapter Installation
Place the adapter on the tonometer pin. Spot Centering
Make sure not to touch the slit lamp Turn on the slit lamp illumination and set
illumination mirror with the SLA it to a narrow slit. Set the
deflection mirror. micromanipulator to its centre position.
Adjust the height of the SLA with the Untighten the horizontal and the vertical
adjustment screw on top of the adapter spot adjustment screw (see arrows). Turn
plate (see arrow). on the aiming beam. Adjust the mirror
until the aiming beam is in the centre of
Check the height of the SLA by the slit on the slit lamp adjustment tool.
monitoring the illumination beam path Affix the adjustment screws. bad good bad
on the deflection mirror of the SLA – the
deflection mirror must be placed in the
centre of illumination beam path.
bad good bad
Affix the height position of the SLA by Final Adjustment
tightening the locking screw (see arrow). Place the slit lamp with the SLA ahead of
a non reflective wall at a distance of 1m.
Spot Adjustment Set the parameters on power = 50 mW;
Set the SLA diameter adjustment to time = 5000 ms; spot diameter = 50 µm.
bad good bad
50 µm. Connect the SLA to the laser Release pulses towards the wall to check
unit and turn on the aiming beam. the shape of the treatment spot. Make
Untighten the two mounting screws of sure the treatment spot is circular and
the SLA main arm (see arrow). Move the well defined. If not re adjust the Spot
SLA main arm until the spot of the Centering and consult the installation
aiming beam on the slit lamp adjustment manual.
tool is clear and as small as possible.
Tighten the two mounting screws.

7005.F9 Ouick Reference SLA-UNI H-Type Page 1 of 1 30313500-A 11.11.2008