Professional Documents
Culture Documents
ABCT3631 Balance Calibration
ABCT3631 Balance Calibration
ABCT3631 Balance Calibration
Balance Calibration
Electronic balance
– same resolution for the whole range, e.g. capacity 200g with 0.01 mg
resolution
Micro Balances 1 μg
3
Electronic balance
(operating principles & features)
© Shimadzu Corporation
4
Environment for using electronic balance
• keep instrument away from
• keep away from
• controlled fluctuation
• protected from (convection)
• low
• robust
the balance by checking the horizontal level
indicator
for 15 to 30 min before measurement
• keep peripheral
5
• applying (standard weights) to the
instrument under specified conditions
the or
, and
the
to be attributed to the results
6
Different for balance
7
• calibration using the
• the tolerance (uncertainty) of the weight used SHOULD
NOT exceed ( ) to ( ) of the
of the balance
• calibration should be performed in a way similar to the
8
Sensitivity (Scale value)
between the value measured by the balance
and the correct value near the
(sensitivity error)
(commonly sensitivity set to 1)
for balance with internal
calibration mass
for balance with external
calibration mass
• for detail operation, refer to the manufacturer’s manual
• if the sensitivity is adjusted before calibration, note this
fact in the calibration results as a condition of calibration
9
• measurement precision under a set of repeatability
conditions of measurement
• the purpose of repeatability test is to determine the
to generate for
Measurement Measurement
(g) error (g) (g) error (g)
= Balance reading
–
11
© Shimadzu Corporation
12
• the (or corner load test) is carried
out to find the in the displayed weighing as
the object is moved to various positions on the
weighing pan
• use weight with of around 50%
at extreme positions to compare with centre
2 5 2 Center 0.01 g
Left
5 1 3 1 Rear
Right
4 3 Front
4
13
Reading entre ight ront eft ack
sequence
14
• The hysteresis measurements are performed to determine the
of the same load
• Performed by using a with 50% to place
in the center of the weighing pan
• After taking the first reading, the weight is lifted for a short
period of time and place back to take a second reading
• The hysteresis is the
15
MPE for 200 g Class E2 weight
= ± 0.3 mg
16
EU Guideline on
(electronic – digital)
General recommendations
17
Qualification of balances (electronic – digital)
General recommendations
(test loads) for qualification of
balances
– The class of reference weights used in the procedure for
calibration of balances shall be chosen depending on the
( ) of the balance
18
Qualification of balances (electronic – digital)
General recommendations
(test loads) for qualification of
balances (cont’d)
– MPE of the balance is defined by the laboratory
• on the basis of the and
• according to the uncertainty of by the
laboratory
– the MPE of the standard weights used for calibration SHALL
BE compatible with the of the balance
19
Qualification of balances (electronic – digital)
– If selection of the weights for calibration of the balance is
based on of the mass of the
used for calibration, then the of the weight
SHALL NOT be greater than the MPE of the balance divided
by 3:
MPE Weight ≤ ⅓ MPE balance
Nominal value Class E₁ Class E₂ Class F₁ Class F₂
500 g 0.25 0.8 2.5 8.0
0.10 1.0 3.0
100 g 0.05 0.5 1.6
Maximum permissible errors for weights (±δm in mg)
21
Level III Periodic and motivated instrument
checks for balances
• Typical requirements for qualification of balances
(periodic and motivated instrument checks)
– requirements and corresponding typical acceptance
criteria for Level III qualification of balances
22
Level III Periodic and motivated instrument
checks for balances
23
Level III Periodic and motivated instrument
checks for balances
– assessed by
(a) determination of repeatability error, i.e. difference between
the minimum and maximum values obtained from the values
of repeated weighing of A WEIGHT 6 times in a range between
50% - 100% of maximum capacity of the balance; or
(b) standard deviation 𝑠𝑠 , calculated from the above 6 repeated
measurements
𝑛𝑛
𝑠𝑠 = 1
𝑛𝑛 −1
�(𝐼𝐼𝑖𝑖 − 𝐼𝐼)̅ 2
𝑖𝑖=1
25
Level III Periodic and motivated instrument
checks for balances
26
Level III Periodic and motivated instrument
checks for balances
27
HOKLAS Calibration Requirements
Type of equipment Recommended Recommended
calibration Interval calibration method
and every The Calibration of
Weights and Balances
by NMIA
*
* using
a known mass
Items marked * are (those minimum) tests that are to be carried out by
laboratory staff if
• it is suitably equipped with calibrated reference equipment, and
• a suitably controlled environment, and
• staff are confirmed competent to perform the calibrations
28
HOKLAS Calibration Requirements
Type of equipment Recommended Recommended
calibration calibration
Interval method
Weights
of initial; ASTME617
, (i.e. consist of a OIML R111
single piece of material), subsequent
or alloy
Working weights, # ASTME617
or alloy OIML R111
Working weights, stainless # ASTME617
steel or nickel chromium alloy OIML R111
30
Thank you
31