T-F-25. Corrective-Preventive Action Format For PT
T-F-25. Corrective-Preventive Action Format For PT
Laboratory FL/T/GEN/F-25
L-17/3, DLF Phase-II, M. G. Road,
Effective From: Jan 2016 Revision Date: 1st Feb 2023
Gurgaon-122002, Haryana
Issue No.: 01 Revision No.: 01
Root Cause Analysis: Cs-137 has gamma energy of 662 keV and the PT item has Cs-134 as well as Sb-125
radionuclides. These two radionuclides also have gamma energy of 600 and 620 keV respectively which will interfere
with the Cs-137 energy level. Our Gamma Spectrometer is based on NaI(Tl) Detector can’t resolve all three elements
peaks at once differently. Therefore Cs-137 has higher value then target value considering the acceptance of Cs-134
and Co-60. The performance can be considered as satisfactory.
Preventive Action Taken: We have participated in MSME ILC and scored a Z Score of -0.88