SAMPLE HEATER / COOLER

TEMPERATURE ACCURACY REPORT


A C Q U I S I TI O N I N F O R M A TI O N

System Name: e2695_2475 MSG Sample Set Id: 8399


Acquired By: Melissa Result Set Id: 8417
Project Path: RC2_SQT _e2695_2475 final

The Sample Heater / Cooler Temperature Accuracy test is a manual test that verifies the internal electronics which
drive the temperature inside the Sample Heater / Cooler to the set point defined by the instrument control software.
This verifies functionality of the Sample Heater / Cooler electronics, internal probe and control software. The
temperature is measured near to the internal temperature probe.

The test requires a calibrated temperature measuring device that is accurate to + or - 1.0°C. Two temperature
set-points per sensor zone are tested.

The Sample Heater / Cooler Temperature Accuracy test passes only if the recorded absolute temperature
inaccuracy is + or - 4.0°C from the set point.

Sample Heater / Cooler Temperature Accuracy Results


Absolute
Temp Set Temp Temp
Temp
Serial Number Sample Name Accuracy Temp Observed Accuracy Result Id
Inaccuracy
Spec (ºC) (ºC) (ºC) Pass/Fail
(ºC)
1 D16SM4475A SHC Temperature 1 + or - 4.0 4.0 3.0 1.0 PASS 8425
2 D16SM4475A SHC Temperature 2 + or - 4.0 40.0 41.0 1.0 PASS 8426

Waters Project Name and Version: SQT_E_2695_2475 v4.00 © 2007 - 2017 Waters Corporation. All rights reserved.
Date: Monday, February 20, 2017