AOAC Official Method 984.06 mark, and shake. Do not use standard solution after 24 h. Determine Dalapon (Magnesium and/or Sodium Salt) percent H2O in dalapon acid reference standard by Karl Fisher in Pesticide Formulations titration. Labeled purity of reference standard is on anhydrous basis Liquid Chromatographic Method and must be appropriately reduced according to water content. First Action 1984 Final Action 1987 D. Extraction of Analyte A. Principle Accurately weigh test portion containing ca 130 mg dalapon acid Dalapon is separated from related compounds on reverse phase equivalent into 50 mL beaker. Quantitatively transfer to 50 mL LC col umn, using paired-ion eluant, and measured by UV volumetric flask with H2O. Fill to mark with H2O and shake. Do not spectrophotometry. use test solution after 24 h. Dalapon acid and salts are hygroscopic; protect from moisture in air by storing in well-sealed bottle. Prepare B. Apparatus test and standard solutions in duplicate. (Note: Do not add mineral (a) Liquid chromatograph.—With 214 nm UV detector, 20 mL acids in extraction of analyte.) in jec tor and chart re corder. Com puting in te gra tor op tional. Operating conditions: temperature, 20°–30°C; eluant flow rate, E. Determination 160 mL/h; detector sensitivity, 0.25 AUFS. Inject standard solution, test solution, and then standard solution. (b) Liquid chromatographic columns.—50 ´ 4.6 (id) mm Co:Pell Calculate results as ODS pellicular guard column (No. 6561-404, Whatman, Inc.); 100 ´ 8 (id) mm C18, 10 mm radial compression main column with RCM-100 column holder (Waters Associates, or equivalent [such as Dalapon acid equivalent, % = (R/R¢) ´ (W¢/W ´ P) Whatman Partisil 10 ODS-3]). Columns slowly degrade in use. Replace when 30% loss of retention occurs and repack first 5 mm of where R and R¢ = average peak height or peak area of dalapon peak for guard column. Radial compression column recommended gives lower test and standard solutions, respectively; W and W¢ = mg test portion back-pressure and longer column life than standard 316 stainless steel and standard, respectively; and P = percent purity of standard. columns. (Note: Initial system stability may be poor. Before injecting test C. Reagents solution, repeat injection of standard solution to confirm system (a) Eluant.—Dilute mixture of 200 mL CH3CN (UV grade), stability. Periodically confirm linearity by analyzing standards 1.6 mL n-octyl amine, and 2.4 g (NH4)2HPO4 to 1 L with H2O containing 100, 130, and 160 mg 99+% dalapon acid/50 mL H2O. (0.4 mm filtered). Adjust pH of eluant to 7.0 with H3PO4. Eluant may Flush injection loop with ³1 mL test or standard solution before be recycled if reservoir is magnetically stirred and no more than operating injection valve.) 200 injections are made per L eluant. References: JAOAC 66, 1390(1983); 70, 265(1987). (b) Dalapon standard solution.—130 mg/50 mL. Accurately weigh ca 130 mg dalapon acid reference standard (99 +% isomer CAS 75-99-0 (dalapon)