Download as pdf or txt
Download as pdf or txt
You are on page 1of 8

PROCEDURE

57 Twelve-Lead Electrocardiogram
Shu-Fen Wung
PURPOSE: A 12-lead electrocardiogram (ECG) provides information about the
electrical activity of the heart from 12 different views or leads. The ECG is the most
commonly performed cardiovascular diagnostic procedure.1 Common uses of a
12-lead ECG include diagnosis of acute coronary syndromes, identification of
arrhythmias, and determination of the effects of medications, electrolytes, or
structural abnormalities on the electrical system of the heart.

PREREQUISITE NURSING • Nurses should be able to operate the 12-lead ECG machine.
KNOWLEDGE Calibration of 1 mV equals 10 mm and paper speed of
25 mm/sec are standards used in clinical practice. For
• Understanding of the anatomy and physiology of the car- ST-segment analysis, filter settings of 0.05 to 100 Hz are
diovascular system, principles of electrophysiology, ECG recommended by the American Heart Association.5 Any
lead placement, basic rhythm interpretation, and electrical variation used for particular clinical purposes should be
safety is necessary. noted on the tracing. Specific information regarding con-
• Advanced cardiac life support (ACLS) knowledge and figuring the ECG machine, troubleshooting, and safety
skills are needed. features is available from the manufacturer and should be
• A 12-lead ECG provides different views or leads of the read before use of the equipment.
electrical activity of the heart. The 12 standard leads • Nurses should be able to interpret recorded ECGs for the
include six limb leads (I, II, III, augmented vector right presence or absence of myocardial ischemia/infarction and
[aVR], augmented vector foot [aVF], and augmented arrhythmias so that patients can be treated appropriately.
vector left [aVL]), and six chest leads (V1 to V6). • Advances in technology have allowed for online or wire-
• The limb leads view the heart from the frontal or vertical less transmission, networking capabilities, and computer-
plane (Fig. 57-1), and the chest leads view the heart from ized interpretation of the 12-lead ECG (Fig. 57-3). The
the horizontal plane (Fig. 57-2). 12-lead ECG cable is attached to a processing device that
• The basic ECG waveforms are labeled with P, Q, R, S, digitizes the 12-lead ECG recording and transfers the
and T waves, which represent electrical activity within the information to the wireless device, which transmits the
heart. information to the medical record. This increases access
• Accuracy in identification of anatomical landmarks for to the 12-lead ECG for review and can assist with rapid
location of electrode sites and knowledge of the impor- interpretation and treatment of the patient.
tance of accurate electrode placement are needed. Accu-
rate ECG interpretation is possible only when the recording EQUIPMENT
electrodes are placed in the proper positions. Slight altera-
tions of the electrode positions may distort significantly • 12-lead ECG machine with patient cable and lead wires
the appearance of the ECG waveforms and can lead to • ECG electrodes
misdiagnosis.2 Reliable comparison of serial (more than Additional equipment, to have available as needed, includes
two ECGs recorded at different times) ECG recordings the following:
relies on accurate and consistent electrode placement. An • Gauze pads or terrycloth washcloth
indelible marker is recommended for clear identification • Cleansing pads or nonemollient soap and water
of the electrode locations to ensure that the same electrode • Skin preparation solution (e.g., skin barrier wipe or tinc-
locations are selected when serial ECGs are recorded. ture of benzoin)
• Nurses should be aware of body-positional changes that • Indelible marker
can alter ECG recordings. Serial ECGs should be recorded • Clippers or scissors to clip hair from the patient’s chest if
with the patient in a supine position to ensure that all needed
recordings are done in a consistent manner. Side-lying
positions and elevation of the torso may change the posi- PATIENT AND FAMILY EDUCATION
tion of the heart within the chest and can change the
waveforms on the ECG recording.3,4 If a position other • Describe the procedure and reasons for obtaining the
than supine is clinically necessary, notation of the altered 12-lead electrocardiogram. Reassure the patient that the
position should be made on the tracing. procedure is painless. Rationale: This communication

494
57 Twelve-Lead Electrocardiogram 495

Figure 57-1 Vertical plane leads: I, II, III, aVR, aVL, aVF.
Figure 57-3 Example of a wireless electrocardiograph (ECG)
device. The 12-lead cable is attached to a processing device that can
then be transmitted to the medical record.

recordings can help clinicians determine whether a change


is acute or chronic.
• Assess for the presence of anginal symptoms, such as
chest pain, pressure, tightness, heaviness, fullness, or
squeezing sensation; radiated pain; or shortness of breath,
nausea, and extreme fatigue. Rationale: This evaluation
correlates ECG changes with patient symptoms.
• Assess the patient’s history of cardiac conditions and
review medication history. Rationale: Knowledge about
Figure 57-2 Horizontal plane leads: V1 to V6. the patient’s cardiac history and medications can help in
interpretation of ECG recordings.

clarifies information, reduces anxiety, and gains coopera- Patient Preparation


tion from the patient. • Verify that the patient is the correct patient using two
• Explain the patient’s role in assisting with the ECG record- identifiers. Rationale: Before performing a procedure, the
ing and emphasize actions that improve the quality of the nurse should ensure the correct identification of the patient
ECG tracing, such as relaxing, avoiding conversation and for the intended intervention.
body movement, and breathing normally. Rationale: This • Ensure that the patient and family understand prepro-
explanation ensures the patient’s cooperation to improve cedural teaching. Answer questions as they arise, and
the quality of the tracing and avoids unnecessary repeating reinforce information as needed. Rationale: This com-
of the ECG because of muscle artifact. munication evaluates and reinforces the understanding of
previously taught information.
• Assist the patient to the supine position and expose the
PATIENT ASSESSMENT AND patient’s torso while maintaining the patient’s modesty.
PREPARATION Rationale: This position enables the recording of a
standard 12-lead ECG and allows comparison of serial
Patient Assessment ECGs and comparison with standard waveforms. Body-
• Interpret previously recorded ECGs. Rationale: Each positional changes, such as elevation and rotation, can
patient has an individual baseline ECG. Previous ECG change recorded amplitudes and axes.
496 Unit II Cardiovascular System

Procedure for 12-Lead Electrocardiogram


Steps Rationale Special Considerations
1. HH
2. Check cables and lead wires for Detects faulty equipment. If the equipment is damaged, obtain
fraying or broken wires. alternative equipment and notify
a biomedical engineer for repair.
3. Check the lead wires for accurate Obtains accurate ECG recordings
labels. and proper placement of leads.
4. Plug the ECG machine into a Maintains electrical safety. Follow manufacturer’s
grounded alternating current (AC) recommendations and institutional
wall outlet or ensure functioning if protocol on electrical safety per
battery operated. the biomedical department.
5. Turn the ECG machine on and Equipment may require self-test Manufacturers provide a calibration
program the ECG machine: paper and warm-up time. Verify check in the machine to identify
speed, 25 mm/sec; calibration, equipment settings in the sensitivity setting. Most
10 mm/mV; filter settings, 0.05– accordance with clinical practice machines have automatic settings.
100 Hz. (Level E*) and recommendations by the
American Heart Association
(AHA).5 Multichannel machines
may require input of information
(e.g., data about the patient) to
store the ECG appropriately.
6. PE
7. Place the patient in a supine Provides adequate support for ECGs should be recorded in the
position. (Level B*) limbs so that muscle activity is same body position to ensure
minimal. Body-position changes ECG changes are not caused by a
can cause ST-segment deviation change in body position. If
and QRS waveform another position is clinically
alteration.3,4,6 necessary, note the altered
position on the ECG recording.
8. Expose only the necessary body Provides privacy and warmth, Ensuring privacy may reduce
parts of the patient (legs, arms, and which reduces shivering. anxiety.
chest) for electrode placement. Shivering may interfere with the
quality of recording.
9. Identify skin electrode locations. Ensures the accuracy of the lead
placement.
A. Limb leads (Fig. 57-4).

Figure 57-4 Limb lead placement in 12-lead ECG.

*Level B: Well-designed, controlled studies with results that consistently support a specific action, intervention, or treatment.
*Level E: Multiple case reports, theory-based evidence from expert opinions, or peer-reviewed professional orga­nizational standards without clinical studies to
support recommendations.
57 Twelve-Lead Electrocardiogram 497

Procedure for 12-Lead Electrocardiogram—Continued


Steps Rationale Special Considerations
• Right arm (RA): inside right Accurate electrode placement is Limb leads should be placed in
forearm essential for obtaining valid and fleshy areas; bony prominences
• Left arm (LA): inside left reliable data for ECG should be avoided. The limb
forearm recordings. The RL electrode is leads need to be placed
• Right leg (RL): anywhere on a ground electrode that does not equidistant from the heart and
the body; by convention, contribute to the ECG tracings. should be positioned in
usually on the right ankle or approximately the same place on
inner aspect of the calf each limb.
• Left leg (LL): left ankle or
inner aspect of the calf
B. Precordial Leads (Fig. 57-5)
• Identify the sternal notch. The angle of Louis assists with Variations in precordial lead
Slide fingers down the center identifying the second rib for placement of as little as 2 cm can
of the sternum to the obvious correct placement of precordial result in important diagnostic
bony prominence, angle of leads in the appropriate ICS. errors, particularly in anteroseptal
Louis, which identifies the Slight alterations in the position of infarction and ventricular
second rib and provides a any of the precordial leads may hypertrophy.6
landmark for noting the second alter the ECG significantly and If precordial leads cannot be
intercostal space (ICS). can affect diagnosis and accurately placed because of
❖ V1: fourth ICS at right treatment.2,7,8 chest wounds, placement of
sternal border defibrillator pads, or other
❖ V2: fourth ICS at left sternal reasons, the alternative site
border should be clearly documented on
❖ V4: fifth ICS at the ECG.9
midclavicular line It is recommended that electrodes
❖ V3: halfway between V2 and be placed under the breast in
V4 women until additional studies
❖ V5: horizontal level to V4 at are available.5
the anterior axillary line
❖ V6: horizontal level to V4 at
the midaxillary line

Figure 57-5 Precordial or chest lead placement.

10. Clean and slightly abrade the skin Removes dead skin cells, Failure to properly prepare the skin
where the electrodes will be applied. promoting impulse may cause artifacts and interfere
A. Wash the skin with soap and transmission.10-12 Moist skin is with interpretation.
water, if needed. not conducive to electrode
B. Abrade the skin with a gauze adherence.
pad or abrader.
C. Ensure that the skin is dry before
skin electrodes are applied.
D. Clipping of chest hair may be
necessary to ensure that adequate
skin contact with the skin
electrodes is made. (Level C*)

*Level C: Qualitative studies, descriptive or correlational studies, integrative reviews, systematic reviews, or randomized controlled trials with inconsistent results.
Procedure continues on following page
498 Unit II Cardiovascular System

Procedure for 12-Lead Electrocardiogram—Continued


Steps Rationale Special Considerations
11. Identify the electrode sites and mark Minimizes ECG changes caused After accurate identification of the
them with an indelible marker. by altered electrode placement.2,6 locations, an indelible marker
should be used to mark the
electrode sites if serial ECGs are
anticipated.
12. For pregelled electrodes, remove the Allows for appropriate conduction Gel must be moist. If pregelled
backing and test for moistness. For of impulses. electrodes are not moist or
adhesive electrodes, remove the adhesive electrodes are not sticky,
backing and check each adhesive replace the electrodes.
pad, as each should be sticky or
moist.
13. Apply the electrodes securely and Electrodes must be secure to If limb plate electrodes are used, do
place the electrodes on the marked prevent external influences from not overtighten to minimize
locations. affecting the ECG. Secure the discomfort.
electrodes to obtain quality ECG
recordings.
14. Fasten the lead wires to the limb Provides for correct lead-to-limb
electrodes, avoiding bending or connection.
strain on the wires, and use the
correct lead-to-electrode connection.
15. Identify the multiple-channel Multiple-channel machines run
machine recording setting several leads simultaneously and
(Fig. 57-6). can be set to run leads in
different configurations.

Figure 57-6 Multiple-channel ECG machine. (Courtesy Philips


Medical Systems, Andover, MA.)
57 Twelve-Lead Electrocardiogram 499

Procedure for 12-Lead Electrocardiogram—Continued


Steps Rationale Special Considerations
16. Obtain a 12-lead ECG recording. Three to six seconds are all that is A multiple-channel machine runs
Most systems record each lead for needed for a permanent record; the limb and chest leads
3–6 seconds and automatically mark a longer strip may be obtained if simultaneously.
the correct lead. a rhythm strip is needed.
17. Examine the quality of the 12-lead While the patient is still connected Reviews the normal conduction
ECG tracing. to the machine, the nurse should sequence and identifies
examine the ECG to see whether abnormalities that may necessitate
any leads need to be repeated. further evaluation or treatment.
18. Disconnect the equipment; clean the Increases patient comfort. Some pregelled electrodes can be
gel off the patient (if necessary), Reduces the transmission of left in place for repeat ECGs.
remove PE , discard used supplies, microorganisms; Standard Follow the manufacturer’s
and prepare the equipment for Precautions. directions and hospital policy for
future use. electrode use and removal in
these cases.
19. HH

Expected Outcomes Unexpected Outcomes


• A clear and accurate 12-lead ECG recording that • Altered skin integrity
allows clinicians to diagnose dysrhythmias and • Inaccurate lead placement or connection (Fig. 57-8)
ischemia (Fig. 57-7) • AC interference, also called 60-cycle interference
• Prompt identification of abnormalities (see Fig. 54-9)
• Wandering baseline (see Fig. 54-10)
• Artifact or waveform interference (see Fig. 54-11)
Procedure continues on following page

Figure 57-7 Clear 12-lead ECG recording.


500 Unit II Cardiovascular System

B
Figure 57-8 Limb lead reversal on 12-lead electrocardiograph (ECG) in lead I. A, Correct place-
ment. B, Incorrect placement.

Patient Monitoring and Care


Steps Rationale Reportable Conditions
These conditions should be reported
if they persist despite nursing
interventions.
1. Obtain a 12-lead ECG as Provides determination of • Angina
prescribed and as needed (e.g., for myocardial ischemia, injury, • Arrhythmias
angina or arrhythmias). and infarction. Aids in • Abnormal 12-lead ECG
diagnosis of arrhythmias.
2. Compare the 12-lead ECG with the Determines normal and • Any abnormal changes in the
previous 12-lead ECGs. abnormal findings. 12-lead ECG
3. Follow institutional standards for Promotes comfort. • Continued pain despite pain
assessing pain. Administer interventions
analgesia and nitrates as prescribed.

Documentation
Documentation should include the following:
• Patient and family education • Pain assessment, interventions, and patient response
• The fact that a 12-lead ECG was obtained to interventions
• The reason for the 12-lead ECG • Follow-up to the 12-lead ECG as indicated
• Any altered lead placement and reason • Unexpected outcomes
• Symptoms that the patient experienced (e.g., chest • Additional interventions
pain, syncope, dizziness, or palpitations)

References and Additional Readings


For a complete list of references and additional readings for
this procedure, scan this QR code with any freely available
smartphone code reader app, or visit
http://booksite.elsevier.com/9780323376624.
57 Twelve-Lead Electrocardiogram 500.e1

References Hancock EW, et al: AHA/ACCF/HRS recommendations for


1. Kligfield P: The centennial of the Einthoven the standardization and interpretation of the
electrocardiogram. J Electrocardiol 35(Suppl):123–129, electrocardiogram: Part V: Electrocardiogram changes
2002. associated with cardiac chamber hypertrophy: A scientific
2. Drew BJ: Pseudo myocardial injury patterns because of statement from the American Heart Association
nonstandard electrocardiogram electrode placement. Electrocardiography and Arrhythmias Committee, Council
J Electrocardiol 41(3):202–204, 2008. on Clinical Cardiology; the American College of
3. Adams MG, Drew BJ: Body position effects on the ECG: Cardiology Foundation; and the Heart Rhythm Society.
Implication for ischemia monitoring. J Electrocardiol Endorsed by the International Society for Computerized
30(4):285–291, 1997. Electrocardiology. J Am Coll Cardiol 53(11):992–1002,
4. Nelwan SP, et al: Correction of ECG variations caused by 2009.
body position changes and electrode placement during Macfarlane PW: Age, sex, and the ST amplitude in health and
ST-T monitoring. J Electrocardiol 34(Suppl):213–216, disease. J Electrocardiol 34(Suppl):235–241, 2001.
2001. Mason JW, et al: Recommendations for the standardization
5. Kligfield P, et al: Recommendations for the and interpretation of the electrocardiogram: Part II:
standardization and interpretation of the Electrocardiography diagnostic statement list a scientific
electrocardiogram: Part I: the electrocardiogram and its statement from the American Heart Association
technology: A scientific statement from the American Electrocardiography and Arrhythmias Committee, Council
Heart Association Electrocardiography and Arrhythmias on Clinical Cardiology; the American College of
Committee, Council on Clinical Cardiology; the American Cardiology Foundation; and the Heart Rhythm Society
College of Cardiology Foundation; and the Heart Rhythm Endorsed by the International Society for Computerized
Society: endorsed by the International Society for Electrocardiology. J Am Coll Cardiol 49(10):1128–1135,
Computerized Electrocardiology. Circulation 2007.
115(10):1306–1324, 2007. Rautaharju PM, et al: AHA/ACCF/HRS recommendations for
6. Baevsky RH, et al: Supine vs semirecumbent and upright the standardization and interpretation of the
12-lead electrocardiogram: Does change in body position electrocardiogram: Part IV: The ST segment, T and U
alter the electrocardiographic interpretation for ischemia? waves, and the QT interval: A scientific statement from the
Am J Emerg Med 25(7):753–756, 2007. American Heart Association Electrocardiography and
7. Drew BJ: Pitfalls and artifacts in electrocardiography. Arrhythmias Committee, Council on Clinical Cardiology;
Cardiol Clin 24(3):309–315, 2006. the American College of Cardiology Foundation; and the
8. Wung SF: Computer-assisted continuous ST-segment Heart Rhythm Society. Endorsed by the International
analysis for clinical research: Methodological issues. Society for Computerized Electrocardiology. J Am Coll
Biol Res Nurs 3(2):65–77, 2001. Cardiol 53(11):982–991, 2009.
9. Drew BJ, et al: Practice standards for Sole M, Klein D, Moseley M: Introduction to critical care
electrocardiographic monitoring in hospital settings: nursing, ed 6, St. Louis, 2013, Elsevier Saunders.
An American Heart Association scientific statement Surawicz B, et al: AHA/ACCF/HRS recommendations for the
from the Councils on Cardiovascular Nursing, Clinical standardization and interpretation of the electrocardiogram:
Cardiology, and Cardiovascular Disease in the Young: Part III: Intraventricular conduction disturbances: A
endorsed by the International Society of Computerized scientific statement from the American Heart Association
Electrocardiology and the American Association of Electrocardiography and Arrhythmias Committee, Council
Critical-Care Nurses. Circulation 110(17):2721–2746, on Clinical Cardiology; the American College of
2004. Cardiology Foundation; and the Heart Rhythm Society.
10. Clochesy JM, Cifani L, Howe K: Electrode site Endorsed by the International Society for Computerized
preparation techniques: A follow-up study. Heart Lung Electrocardiology. J Am Coll Cardiol 53(11):976–981,
20(1):27–30, 1991. 2009.
11. Tam HW, Webster JG: Minimizing electrode motion Wagner GS, et al: AHA/ACCF/HRS recommendations for the
artifact by skin abrasion. IEEE Trans Biomed Eng standardization and interpretation of the electrocardiogram:
24(2):134–139, 1977. Part VI: Acute ischemia/infarction: A scientific statement
12. Lawler JC, Davis MJ, Griffith EC: Electrical from the American Heart Association Electrocardiography
characteristics of the skin. The impedance of the surface and Arrhythmias Committee, Council on Clinical
sheath and deep tissues. J Invest Dermatol 34:301–308, Cardiology; the American College of Cardiology
1960. Foundation; and the Heart Rhythm Society: endorsed by
the International Society for Computerized
Additional Readings Electrocardiology. Circulation 119(10):e262–e270, 2009.
Burns S: AACN essentials of critical care nursing, New York,
2014, McGraw-Hill Companies.

You might also like