LOAD RESISTANCE
Variable: 0 Ω, 10 Ω, 20 Ω, 25 Ω to 2500 Ω (by 25 Ω), 2500 Ω to 5200 Ω (by 100 Ω)
DC Accuracy: ± 2.5 %
POWER
Ranges: 0.0 W to 99.9 W, 100 W to 500 W
Accuracy: < 50 W: +/- (6.25 % + 1 W) , ≥ 50 W: +/- 6.25 %
CURRENT
RMS: 0 mA to 5,500 mA
Accuracy: < 50 mA: +/- (3.5 % + 1 mA), ≥ 50 mA: +/- (2.5 % + 1 mA)
VOLTAGE
Peak: 10 kV Peak to Peak
Accuracy: ±(10 % of reading + 50 V)
Crest factor:1.4 to 16.0 Defined as the ratio of Peak voltage to RMS voltage (Vpk /Vrms), using the larger of the 2 peaks (positive or
negative)
HF LEAKAGE CURRENT
Fixed load: 200 Ω
V Accuracy: ± 2.5 %
Power rating: 400 W
Additional fixed load: 200 Ω
Current, RMS: 0 mA to 5500 mA
Accuracy: < 50 mA: +/- (3.5 % + 1 mA), ≥ 50 mA: +/- (2.5 % + 1 mA)
COMMUNICATIONS
USB device port: Micro B connector, full speed
Accuracy: ± 5 dB(A)
Display Resolution: 0.1 dB(A)
IEC-61672-1 Class 2 from 31.5Hz to 8kHz
Surface Temperature
Range: -5 °C to 60 °C
Accuracy: ± 0.5 °C
Display resolution: 0.05 °C
SAFETY
IEC 61010-1 Overvoltage category none, pollution degree 2
IEC 61326-1 Basic
WIRELESS RADIO
Frequency range: 2412 MHz to 2483 MHz
Output power: 10 mW
INTERFACE
Push navigation buttons: Power on/off, test, select, back, and arrow keys
User Preferences: Adjust backlight, display brightness, and set clock
TECHNICAL SPECIFICATIONS
Large LCD
Accurate to ±5% of full scale
Spectral range: 429 - 473 nm (max. 97% response at 453 nm)
Measurement range: 0-1999 µW/cm
Resolution: 1 µW/cm
Probe: Lens matches the cosine receiving function of human skin
Power: 9 V battery; an arrow appears on display for battery replacement; 150 continuous hours operation
Weight: 8.8 oz (250 g) ± 10%
Dimensions: (14.5 cm x 8.0 cm x 3.8 cm) ±5cm
STANDARD ACCESSORIES
Hard case for the Phototherapy Radiometer
Probe
9V Rechargeable Battery with Charger
SPECIAL CONDITIONS
Application Software
- Must provide fully activated Centralize Maintenance Management Software (CMMS) applicable with the use of test equipment.
Calibration
-Must provide calibration certificate traceable to the International System of Units (SI) through the appropriate National Metrology
Institutes such as NIST or through intrinsic standards.
- Annual calibration free of charge within the warranty period
Warranty
- Three (3) years warranty on services, parts, and accessories
- Able to provide after sales support or response within 24 hours
- Must provide loaner unit in case of breakdown or need to pull-out for repair or calibration.
Training
-Must conduct technical and calibration training with certificate for all Medical Equipment Technician (MET) in accordance with ISO 17025
Temperature Sensor: 100Ω Platinum RTD
Temperature Range (Body): -40 to 140ºC (-40 to 284ºF)
Temperature Measurement Range (Probe): -200 to 260ºC (-328 to 500ºF)
Temperature Resolution: 0.01ºC (0.02ºF)
Calibrated Accuracy: ±0.1ºC (±0.18ºF) [20 to 140ºC (68 to 284ºF)]
Start Modes:
• Software programmable immediate start
• Delay start up to eighteen months in advance
Stop Modes:
• Manual through Software
• Time (specific date and time)
Real Time Recording: May be used with PC to monitor and record data in real time
Memory: 32,700 readings
Reading Rate: One second up to once every 24 hours
Battery Type: 3.6V high-temperature lithium battery (included); user-replaceable
Battery Life: 1 year typical [1 minute reading rate at 25ºC (77ºF)]
Calibration: Digital calibration through software
Calibration Date: Automatically recorded within device
Data Format: Date and time stamped ºC, ºF, K, ºR
Time Accuracy: ±1 minute/month at 20 to 30ºC (68 to 86ºF) (RS232 cable not in use)
Computer Interface: USB docking station, 125,000 baud
Software: Windows XP SP3/Vista/7 and 8 (32- and 64-bit)
Material: 316 stainless steel
Connection Type: USB to PC
STANDARD ACCESSORIES
Hard case for Temperature Data Logger
Docking station
Connecting cable
Tablet PC
SPECIAL CONDITIONS
Application Software
- Must provide fully activated Centralize Maintenance Management Software (CMMS) applicable with the use of test equipment.
Calibration
-Must provide calibration certificate traceable to the International System of Units (SI) through the appropriate National Metrology
Institutes such as NIST or through intrinsic standards.
- Annual calibration free of charge within the warranty period
Warranty
- Two (2) years warranty on services, parts, and accessories
- Able to provide after sales support or response within 24 hours
10,10, 20,20
Unit of Measure Total Quantity Price Catalogue Total Amount for the Year
for the Year
Piece 20.00
250.00 5,000.00
Piece 10.00
350.00 3,500.00
Piece 20.00
700.00 14,000.00
Piece 10.00
650.00 6,500.00
Piece 50.00 500.00 25,000.00
Piece 50.00 100.00 5,000.00
Piece 20.00 150.00 3,000.00
Piece 20.00 250.00 5,000.00
Piece 10.00 350.00 3,500.00
Piece 10.00 15.00 150.00
Piece 10.00 400.00 4,000.00
Piece 10.00 400.00 4,000.00
Piece 5.00 2,500.00 12,500.00
Piece 30.00 50.00 1,500.00
Piece 5.00 650.00 3,250.00
Piece 20.00 45.00 900.00
Piece 40.00 5.00 200.00
Box 10.00 500.00 5,000.00
Set 5.00 500.00 2,500.00
Piece 5.00 500.00 2,500.00
Piece 10.00 100.00 1,000.00
Piece 2.00 1,500.00 3,000.00
Piece 5.00 500.00 2,500.00
Piece 5.00 500.00 2,500.00
Box 5.00 100.00 500.00
Piece 5.00 500.00 2,500.00
Piece 5.00 500.00 2,500.00
Piece 5.00 500.00 2,500.00
Piece 5.00 250.00 1,250.00
Piece 2.00 700.00 1,400.00
Piece 10.00 600.00 6,000.00
Piece 10.00 150.00 1,500.00
Piece 5.00 300.00 1,500.00
Piece 10.00 300.00 3,000.00
Piece 5.00 300.00 1,500.00
Piece 20.00 30.00 600.00
Piece 5.00 30.00 150.00
Piece 50.00 50.00 2,500.00
Piece 10.00 50.00 500.00
Set 5.00 1,200.00 6,000.00
Set 3.00 500.00 1,500.00
box 2.00 250.00 500.00
roll 1.00 5,000.00 5,000.00
piece 20.00 500.00 10,000.00
piece 50.00 100.00 5,000.00
Box 5.00 850.00 4,250.00
Box 5.00 950.00 4,750.00
roll 5.00 700.00 3,500.00
Bottle 5.00 950.00 4,750.00
Piece 5.00 650.00 3,250.00
Piece 5.00 150.00 750.00