Professional Documents
Culture Documents
Manual CTC - Enus
Manual CTC - Enus
Manual CTC - Enus
CONPROVE
TEST
CENTER
(CTC)
SOFTWARE USE MANUAL
Statement of Responsibility
The information contained in this tutorial is constantly checked. However, differences in the description cannot
be completely excluded; therefore, CONPROVE disclaims any responsibility for errors or omissions contained
in the information transmitted.
Suggestions for improving this material are welcome, simply by contacting the email
suporte@conprove.com.br
The manual contains knowledge obtained from the resources and technical data at the time of writing.
Therefore CONPROVE reserves the right to make changes to this document without prior notice.
This document is intended to be a guide only, the manual of the equipment to be Test set should always be
consulted.
WARNING!
The equipment generates high current and voltage values during its operation.
Improper use of the equipment can result in material and physical damage.
Only persons with adequate qualifications should handle the instrument. It is observed that the user must have
satisfactory training in maintenance procedures, a good knowledge of the equipment to be Test set and also be
aware of safety rules and regulations.
Copyright
Copyright © CONPROVE. All rights reserved. The disclosure, total or partial reproduction of its content is not
authorized, unless they are expressly allowed. Violations are subject to sanctions by law.
March 2022
Version control
Summary
1 Introduction ......................................................................................................................42
1.1 Minimum software requirements ..............................................................................42
2 Hardware and Software – Operating instructions for CE-6003/CE-6006/CE-6707/CE-
6710/CE-7012/CE-7024 .........................................................................................................44
2.1 Important information ...............................................................................................44
2.2 Description of the CE-6003 panel .............................................................................44
2.3 Description of the CE-6006 panel .............................................................................47
2.4 Description of the CE-6707 panel .............................................................................49
2.5 Description of the CE-6710 panel .............................................................................52
2.6 Description of the CE-7012 panel .............................................................................55
2.7 Description of the CE-7024 panel .............................................................................59
3 Software - Instructions on installing the Conprove Test Center (CTC) ..........................63
3.1 Installing CTC applications.......................................................................................64
3.2 Installing PS Simul (demo version) ..........................................................................66
3.3 Installing the Tutorials ..............................................................................................69
3.4 Installing I/O Drivers ................................................................................................71
4 Conprove Test Center (CTC) – Application Manager for CE-6003 / CE-6006 / CE-6707
/ CE-6710 /CE-7012 / CE-7024 – General Notes...................................................................75
4.1 General characteristics ..............................................................................................75
4.2 GENERAL ................................................................................................................76
4.2.1 MANUAL................................................................................................................................... 76
4.2.2 AUXILIARY SOURCE ............................................................................................................. 76
4.2.3 CALIBRAC ................................................................................................................................ 76
4.2.4 TEST PLAN ............................................................................................................................... 76
4.2.5 GERAÇÃO REMOTA ............................................................................................................... 76
4.3 PRIMARY .................................................................................................................76
4.3.1 CT ............................................................................................................................................... 76
4.3.2 VT ............................................................................................................................................... 77
4.3.3 TRANSFORMER ....................................................................................................................... 77
4.3.4 RESISTANCE ............................................................................................................................ 77
4.3.5 PMASTER .................................................................................................................................. 77
4.4 SECONDARY ..........................................................................................................77
4.4.1 DIFERENC................................................................................................................................. 77
4.4.2 DIRECPOT................................................................................................................................. 77
4.4.3 DISTANC ................................................................................................................................... 78
4.4.4 MASTER .................................................................................................................................... 78
4.4.5 PSB-OoS..................................................................................................................................... 78
4.4.6 RAMPA ...................................................................................................................................... 78
4.4.7 RESTHARM .............................................................................................................................. 78
4.4.8 SEQUENC .................................................................................................................................. 79
4.4.9 SYNCHRONISM ....................................................................................................................... 79
4.4.10 SOBRECOR ............................................................................................................................... 79
4.4.11 TRANSIENT .............................................................................................................................. 79
4.4.12 VOLTSPHZ ................................................................................................................................ 79
4.4.13 TRANSDUT ............................................................................................................................... 80
4.4.14 MEDIDOR ................................................................................................................................. 80
4.4.15 POWER QUALITY ................................................................................................................... 80
4.5 OTHER......................................................................................................................80
4.5.1 TRANSIENT VIEW................................................................................................................... 80
4.5.2 VALIDATE PDF REPORTS ..................................................................................................... 80
CONPROVE INDÚSTRIA E COMÉRCIO LTDA 5
CTC SOFTWARE MANUAL
Picture Index
Figure 1 – Hardware – CE-6003 Panel ............................................................................................................... 45
Figure 2 – Hardware – CE-6006 Panel ............................................................................................................... 47
Figure 3 – Hardware – CE-6707 Panel ............................................................................................................... 50
Figure 4 – Hardware – CE-6710 panel ............................................................................................................... 53
Figure 5 – Hardware – CE-7012 panel ............................................................................................................... 56
Figure 6 – Hardware – CE-7024 panel ............................................................................................................... 59
Figure 7 – Installation – Installation Manager .................................................................................................... 63
Figure 8 – Installation – CTC Installation .......................................................................................................... 64
Figure 9 – Installation – CTC license ................................................................................................................. 64
Figure 10 – Installation – CTC destination folder .............................................................................................. 65
Figure 11 – Installation – Confirmation of the beginning of the installation of the CTC ................................... 65
Figure 12 – Installation – CTC successfully installed ........................................................................................ 66
Figure 13 – Installation – PS Simul Installation ................................................................................................. 66
Figure 14 – Installation – PS Simul license ........................................................................................................ 67
Figure 15 – Installation – PS Simul destination folder ....................................................................................... 67
Figure 16 – Installation – Confirmation of the beginning of the installation of the CTC ................................... 68
Figure 17 – Installation – PS Simul successfully installed ................................................................................. 68
Figure 18 – Installation – Installation off Tutorials ............................................................................................ 69
Figure 19 – Installation – Tutorial destination folder ......................................................................................... 69
Figure 20 – Installation – Confirmation of the beginning of the installation of the tutorials ............................. 70
Figure 21 – Installation – Tutorials successfully installed ................................................................................. 70
Figure 22 – Installation – National Instruments installation start screen ............................................................ 71
Figure 23 – Installation – National Instruments destination folder .................................................................... 71
Figure 24 – Installation – Verification if there upgrade to the National Instruments ......................................... 72
Figure 25 – Installation – Message showing that there is no update for National Instruments .......................... 72
Figure 26 – Installation – National Instruments license ..................................................................................... 73
Figure 27 – Installation – National Instruments installation begins ................................................................... 73
Figure 28 – Installation – National Instruments successfully installed .............................................................. 74
Figure 29 – Installation – Message to restart the computer ................................................................................ 74
Figure 30 – CTC Main screen ............................................................................................................................ 75
Figure 31 – Manual – Open Settings when Starting ........................................................................................... 84
Figure 32 – Manual – Main Screen .................................................................................................................... 85
Figure 33 – Manual – Menu – Software Options ............................................................................................... 85
Figure 34 – Manual – General Button Features .................................................................................................. 86
Figure 35 – Manual – Home Menu .................................................................................................................... 87
Figure 36 – Manual – Home Menu – Hardware, Results and Generation .......................................................... 87
Figure 37 – Manual – Home Menu – Options, Report, Units e Layout.............................................................. 88
Figure 38 – Manual – View Menu – Windows and Layout ............................................................................... 89
Figure 39 – Manual – Software Options Menu – General .................................................................................. 90
Figure 40 – Manual – Adjustments button ......................................................................................................... 90
Figure 41 – Manual – Settings screen – General ................................................................................................ 91
Figure 42 – Manual – Pre-Fault – Sub tab NO ................................................................................................... 92
Figure 43 – Manual – Pre-fault tab – Sub tab NO – Description of the Direct Mode ........................................ 92
Figure 44 – Manual – Pre-Fault tab – Resources ............................................................................................... 93
Figure 45 – Manual – Pre-Fault – Resources – Voltages, Currents or Frequency .............................................. 93
Figure 46 – Manual – Pre-Fault tab – Sub tab NO – Description of the Harmonics mode ................................ 94
Figure 47 – Manual – Pre-Fault – Resources ..................................................................................................... 94
Figure 48 – Manual – Pre-Fault tab – Formatting features ................................................................................. 95
Figure 49 – Manual – Pre-Fault tab – Sub tab NO – Description of the Symmetric Components mode ........... 95
Figure 50 – Manual – Pre-Fault – Resources ..................................................................................................... 96
Figure 51 – Manual – Pre-Fault tab – Sub tab NO – Description of the Phase-Phase mode .............................. 96
Figure 52 – Manual – Pre-Fault – Resources ..................................................................................................... 97
Figure 53 – Manual – Pre-Fault tab – Sub tab NO – Description of the Power mode ....................................... 98
Figure 54 – Manual – Pre-Fault – Resources ..................................................................................................... 98
Figure 55 – Manual – Pre-Fault – Sub tab NO – Fault mode description .......................................................... 99
CONPROVE INDÚSTRIA E COMÉRCIO LTDA 20
CTC SOFTWARE MANUAL
Figure 168 – Power Quality – Waveform Tab: Event – Resources – Display ................................................. 180
Figure 169 – Power Quality – Waveform Tab: Detail...................................................................................... 180
Figure 170 – Power Quality – Waveform Tab: Detail – Resources ................................................................. 181
Figure 171 – Power Quality – Waveform Tab: Detail – Resources – Display ................................................. 181
Figure 172 – Power Quality – Phasor Tab........................................................................................................ 182
Figure 173 – Power Quality – Phasors Tab – Resources .................................................................................. 182
Figure 174 – Power Quality – Phasors Tab – Show Columns .......................................................................... 183
Figure 175 – Power Quality – Harmonics Tab ................................................................................................. 183
Figure 176 – Power Quality – Harmonics Tab – Resources ............................................................................. 184
Figure 177 – Power Quality – Main Screen – Test........................................................................................... 184
Figure 178 – Power Quality – Test – Insert/ Edit Points – General Options – Geral – Frequency .................. 185
Figure 179 – Power Quality – Test – Insert/ Edit Points – General Options – Definitions – Frequency ......... 185
Figure 180 – Power Quality – Test – Insert/ Edit Points – General Options – Geral – Magnitude .................. 186
Figure 181 – Power Quality – Test – Insert/ Edit Points – General Options – Definitions – Magnitude ......... 186
Figure 182 – Power Quality – Test – Insert/ Edit Points – General Options – Geral – Flicker ........................ 187
Figure 183 – Power Quality – Test – Insert/ Edit Points – General Options – Definitions – Flicker ............... 187
Figure 184 – Power Quality – Test – Insert/ Edit Points – General Options – Geral – Sag ............................. 188
Figure 185 – Power Quality – Test – Insert/ Edit Points – General Options – Definitions – Sag .................... 188
Figure 186 – Power Quality – Test – Insert/ Edit Points – General Options – Geral – Swell .......................... 189
Figure 187 – Power Quality – Test – Insert/ Edit Points – General Options – Definitions – Swell ................. 189
Figure 188 – Power Quality – Test – Insert/ Edit Points – General Options – Geral – Interruption ................ 190
Figure 189 – Power Quality – Test – Insert/ Edit Points – General Options – Definitions – Interruption ....... 190
Figure 190 – Power Quality – Test – Insert/ Edit Points – General Options – Geral – Transients ................... 191
Figure 191 – Power Quality – Test – Insert/ Edit Points – General Options – Definitions – Transients .......... 191
Figure 192 – Power Quality – Test – Insert/ Edit Points – General Options – Geral – Unbalance .................. 192
Figure 193 – Power Quality – Test – Insert/ Edit Points – General Options – Definitions – Unbalance ......... 192
Figure 194 – Power Quality – Test – Insert/ Edit Points – General Options – Geral – Harmonics .................. 193
Figure 195 – Power Quality – Test – Insert/ Edit Points – General Options – Definitions – Harmonics ......... 193
Figure 196 – Power Quality – Test – Insert/ Edit Points – General Options – Geral – Inter-Harmonics ......... 194
Figure 197 – Power Quality – Test – Insert/ Edit Points – Definitions – Inter-Harmonics .............................. 194
Figure 198 – Power Quality – Test – Insert/ Edit Points – General Options – Geral – Quick Change ............ 195
Figure 199 – Power Quality – Test – Insert/ Edit Points – General Options – Quick Change ......................... 195
Figure 200 – Power Quality – Test – Insert/ Edit Points – General Options – Geral – Multi-Incident ............ 196
Figure 201 – Power Quality – Test – Insert/ Edit Points – General Options – Geral – Assessment ................ 196
Figure 202 – Power Quality – Test – Test Points ............................................................................................. 197
Figure 203 – Power Quality – Main Screen – Test Settings ............................................................................. 198
Figure 204 – Power Quality – Main Screen – Test Settings – Direc. Generation Channel .............................. 198
Figure 205 – Power Quality – Trigger Interf. – Trigger Logic ......................................................................... 199
Figure 206 – CT – Open Settings at Star .......................................................................................................... 202
Figure 207 – CT – Main screen ........................................................................................................................ 202
Figure 208 – CT – Quick Access Bar ............................................................................................................... 203
Figure 209 – CT – General Button Resources .................................................................................................. 203
Figure 210 – CT – Home Menu ....................................................................................................................... 204
Figure 211 – CT – Home Menu – Hardware, Results and Generation ............................................................. 204
Figure 212 – CT – Home Menu – Options, Report and Layout ....................................................................... 205
Figure 213 – CT – View Menu – Windows and Layout .................................................................................. 206
Figure 214 – CT – Software Options Menu – General ..................................................................................... 207
Figure 215 – CT – Adjustments button ............................................................................................................ 207
Figure 216 – CT – Settings Screen – General .................................................................................................. 208
Figure 217 – CT – Relation and Connection Scheme ...................................................................................... 209
Figure 218 – CT – Sub-tab - Ratio ................................................................................................................... 209
Figure 219 – CT – Voltage Relation and Connection Scheme ......................................................................... 210
Figure 220 – CT – Sub tab - Relation to Voltage ............................................................................................. 210
Figure 221 – CT – Rogowski Ratio and Connection Scheme .......................................................................... 211
Figure 222 – CT – Sub tab - Rogowski Ratio .................................................................................................. 211
Figure 223 – CT – Low Power Ratio and Connection Scheme ........................................................................ 212
Figure 504 – Sequenc – Sequence Tab – NO sub tab – Description Fault Mode ............................................. 402
Figure 505 – Sequenc – Sequence Tab – Resources – Part 10 ......................................................................... 402
Figure 506 – Sequenc – Sequence Tab – NO sub tab – Description Constant ZI Mode .................................. 403
Figure 507 – Sequenc – Sequence Tab – Resources – Part 12 ......................................................................... 403
Figure 508 – Sequenc – Sequence Tab – NO sub tab – Description Constant ZV Mode................................. 404
Figure 509 – Sequenc – Sequence Tab – Resources – Part 12 ......................................................................... 404
Figure 510 – Sequenc – Sequence Tab – NO sub tab – Z mode description – constant source ....................... 405
Figure 511 – Sequenc – Sequence Tab – Resources – Part 13 ......................................................................... 405
Figure 512 – Sequenc – Sequence Tab – NO sub tab – Description of the Z Mode - Smart............................ 406
Figure 513 – Sequenc – Sequence Tab – Resources – Part 14 ......................................................................... 406
Figure 514 – Sequenc – Sequence Tab – NO sub tab – Description of the Standby Mode .............................. 407
Figure 515 – Sequenc – Sequence Tab – Resources – Part 15 ......................................................................... 407
Figure 516 – Sequenc – Sequence Tab – Resources – Part 16 ......................................................................... 408
Figure 517 – Sequenc – Sequence Tab – NO sub tab – Description of the Accommodation Mode ................ 409
Figure 518 – Sequenc – Sequence Tab – Output Analog. DC Sub tab – General Description ........................ 409
Figure 519 – Sequenc – Sequence Tab – Binary Outputs sub tab – General Description ................................ 409
Figure 520 – Sequenc – Sequence Tab – GOOSE Outputs sub tab – General Description.............................. 410
Figure 521 – Sequenc – Sequence Tab – Time and Advance sub tab – General Description .......................... 410
Figure 522 – Sequenc – Waveform Tab ........................................................................................................... 411
Figure 523 – Sequenc – Waveform Tab – Resources ....................................................................................... 411
Figure 524 – Sequenc – Waveform Tab – Resources – View .......................................................................... 412
Figure 525 – Sequenc – Phasor Tab ................................................................................................................. 412
Figure 526 – Sequenc – Phasor Tab – Resources ............................................................................................. 413
Figure 527 – Sequenc – Waveform Tab – Resources – Show Columns .......................................................... 413
Figure 528 – Sequenc – Trajectories tab .......................................................................................................... 414
Figure 529 – Sequenc – Trajectories tab – Resources ...................................................................................... 414
Figure 530 – Sequenc – Harmonics Tab........................................................................................................... 415
Figure 531 – Sequenc – Harmonics Tab – Resources ...................................................................................... 415
Figure 532 – Sequenc – Synchronoscope tab ................................................................................................... 416
Figure 533 – Sequenc – Plane Z Tab ................................................................................................................ 417
Figure 534 – Sequenc – Plane Z Tab – Resources ........................................................................................... 417
Figure 535 – Sequenc – Plane Z Tab – Resources ........................................................................................... 418
Figure 536 – Rampa – Open Settings at Start................................................................................................... 420
Figure 537 – Rampa – Main screen .................................................................................................................. 421
Figure 538 – Rampa – Quick Access Bar ......................................................................................................... 421
Figure 539 – Rampa – General Button Resources ............................................................................................ 422
Figure 540 – Rampa – Home Menu ................................................................................................................. 422
Figure 541 – Rampa – Home Menu – Hardware and Sequence ....................................................................... 423
Figure 542 – Rampa – Home Menu – Result and Generation .......................................................................... 423
Figure 543 – Rampa – Home Menu – Options, Report, Units and Layout ...................................................... 424
Figure 544 – Rampa – View Menu – Windows and Layout ............................................................................ 425
Figure 545 – Rampa – Software Options Menu – General ............................................................................... 425
Figure 546 – Rampa – Adjustments button ...................................................................................................... 426
Figure 547 – Rampa – Adjustment Screen - General ....................................................................................... 427
Figure 548 – Rampa – Parameters screen – Distance – Adjustment ................................................................ 428
Figure 549 – Rampa – Distance Protection Adjustment – Insert – Zone – Part 1 ............................................ 429
Figure 550 – Rampa – Distance Protection Adjustment – Insert – Zone – Part 2 ............................................ 430
Figure 551 – Rampa – Distance Protection Adjustment - Insert - Zone ........................................................... 431
Figure 552 – Rampa – Distance Protection Adjustment – Insert – Drawn Area .............................................. 432
Figure 553 – Rampa – Distance Protection Adjustment – Insert – Quadrilateral Zone.................................... 433
Figure 554 – Rampa – Adjustment screen - Synchronism - Systems ............................................................... 433
Figure 555 – Rampa – Adjustment Screen – Synchronism – Systems – Part 1................................................ 434
Figure 556 – Rampa – Adjustment Screen - Synchronism - Systems - Part 2.................................................. 435
Figure 557 – Rampa – Adjustment screen - Synchronism - Synchronization adjustment ................................ 436
Figure 558 – Rampa – Adjustment Screen - Synchronism - Adjust Synchronization - Part 1 ......................... 436
Figure 559 – Rampa – Adjustment Screen - Synchronism - Adjust Synchronization - Part 2 ......................... 436
Figure 560 – Rampa – Adjustment Screen - Synchronism - Adjust Synchronization - Part 3 ......................... 437
Figure 561 – Rampa – Adjustment Screen - Synchronism - Adjust Synchronization - Part 4 ......................... 437
Figure 562 – Rampa – Sequence Tab – NODE sub tab – Two Nodes and three Sequences ............................ 438
Figure 563 – Rampa – Sequence Tab ............................................................................................................... 438
Figure 564 – Rampa – Sequence Tab – Resources Ramp in Module mode ..................................................... 439
Figure 565 – Rampa – Sequence Tab – Resources Ramp in Angle mode........................................................ 440
Figure 566 – Rampa – Sequence Tab – Resources Ramp in order Frequency ................................................. 441
Figure 567 – Rampa – Sequence Tab – Resources in the way and Angle Module .......................................... 442
Figure 568 – Rampa – Sequence Tab – Resources Ramp in Module and Frequency mode ............................ 443
Figure 569 – Rampa – Sequence Tab – Resources Ramp in dF/dt mode ......................................................... 444
Figure 570 – Rampa – Sequence Tab – Resources Ramp in order Symmetrical Component .......................... 445
Figure 571 – Rampa – Sequence Tab – Resources Ramp in Phase-phase mode .............................................. 446
Figure 572 – Rampa – Sequence Tab – Resources Ramp in Power mode ....................................................... 447
Figure 573 – Rampa – Sequence Tab – Resources Ramp in mode impedance Icc and Zlin ............................ 448
Figure 574 – Rampa – Sequence Tab – Resources Ramp mode impedance at Vcc and Zlin ........................... 449
Figure 575 – Rampa – Sequence Tab – Resources Ramp mode impedance in SIR and Zlin........................... 450
Figure 576 – Rampa – Sequence Tab – DC Analog Outputs sub tab – General Description ........................... 451
Figure 577 – Rampa – Sequence Tab – Binary Outputs sub tab – General Description .................................. 451
Figure 578 – Rampa – Sequence Tab – GOOSE Outputs sub tab – General Description ................................ 452
Figure 579 – Rampa – Sequence Tab – Sub tab Time and Advance – General Description ............................ 452
Figure 580 – Rampa – Waveform Tab ............................................................................................................. 453
Figure 581 – Rampa – Waveform Tab - Resources.......................................................................................... 453
Figure 582 – Rampa – Waveform Tab – Resources - View ............................................................................. 454
Figure 583 – Rampa – Phasor Tab ................................................................................................................... 454
Figure 584 – Rampa – Phasor Tab- Resources ................................................................................................. 455
Figure 585 – Rampa – Waveform Tab – Resources – Show Columns............................................................. 455
Figure 586 – Rampa – Trajectories tab ............................................................................................................ 456
Figure 587 – Rampa – Trajectories tab - Resources ......................................................................................... 456
Figure 588 – Rampa – Harmonics Tab ............................................................................................................. 457
Figure 589 – Rampa – Harmonics Tab - Resources ......................................................................................... 457
Figure 590 – Rampa – Synchronoscope tab ..................................................................................................... 458
Figure 591 – Rampa – Plane Z Tab .................................................................................................................. 458
Figure 592 – Rampa – Plane Z Tab - Resources .............................................................................................. 459
Figure 593 – Rampa – Plane Z Tab - Resources .............................................................................................. 459
Figure 594 – Master – Open Settings at Start ................................................................................................... 461
Figure 595 – Master – Main screen .................................................................................................................. 462
Figure 596 – Master – Quick Access Bar ......................................................................................................... 462
Figure 597 – Master – Resources General Button ............................................................................................ 463
Figure 598 – Master – Home Menu.................................................................................................................. 463
Figure 599 – Master – Home Menu – Hardware and Sequence ....................................................................... 464
Figure 600 – Master – Home Menu – Result and Generation .......................................................................... 464
Figure 601 – Master – Home Menu – Options, Report, Units and Layout ....................................................... 465
Figure 602 – Master – View Menu – Windows and Layout............................................................................. 466
Figure 603 – Master – Software Options Menu – General ............................................................................... 466
Figure 604 – Master – Adjustments button ...................................................................................................... 467
Figure 605 – Master – Adjustment Screen – General ....................................................................................... 468
Figure 606 – Master – Parameters Screen – Distance – Adjustment ................................................................ 469
Figure 607 – Master – Parameters Screen – Distance – Distance Protection Adjustment – Insert – Zone ...... 470
Figure 608 – Master – Parameters Screen – Distance – Distance Protection Adjustment – Insert – Zone ...... 471
Figure 609 – Master – Distance Protection Adjustment – Insert – Zone .......................................................... 472
Figure 610 – Master – Distance Protection Adjustment – Insert – Drawn Area .............................................. 473
Figure 611 – Master – Distance Protection Adjustment – Insert – Quadrilateral Zone .................................... 474
Figure 612 – Master – Adjustment Screen – Synchronism – Systems ............................................................. 475
Figure 613 – Master – Adjustment Screen – Synchronism – Systems – Part 1 ................................................ 475
Figure 614 – Master – Adjustment Screen – Synchronism – Systems – Part 2 ................................................ 476
Figure 615 – Master – Adjustment screen – Synchronism – Synchronization adjustment............................... 477
Figure 616 – Master – Adjustment Screen – Synchronism – Synchronization Adjustment – Part 1 ............... 477
Figure 617 – Master – Adjustment Screen – Synchronism – Synchronization Adjustment – Part 2 ............... 478
Figure 618 – Master – Adjustment Screen – Synchronism – Synchronization Adjustment – Part 3 ............... 478
Figure 619 – Master – Adjustment Screen – Synchronism – Synchronization Adjustment – Part 4 ............... 478
Figure 620 – Master – Sequence Tab - NODE sub tab - Two Nodes and three Sequences ............................. 479
Figure 621 – Master – Sequence Tab - NODE sub tab - Description of the direct mode ................................. 480
Figure 622 – Master – Sequence Tab – Resources – Part 1.............................................................................. 480
Figure 623 – Master – Sequence Tab – Resources – Part 2.............................................................................. 481
Figure 624 – Master – Sequence Tab – NODE sub tab – Description of the Harmonics mode ....................... 482
Figure 625 – Master – Sequence Tab – Resources – Part 3.............................................................................. 482
Figure 626 – Master – Sequence Tab – Resources – Part 4.............................................................................. 483
Figure 627 – Master – Sequence Tab - NODE sub tab - Interharmonic mode description .............................. 483
Figure 628 – Master – Sequence Tab – Resources – Part 5.............................................................................. 484
Figure 629 – Master – Sequence Tab – Resources – Part 6.............................................................................. 484
Figure 630 – Master – Sequence Tab – NODE sub tab – Description of the Symmetric Components Mode . 485
Figure 631 – Master – Sequence Tab – Resources – Part 7.............................................................................. 485
Figure 632 – Master – Sequence Tab – NODE sub tab – Description of the Phase-to-Phase Mode ................ 486
Figure 633 – Master – Sequence Tab – Resources – Part 8.............................................................................. 486
Figure 634 – Master – Sequence Tab – NODE sub tab – Description of the Power Mode .............................. 487
Figure 635 – Master – Sequence Tab – Resources – Part 9.............................................................................. 487
Figure 636 – Master – Sequence Tab - NODE sub tab– Description of the Fault Mode ................................. 488
Figure 637 – Master – Sequence Tab – Resources – Part 10............................................................................ 488
Figure 638 – Master – Sequence Tab - NODE sub tab– Description of the Constant ZI Mode ....................... 489
Figure 639 – Master – Sequence Tab – Resources – Part 11............................................................................ 489
Figure 640 – Master – Sequence Tab - NODE sub tab– Description of the Constant ZV Mode ..................... 490
Figure 641 – Master – Sequence Tab – Resources – Part 12............................................................................ 490
Figure 642 – Master – Sequence Tab – NODE sub tab – Description of the Z mode – constant source ......... 491
Figure 643 – Master – Sequence Tab - Resources– Part 13 ............................................................................. 491
Figure 644 – Master – Sequence Tab - NODE sub tab– Description of the Z Mode - Smart .......................... 492
Figure 645 – Master – Sequence Tab – Resources – Part 14............................................................................ 492
Figure 646 – Master – Sequence Tab - NODE sub tab– Description of the Standby Mode............................. 493
Figure 647 – Master – Sequence Tab – Resources – Part 15............................................................................ 493
Figure 648 – Master – Sequence Tab – Resources – Part 16............................................................................ 494
Figure 649 – Master – Sequence Tab - NODE sub tab– Description of the Accommodation Mode ............... 494
Figure 650 – Master – Sequence Tab ............................................................................................................... 495
Figure 651 – Master – Sequence Tab – Resources the ramp in Module mode ................................................ 496
Figure 652 – Master – Sequence Tab – Resources of the ramp in the Angle mode ......................................... 497
Figure 653 – Master – Sequence Tab – Resources of the ramp in order Frequency ........................................ 498
Figure 654 – Master – Sequence Tab – Resources of the ramp on the way and Angle Module ...................... 499
Figure 655 – Master – Sequence Tab – Resources the ramp in mode Module and Frequency ........................ 500
Figure 656 – Master – Sequence Tab – Resources of the ramp in mode dF/dt ................................................ 501
Figure 657 – Master – Sequence Tab – Resources of the ramp in mode Symmetrical Component ................. 502
Figure 658 – Master – Sequence Tab – Resources of the ramp in the Phase-to-Phase Mode .......................... 503
Figure 659 – Master – Sequence Tab – Resources the ramp in Power Mode .................................................. 504
Figure 660 – Master – Sequence Tab – Resources of the ramp in Impedance Icc and Zlin mode ................... 505
Figure 661 – Master – Sequence Tab – Resources of the ramp in Vcc and Zlin Impedance mode ................. 506
Figure 662 – Master – Sequence Tab – Resources of the ramp in SIR and Zlin Impedance mode .................. 507
Figure 663 – Master – Sequence Tab ............................................................................................................... 508
Figure 664 – Master – Sequence Tab – Channel Direction .............................................................................. 509
Figure 665 – Master – Sequence Tab – Channel Direction .............................................................................. 510
Figure 666 – Master – Sequence Tab – Output Analog. DC Sub tab – General Description ........................... 511
Figure 667 – Master – Sequence Tab – Binary Outputs sub tab – General Description .................................. 511
Figure 668 – Master – Sequence Tab – GOOSE Outputs Sub tab – General Description ............................... 512
Figure 669 – Master – Sequence Tab –Time and Advance Sub tab – General Description ............................. 512
Figure 670 – Master – Waveform Tab ............................................................................................................. 513
Figure 671 – Master – Waveform Tab - Resources .......................................................................................... 513
Figure 726 – Sobrecor – Main screen – Test Settings – Adjustments .............................................................. 548
Figure 727 – Sobrecor – Main screen – Test Settings – Other Adjustments .................................................... 548
Figure 728 – Sobrecor – Main screen – Test Settings –Interfaces.................................................................... 549
Figure 729 – Sobrecor – Main screen – Test Settings – Pickup Limits ............................................................ 549
Figure 730 – Diferenc – Open Settings at Start ................................................................................................ 551
Figure 731 – Diferenc – Main screen ............................................................................................................... 552
Figure 732 – Diferenc – Quick Access Bar ...................................................................................................... 552
Figure 733 – Diferenc – Resources General Button ......................................................................................... 553
Figure 734 – Diferenc – Home Menu ............................................................................................................... 554
Figure 735 – Diferenc – Home Menu – Hardware and Generation .................................................................. 555
Figure 736 – Diferenc – Home Menu – Options, Report, Units and Layout .................................................... 556
Figure 737 – Diferenc – View Menu – Windows and Layout .......................................................................... 557
Figure 738 – Diferenc – Software Options Menu – General ............................................................................ 557
Figure 739 – Diferenc – Adjustments button ................................................................................................... 558
Figure 740 – Diferenc – Adjustment Screen – General .................................................................................... 559
Figure 741 – Diferenc – Protect Equipment – Differential – Settings .............................................................. 560
Figure 742 – Diferenc – Adjust Prot. Differential – Differential – Adjustment ............................................... 561
Figure 743 – Diferenc – Adjust Prot. Differential – Differential – Slope Definition ....................................... 562
Figure 744 – Diferenc – Maximized Systems Tab ........................................................................................... 563
Figure 745 – Diferenc – Relay Details tab ....................................................................................................... 564
Figure 746 – Diferenc – Graphic Tab ............................................................................................................... 564
Figure 747 – Diferenc – Graphic tab maximized. ............................................................................................ 565
Figure 748 – Diferenc – Graph Tab - Resources .............................................................................................. 565
Figure 749 – Diferenc – Waveform Tab........................................................................................................... 566
Figure 750 – Diferenc – Waveform Tab - Resources ....................................................................................... 566
Figure 751 – Diferenc – Waveform Tab – Resources - Display....................................................................... 567
Figure 752 – Diferenc – Phasor Tab ................................................................................................................. 567
Figure 753 – Diferenc – Phasor Tab- Resources .............................................................................................. 568
Figure 754 – Diferenc – Waveform Tab – Resources – Show Columns .......................................................... 568
Figure 755 – Diferenc – Main screen – Test Set .............................................................................................. 570
Figure 756 – Diferenc – Main screen – Test Set – Insert / Edit Points ............................................................. 570
Figure 757 – Diferenc – Main screen – Test Set – Test sets Points .................................................................. 571
Figure 758 – Diferenc – Main screen – Test Set – General options – Short current ........................................ 572
Figure 759 – Diferenc –General Options – Winding Current ........................................................................... 573
Figure 760 – Diferenc – Main screen – Test Set –General Options – IDif and IRest....................................... 573
Figure 761 – Diferenc – Main screen – Test Set – System............................................................................... 574
Figure 762 – Diferenc – Main screen – Test Set – Relay Details (Diagram) ................................................... 574
Figure 763 – Diferenc – Main screen – Test Set – Characteristic .................................................................... 575
Figure 764 – Diferenc – Main screen – Test Set – Waveform ......................................................................... 575
Figure 765 – Diferenc – Main screen – Test – Phasors .................................................................................... 576
Figure 766 – Diferenc – Main screen – Point Test ........................................................................................... 577
Figure 767 – Diferenc – Main screen – Point Test– Test set Points ................................................................. 577
Figure 768 – Diferenc – Main screen – Point Test – Insert / Edit Points ......................................................... 578
Figure 769 – Diferenc – Main screen – Point Test - General Options.............................................................. 578
Figure 770 – Diferenc – Main screen – Point Test – Characteristic ................................................................. 579
Figure 771 – Diferenc – Main screen – Search Test......................................................................................... 580
Figure 772 – Diferenc – Main screen – Search Test......................................................................................... 580
Figure 773 – Diferenc – Main screen – Search Test – Test set Points ............................................................. 581
Figure 774 – Diferenc – Main screen – Search Test – Insert / Edit Points ....................................................... 581
Figure 775 – Diferenc – Main screen – Search Test – General Options .......................................................... 582
Figure 776 – Diferenc – Main screen – Search Test – Characteristic............................................................... 582
Figure 777 – Diferenc – Main screen – Test Settings....................................................................................... 583
Figure 778 – Diferenc – Main screen – Generation Channels Direc – Current ................................................ 584
Figure 779 – Diferenc – Main screen – Pre-Fault Time and Binary Outputs and GOOSE .............................. 585
Figure 780 – Diferenc – Main screen – Test Settings – Interfaces ................................................................... 585
Figure 781 – Diferenc – Main screen – Test Settings....................................................................................... 586
Figure 942 – PSB_OoS – System – Location of the Relay in the System ........................................................ 692
Figure 943 – PSB_OoS – Waveform Tab ........................................................................................................ 692
Figure 944 – PSB_OoS – Waveform Tab - Resources ..................................................................................... 693
Figure 945 – PSB_OoS – Waveform Tab – Resources - View ........................................................................ 693
Figure 946 – PSB_OoS – Phasor Tab .............................................................................................................. 694
Figure 947 – PSB_OoS – Phasor Tab- Resources ............................................................................................ 694
Figure 948 – PSB_OoS – Resources – Show Columns .................................................................................... 695
Figure 949 – PSB_OoS – Trajectories tab ........................................................................................................ 695
Figure 950 – PSB_OoS – Trajectories tab - Resources .................................................................................... 696
Figure 951 – PSB_OoS – Main screen – System Simulation Test ................................................................... 697
Figure 952 – PSB_OoS – Main screen – System Simulation – Options Oscillations Test set ......................... 697
Figure 953 – PSB_OoS – Main screen – System Simulation– Result Test set Oscillations............................. 698
Figure 954 – PSB_OoS – Main screen – System Simulation – Insert / Edit Oscillations ................................ 698
Figure 955 – PSB_OoS – Main screen – System Simulation - General Options - Synchronous ..................... 699
Figure 956 – PSB_OoS – Main screen – System Simulation - General Options - Asynchronous ................... 699
Figure 957 – PSB_OoS –Sequence .................................................................................................................. 700
Figure 958 – PSB_OoS – Define Sequence ..................................................................................................... 700
Figure 959 – PSB_OoS – System Simulation - System ................................................................................... 701
Figure 960 – PSB_OoS – System Simulation – Sequence – Points Created .................................................... 701
Figure 961 – PSB_OoS – Main screen – Trajectory Simulation ...................................................................... 702
Figure 962 – PSB_OoS – Main screen – Trajectory Simulation – Test set Trajectory Options ....................... 702
Figure 963 – PSB_OoS – Main screen – Trajectory Simulation – Result of Trajectories Test set .................. 703
Figure 964 – PSB_OoS – Main screen – System Simulation – Insert / Edit Trajectories ................................ 703
Figure 965 – PSB_OoS – Main screen – System Simulation – General Options – Trajectory ........................ 704
Figure 966 – PSB_OoS – Main screen – System Simulation – General Options – System ............................. 704
Figure 967 – PSB_OoS – Main screen – System Simulation – General Options – Fault................................. 705
Figure 968 – PSB_OoS – Main screen – System Simulation – General Options – Evaluation........................ 705
Figure 969 – PSB_OoS – Main screen – Test Settings .................................................................................... 706
Figure 970 – PSB_OoS – Main screen – Test Settings – Direction of Generation Channel – Pre-Simulation and
Post-Simulation ................................................................................................................................................ 707
Figure 971 – PSB_OoS – Main screen – Test Settings – Ratio between CTs and TPs - Polarities of CTs and
TPs.................................................................................................................................................................... 707
Figure 972 – PSB_OoS – Main screen – Test Settings – Pre- Simulation, Post- Simulation Time and Binary
Outputs and GOOSE ........................................................................................................................................ 708
Figure 973 – PSB_OoS – Main screen – Test Settings – Enable Interfaces ..................................................... 708
Figure 974 – Synchronism – Open Settings at Start ......................................................................................... 710
Figure 975 – Synchronism – Settings Screen ................................................................................................... 711
Figure 976 – Synchronism – Quick Access Bar ............................................................................................... 711
Figure 977 – Synchronism – Resources General Button .................................................................................. 712
Figure 978 – Synchronism – Home Menu ........................................................................................................ 712
Figure 979 – Synchronism – Home Menu – Hardware and Generation ........................................................... 713
Figure 980 – Synchronism – Home Menu – Options, Report, Units and Layout ............................................. 714
Figure 981 – Synchronism – View Menu – Windows and Layout ................................................................... 715
Figure 982 – Synchronism – Software Options Menu – General ..................................................................... 715
Figure 983 – Synchronism – Adjustments button ............................................................................................ 716
Figure984 - Synchronism – Adjustment Screen – General – General Information. ......................................... 716
Figure 985 – Synchronism – Adjustment Screen – General - Systems ............................................................ 717
Figure 986 – Synchronism – Adjustment Screen - Synchronism – Systems .................................................... 718
Figure 987 – Synchronism – Adjustment Screen - Synchronism – Systems – Part 1 ...................................... 719
Figure 988 – Synchronism – Adjustment Screen - Synchronism - Systems - Part 2 ........................................ 719
Figure 989 – Synchronism – Adjustment screen - Synchronism - Synchronization adjustment ...................... 720
Figure 990 – Synchronism – Adjustment Screen - Synchronism – Sync Settings – Part 1 .............................. 720
Figure 991 – Synchronism – Adjustment Screen - Synchronism – Sync Settings – Part 2 .............................. 721
Figure 992 – Synchronism – Adjustment Screen - Synchronism – Sync Settings – Part 3 .............................. 721
Figure 993 – Synchronism – Adjustment Screen - Synchronism – Sync Settings – Part 4 .............................. 721
Figure 994 – Synchronism – Graph Tab .......................................................................................................... 722
Figure 1049 – VoltsPHz – Adjustment Screen – Volts / Hertz – Part 2 ........................................................... 755
Figure 1050 – VoltsPHz – Adjustment Screen Volts / Hertz – Part 3 .............................................................. 755
Figure 1051 – VoltsPHz – Adjustment Screen Volts / Hertz – Graph (V/Hz) x t ............................................ 755
Figure 1052 – VoltsPHz – Adjustment Screen Volts / Hertz – Graph Pickup ................................................. 756
Figure 1053 – VoltsPHz – Graph Tab .............................................................................................................. 756
Figure 1054 – VoltsPHz – Maximized Graph Tab ........................................................................................... 757
Figure 1055 – VoltsPHz – Graph Tab - Resources .......................................................................................... 757
Figure 1056 – VoltsPHz – Waveform Tab ....................................................................................................... 758
Figure 1057 – VoltsPHz – Waveform Tab - Resources ................................................................................... 758
Figure 1058 – VoltsPHz – Waveform Tab – Resources - View ....................................................................... 759
Figure 1059 – VoltsPHz – Phasor Tab ............................................................................................................. 759
Figure 1060 – VoltsPHz – Phasor Tab- Resources........................................................................................... 760
Figure 1061 – VoltsPHz – Waveform Tab – Resources – Show Columns ...................................................... 760
Figure 1062 – VoltsPHz – Main screen – Time Test ....................................................................................... 761
Figure 1063 – VoltsPHz – Main screen – Time Test– Options ........................................................................ 761
Figure 1064 – VoltsPHz – Main screen – Time Test – Test set Lines ............................................................. 762
Figure 1065 – VoltsPHz – Main screen – Time Test – Insert / Edit Lines ....................................................... 762
Figure 1066 – VoltsPHz – Main screen – Time Test - General Options – New Line ...................................... 763
Figure 1067 – VoltsPHz – Main screen – Time Test - General Options – Sequence ....................................... 763
Figure 1068 – VoltsPHz – Main screen – Time Test – Graph.......................................................................... 764
Figure 1069 – VoltsPHz – Main screen – Time Test –Legend......................................................................... 764
Figure 1070 – VoltsPHz – Main screen – Search Test ..................................................................................... 765
Figure 1071 – VoltsPHz – Main screen – Search Test– Options ..................................................................... 765
Figure 1072 – VoltsPHz – Main screen – Search Test – Test set Points .......................................................... 766
Figure 1073 – VoltsPHz – Main screen – Search Test – Insert / Edit Lines .................................................... 766
Figure 1074 – VoltsPHz – Main screen –Search Test – General Options ........................................................ 767
Figure 1075 – VoltsPHz – Main screen – Search Test – General Options - Sequence .................................... 767
Figure 1076 – VoltsPHz – Main screen – Search Test – Graph ....................................................................... 768
Figure 1077 – VoltsPHz – Main screen – Search Test – Legend. .................................................................... 768
Figure 1078 – VoltsPHz – Main screen – Trajectory ....................................................................................... 769
Figure 1079 – VoltsPHz – Main screen – Trajectory – Options....................................................................... 769
Figure 1080 – VoltsPHz – Main screen – Trajectory – Test set Points ............................................................ 770
Figure 1081 – VoltsPHz – Main screen – Trajectory – Insert / Edit Lines....................................................... 770
Figure 1082 – VoltsPHz – Main screen – Trajectory – General Options ......................................................... 771
Figure 1083 – VoltsPHz – Main screen –Trajectory – Graph .......................................................................... 771
Figure 1084 – VoltsPHz – Main screen – Trajectory – Legend ....................................................................... 772
Figure 1085 – VoltsPHz – Main screen – Test Settings ................................................................................... 773
Figure 1086 – VoltsPHz – Main screen – Test Settings – Direction of Generation Channel– Pre-Simulation and
Post-Simulation ................................................................................................................................................ 773
Figure 1087 – VoltsPHz – Main screen – Test Settings – Pre-Simulation, Post-Simulation Time and Binary
Outputs and GOOSE ........................................................................................................................................ 774
Figure 1088 – VoltsPHz – Main screen – Test Settings – Search Test ............................................................ 774
Figure 1089 – VoltsPHz – Main screen – Test Settings – Enable Interfaces ................................................... 775
Figure 1090 – Transdut – Open Settings at Start .............................................................................................. 777
Figure 1091 – Transdut – Settings Screen ........................................................................................................ 778
Figure 1092 – Transdut – Quick Access Bar .................................................................................................... 778
Figure 1093 – Transdut – Resources General Button ....................................................................................... 779
Figure 1094 – Transdut – Home Menu ............................................................................................................ 779
Figure 1095 – Transdut – Home Menu – Hardware and Generation................................................................ 780
Figure 1096 – Transdut – Home Menu – Options, Report, Units and Layout .................................................. 781
Figure 1097 – Transdut – Display Menu – Windows and Layout .................................................................... 782
Figure 1098 – Transdut – Software Options Menu – General .......................................................................... 782
Figure 1099 – Transdut – Adjustments button ................................................................................................. 783
Figure 1100 – Transdut – Transducer – Transducer Adjustment ..................................................................... 784
Figure 1101 – Transdut – Graph Tab ............................................................................................................... 784
Figure 1102 – Transdut – Maximized Graph Tab ............................................................................................ 785
Figure 1154 – TransientView – Sequence Tab – GOOSE Outputs sub tab – General Description ................. 818
Figure 1155 – TransientView – Sequence Tab – Time and Advance sub tab – General Description .............. 818
Figure 1156 – TransientView – Waveform Tab ............................................................................................... 819
Figure 1157 – TransientView – Waveform Tab - Resources ........................................................................... 819
Figure 1158 – TransientView – Waveform Tab – Resources - View............................................................... 820
Figure 1159 – TransientView – Phasor Tab ..................................................................................................... 820
Figure 1160 – TransientView – Phasor Tab- Resources .................................................................................. 821
Figure 1161 – TransientView – Waveform Tab – Resources – Show Columns .............................................. 821
Figure 1162 – TransientView – Trajectories tab .............................................................................................. 822
Figure 1163 – TransientView – Trajectories tab - Resources........................................................................... 822
Figure 1164 – TransientView – Harmonics Tab .............................................................................................. 823
Figure 1165 – TransientView – Harmonics Tab - Resources ........................................................................... 823
Figure 1166 – TransientView – Synchronoscope tab ....................................................................................... 824
Figure 1167 – TransientView – Plane Z Tab.................................................................................................... 824
Figure 1168 – TransientView – Plane Z Tab - Resources ................................................................................ 825
Figure 1169 – TransientView – Plane Z Tab - Resources ................................................................................ 825
Figure 1170 – Statistical Analysis – Initial screen ........................................................................................... 827
Figure 1171 – Statistical Analysis– Quick Access Bar .................................................................................... 828
Figure 1172 – Statistical Analysis– File Button Resources .............................................................................. 828
Figure 1173 – Statistical Analysis– General Information – Identification of the Analysis and Responsible ... 829
Figure 1174 – Statistical Analysis– Adding a Table ........................................................................................ 829
Figure 1175 – Statistical Analysis – Error opening file .................................................................................... 830
Figure 1176 – Statistical Analysis – Table Tab ................................................................................................ 830
Figure 1177 – Statistical Analysis – Tables Tab - Definition of values for variables not defined or defined by
labels................................................................................................................................................................. 831
Figure 1178 – Statistical Analysis– Tables Tab - Imported table. .................................................................... 832
Figure 1179 – Statistical Analysis – Analysis Tab - Analysis List ................................................................... 833
Figure 1180 – Statistical Analysis – Analysis Tab - Filters / Weights List - Filters Tab ................................. 833
Figure 1181 – Statistical Analysis – Analysis Tab - Filters / Weights List - Weights Tab .............................. 834
Figure 1182 – Statistical Analysis – Analysis - PlotXY Graph ........................................................................ 834
Figure 1183 – Statistical Analysis – Analysis - Results - Definitions PlotXY Chart ....................................... 835
Figure 1184 – Statistical Analysis – Analysis - Results ( PlotXY Graph ) - Sub Tab Table ............................ 836
Figure 1185 – Statistical Analysis– Analysis – Graphic Bar ............................................................................ 837
Figure 1186 – Statistical Analysis – Bar Graph– Definitions for graph construction ...................................... 837
Figure 1187 – Statistical Analysis – Analysis – Pizza Chart ............................................................................ 839
Figure 1188 – Statistical Analysis – Present Report - Configure Presentation................................................. 840
Figure 1189 – Autodiagnóstico– Initial Screen ................................................................................................ 842
Figure 1190 – Autodiagnóstico – Saving File .................................................................................................. 843
Figure 1191 – Autodiagnóstico –Initial File ..................................................................................................... 843
Figure 1192 – Statistical Analysis – Quick Access Bar ................................................................................... 844
Figure 1193 – Remote Access – Initial Screen ................................................................................................. 846
Figure 1194 – Remote Access – Resources ...................................................................................................... 847
Figure 1195 – Transient power curve ............................................................................................................... 852
Figure 1196 - CE-4000 panel layout ................................................................................................................ 853
Index Table
Table 1 – .........................................................................................................325
1 Introduction
This manual contains information on how to work with the software inserted within
the CONPROVE TEST CENTER (CTC) application manager. This software is used to
control the CE-6003; CE-6006; CE-6707; CE-6710 CE-7012 and CE-7024. For the software
to behave satisfactorily, make sure that your computer has the following conditions:
The manual was designed in an objective way, aiming to immediately put the user in
contact with the equipment.
The following chapter contains information about CE-6003; CE-6006; CE-6707; CE-
6710; CE-7012 and CE-7024.
CONPROVE
TEST
CENTER
HARDWARE AND SOFTWARE
BOOKMARK MANAGER OR APPLICATIONS
FOR CE-6003/CE-6006/CE-6707/CE-
6710/CE7012/CE-7024
Before starting the power generation in manual or automatic mode ensures that the
outputs of the current amplifiers are connected to the relay coils to be Test or any inductive
or resistive load that complies with the power limit maximum allowed for each channel. In
voltage amplifiers, make sure that there are no shorts in the output terminals or loads that
cause overload due to excess power.
Current measurement
Current generators channels. Binary inputs
channels.
interface.
The maximum amplitude that can be generated in each voltage channel is 300 V rms,
the maximum power for each channel is 90 VA and the maximum current is 0.3 A rms. So
that associates can reach up to a voltage with maximum amplitude of 300 Vrms or a
maximum power of 200 VA. The amplitude accuracy is better than or equal to 0.1% and the
angle is better or equal to 0.1°.
The maximum amplitude that can be generated in each current channel is 20 A rms,
the maximum power for each channel is 90 VA and the maximum voltage is 4.5 V rms. So
that members can reach a current with maximum amplitude of 60 Arms or a maximum
power of 300 VA. The amplitude accuracy is better than or equal to 0.1% and higher than or
equal to 0.15°.
The CE-6003 has three channels for measuring voltage. These channels are used to
process the acquisition of voltage signals so that energy quality analyzes can be made in the
measurement software.
The CE-6003 has three channels for measuring current. These channels are used to
process the acquisition of current signals, for the analysis of power quality made by the
measurement software.
The current measurement should be done only with the use of clamps recommend by
the manufacturer (ammeter pliers), never directly. Clamps are optional to the CE-6003, and
can be adjusted to work in three scales, 100mV/A for currents from 100mA to 10 A, 10mV /
A for currents from 100mA to 100A and 1mV/A for currents of 1 at 1000 A. insulation class
of the clamp is 600V ac rms. Before using the clamps, consult the manual of the clamp that
comes with the equipment.
The CE-6003 has six digital output interfaces, four by relays on the front and with
the aid of a cable, two transistorized outputs on the rear (optional cable).
The CE-6003 has eight digital input interfaces, being configured by dry or wet
contact.
Attention:
If it is at the binary input it is configured as wet, the voltage level used must be
higher than the adjustment level of the pickup (change level) of the input plus the maximum
precision limit of the input (according to the scale corresponding to the adjustment ). In most
cases, an easy way to respect this rule is to work with the pickup setting, at most, at half the
voltage level. For example, injected voltage of 100.0V is used as a “pickup” a value less
than or equal to 50.0V.
External hardware is attached to this connector, used in tests that require more
power, amplitude or number of channels. External hardware that can be coupled to the
CE6003 is the Amplifier CE-4000, the test set CE-6003 and the test set CE-6006 (produced
by CONPROVE).
The software that comes with the CE-6003 uses these analog interfaces for
measuring transducers. These interfaces allow the connection of several models and types of
transducers existing on the market, whose response signals are given in currents or DC
voltages. The interface allows DC currents up to +/- 25 mA with 0.05% accuracy at the full
scale and DC voltages up to +/- 10 V with 0.05% accuracy at the full scale.
The maximum amplitude that can be generated in each voltage channel is 300 V rms,
the maximum power for each channel is 90 VA and the maximum current is 0.3 A rms. So
that associates can reach up to a voltage with maximum amplitude of 600 Vrms or a
maximum power of 400 VA. The amplitude accuracy is better than or equal to 0,1 %
and better angle than or equal to 0.1°.
The maximum amplitude that can be generated in each current channel is 20 A rms,
the maximum power for each channel is 90 VA and the maximum voltage is 4.5 V rms. So
that members can reach a current with maximum amplitude of 120Arms or a maximum
power of 1100 VA. The amplitude accuracy is better than or equal to 0.1% and best angle
equal to 0.15°.
The CE-6006 has three channels for measuring voltage. These channels are used to
process the acquisition of voltage signals, for the analysis of power quality made by the
measurement software.
The CE-6006 has three channels for measuring current. These channels are used to
process the acquisition of current signals, for the analysis of energy quality made by the
measurement.
The current measurement should be made only with the use of clamps
recommended by the manufacturer (clamp meters), never directly. Clamps are optional to
the CE-6006, and can be adjusted to work in three scales, 100mV / A for currents from 100
mA to 10 A, 10mV / A for currents from 100 mA to 100 A and 1mV / A for currents of 1 at
1000 A. insulation class of the clamp is 600 V ac rms. Before using the clamps, consult the
manual of the clamp that comes with the equipment.
The CE-6006 has six digital output interfaces, four by relays on the front and with
the aid of a cable, two transistorized outputs on the rear (optional cable). There
are also six virtual outlets for sending GOOSE messages.
The CE-6006 has eight digital input interfaces being configured by dry or wet
contact. There are also eight virtual entries for receiving GOOSE messages.
Attention:
If it is at the binary input it is configured as wet, the voltage level used must be
higher than the adjustment level of the pickup (change level) of the input plus the maximum
precision limit of the input (according to the scale corresponding to the adjustment ). In most
cases, an easy way to respect this rule is to work with the pickup setting, at most, at half the
voltage level. For example, injected voltage of 100.0V is used as a “pickup” a value less
than or equal to 50.0V.
External hardware is attached to this connector, used in tests that require more
channels or more current. External hardware that can be coupled to the CE6006 is the
Amplifier CE-4000, CE-6006 and CE-6003 (produced by CONPROVE).
The software that comes with the CE-6006 uses these analog interfaces for
measuring transducers. These interfaces allow the connection of several models and types of
transducers existing on the market, whose response signals are given in currents or DC
voltages. The interface allows DC currents up to +/- 25 mA with 0.05% accuracy at the full
scale and DC voltages up to +/- 10 V with 0.05% accuracy at the full scale.
Digital input /
voltage and current Special secondary current and voltage
measurement. analog input.
The CE-670 7 has four fixed voltage generation channels. The maximum amplitude
that can be generated in each voltage channel is 300 V rms, the maximum power for each
channel is 100 VA and the maximum current is 1.5A rms. So that associates can reach a
voltage with a maximum amplitude of 600 Vrms or a maximum power of 350 VA. In the
"AC" the accuracy of the amplitude is less than or equal to 0.04% of reading plus 0.01% of
full scale and the angle accuracy is less than or equal to 0.1 °. For the “DC” mode, the
amplitude accuracy is less than or equal to 0.05% of the reading plus 0.05% of the full scale.
The CE-6707 has three current generation channels. The maximum amplitude that
can be generated in each current channel is 32A rms, the maximum power for each channel
is 130W / 210 VA and the maximum voltage is 10V rms. So that members can reach a
current with maximum amplitude of 96 Arms with a maximum power of 350 W / 550 VA.
The amplitude accuracy is less than or equal to 0.04% of the reading plus 0.01% of the full
scale. For the angle accuracy is less than or equal to 0.1°.
In the DC generation, the precision is defined by 0.1% of the reading plus 0.1% of
the full scale.
The CE-67 07 has twelve channels for measuring voltage and / or current. When
configured as voltage, these channels are used to process the acquisition of voltage signals,
for the analysis of power quality made by the measurement software. There are several
measurement ranges (200mV rms with a tolerance of 0.3% AC and 0.45% DC / 2V rms with
a tolerance of 0.15% AC and 0.25% DC / 20Vrms with a tolerance of 0.15% AC and 0.25%
DC and 600Vrms with tolerance of 0.15% AC and 0.25% DC, for voltages up to 400Vac) to
better match the acquired voltages at levels of amplitude suitable for processing in the
ADCs. All the tolerances mentioned above are related to the full scale (range).
When configured as current, these channels are used to process the acquisition of
current signals, for the analysis of power quality made by the measurement software. The
current measurement can be done using clamp pliers, shunts, transducers, Rogowski coils or
any other transducer that has voltage output in the ranges described above. To do this,
simply inform the full scale and the transformation ratio of the transducer used. Optionally
CONPROVE can supply Clamp pliers to meet the customer's needs.
The CE-670 7 has eight digital output interfaces (four relay outputs with potential-
free NO / NC contacts and four outputs in open collector or in 5V TTL voltage), in addition
to several outputs for sending GOOSE messages.
The CE-6707 has twelve digital input interfaces per NO / NC contact and voltage, in
addition to several inputs for receiving GOOSE messages.
Attention:
If it is at the binary input it is configured as wet, the voltage level used must be
higher than the adjustment level of the pickup (change level) of the input plus the maximum
precision limit of the input (according to the scale corresponding to the adjustment ). In most
cases, an easy way to respect this rule is to work with the pickup setting, at most, at half the
voltage level. For example, injected voltage of 100.0V is used as a “pickup” a value less
than or equal to 50.0V.
There are two connectors for coupling external hardware. Both the plug (S1) and
(S2) enables m control over six external analog channels each. The external hardware that
can be coupled to the CE-70 1 2 are the Amplifier CE-4000, the Test CE-7012, the Test CE
6710, the Test CE 6707, the Test CE-6006 and the Test CE-6003 (all produced by
CONPROVE).
Signal meter (DC) for testing voltage and current transducers. The current ranges are
± 2 mA/ ± 25 mA with an accuracy of 0.05% of the range. In the case of voltage, the ranges
are ±1V / ± 10 V with 0.05% accuracy of the range, ± 100mV with 0.03 % accuracy plus
0.05 % of the range and ± 10m V with 0 accuracy, 05% of reading plus 0.15 % of range.
These interfaces allow the connection of several models and types of transducers existing on
the market, whose response signals are given in currents or DC voltages.
Auxiliary source:
Auxiliary source for generating DC voltage up to 300V. The level is adjusted in the
software, in the hardware configuration screen. Adjustment range is from 5 to 300Vdc with
typical accuracy of 2% of range.
The CE- 6710 has four fixed voltage generation channels. The maximum amplitude
that can be generated in each voltage channel is 30 0 V rms, the maximum power of each
channel is 10 0 VA and the maximum current is 1.5 A rms. So that associates can reach a
voltage with a maximum amplitude of 6 00 Vrms or a maximum power of 35 0 VA. In the
“AC” the accuracy of the amplitude is less than or equal to 0.04% of the reading plus
0.01% of the full scale. For the angle, the accuracy is less than or equal to 0.05 °. For the
“DC” mode, the amplitude accuracy is less than or equal to 0.05% of the reading plus 0.05%
of the full scale.
The CE- 6710 has six current generation channels. The maximum amplitude that can
be generated in each current channel is 32 A rms, the maximum power for each channel
is 1 3 0 W / 2 1 0 VA and the maximum voltage is 10 V rms. So that members can reach a
current with a maximum amplitude of 192 A rms with a maximum power of 8 00W / 11 00
VA. The amplitude accuracy is less than or equal to 0.04% of the reading plus 0.01% of the
full scale. For the angle, the accuracy is less than or equal to 0.05°.
In the DC generation, the precision is defined by 0.1% of the reading plus 0.1% of
the full scale.
The CE-6710 has twelve channels for measuring voltage and / or current. When
configured as voltage, these channels are used to process the acquisition of voltage signals,
for the analysis of power quality made by the measurement software. There are several
measurement ranges (200mV rms with a tolerance of 0.3% AC and 0.45% DC / 2V rms with
a tolerance of 0.15% AC and 0.25 % DC / 20Vrms with a tolerance of 0.15 % AC and
0.25 % DC , and 600V rms with tolerance 0.15 % CA and 0.25 % DC, for voltages up to
400Vac) to better match the aquisitadas voltage amplitude levels in suitable for processing
in the ADC. All the tolerances mentioned above are related to the full scale (range).
When configured as current, these channels are used to process the acquisition of
current signals, for the analysis of power quality made by the measurement software. The
current measurement can be done using clamp pliers, shunts, transducers, Rogowski coils or
any other transducer that has voltage output in the ranges described above. To do this,
simply inform the full scale and the transformation ratio of the transducer used. Optionally
CONPROVE can supply Clamp pliers to meet the customer's needs.
The CE-6710 has eight digital output interfaces (four relay outputs with potential-
free NO / NC contacts and four outputs in open collector or in 5V TTL voltage), in addition
to several outputs for sending GOOSE messages.
The CE-6710 has twelve digital input interfaces per NO / NC contact and voltage, in
addition to several inputs for receiving GOOSE messages.
Attention:
If it is at the binary input it is configured as wet, the voltage level used must be
higher than the adjustment level of the pickup (change level) of the input plus the maximum
precision limit of the input (according to the scale corresponding to the adjustment ). In most
cases, an easy way to respect this rule is to work with the pickup setting, at most, at half the
voltage level. For example, injected voltage of 100.0V is used as a “pickup” a value less
than or equal to 50.0V.
There are two connectors for coupling external hardware. Both the plug (S1) and
(S2) enables m control over six external analog channels each. External hardware that can be
coupled to the CE-7012 is the Amplifier CE-4000, the test set CE-7012, the test set CE
6710, the test set CE-6006 and the test set CE-6003 (all produced by CONPROVE).
Process generator for simulations of transducer outputs and low level DC signals,
both voltage and current. The current and voltage ranges are respectively ± 20 mA and ± 10
V. The accuracy is less than or equal to 0.01% of the reading plus 0.5% of the range.
Signal meter (DC) for testing voltage and current transducers. The current ranges are
± 2 mA / ± 25 mA with an accuracy of 0.05% of the range, ± 100 mA / 1.0A / ± 10 A with
an accuracy of 0.03% of the reading plus 0.08% of the range. In the case of voltage, the
ranges are ± 1V / ± 10 V with 0.05% accuracy of the range, ± 100mV
with 0.03 % accuracy plus 0.05 % of the range and ± 10m V with accuracy, 0.05% of
reading plus 0.15 % of range. These interfaces allow the connection of several models and
types of transducers existing on the market, whose response signals are given in currents or
DC voltages.
Auxiliary source:
Auxiliary source for generating DC voltage up to 300V. The level is adjusted in the
software, in the hardware configuration screen. Adjustment range is from 5 to 300Vdc with
typical accuracy of 2% of range.
The CE-7012 has six fixed voltage generation channels. The maximum amplitude
that can be generated in each voltage channel is 330 V rms, the maximum power for each
channel is 10 0 VA and the maximum current is 1.8 A rms. So that associates can reach up
to a voltage with maximum amplitude of 2000 Vrms or a maximum power of 400 VA. The
amplitude accuracy is less than or equal to 0.04% of the reading plus 0.01% of the range and
for the angle the precision is less than or equal to 0.05°.
In the DC generation, the precision is defined by 0.05% of the reading plus 0.05% of
the DC range.
For Primary tests, the voltage generation channels can be grouped in series allowing
the generation of:
The CE-7012 has 6 channels of current generation. The maximum amplitude that can
be generated in each current channel is 50 A rms, the maximum power for each channel
is 250W / 430 VA and the maximum voltage is 14 V rms. So that members can reach a
current with a maximum amplitude of 300 A rms or a maximum power of 1000W /1500
VA. The amplitude accuracy is less than or equal to 0.04% of the reading plus 0.01% of the
range (for generations up to 90% of the range) and for the angle the precision is less than or
equal to 0.05°.
In the DC generation, the precision is defined by 0.1% of the reading plus 0.1% of
the DC range.
For Primary tests, the current generation channels can be grouped in parallel
allowing the generation of:
The CE-7012 has twelve channels for measuring voltage and / or current. When
configured as voltage, these channels are used to process the acquisition of voltage signals,
for the analysis of power quality made by the measurement software. There are several
measurement ranges (200mV rms with a tolerance of 0.3% AC and 0.45% DC / 2V rms with
a tolerance of 0.15% AC and 0.25 % DC / 20Vrms with a tolerance of 0.15% AC and
0.25 % CO and 600Vrms with a tolerance of 0.15 AC, 0.25 % DC, for voltages up to
400Vac) to better match the aquisitadas voltage amplitude levels in suitable for processing
in the ADC. All the tolerances mentioned above are related to the full scale (range).
When configured as current, these channels are used to process the acquisition of
current signals, for the analysis of power quality made by the measurement software. The
current measurement can be done using clamp pliers, shunts, transducers, Rogowski coils or
any other transducer that has voltage output in the ranges described above. To do this,
simply inform the full scale and the transformation ratio of the transducer used. Optionally
CONPROVE can supply Clamp pliers to meet the customer's needs.
The CE-7012 has eight digital output interfaces (four relay outputs with potential-
free NO / NC contacts and four outputs in open collector or in 5V TTL voltage), in addition
to several outputs for sending GOOSE messages.
The CE-7012 has twelve digital input interfaces per NO / NC contact and voltage, in
addition to several inputs for receiving GOOSE messages.
Attention:
If it is at the binary input it is configured as wet, the voltage level used must be
higher than the adjustment level of the pickup (change level) of the input plus the maximum
precision limit of the input (according to the scale corresponding to the adjustment ). In most
cases, an easy way to respect this rule is to work with the pickup setting, at most, at half the
voltage level. For example, injected voltage of 100.0V is used as a “pickup” a value less
than or equal to 50.0V.
There are two connectors for coupling external hardware. Both the plug (S1) and
(S2) enables m control over six external analog channels each. The external hardware that
can be coupled to the CE-7012 are the Amplifier CE-4000, the test set CE-7012, the test set
CE 6707, the test set CE 6710, the test set CE-6006 and the test set CE-6003 (all produced
by CONPROVE).
Process generator for simulations of transducer outputs and low level DC signals,
both voltage and current. The ranges of currents and voltages are respectively ± 25 mA and
± 10 V. The accuracy is less than or equal to 0.01% of the reading plus 0.5% of the range.
Signal meter (DC) for testing voltage and current transducers. Current ranges are ±
2 mA / ± 25 mA with 0.05% accuracy of the full scale plus 0.05% of the point and ± 100
mA / 1.0A / ± 10 A with 0.03 accuracy % of reading plus 0.08% of the range. In the case of
voltage, the ranges are ± 1V / ± 10V with an accuracy of 0.05% of the range plus 0.05% of
the point, ± 100mV with an accuracy of 0.1% of the reading plus 0.1% of the range and ±
10m V with accuracy of 0.05% of the reading plus 0.2% of the range. These interfaces allow
the connection of several models and types of transducers existing on the market, whose
response signals are given in currents or DC voltages.
Auxiliary source:
Auxiliary source for generating DC voltage up to 300V. The level is adjusted in the
software, in the hardware configuration screen. Adjustment range is from 5 to 300Vdc with
typical accuracy of 2% of range.
Digital input /
voltage and current Special analog voltage Special secondary / primary analog
measurement. and current output. input for voltage and current.
.
Figure 6 – Hardware – CE-7024 panel
The maximum amplitude that can be generated in each voltage channel is 330 V rms,
the maximum power for each channel is 100 VA and the maximum current is 1.8 A rms. So
that associates can reach up to a voltage with maximum amplitude of 2000 Vrms or a
maximum power of 400 VA. The amplitude accuracy is less than or equal to 0.04% of the
reading plus 0.01% of the range and for the angle the precision is less than or equal to 0.05°.
In the DC generation, the precision is defined by 0.05% of the reading plus 0.05% of
the DC range.
For Primary tests, the voltage generation channels can be grouped in series allowing
the generation of:
The maximum amplitude that can be generated in each current channel is 50 A rms,
the maximum power for each channel is 250W / 430 VA and the maximum voltage is 14 V
rms. So that members can reach a current with maximum amplitude of 800 Arms or
a maximum power of 5000W / 5000 VA. The amplitude accuracy is less than or equal to
0.04% of the reading plus 0.01% of the range (for generations up to 90% of the range) and
for the angle the precision is less than or equal to 0.05°.
In the DC generation, the precision is defined by 0.1% of the reading plus 0.1% of
the DC range.
For Primary tests, the current generation channels can be grouped in parallel
allowing the generation of:
The CE-7024 has twelve channels for measuring voltage and / or current. When
configured as voltage, these channels are used to process the acquisition of voltage signals,
for the analysis of power quality made by the measurement software. There are several
measuring ranges (200mV RMS with a tolerance of 0.3% and 0.45% DC AC / 2V rms with
a tolerance of 0.15% CA and 0.25 % DC / 20Vrms with a tolerance of 0.15 % BC and
0.25 % DC 600Vrms and tolerance of 0.15 % CA and 0.25 % DC, for voltages up to
400Vac) to better match the aquisitadas voltage amplitude levels in suitable for processing
in the ADC. All the tolerances mentioned above are related to the full scale (range).
When configured as current, these channels are used to process the acquisition of
current signals, for the analysis of power quality made by the measurement software. The
current measurement can be done using clamp pliers, shunts, transducers, Rogowski coils or
any other transducer that has voltage output in the ranges described above. To do this,
simply inform the full scale and the transformation ratio of the transducer used. Optionally
CONPROVE can supply Clamp pliers to meet the customer's needs.
The CE-7024 has eight digital output interfaces (with four outputs to relay contacts
NO / NC potential-free and four outputs in open collector or TTL voltage 5V), in addition,
several outputs for sending messages GOOSE.
The CE-7024 has twelve digital input interfaces per NO / NC contact and voltage, in
addition to several inputs for receiving GOOSE messages.
Attention:
If it is at the binary input it is configured as wet, the voltage level used must be
higher than the adjustment level of the pickup (change level) of the input plus the maximum
precision limit of the input (according to the scale corresponding to the adjustment ). In most
cases, an easy way to respect this rule is to work with the pickup setting, at most, at half the
voltage level. For example, injected voltage of 100.0V is used as a “pickup” a value less
than or equal to 50.0V.
There are two connectors for coupling external hardware. The first connector (S1)
allows the control of more external analog channels and the second connector (S2) allows an
increase in the number of digital inputs and outputs. The external hardware that can be
coupled to the CE-7024 are the Amplifier CE-4000, the Test set CE-6707, the Test set CE-
6710, the Test set CE-6006 and the Test set CE-6003 (produced by CONPROVE).
Process generator for simulations of transducer outputs and low level DC signals,
both voltage and current. The ranges of currents and voltages are respectively ± 25 mA and
± 10 V. The accuracy is less than or equal to 0.01% of the reading plus 0.5% of the range.
Signal meter (DC) for testing voltage and current transducers. The current ranges are
± 2 mA / ± 25 mA with an accuracy of 0.0 5 % of the range and ± 100 mA / 1.0A / ± 10 A
with an accuracy of 0.03% of the reading plus 0.08 % of the range. In the case of voltage the
ranges are ± 1V / ± 10V with 0.03% accuracy of the range, ± 100mV with 0.03% accuracy
of the reading plus 0.05% of the range and ± 10m V with accuracy of 0.03% of the reading
plus 0.15% of the range. These interfaces allow the connection of several models and types
of transducers existing on the market, whose response signals are given in currents or DC
voltages.
Auxiliary source:
Auxiliary source for generating DC voltage up to 300V. The level is adjusted in the
software, in the hardware configuration screen. Adjustment range is from 5 to 300Vdc with
typical accuracy of 2% of range.
First make sure that your computer has the minimum requirements for the software
to work properly.
Windows: 7, 8 or 10.
Framework: 4.8 or higher.
Processor: minimum Intel CORE i3.
Memory (RAM): minimum 4GB;
Operating System: 64 bits.
If the computer was supplied with the equipment, the software will already be
properly installed at the factory. However, the software installation DVD is always
provided. In this DVD, the user will find in 3 different directories, the setup for installing the
CTC applications, the setup of the tutorials and the setup for installing the communication
drivers, having the freedom to install each one separately. However, in the first installation,
just insert the DVD and the following window will be shown:
Let marked all options check box that has come selected and click the
icon Install and follow only confirming the options that appear on the screen.
Choose the directory for installing the software. Try to choose the option already
suggested by the software. In the next window click Next again.
Choose the directory for installing the software. Try to choose the option already
suggested by the software. In the next window click Next again.
Choose the directory for installing the software. Try to choose the option already
suggested by the software. In the next window click Next.
Figure 25 – Installation – Message showing that there is no update for National Instruments
4.2 GENERAL
Set of software for various tests are:
4.2.1 MANUAL
Software that performs tests in manual mode at both the secondary and primary
levels. Here the user has total control of what he is generating and acquiring, demanding
more knowledge in the desired test. It is recommended for quick tests where the user wants
to check if the Test set equipment is really working.
4.2.3 CALIBRAC
Software used to perform the calibration of all generation and measurement channels
such as voltage and current of equipment CE-6003 / CE-6006 / CE-7012 / CE-7024. For
both voltage and current, the software allows you to choose a base value for the module
and frequency. It is recommended to keep the default value.
4.3 PRIMARY
Set of automatic software for testing at primary levels are:
4.3.1 CT
Specific software for testing current transformers (CT). Evaluates the transformation
ratio using current, voltage or low power signal. Measures the CT load (load), raises the CT
excitation curve in a fully automated way. Check the resistance of the windings performs
insulation tests and analyze the polarity.
4.3.2 VT
Specific software for testing potential transformers (PT). Evaluates the
transformation ratio using either voltage or low power signal. Performs the measurement of
TP loading (load). Check the resistance of the windings performs insulation tests and
analyze the polarity.
4.3.3 TRANSFORMER
Specific software for testing Power Transformers. Evaluates the transformation ratio
monophonically or three-phase. Checks the resistance of the windings and performs an
insulation test.
4.3.4 RESISTANCE
Specific software for measuring resistance. Perform testing on windings and has the
functionality of a micro meter.
4.3.5 PMASTER
Software used for testing equipment in general. It has the functionality of CT, VT,
TRANSFORMER, RESISTANCE software in single software.
4.4 SECONDARY
Set of automatic software for testing at secondary levels are:
4.4.1 DIFERENC
Specific automatic software for differential tests. In this test, the user only informs
the configuration and adjustment of the Test set equipment and the software performs all
generation and acquisition automatically. In DIFERENC, three types of tests are performed:
TEST SET, where it is evaluated whether the equipment configuration informed to the
software is correct, POINT TEST, where the user chooses points on the Idiff x
Irest curve and checks whether there is performance or not and o SEARCH TEST, where
the user chooses test lines n the differential curve in the Idiff x Irest plane to be searched
4.4.2 DIRECPOT
Specific automatic software for power direction tests. In this test, the user only
informs the configuration of the Test set equipment and the software performs all generation
and acquisition automatically. DIRECPOT performs three types of tests: SHOTTING,
where the user inserts points in the regions of operation and non-operation in order to test
the performance or non-performance as well as the time of operation, SEARCH, where the
user defines lines that cross the region of operation and non-operation in order to test the
starting values of the region of operation and the ROUTE, where a variation of powers is
inserted in order to evaluate the region of operation in a dynamic way.
4.4.3 DISTANC
Specific automatic software for distance testing. In this test, the user only informs the
configuration of the Test set equipment and the software performs all generation and
acquisition automatically. At DISTANC, two types of tests are carried out: POINT TEST,
where the user inserts points in regions inside and outside the protection zones in order to
test the performance or non-performance as well as the performance time of each zone,
SEARCH TEST , where the user defines lines that cut the protection zones in order to test
the limits of each operating zone.
4.4.4 MASTER
Software which can create sequence s voltage and current comprising: ramps
increase or decrease in frequency, module angle and angle module, and
the module frequency, df/dt and ATP generation and COMTRADE file. One can create
reviews of the pickup time and Test set greatness. It is recommended in the case where the
user wants to model a very complex situation for testing the equipment.
4.4.5 PSB-OoS
Specific automatic software for tests of power balance and loss of system
synchronism. In this test, the user only informs the configuration of the Test set equipment
and the software performs all generation and acquisition automatically. Two types of tests
are performed on the PSB-OoS: SYSTEM SIMULATION, where the user through an
equivalent system defined by some parameters creates an oscillation and verifies the
behavior of the relay that can block (power balance) or act (loss of synchronism),
TRAJECTORIES SIMULATION, where the user freely defines in an impedance diagram
containing the zones in which function 21 functions if he wants a trajectory that simulates a
power balance or a trajectory that simulates a loss of synchronism.
4.4.6 RAMPA
Software where it is possible to create voltage and current ramp where
the frequency increases or decreases, module, angle, angle and module, module
and frequency, df/dt. One can create reviews of the pickup time and Test set greatness. It is
recommended in the case where the user wants to test the frequency and voltage functions
that the equipment has.
4.4.7 RESTHARM
Automatic software specifically for testing the HARMONIC RESTRICTION
function. In this test, the user only informs the configuration of the Test set equipment and
the software performs all generation and acquisition automatically. In RESTHARM, three
types of tests are performed: POINT TEST, where the user chooses points on the Idiff x%
harm curve and checks whether there is a block or not blocking function 87, the SEARCH
TEST, where the user chooses test lines and the Differential curve in the Idiff x% harm
plane will be sought and the CROSSBLOCK TEST, where harmonic is generated only in
one or two phases in order to check if the relay blocks function 87 by “cross blocking”.
4.4.8 SEQUENC
Software where it is possible to create a large amount of voltage and current
sequences directly, by symmetrical components, including harmonics (up to the fiftieth
order), inter-harmonics, by symmetric components, by constant power, by faults, by
constant VI, by constant ZV, by Z from constant source and by intelligent mode. One can
create reviews of the pickup time and Test set greatness. It is recommended in the case
where the user is to model situations such as: reclosing, logic testing, logical selectivity and
etc.
4.4.9 SYNCHRONISM
Automatic software specific for synchronism tests. In this test, the user only informs
the configuration of the Test set equipment and the software performs all generation and
acquisition automatically. In SYNCHRONISM, three types of tests are performed:
TRIGGER, where the user inserts points in regions inside and outside the synchronism
zones in order to test the synchronization within a range of ∆V and ∆f and a certain fixed
angle, SEARCH, where the user defines lines that cut the synchronization zone keeping the
angle constant and varying voltage and frequency in order to test the limits of the
synchronization zone and the TRAJECTORY, where a variation of the three
quantities is inserted in order of evaluating the synchronism region, this last test is more real
and dynamic because it includes the angle variation.
4.4.10 SOBRECOR
Specific automatic software for directionality and overcurrent tests. In this test, the
user only informs the configuration of the Test set equipment and the software performs all
generation and acquisition automatically. At SOBRECOR, two types of tests are performed
(both for phase, negative and zero): PICKUP, where the user enters search points in order to
test the pickup value of both the timed curve and the curve (s) instant (s), TIME, where the
user defines lines that cut the overcurrent curves in order to test the operating times. It is also
possible to include the directionality feature in the PICKUP and TIME tests.
4.4.11 TRANSIENT
Software where it is possible to create a sequence to reproduce COMTRADE (. Cfg)
and ATP (.lis) files. One can create reviews of the pickup time and Test set greatness. It is
recommended in the case where the user wants to reproduce a real field situation
(COMTRADE) or coming from a simulation situation (ATP).
4.4.12 VOLTSPHZ
Automatic software specifically for overexcitation tests. In this test, the user only
informs the configuration of the Test set equipment and the software performs all generation
and acquisition automatically. In VOLTSPHZ, three types of tests are performed: TIME,
where the user defines lines that cut the overexcitation curves in order to test the operating
times, SEARCH, where the user defines lines that cut the VxHz curve with the objective of
testing the limits of the function pickup, in the TRAJECTORY, the user defines lines that
cut the VxHz curve with the objective of testing the limits of the function pickup, the detail
is that in this test both the voltage and the frequency vary, making the test more dynamic.
4.4.13 TRANSDUT
Software used to calibrate the following transducers: active power, reactive power,
apparent power, frequency , current, voltage, power factor, angle between voltages, angle
between voltage and current, angle between currents, average current, DC current, voltage
DC, DC power and line voltage.
4.4.14 MEDIDOR
Software used to perform the calibration of meters: multifunctional, Watt hour, Var
hour, VA hour, Q hour, V 2 hour, I 2 hour, V hour, I hour. Be it 3 wires or 4.
4.5 OTHER
Set of software for additional applications are:
4.6 MEASURAMENT
Set of software dedicated to measurement are:
4.6.1 MULTIM
Specific software for measuring voltage, current, power and impedance in addition to
harmonics up to 3000 Hz. Generates a report with the measurements taken at the moment
you stop acquiring data.
4.7 SETUP
Set of configurations of the test set equipment and the test being:
4.7.1 PREFERENCES
Software that allows you to choose the test directories, explanatory figures,
overcurrent curves, COMTRADE file, ATP file, RIO file and IEC61850 substations. You
can choose the default language for the report, add the company logo, choose colors and add
effects to the chart. Finally, you have the option to choose the default values for general
information, system, Notes & obs, Explanatory figures, Check List, Synchronization and
Others.
4.8 SUPPORT
Applications that are intended to assist the user of the test case.
4.8.1 TUTORIALS
The user accesses the folder with all the tutorials for secondary and primary tests.
4.8.2 CONTACT
The user checks the address, website, email, and telephone number of CONPROVE.
4.8.3 FORUM
The user accesses the electrical systems protection forum managed by CONPROVE.
4.8.4 MANUAL
The user accesses the instruction manual for the test cases.
4.8.6 AUTODIAGNÓSTICO
Software used to detect abnormalities in the test cases.
MANUAL
SOFTWARE FOR MANUAL TESTING ON
RELAYS AND TRANSDUCERS
The options on the Settings screen are described in detail in chapter 21.
Creates a new
test file.
Open a saved
file.
Password
protects certain
parameters of the
test file.
Saves the current
file.
In the start menu, several resources are available and are detailed in the following
figures. The following figure describes features of hardware configurations, editing of
results and generation control.
Opens a Channel Opens a screen Reissue Shows the tests performed or Initiates
Orientation screen. whether it is being performed
for Hardware the current at the current time.
signal
Configuration. test. generation.
Opens a screen Opens a screen for Insert a Exclude Deletes Stop signal
for GPS Setup. GOOSE and SV new test. s all the generation.
Configuration. Tests. current
Test.
Figure 36 – Manual – Home Menu – Hardware, Results and Generation
In the figure below, the resources for analysis options of the generated quantities are
shown, as well as the options for general adjustments and the creation of an
evaluation. Access to the report, options for viewing the units involved (time and module)
and general options for restoring and viewing the layout are also displayed.
Calls the Shows Show the Shows the Shows the Shows the Returns to
Settings the Harmonic time in time in the standard
Evaluation
Tab.
Screen. Wavefor tab. seconds. cycles. layout.
m Tab.
Enables /
disables tabs
for viewing.
Shows the Shows the Shows Shows Shows Shows values
Phasor configuration primary absolute relative based on the
Tab. of the report. values. values. values. secondary.
Position the selected Superimposes the Hide the windows Returns to Enables /
window: on the left; selected window. the standard
for better viewing. disables tabs
to the right, above, layout.
for viewing.
below or as a
document.
Hide the selected Hide the selected Close all open Closes all open
tabs. tabs. windows. documents.
Insert password in Opens the Enables binary It separates the Shows the
the Test file so as Settings inputs and visualization of the manufacturer’s
not to allow the Screen when GOOSES inputs to analog signals and address, phone,
editing of certain starting the perform the test. the binary outputs in email and fax.
two tabs.
settings. software.
It is also opened at software startup, if the Opens Settings when Starting option in
the Software Options tab is selected.
In General mode , there is the General Information tab where the user enters all the
data related to the test, the Test set device and the person responsible for the test. There is
also the Systems tab where the user informs the nominal values of frequency,
voltage, current and power, among others. And, there is a tab for Notes and Observations for
miscellaneous records.
There is an Explanatory Figures tab where the user can insert up to three figures
relevant to the test, a Check List tab to create steps to be followed during the test, there is a
tab called Others for configuring digital signal filters and finally a tab Connections allowing
the user to structure all connections between the test case and the equipment to be Test set.
In the figures below, all the ways to insert a pre-fault will be shown.
Choice of
Module operating
value. mode.
Generic
Frequency
point of the
value.
System.
Angle
Description of value.
the targeted
channel.
Figure 43 – Manual – Pre-fault tab – Sub tab NO – Description of the Direct Mode
Default number of
Voltage input mode. decimal places (two
places).
Current input mode.
Reduced number of
decimal places.
Frequency entry User-defined number
mode. of decimal places.
Choice of
Description of
the targeted
operating
channel. mode.
Figure 46 – Manual – Pre-Fault tab – Sub tab NO – Description of the Harmonics mode
Harmonic Order.
Pre-fault duration
Voltage and Enable pre- time.
current fault.
values that
will be
generated to Choice of
meet the operating
condition of mode.
symmetrical
component.
Figure 49 – Manual – Pre-Fault tab – Sub tab NO – Description of the Symmetric Components mode
Definition of
Values of phase-to-phase
the Phase-
voltage, Sequence zero
Phase value. voltage and current in
the three phases.
Figure 51 – Manual – Pre-Fault tab – Sub tab NO – Description of the Phase-Phase mode
CONPROVE INDÚSTRIA E COMÉRCIO LTDA 96
CTC SOFTWARE MANUAL
Enables / disables
Pre-fault.
Definition of
the Power Phase-to-ground
values. voltage values, single-
phase active and
reactive powers.
Figure 53 – Manual – Pre-Fault tab – Sub tab NO – Description of the Power mode
Enables / disables
Pre-fault.
Defining the
fault values. Fault types (three-
phase, two-phase or
single-phase), angle
between voltage and
current and module /
angle of fault voltage
and fault current.
Figure 57 – Manual – Pre-fault tab – Sub tab NO – Description of the constant ZI mode
Figure 59 – Manual – Pre-fault tab – NO sub tab – Description of the constant ZV mode
Pre-fault
duration time.
DC Voltage
Value.
DC current
value.
Figure 66 – Manual – Sequence Tab – Output Analog. DC Sub tab – Description General
Monitored
binary input.
Binary actuation
Monitored time.
GOOSE input.
Waiting time
Disregards the between
first operation. stopwatches.
Clears all
Manual increment
of the amplitude or incremental
angle. values.
Clicking on the Show Columns option in the previous figure brings up the following
options:
It allows 4 different
sizes for visualization
of the graphs.
Consideration for
Evaluation between
module / angle, R / X, Rs / Constant Tolerance
Rp, Ls / Lp, Powers. It multiplied taken into
Channels that Theoretical
depends on what is by the Ratio account in the
will be parameterized on the obtained. value for
Evaluation.
compared. “Type” screen. Evaluation.
6 Auxiliary Source
It is software specific to control the source helper. One can adjust the value of the
output voltage to a value of 265 V (60 W) for equipment CE-6003 / CE6006 and 300 V
(60W - planned) to the equipment CE-7012/ CE-7024. The detail in this software is that it is
opened automatically every time other software is closed, so that the user always has control
of the voltage. Below are the details of the software settings:
Value
measured
at the
source
Starts or output and
stops its
generatin respective
g voltage graphic
from the indication.
source.
CALIBRAC
SOFTWARE FOR CALIBRATION OF CE-
6003/CE-6006/CE-6707/CE-6710/CE-7012/CE-7024
Serial number of the equipment. Displays the channel to Displays the time and date
be calibrated. of the last saved calibration.
The user must read the generated value in the standard and type it in the input
box. For example, suppose the voltage base chosen is 100 V and the standard indicates 98 V,
the user must type 98 in the text box and click on the “Reset” button. The voltage level
immediately increases, approaching the chosen 100 V. The calibration will be correct if the
voltage level chosen, in this case 100 V, is equal to the voltage level actually generated, that
CONPROVE INDÚSTRIA E COMÉRCIO LTDA 117
CTC SOFTWARE MANUAL
is, when the user reads 100 V in the standard. When the value is equal to the chosen one,
100 V, it means that the desired generation level has been reached and the calibration
process can now move on to the next step. The calibration of the acquisition gains the above
process is repeated for the frequency s harmonic orders 10, 20, 30, 40 and 50.
Defines
base
frequency:
50/60 Hz.
Base
amplitude
value used
in the
calibration.
Base frequency Amplifier being Describes the Field where the value read in
defined. calibrated. step in progress. the standard must be entered.
Base
Amplitude
value defined.
Cancels the
calibration process.
In the next step, the “Correcting Measurement Linearity” process begins. In this step,
the linearity correction of the amplifiers Transducers will be done in five amplitude levels
for the same base frequency , the user must read the measured value in the standard and
enter it only once for each amplitude level, and click on the button "Readjust".
Base value of Base frequency Amplifier being Mostra o nível de tensão cuja
Defined amplitude. defined. calibrated. Linearidade está sendo corrigida e
Ca Process descreve o processo.
steps
completed.
sem
Salva a
calibração da
AmplitListField
of where the
value read in
readjusted the standard
earnings. must be
entered.
i o
arquivo de
Reajusta a linearidade
calibração Cancels the
para o nível de calibration.
anterior).
amplitude dado.
altera
r a calibração
Indicates the process step in progress (red arrow)
and completed steps (in green). anterior.
After completing the steps above, the screen illustrated in the following figure will
be active starting the automatic step of the process.
Base Amplitude
value defined.
Cancels the
calibration.
Automatic process
status bar.
Figure 87 – Calibrac – Adjustment of the multipliers of the 2nd the 50th harm order.
CONPROVE INDÚSTRIA E COMÉRCIO LTDA 120
CTC SOFTWARE MANUAL
At the end of the process, a list of the earnings and values generated is
presented. The user can save these gains or cancel the calibration.
Base frequency Amplifier that has Cancels the calibration process, without
changing the previous calibration.
defined. been calibrated.
Base
Amplitude
value.
List of
readjusted
earnings.
Shows connection
scheme for
calibration of the
lags.
Displays stage
of the process Cancels the
in progress. calibration process.
The previous screen is active when the user selects the option on the main screen:
“Voltage and Current Generation - Lag”, and clicks on “Calibrate”.
Calibration of the Gap between Amplifiers is done in stages. In the first, the lags of
the voltage channel measurement transducers are calibrated, in the next the same is done for
the current channels, both made automatically after the user clicks the “Start” button.
Base frequency
defined.
Description of the
Indicates process step in progress.
Process step
that has
been
completed.
After processing these steps, the software is preparing to correct the gap between
voltage and current, it is necessary for the user to enter the value shown in the standard
and presses the “Reset” key.
In the last step of the process, the software automatically calibrates the lags of the
voltage and current generations in relation to channels V1 and I1.
At the end of the process, a list of the measured and generated lag values is
presented. The user can save these gains or cancel the calibration, keeping
the old calibration.
Base frequency
specified.
Cancels the
calibration process.
Figure 91 – Calibrac – Stage “Correcting the gap between voltage and current”
Base frequency
defined.
Cancels the
calibration
process,
without
changing the
previous
calibration.
Completed steps.
Defined current
base amplitude.
Scheme of equipment
connection for this
Cancels the
process.
calibration
process.
When pressing the “start” button, the transitory mode calibration process begins. The
user must enter the current value measured in the standard ammeter, until it is equal to the
base amplitude value, when this occurs, the software moves on to the next step. This process
is repeated for currents I1 through I6.
Base frequency
defined.
Cancels the
calibration
process.
Then the software moves to the Calibration Phase Generation Gain step in Transient
mode.
When you click on “Start”, the automatic steps to calibrate the Generation Gains
begin.
At the end of the calibration process, a list of gains and the values generated based on
these gains is presented. The user can save these gains or cancel the calibration
Describes
the step in
Base frequency progress
defined. and gives
instructions
to the user.
Base current amplitude
value.
Cancels
Show Scheme of the
equipment connection calibration
for this process. process.
Starts calibration
process.
Cancels the
calibration
process,
without
changing
the previous
calibration.
Saves readjusted lag
values.
Completed steps.
When clicking on the “Start” button, the step “Calibrating Aqs Gain and Def. In
Fundamental” is enabled. The user must inform the value read on the standard multimeter
and press the “Reset” button. This procedure is repeated for the frequency’s harmonic order
of 10, 20, 30, 40 and 50.
Then the linearity correction step will be enabled, which will be carried out in five
voltage levels, the user must inform for each level the value read on the standard multimeter
and click on "Reset".
At the end of the process, a list of the gains, corrections and lags of the Voltage
Measurement is presented. The user can save this data or cancel the calibration.
Base frequency
defined.
Cancels the
Indicates step in progress (red), calibration
steps completed (green). process.
Figure 99 – Calibrac – Stage “Cal. mild Gain of Aqs. and Defas. in Fundamental Frequency”
Base voltage
amplitude
value.
Describes the
step in
Base frequency progress and
gives
defined. instructions to
the user.
Cancels the
Values read on calibration
the standard process.
multimeter.
Calibrated
voltage range.
Base frequency
defined.
The user must leave the transducer calibration input that will be calibrated open, and
press the “Correct Offset” button, the correction will be carried out automatically.
In the next step, the user must connect the Transducer calibration input to a
controlled DC voltage source, or he can still use the CE6006 / 3 auxiliary source, as long as
it is conditioned to the maximum level of 10V at the input (for this use a resistive divider)
for calibration of Voltage calibration input V dc; or a controlled DC current source if the
calibration is from the mA dc current calibration input.
The step for calibrating the Acquisition Gains will be enabled. The user must inform,
for each suggested DC amplitude level, the value read on the standard DC multimeter /
milli-ammeter and press the “Reset” button.
After readjusting the 5 levels of the Positive Response, the user must invert the
polarity of the source and press the "Start" button.
For each of the levels of Negative answer, the user must inform the value read in the
standard and press the “Reset” button.
At the end of the process, a list of the gains and corrections obtained will be
presented. The user can save these gains and corrections, or cancel the calibration.
Figure 103 – Calibrac – Stage “Calibrating acquisition gain to level approx. 2V”
Process steps
completed.
The user must wait for the Source to stabilize and then inform the measured value by
clicking and then on the “Reset” button. The procedure should be repeated for nine levels.
At the end of the process, a list of the generated gains and corrections will be
presented. The user can save these gains and corrections or cancel the calibration.
Shows the
connection
Describes the step in progress and
scheme of the gives instructions to the user.
equipment.
Indicates Process
step in progress.
Cancels the
calibration
Starts calibration
process.
process.
Saves readjusted
gains and
corrections.
Process steps completed.
Cancels the
calibration
process.
TEST PLAN
SOFTWARE TO PERFORM SEVERAL TESTS
IN SEQUENCE
It is possible to test all functions of a relay using a single test file. This software
opens the tests to be done by other applications in sequence, automating the way of testing
the equipment.
This software has the feature “ Insert Test Plan when starting ” (Software
Options menu) , if it is active, when starting the Test Plan software , the option “Test Plan 1”
is already created automatically.
Creates a new
test file.
Open a saved
file.
Password protect
certain
parameters of the
test file.
Saves the current
file.
In the start menu, several resources are available, detailed in the following figures.
The following figure describes hardware configuration features and generation controls such
as start, pause, end, next test module, next test plan and partial or total exclusion.
Paste a test Cut out a test Starts the Pause or stop Go to the next Clears the
module. module. signal signal test module. selected
generation generation. module and
process. returns to the
status of Not
Tested.
Deletes the selected Copies the Move on to the Clears all modules and
test plan or module. selected next test plan. returns to the status of
module. Not Tested.
In the figure below, the features of inserting a test plan, inserting adjustments,
inserting hardware configuration and inserting test modules are shown. Access to the report,
options for viewing the units involved (time and module) and general options for restoring
and viewing the layout are also displayed.
Enables /
disables
tabs for
viewing.
Figure 115 – Test Plan – Home Menu – Options, Report, Units and Layout
Position the selected Superimposes the Hide the windows Returns to Enables /
window: on the left; selected window. for better viewing. the disables tabs
to the right, above, standard for viewing.
below or as a layout.
document.
Hide the selected Hide the selected Close all open Closes all open
tabs. window. windows. documents.
Insert password in Enables / disables the Opens the Settings Shows the
Screen when the manufacturer’s address,
the Test file so as simulation mode. Since
software is started. phone, email and fax.
not to allow the in this mode there is no
editing of certain insertion of voltage and
settings. current.
REMOTE
GENERATION
SOFTWARE TO CONTROL OTHER TEST SET
It is defined if the Identification created The user can create a Allows remote
connection is by the software itself. password to allow remote access.
through the internet access.
or local.
METER
SOFTWARE FOR TESTINGS METERS
The user must configure the tests specifying which generation channels are directed
to the voltages and currents, he can also enable a reference table to correct the generation
errors, in this case, the reference is no longer the generation of the suitcase but yes that table.
The tests available are: Load test, mechanism, injection, no load and creep. For all
tests it is possible to visualize the waveforms, phasors and harmonics.
This software has the feature “Open Settings at Start” (Software Options menu) , if it
is active, when starting the Meter software the Settings screen is active immediately, this
feature is interesting since, before starting the tests it is Important to specify the parameters
related to the test.
The options on the Settings screen are described in detail in chapter 21. The Meter
field will be discussed throughout this chapter.
Creates a new Opens a saved Saves the Customize the Software name
test file. Test. current test. Quick Access and version.
Bar.
Equipment used
Opens the options: and serial
Restore, Move, Size,
number.
Minimize, Maximize
and Close.
Manufacturer data is
displayed.
Shut down the software.
In the start menu, several resources are available. The following figure describes
hardware configuration features and generation controls such as start, stop, move to the next
point / line, deletion of one or all of the Test set points.
Opens a Channel Opens a screen Starts the Pauses signal Move on to the
Orientation screen. for Hardware signal generation. next point.
Configuration. generation
process.
Clears the
selected
point and
returns to
the status
of Not
Tested.
Connects between Opens a screen for Setup of sample Clears all points and
software and GPS Setup. value and goose returns to untested status.
hardware. messages.
In the figure below, the static generation, the resources for analysis options of the
generated quantities (waveforms and phasors), as well as the adjustment options (General
Settings and Meter) are displayed. Access to the report is also displayed, options for viewing
the units involved (time and module) and general options for restoring the layout, viewing
the layout and recreating graphs.
Starts the Shows the Shows Shows the time Shows the error Recreates Enables /
static Waveform the in seconds or or accuracy of the disables tabs
generation Tab. Phasor cycles. the meter. graphics. for viewing.
mode. Tab.
Opens the
report Shows values
Calls the configuration based on the Shows absolute Restores the
Settings Shows window. primary or or relative default layout
Screen. Harmonic Tab. secondary. values. of the screen.
Figure 127 – Meter – Home Menu – Static generation, Options, Report, Units and Layout
Shows the
Voltage, current, angle, Waveform
power factor and power Tab and
values to be generated. phasors.
Position the selected Superimposes the Hide the windows Returns to Enables /
window: on the left; selected window. for better viewing. the disables tabs
to the right, above, standard for viewing.
below or as a layout.
document.
Hide the selected Hide the selected Close all open Closes all open
tabs. window. windows. documents.
Shows the
manufacturer’s
address, phone,
email and fax.
It is also opened at software startup, if the Opens Settings when Starting option in
the Software Options tab is selected.
In the General mode, there is the “General Information” tab, where the user enters all
the data relating to: the Test, the Test Device and the person responsible for the
Test. There is also the tab “Reference Values”, where the user informs the nominal values of
voltage, current and power for system 1. There is also a tab for “Notes and Observations”. In
the other tab, “Explanatory Figures”, it is possible to insert up to 3 figures in the most used
image extensions, such as: Bitmap; JPGE; PNG; JPG; etc.
In Meter mode, there is the “Meter Adjustment” tab. In it the user chooses the type of
meter. It allows the user to adjust the export angle, the directionality mode (bidirectional,
importing or exporting), and the analyzed quadrant. It also allows the user to enter specific
data, such as: Meter class, pulse / rotation (electromechanical meters).
Other resources are related to the constants inherent to the meter (pulse / quantity or
quantity per pulse) in primary or secondary format using the voltage and current
transformation ratios already parameterized in the “General” tab.
Extract the
COMTRADE or CSV
Allows you to Insert or file from the waveforms.
Edit either Markings or Take a photo of the chart
Comments in graphics. by storing it on the
clipboard.
Clicking on the Display option in the previous figure brings up the following options:
Clicking on the Show Columns option in the previous figure brings up the following
options:
Enables or disables the
display of the
following values:
Module, angle, actual
value, imaginary value,
lag, instantaneous
amplitude, frequency
and frequency
variation.
10.15 Test
The test of the transducer aims to inject voltage and current and measure, through the
CE-OSU, the amount of pulses / energy in a given time, comparing with theoretical and
tolerances.
The previous figure shows the result of a test of an active power meter. For a better
explanation of this test, this figure is divided into parts.
For each Test set point the user can choose what type of data will be shown in the
table according to the figure below.
Shows the number of Reports nominal values Warns the meter Provides voltage
phases in Test, the of frequency, number tolerance and the and current modulus
current voltage value of pulses, energy and error found. and angle values.
and the power factor. time.
Reports the start Shows measured Warns the standard Provides the
and end date and values of frequency, deviation for Test module and angle
time of the test. number of pulses, Points more than of the line
energy and time. once. voltages.
Removes
selected point.
When the user clicks on “New Point” the following fields can be parameterized.
Creep current
and tolerance
considered. Current,
voltage and
power factor
Harmonic values to be
injection. adjusted.
Active,
reactive and
apparent
power
values to be
adjusted.
Load - The user can set the number of pulses, the total energy generated or the
amount of time that the load will be applied. Current and voltage is injected into the meter
through the test case and the number of pulses emitted by the meter is captured by the CE-
OSU. At the end of the software will evaluate the amount of pulses captured s with a par.
Mechanism - In this test, pulses are not read through the CE-OSU. In this case, the
user will inform the software of the initial reading on the meter, start the generation of an
energy previously stipulated and after the generation the user again informs the software of
the value of the new energy read. After this process, the software processes and validates the
Test set point.
Injection - In this test, only the energy or generation time is stipulated for applying
voltage and current to the meter. No type of evaluation is done, as the purpose is only to
check the functionality of the equipment. The final evaluation of the software is given
as “Generated”.
No Load - The test case injects only voltage into the meter or a current outside the
meter's directionality or functionality (for example, if it is a Wh meter, a fully inductive or
capacitive current is injected). The test is passed if no pulse is measured by the test
bag. There is also the possibility of considering that a pulse is emitted during this test.
Creep - The user defines in the software the minimum current in percentage of the
nominal current so that the meter is sensitized and sends a pulse. The test is processed and
the meter will be evaluated positively if the pulse is emitted within the expected time
In the “Test Settings” tab, the user directs the voltage and current generation
channels, controls the binary outputs, directs the measurement acquisition channels, has the
option of considering the total RMS value in the calculations, adjusts the trigger interface
and you can load a reference table for generation errors
Directs the
binary outputs to
each phase.
This association
is vital to make Enables /
the test possible. disables the
calculation of
the RMS value
in the single-
phase test.
Figure 150 – Meter – Main screen – Test Settings – Direction. Generation Channels
POWER
QUALITY
SOFTWARE FOR TESTING ON POWER
QUALITY ANALYZER
The user must configure the tests specifying which generation channels are directed
to voltages and currents, configure the states of the binary outputs at the time of pre-
incident, incident and post-incident. Configure the generation trigger method and still enable
the use of binary inputs if necessary.
This software has the feature “Open Settings when Start” (Software Options menu),
if it is active, when starting the Power Quality software the Settings screen is active
immediately, this feature is interesting since, before starting the tests it is important to
specify the parameters related to the test.
The options on the Settings screen are described in detail in the chapter “General
Settings”.
Create a new Open a saved Save current Customize the Software name
test file. test. test. quick access and version.
bar.
Opens the
options: Equipment used
Restore, and serial
Move, Size,
number.
Minimize,
Maximize
and Close.
Password protect
certain test file
parameters.
Enable/disable options for:
Simulation Mode;
Open Settings when Start.
Manufacturer data is
displayed.
Closes the software.
In the home menu several features are available. The following figure describes
hardware configuration resources and generation controls such as start, stop, move to the
next point/line, delete one or all tested points.
Opens a Channel Opens a Hardware Starts signal Pause signal Go to the next
Steering screen. Configuration generation generation. point.
screen. process.
Clears
selected
point and
returns to
Untested
status.
Makes connection Opens a GPS Setup Configuration of Clears all points and
between software screen. GOOSE and SV returns to Untested status.
and hardware. messages.
In the figure below, the resources of the analysis options of the generated quantities
(waveforms, phasors and harmonics) are shown, as well as the General Settings options. It
also displays access to the report, viewing options for the units involved (time and module)
and general options for restoring the layout, viewing layout and recreating graphics.
Opens the
report Show values
Calls the configuration based on Show absolute Restores the
Settings Shows the window. primary or or relative default Screen
screen. harmonics. secondary. values. Layout.
Figure 159 – Power Quality – Home Menu – Options, Report, Units and Layout
Position the selected Overrides the Hides windows for Return to Enable/disable
window: to the left; selected window. better viewing. default tabs for
on the right, above, Layout. viewing.
below or as a
document.
Rebuild the
graphics.
Hide selected tabs. Hides the selected Close all open Close all open
window. windows. documents.
Shows the
manufacturer's
address, phone,
email and fax.
Extracts the
COMTRADE or CSV
Allows Inserting or file from the waveforms.
Editing either Markup Take a picture of the
or Comments in chart storing in the
graphics. clipboard.
By clicking on the Display option in the previous figure, you reach the following
options:
Extracts the
COMTRADE or CSV
Allows Inserting or file from the waveforms.
Editing either Markup Take a picture of the
or Comments in chart storing in the
graphics. clipboard.
By clicking on the Display option in the previous figure, you reach the following
options:
Extracts the
COMTRADE or CSV
Allows Inserting or file from the waveforms.
Editing either Markup Take a picture of the
or Comments in chart storing in the
graphics. clipboard.
By clicking on the Display option in the previous figure, you reach the following
options:
Shows the
Trajectories.
By clicking on the Show Columns option in the previous figure, you reach the
following options:
11.13 Test
The power quality analyzer test can be done using several input signals by clicking
on the “New Point” button, the user has the following options.
An important detail is that the pre-incident and post-incident always use nominal
values of voltage, current and frequency.
Type of Incident.
Time of each
Incident.
Figure 178 – Power Quality – Test – Insert/ Edit Points – General Options – Geral – Frequency
Event name.
Minimum frequency
value during the Frequency value to open
Incident. event count.
Figure 179 – Power Quality – Test – Insert/ Edit Points – General Options – Definitions – Frequency
Type of Incident.
Time of each
Incident.
Figure 180 – Power Quality – Test – Insert/ Edit Points – General Options – Geral – Magnitude
Figure 181 – Power Quality – Test – Insert/ Edit Points – General Options – Definitions – Magnitude
Type of Incident.
Time of each
Incident.
Figure 182 – Power Quality – Test – Insert/ Edit Points – General Options – Geral – Flicker
Event name.
Flicker frequency
value.
Sinusoidal or
square flicker.
Values in absolute
Peak voltage form or percentage
and/or current of fundamental.
values.
Figure 183 – Power Quality – Test – Insert/ Edit Points – General Options – Definitions – Flicker
Type of Incident.
Time of each
Incident.
Figure 184 – Power Quality – Test – Insert/ Edit Points – General Options – Geral – Sag
Event name.
Total ramp
time.
Final value of
voltage and
limit voltage Final current and
for counting the limit current value
event. for event
counting.
Event time.
Figure 185 – Power Quality – Test – Insert/ Edit Points – General Options – Definitions – Sag
Type of Incident.
Time of each
Incident.
Figure 186 – Power Quality – Test – Insert/ Edit Points – General Options – Geral – Swell
Event name.
Total ramp
time.
Final value of
voltage and
limit voltage Final current and
for event limit current value
counting. for event
counting.
Event time.
Figure 187 – Power Quality – Test – Insert/ Edit Points – General Options – Definitions – Swell
Type of Incident.
Time of each
Incident.
Figure 188 – Power Quality – Test – Insert/ Edit Points – General Options – Geral – Interruption
Event name.
Total ramp
time.
Final value of
voltage and
limit voltage Final current and
for counting the limit current value
event. for event
counting.
Event time.
Figure 189 – Power Quality – Test – Insert/ Edit Points – General Options – Definitions – Interruption
Type of Incident.
Time of each
Incident.
Figure 190 – Power Quality – Test – Insert/ Edit Points – General Options – Geral – Transients
Event name.
Cutting moment angle.
Figure 191 – Power Quality – Test – Insert/ Edit Points – General Options – Definitions – Transients
Type of Incident.
Time of each
Incident.
Figure 192 – Power Quality – Test – Insert/ Edit Points – General Options – Geral – Unbalance
* Natural Components.
* Symmetrical Components.
* Rel. Sym. Components.
Figure 193 – Power Quality – Test – Insert/ Edit Points – General Options – Definitions – Unbalance
Type of Incident.
Time of each
Incident.
Figure 194 – Power Quality – Test – Insert/ Edit Points – General Options – Geral – Harmonics
Harmonic
order,
modulus
and angle
of analog Add or remove
signals. harmonic
orders.
Figure 195 – Power Quality – Test – Insert/ Edit Points – General Options – Definitions – Harmonics
Type of Incident.
Time of each
Incident.
Figure 196 – Power Quality – Test – Insert/ Edit Points – General Options – Geral – Inter-Harmonics
Event name.
Voltage or
current
channel.
Inter-
harmonics,
modulus
and angle Add or remove
of analog inter-harmonic
signals. orders.
Figure 197 – Power Quality – Test – Insert/ Edit Points – Definitions – Inter-Harmonics
Type of Incident.
Time of each
Incident.
Figure 198 – Power Quality – Test – Insert/ Edit Points – General Options – Geral – Quick Change
Event name.
Time of
change, initial Current initial
voltage value, value, current
voltage increment
increment value and
value and current change
speed of speed.
voltage
change.
Figure 199 – Power Quality – Test – Insert/ Edit Points – General Options – Quick Change
Figure 200 – Power Quality – Test – Insert/ Edit Points – General Options – Geral – Multi-Incident
Clicking on “New Point” and choosing “Assessment” inserts the instructions for a
given event to be approved or disapproved.
Figure 201 – Power Quality – Test – Insert/ Edit Points – General Options – Geral – Assessment
Directs the
voltage and
current
channels.
Binary Output
States in each
Incident.
Figure 204 – Power Quality – Main Screen – Test Settings – Direc. Generation Channel
CT
CURRENT
TRANSFORMER
TESTING
SOFTWARE
In it, the user defines, for each signal, a generation channel and a measurement
channel according to the configuration of the test equipment. This software is
only compatible for the CE-7012 and CE-7024.
Transformation Ratio;
Transformation Ratio (using tension);
Transformation Ratio for Rogowski coil;
Transformation Ratio for low power;
Burden or Load and Power Factor;
Lifting of the magnetization curve;
Winding resistance with temperature compensation;
Insulation test;
Polarity test.
The user has the possibility to carry out the capture of the tests, fully
automatically, so that the software controls the generation time and automatically captures
the points. The second semi-way automatic user controls generation software and capture
points and the third way the user manual controls the generation time and the time of capture
points.
The user can configure the adjustment screen to facilitate the preparation of the
report. There are icons to start or to generate the test.
This software has the feature “Opens Settings when Starting” ( Software Options
menu ) , if it is active, when starting the CT software the Settings screen is active
immediately, this feature is interesting since, before starting the tests it is Important to
specify the parameters related to the test.
The options on the Settings screen are described in detail in chapter 21.
Creates a new
test file.
Open a saved
file.
Password protect
certain
parameters of the
test file.
Saves the current
file.
In the start menu, several resources are available. The following figure describes
features of hardware configurations, editing of results and generation control.
Opens a screen Opens a screen for Insert a Excludes Deletes Stop signal
for GPS Setup. GOOSE and SV new test. all tests. the generation.
Configuration. current
test.
Figure 211 – CT – Home Menu – Hardware, Results and Generation
The figure below shows the general settings resources, access to the report and general
options for restoring and viewing the layout.
Position the selected Superimposes the Hide the windows Returns to Enables /
window: on the left; selected window. the standard
for better viewing. disables tabs
to the right, above, layout.
for viewing.
below or as a
document.
Hide the selected Hide the selected Close all open Closes all open
tabs. window. windows. documents.
Insert password in Opens the Enables binary It separates the Shows the
Settings Screen
the Test file in inputs and visualization of the manufacturer’s
when starting analog signals and
order not to allow the Manual GOOSES inputs to address, phone,
editing of certain perform the test. the binary outputs in email and fax.
software.
two tabs.
settings.
It is also opened at software startup, if the Opens Settings when Starting option in
the Software Options tab is selected.
In General mode , there is the General Information tab where the user enters all the
data related to the test, the device Test set and the person responsible for the
test. There is also the Systems tab, where the user informs the nominal values of frequency,
voltage, current and power, among others. And, there is a tab for Notes and Observations for
miscellaneous records.
There is an Explanatory Figures tab where the user can insert up to three figures
relevant to the test, a Check List tab to create steps to be followed during the test, there is a
tab called Others for configuring digital signal filters and finally a tab Connections allowing
the user to structure all connections between the test case and the equipment to be Test set.
12.9 TC Tab
The following will explain all the resources available in the sub- tabs.
12.9.1 Relationship
In this tab, the user verifies the transformation ratio of the CTs, both automatically
and manually. You can also perform load (Burden) and polarity tests. There is also a tab
with the connection diagram between the TC and the test case. This connection is changed
according to the configuration set in the CE-7012 or CE-7024.
In the figure below, all fields of the Ratio test are detailed.
Designation Nominal value Multiplying Value used Module and If enabled, the
of channels of the primary factor of in the test angle values measurements
for and secondary nominal and its (Manual Input). are entered
generation current. values. increment. manually.
and
measurement
.
Enable
automatic test.
Designation Nominal value Multiplying Value used Module and If enabled, the
of channels of the primary factor of in the test angle values measurements
for and secondary nominal and its (Manual Input). are entered
generation current. values. increment. manually.
and
measurement
.
Enable
automatic
test.
Designation of Nominal value of Multiplying Value used Module and If enabled, the
channels for the primary current factor of angle values
in the test measurements
generation and and secondary nominal (Manual Input).
voltage. and its are entered
measurement. values. increment. manually.
In the figure below, all fields of the low power ratio test are detailed.
Designation of Nominal value of Multiplying Value used Module and If enabled, the
channels for the primary current factor of angle values
in the test measurements
generation and and secondary nominal (Manual Input).
voltage. and its are entered
measurement. values. increment. manually.
12.9.5 Burden
In this test, an evaluation of the burden. This test can also be done together with the
transformation relation in the sub tab relation.
Designation of Nominal value of Multiplying Value used Module and If enabled, the
channels for the primary current factor of angle values
in the test measurements
generation and and secondary nominal (Manual Input).
voltage. and its are entered
measurement. values. increment. manually.
Enable
automatic
Maximum Burden test. Final test result showing Saves the
value and minimum loading and power factor. values during
power factor value. the test.
In the figure below, all fields of the magnetization curve test are detailed.
Enable
automatic test.
In the figure below, all fields of the Insulation test are detailed.
Enable
automatic test.
12.9.9 Polarity
In this tab, the user checks the polarity of the CT.
In the figure below, all fields of the Polarity test are detailed.
Designation of Nominal value Multiplying Value used Module and If enabled, the
channels for factor of angle values
of the primary in the test measurements
generation and nominal (Manual Input).
and secondary and its are entered
measurement. currents. values. increment. manually.
Saves the
Average values during
number of Enable automatic test. Polarity final result and whether the test.
purchases. it is approved or disapproved.
VT
SOFTWARE FOR TESTING POTENTIAL
TRANSFORMERS
In it, the user defines, for each signal, a generation channel and a measurement
channel according to the configuration of the test equipment. This software is only
compatible for the CE-7012 and CE-7024.
Transformation Ratio;
Transformation Ratio with electronic output;
Burden or Load and Power Factor;
Insulation test;
Polarity test.
The user has the possibility to carry out the capture of the tests, fully automatically,
so that the software controls the generation time and automatically captures the points. The
second semi-way automatic user controls generation software and captures points and the
third way the user manual controls the generation time and the time of capture points.
The user can configure the adjustment screen to facilitate the preparation of the
report. There are icons to start or to generate the test.
This software has the feature “Open Settings at Start” ( Software Options menu ) , if
it is active, when starting the VT software the Settings screen is active immediately, this
feature is interesting since, before starting the tests it is Important to specify the parameters
related to the test
The options on the Settings screen are described in detail in chapter 21.
Creates a new
test file.
Open a saved
file.
Password protect
certain
parameters of the
test file.
Saves the current
file.
In the start menu, several resources are available. The following figure describes
features of hardware configurations, editing of results and generation control.
Opens a screen Opens a screen for Insert a Exclude Deletes Stop signal
for GPS Setup. GOOSE and SV new test. s all the generation.
Configuration. tests. current
test.
Figure 240 – VT – Home Menu – Hardware, Results and Generation
The figure below shows the general settings resources, access to the report and general
options for restoring and viewing the layout.
Position the selected Superimposes the Hide the windows Returns to Enables /
window: on the left; selected window. the standard
for better viewing. disables tabs
to the right, above, layout.
for viewing.
below or as a
document.
Hide the selected Hide the selected Close all open Closes all open
tabs. window. windows. documents.
Insert password in Opens the Enables binary Enables / disables a Capture the
the Test file so as Settings inputs and filter via software to point and stop
not to allow the Screen when GOOSES inputs to eliminate the generation
editing of certain starting the perform the test. frequencies other
VT software. than what is desired.
settings.
It is also opened at software startup, if the Opens Settings when Starting option in
the Software Options tab is selected.
In General mode , there is the General Information tab where the user enters all the
data related to the test, the device Test set and the person responsible for the
test. There is also the Systems tab where the user informs the nominal values of frequency,
voltage, current and power, among others. And, there is a tab for Notes and Observations for
miscellaneous records.
There is an Explanatory Figures tab where the user can insert up to three figures
relevant to the test, a Check List tab to create steps to be followed during the test, there is a
tab called Others for configuring digital signal filters and finally a tab Connections allowing
the user to structure all connections between the test case and the equipment to be Test set.
13.9 TP tab
The following will explain all the resources available in the sub tab.
13.9.1 Ratio
In this tab, the user verifies the transformation ratio of the TPs both automatically
and manually. You can also perform burden and polarity tests. There is also a tab with the
connection diagram between the TP and the test case. This connection is changed according
to the configuration set in the CE-7012 or CE-7024.
In the figure below, all fields of the Ratio test are detailed.
Designation Nominal value Multiplying Value used Module and If enabled, the
of channels of the primary factor of in the test angle values measurements
for and secondary nominal and its (Manual Input). are entered
generation voltage. values. increment. manually.
and
measurement
.
Enable automatic
Adjustment of
test.
Burden and
power factor
values.
Next, all fields of the Ratio test with electronic output are detailed.
Designation of Nominal value Multiplying Value used Module and If enabled, the
channels for of the primary factor of in the test angle values measurements
generation and and secondary nominal and its (Manual Input). are entered
measurement. current. values. increment. manually.
13.9.3 Burden
In this test, an evaluation of the burden. This test can also be done together with
the transformation ratio in the sub tab.
In the figure below, all fields of the Burden test are detailed.
Designation of Nominal value Multiplying Value used Module and If enabled, the
channels for of the factor of in the test angle values measurements
generation and secondary nominal and its (Manual Input). are entered
measurement. voltage. values. increment. manually.
Enable
automatic test.
Adjustment of Final test result showing the Saves the
Burden and Burden and power factor values during
power factor found. the test.
values.
13.9.4 Insulation
In this mode, the user evaluates the insulation level.
In the figure below, all fields of the insulation test are detailed.
Enable
automatic test.
13.9.5 Polarity
This tab the user checks the polarity of TP.
In the figure below, all fields of the polarity test are detailed.
Designation of Nominal value of Multiplying Value used Module and If enabled, the
channels for the primary current factor of angle values
in the test measurements
generation and and secondary nominal (Manual Input).
voltage. and its are entered
measurement. values. increment. manually.
Saves the
values during
Average number Enable Polarity final result and whether the test.
of purchases. automatic test. it is approved or disapproved.
TRANSFOR-
MER
POWER TRANSFORMER TESTING
SOFTWARE
In it, the user defines, for each signal, a generation channel and a measurement
channel according to the configuration of the test equipment. This software is
only compatible for the CE-7012 and CE-7024.
Transformation Ratio;
Transformation Ratio in a three-phase manner;
Winding resistance with temperature compensation;
Insulation test;
The user has the possibility to carry out the capture of the tests, fully automatically,
so that the software controls the generation time and automatically captures the points. The
second semi-way automatic user controls generation software and capture points and the
third way the user manual controls the generation time and the time of capture points.
The user can configure the adjustment screen to facilitate the preparation of the
report. There are icons to start or to generate the test.
This software has the feature “Opens Settings when Starting” ( Software Options
menu ) , if it is active, when starting the Transformer software the Settings screen is active
immediately, this feature is interesting since, before starting the tests it is Important to
specify the parameters related to the test.
The options on the Settings screen are described in detail in chapter 21.
Opens the
options: Restore,
Move, Size,
Minimize, Equipment
Maximize and used and serial
Close. number.
Creates a new
test file.
Open a saved
file.
Password protect
certain
parameters of the
test file.
Saves the current
file.
In the start menu dive resources are available. The following figure describes
features of hardware configurations, editing of results and generation control.
Opens a screen Opens a screen for Insert a Excludes Deletes Stop signal
for GPS Setup. the current generation.
GOOSE and SV new test. all tests. test.
Configuration.
The figure below shows the general settings resources, access to the report and general
options for restoring and viewing the layout.
CONPROVE INDÚSTRIA E COMÉRCIO LTDA 236
CTC SOFTWARE MANUAL
Position the selected Superimposes the Hide the windows Returns to Enables /
window: on the left; selected window. the standard
for better viewing. disables tabs
to the right, above, layout.
for viewing.
below or as a
document.
Hide the selected Hide the selected Close all open Closes all open
tabs. window. windows. documents.
Insert password in Opens the Enables binary Enables / disables a Capture the
Settings Screen filter via software to
the Test file so as inputs and point and stop
when starting eliminate frequencies
not to allow the the Manual GOOSES inputs to other than what is the generation.
editing of certain software. perform the test. desired.
settings.
It is also opened at software startup, if the Opens Settings when Starting option in
the Software Options tab is selected.
In General mode, there is the General Information tab where the user enters all the
data related to the test, the device Test set and the person responsible for the
test. There is also the Systems tab where the user informs the nominal values of frequency,
voltage, current and power, among others. There is a tab for Notes and Observations for
miscellaneous records.
There is an Explanatory Figures tab where the user can insert up to three figures
relevant to the test, a Check List tab to create steps to be followed during the test, there is a
tab called Others for configuring digital signal filters and finally a tab Connections allowing
the user to structure all connections between the test case and the equipment to be Test set.
14.9.1 Relationship
In this tab, the user verifies the transformation Ratio of the power transformers both
automatically and manually. You can also perform winding and insulation
resistance tests. There is also a tab with the connection diagram between the TP and the test
case. This connection is changed according to the configuration set in the CE-7012 or CE-
7024.
In the figure below, all fields of the Ratio test are detailed.
14.9.2 Ratio – 3P
In this tab, the user verifies the transformation ratio of the TPs both automatically
and manually.
In the figure below, all fields of the three-phase ratio test are detailed.
In the figure below, all fields of the winding resistance test are detailed.
14.9.4 Insulation
In this mode, the user evaluates the insulation level.
In the figure below, all fields of the insulation test are detailed.
Enable
automatic test.
RESISTAN-
CE
RESISTANCE TESTING SOFTWARE
In it, the user defines, for each signal, a generation channel and a measurement
channel according to the configuration of the test equipment. This software is only
compatible for the CE-7012 and CE-7024.
Microhmmeter;
Winding resistance with temperature compensation.
The user has the possibility to carry out the capture of the tests, fully automatically,
so that the software controls the generation time and automatically captures the points. The
second semi-way automatic user controls the temp the generation and the software captures
points and the third way manual user controls the generation time and the time of capture
points.
The options on the Settings screen are described in detail in chapter 21.
Equipment used
Opens the and serial
options: number.
Restore, Move,
Size, Minimize,
Maximize and
Close.
Creates a new
test file.
Open a saved
file.
Password protect
certain
parameters of the
test file.
Saves the current
file.
In the start menu, several resources are available. The following figure describes
features of hardware configurations, editing of results and generation control.
Opens a screen Opens a screen for Insert a Exclude Deletes Stop signal
for GPS Setup. GOOSE and SV new test. s all the generation.
Configuration. tests. current
test.
Figure 280 – Resistance – Hardware, Results and Generation
The figure below shows the general settings resources, access to the report and
general options for restoring and viewing the layout.
Position the selected Superimposes the Hide the windows Returns to Enables /
window: on the left; selected window. the standard
for better viewing. disables tabs
to the right, above, layout.
for viewing.
below or as a
document.
Hide the selected Hide the selected Close all open Closes all open
tabs. window. windows. documents.
Insert password in Opens the Enables binary Enables / disables a Capture the
Settings Screen filter via software to
the Test file so as inputs and point and stop
when starting eliminate
not to allow the the Manual GOOSES inputs to the generation.
editing of certain perform the test. frequencies other
software.
than what is desired.
settings.
It is also opened at software startup, if the “Opens Settings when Starting” option in
the “Software Options” tab is selected.
In General mode there is the General Information tab, where the user enters all the
data related to the test, the device Test set and the person responsible for the test, there is the
Systems tab, where the user informs the nominal values of frequency, voltage, current and
power among others, there is a tab for Notes and Observations for various records.
There is an Explanatory Figures tab where the user can insert up to three figures
relevant to the test, a Check List tab to create steps to be followed during the test, there is a
tab called Others for configuring digital signal filters and finally a tab Connections allowing
the user to structure all connections between the test case and the equipment to be Test set.
CONPROVE INDÚSTRIA E COMÉRCIO LTDA 252
CTC SOFTWARE MANUAL
15.9.1 Micro-ohmmeter
In this tab the user checks the resistance values. There is also a tab with the
connection diagram between the resistance and the test case. This connection is changed
according to the configuration set in the CE-7012 or CE-7024.
In the figure below is detailed all the fields of the microhmmeter test.
Enable
automatic test.
In the figure below, all fields of the winding resistance test are detailed.
Enable
automatic test.
PMASTER
SOFTWARE FOR TESTING IN TCS, TPS,
POWER TRANSFORMER AND RESISTANCE
In it, the user defines, for each signal, a generation channel and a measurement
channel according to the configuration of the test equipment. This software is only
compatible for the CE-7012 and CE-7024.
Transformation Ratio;
Transformation Ratio (using voltage);
Transformation Ratio for Rogowski coil;
Transformation Ratio for low power;
Burden or Load and Power Factor;
Survey of the Magnetization Curve;
Winding resistance with temperature compensation;
Insulation test;
Polarity test.
Transformation Ratio;
Transformation Ratio with electronic output;
Burden or Load and Power Factor;
Insulation test;
Polarity test.
Transformation Ratio;
Transformation Ratio in a three-phase manner;
Micro-ohmmeter;
Winding resistance with temperature compensation.
The user has the possibility to carry out the capture of the tests, fully automatically,
so that the software controls the generation time and automatically captures the points. The
second semi-way automatic user controls generation software and captures points and the
third way the user manual controls the generation time and the time of capture points.
The user can configure the adjustment screen to facilitate the preparation of the
report. There are icons for starting or generating the test.
This software has the feature “Opens Settings when Starting” (Software Options
menu) , if it is active, when starting the PMaster software the Settings screen is active
immediately, this feature is interesting since, before starting the tests it is Important to
specify the parameters related to the test.
The options on the Settings screen are described in detail in chapter 21.
Equipment used
Opens the and serial
options: Restore,
Move, Size, number.
Minimize,
Maximize and
Close.
Creates a new
test file.
Open a saved
file.
Password protect
certain
parameters of the
test file.
Saves the current
file.
In the start menu dive resources are available. The following figure describes
features of hardware configurations, editing of results and generation control.
Opens a screen Opens a screen for Insert a Excludes Deletes Stop signal
for GPS Setup. the generation.
GOOSE and SV new test. all tests. current
Configuration. test.
The figure below shows the general settings resources, access to the report and general
options for restoring and viewing the layout.
Position the selected Superimposes the Hide the windows Returns to Enables /
window: on the left; selected window. the standard
for better viewing. disables tabs
to the right, above, layout.
for viewing.
below or as a
document.
Hide the selected Hide the selected Close all open Closes all open
tabs. window. windows. documents.
Insert password in Opens the Enables binary Enables / disables a Capture the
Settings filter via software to
the Test file so as inputs and point and stop
Screen when eliminate frequencies
not to allow the GOOSES inputs to other than what is the generation.
editing of certain starting the perform the test. desired.
software. tabs.
settings.
It is also opened at software startup, if the Opens Settings when Starting option in the
Software Options tab is selected.
In General mode , there is the General Information tab where the user enters all the
data related to the test, the Test set device and the person responsible for the
test. There is also the Systems tab where the user informs the nominal values of frequency,
voltage, current and power, among others. And, there is a tab for Notes and Observations for
miscellaneous records.
There is an Explanatory Figures tab where the user can insert up to three figures
relevant to the test, a Check List tab to create steps to be followed during the test, there is a
tab called Others for configuring digital signal filters and finally a tab Connections allowing
the user to structure all connections between the test case and the equipment to be Test set.
Allows access to Inform tabs. General, System, Notes & Obs.,
Explanatory Figures, Check List, Others and Connections.
16.9 TC Tab
The following will explain all the resources available in the sub tab.
16.9.1 Ratio
In this tab, the user checks the transformation ratio of the TCs both automatically and
manually. You can also perform Burden and Polarity tests. There is also a tab with the
connection diagram between the TC and the Test set. This connection is changed according
to the configuration set in the CE-7012 or CE-7024.
In the figure below, all fields of the Ratio test are detailed.
Designation of Nominal value Multiplying Value used Module and If enabled, the
channels for of the primary factor of in the test angle values measurements
generation and and secondary nominal and its (Manual Input). are entered
measurement. current. values. increment. manually.
Enable
automatic test.
In the figure below, all fields of the voltage relation test are detailed.
Designation of Nominal value Multiplying Value used Module and If enabled, the
channels for of the primary factor of in the test angle values measurements
generation and and secondary nominal and its (Manual Input). are entered
measurement. current. values. increment. manually.
Enable
automatic
test.
In the figure below, all fields of the Rogowski ratio test are detailed.
Designation of Nominal value of Multiplying Value used Module and If enabled, the
channels for the primary current factor of angle values
in the test measurements
generation and and secondary nominal (Manual Input).
voltage. and its are entered
measurement. values. increment. manually.
Figure 307 – PMaster –TC – Low Power Ratio and Connection Scheme
In the figure below, all fields of the low power ratio test are detailed.
Designation of Nominal value of Multiplying Value used Module and If enabled, the
channels for the primary current factor of angle values
in the test measurements
generation and and secondary nominal (Manual Input).
voltage. and its are entered
measurement. values. increment. manually.
16.9.5 Burden
In this test, an evaluation of the loading is performed (Burden). This test can also be
done together with the transformation relation in the sub tab relation.
In the figure below, all fields of the Burden test are detailed.
Designation of Nominal value of Multiplying Value used Module and If enabled, the
channels for the primary current factor of angle values
in the test measurements
generation and and secondary nominal (Manual Input).
voltage. and its are entered
measurement. values. increment. manually.
Enable
automatic test.
Maximum Burden Final test result showing Saves the
value and minimum loading and power factor. values during
power factor value. the test.
In the figure below, all fields of the magnetization curve test are detailed.
In the figure below, all fields of the winding resistance test are detailed.
Enable
automatic test.
In the figure below, all fields of the insulation test are detailed.
Enable
automatic test.
16.9.9 Polarity
In this tab, the user checks the polarity of the CT.
In the figure below, all fields of the polarity test are detailed.
Designation of Nominal value Multiplying Value used Module and If enabled, the
channels for factor of angle values
of the primary in the test measurements
generation and nominal (Manual Input).
and secondary and its are entered
measurement. values.
currents. increment. manually.
Saves the
values during
Enable Polarity final result and whether the test.
automatic it is approved or disapproved.
test.
16.10 TP tab
The following will explain all the resources available in the sub tab.
16.10.1 Ratio
In this tab, the user verifies the transformation ratio of the TPs both automatically
and manually. You can also perform load (Burden) and polarity tests. There is also a tab
with the connection diagram between the TP and the test case. This connection is changed
according to the configuration set in the CE-7012 or CE-7024.
In the figure below, all fields of the Ratio test are detailed.
Designation of Nominal value Multiplying Value used Module and If enabled, the
channels for of the primary factor of in the test angle values measurements
generation and and secondary nominal and its (Manual Input). are entered
measurement. voltage. values. increment. manually.
Enable automatic
test.
Figure 321 – PMaster –TP – Ratio with Electronic Output and Connection Scheme
In the figure below, all fields of the electronic ratio test are detailed.
Designation of Nominal value Multiplying Value used Module and If enabled, the
channels for of the primary factor of in the test angle values measurements
generation and and secondary nominal and its (Manual Input). are entered
measurement. current. values. increment. manually.
16.10.3 Burden
In this test, an evaluation of the loading is performed (Burden). This test can also be
done together with the transformation relation in the sub tab relation.
In the figure below, all fields of the Burden test are detailed.
Designation Nominal value Multiplying Value used Module and If enabled, the
of channels of the factor of in the test angle values measurements
for secondary nominal and its (Manual Input). are entered
generation voltage. values. increment. manually.
and
measurement
.
Enable
automatic
Adjustment of test. Final test result showing the Saves the
Burden and Burden and power factor values during
power factor found. the test.
values.
16.10.4 Insulation
In this mode, the user evaluates the insulation level.
In the figure below, all fields of the insulation test are detailed.
Enable
automatic test.
16.10.5 Polarity
This tab the user checks the polarity of TP.
In the figure below, all fields of the polarity test are detailed.
Designation of Nominal value of Multiplying Value used Module and If enabled, the
channels for the primary current factor of angle values
in the test measurements
generation and and secondary nominal (Manual Input).
voltage. and its are entered
measurement. values. increment. manually.
Saves the
values during
Enable Polarity final result and whether the test.
automatic test. it is approved or disapproved.
16.11.1 Ratio
In this tab, the user verifies the transformation ratio of the power transformers both
automatically and manually. You can also perform winding and insulation resistance
tests. There is also a tab with the connection diagram between the TP and the test case. This
connection is changed according to the configuration set in the CE-7012 or CE-7024.
In the figure below, all fields of the Ratio test are detailed.
16.11.2 Ratio – 3P
In this tab, the user verifies the transformation ratio of the TPs both automatically
and manually.
In the figure below, all fields of the three-phase relation test are detailed.
In the figure below, all fields of the winding resistance test are detailed.
16.11.4 Insulation
In this mode, the user evaluates the insulation level.
In the figure below, all fields of the insulation test are detailed.
Enable
automatic test.
16.12.1 Micro-ohmmeter
In this tab the user checks the resistance values. There is also a tab with the diagram
of the connection between the resistance and the test case. This connection is changed
according to the configuration set in the CE-7012 or CE-7024.
In the figure below, all fields of the micro-ohmmeter test are detailed.
Enable
automatic test.
16.12.2Winding resistance
Choosing the current and voltage channels, the software evaluates the winding
resistance value even considering the copper temperature value.
In the figure below, all fields of the winding resistance test are detailed.
GENERAL
SETTINGS
Standard
configura
tions of
voltage
channels.
Standard
configura Schematic
tions of diagram of
current External PT. connections.
channels.
Mode of operation
of the binary
inputs (contact or
Configur voltage).
ation in
electrom
echanical External CT.
mode.
To access the settings of the external TPs and TCs, just click on the three points then
the following screen, for TP and TC respectively, will open:
Enables / disables.
Description.
Max module.
Transformation
Ratio.
Max./Min.
Frequency of
operation.
Polarity Polarity
inversion. inversion.
Delay time related
to angular lag.
It allows special
configurations
for transient
exits.
Generation of 6
amplifiers defined
Instruction
by the user.
s and obs.
Figure 343 – Hardware Configurations – Working mode with free analog outputs
CONPROVE INDÚSTRIA E COMÉRCIO LTDA 290
CTC SOFTWARE MANUAL
If the user has an external CONPROVE amplifier to work with his CE-6003 / CE-
6006, it is necessary to inform some configurations of the case that will work in “SLAVE”
mode:
Figure 344 – Hardware Configurations – Working mode with Amplif. Ext. CONPROVE
If the user has any external amplifier to work together with his CE-6003 / CE-6006
working as an analog output, it is necessary to inform some settings of the equipment that
will work in “SLAVE” mode:
Choose Descripti List RMS Nom. - Value of the input module in the Amplifier.
on of the
between of RMS Excit. - Output Module value in the Amplifier.
amplifier
Analog others Offset Excit. - Correction of the offset value.
Output. or Ampl. t Def. - Angular lag that the Amplifier causes in the 60 Hz output
Anlg Entry. signal. Value given in seconds.
Thermal limits
of ampl. of
voltage.
Thermal limits
of ampl. of
current.
Figure 345 – Hardware Configurations – Working mode with Amplif. Ext. Output set p / user
If the user has any external amplifier to work together with his CE-6003 / CE-6006
working as an analog input, it is necessary to inform some settings of the equipment that
will work in “SLAVE” mode:
Choose Descri List RMS Nom. - Value of the Output module in the Amplifier.
between ption of RMS Cond. - Value of the Input Module in the Amplifier.
Anlg Output. of the others Offset Cond. - Correction of the offset value.
or Anlg amplifi Ampl. t Def. - Angular lag that the Amplifier causes in the 60 Hz output
Entry. er signal. Value given in seconds.
Inverts the
signal of each
amplifier.
Operating
frequencies.
Figure 346 – Hardware Configurations – Working mode w / Amplif. User defined input extension
The Settings screen is illustrated in the Figure below. This screen is active when you
click on the “Config. Hrd” located in the “home” menu in the “Hardware” part. In this
screen the user has the option to choose the working mode, whether the connection will be
made with or without External Hardware, whether the box will operate with all analog
outputs (standard or free mode) generating voltage, or generating current, or voltage and
current, the user can configure the initial state of the binary outputs, the operation mode of
the binary inputs (by contact or by voltage) and the value of the auxiliary source (default
options or “other” options).
Master Model of Serial number of Initial Value at
equipment equipment used the equipment state of which
configurations. as master. used as a master. the binary Auxiliary
outputs. Source will
be adjusted.
Standard
configura
tions of
voltage
channels.
Schematic
diagram of
connections.
Standard
External PT.
configura
tions of Operation mode of
current the binary inputs
channels. (contact, voltage or
measurement).
External CT.
To access the settings of the external TPs and TCs, just click on the three points then the
following screen, for TP and TC respectively, will open:
Enables / disables.
Description.
Max module.
Transformati
on Ratio.
Max./Min.
Frequency of
operation.
Polarity Polarity
inversion. inversion.
Delay time related
to angular lag.
If the user has an external CONPROVE amplifier to work with his CE-6003 / CE-
6006, it is necessary to inform some configurations of the case that will work in “SLAVE”
mode:
Figure 349 – Hardware Configurations – Work mode with Amplif. Ext. CONPROVE
If the user has any external amplifier to work together with his CE-6003 / CE-6006
working as an analog output, it is necessary to inform some settings of the equipment that
will work in “SLAVE” mode:
Choose Descri List RMS Nom. - Value of the input module in the Amplifier.
between ption of RMS Excit. - Output Module value in the Amplifier.
Anlg Output. of the others Offset Excit. - Correction of the offset value.
or Anlg amplifi Ampl. t Def. - Angular lag that the Amplifier causes in the 60 Hz output
Entry. er signal. Value given in seconds.
Inverts the
signal of each
amplifier.
Operating
frequencies.
Thermal
limits of
ampl. of
voltage.
Thermal
limits of
ampl. of
current.
Figure 350 – Hardware Configurations – Working mode with Amplif. Ext. Output set for user
If the user has any external amplifier to work together with his CE-6003 / CE-6006
working as an analog input, it is necessary to inform some settings of the equipment that
will work in “SLAVE” mode
Choose Descri List RMS Nom. - Value of the Output module in the Amplifier.
between ption of RMS Cond. - Value of the Input Module in the Amplifier.
Anlg Output. of the others Offset Cond. - Correction of the offset value.
or Anlg amplifi Ampl. t Def. - Angular lag that the Amplifier causes in the 60 Hz output
Entry. er signal. Value given in seconds.
Inverts the
signal of each
amplifier.
Operating
frequencies.
Figure 351 – Hardware Configurations – Working mode w / Amplif. User defined input extension
The concept of channel targeting was created at CTC. In this resource it is possible to
create voltage and current nodes in the software that is totally unrelated to the
hardware. With that, the whole sequence of data created in a specific test and with a specific
bag model is not lost when the equipment is changed. It is enough for the user to associate
the new hardware with the existing nodes. To access the settings, click on "Channel
Direction”, in the "Hardware" part located on the "Home" tab.
Access
Hardware Adjust I / Os - Adjust the number of Autoassociate -
configuration. channels according to the parameterized Automatically associates
hardware. the created channels with
Autoassociate - Automatically the respective virtual
Equipment
“Nodes”.
model. associates the channels created with the Clear - Clears all
existing Hardware. channels associated with
Serial number of Clean - Clears all channels.
the equipment.
a node.
Adding or
Removing
Nodes.
System
nominal
values. Manual association of a
Line and Association of “Hardware” with an
source Points with “Node” and an existing
parameters channels. “Point”.
equivalent.
In the figure above a direction was made for the exit channels of the suitcase. Note in
the figure below the similarity of the data for the input direction (measurement). The basic
difference is that in CE-6003 / CE-6006 equipment there are only 3 voltage measurement
channels and 3 current measurement channels. In the CE-7012 / CE-7024, there are twelve
channels that can be parameterized for both voltage measurement and current measurement.
In channel targeting there is also the ability to inject a current value that depends on
the sum of other channels or that other channels inject values that depend on a third
party. This feature is used, for example, for tests on distance relays whose measurement is
carried out at two points (a circuit breaker and a half system). To use this feature, associate a
different node for each hardware and keep the same point in the system, as shown in the
figure below (Three different nodes, three current channels and a single point in the system):
Three different
Three different nodes.
hardware.
Same
point in
System.
The third node, which represents the third current channel, is tied to the other two
nodes so that this channel defines the value of the other two or the other two channels define
the value of that third.
Access to
summation
configuration.
Node values
01 and 02
define the
value of the NODE 03
third channel. value defines
the value of
NODS 01 and
02, following a
proportion.
Desired
proportion of
each NODE in
the total sum.
Current channel of
NO03 that defines the
current value of the
other Nodes.
The figure below shows the direction of the channels of the binary outputs and
the GOOSE outputs, very similar to the process of the voltage and current generation
channels. The initial state of the binary outputs is defined by the hardware
configuration. The routing is done by associating a channel with a physical output. In the
case of GOOSE messages, in addition to associating a physical output to a channel, it is also
necessary to choose the dataset, for that click on the three buttons. In the case of CE-7012 /
CE-7024 equipment, the option of DC analog output is contemplated, both for voltage and
current.
Targeting the binary inputs and GOOSE inputs. Note in the figure below that the
channel direction of the transducer inputs (voltage and current) appears. The user simply
associates a channel with a physical entry:
21 General Settings
Substation: Specific name of the location where Name: Name of the person responsible for
the relay is located. executing the test.
Bay: Name of the set of equipment that the relay Sector: Location of the company where the
makes Part. person responsible works.
Address, City, State: Exact location of the Registration: Secondary identification of the
substation. person responsible.
The figure below shows the parameters contained in the “SYSTEMS” tab. In it, the
system's base values, impedance data of the line where the relay is connected and data from
the system's equivalent source are parameterized.
Impedance data
of a possible line
/ feeder where
the relay can be
connected:
Reference from
Z to primary,
Current
correction, line
length, line
angle, ground
compensation
parameters
(Earth Factor),
parallel and
mutual line
compensation
parameters.
The figure below shows the parameters contained in the “NOTES & OBS” tab. In it,
the user has a space to report any abnormality, observation, opinion about the test. There is
also the option of extra customizable fields in order to offer more options for general
information.
Dedicated space
for writing user
notes and
observations.
Customizable
extra fields.
Image Description.
Inserting the
figure.
The figure below shows the parameters contained in the “CHECK LIST” tab. In it,
the user can include several tasks to be performed during commissioning
and, when completed, a check box can be checked confirming the execution.
Check box to
be checked
when Insert ITEM in the
necessary. list or remove an
item from the list.
The figure below shows the parameters contained in the “OTHERS” tab. In it the
user can set the deglitch time (the input signal is registered with a time delay)
and debounce (the signal is only registered after a certain time) for binary inputs and inputs
via GOOSE message.
Deglitch / debounce
time for Hardware
and Software of
GOOSE inputs.
22 Preferences
The “PREFENCES” option will be detailed below. This feature appears in all CTC
software and has details that will be part of the report along with the tests performed by the
user. This feature allows you to adjust default values for general adjustments that can be
loaded in their respective parameter windows by simply clicking on “Default”. It is possible
to choose directories from which the software can search for a file, choose the company logo
that will go to the report or define colors of the graph. Following is the directory screen.
Default
Choice of language
directories
(where the
for all
software software.
automatically
searches for a
file, when
requested) of
Tests,
explanatory
figures,
overcurrent
curves, comtrade
files (.CFG),
ATP files (.LIS), Choice of an audible
RIO file and alarm for indication of
IEC-61850 overload in the channels.
Substations.
The figure below shows the parameters contained in the “REPORT” tab. In it, the
user can load the company logo on the report and can also choose the standard language in
which the report would be issued.
Choice of the
Choice of the logo default language.
of the company that
owns the
equipment.
The figure below shows the parameters contained in the “GRAPHICS” tab. In it, the
user can choose the gradient effect, grid types and colors of the graphics.
Choice of
the gradient
effect and
the colors
used.
Choose
the type of
grid and
the colors
used (grid
and
zoom).
The figure below shows the parameters contained in the “VALUES DEFAULT”
tab. In it, the user can choose default values of “GENERAL SETTINGS”. For the values
parameterized here to be used in the “GENERAL SETTINGS”, the user simply clicks the
“Default” button.
23 Firmware
24 Data Association
The field in question can be associated with a multitude of quantities that are present
in the software. Even, you can even add a value multiplier and / or adder easing further test
file.
Multiplier.
Adder.
GOOSE MESSAGES
SETTINGS
The GOOSE Messages Settings screen has as its main functionality the configuration
of GOOSE messages exchanged between Intelligent Electronic Devices, or Intelligent
Electronic Devices (IED), defined by the IEC 61850 standard. To view it, just click on the
“Home” tab and on the “Config GOOSE”.
Creation of new GOOSE messages, or new sets of DataSet, with attributes also
defined by the user;
Mapping attributes of the DataSet to receive, with 8 input channels, each capable of
monitoring an attribute, allowing the receipt of 8 attributes for the CE-6006. The CE-
7012 and CE-7024, on the other hand, allow the receipt of 44 attributes.
Mapping of DataSet attributes for sending, with 6 output channels, each capable of
sending up to 1 attribute, allowing the sending of up to 6 attributes to the CE-6006. The CE-
7012 and CE-7024, on the other hand, allow the sending of up to 16 attributes.
Imports the file Defines that all attributes Defines that all Shows the status
with the GOOSE come from a simulation. attributes come from of each attribute.
mgs. test.
On the Settings screen of messages GOOSE principal fields Receive and Send are
displayed and manipulated the attributes to be received and those mapped to shipping.
On the Settings screen of messages GOOSE principal fields Receive and Send are
displayed and manipulated the attributes to be received and those mapped to shipping.
Changes the
ordering of a
dataset or
attribute.
Descriptio
n of each
dataset
with its
attributes.
There is the option of mapping 4 parameters for sending GOOSE messages: Review,
Simulation Bit, Test and Commissioning Flag. To select these fields as attributes of a
GOOSE message is just put '#-attribute name'. The figure below exemplifies this
PROCEDURE n to:
The main functionality of the SMV Message Settings screen is the configuration
of Sampled Values messages exchanged between Intelligent Electronic Devices, or
Intelligent Electronic Devices (IED), defined by the IEC 61850 standard. “Config SV” icon.
Creation of new SV messages, or new sets of DataSet, with the attributes also
defined by the user.
On the main SMV Message Settings screen, in the Receive and Send fields, the SMV
messages to be received and those mapped for sending are displayed and manipulated.
Changes the
ordering of an
SMV
message and /
or attribute.
Descriptio
n of each
SMV
message
with its
attributes.
Looking at the previous figure, it appears that for each SMV Value (TCTR or TVTR)
there is a corresponding bitstring (BitString [14]). This field, according to IEC 61850 -7-3,
is called 'Quality Bits' and consists of 14 bits, each of which represents a parameter. The
parameters and their respective values will be explained below:
* The Derived parameter is not originally found in IEC 61850-7-3, as it was added to
the Implementation Guideline for Digital Interface to Instrument Transformers using IEC
61850-9-2, being better known as Light Edition or 9-2LE.
is represented by the 16-bit unsigned integer data type and its content exceeds
65535.
Out of Range: indicates that the value is beyond the defined limits (used
only for measured values). For example, the expected value is between 0 and
40,000; however the measured value is 41000.
Bad Reference: indicates that the value may be incorrect, as the
reference is not calibrated (used only for measured values and binary
information).
Oscillatory: indicates that the value is fluctuating unstably within a certain
time interval (used only for status information).
Failure: indicates that the supervision function has detected an internal or
external failure.
Old Data: indicates that the value has not been updated within a given time
interval.
Inconsistent: indicates an inconsistency.
Inaccurate: indicates that the value is outside the expected precision.
Source - informs the origin of the value:
o Process: indicates that the value originates from an input function or
is calculated by some application function.
o Substituted: indicates that the value comes from an operator input or
an automatic function.
Test: indicates that the value is for testing and should not be used for
operational purposes.
Operator Blocked: indicates that the operator has blocked the update of the
value.
Derived: this parameter must be set (TRUE) if the value was not obtained
from a physical sensor, but from calculations based on other values. For
example, the value of Neutral = Phase A + Phase B + Phase C. The parameter
must be reset (FALSE) if the value was obtained from a physical sensor.
In the configuration screen of the previous figure, it is observed that next to each
Quality there is a field ('Default') to configure each of the parameters. The value default is
all '0', and if the user wishes to configure the values, it must be done bit by bit in the MSB -
LSB direction. For example, for the Quality of the TCTR1 value, if you want to set bits 13
and 11, you must do it as: 10100000000000. The figure below exemplifies the procedure:
Quality
configuration
example.
Advanced settings
Virtual LAN ID and SMV
and priority message buffer.
information.
MULTIM
MEASUREMENT SOFTWARE
INSTANT
To make the acquisitions, simply connect the equipment's measurement inputs to the
outputs and / or generators (for example, a three-phase source) whose values will be
measured, and click on the “Acquire” button, which changes to “Stop Acquisition”. The user
can choose the Clamp scale (for current measurements). You can also define whether you
are working with a frequency of 50 or 60 Hz.
With MULTIM, the user can view the waveforms of the generated signal, see its rms
values of amplitude and angle. You can also view all the harmonic components present in
the measured signal in the form of a graph and table.
This software has the feature “Open Settings at Start” (Software Options menu), if it
is active, when starting the Manual software, the Settings screen is active immediately, this
feature is interesting since, before starting the tests it is Important specify the parameters
related to the test
The options on the Settings screen are described in detail in chapter 21.
Creates a new
test file.
Open a saved
file.
Password protect
certain
parameters of the
test file.
Saves the current
file.
In the start menu, several resources are available and are detailed in the following
figures. The following figure describes features of hardware configurations, editing of
results and generation control.
Opens a Channel Opens a screen Reissue Shows the tests performed or Initiates
whether it is being performed
Orientation for Hardware the test signal
at the current time.
screen. Configuration. acts. generation.
Opens a screen Opens a screen for Insert a Excludes Deletes Stop signal
for GOOSE and connecting to the new test. all tests. the current generation.
SV Configuration. suitcase. test.
In the figure below, the resources for analysis options of the generated quantities are
shown, as well as the options for general adjustments and the creation of an
evaluation. Access to the report, options for viewing the units involved (time and module)
and general options for restoring and viewing the layout are also displayed.
Calls the Shows the Shows the Shows the Shows the Shows the Returns to
Waveform Phasor Harmonics Tab. time in the standard
Settings time in
Tab. Tab.
Screen. seconds. cycles. layout.
Habilita/
desabilita
abas para
Shows the Shows the Shows Shows Shows Shows values visualização
Evaluation configuration relative based on the
primary absolute
tab. secondary.
of the report. values. values. values.
Figure 398 – Multim – Home Menu – Options, Report, Units and Layout
Position the selected Superimposes the Hide the windows Returns to Enables /
window: on the left; selected window. the standard
for better viewing. disables tabs
to the right, above, layout.
for viewing.
below or as a
document.
Hide the selected Hide the selected Close all open Closes all open
tabs. window. windows. documents.
It is also opened at software startup, if the Opens Settings when Starting option in the
Software Options tab is selected.
In General mode there is the General Information tab, where the user enters all the
data related to the test, the Test set device and the person responsible for the test, there is the
Systems tab, where the user informs the nominal values of frequency, voltage, current and
power among others, there is a tab for Notes and Observations for various records.
There is an Explanatory Figures tab where the user can insert up to three figures
relevant to the test, a Check List tab to create steps to be followed during the test, there is a
tab called Others for configuring digital signal filters and finally a tab Connections allowing
the user to structure all connections between the test case and the equipment to be Test set.
Enables / disables
special inputs for
transducer calibration.
27.7.2
It allows 4 different
sizes for
visualization of the
graphs.
It allows changing the
following parameters
Opens the Assign of a single graphic or
Signals window, all of them: Title,
where the user Legend, Scale, Grid,
determines which Table and
values will be shown Dimensions.
in the graphs.
Clicking on the Show Columns option in the previous figure brings up the following
options:
It allows 4 different
sizes for visualization
of the graphs.
Consideration for
Evaluation between
module / angle, R / X, Rs / Multiplied Tolerance
Rp, Ls / Lp, Powers. It constant taken into
Channels that Theoretical
depends on what is Ratio account in the
will be parameterized on the value for Evaluation.
compared. obtained. Evaluation.
“Type” screen.
TRANSIENT
SOFTWARE FOR SIMULATION OF
MEASUREMENT SIGNALS IN TRANSITORY
REGIME
The options on the Settings screen are described in detail in chapter 21.The Distance,
Timing, Import and Export fields will be discussed throughout this chapter.
Creates a new Opens a saved Saves the Customize the Software name and
test file. test. current test. Quick Access version.
Bar.
Equipment used
Opens the options: and serial
Restore, Move, number.
Size, Minimize,
Maximize and
Close.
Creates a new
test file.
Open a saved
file.
Password protect
certain
parameters of the
test file.
Saves the current
file.
In the start menu, several resources are available. In the following figure are
described features of the hardware settings and editing sequence s such as insert, copy,
delete, scroll and select the reference.
Connects between Opens a screen for Deletes all Shifts the selected
software and GPS Setup. Sequences. Sequence to the beginning
hardware. or end.
The following are the settings related to the management of results (reissue of tests,
exclusion of tests and choice of test for analysis) and generation control.
In the figure below, the resources for analysis options of the generated quantities
(waveforms, phasors, trajectories, harmonics, synchronoscope and impedance plane) are
displayed, as well as the adjustment options (General, distance and synchronism
settings). Access to the report, options for viewing the units involved (time and module) and
general options for restoring and viewing the layout are also displayed.
Calls the Shows the Shows the Shows the Opens the Shows the Shows the
Settings Waveform Phasor Harmonics report time in time in
Tab. Tab. Tab.
Screen. configuration seconds. cycles.
window.
Shows
values
based on the
secondary.
Tabs for
viewing and
Standard
layout.
Shows values
Shows the Shows the based on the Shows
Synchronoscope Impedance Shows
Trajectory primary. relative
plan absolute
tab. tab.
configuration. values. values.
Figure 421 – Transient – Home Menu – Options, Report, Units and Layout
Position the selected Superimposes the Hide the windows Returns to Enables /
window: on the left; selected window. the standard
for better viewing. disables tabs
to the right, above, layout.
for viewing.
below or as a
document.
Hide the selected Hide the selected Close all open Closes all open
tabs. window. windows. documents.
It is also opened at software startup, if the Opens Settings when Starting option in the
Software Options tab is selected.
In General mode there is the General Information tab, where the user enters all the
data related to the test, the Test set device and the person responsible for the test, there is the
Systems tab, where the user informs the nominal values of frequency, voltage, current and
power among others, there is a tab for Notes and Observations for various records.
There is an Explanatory Figures tab where the user can insert up to three figures
relevant to the test, a Check List tab to create steps to be followed during the test, there is a
tab called Others for configuring digital signal filters and finally a tab Connections allowing
the user to structure all connections between the test case and the equipment to be Test set.
In Distance mode, there is the Distance Protection Adjustment tab. In it the user sets
the line length and the angle. Choose how the earth factor will be inserted. You can still
enable whether to remove the Arc Resistance and use Impedance Correction for Rated
Current. To facilitate the adjustment of the zones, there is an Import option whose purpose is
to read files .RIO. Some distance relays record line, zone and operating time settings in this
format. Importing this file, Transient reads automatically so that the user does not need to
adjust it manually. Another option would be to export a file. RIO once the feature was
entered by the user.
The Insert option opens a new window for the user to enter the name and number of
the zone, adjust the Trigger Time, Absolute and Relative Tolerances of the impedance and
the Absolute and Relative Tolerances for the time. You can use a specific earth factor and
enable the option to Remove Arc Resistance. If the relay does not generate a file .RIO there
is the option to design the zone to be Test set or to choose between predefined
zones. Depending on the type of zone chosen, options for load compensation and
directionality are opened.
Displays the
tab for
adjusting the
sync function.
Import
.RIO file
Export
.RIO Files
Choose the
mode of the
earth factor:
(Mod;Ang)
(R;X)
(R0;X0)
(K0)
(Rt/Rl;Xt/Xl)
(Z0/Z1).
Shows which
zones are
available.
Insert a Edit a zone Deletes a zone that has Shows the design of
already inserted. already been inserted. the chosen zone.
new zone.
Let us see the window that opens when clicking on Insert, where the adjustments of
the protection zones are made.
To facilitate the visualization of the previous figure, it was divided into parts. In this
first part the user can enter the description of the added zone, choose the type of actuation,
select which loop (detection algorithm) that zone belongs to and enter the absolute and
relative tolerances of time and impedance.
Zone description.
Zone type registration:
Zone index. Departure, Performance,
Extended or Block.
Links the zone to a (Informative data).
specific type of
fault. Absolute (Ω) and
relative (%) tolerance
of Impedance.
Trip time of the
zone.
Absolute tolerance
(more or less) in
seconds and relative
tolerance in% of the
time of operation.
Points inserted
when using the
option to draw.
In this second part the user has access to the various ways of drawing the desired
zone. The first form is drawing segment by segment, the second form uses known zones
(Mho, Quadrilateral and tomato / lens) where the user by comparison changes some intuitive
segments and the third form is using the presets of the relays themselves (Masks).
Finalizes the
drawing
process.
The following figure shows a quadrilateral area which was drawn using line
segments.
In the next figure, the quadrilateral option was chosen directly. Note that the settings
to be entered are more intuitive with this option.
Then the screen where system data 1 and 2 and the settings for synchronism are
inserted.
To facilitate understanding, the previous figure will be divided into two parts. Note
in the figure below that system one was adopted as the one to be synchronized. It is also
shown that it is possible to choose whether it is single-phase or three-phase, including the
reference phase, the angular offset of the elevation transformer (if the synchronism is done
in a Generator + Transformer system). Finally, the user can choose the way to start the
synchronism and parameterize the closing time of the circuit breaker (which will be taken
into account in the calculation of the relay).
Then system 2 is adopted as the side of the system that will receive the generation or
simply the infinite bar. In the figure below you can choose a three-phase or single-phase
system, the angular reference and the voltage and current levels in primary and secondary
values. Finally, you can change the phase sequence.
Figure 440 – Transient – Sequence Tab – NO sub tab – Two Nodes and three Sequences
In the following explanations, we will adopt that only one node (NO01) was created
containing three voltage channels and three current channels. Remember that to
add sequence s click the Insert New Sequence. The figures below show the sequence modes
that can be entered in the test.
Choose between
COMTRADE
file (.cfg
extension) or Access button for
ATP file (.lis parameterizing the
extension). comtrade / ATP file.
By clicking on the access button to parameterize the COMTRADE mode and load a
file .cfg appears a screen where basically the user will redirect the COMTRADE signal
(voltage, current or binary signals) to a specific channel (analog or binary output). See the
figure with details below:
Shows information
Picture Search for file to upload. about the uploaded
box. file.
By clicking on the access button to parameterize the ATP mode and load
a file. lis appears a screen where basically the user will redirect the simulation signal from
the ATP file (voltage, current) to a given analog channel. See the figure with details below:
Shows information
Picture Search for file to upload. about the file.
box. carregado.
CE-7024 equipment have this hardware feature. Every time a sequence is added, a DC
generation sub tab opens, both current and voltage.
Sequence duration
time.
Sequence name.
Voltage and
current module.
Figure 444 – Transient – Sequence Tab – Output Analog. DC Sub tab – General Description
Sequence duration
time.
Sequence
name.
Figure 445 – Transient – Sequence Tab – Binary Outputs sub tab – General Description
Sequence duration
time.
Sequence
name.
Figure 446 – Transient – Sequence Tab – GOOSE Outputs sub tab – General Description
Figure 447 – Transient – Sequence Tab – Sub tab Time and Advance – General Description
Extract the
COMTRADE or CSV
Allows you to file from the waveforms.
Take a photo of the chart
Insert or Edit either
by storing it on the
Markings or clipboard.
Comments in
graphics.
Figure 449 – Transient – Waveform Tab - Resources
Clicking on the Display option in the previous figure brings you to the following
options:
28.9.11Phasor Tab
In this screen it is possible to view for each channel the module and the value of the
voltage or current angle, as well as the Phasors graph.
This screen has the Transparency property, and can remain open while points are
inserted or edited in the test. The following screen is for a line Test set using Search Test.
Clicking on the Show Columns option in the previous figure brings up the following
options:
28.9.13Trajectories Tab
With the voltage and current channels already directed with COMTRADE or ATP
file signals, we can represent an impedance path, so that if the user adds protection zones in
the Settings, it is possible to view the path of that COMTRADE or ATP in a fault region.
28.9.17Synchronoscope tab
With the voltage and current channels already directed with signals from the
COMTRADE or ATP file, we can verify through a synchronoscope if these signals are able
to perform a synchronism. Note in the figure below that the light blue region is the
synchronism region and is determined in the Settings.
Choice of
voltages that
represented
System 1 (to be
Synchronoscope synchronized)
of cursor 1 and 2. and System 2
(already
synchronized).
Type of fault
to be inserted.
Button to
Table with the insert a fault
various faults that sequence.
can be inserted.
When clicking on the insert button, the following screen will appear:
Fault insertion
Description of the mode (smart, Z-I
Sequences that cte, Z-V cte and Z-
represented a fault. Source cte)
After choosing the data in the figure above, just allocate the fault point with the
mouse in the impedance diagram, as shown in the figure below:
Confirmation
of the chosen
Table with several points point.
chosen and their
respective information.
SEQUENC
FAULT SEQUENCE SIMULATION SOFTWARE
The options on the Settings screen are described in detail in chapter 21. The
Distance, Timing, Import and Export fields will be discussed throughout this chapter.
Opens the
Equipment used
options: Restore, and serial
Move, Size, number.
Minimize,
Maximize and
Close.
Creates a new
test file.
Open a saved
file.
Password protect
certain
parameters of the
test file.
Saves the current
file.
In the start menu, several resources are available. In the following figure are
described features of the hardware settings and editing sequence s such as insert, copy,
delete, scroll and select the reference.
Connects between Opens a screen for Deletes all Shifts the selected
software and GPS Setup. Sequences. Sequence to the beginning
hardware. or end.
The following are the settings related to the management of the results (reissue of
tests, exclusion of tests and choice of the test for analysis) and generation control.
In the figure below, the resources for analysis options of the generated quantities
(waveforms, phasors, trajectories, harmonics, synchroscope and impedance plane) are
displayed, as well as the adjustment options (General, distance and synchronism settings).
CONPROVE INDÚSTRIA E COMÉRCIO LTDA 378
CTC SOFTWARE MANUAL
Access to the report is also displayed, options for viewing the units involved (time and
module) and general options for restoring and viewing the layout.
Calls the Shows Shows the Shows the Opens the Shows the Shows the
Settings the Waveform Harmonics report time in time in
Screen. Phasor Tab. Tab. configuration seconds. cycles.
Tab.
window.
Shows
values
based on the
secondary.
Tabs for
viewing and
Standard
layout.
Shows values
Shows the Shows the based on the Shows
Impedance Shows
Synchronos Trajectory tab. primary. relative
plan absolute
cope tab.
configuration. values. values.
Figure 469 – Sequenc – Home Menu – Options, Report, Units and Layout
Position the selected Superimposes the Hide the windows Returns to Enables /
window: on the left; selected window. the standard
for better viewing. disables tabs
to the right, above, layout.
for viewing.
below or as a
document.
Hide the selected Hide the selected Close all open Closes all open
tabs. window. windows. documents.
Insert password in Enables / disables the Opens the Settings Shows the
manufacturer’s address,
the Test file so as simulation mode. Since Screen when the
phone, email and fax.
not to allow the in this mode there is no software is started.
editing of certain insertion of voltage and
settings. current.
It is also opened at software startup, if the Opens Settings when Starting option in the
Software Options tab is selected.
In General mode there is the General Information tab, where the user enters all the
data related to the test, the Test set device and the person responsible for the test, there is the
Systems tab, where the user informs the nominal values of frequency, voltage, current and
power among others, there is a tab for Notes and Observations for various records.
There is an Explanatory Figures tab where the user can insert up to three figures
relevant to the test, a Check List tab to create steps to be followed during the test, there is a
tab called Others for configuring digital signal filters and finally a tab Connections allowing
the user to structure all connections between the test case and the equipment to be Test set.
In Distance mode, there is the Distance Protection Adjustment tab. In it the user sets
the line length and the angle. Choose how the earth factor will be inserted. You can still
enable whether to remove the Arc Resistance and use Impedance Correction for Rated
Current. To facilitate the adjustment of the zones, there is an Import option whose purpose is
to read files .RIO. Some distance relays record line, zone and operating time settings in this
format. Importing this file, Sequenc automatically reads it so that the user does not need to
adjust it manually. Another option would be to export a file .RIO once the feature was
entered by the user.
The Insert option opens a new window for the user to enter the name and number of
the zone, adjust the Trigger Time, Absolute and Relative Tolerances of the impedance and
the Absolute and Relative Tolerances for the time. You can use a specific earth factor and
enable the option to Remove Arc Resistance. If the relay does not generate a file .RIO there
is the option to design the zone to be Test set or to choose between predefined
zones. Depending on the type of zone chosen, options for load compensation and
directionality are opened.
Displays the
tab for
adjusting the
sync
function.
Import
Files
.RIO
Export
Files
.RIO
Choose the
mode of the
earth factor:
(Mod;Ang)
(R;X)
(R0;X0)
(K0)
(Rt/Rl;Xt/Xl)
(Z0/Z1).
Shows which
zones are
available.
Let's look at the window that opens when you click Insert and that
it is performed s settings of protection zones.
To facilitate the visualization of the previous figure, it was divided into parts. In this
first part the user can enter the description of the added zone, choose the type of actuation,
select which loop (detection algorithm) that zone belongs to and enter the absolute and
relative tolerances of time and impedance.
Zone description.
Zone type registration:
Zone index. Departure, Performance,
Extended or Block.
Links the zone to a (Informative data).
specific type of
fault. Absolute (Ω) and
relative (%) tolerance
of Impedance.
Trip time of the
zone.
Absolute tolerance
(more or less) in
seconds and relative
tolerance in% of the
time of operation.
Points inserted
when using the
option to draw.
In this second part the user has access to the various ways of drawing the desired
zone. The first form is drawing segment by segment, the second form uses known zones
(Mho, Quadrilateral and tomato / lens) where the user by comparison changes some intuitive
segments and the third form is using the presets of the relays themselves (Masks).
Finalizes the
drawing
process.
The following figure shows a quadrilateral area which was drawn using line
segments.
In the next figure, the quadrilateral option was chosen directly. Note that the settings
to be entered are more intuitive with this option.
Then the screen where system data 1 and 2 and the settings for synchronism are inserted.
To facilitate understanding, the previous figure will be divided into two parts. Note
in the figure below that system one was adopted as the one to be synchronized. It is also
shown that it is possible to choose whether it is single-phase or three-phase, including the
reference phase, the angular offset of the elevation transformer (if the synchronism is done
in a Generator + Transformer system). Finally, the user can choose the way to start the
synchronism and parameterize the closing time of the circuit breaker (which will be taken
into account when calculating the relay).
Then system 2 is adopted as the side of the system that will receive the generation or
simply the infinite bar. In the figure below you can choose a three-phase or single-phase
system, the angular reference and the voltage and current levels in primary and secondary
values. Finally, you can change the phase sequence.
Figure 488 – Sequenc – Sequence tab – NO sub tab – Two Nodes and three sequences
In the following explanations, we will adopt that only one node (NO01) was created
containing three voltage channels and three current channels. Remember that to
add sequence s click the Insert New Sequence. The figures below show the sequence modes
that can be entered in the test.
In this mode, the user directly enters the value of the module, angle and frequency of
each channel in the desired sequence.
Module
value. Choice of
operating
mode.
Description of
the targeted
channel.
Frequency
value.
Generic
point of the
System. Angle
value.
Figure 489 – Sequenc – Sequence Tab – NO sub tab – Direct Mode Description
Choice of
operating
mode.
Description of
the targeted
Access to the
channel.
Module and
Angle of 50
harm orders.
Generic
point of the
System. RMS value and
Total Harmonic
Distortion
(THD) of the
entered
harmonic
orders.
Figure 492 – Sequenc – Sequence Tab – NO sub tab – Description Harmonic mode
By right-clicking on the text box on the harmonics input screen, the following
formatting features of the input data fields are available.
In this mode, the user directly enters the value of the module and angle of each
interharmonic component (up to 10 interharmonic).
Choice of
operating
mode.
Description of
the targeted Access to the Module
channel. and Angle of
interharmonical orders.
Generic
point of the RMS value and
System. Total Harmonic
Distortion (DHT) of
the inserted
interharmonic
orders.
Figure 495 – Sequenc – Sequence Tab – NO sub tab – Description interharmonic mode
When clicking on the access button for the parameterization of the module and angle
of the harmonics, the following screen appears.
Limpa valores de
módulo e ângulo já
parametrizados
Frequency of the
interharmonic of Module value and
interest. angle of the inter-
harmonic order in
question.
Reduced number of
Clears all modules
decimal places.
and angles.
User-defined number
Default number of of decimal places.
decimal places (two
places).
Definition of
symmetric Voltage and current
values. values in
symmetrical
component format.
Both module and
angle.
Figure 498 – Sequenc – NO sub tab – Description of the Symmetric Components Mode
Default number of
Voltage input mode.
decimal places (two
places).
Current input mode.
Reduced number of
decimal places.
Frequency entry
mode.
User-defined number
of decimal places.
Definition of
Values of phase-to-phase
the Phase-
voltage, Sequence zero
Phase value. voltage and current in
the three phases.
Figure 500 – Sequenc – Sequence Tab – NO sub tab – Description of Phase-to-Phase Mode
Reduced number of
Frequency entry decimal places.
mode.
User-defined number
of decimal places.
Definition of
the Power Phase-to-ground
values. voltage values, single-
phase active and
reactive powers.
Figure 502 – Sequenc – Sequence Tab – NO sub tab – Description Mode Power
Defining the
fault values. Fault types (three-
phase, two-phase or
single-phase), angle
between voltage and
current and module /
angle of fault voltage
and fault current.
Figure 504 – Sequenc – Sequence Tab – NO sub tab – Description Fault Mode
29.9.16Constant ZI Mode
Nesse modo, o usuário define valores de Impedance e corrente que serão mantidas
constantes. O software então calcula quais os valores que tem que ser gerados pelos canais
para que essas quantidades de Z/I constante e V variável sejam atendidas.
Figure 506 – Sequenc – Sequence Tab – NO sub tab – Description Constant ZI Mode
Figure 508 – Sequenc – Sequence Tab – NO sub tab – Description Constant ZV Mode
Reduced number of
Frequency entry decimal places.
mode.
User-defined number
of decimal places.
Figure 510 – Sequenc – Sequence Tab – NO sub tab – Z mode description – constant source
Reduced number of
Frequency entry
decimal places.
mode.
User-defined number
of decimal places.
Nesse modo, o usuário define a Impedance de falta e os valores de falta são gerados
respeitando um limite definido pelo usuário, nesse caso, portanto, tanto a corrente como a
tensão variam.
Voltage and Sequence Sequence duration Choice of
current name. time. operating
values that mode.
will be
generated to
meet the
condition of Fault types (three-phase, two-
constant Z- phase or single-phase),
source. enabling DC offset, start of the
fault (together with the
maximum / minimum offset or
Definition of Z- a user-defined offset angle),
source constant definition of fault impedance,
values. definition of limits (voltage
and current) and definition of
line parameters (manually, by
the node or by zone
adjustments).
Figure 512 – Sequenc – Sequence Tab – NO sub tab – Description of the Z Mode - Smart
Reduced number of
Frequency entry decimal places.
mode.
User-defined number
of decimal places.
Sequence Choice of
name. operating
mode.
Definition of
the waiting
sequence.
Access button
for adding
image and
auxiliary text.
Figure 514 – Sequenc – Sequence Tab – NO sub tab – Description of the Standby Mode
Deletes an existing
image.
At the time of the test when this waiting sequence is reached, the following screen
will appear:
Choice of
operating
mode.
Sequence
name.
Figure 517 – Sequenc – Sequence Tab – NO sub tab – Description of the Accommodation Mode
Sequence duration
time.
Sequence name.
Voltage and
current module.
Figure 518 – Sequenc – Sequence Tab – Output Analog. DC Sub tab – General Description
Sequence duration
time.
Sequence
name.
Figure 519 – Sequenc – Sequence Tab – Binary Outputs sub tab – General Description
Sequence duration
time.
Sequence
name.
Figure 520 – Sequenc – Sequence Tab – GOOSE Outputs sub tab – General Description
Figure 521 – Sequenc – Sequence Tab – Time and Advance sub tab – General Description
CONPROVE INDÚSTRIA E COMÉRCIO LTDA 410
CTC SOFTWARE MANUAL
Extract the
COMTRADE or CSV
Allows you to Insert or file from the waveforms.
Edit either Markings or Take a photo of the chart
Comments in graphics. by storing it on the
clipboard.
Clicking on the Display option in the previous figure brings you to the following
options:
This screen has the Transparency property, and can remain open while points are
inserted or edited in the test. The following screen is for a line Test set using Search Test.
Clicking on the Show Columns option in the previous figure brings up the following
options:
It allows 4 different
sizes for visualization of
the graphs.
It allows the
visualization of
Opens the Assign values in module,
Signals window where angle, real value and
the user determines
imaginary value.
which values will be
shown on the graphs.
Type of fault
to be inserted.
Button to
Table with the insert a fault
various faults that sequence.
can be inserted.
When clicking on the insert button, the following screen will appear:
Fault insertion
Description of the mode (smart, Z-I
Sequences that cte, Z-V cte and Z-
represented a fault. Source cte).
After choosing the data in the figure above, just allocate the fault point with the
mouse in the impedance diagram, as shown in the figure below:
Confirmation
of the chosen
Table with several points point.
chosen and their
respective information.
RAMPA
SOFTWARE TO GENERATE SIGNAL
SEQUENCES WITH RAMP INCREMENTATION
The options on the Settings screen are described in detail in chapter 21. The
Distance, Timing, Import and Export fields will be discussed throughout this chapter.
Creates a new Opens a saved Saves the Customize the Software name
test file. test. current test. Quick Access and version.
Bar.
Equipment used
Opens the options: and serial
Restore, Move, number.
Size, Minimize,
Maximize and
Close.
Creates a new
test file.
Open a saved
file.
Password protect
certain
parameters of the
test file.
Saves the current
file.
In the start menu, several resources are available. In the following figure are
described features of the hardware settings and editing sequences such as insert, copy,
delete, and scroll and select the reference.
Connects between Opens a screen for Deletes all Shifts the selected
software and GPS Setup. Sequences. Sequence to the beginning
hardware. or end.
The following are the settings related to the management of results (reissue of tests,
exclusion of tests and choice of test for analysis) and generation control.
In the figure below, the resources for analysis options of the generated quantities
(waveforms, phasors, trajectories, harmonics, syncoscope and impedance plane) are
displayed, as well as the adjustment options (General, distance and synchronism
settings). Access to the report, options for viewing the units involved (time and module) and
general options for restoring and viewing the layout are also displayed.
Calls the Shows Shows the Shows the Opens the Shows the Shows the
Settings the Waveform Harmonics report time in time in
Screen. Phasor Tab. Tab. configuration seconds. cycles.
Tab. window.
Shows
values
based on the
secondary.
Tabs for
viewing and
Standard
layout.
Shows values
Shows the Shows the based on the Shows
Synchronoscope Impedance Shows
Trajectory plan primary. relative
tab. absolute
tab. configuration. values. values.
Figure 543 – Rampa – Home Menu – Options, Report, Units and Layout
Position the selected Superimposes the Hide the windows Returns to Enables /
window: on the left; selected window. the standard
for better viewing. disables tabs
to the right, above, layout.
for viewing.
below or as a
document.
Hide the selected Hide the selected Close all open Closes all open
tabs. window. windows. documents.
Insert password in Enables / disables the Opens the Settings Shows the
manufacturer’s address,
the Test file so as simulation mode. Since Screen when the
phone, email and fax.
not to allow the in this mode there is no software is started.
editing of certain insertion of voltage and
settings. current.
It is also opened at software startup, if the Opens Settings when Starting option in the
Software Options tab is selected.
In General mode there is the General Information tab, where the user enters all the
data related to the test, the Test set device and the person responsible for the test, there is the
Systems tab, where the user informs the nominal values of frequency, voltage, current and
power among others, there is a tab for Notes and Observations for various records.
There is an Explanatory Figures tab where the user can insert up to three figures
relevant to the test, a Check List tab to create steps to be followed during the test, there is a
tab called Others for configuring digital signal filters and finally a tab Connections allowing
the user to structure all connections between the test case and the equipment to be Test set.
In Distance mode, there is the Distance Protection Adjustment tab. In it the user sets
the line length and the angle. Choose how the earth factor will be inserted. You can still
enable whether to remove the Arc Resistance and use Impedance Correction for Rated
Current. To facilitate the adjustment of the zones, there is an Import option whose purpose is
to read files .RIO. Some distance relays record line, zone and operating time settings in this
format. Importing this file, Sequenc automatically reads it so that the user does not need to
adjust it manually. Another option would be to export a file .RIO once the feature was
entered by the user.
The Insert option opens a new window for the user to enter the name and number of
the zone, adjust the Trigger Time, Absolute and Relative Tolerances of the impedance and
the Absolute and Relative Tolerances for the time. You can use a specific earth factor and
enable the option to Remove Arc Resistance. If the relay does not generate a file .RIO there
is the option to design the zone to be Test set or to choose between predefined
zones. Depending on the type of zone chosen, options for load compensation and
directionality are opened.
Displays the
tab for
adjusting the
sync
function.
Import
Files
.RIO
Export
Files
.RIO
Choose the
mode of the
earth factor:
(Mod;Ang)
(R;X)
(R0;X0)
(K0)
(Rt/Rl;Xt/Xl)
(Z0/Z1).
Shows which
zones are
available.
Let us see the window that opens when clicking on Insert, where it is made the
adjustments of the protection zones.
To facilitate the visualization of the previous figure, it was divided into parts. In this
first part the user can enter the description of the added zone, choose the type of actuation,
select which loop (detection algorithm) that zone belongs to and enter the absolute and
relative tolerances of time and impedance.
Zone description.
Zone type registration:
Zone index. Departure, Performance,
Extended or Block.
Links the zone to a (Informative data).
specific type of
fault. Absolute (Ω) and
relative (%) tolerance
of Impedance.
Trip time of the
zone.
Absolute tolerance
(more or less) in
seconds and relative
tolerance in% of the
time of operation.
Points inserted
when using the
option to draw.
In this second part the user has access to the various ways of drawing the desired
zone. The first form is drawing segment by segment, the second form uses known zones
(Mho, Quadrilateral and tomato / lens) where the user by comparison changes some intuitive
segments and the third form is using the presets of the relays themselves (Masks).
Finalizes the
drawing
process.
The following figure shows a quadrilateral area which was drawn using line
segments.
In the next figure, the quadrilateral option was chosen directly. Note that the settings
to be entered are more intuitive with this option.
Then the screen where system data 1 and 2 and the settings for synchronism are inserted.
To facilitate understanding, the previous figure will be divided into two parts. Note
in the figure below that system one was adopted as the one to be synchronized. It is also
shown that it is possible to choose whether it is single-phase or three-phase, including the
reference phase, the angular offset of the elevation transformer (if the synchronism is done
in a Generator + Transformer system). Finally, the user can choose the way to start the
synchronism and parameterize the closing time of the circuit breaker (which will be taken
into account when calculating the relay).
Then system 2 is adopted as the side of the system that will receive the generation or
simply the infinite bar. In the figure below you can choose a three-phase or single-phase
system, the angular reference and the voltage and current levels in primary and secondary
values. Finally, you can change the phase sequence.
Figure 562 – Rampa – Sequence Tab – NODE sub tab – Two Nodes and three Sequences
In the following explanations, we will adopt that only one node (NO01) was created
containing three voltage channels and three current channels. Remember that to
add sequences click the Insert New Sequence. The figures below show the sequence modes
that can be entered in the test.
Choosing the Choose between Direct (no return for a specific Time in which each
type of ramp. value) or Pulsed (at the end of a specific increment is being generated
generation time the ramp goes to a specific value). and the time in which each
value of the reset time.
When clicking on the access button for the ramp parameterization, the following
screen appears. Choose the "Angle type" option in the "Ramp Type”. In the figure below, all
features that can be parameterized in this mode are characterized, such as direct or pulsed
type, initial ramp values, final values, increment, reset values (in case the pulsed type is
chosen), generation and reset times.
Choosing the Choose between Direct (no return for a specific value) Time in which each increment is
type of ramp. or Pulsed (at the end of a specific generation time the being generated and the time in
ramp goes to a specific value). which each value of the reset
time.
When clicking on the access button for the ramp parameterization, the following
screen appears. Choose in the "Ramp Type" the Frequency option. In the figure below, all
features that can be parameterized in this mode are characterized, such as direct or pulsed
type, initial ramp values, final values, increment, reset values (in case the pulsed type is
chosen), generation and reset times.
Choosing the Choose between Direct (no return for a specific Time in which each
type of ramp. value) or Pulsed (at the end of a specific generation increment is being generated
time the ramp goes to a specific value). and the time in which each
value of the reset time.
When clicking on the access button for the ramp parameterization, the following
screen appears. Choose the "Module and Angle" option in the "Ramp Type”. In the figure
below, all features that can be parameterized in this mode are characterized, such as direct or
pulsed type, initial ramp values, final values, increment, reset values (in case the pulsed type
is chosen), generation and reset times.
Choice of Choose between Direct (no return for a specific Time in which each increment is
ramp type. value) or Pulsed (at the end of a specific generation being generated and the time in
time the ramp goes to a specific value). which each value of the reset time.
Figure 567 – Rampa – Sequence Tab – Resources in the way and Angle Module
When clicking on the access button for the ramp parameterization, the following
screen appears. Choose the option Module and frequency in the "Ramp Type”. In the figure
below, all features that can be parameterized in this mode are characterized, such as direct or
pulsed type, initial ramp values, final values, increment, reset values (in case the pulsed type
is chosen), generation and reset times.
Choice of Choose between Direct (no return for a specific Time in which each increment
ramp type. value) or Pulsed (at the end of a specific is being generated and the time
generation time the ramp goes to a specific value). in which each value of the
reset time.
Figure 568 – Rampa – Sequence Tab – Resources Ramp in Module and Frequency mode
When clicking on the access button for the ramp parameterization, the following
screen appears. Choose in the "Ramp Type" the option dF/dt. In the figure below, all
features that can be parameterized in this mode are characterized, such as direct or pulsed
type, initial ramp values, final values, increment, reset values (in case the pulsed type is
chosen), generation and reset times.
Choice of Choose between Direct (no return for a specific Time in which each increment is
ramp type. value) or Pulsed (at the end of a specific generation being generated and the time in
time the ramp goes to a specific value). which each value of the reset time.
When clicking on the access button for the ramp parameterization, the following
screen appears. Choose “Symmetrical components” in the “Ramp type”. In the figure below,
all features that can be parameterized in this mode are characterized, such as direct or pulsed
type, initial ramp values, final values, increment, reset values (in case the pulsed type is
chosen), generation and reset times.
Choice of Choose between Direct (no return for a specific Time in which each increment is
ramp type. value) or Pulsed (at the end of a specific generation being generated and the time in
time the ramp goes to a specific value). which each value of the reset time.
Figure 570 – Rampa – Sequence Tab – Resources Ramp in order Symmetrical Component
When clicking on the access button for the ramp parameterization, the following
screen appears. Choose the “Phase-Phase” option in the “Ramp Type”. In the figure below,
all features that can be parameterized in this mode are characterized, such as direct or pulsed
type, initial ramp values, final values, increment, reset values (in case the pulsed type is
chosen), generation and reset times.
Choice of Choose between Direct (no return for a specific Time in which each increment is
ramp type. value) or Pulsed (at the end of a specific being generated and the time in
generation time the ramp goes to a specific value). which each value of the reset time.
When clicking on the access button for the ramp parameterization, the following
screen appears. Choose the “Power Ramp Type” option. In the figure below, all features that
can be parameterized in this mode are characterized, such as direct or pulsed type, initial
ramp values, final values, increment, reset values (in case the pulsed type is chosen),
generation and reset times.
Choice of Choose between Direct (no return for a specific Time in which each increment is
ramp type. value) or Pulsed (at the end of a specific being generated and the time in
generation time the ramp goes to a specific value). which each value of the reset time.
When clicking on the access button for the ramp parameterization, the following
screen appears. In the "Ramp Type", choose the Icc and Zlin Impedance option. In the figure
below, all features that can be parameterized in this mode are characterized, such as direct or
pulsed type, initial ramp values, final values, increment, reset values (in case the pulsed type
is chosen), generation and reset times.
/
Choice of Choose between Direct (no return for a specific Time in which each increment is
ramp type. value) or Pulsed (at the end of a specific generation being generated and the time in
time the ramp goes to a specific value). which each value of the reset
time.
Figure 573 – Rampa – Sequence Tab – Resources Ramp in mode impedance Icc and Zlin
When clicking on the access button for the ramp parameterization, the following
screen appears. Choose in the "Ramp Type" the option Vcc and Zlin impedance. In the
figure below, all features that can be parameterized in this mode are characterized, such as
direct or pulsed type, initial ramp values, final values, increment, reset values (in case the
pulsed type is chosen), generation and reset times.
Choose between Direct (no return for a specific Time in which each
Choice of
value) or Pulsed (at the end of a specific increment is being
ramp type.
generation time the ramp goes to a specific generated and the time
value). in which each value of
the reset time.
Figure 574 – Rampa – Sequence Tab – Resources Ramp mode impedance at Vcc and Zlin
When clicking on the access button for the ramp parameterization, the following
screen appears. Choose in the "Ramp Type" the SIR and Zlin Impedance option. In the
figure below, all features that can be parameterized in this mode are characterized, such as
direct or pulsed type, initial ramp values, final values, increment, reset values (in case the
pulsed type is chosen), generation and reset times.
Choose between Direct (no return for a specific Time in which each
Choice of
value) or Pulsed (at the end of a specific increment is being
ramp type.
generation time the ramp goes to a specific generated and the time
value). in which each value of
the reset time.
Figure 575 – Rampa – Sequence Tab – Resources Ramp mode impedance in SIR and Zlin
Sequence duration
time.
Sequence name.
Voltage and
current module.
Figure 576 – Rampa – Sequence Tab – DC Analog Outputs sub tab – General Description
Sequence duration
time.
Sequence
name.
Figure 577 – Rampa – Sequence Tab – Binary Outputs sub tab – General Description
Sequence duration
time.
Sequence
name.
Figure 578 – Rampa – Sequence Tab – GOOSE Outputs sub tab – General Description
Figure 579 – Rampa – Sequence Tab – Sub tab Time and Advance – General Description
Extract the
COMTRADE or CSV
Allows you to Insert or file from the waveforms.
Edit either Markings or Take a photo of the chart
Comments in graphics. by storing it on the
clipboard.
Clicking on the View option in the previous figure brings up the following options:
Clicking on the Show Columns option in the previous figure brings up the following
options:
Enables or disables
the display of the
following values:
Module, angle,
actual value,
imaginary value, lag,
instantaneous
amplitude, frequency
and frequency
variation.
Enables or disables the
display of the following
values for the difference
between the cursors:
Minimum, maximum or
average Instantaneous
Amplitude and the
RMS value in the
period.
Figure 585 – Rampa – Waveform Tab – Resources – Show Columns
It allows 4 different
sizes for visualization
of the graphs.
Type of fault
to be inserted.
Button to
Table with the insert a fault
various faults that sequence.
can be inserted.
When clicking on the insert button, the following screen will appear:
Fault insertion
Description of the mode (smart, Z-I
strings that cte, Z-V cte and Z-
represented a fault. Source cte).
After choosing the data in the figure above, just allocate the fault point with the
mouse in the impedance diagram, as shown in the figure below:
Confirmation
of the chosen
Table with several points point.
chosen and their
respective information.
MASTER
SOFTWARE TO GENERATE SEQUENCES OF
ALL TYPES OF SIGNS
The options on the Settings screen are described in detail in chapter 21. The
Distance, Timing, Import and Export fields will be discussed throughout this chapter.
Equipment used
and serial
Opens the
options: Restore, number.
Move, Size,
Minimize,
Maximize and
Close.
Creates a new
test file.
Open a saved
file.
Password protect
certain
parameters of the
test file.
Saves the current
file.
In the start menu dive resources are available. In the following figure are described
features of the hardware settings and editing sequence s such as insert, copy, delete, scroll
and select the reference.
Connects between Opens a screen for Deletes all Shifts the selected
software and GPS Setup. strings. Sequence to the beginning
hardware. or end.
The following are the settings related to the management of results (reissue of tests,
exclusion of tests and choice of test for analysis) and generation control.
In the figure below, the resources for analysis options of the generated quantities
(waveforms, phasors, trajectories, harmonics, syncoscope and impedance plane) are
displayed, as well as the adjustment options (General, distance and synchronism
settings). Access to the report, options for viewing the units involved (time and module) and
general options for restoring and viewing the layout are also displayed.
Calls the Shows the Shows the Shows the Opens the Shows the Shows the
Settings Phasor Waveform Harmonics report time in time in
Screen. Tab. Tab. Tab. configuration seconds. cycles.
window.
Shows
values
based on the
secondary.
Tabs for
viewing and
Standard
layout.
Shows values
Shows the Shows the based on the Shows
Synchronoscope Impedance Shows
Trajectory plan primary. relative
tab. absolute
tab. configuration. values. values.
Figure 601 – Master – Home Menu – Options, Report, Units and Layout
Position the selected Superimposes the Hide the windows Returns to Enables /
window: on the left; selected window. the standard
for better viewing. disables tabs
to the right, above, layout.
for viewing.
below or as a
document.
Hide the selected Hide the selected Close all open Closes all open
tabs. window. windows. documents.
It is also opened at software startup, if the Opens Settings when Starting option in the
Software Options tab is selected.
In General mode there is the General Information tab, where the user enters all the
data related to the test, the Test set device and the person responsible for the test, there is the
Systems tab, where the user informs the nominal values of frequency, voltage, current and
power among others, there is a tab for Notes and Observations for various records.
There is an Explanatory Figures tab where the user can insert up to three figures
relevant to the test, a Check List tab to create steps to be followed during the test, there is a
tab called Others for configuring digital signal filters and finally a Connections tab allowing
the user to structure all the connections between the test bag and the equipment to be Test
set.
In Distance mode, there is the Distance Protection Adjustment tab. In it the user sets
the line length and the angle. Choose how the earth factor will be inserted. You can still
enable whether to remove the Arc Resistance and use Impedance Correction for Rated
Current. To facilitate the adjustment of the zones, there is an Import option whose purpose is
to read files .RIO. Some distance relays record line, zone and operating time settings in this
format. When importing this file, the Master automatically reads it so that the user does not
need to adjust it manually. Another option would be to export a file .RIO once the feature
was entered by the user.
The Insert option opens a new window for the user to enter the name and number of
the zone, adjust the Trigger Time, Absolute and Relative Tolerances of the impedance and
the Absolute and Relative Tolerances for the time. You can use a specific earth factor and
enable the option to Remove Arc Resistance. If the relay does not generate a file .RIO there
is the option to design the zone to be Test set or to choose between predefined
zones. Depending on the type of zone chosen, options for load compensation and
directionality are opened.
Displays the
tab for
adjusting the
sync function.
Import
Files
.RIO
Export
Files
.RIO
Choose the
mode of the
earth factor:
(Mod;Ang)
(R;X)
(R0;X0)
(K0)
(Rt/Rl;Xt/Xl)
(Z0/Z1).
Shows which
zones are
available.
Insert a Edit a zone Deletes a zone that has Shows the design of
already inserted. already been inserted. the chosen zone.
new zone.
Let us see the window that opens when clicking on Insert, where it is made the
adjustments of the protection zones.
Figure 607 – Master – Parameters Screen – Distance – Distance Protection Adjustment – Insert – Zone
To facilitate the visualization of the previous figure, it was divided into parts. In this
first part the user can enter the description of the added zone, choose the type of actuation,
select which loop (detection algorithm) that zone belongs to and enter the absolute and
relative tolerances of time and impedance.
Zone description.
Zone type registration:
Zone index. Departure, Performance,
Extended or Block.
Links the zone to a (Informative data).
specific type of
fault. Absolute (Ω) and
relative (%) tolerance
of Impedance.
Trip time of the
zone.
Absolute tolerance
(more or less) in
seconds and relative
tolerance in% of the
time of operation.
Points inserted
when using the
option to draw.
Figure 608 – Master – Parameters Screen – Distance – Distance Protection Adjustment – Insert – Zone
In this second part the user has access to the various ways of drawing the desired
zone. The first form is drawing segment by segment, the second form uses known zones
(Mho, Quadrilateral and tomato / lens) where the user by comparison changes some intuitive
segments and the third form is using the presets of the relays themselves (Masks).
Finalizes the
drawing
process.
The following figure shows a quadrilateral zone that was drawn using straight line
segments.
In the next figure, the quadrilateral option was chosen directly. Note that the settings
to be entered are more intuitive with this option.
Then the screen where system data 1 and 2 and the settings for synchronism are inserted.
To facilitate understanding, the previous figure will be divided into two parts. Note
in the figure below that system one was adopted as the one to be synchronized. It is also
shown that it is possible to choose whether it is single-phase or three-phase, including the
reference phase, the angular offset of the elevation transformer (if the synchronism is done
in a Generator + Transformer system). Finally, the user can choose the way to start the
synchronism and parameterize the closing time of the circuit breaker (which will be taken
into account when calculating the relay).
Then system 2 is adopted as the side of the system that will receive the generation or
simply the infinite bar. In the figure below you can choose a three-phase or single-phase
system, the angular reference and the voltage and current levels in primary and secondary
values. Finally, you can change the phase sequence.
Relative and
absolute frequency
tolerances given by
the relay
manufacturer.
Relative and
absolute time
tolerances given
by the relay
manufacturer.
Figure 620 – Master – Sequence Tab - NODE sub tab - Two Nodes and three Sequences
In the following explanations, we will adopt that only one node (NO01) was created
containing three voltage channels and three current channels. Remember that to
add sequences click the Insert New Sequence. The figures below show the sequence modes
that can be entered in the test.
In this mode, the user directly enters the value of the module, angle and frequency of
each channel in the desired sequence.
Module
Choice of
value. operating
mode.
Description
of the
targeted
channel. Frequency
value.
Generic
point of the
System. Angle
value.
Figure 621 – Master – Sequence Tab - NODE sub tab - Description of the direct mode
In this mode, the user directly enters the module value and angle of each harmonic
component (up to 50th order). Note in the figure below that only the total RMS (true RMS)
and the total harmonic distortion of the quantity involved are displayed, to enter the values
as described, the user has to access the modules and angles as described below.
Choice of
operating
mode.
Description
of the Access to the
targeted Module and
channel. Angle of 50
harm orders.
Generic
point of the
System. RMS value
and Total
Harmonic
Distortion
(THD) of
the entered
harmonic
orders.
Figure 624 – Master – Sequence Tab – NODE sub tab – Description of the Harmonics mode
Harmonic Order.
By right-clicking on the text box on the Harmonics input screen in the Harmonics
mode, the following input data formatting features are available.
In this mode, the user directly enters the value of the module and angle of each
interharmonic component (up to 10 interharmonic).
Choice of
operating
mode.
Description of
the targeted Access to the Module
channel. and Angle of
interharmonical orders.
Generic
point of the RMS value and
System. Total Harmonic
Distortion (DHT) of
the inserted
interharmonic
orders.
Figure 627 – Master – Sequence Tab - NODE sub tab - Interharmonic mode description
When clicking on the access button for the parameterization of the module and angle
of the harmonics, the following screen appears.
Clears values of
module and angle
already
parameterized.
Frequency of the
interharmonic of Module value and angle
interest. of the inter-harmonic
order in question.
User-defined number
Default number of
of decimal places.
decimal places (two
places).
Definition of
symmetric Voltage and current
values. values in
symmetrical
component format.
Both module and
angle.
Figure 630 – Master – Sequence Tab – NODE sub tab – Description of the Symmetric Components
Mode
Values of phase-to-phase
Definition of voltage, Sequence zero
the Phase- voltage and current in
Phase value. the three phases.
Figure 632 – Master – Sequence Tab – NODE sub tab – Description of the Phase-to-Phase Mode
Reduced number of
Frequency entry
decimal places.
mode.
User-defined number
of decimal places.
Definition of
the Power Phase-to-ground
values. voltage values, single-
phase active and
reactive powers.
Figure 634 – Master – Sequence Tab – NODE sub tab – Description of the Power Mode
Reduced number of
Frequency entry
decimal places.
mode.
User-defined number
of decimal places.
Defining the
fault values. Fault types (three-
phase, two-phase or
single-phase), angle
between voltage and
current and module /
angle of fault voltage
and fault current.
Figure 636 – Master – Sequence Tab - NODE sub tab– Description of the Fault Mode
Reduced number of
Frequency entry decimal places.
mode.
User-defined number
of decimal places.
Figure 638 – Master – Sequence Tab - NODE sub tab– Description of the Constant ZI Mode
Reduced number of
Frequency entry
decimal places.
mode.
User-defined number
of decimal places.
Figure 640 – Master – Sequence Tab - NODE sub tab– Description of the Constant ZV Mode
User-defined number
of decimal places.
Figure 642 – Master – Sequence Tab – NODE sub tab – Description of the Z mode – constant source
Reduced number of
Frequency entry decimal places.
mode.
User-defined number
of decimal places.
Figure 644 – Master – Sequence Tab - NODE sub tab– Description of the Z Mode - Smart
Reduced number of
Frequency entry decimal places.
mode.
User-defined number
of decimal places.
Sequence Choice of
name. operating
mode.
Definition of
the waiting
sequence.
Access button
for adding
image and
auxiliary text.
Figure 646 – Master – Sequence Tab - NODE sub tab– Description of the Standby Mode
Insert an image or
change an existing
Area for text editing. one.
Deletes an existing
image.
At the time of the test when this waiting sequence is reached, the following screen
will appear.
Choice of
operating
mode.
Sequence
name.
Duration of the
transition in cycles.
Reference channel
for starting
damping.
Figure 649 – Master – Sequence Tab - NODE sub tab– Description of the Accommodation Mode
CONPROVE INDÚSTRIA E COMÉRCIO LTDA 494
CTC SOFTWARE MANUAL
Choice of Choose between Direct (no return for a specific value) Time in which each increment is
ramp type being generated and the time in
or Pulsed (at the end of a specific generation time the
which each value of the reset time.
ramp goes to a specific value).
Figure 651 – Master – Sequence Tab – Resources the ramp in Module mode
Choice of Choose between Direct (no return for a specific value) Time in which each increment is
being generated and the time in
ramp type. or Pulsed (at the end of a specific generation time the
which each value of the reset time.
ramp goes to a specific value).
Figure 652 – Master – Sequence Tab – Resources of the ramp in the Angle mode
Choosing the Choose between Direct (no return for a specific Time in which each increment is
being generated and the time in
type of ramp. value) or Pulsed (at the end of a specific generation
which each value of the reset time.
time the ramp goes to a specific value).
Figure 653 – Master – Sequence Tab – Resources of the ramp in order Frequency
Choosing the Choose between Direct (no return for a specific Time in which each increment is
type of ramp. being generated and the time in which
value) or Pulsed (at the end of a specific
each value of the reset time.
generation time the ramp goes to a specific value).
Figure 654 – Master – Sequence Tab – Resources of the ramp on the way and Angle Module
Choosing the Choose between Direct (no return for a specific Time in which each increment is
being generated and the time in
type of ramp. value) or Pulsed (at the end of a specific generation
which each value of the reset time.
time the ramp goes to a specific value).
Figure 655 – Master – Sequence Tab – Resources the ramp in mode Module and Frequency
Choice of Choose between Direct (no return for a specific value) Time in which each increment is
being generated and the time in
ramp type. or Pulsed (at the end of a specific generation time the
which each value of the reset time.
ramp goes to a specific value).
Figure 656 – Master – Sequence Tab – Resources of the ramp in mode dF/dt
Choosing the Choose between Direct (no return for a specific Time in which each increment is
being generated and the time in
type of ramp. value) or Pulsed (at the end of a specific generation
which each value of the reset time.
time the ramp goes to a specific value).
Figure 657 – Master – Sequence Tab – Resources of the ramp in mode Symmetrical Component
Choosing the Choose between Direct (no return for a specific Time in which each increment is
type of ramp. being generated and the time in
value) or Pulsed (at the end of a specific generation
which each value of the reset time.
time the ramp goes to a specific value).
Figure 658 – Master – Sequence Tab – Resources of the ramp in the Phase-to-Phase Mode
Choosing the Choose between Direct (no return for a specific value) Time in which each increment is
being generated and the time in
type of ramp. or Pulsed (at the end of a specific generation time the
which each value of the reset time.
ramp goes to a specific value).
Figure 659 – Master – Sequence Tab – Resources the ramp in Power Mode
Choosing the
Choose between Direct (no return for a specific Time in which each
type of ramp. value) or Pulsed (at the end of a specific increment is being
generation time the ramp goes to a specific generated and the time
value). in which each value of
the reset time.
Figure 660 – Master – Sequence Tab – Resources of the ramp in Impedance Icc and Zlin mode
Choose between Direct (no return for a specific Time in which each
Choosing the
value) or Pulsed (at the end of a specific increment is being
type of ramp.
generation time the ramp goes to a specific generated and the time
value). in which each value of
the reset time.
Figure 661 – Master – Sequence Tab – Resources of the ramp in Vcc and Zlin Impedance mode
Choose between Direct (no return for a specific Time in which each
Choosing the
value) or Pulsed (at the end of a specific increment is being
type of ramp.
generation time the ramp goes to a specific generated and the time
value). in which each value of
the reset time.
Figure 662 – Master – Sequence Tab – Resources of the ramp in SIR and Zlin Impedance mode
Choose between
COMTRADE file Access button for
(.cfg extension) parameterizing the
or ATP file (.lis comtrade / ATP file.
extension).
Shows information
Picture Search for file to upload. about the uploaded
box. file.
Shows information
Picture Search for file to upload. about the file.
box. carregado.
Sequence duration
time.
Sequence name.
Voltage and
current module.
Figure 666 – Master – Sequence Tab – Output Analog. DC Sub tab – General Description
Sequence duration
time.
Sequence
name.
Figure 667 – Master – Sequence Tab – Binary Outputs sub tab – General Description
relay and so on. It has the ability to change the state of the message between NO and NC
in any sequence (through a graphic resource), delay or advance the time of changing that
state.
Sequence duration
time.
Sequence
name.
Figure 668 – Master – Sequence Tab – GOOSE Outputs Sub tab – General Description
Figure 669 – Master – Sequence Tab –Time and Advance Sub tab – General Description
Clicking on the View option in the previous figure brings you to the following
options:
This screen has the Transparency property, and can remain open while points are
inserted or edited in the test. The following screen is for a line Test set using Search Test.
Clicking on the Show Columns option in the previous figure brings up the following
options:
Enables or disables
the display of the
following values:
Module, angle,
actual value,
imaginary value, lag,
instantaneous
amplitude, frequency
and frequency
variation. Enables or disables the
display of the following
values for the difference
between the cursors:
Minimum, maximum or
average Instantaneous
Amplitude and the
RMS value in the
period.
31.9.55Synchronoscope tab
With the voltage and current channels already directed with signals from the
COMTRADE / ATP file, sequence s or ramps, we can check through a synchronoscope if
these signals are able to perform a synchronism. Note in the figure below that the light blue
region is the synchronism region and is determined in the Settings.
Type of fault
to be inserted.
Button to
Table with the insert a fault
various faults that sequence.
can be inserted.
When clicking on the insert button, the following screen will appear:
Fault insertion
Description of the mode (smart, Z-I
Sequences that cte, Z-V cte and Z-
represented a fault. Source cte).
After choosing the data in the figure above, just allocate the fault point with the
mouse in the impedance diagram, as shown in the figure below:
Confirmation
of the chosen
Table with several points point.
chosen and their
respective information.
SOBRECOR
SOFTWARE FOR TESTING OVERCURRENT
RELAYS
The options on the Settings screen are described in detail in chapter 21.
The field Sobrecor is will discussed throughout this chapter.
Creates a new Opens a saved Saves the Customize the Software name and
test. current test. Quick Access version.
test file.
Bar.
Creates a new
test file.
Open a saved
file.
Password protect
certain
parameters of the
test file.
Saves the current
file.
In the start menu, several resources are available . The following figure describes
hardware configuration features and generation controls such as start, stop, move to the next
point / line, deletion of one or all of the Test set points.
Opens a Channel Opens a screen Starts the Pauses signal Move on to the
Orientation screen. for Hardware signal generation. next point.
Configuration. generation
process.
Clears the
selected
point and
returns to
the status of
Not Tested.
In the figure below, the resources for analysis options of the generated quantities
(waveforms and phasors) are displayed, as well as the adjustment options (General Settings
and overcurrent). Access to the report, options for viewing the units involved (time and
module) and general options for restoring and viewing the layout are also displayed.
Shows
values
based on the
secondary.
Opens the
Calls the report Shows values Shows Shows
Settings configuration based on the absolute relative
Screen. window. primary. values. values.
Figure 690 – Sobrecor – Home Menu – Options, Report, Units and Layout
Hide the selected Hide the selected Close all open Closes all open
tabs. window. windows. documents.
Insert password in Enables / disables the Opens the Settings Shows the
the Test file so as simulation mode. Since Screen when the manufacturer’s
not to allow the in this mode there is no software is started. address, phone,
editing of certain insertion of voltage and email and fax.
settings. current.
It is also opened at software startup, if the Opens Settings when Starting option in the
Software Options tab is selected.
In General mode, there is the “General Information” tab, where the user enters all the
data relating to: the Test, the Test set Device and the person responsible for the Test, there is
the “Reference Values” tab, where the user informs the nominal values of voltage, current
and power for system 1, there is also a tab for “Notes and Observations”. In the other tab
“Explanatory Figures” it is possible to insert up to 3 figures in the most used image
extensions, such as: Bitmap; JPGE; PNG; JPG; etc.
In the Sobrecor mode, there is the “Settings” tab. Here the user can skilful it air
directionality and polarization mode. For the time test, it is possible to define whether the
time considered is that of the shortest curve or if it will respect the characteristic of each one.
You can choose between automatic or user-defined scale, adjust the values
of the negative sequence and zero sequence multipliers and configure the current, time and
angle tolerances associated with the test.
There is another tab called "overcurrent elements” which configures the phase
elements, residual sequence and negative sequence zero. Considering a make element, for
example, you can adjust it as function 51, 50 or 67.
following figure shows three phase elements, the graph of which is shown in the Total
Characteristic tab.
Selecting one of the elements, you can adjust particular characteristics for each
element. In the figure below, the individual characteristic of the inverse time unit (51) is
shown, on this screen the description, the choice of the standardized curve, the adjustment of
the time dial, the pickup value and the drop-out ratio are made.
Note two specific settings: pickup test multiple (some relays apply this multiple to
their already parameterized pickup) and ∆t (some relays apply this waiting time after the
curve time is reached).
The figure below shows the screen where the individual directionality of the selected
element is parameterized. Note that the polarization quantity is not available, as it has
already been selected in the "Settings" tab. It is observed that for each element (50 or 51)
different adjustments can be applied (directionality, ATM, Vmin and etc.). This is due to
some relays that allow such an adjustment in the 50/51 directionality.
After entering the settings, these are shown in the Overcurrent Chart tab.
Extract the
COMTRADE or CSV
file from the waveforms.
Allows you to Insert or
Take a photo of the chart
Edit either Markings or by storing it on the
Comments in graphics. clipboard.
Clicking on the Display option in the previous figure brings you to the following
options:
Clicking on the Show Columns option in the previous figure brings up the following
options:
Enables or disables
the display of the
following values:
Module, angle, actual
value, imaginary
value, lag,
instantaneous
amplitude, frequency
and frequency Enables or disables the
variation. display of the following
values for the difference
between the cursors:
Minimum, maximum or
average Instantaneous
Amplitude and the
RMS value in the
period.
The previous figure shows the result of a Pickup test. For a better explanation of this
test, this figure is divided into parts.
For each Test set point the user can choose what type of data will be shown in the
table according to the figure below.
Choose how the tests Shows which element Shows the nominal Shows the percentage
will be shown. Each is being tested. and actual DropOut and absolute errors of
point individually or in value, that is, what Pickup and DropOut.
a group. was found.
Shows the type Informs the value of the Shows the nominal Shows the voltage
of fault and the nominal, minimum, DropOut factor and current values
angle when there maximum pickup and and the one found. injected in the test.
is directionality. what was found.
As an example leaving the options “Fault & Ang”, “Ref Curve” and “Info Pickup” of
the previous figure marked, the following data are shown:
Edit a point
already tested.
Creates a
new point
Not available for test.
for this test.
Removes the
Removes all selected
tested points point.
Figure 711 – Sobrecor – Main screen – Pickup Test – Insert / Edit Points
When the user clicks on “New Point” it is possible to choose whether DropOut will
be Test set and for which type of fault it is desired to do the Pickup and / or DropOut test.
The following figure shows the plot of the points before they are Test set. The
moment they are Test set they turn green if they pass or red if they fail.
The next illustration shows the legend for the graph of the performance of the inverse
curve and the time defined in the software.
Colors representing
status of points:
Orange: Not Tested;
Green: Approved;
Red: Failed.
The previous figure shows the result of a search test. To facilitate the understanding,
each tab of this test will be shown in detail in the next figures. For each row the user can
choose what type of data will be shown in the table according to the figure below.
Choose how the test will Shows which Shows Shows the expected Shows the voltage
be shown. Each point multiple is which time and the time and current values
individually or in a being tested. element is that was found. injected in the test.
group. being tested.
Shows the type Shows the fault Shows the nominal Shows the
of fault of the current and the angle DropOut factor percentage and
tested point. when there is and the one found. absolute errors of
directionality. the operating time.
As an example only leaving Options "missing" "Curve ref ", "Time" and "Error” of
the previous figure are shown the following data:
Test Type Element Time set Time Percentage and absolute Test
point of under in the found in error between nominal validati
number. fault. test. relay. the test. and real time. on.
Figure 717 – Sobrecor – Main screen – Time Test – Test set Points
Creates a new
line.
Creates a
sequence of
lines.
Removes
selected line.
Removes all
added lines.
Figure 718 – Sobrecor – Main screen – Time Test – Insert / Edit Lines
When the user clicks on “New Line” the settings of the fields shown below are
available.
Enables / disables the option to Type of fault Value of the Current value in
choose which element your pickup to be tested. multiple of the amps for the test.
will be used as a multiple base. pickup for the
test.
When clicking on “Sequence” the user defines an initial, final and step current value.
In this way, several lines are automatically inserted.
Choose one
or more types
of fault.
Figure 720 – Sobrecor – Main screen – Time Test – General Options - Sequenc
The following figure shows the graph of the lines before they are Test set. The
moment they are Test set they become green points if they pass or red points if they fail.
Colors representing
status of points:
Orange: Not Tested;
Green: Approved;
Red: Failed.
In the following figure, the designation of channels and the insertion of pre and post
fault are detailed.
Enables pre-simulation
1 and 2 and post-
simulation conditions.
Designates the generation Module and angle values of each Module and angle values of
channels either current or channel for pre-simulation each channel for post-
voltage. condition - 1 and 2. simulation condition.
Figure 724 – Sobrecor – Main screen – Test Settings – Direction Generation Channels – Pre-Simulation
and Post-Simulation
The figure below shows the location in the Test Configuration tab where the user
determines the status of each binary output (BO), each GOOSE output (GO) and the pre-
and post-fault time.
Defines the states of the Defines the states of the Defines the states of the Defines the states of the
Binary Outputs and Binary Outputs and Binary Outputs and Binary Outputs and
GOOSE Outputs (when GOOSE Outputs (when GOOSE Outputs (when GOOSE Outputs (when
configured) in the configured) in the Pre- configured) in the Pre- configured) in the Post-
Simulation cycle. Simulation 1 cycle. Simulation cycle 2. Simulation cycle.
Figure 725 – Sobrecor – Main screen – Test Settings – Pre-Simulation, Post-Simulation Time and Binary
and GOOSE Outputs.
The figure below shows the line voltage / phase voltage adjustments applied to three-
phase, two-phase and ground-to-earth faults in directional protections. The ratio between the
CONPROVE INDÚSTRIA E COMÉRCIO LTDA 547
CTC SOFTWARE MANUAL
earth and phase transformation ratios for TPs and TCs is adjusted. Finally, the user can
invert the polarities of the TCs and TPs.
Group of
phase
elements,
residual,
positive,
negative
and zero
sequence.
Stop
interface
Element for the
number. trip
signal.
Stopwatch trigger interface by: Insert logic between binary Trigger time
software, GPS, or by binary logic. inputs or GOOSE messages. delay.
Allows “E” or “OR” logic.
DIFERENC
SOFTWARE FOR TESTINGS IN
DIFFERENTIALS RELAYS
This software has the feature "Open settings at startup" ( Software Options menu ) ,
if it is active when starting the software Diferenc the screen settings becomes active
immediately, this feature is interesting since, before starting the tests it is Important to
specify the parameters related to the test.
The options on the screen settings are described in detail in Chapter 21. Field
Diferencial will be discussed throughout this chapter.
Creates a new Opens a saved Saves the Customize the Software name
test file. test. current test. Quick Access and version.
Bar.
Opens the
options: Equipment used
Restore, and serial
Move, Size, number.
Minimize,
Maximize and
Close.
Creates a new
test file.
Open a saved
file.
Password protect
certain
parameters of the
test file.
Saves the current
file.
In the start menu dive resources are available. The following figure describes
hardware configuration features and generation controls such as start, stop, move to the next
point / line, deletion of one or all of the Test set points.
Opens a Channel Opens a screen Starts the Pauses signal Move on to the
Orientation for Hardware signal generation. next point.
screen. Configuration. generation
process.
Clean the
spot
Selected
and returns
to Untested
status.
Clears all
spots
and
returns to
the status
Opens a screen Setup of sample of Not
Connects between for GPS Setup. value and goose Tested.
software and messages.
hardware.
In the figure below, the resources for analysis options of the generated quantities
(waveforms, phasors, System represented by the case, internal currents to the relay and
special adjustments for tests in the SEL relay) are displayed, as well as the adjustment
options (General Settings and Differential). Access to the report, options for viewing the
units involved (time and module) and general options for restoring and viewing the layout
are also displayed.
Calls the Shows the Shows the Shows the Shows the Shows the Returns to
Settings Waveform Phasor System that a time in time in the standard
Screen. Tab. Tab. test set will seconds. cycles. layout.
simulate.
Enables /
Shows the Opens the Restore all disables tabs
Special details of the report graphics.
settings for for viewing.
relay before configuration
SEL relays. and after window.
compensation.
Figure 736 – Diferenc – Home Menu – Options, Report, Units and Layout
Position the selected Superimposes the Hide the windows Returns to Enables /
window: on the left; selected window. for better viewing. the standard disables tabs
to the right, above, layout. for viewing.
below or as a
document.
Hide the selected Hide the selected Close all open Closes all open
tabs. window. windows. documents.
Insert password in Enables / disables the Opens the Settings Shows the
the Test file so as simulation mode. Since Screen when the manufacturer’s
not to allow the in this mode there is no software is started. address, phone,
editing of certain insertion of voltage and email and fax.
settings. current.
Displays
information
about which
trip signal
should be
captured on
earth faults.
It is also opened at software startup, if the Opens Settings when Starting option in the
Software Options tab is selected.
In General mode , there is the General Information tab where the user enters all the
data related to the test, the Test set device and the person responsible for the
test. There is also the Systems tab where the user informs the nominal values of frequency,
voltage, current and power, among others. And, there is a tab for Notes and Observations for
miscellaneous records.
There is an Explanatory Figures tab where the user can insert up to three figures
relevant to the test, a Check List tab to create steps to be followed during the test, there is a
tab called Others for configuring digital signal filters and finally a Connections tab allowing
the user to structure all the connections between the test bag and the equipment to be Test
set.
In Differential mode, there is the Protected Equipment / CTs tab and the Differential
Protection Adjustment tab. In the Protected Equipment / CTs tab, the user chooses the
protected equipment (Transformer, Generator, Line, Bus or Motor), the rated power, and the
rated voltage. If the equipment is a transformer, it is necessary to inform the connection, the
vector group and on which side the ground is.
In this same tab it is possible to parameterize the CTs that make the measurements on
both sides and the data that must be entered are: primary current, secondary current and the
connection. If the connection of these CT banks is for electromechanical relays (option D or
Y), their vector group can be adjusted. Finally, it is possible to parameterize the connection
of auxiliary CTs, which are used for the correction of mismatch in the electromechanical
differential protection, the parameterization data are: primary current, secondary current,
connection and vector group.
In the Protection Adjustment tab, there are two “sub-tabs”: Adjustments and Slope
Definition. In the Settings sub tab, the user can define a pre-registered relay in the “data
entry” field, in addition to the differential and instantaneous adjustments along with their
operating times.
If the user tests equipment that does not exist in the register, it is possible and
necessary for the user to enter data such as: restriction current, lag compensation, tap
compensation and zero sequence elimination. Finally, in this tab the user can enter time,
current and angle.
In the “Slope definition” sub tab, the user defines between a single Slope (slope) or
several slopes and defines what the Idiff x Irest characteristic is through points in the
graph. If the user uses a relay already registered, this tab will show parameterization options
with the same name that appears in the relay parameterization software.
In the figure are shown are the Adjustments dealing with the nominal data of the
protected equipment and the set of CTs that perform measurements for protection, as
previously mentioned.
Voltage level.
Equipment
Winding
power. Vector or lag
connection
group
(Transf. only).
(Transformers
only).
Primary current
of the CTs. CT bank connection
(neutral point
Secondary closing side).
Current of the
CTs.
In the figure are shown are the adjustments dealing with the nominal data of the
protective relay and adjustment that are part of the differential function, as previously
mentioned.
Choose between a Differential pickup Snapshot pickup and Current, time and
preset relay or user and time setting time setting (87-2). angle tolerance
mode (all (87-1). (Relative and
compensations are Absolute).
editable).
Choosing the
method of
calculating the
restraint current.
Parameterization
of the module
compensation.
Parameterizati
Enables or does not on of the
enable Zero angle
Sequence compensation.
compensation.
In the figure is shown below the screen where they are entered are the data of the
differential characteristic (Irest x Idiff).
Definition of the Fixed slope If the user use the pre-setting, the
number of slopes (if value, if only 1 fields below will come according to
the user uses the user slope is used. the relay (number of slopes, specific
mode). names and etc)
Curve editing
Characteristic
Idiff x Irest.
To appear this tab, click on the top Home tab and click on the Systems icon. To
maximize the visualization, double click on the created Zxt tab. After inserting the desired
characteristic, a diagram is shown in the Systems tab in which the relay is inserted. When
test points are inserted, the current flow is indicated.
Double clicking on the Graph tab maximizes this screen. In it, the user sees the
differential characteristic parameterized in the settings.
Maximizes the
Place a test point in graphics tab.
the desired position.
Extract the
COMTRADE or CSV
Allows you to Insert or file from the waveforms.
Edit either Markings or Take a photo of the chart
Comments in graphics. by storing it on the
clipboard.
Clicking on the Display option in the previous figure brings you to the following
options:
This screen has the Transparency property, and can remain open while points are
inserted or edited in the test. The following screen is for a point Test set using Point Test (It
will be explained below).
Clicking on the Show Columns option in the previous figure brings up the following
options:
When clicking on the New Point button on the Insert / Edit Points tab, the
selections in the General Options field are active, where the user can select the type of
Data Entry:
Short current;
Winding currents;
IDif and IRest.
The user can also define which winding the Source is in and, consequently, the Short.
Also in this tab, the user can insert one of points, or edit the inserted points, or even
remove them.
It is also possible to view the illustrative drawing of the system under test, details of
the Relay in the form of Diagram, Characteristic Curve, injected waveform and phasors.
On the left in the Main Screen, a table is shown with the Test set Points and the
current status of each one of them.
After inserting the test points, just click on generate, and the software automatically
tests all points.
The figure below shows how the insertion, partial / total removal and editing of test
points are done.
Insert a new
point for test in
the table.
Insert a Sequence of points in
the table for the test.
Removes the
selected point
from the tested Removes all points from
points table. the test table.
Figure 756 – Diferenc – Main screen – Test Set – Insert / Edit Points
In the figure below, a test point was inserted in the configuration test, all fields are
detailed:
Figure 757 – Diferenc – Main screen – Test Set – Test sets Points
The way in which data is entered for a test point of the type “Short currents” is
detailed below.
Figure 758 – Diferenc – Main screen – Test Set – General options – Short current
The way in which data is entered for a test point of the type “Currents in windings” is
detailed below.
Cancel point or
sequence of points,
that is, do not insert
Confirms the Point in the Table of
or the Sequence of Tested Points.
Points entered.
Below is detailed the way in which data is entered for a test point of the type “Idiff
and Irest (differential current and restraint current)”.
Enables Load and Windings 3 and 4 if any. They are active when
windings are selected on the Parameters Screen. Confirms the Point
or the Sequence of
Points entered.
The user selects which winding is
short and which is the Source.
Cancel point or sequence of points, that is, do not insert
in the Table of Tested Points.
Figure 760 – Diferenc – Main screen – Test Set –General Options – IDif and IRest
Following figures below is shown the manner already described that the flaps mind
remained after inserting a test point of any type (short currents, currents in the windings or
differential current and constraint current).
Figure 762 – Diferenc – Main screen – Test Set – Relay Details (Diagram)
For this test the Sequence button is disabled, it is only possible to insert points,
inserting them one by one all the points to be Test set. This test allows the user to test points
in the Operating and Non-Operating Regions of the relay characteristic. After entering the
test points, just click on generate, and the software automatically tests all points. In
the figure below is detailed all the items that appear in the Test set points table.
Figure 767 – Diferenc – Main screen – Point Test– Test set Points
The figure below shows how the insertion, partial / total removal and editing of test
points are done.
Edit selected
point.
Insert a new test
point.
Insert a sequence
of points.
Removes selected
point. Removes all
stitches.
Figure 768 – Diferenc – Main screen – Point Test – Insert / Edit Points
The way in which data is entered for a point in the point test is detailed below.
In the figure below, all the items that appear in the Test set thread table are detailed.
Figure 773 – Diferenc – Main screen – Search Test – Test set Points
The figure below shows how the insertion, partial / total removal and editing of test
points are done.
Figure 774 – Diferenc – Main screen – Search Test – Insert / Edit Points
The way in which data is entered for a test point in the point test is detailed below.
Restraint current
reported by the user.
Confirms
inserted
points.
Selects the Type
of fault. Cancels the insertion of the
point or the Sequence.
In the Test Settings tab, the user directs the current and voltage generation channels
for each winding, being able to enable up to two pre-fault conditions.
It is also in this tab that the user defines the waiting time for entering the
Configuration Test data and the Tolerance multiples, the absolute and relative resolutions for
the Search Test, and the waiting and reset times for all tests.
The user can also direct Fault and Pre-fault situations to the Binary
and GOOSE outputs associated with each condition.
In this tab, the nominal working frequency and the phase sequence of the relay are
selected. Selections are also made for the Stop and Trigger interfaces, which, just like in the
other software CE-600X and CE-70XX, can be performed by one of the Binary
Inputs, GOOSE, or by logical combinations.
The figure below shows the test settings tab. This screen will be divided into the
following figures to facilitate understanding.
Below is shown how to designate the channels for each winding, and enabling two
pre-faults along with the mode, modules, angles and times.
The user enters the amplitude and The user enters the amplitude and
Winding angle values of each channel for angle values of each channel for
identification. the Pre-fault 1 condition. the Pre-fault 2 condition..
Identifies each
phase of each
winding. The user selects which
generation channel to direct to Time of
each phase of each winding. each Pre-
fault.
Below is shown the way to change the states of the binary outputs
and GOOSE outputs of the case in the fault and pre fault situations.
Defines the states of the Defines the states of the Binary Defines the states of the Binary
Binary Outputs and GOOSE Outputs and GOOSE Outputs Outputs and GOOSE Outputs
Outputs (when configured) (when configured) in the Fault (when configured) in the Pre-
in the Fault cycle. cycle. fault 2 cycle.
Figure 779 – Diferenc – Main screen – Pre-Fault Time and Binary Outputs and GOOSE
Below is shown the way to change the software interface stopping, starting interface
and access for use of logic in these two resources (stopping and starting).
The following figure details general options for the configuration test, point test,
search test and other configurations.
Operation Timeout.
After the theoretical
time has exceeded, the
suitcase continues to
generate for that time. Defines Absolute
Resolution for
Search Test.
Defines Relative
Operation Resolution for
Timeout. After Search Test.
the theoretical
time has
exceeded, the Reset time for all Tests.
suitcase continues
to generate for
that time.
RESTHARM
SOFTWARE FOR HARMONIC RESTRICTION
TESTS
The options on the Settings screen are described in detail in chapter 21. The
Harmonic Restrain fields will be discussed throughout this chapter.
Creates a new
test file.
Open a saved
file.
Password protect
certain
parameters of the
test file.
Saves the current
file.
In the start menu, several resources are available . The following figure describes
hardware configuration features and generation controls such as start, stop, move to the next
point / line, deletion of one or all of the Test set points.
Opens a Channel Opens a screen Starts the signal Pauses signal Move on to the
Orientation screen. for Hardware generation generation. next point.
Configuration. process.
Clean the
spot
Selected
and returns
to the status
of Not
Tested.
Clears all
spots
and
returns to
the status
of Not
Connects between Tested.
Opens a screen
software and
for GPS Setup.
hardware.
In the figure below, the resources for analysis options of the generated quantities
(waveforms and phasors) are displayed, as well as the adjustment options (General
Settings). Access to the report, options for viewing the units involved (time and module) and
general options for restoring and viewing the layout are also displayed.
Calls the Shows the Shows the Shows the Shows the Returns to
Settings Waveform Phasor Tab. time in time in the standard
Screen. Tab. seconds. cycles. layout.
Enables /
disables tabs
Opens the report for viewing.
configuration
window.
Figure 788 – RestHarm – Home Menu – Options, Report, Units and Layout
Position the selected Superimposes the Hide the windows Returns to Enables /
window: on the left; selected window. for better viewing. the standard disables tabs
to the right, above, layout. for viewing.
below or as a
document.
Hide the selected Hide the selected Close all open Closes all open
tabs. window. windows. documents.
Insert password in Enables / disables the Opens the Settings Shows the
manufacturer’s address,
the Test file so as simulation mode. Since Screen when the
phone, email and fax.
not to allow the in this mode there is no software is started.
editing of certain insertion of voltage and
settings. current.
It is also opened at software startup, if the Opens Settings when Starting option in the
Software Options tab is selected.
In General mode there is the General Information tab, where the user enters all the
data related to the test, the Test set device and the person responsible for the test, there is the
Systems tab, where the user informs the nominal values of frequency, voltage, current and
power among others, there is a tab for Notes and Observations for various records.
There is an Explanatory Figures tab where the user can insert up to three figures
relevant to the test, a Check List tab to create steps to be followed during the test, there is a
tab called Others for configuring digital signal filters and finally a tab Connections allowing
the user to structure all connections between the test case and the equipment to be Test set.
In Harmonic Restriction mode there is the Harmonic Restriction Adjustment tab. For
each harmonic (second, third or fifth order) Test set there is a sub- tab. In these sub-tabs the
user chooses in which reference the markings in the% Harm x Idiff graph (In or TAP) will
be made, edit the curve informing from which percentage of harmonic the relay does the
blocking. For CrossBlock it is possible to parameterize the blocking time and minimum
percentage for blocking for infinite time. It is necessary for the user to parameterize the
initial (87-1) and final (87-2) limits, as well as their operating times. Finally, the user has the
option of entering current and time tolerances.
In the figure below, the entire window where the parameterization of the harmonic
restriction test is performed is detailed. The user can see that what is done for the 2nd
harmonic is also repeated for the 3rd and 5th harmonic.
Choice of the reference Sub tab with the parameters Reference value, which
for marking on the of the harmonic that will be can be by In or TAP.
graph. In or TAP. tested.
Crossblock
time and the%
harm value for
which the
block is
infinite (some
relays).
Initial value
(87-1) and
final value
(87-2) with
their
respective
times.
Relative and absolute tolerances
of current and time.
Extract the
COMTRADE or CSV
Allows you to Insert or file from the waveforms.
Edit either Markings or Take a photo of the chart
Comments in graphics. by storing it on the
clipboard.
Clicking on the Display option in the previous figure brings you to the following
options:
Clicking on the Show Columns option in the previous figure brings up the following
options:
Enables or disables
the display of the
following values:
Module, angle,
actual value,
imaginary value, lag,
instantaneous
amplitude, frequency
and frequency Enables or disables the
variation. display of the following
values for the difference
between the cursors:
Minimum, maximum or
average Instantaneous
Amplitude and the
RMS value in the
period.
Current, percentage ratio between harmonic current and applied current and the angle of the
inserted point will be filled in the Currents or the user can enter each of these values in their
respective fields and the inserted point will be marked on the Characteristic graph.
After entering all the points he wants to test, the user must click on the Generate
button for the software to start the tests automatically.
In the figure below it is detailed all the items that appear in the Test set points table.
Current situation of
Restraint current Location of the Performance time
the point in test.
value, percent point on the found.
harmonic at each Characteristic curve.
point.
The figure below shows how the insertion, partial / total removal and editing of test
points are done.
Edit selected
point.
Insert a new test
point.
Copy selected point.
Removes selected
point.
Removes all
stitches.
Figure 802 – RestHarm – Main screen – Point Test – Insert / Edit Points
The way in which data is entered for a test point in the point test is detailed below.
In the figure below an overview of the search test is displayed, it will be detailed in
the following parts in order to facilitate the understanding.
In the figure below, all the items that appear in the Test set thread table are detailed.
Theoretical percentage
value.
Figure 807 – RestHarm – Main screen – Search Test – Test set Points
The figure below shows how the insertion, partial / total removal and editing of test
points are done.
Figure 808 – RestHarm – Main screen – Search Test – Insert / Edit Points
In the figure below, all the items that appear in the Crossblock test points table
are detailed.
Number of Tested
Points.
Current
situation of
each point.
Value of the applied
current.
Real time elapsed between the
start of the Generation and the
Percent harmonic restriction current calculated relay blocking action.
by the software for each point. Expected Lockout
Time.
Figure 811 – RestHarm – Main screen – Crossblock Test – Test set Points
Removes selected
point. Removes all
stitches.
Figure 812 – RestHarm – Main screen – Crossblock Test – Insert / Edit Points
Detailed below is a man threshing and data entry is made to a point test in
test Crossblock.
time for all tests. The user can also direct Fault, pre-fault and post-fault situations to the
Binary and GOOSE outputs associated with each condition.
In this tab, the nominal working frequency and the phase sequence of the relay are
selected. Selections are also made for the Stop and Trigger interfaces, which, just like in the
other software CE-600X and CE-70XX, can be performed by one of the Binary
Inputs, GOOSE, or by logical combinations. In the figure below, all channel routing,
enabling two pre- faults and enabling a post fault is detailed. It also indicates the location
from which to set the time for each pre / post fault.
Cycle duration.
Lag angle between Selects whether there will be a post-fault cycle, and defines the
voltage and current in the voltage and current conditions for this cycle. Attention: The
pre / post fault. performance in the Post-Fault is only considered and analyzed in
the condition in which the fault currents are maintained, removing
only the Harmonics (Missing without Harm.).
Selects the current channel for
generating each phase. (The
connections of the CE-600X must
be made respecting this selection).
N these fields the user defines the modules and angles of the values that will be generated in each channel
for the Pre-fault cycles. (In case the mode is NOMINAL, that field is blocked).
Figure 815 – RestHarm – Main screen – Test Settings – Direction of Generation Channel
In the figure below is showing the process to change the state of binary outputs
and GOOSE each pre - fault cycle.
Defines the state of the Defines the state of the Defines the status of the Binary
Binary and GOOSE outputs Binary and GOOSE outputs and GOOSE Outputs for the
for the Pre-fault cycles. post-fault cycle.
for the Fault cycle.
Figure 816 – RestHarm – Main screen – Test Settings – Binary Outputs & GOOSE
Opens the
Stopwatch Logic
Screen.
Defines a delay
Selects whether the interfaces (trip and stop) will be between the
activated via change of state or voltage level at the Opens the Logic activation of the
trigger interface and
Binary Inputs, or will be by Logic, or disable the Shooting
the effective start of
Stop and enable the software trip. Screen. generation.
In the figure below, all the configurations and parameterizations inherent to the
specific search test and to the other tests are detailed.
Disregard
Characteristic%
Harm x IDiff
Enables
resetting the
timer at each
search point.
Defines current
multiplier for generation Sets the reset
in phases without time for all Tests.
Harmonics for the
CrossBlock Test.
DIRECPOT
SOFTWARE FOR TESTING DIRECTIONAL
POWER RELAYS
The options on the screen settings are described in detail in Chapter 21. The Field
Directional Power is will discussed throughout this chapter.
Creates a new test file. Opens a Saves the Customize the Name and
saved test. current test. Quick Access version of the
Bar. software.
Creates a new
test file.
Open a saved
file.
Password protect
certain
parameters of the
test file.
Saves the current
file.
In the start menu, several resources are available. In the following figure are
described hardware configurations of resources and generation of controls such as
in t cesium, stop, transition to next point / line, excluding one or all the Test set points.
Clean the
spot
selected and
returns to
the status of
Not Tested.
Clears all
spots
and
returns to
the status
Connects between Opens a screen Setup of sample Go to the of Not
software and for GPS Setup. value and goose next line. Tested.
hardware. messages.
In the figure below, the resources for analysis options of the generated quantities
(waveforms and phasors) are displayed, as well as the adjustment options. Access to the
report, options for viewing the units involved (time and module) and general options for
restoring and viewing the layout are also displayed.
Calls the Shows the Shows the Shows the Shows the Restore all Returns to
Settings Waveform Phasor time in time in graphics. the standard
Screen. Tab. Tab. seconds. cycles. layout.
Enables /
disables
tabs for
viewing.
Shows
values
based on the
secondary.
Opens the
report
configurati Shows values Shows
Shows the on based on the absolute Shows relative
Synchronism tab. window. primary. values. values.
Figure 825 – DirecPot – Home Menu – Options, Report, Units and Layout
Position the selected Superimposes the Hide the windows Returns to Enables /
window: on the left; selected window. the standard
for better viewing. disables tabs
to the right, above, layout.
for viewing.
below or as a
document.
Hide the selected Hide the selected Close all open Closes all open
tabs. window. windows. documents.
Insert password in Enables / disables the Opens the Settings Shows the
manufacturer’s address,
the Test file so as simulation mode. Since Screen when the
phone, email and fax.
not to allow the in this mode there is no software is started.
editing of certain insertion of voltage and
settings. current.
It is also opened at software startup, if the Opens Settings when Starting option in the
Software Options tab is selected.
In General mode, there is the “General Information” tab, where the user enters all the
data relating to: the Test, the Test set Device and the person responsible for the Test, there is
the “Reference Values” tab, where the user informs the nominal values of voltage, current
and power for system 1 , there is also a tab for “Notes and Observations”. In the other tab
“Explanatory Figures” it is possible to insert up to 3 figures in the most used image
extensions, such as: Bitmap; JPGE; PNG; JPG; etc.
In the Directional Power mode there is the “Settings” tab. In it, the user
configures the pickup adjustment mode. Adjusts the relative and absolute tolerance
values for power time and angle. You can also limit a maximum or minimum value of
voltage and current together with the relative and absolute tolerances.
There is another tab called “Directional Power Elements” where the active power,
reactive power and apparent power elements are configured. For each element, the pickup
value, operating time and directionality must be configured for each element.
In this screen the user can register the elements of Active power, Reactive power and
Apparent power. It should If configure individual directionality each element , forward or
reverse, trip or trip the minimum to the maximum, the maximum torque angle or ATM such
as normal or reversed define a specific angle. You can configure the W0 factor to
compensate for measurement errors of the angle and finally define the positive and negative
angular lag in relation to the ATM.
Enables / disables
Maximizes the the display of the
graphics tab. line in the Search
Test.
The origin is
centered within the
Zooms in or out for
graph tab.
easy viewing of the
test.
Shows a grid for
easy viewing.
Central axes are in
bold for greater
emphasis.
Extract the
COMTRADE or CSV
Allows you to Insert or file from the waveforms.
Edit either Markings or Take a photo of the chart
Comments in graphics. by storing it on the
clipboard.
Clicking on the Display option in the previous figure brings you to the following
options:
Clicking on the Show Columns option in the previous figure brings up the following
options:
35.14 Shooting
This test aims to allow the user to test whether the operation occurs for a given
power or not. When in a region of operation it must operate within a predicted time plus the
tolerances.
The previous figure shows the result of a Shooting. For a better explanation of this
test, this figure is divided into parts.
For each test set point the user can choose what type of data will be shown in the
table according to the figure below.
Shows the Shows the Shows Shows the Shows Operating Voltage
phases power the reference whether the time. and
involved in values. power element. point acted current
the Test and factor. or not. values.
the angle of
the System.
For example leaving only the s options "Performance" and "Time" of the previous
figure selected the following data are shown.
The figure below shows how the insertion, partial / total removal and editing of test
points are done.
Edit a point
already tested.
Creates a
new point
for test.
Creates a
Sequence of
points.
Removes the
selected
point.
Removes all
tested points
When the user clicks "New Point" settings d the type of test, apparent power, angle,
capacitive or inductive load, power factor, active power, reactive power, current, voltage and
angle system are requested.
Apparent Power
Power. factor.
Reactive
Load Power.
type.
Voltage
value.
Active
Power. System angle.
Current value. Confirm the point for the test. Cancel the point for the Test.
Figure 846 – DirecPot – Main screen – Shooting - General Options – New Point
The following figure shows the graph with the points already inserted. The moment
they are Test set they turn green if they pass or red if they fail.
Next illustration is shown and the meaning of subtitle display mode of the graph.
Colors representing
status of points:
Test points and Orange: Not Tested;
untested icons. Green: Approved;
Red: Failed.
The previous figure shows the result of a search test. To facilitate the understanding,
each tab of this test will be shown in detail in the next figures. For each row the user can
choose what type of data will be shown in the table according to the figure below.
When Shows Shows the Shows the Shows the real Shows the
enabled as whether the apparent reference time when the voltage and
Line it test is three- power element for relay is current values
shows the phase or values and the test. activated. found at the
initial and single- their angle edge of the
final power phase. and the operating
values. active and region.
reactive
power.
For example leaving only the s options "General Info" and "Power" of the previous
figure selected the following data are shown:
Test point Test Type: Apparent power Active and Final test
number. three-phase or value and its found Reactive power result.
single-phase. angle. value found.
Figure 852 – DirecPot – Main screen – Search Test – Test set Points
The figure below shows how the insertion, partial / total removal and editing of test
points are done.
Figure 853 – DirecPot – Main screen – Search Test – Insert / Edit Lines
When the user clicks on “New Line”, the settings of the test type, apparent power,
angle, active power, reactive power, current and voltage of the start and end points are
requested.
Final values of apparent power and angle, active Confirm the line Cancel the line
and reactive power, current and voltage. for the test. for the test.
When clicking on "Sequence" the user defines the line variable and creates a
sequence of lines according to the chosen definition.
Figure 855 – DirecPot – Main screen – Search Test – General Options - Sequence
The following figure shows the graph of the lines before they are Test set. The
moment they are Test set they become green points if they pass or red points if they fail.
The following figure details the meaning of the legend and how to view the graph.
35.16 Route
This test aims to allow the user to test whether the operation occurs when both the
voltage and the current vary.
The previous figure shows the result of a route test. To facilitate the understanding,
each tab of this test will be shown in detail in the next figures. For each row the user can
choose what type of data will be shown in the table according to the figure below.
Type: Course Shows the Start and Step Size Reference If Total
or Group. type and end and Time of element. operation time of
When in the mode of points of Duration of has the
Group omits the test. the Power occurred. power
the last 3 powers. Variation. variation.
options.
The figure below details the items when the options “Step” and “Duration” are
selected.
Line number of Variable used: Step value Step time Total time of Final test
the route tested. apparent power, angle, used. used. the power result.
active or reactive variation.
power.
The figure below shows how the insertion, partial / total removal and the editing of
test path lines are carried out.
When the user clicks on “New Line” the settings of the fields shown in the following
figure are available.
Start and
end points
of the Step size
power of and time.
the
created
route.
Total
travel
time.
In the next figure, the light region represents a non-operating region and the dark
region the operating region since the operating time has been exceeded.
The following figure details the legend along with the graph display.
Values of apparent
power, angle, active
power and Reactive Graph display mode.
Power.
A greater detail of the previous figure is shown below. The figure below shows how
to insert pre and post fault conditions, in addition to adjusting the voltage and current
channels.
It associates the Module and angle values of Module and angle values
voltage and current each channel for pre- of each channel for post-
generation channels. simulation condition - 1 and simulation condition.
2.
Figure 866 – DirecPot – Main screen – Test Settings – Direction of Generation Channel – Pre -
Simulation and Post -Simulation
The figure below shows how to insert the pre and post fault timings and also how to
change the status of the binary and GOOSE outputs (if configured) .
Defines the states of the Defines the states of the Defines the states of the Defines the states of the
Binary Outputs and Binary Outputs and Binary Outputs and Binary Outputs and
GOOSE Outputs (when GOOSE Outputs (when GOOSE Outputs (when GOOSE Outputs (when
configured) in the configured) in the Pre- configured) in the Pre- configured) in the Post-
Simulation cycle. Simulation 1 cycle. Simulation cycle 2. Simulation cycle.
Figure 867 – DirecPot – Main screen – Test Settings – Pre-Simulation, Post-Simulation Time and Binary
and GOOSE Outputs
The following shows where to set the voltage or constant current test mode, as well
as the polarity of the potential transformers and current transformers.
In the figure below, specific settings of the search test and the general reset time are
detailed.
Search Test for: Absolute and relative current Waiting time when the point
minimum resolution or resolution. is inserted in a non-
tolerance. operating region.
Additional waiting
time when the point
is inserted in an
operating region.
In the figure below, the direction of the interfaces and the software stop/
start interface are detailed.
DISTANC
SOFTWARE FOR TESTINGS DISTANCE
RELAY
The options on the Settings screen are described in detail in chapter 21.
The Distance, Import and Export fields will be discussed throughout this chapter.
Creates a new
test file.
Open a saved
file.
Password protect
certain
parameters of the
test file.
Saves the current
file.
In the start menu, several resources are available. The following figure describes
hardware configuration features and generation controls such as start, stop, move to the next
point / line, deletion of one or all of the Test set points.
Opens a Channel Opens a screen Starts the Pauses signal Move on to the
Orientation screen. for Hardware signal generation. next point.
Configuration. generation
process.
Clean the
spot
Selected
and returns
to Untested
status.
Clears all
spots
and
returns to
the status
Connects between Opens a screen Go to the Go to the next of Not
software and for GPS Setup. next line. Type of fault. Tested.
hardware.
In the figure below, the resources for analysis options of the generated quantities
(waveforms, phasors, impedance trajectory and Zxt graph) are displayed, as well as the
adjustment options (General and Differential Settings). Access to the report is also
displayed, options for viewing the units involved (time and module) and general options for
restoring and viewing the layout.
Calls the Shows the Shows the Shows the Shows the Shows the Returns to
Settings Waveform Phasor Zxt Tab. time in time in the standard
Screen. Tab. Tab. seconds. cycles. layout.
Enables /
disables
tabs for
viewing.
Shows
values
based on the
secondary.
Opens the
report
Shows the Special configurati
settings for Shows values Shows
Trajectory on
SEL relays. based on the absolute Shows relative
tab. window. values.
primary. values.
Figure 877 – Distanc – Home Menu – Options, Report, Units and Layout
Position the selected Superimposes the Hide the windows Returns to Enables /
window: on the left; selected window. the standard
for better viewing. disables tabs
to the right, above, layout.
for viewing.
below or as a
document.
Hide the selected Hide the selected Close all open Closes all open
tabs. window. windows. documents.
Insert password in Enables / disables the Opens the Settings Shows the
manufacturer’s address,
the Test file so as simulation mode. Since Screen when the
phone, email and fax.
not to allow the in this mode there is no software is started.
editing of certain insertion of voltage and
settings. current.
It is also opened at software startup, if the Opens Settings when Starting option in the
Software Options tab is selected.
In mode General 's tab General Information , where the user enters all of for the test,
the dispositive state and responsible by t this, there is the flap systems , where the user enters
the nominal values of frequency, voltage, current and power, among others , there is a tab
for Notes and Observations for various records.
There is an Explanatory Figures tab where the user can insert up to three figures
relevant to the test, a Check List tab to create steps to be followed during the test, there is a
tab called Others for configuring digital signal filters and finally a Connections tab allowing
the user to structure all the connections between the test bag and the equipment to be Test
set.
In Distance mode, there is the Distance Protection Adjustment tab. In it the user sets
the line length and the angle. Choose how the earth factor will be inserted. You can still
enable whether to remove the Arc Resistance and use Impedance Correction for Rated
Current.
To facilitate the adjustment of the zones, there is an Import option whose purpose is
to read files .RIO. Some distance relays record line, zone and operating time settings in this
format. Importing this file automatically reads the Distanc of so that the user does not need
to manually adjust it. Another option would be to export a file .RIO once the feature was
entered by the user.
The Insert option opens a new window for the user to enter the name and number of
the zone, adjust the Trigger Time, Absolute and Relative Tolerances of the impedance and
the Absolute and Relative Tolerances for the time. You can use a specific earth factor and
enable the option to Remove Arc Resistance.
If the relay does not generate a file .RIO there is the option to design the zone to be
Test set or to choose between predefined zones. Depending on the type of zone chosen,
options for load compensation and directionality are opened.
CONPROVE INDÚSTRIA E COMÉRCIO LTDA 648
CTC SOFTWARE MANUAL
Import
Files
.RIO
Export
Files
.RIO
Choose the
mode of the
earth factor:
(Mod;Ang)
(R;X)
(R0;X0)
(K0)
(Rt/Rl;Xt/Xl)
(Z0/Z1).
Shows which
zones are
available.
Insert a Edit a zone Deletes a zone that has Shows the design of
already inserted. already been inserted.
new zone. the chosen zone.
Let us see the window that opens when clicking on Insert, where it is made the
adjustments of the protection zones.
Figure 883 – Distanc – Parameters Screen – Distance – Distance Protection Adjustment – Insert – Zone
To facilitate the visualization of the previous figure, it was divided into parts. In this
first part the user can enter the description of the added zone, choose the type of actuation,
select which loop (detection algorithm) that zone belongs to and enter the absolute and
relative tolerances of time and impedance.
Zone description.
Zone type registration:
Zone index. Departure, Performance,
Extended or Block.
Links the zone to a (Informative data).
specific type of
fault. Absolute (Ω) and
relative (%) tolerance
of Impedance.
Trip time of the
zone.
Absolute tolerance
(more or less) in
seconds and relative
tolerance in% of the
time of operation.
Points inserted
when using the
option to draw.
Figure 884 – Distanc – Parameters Screen – Distance – Distance Protection Adjustment – Insert – Zone
In this second part the user has access to the various ways of drawing the desired
zone. The first form is drawing segment by segment, the second form uses known zones
(Mho, Quadrilateral and tomato / lens) where the user by comparison changes some intuitive
segments and the third form is using the presets of the relays themselves (Masks).
Finalizes the
drawing
process.
Figure 885 – Distanc – Parameters Screen – Distance – Distance Protection Adjustment – Insert – Zone
The following figure shows a quadrilateral area which was drawn using line
segments.
Figure 886 – Distanc – Parameters Screen – Distance – Distance Protection Adjustment – Insert –
Drawn Area
In the next figure, the quadrilateral option was chosen directly. Note that the settings
to be entered are more intuitive with this option.
Figure 887 – Distanc – Parameters Screen – Distance – Distance Protection Adjustment – Insert –
Quadrilateral Zone
After inserting all the desired zones, these are shown in the Graphics tab.
Enables / disables
Shows the grid to facilitate values to be shown in
the visualization of values. Y.
Extract the
COMTRADE or CSV
Allows you to Insert or file from the waveforms.
Edit either Markings or Take a photo of the chart
Comments in graphics. by storing it on the
clipboard.
Clicking on the Display option in the previous figure brings you to the following
options:
This screen has the Transparency property, and can remain open while points are
inserted or edited in the test. The following screen is for a line Test set using Point Test.
Clicking on the Show Columns option in the previous figure brings up the following
options:
The solid lines of the zones in the graph represent the limit of each zone given by its
nominal value. The dotted lines represent the transition region from one zone to
another considering the tolerance given by the manufacturer.
For this test the Sequence button is disabled, it is only possible to insert points,
inserting them one by one all the points to be Test set.
This test allows the user to test points in the Operating and Non-Operating Regions
of the relay characteristic.
After inserting the test points, just click on generate, and the software automatically
tests all points.
For each point the user can choose what type of data will be shown in the
table according to the figure below.
Figure 902 – Distanc – Main screen – Point Test –Test set Points Options
In the figure below it is detailed all the items that appear in the Test set points table.
Figure 903 – Distanc – Main screen – Point Test– Test set Points
The figure below shows how the insertion, partial / total removal and editing of test
points are done.
Edit selected
point.
Insert a new test
point.
Insert a sequence of
points.
Removes selected
point. Removes all
stitches.
Figure 904 – Distanc – Main screen – Point Test – Insert / Edit Points
The way in which data is entered for a test point in the point test is detailed below.
This test aims to find the border between one zone and another. Test points are
entered in two ways:
a) Click on the button New Line; then choose the type of fault, click on the
region of the graph choosing a starting point and ending point, a line is
automatically created between the points.
b) b) Click on the button New Line; then enter the values of the module or
angle in the corresponding fields. This is if Data Entry is selected Z and
θ. Otherwise, the resistance and impedance values are entered. The next
step is to insert the module and Angle of the length.
The figure below shows some points already found and some lines of research that
have not yet been Test set.
In the figure below, all the items that appear in the test set thread table are detailed.
Nominal
impedance Impedance
Zone
module and module and angle
Number of the boundary.
angle. used in the test.
tested point.
Figure 909 – Distanc – Main screen – Search Test – Test set Points
The figure below shows how the insertion, partial / total removal and editing of test
points are done.
Figure 910 – Distanc – Main screen – Search Test – Insert / Edit Points
The way in which data is entered for a test point in the point test is detailed below.
The top screen will be divided into several parts for easy viewing. The designation of
each channel and the insertion of pre / post faults are detailed in the figure below.
Figure 914 – Distanc – Main screen – Test Settings – Direction Generation Channel
In the Figure below, the fields dealing with relations between earth and phase of TPs
and CTs, inversion of TPs and phase and ground CTs and the choice of the start of the
timing are detailed.
Figure 915 – Distanc – Main screen – Test Settings – Relationship between CTs and TPs - Polarities of
CTs and TPs
In the figure below, the fields dealing with the relationship between earth and phase
of TPs and CTs, inversion of TPs and phase and ground CTs and the choice of the start of
the timing are detailed.
Defines the states of Defines the states of Defines the states of Defines the states of
the Binary Outputs the Binary Outputs and the Binary Outputs and the Binary Outputs and
and GOOSE Outputs GOOSE Outputs GOOSE Outputs GOOSE Outputs
(when configured) in (when configured) in (when configured) in (when configured) in
the Fault cycle. the Pre-fault1 cycle. the Pre-fault2 cycle. the Post-Fault cycle.
Figure 916 – Distanc – Main screen – Test Settings – Pre-Fault, Post-Fault Time and Binary
and GOOSE Outputs
The test mode and the characteristics of each mode are detailed in the figure below.
Maximu
m
multiple
of the
voltage
used in
the fault.
Figure 917 – Distanc – Main screen – Test Settings – Pickup Generation Limits - Test Mode
In the figure below, all the specific configurations of the search test and other tests
are detailed.
Below is shown the way to change the software interface stopping, starting interface,
access to the use of logic in these two resources (stopping and starting) and the manner of
early faults.
Configures whether the Shows the logic for Selects the fault start mode:
generation trigger triggering the generation. Random, Maximum off Set,
through: Software, Logic, Which outputs are Minimum Off Set and Fixed
Binary Output, GPS or involved and the type of Angle.
external signal. logic: AND // OR.
Figure 919 – Distanc – Main screen – Test Settings – Interface –Fault Start
The figure below details all possible ways of generating voltage and current that the
equipment provides through the software.
PSB_OoS
SOFTWARE FOR TESTING DISTANCE
RELAYS
The options on the Settings screen are described in detail in chapter 21. The
Distance, Import and Export fields will be discussed throughout this chapter.
Creates a new Opens a saved Saves the Customize the Software name and
test file. test. current test. Quick Access version.
Bar.
Opens the
options: Equipment used
Restore,
and serial
Move, Size,
Minimize, number.
Maximize
and Close.
Creates a new
test file.
Open a saved
file.
Password protect
certain
parameters of the
test file.
Saves the current
file.
In the start menu, several resources are available . The following figure describes
hardware configuration features and generation controls such as start, stop, move to the next
point / line, deletion of one or all of the Test set points.
Opens a Channel Opens a screen Starts the signal Pauses signal Move on to the
Orientation screen. for Hardware generation generation. next point.
Configuration. process.
Clean the
spot
Selected
and returns
to Untested
status.
Clears all
spots
and
returns to
the status
Connects between Opens a screen for Go to the of Not
software and GPS Setup. next group. Tested.
hardware.
In the figure below, the resources for analysis options of the generated quantities
(waveforms, phasors and impedance trajectories) are displayed, as well as the adjustment
options ( General and Differential Settings ). Access to the report is also displayed, options
for viewing the units involved (time and module) and general options for restoring and
viewing the layout.
Calls the Shows the Shows the Shows the Shows the Returns to
Settings Waveform Phasor time in time in the standard
Screen. Tab. Tab. seconds. cycles. layout.
Enables /
disables
tabs for
viewing.
Shows
values
based on the
secondary.
Opens the
report
Shows the configurati Shows values Shows
Trajectory on based on the absolute Shows relative
tab. window. primary. values.
values.
Figure 927 – PSB_OoS – Home Menu – Options, Report, Units and Layout
Hide the selected Hide the selected Close all open Closes all open
tabs. window. windows. documents.
Insert password in Enables / disables the Opens the Settings Shows the
manufacturer’s address,
the Test file so as simulation mode. Since Screen when the
phone, email and fax.
not to allow the in this mode there is no software is started.
editing of certain insertion of voltage and
settings. current.
It is also opened at software startup, if the Opens Settings when Starting option in the
Software Options tab is selected.
In General mode there is the General Information tab, where the user enters all the
data related to the test, the Test set device and the person responsible for the test, there is the
Systems tab, where the user informs the nominal values of frequency, voltage, current and
power among others, there is a tab for Notes and Observations for various records.
There is an Explanatory Figures tab where the user can insert up to three figures
relevant to the test, a Check List tab to create steps to be followed during the test, there is a
tab called Others for configuring digital signal filters and finally a tab Connections allowing
the user to structure all connections between the test case and the equipment to be Test set.
In Distance mode, there is the Distance Protection Adjustment tab. In it the user sets
the line length and the angle. Choose how the earth factor will be inserted. You can still
enable whether to remove the Arc Resistance and use Impedance Correction for Rated
Current.
To facilitate the adjustment of the zones, there is an Import option whose purpose is
to read .RIO files. Some distance relays record line, zone and operating time settings in this
format. Importing this file, Distanc reads automatically so that the user does not need to
adjust it manually. Another option would be to export a .RIO file once the feature has been
entered by the user.
The Insert option opens a new window for the user to enter the name and number of
the zone, adjust the Trigger Time, Absolute and Relative Tolerances of the impedance and
the Absolute and Relative Tolerances for the time. You can use a specific earth factor and
enable the option to Remove Arc Resistance.
If the relay does not generate a .RIO file, there is the option to design the zone to be
Test set or to choose between predefined zones. Depending on the type of zone chosen,
options for load compensation and directionality are opened.
Displays the
tab for
adjusting the
Distance
function.
Import
Files
.RIO
Export
Files
.RIO
Choose the
mode of the
earth factor:
(Mod;Ang)
(R;X)
(R0;X0)
(K0)
(Rt/Rl;Xt/Xl)
(Z0/Z1).
Shows which
zones are
available.
Let us see the window that opens when clicking on Insert, where it is made the
adjustments of the protection zones.
Figure 933 – PSB_OoS – Parameters Screen – Distance – Distance Protection Adjustment – Insert –
Zone
To facilitate the visualization of the previous figure, it was divided into parts. In this
first part the user can enter the description of the added zone, choose the type of actuation,
select which loop (detection algorithm) that zone belongs to and enter the absolute and
relative tolerances of time and impedance.
Zone description.
Zone type registration:
Zone index. Departure, Performance,
Extended or Block.
Links the zone to a (Informative data).
specific type of
fault. Absolute (Ω) and
relative (%) tolerance
of Impedance.
Trip time of the
zone.
Absolute tolerance
(more or less) in
seconds and relative
tolerance in% of the
time of operation.
Points inserted
when using the
option to draw.
Figure 934 – PSB_OoS – Parameters Screen – Distance – Distance Protection Adjustment – Insert –
Zone
In this second part the user has access to the various ways of drawing the desired
zone. The first form is drawing segment by segment, the second form uses known zones
(Mho, Quadrilateral and tomato / lens) where the user by comparison changes some intuitive
segments and the third form is using the presets of the relays themselves (Masks).
Finalizes the
drawing
process.
Figure 935 – PSB_OoS – Parameters Screen – Distance – Distance Protection Adjustment – Insert –
Zone
The following figure shows a quadrilateral area which was drawn using line
segments.
Figure 936 – PSB_OoS – Parameters Screen – Distance – Distance Protection Adjustment – Insert –
Drawn Area
In the next figure, the quadrilateral option was chosen directly. Note that the settings
to be entered are more intuitive with this option.
Figure 937 – PSB_OoS – Parameters Screen – Distance – Distance Protection Adjustment – Insert –
Zone Quadrilateral
Choose between:
Fault starts
Fixed, Maximum, between 0 and
Minimum or 360 °.
Random.
Fault Angle.
Reference phase
for the fault.
Opening
moment to
Enables / count the
disables Time time.
Evaluation.
Tolerances
given by the
Expected time manufacture
for r of the
relay.
performance.
Extract the
Allows you to COMTRADE or CSV
Insert or Edit either file from the waveforms.
Markings or Take a photo of the chart
Comments in by storing it on the
graphics. clipboard.
Clicking on the Display option in the previous figure brings you to the following
options:
Clicking on the Show Columns option in the previous figure brings up the following
options:
37.12.2
37.13 System Simulation Test
This test aims to allow the user to create power fluctuations and thus test situations
where there is a Power swing, the distance function or the Out-of-Step function . To start a
test, follow these steps:
Click on the “New Oscillation” button; then choose the type of oscillation:
"Synchronous" or "Asynchronous".
a. Synchronous: In this case, a slip angle value greater than -180° and less
than 180° must be defined, the duration time, the data from the two
sources, the evaluations of whether a function will act and how long,
finally the location of the relay in the system must be adjusted. NOTE: In
synchronous oscillation, it is not allowed to insert a fault.
b. Asynchronous: In this situation, the number of turns, the sliding
frequency, the data from the two sources, the evaluations of whether a
function is going to act and how long, must finally be set, the location of
the relay in the system must finally be adjusted. A fault can be inserted
together with the oscillation.
The previous figure shows the result of a System Simulation test. To facilitate the
understanding, each adjustment of this test will be shown in detail in the following figures.
For each point the user can choose what type of data will be shown in the table
according to the figure below.
Figure 952 – PSB_OoS – Main screen – System Simulation – Options Oscillations Test set
In the figure below it is detailed all the items that appear in the table of simulations in
the test set system.
Figure 953 – PSB_OoS – Main screen – System Simulation– Result Test set Oscillations
The figure below shows how the insertion, partial / total removal and editing of test
points are done.
Edit an
already tested
oscillation.
Creates a
new
oscillation.
Creates a
sequence of
oscillations. Removes the
selected
Removes all oscillation.
tested
oscillations.
Figure 954 – PSB_OoS – Main screen – System Simulation – Insert / Edit Oscillations
The following screens show the settings when the user chooses the option "New
Oscillation" and "Type of Oscillation" Synchronous.
If enabled, it
defines the
module and
disb angle of the
electrical center.
Difference
between
the angles If enabled, it
of the two limits the
Current value.
sources.
Oscillation
time.
Confirm Test Cancel test
point. point.
Figure 955 – PSB_OoS – Main screen – System Simulation - General Options - Synchronous
Allows you to choose the Defines the type of If enabled, it allows If enabled, it
Oscillation duration. Whether oscillation. Being inserting fault in the defines the module
by number of spins or by time. Synchronous or asynchronous and angle of the
oscillation. electrical center.
Asynchronous.
The number
of times the
angle varies
from 0 ° to
360 °.
If enabled, it
Time in limits the
seconds of current value
the in the test.
oscillation.
Figure 956 – PSB_OoS – Main screen – System Simulation - General Options - Asynchronous
CONPROVE INDÚSTRIA E COMÉRCIO LTDA 699
CTC SOFTWARE MANUAL
There is the option "Sequence", where the user can vary a parameter with a certain
step, from an initial value to a final value and automatically creates several tests. Can be
varied slip angle and / or the duration of oscillation and the number of cycles (only
asynchronous oscillation). Another possibility is the variation of the parameters of sources 1
and 2. It is possible to change the values of modulus, voltage angle, resistance s and
impedance s, either phase or ground.
The following figure shows a test for synchronous simulation where the sliding angle
is varied from 0° to 180° using a 10 step. In this test, the performance of the PSB alarm is
evaluated, which, according to the setting, acts from 50° angles. There is the option of
inserting a second Sequence in addition to the possibility of evaluating the performance time
of each test allowing a more complete test.
The previous figure shows the result of a Simulation of Trajectories test. To facilitate
the understanding, each option of this test will be shown in detail in the next figures.
It has two types: When enabled, shows When enabled, it When enabled, it shows
Individual or the Type and Time of shows in the Test “Time Reference”,
Group. When in a the Oscillation result the “Interface”, “Nominal Time” and
group it shows the duration in the Test the “Nominal “Real Time” in the Test
result. result.
time and status of Performance” and if it
the Trajectory. really acted or not.
Figure 962 – PSB_OoS – Main screen – Trajectory Simulation – Test set Trajectory Options
In the figure below, all the items that appear in the test set trajectories table are
detailed.
Figure 963 – PSB_OoS – Main screen – Trajectory Simulation – Result of Trajectories Test set
The figure below shows how the insertion, partial / total removal and editing of test
points are done.
Edit an already
tested
oscillation.
Creates a new
Trajectory.
Creates a group
of Trajectories.
Remove
selected
Removes all Trajectory.
tested
Trajectories.
Figure 964 – PSB_OoS – Main screen – System Simulation – Insert / Edit Trajectories
The way in which data is entered for a test point in the trajectory test is detailed
below.
The seven values (in this case) to Impedance module and Variation of Time between
angle values or
create the Impedance Trajectory Impedance one
resistance and
reproducing oscillations. reactance. over time. Impedance
and another.
Data entry | Z |
and θ, or R and
X.
Number of
points inserted.
Variation of
Impedance over
time. Deve ser
diferente de
zero.
Figure 965 – PSB_OoS – Main screen – System Simulation – General Options – Trajectory
Below is detailed the way in which data is entered for a test point in the trajectory
test, informing data from the simulated system.
Ground impedance.
(In this case Ze = Zs
x Ks).
Phase impedance.
Figure 966 – PSB_OoS – Main screen – System Simulation – General Options – System
Below is detailed the way in which data is entered for a test point in the trajectory
test, informing data of a fault in the system.
Figure 967 – PSB_OoS – Main screen – System Simulation – General Options – Fault
The way in which the evaluation of the trajectory test is performed is detailed below.
Enables the
Evaluation of
the
performance
time.
Nominal
adjustment
time plus the
tolerances Enables the moment
given by the
to start counting the
manufacturer.
time.
Figure 968 – PSB_OoS – Main screen – System Simulation – General Options – Evaluation
“Vdelta” = displacement Module and angle values for Module and angle values of
tension. “IE” = Neutral each channel for Pre-fault each channel for post-fault
current. condition 1 and 2. condition.
Figure 970 – PSB_OoS – Main screen – Test Settings – Direction of Generation Channel – Pre-
Simulation and Post-Simulation
In the figure below, the fields dealing with the relationship between earth and phase
of TPs and CT’s, inversion of TPs and phase and ground CT’s and the choice of the start of
the timing are detailed.
Enables reversed
connected phase and / or
earth CT polarities.
Figure 971 – PSB_OoS – Main screen – Test Settings – Ratio between CTs and TPs - Polarities of CTs
and TPs
In the figure below, the initial status of the binary outputs and GOOSE outputs, pre /
post fault times and the option to maintain pre-faults in phases not affected by the fault are
detailed.
Defines the states of Defines the states of the Defines the states of the Defines the states of the
the Binary Outputs and Binary Outputs and Binary Outputs and Binary Outputs and
GOOSE Outputs GOOSE Outputs (when GOOSE Outputs (when GOOSE Outputs (when
configured) in the Fault configured) in the Pre- configured) in the Post-
(when configured) in
cycle. fault 2 cycle. Fault cycle.
the Fault cycle.
Figure 972 – PSB_OoS – Main screen – Test Settings – Pre- Simulation, Post- Simulation Time and
Binary Outputs and GOOSE
SYNCHRONISM
SOFTWARE FOR TESTING
SYNCHRONISM RELAY
The options on the Settings screen are described in detail in chapter 21. The
Synchronism field will be discussed throughout this chapter.
Creates a new Opens a saved Saves the Customize the Software name and
test. current test. Quick Access version.
test file.
Bar.
Equipment used
and serial
Opens the options:
number.
Restore, Move, Size,
Minimize, Maximize
and Close.
Creates a new
test file.
Open a saved
file.
Password protect
certain
parameters of the
test file.
Saves the current
file.
In the start menu, several resources are available . The following figure describes
hardware configuration features and generation controls such as start, stop, move to the next
point / line, deletion of one or all of the Test set points.
Opens a Channel Opens a screen Starts the Pauses signal Move on to the
Orientation screen. for Hardware signal generation. next point.
Configuration. generation
process.
Clean the
spot
Selected
and returns
to Untested
status.
Clears all
spots
and
returns to
the status
Connects between Opens a screen for Go to the of Not
software and GPS Setup. next line. Tested.
hardware.
In the figure below, the resources for analysis options of the generated quantities
(waveforms, phasors and synchronoscope) are displayed, as well as the adjustment options
(General Settings and Synchronism). Access to the report, options for viewing the units
involved (time and module) and general options for restoring and viewing the layout are also
displayed.
Calls the Shows the Shows the Shows the Shows the Shows the Returns to
Phasor the standard
Settings Waveform time in time in values related
Tab. layout.
Screen. Tab. seconds. cycles. to
synchronism.
Enables /
disables
tabs for
viewing.
Shows
values
based on the
secondary.
Opens the
report
configurati Shows values Shows
Shows the on based on the absolute Shows relative
Synchronism tab. window. primary. values. values.
Figure 980 – Synchronism – Home Menu – Options, Report, Units and Layout
Position the selected Superimposes the Hide the windows Returns to Enables /
window: on the left; selected window. the standard
for better viewing. disables tabs
to the right, above, layout.
for viewing.
below or as a
document.
Hide the selected Hide the selected Close all open Closes all open
tabs. window. windows. documents.
Insert password in Enables / disables the Opens the Settings Shows the
manufacturer’s address,
the Test file so as simulation mode. Since Screen when the
phone, email and fax.
not to allow the in this mode there is no software is started.
editing of certain insertion of voltage and
settings. current.
It is also opened at software startup, if the Opens Settings when Starting option in the
Software Options tab is selected.
In General mode, there is the “General Information” tab, where the user enters all the
data relating to: the Test, the Test set Device and the person responsible for the Test, there is
the “Reference Values” tab, where the user informs the nominal values of voltage, current
and power for system 1, there is also a tab for “Notes and Observations”. In the other tab
“Explanatory Figures” it is possible to insert up to 3 figures in the most used image
extensions, such as: Bitmap; JPGE; PNG; JPG; etc.
In Synchronism mode, there is the “Systems” tab. In it, the user configures whether
the voltage of system 1 used as a reference is single-phase, or three-phase. In the latter case,
it is possible to choose between phase-to-phase and phase-to-neutral voltages. The same
adjustments are made for system 2, which also has fields for adjusting the primary voltage,
primary and secondary current in addition to the phase sequence. There is an option that
copies the settings from system 1 to system 2.
For the circuit breaker there are time adjustments for effective closing to occur, and
for reporting whether the signal emitted is a pulse or continuous. If the option “none” is set,
the circuit breaker closing time is disregarded.
Adjustment
of phase-to-
phase and
neutral phase
voltages for
System 1.
Then the screen where system data 2 and the settings for synchronism are inserted.
To facilitate understanding, the previous figure will be divided into two parts. Note
in the figure below that system one was adopted as the one to be synchronized. It is also
shown that it is possible to choose whether it is single-phase or three-phase, including the
reference phase, the angular offset of the elevation transformer (if the synchronism is done
in a Generator + Transformer system). Finally, the user can choose the way to start the
synchronism and parameterize the closing time of the circuit breaker (which will be taken
into account in the calculation of the relay).
Then system 2 is adopted as the side of the system that will receive the generation or
simply the infinite bar. In the figure below you can choose a three-phase or single-phase
system, the angular reference and the voltage and current levels in primary and secondary
values. Finally, you can change the phase sequence.
angle given by the manufacturer for reporting purposes, that is, validation of the points Test
set.
Relative and
absolute frequency
tolerances given by
the relay
manufacturer.
Enables / disables
Zooms in or out for the display of the
easy viewing of the line in the Search
test. Test.
The origin is
centered within the
graph tab.
Shows the frequency
and voltage values on
the x and y axes. Shows a grid for
easy viewing.
Central axes are in
bold for greater
Hides tolerances from
emphasis. the synchronism graph.
Extract the
Allows you to Insert or COMTRADE or CSV
Edit either Markings or file from the waveforms.
Comments in graphics. Take a photo of the chart
by storing it on the
clipboard.
Clicking on the Display option in the previous figure brings you to the following
options:
This screen has the Transparency property, and can remain open while points are
inserted or edited in the test. The following screen is for a line Test set using Trigger.
Clicking on the Show Columns option in the previous figure brings up the following
options:
38.12.1 Trigger
This test aims to allow the user to test whether a system can be synchronized to
another, according to the values of voltage, frequency and angle, respecting the adjusted
differences.
The previous figure shows the result of a Trigger test. For a better explanation of this
test, this figure is divided into parts.
CONPROVE INDÚSTRIA E COMÉRCIO LTDA 727
CTC SOFTWARE MANUAL
For each Test set point the user can choose what type of data will be shown in the
table according to the figure below.
As an example, leaving only the “Resp” option in the previous figure, the following
ones are shown.
The figure below shows how the insertion, partial / total removal and editing of test
points are done.
Edit a point
already tested.
Creates a
new point
for test.
Creates a
Sequence of
points. Removes the
selected
point.
Removes all
tested points.
Figure 1007 – Synchronism – Main screen – System Simulation – Insert / Edit Points
When the user clicks on “New Point” the voltage, frequency and angle adjustments
must be entered. The user first adjusts the values of system 1 and can choose between two
options: enter the value of the difference between system 1 and 2, or enter the absolute
values of voltage, frequency and angle of system 2.
Angle difference
Voltage difference Frequency between the two
difference between
Cancel the
between the two Systems.
the two Systems. point.
Systems.
Confirm
the point.
Figure 1008 – Synchronism – Main screen – Shooting - General Options – New Point
Angle
pitch.
Total points
calculated
automatically.
The following figure shows the plot of the points before they are Test set. The
moment they are Test set they turn green if they pass or red if they fail.
The next illustration shows If the area where the timing is possible. The green arrow
shows the moment of the circuit breaker closing command. If the arrow is red it means that
with the current values of the two systems, synchronism is not allowed.
The previous figure shows the result of a test of search. To facilitate the
understanding, each tab of this test will be shown in detail in the next figures. For each row
the user can choose what type of data will be shown in the table according to the figure
below.
For example leaving Only Smarter and the "Reply" option of the previous figure are
shown the following data:
Figure 1014 – Synchronism – Main screen – Search Test – Test set Points
The figure below shows how the insertion, partial / total removal and editing of test
points are done.
Figure 1015 – Synchronism – Main screen – Search Test – Insert / Edit Lines
When the user clicks on “New Line” the settings of the fields shown below are
available.
To facilitate the visualization and explanation of each field, the previous figure was
divided in two. In the figure below, the insertion of the starting point of a new test line is
detailed.
Voltage, frequency and angle values Difference in voltage and Absolute voltage and
frequency between the two
for System 1. They are configured in frequency values for the
Systems for the starting
“Settings”> “General”> “Reference point of the search. System 2 start point.
Values”.
Figure 1017 – Synchronism – Main screen –Search Test – General Options – First Half
In the figure below, the end point of the added line is shown, as well as the enabling
of the synchronism check feature after searching.
Figure 1018 – Synchronism – Main screen –Search Test – General Options – Second Half
By clicking “Sequence” the user sets a value of voltage and frequency and just pick
one step to the angle different s lines are inserted automatically.
Step
angle.
Figure 1019 – Synchronism – Main screen – Search Test – General Options - Sequence
The following figure shows the graph of the lines before they are Test set. The
moment they are Test set they become green dots if they pass, or red dots if they fail.
The following figure details the legend along with the synchronoscope.
38.14 Trajectory
This test aims to allow the user to test whether a system reaches synchrony even if
there is a variation in voltage over time and / or variation in frequency over time. Voltage,
frequency are defined for system 1. For system 2, a voltage, frequency and an initial and
CONPROVE INDÚSTRIA E COMÉRCIO LTDA 736
CTC SOFTWARE MANUAL
final angle are defined . Finally, the values of the variation of voltage and frequency over
time are adjusted.
The previous figure shows the result of a route test. To facilitate the understanding,
each tab of this test will be shown in detail in the next figures. For each row the user can
choose what type of data will be shown in the table according to the figure below.
In the figure below, all the items that appear in the Test set route line table are
detailed.
The figure below shows how the insertion, partial / total removal and the editing of
test path lines are carried out.
When the user clicks on “New Line” the settings of the fields shown in the following
figure are available.
To facilitate the visualization and explanation of each field, the previous figure was
divided in two. In the figure below, the insertion of the conditions of the initial system of the
path test is detailed.
Voltage, frequency and angle Difference in voltage, frequency Voltage, angle and
and angle between the two frequency values for
values for System 1. They are
Systems for the starting point of the System 2 start
configured in “Settings”> the search.
“General”> “Reference Values”. point.
Figure 1027 – Synchronism – Main screen –Trajectory – General Options – First Half
In the figure below, the insertion of the conditions of the final system of the route test
is detailed.
Figure 1028 – Synchronism – Main screen –Trajectory – General Options – Second Half
In the next figure, the region within the square of the full line is the region of
synchronism and the external region is of non-synchronism . The region defined between the
two squares of dotted lines represents the tolerances given by the manufacturer
of each relay.
The following figure details the legend along with the synchronoscope.
Colors representing
Tested and line status:
untested points Blue: Not Tested;
icons and search Green: Approved;
line. Red: Failed.
Synchronism region
highlighted in blue according
to tension and angle criteria.
In order for the synchronism
to occur effectively, the
frequency criterion must also
be obeyed.
A greater detail of the previous figure is shown below. The figure below shows how
to insert pre-and post-fault conditions, in addition to, of course, the designations of each
voltage channel.
Enables pre-simulation
1 and 2 and post-
simulation conditions.
Designates the Module and angle values of each Module and angle values of
generation channels channel for pre-simulation each channel for post-
for each System. condition - 1 and 2. simulation condition.
Figure 1032 – Synchronism – Main screen – Test Settings – Direction of Generation Channel– Pre-
Simulation and Post-Simulation
The figure below shows how to insert the pre and post fault timings and also how to
change the state of the binary and GOOSE outputs .
Defines the states of Defines the states of the Defines the states of the Defines the states of the
the Binary Outputs and Binary Outputs and Binary Outputs and Binary Outputs and
GOOSE Outputs GOOSE Outputs (when GOOSE Outputs (when GOOSE Outputs (when
configured) in the Pre- configured) in the Pre- configured) in the Post-
(when configured) in
Simulation 1 cycle. Simulation cycle 2. Simulation cycle.
the Simulation cycle.
Figure 1033 – Synchronism – Main screen – Test Settings – Pre-Simulation, Post-Simulation Time and
Binary Outputs and GOOSE
In the figure below, specific settings for the search test and the general reset
time are detailed.
Time
between
increments.
Time to
restart.
In the figure below, the direction of the V>, V <, f> and f <interfaces (Important for
the search test) and the software stop / start interface is detailed.
Stop interface
for reduced
voltage.
Stopwatch stop
interface. By a
single binary or by
binary logic.
Stopwatch trigger interface by: Insert logic between binary Trigger time
software, GPS, or by binary logic. inputs or GOOSE messages. delay.
Allows “AND” or “OR” logic.
VOLTS P
HERTZ
SOFTWARE FOR TESTING
OVEREXCITATION RELAYS
The options on the Settings screen are described in detail in chapter 21. The Volts /
Hertz field will be discussed throughout this chapter.
Creates a new Opens a saved Saves the Customize the Software name and
test. current test. Quick Access version.
test file.
Bar.
Equipment used
Opens the options: and serial
Restore, Move, number.
Size, Minimize,
Maximize and
Close.
Figure 1038 – VoltsPHz – Quick Access Bar
Creates a new
test file.
Open a saved
file.
Password protect
certain
parameters of the
test file.
Saves the current
file.
In the start menu, several resources are available. The following figure describes
hardware configuration features and generation controls such as start, stop, move to the next
point / line, deletion of one or all of the Test set points.
Opens a Channel Opens a screen Starts the signal Pauses signal Move on to the
Orientation for Hardware generation generation. next point.
screen. Configuration. process.
In the figure below, the resources for analysis options of the generated quantities
(waveforms, phasors) are displayed, as well as the adjustment options (General Settings and
Volts / Hertz ). Access to the report is also displayed, options for viewing the units involved
(time and module) and general options for restoring and viewing the layout.
Enables /
disables
tabs for
viewing.
Shows
values
based on the
secondary.
Opens the
report
Calls the configurati Shows values Shows
Settings on based on the absolute Shows relative
Screen. window. primary. values. values.
Figure 1042 – VoltsPHz – Home Menu – Options, Report, Units and Layout
Position the selected Superimposes the Hide the windows Returns to Enables /
window: on the left; selected window. the standard
for better viewing. disables tabs
to the right, above, layout.
for viewing.
below or as a
document.
Hide the selected Hide the selected Close all open Closes all open
tabs. window. windows. documents.
Insert password in Enables / disables the Opens the Settings Shows the
manufacturer’s address,
the Test file so as simulation mode. Since Screen when the
phone, email and fax.
not to allow the in this mode there is no software is started.
editing of certain insertion of voltage and
settings. current.
It is also opened at software startup, if the Opens Settings when Starting option in the
Software Options tab is selected.
In General mode, there is the “General Information” tab, where the user enters all the
data relating to: the Test, the Test set Device and the person responsible for the Test, there is
the “Reference Values” tab, where the user informs the nominal values of voltage, current
and power for system 1, there is also a tab for “Notes and Observations”. In the other tab
“Explanatory Figures” it is possible to insert up to 3 figures in the most used image
extensions, such as: Bitmap; JPGE; PNG; JPG; etc.
In General mode, there is the “General Information” tab, where the user enters all the
data relating to: the Test, the Test set Device and the person responsible for the Test, there is
the “Reference Values” tab, where the user informs the nominal values of voltage, current
and power for system 1, there is also a tab for “Notes and Observations”. In the other tab
“Explanatory Figures” it is possible to insert up to 3 figures in the most used image
extensions, such as: Bitmap; JPGE; PNG; JPG; etc.
In Volts / Hertz mode, the user sets curve 1 and can insert points, adjust the pick-up
value and time dial. In curve 2, the user adjusts the pick-up value and the actuation time if
using a defined time curve.
In this screen, the voltage and frequency limits are set as well as the relative and
absolute tolerances for both voltage, frequency and time.
There are also two tabs for viewing the timed curve (V/f x t) and another for viewing
the pickup (V/f / Vn/Fn x t)
To facilitate understanding, the previous figure will be divided into parts. In the
figure below, parameterizes if the type, pick up and dial curve time.
Curve 2 pickup.
Curve 1
pickup.
In the figures below, you can enter the voltage and frequency limits of a given relay,
all in conjunction with the tolerance of time, voltage and frequency.
Minimum
Maximum
frequency.
frequency.
Absolute
Relative frequency
frequency tolerance.
tolerance.
The following figures show the graph of the curve (V/f) x t and the graph of the
frequency x voltage pick-up.
Insert a new
test point.
Extract the
COMTRADE or CSV
Allows you to Insert or file from the waveforms.
Edit either Markings or Take a photo of the chart
Comments in graphics. by storing it on the
clipboard.
Clicking on the View option in the previous figure brings you to the following
options:
CONPROVE INDÚSTRIA E COMÉRCIO LTDA 758
CTC SOFTWARE MANUAL
Clicking on the Show Columns option in the previous figure brings up the following
options:
Enables or disables
the display of the
following values:
Module, angle, actual
value, imaginary
value, lag,
instantaneous
amplitude, frequency
and frequency
variation.
Enables or disables the
display of the following
values for the difference
between the cursors:
Minimum, maximum or
average Instantaneous
Amplitude and the RMS
value in the period.
The previous figure shows the result of this time. For a better explanation of this test,
this figure is divided into parts.
For each Test set point the user can choose what type of data will be shown in the
table according to the figure below.
Shows the Test Voltage values in Shows whether or not it Shows the nominal
mode and the volts and is an operating region. time and the
multiple (In / End) frequency in hertz And finally it shows measured time in
of the Tested whether the operation
of the tested point. seconds.
took place or not.
Points.
For example leaving the Options “General Info” and “Performance” of the previous
figure enabled the following data are shown:
Figure 1064 – VoltsPHz – Main screen – Time Test – Test set Lines
In the following figure are detalhad the insertion, the editing or the removal test
lines.
Figure 1065 – VoltsPHz – Main screen – Time Test – Insert / Edit Lines
When the user clicks on “New Line” he defines between voltage or constant
frequency as the test mode, the multiple of the division between nominal voltage and
frequency and the value of phase-phase voltage or frequency.
Confirm or Cancel
Frequency value. the line for test.
Figure 1066 – VoltsPHz – Main screen – Time Test - General Options – New Line
When clicking on “Sequence” the user defines the test mode, an initial and final
multiple, as well as the step. The different lines are created automatically.
Figure 1067 – VoltsPHz – Main screen – Time Test - General Options – Sequence
The following figure shows a zoom of the lines before being Test set. At the
moment to that are Test set are torn am up green if approved, or red s if they disapprove.
The next illustration shows the legend with the icons of the test line and the points
Test set. There is also a color legend and finally the current point value as the user moves
the mouse over the previous graph.
Colors representing
status of points:
Test line icons Orange: Not Tested;
and tested points. Green: Approved;
Red: Failed.
The previous figure shows the result of a search test. To facilitate the understanding,
each tab of this test will be shown in detail in the next figures. For each row the user can
choose what type of data will be shown in the table according to the figure below.
As an example, leaving the three options enabled in the previous figure, the
following data are shown:
Figure 1072 – VoltsPHz – Main screen – Search Test – Test set Points
In the figure are detailed following the insertion, the editing or the removal test lines.
Figure 1073 – VoltsPHz – Main screen – Search Test – Insert / Edit Lines
When the user clicks on “New Line” the settings of the fields shown below are
available.
By clicking "Sequence" the user sets the test mode and inserts the value initial, final
and the step and the number of points is calculated automatically by the software. The
following figure shows a case for the constant frequency.
Figure 1075 – VoltsPHz – Main screen – Search Test – General Options - Sequence
The following figure shows a zoom of the lines before they are Test set. The moment
they are Test set they become green points if they pass or red points if they fail.
The next illustration shows the legend with the icons of the test line and the points
Test set. There is also a color legend and finally the current point value as the user moves
the mouse over the previous graph.
Colors representing
status of points:
Test line icons Orange: Not Tested;
and tested points. Green: Approved;
Red: Failed.
Voltage and
frequency value of
the point pointed by
the mouse in the
previous graph.
39.14 Trajectory
This test is intended to allow the user to test the function overexcitation for situations
where occur variations voltage and variation in frequency at the same time.
The previous figure shows the result of a route test. To facilitate the understanding,
each tab of this test will be shown in detail in the next figures. For each row the user can
choose what type of data will be shown in the table according to the figure below.
As an example, leaving the option “Def. Route” of the previous figure, the following
data are shown:
Test point Test definition Variation of Variation of Total time Shows the
mode: final
number. voltage frequency of the tested
dV / dt, df / dt or Evaluation of
time. over time. over time. line.
the tested line.
In the figure are detailed following the insertion, the editing or the removal test lines.
When the user clicks on “New Line” the settings of the fields shown in the following
figure are available.
Phase-to-phase
Phase-to-phase
voltage.
voltage.
Mode: dV / dt, df
Voltage variation / dt or time.
over time.
In the next figure, the region within the square of the full line is the region of
synchronism and the external region is of non-synchronism. The region defined between the
two squares of dotted lines represents the tolerances given by the manufacturer of each
relay.
The next illustration shows the legend with the icons of the test line and the points
Test set. There is also a color legend and finally the current point value as the user moves
the mouse over the previous graph.
Colors representing
status of points:
Test line icons Orange: Not Tested;
and tested points Green: Approved;
Red: Failed.
A greater detail of the previous figure is shown below. In the figure below it is
detailed all the fields of pre and post fault, as well as the designation of the channels.
Enables pre-simulation
1 and 2 and post-
simulation conditions.
Designates the Module and angle values of each Module and angle values of
generation channels channel for pre-simulation each channel for post-
for each phase. condition - 1 and 2. simulation condition.
Figure 1086 – VoltsPHz – Main screen – Test Settings – Direction of Generation Channel– Pre-
Simulation and Post-Simulation
In the figure below, the initial status of the binary outputs and GOOSE outputs, pre /
post fault times is detailed.
Defines the states of Defines the states of the Defines the states of the Defines the states of the
the Binary Outputs and Binary Outputs and Binary Outputs and Binary Outputs and
GOOSE Outputs GOOSE Outputs (when GOOSE Outputs (when GOOSE Outputs (when
configured) in the Pre- configured) in the Pre- configured) in the Post-
(when configured) in
Simulation 1 cycle. Simulation cycle 2. Simulation cycle.
the Simulation cycle.
Figure 1087 – VoltsPHz – Main screen – Test Settings – Pre-Simulation, Post-Simulation Time and
Binary Outputs and GOOSE
In the figure below, all the specific configurations of the search test and other tests
are detailed.
Time
between
increments.
Extra time.
Time to
restart.
In the figure below, all the pickup, stop and trip interfaces required in the tests are
detailed, as well as the access of their respective logics.
Allows triggering by
binary, software,
logic or GPS.
TRANSDUT
SOFTWARE FOR TESTING TRANSDUCER
The options on the Settings screen are described in detail in chapter 21. The
Transducer field will be discussed throughout this chapter.
Creates a new Opens a saved Saves the Customize the Software name and
test. current test. Quick Access version.
test file.
Bar.
Equipment used
Opens the options: and serial
Restore, Move, Size, number.
Minimize, Maximize
and Close.
In the start menu, several resources are available. The following figure describes
hardware configuration features and generation controls such as start, stop, move to the next
point / line, deletion of one or all of the Tested points.
Opens a Channel Opens a screen Starts the signal Pauses signal Move on to the
Orientation screen. for Hardware generation generation. next point.
Configuration. process.
Clean the
spot
Selected
and returns
to Untested
status.
In the figure below, the resources for analysis options of the generated quantities
(waveforms and phasors) are displayed, as well as the adjustment options (General Settings
and Transducer). Access to the report is also displayed, options for viewing the units
involved (time and module) and general options for restoring and viewing the layout.
Shows the Shows the Shows the Shows the Returns to Enables /
Waveform Phasor Tab. time in time in the standard disables
Tab. seconds. cycles. layout. tabs for
viewing.
Shows
values
based on the
secondary.
Figure 1096 – Transdut – Home Menu – Options, Report, Units and Layout
Position the selected Superimposes the Hide the windows Returns to Enables /
window: on the left; selected window. for better viewing. the standard disables tabs
to the right, above, layout. for viewing.
below or as a
document.
Hide the selected Hide the selected Close all open Closes all open
tabs. window. windows. documents.
Insert password in Enables / disables the Opens the Settings Shows the
manufacturer’s address,
the Test file so as simulation mode. Since Screen when the
phone, email and fax.
not to allow the in this mode there is no software is started.
editing of certain insertion of voltage and
settings. current.
Enables /
Disables
automatic
control gain.
It is also opened at software startup, if the Opens Settings when Starting option in the
Software Options tab is selected.
In the General mode, there is the “General Information” tab, where the user enters all
the data relating to: the Test, the Test set Device and the person responsible for the
Test. There is also the tab “Reference Values” where the user informs the nominal values of
voltage, current and power for system 1. There is also a tab for “Notes and Observations”. In
the other tab, “Explanatory Figures”, it is possible to insert up to 3 figures in the most used
image extensions, such as: Bitmap; JPGE; PNG; JPG; etc.
In Transducer mode, there is the “Transducer Adjustment” tab. In it, the user chooses
the type of transducer, the minimum and maximum values of input and output of these
transducers. There is the possibility, depending on the transducer, to register it as single-
phase or three-phase. You must adjust a tolerance and choose the type of curve behavior
between: Linear, Composite or Quadratic.
Other features are as to the transducer output if it is given by voltage or current, the
closing of the star point of the CT if it is facing bar or line, how the angle is calculated if
voltage minus current or current minus voltage and finally, the stabilization time for
the transducer software to acquire the Test value.
The figure below shows all transducer test options, once a specific type of transducer
is chosen, a new tab with the name of the transducer is created with parameters that vary
from transducer to transducer.
Insert a new
Test point.
Maximizes the graphics tab.
Extract the
COMTRADE or CSV
Allows you to Insert or file from the waveforms.
Edit either Markings or Take a photo of the chart
Comments in graphics. by storing it on the
clipboard.
Clicking on the Display option in the previous figure brings you to the following
options:
Clicking on the Show Columns option in the previous figure brings up the following
options:
40.14 Test
The test of the transducer aims to inject the quantity at the input and measure the
output value and compare it with the expected value as theoretical plus the tolerances.
The previous figure shows the result of a test of an active power transducer. For a
better explanation of this test, this figure is divided into parts.
For each tested point the user can choose what type of data will be shown in the table
according to the figure below.
Reports the Shows the Shows the errors: Shows the standard Shows the voltage
nominal input deviation errors: absolute,
actual absolute, relative and current values
and output relative and total of the
value in the and total of the tested point. injected in the test.
values. output. tested point.
As an example, leaving the options “General Info” and “Actual Value” of the
previous figure marked, the following data are shown:
Creates a new
Creates a Sequence point.
of points.
Removes
Removes all selected point.
added points.
When the user clicks on “New Point” the following fields can be parameterized.
Fundamental frequency
Possibility of
of the input variable.
inserting harmonics
of up to 50th order.
The following figure shows the plot of the points before they are tested. The moment
they are tested they turn green if they pass or red if they fail.
The next illustration shows the legend for the graph of the performance of the inverse
curve and the time defined in the software.
Colors representing
status of points:
Orange: Not Tested;
Green: Approved;
Red: Failed.
Associates hardware Load table to use as a Clears the table used Enables the loaded
channels with software. reference. as a reference. table.
Figure 1122 – Transdut – Main screen – Test Settings – Direction Generation Channels
TRANSIENT
VIEW
SOFTWARE FOR VISUALIZATION OF
MEASUREMENT SIGNALS IN TRANSIENT
REGIME
The options on the Settings screen are described in detail in chapter 21. The
Distance, Synchronism, Import and Export fields will be discussed throughout this chapter.
Opens the
options:
Restore,
Move, Size, Software name and
Minimize,
version.
Maximize
and Close.
Creates a new
file.
Open a saved
file.
Password protect
certain
parameters of the
test file.
Saves the current
file.
In the start menu, several resources are available and are detailed in the following
figures.
Connects between Opens a screen for Deletes all Shifts the selected
software and GPS Setup. Sequences. Sequence to the beginning
hardware. or end.
Calls the Shows the Shows the Shows the Opens the report Shows the time in
Settings Waveform Phasor Harmonics configuration seconds or in
Screen. Tab. Tab. Tab. window. cycles.
Returns to
the standard
layout.
Enables / disables
tabs for viewing.
Figure 1129 – TransientView – Home Menu – Options, Report, Units and Layout
Hide the selected Hide the selected Close all open Closes all open
tabs. window. windows. documents.
It is also opened at software startup, if the Opens Settings when Starting option in the
Software Options tab is selected.
In General mode there is the General Information tab, where the user enters all the
data related to the test, the tested device and the person responsible for the test, there is the
Systems tab, where the user informs the nominal values of frequency, voltage, current and
power among others, there is a tab for Notes and Observations for various records.
There is an Explanatory Figures tab where the user can insert up to three figures
relevant to the test, a Check List tab to create steps to be followed during the test, there is a
tab called Others for configuring digital signal filters and finally a tab Connections allowing
the user to structure all connections between the test case and the equipment to be tested.
In Distance mode, there is the Distance Protection Adjustment tab. In it the user sets
the line length and the angle. Choose how the earth factor will be inserted. You can still
enable whether to remove the Arc Resistance and use Impedance Correction for Rated
Current. To facilitate the adjustment of the zones, there is an Import option whose purpose is
to read .RIO files. Some distance relays record line, zone and operating time settings in this
format. Importing this file, Transient reads automatically so that the user does not need to
adjust it manually. Another option would be to export a .RIO file once the feature has been
entered by the user.
The Insert option opens a new window for the user to enter the name and number of
the zone, adjust the Trigger Time, Absolute and Relative Tolerances of the impedance and
the Absolute and Relative Tolerances for the time. You can use a specific earth factor and
enable the option to Remove Arc Resistance. If the relay does not generate a .RIO file, there
is the option to design the zone to be tested or to choose between predefined
zones. Depending on the type of zone chosen, options for load compensation and
directionality are opened.
Displays the
tab for
adjusting the
sync function.
Import
Files
.RIO.
Export
Files
.RIO.
Choose the
mode of the
earth factor:
(Mod;Ang)
(R;X)
(R0;X0)
(K0)
(Rt/Rl;Xt/Xl)
(Z0/Z1).
Shows which
zones are
available.
Insert a Edit a zone Deletes a zone that has Shows the design of
already inserted. already been inserted. the chosen zone.
new zone.
Let us see the window that opens when clicking on Insert, where it is made the
adjustments of the protection zones.
Figure 1135 – TransientView – Parameters Screen – Distance – Distance Protection Adjustment - Insert
- Zone
To facilitate the visualization of the previous figure, it was divided into parts.
Zone description.
Zone type registration:
Zone index. Departure, Performance,
Extended or Block.
Links the zone to a (Informative data).
specific type of
fault. Absolute (Ω) and
relative (%) tolerance
of Impedance.
Trip time of the
zone.
Absolute tolerance
(more or less) in
seconds and relative
tolerance in% of the
time of operation.
Points inserted
when using the
option to draw.
Figure 1136 – TransientView – Parameters Screen – Distance – Distance Protection Adjustment – Insert
– Zone
Finalizes the
drawing
process.
Figure 1137 – TransientView – Parameters Screen – Distance – Distance Protection Adjustment – Insert
– Zone
The following figure shows a quadrilateral zone that was drawn using straight line
segments.
Figure 1138 – TransientView – Parameters Screen – Distance – Distance Protection Adjustment – Insert
– Drawn Area
In the next figure, the quadrilateral option was chosen directly. Note that the settings
to be entered are more intuitive with this option.
Figure 1139 – TransientView – Parameters – Distance – Distance Protection Adjustment – Insert – Zone
Quadrilateral
Then the screen where system data 1 and 2 and the settings for synchronism are
inserted.
To facilitate understanding, the previous figure will be divided into two parts.
Figure 1148 – TransientView – Sequence Tab - NODE sub tab - Two Nodes and three Sequences
In the following explanations, we will adopt that only one node (NO01) was created
containing three voltage channels and three current channels. Remember that to
add sequence s click the Insert New Sequence. The figures below show the sequence modes
that can be entered in the test.
Choose between
COMTRADE
file (.cfg
extension) or Access button for
ATP file (.lis parameterizing the
extension). comtrade / ATP file.
When clicking on the access button for the parameterization of the COMTRADE
mode and loading a .cfg file, the following screen appears:
Shows information
Picture Search for file to upload. about the file
box. loaded.
When clicking on the access button for the parameterization of the ATP mode and
loading a .lis file, the following screen appears:
Shows information
Picture Search for file to upload. about the file
box. loaded.
Sequence duration
time.
Sequence name.
Voltage and
current module.
Figure 1152 – TransientView – Sequence Tab – Output Analog. DC Sub tab – General Description
Sequence duration
time.
Sequence
name.
Figure 1153 – TransientView – Sequence Tab – Binary Outputs sub tab – General Description
Sequence duration
time.
Sequence
name.
Figure 1154 – TransientView – Sequence Tab – GOOSE Outputs sub tab – General Description
Figure 1155 – TransientView – Sequence Tab – Time and Advance sub tab – General Description
Extract the
COMTRADE or CSV
Allows you to Insert or file from the waveforms.
Edit either Markings or Take a photo of the chart
Comments in graphics. by storing it on the
clipboard.
Clicking on the View option in the previous figure brings you to the following
options:
41.9.11Phasor Tab
In this screen it is possible to view for each channel the module and the value of the
voltage or current angle, as well as the Phasors graph.
This screen has the Transparency property, and can remain open while points are
inserted or edited in the test. The following screen is for a line tested using Search Test.
Clicking on the Show Columns option in the previous figure brings up the following
options:
41.9.13Trajectories tab
41.9.15Harmonics Tab
Type of fault
to be inserted.
Button to
Table with the insert a fault
various faults that sequence.
can be inserted.
When clicking on the insert button, the following screen wills appear:
Fault insertion
Description of the mode (smart, Z-I
Sequences that cte, Z-V cte and Z-
represented a fault. Source cte).
After choosing the data in the figure above, just allocate the fault point with the
mouse in the impedance diagram, as shown in the figure below:
Confirmation
of the chosen
Table with several points point.
chosen and their
respective information.
STATISTICAL
ANALYSIS
SOFTWARE FOR STATISTICAL ANALYSIS OF
A DATA SET
Figure 1173 – Statistical Analysis– General Information – Identification of the Analysis and Responsible
In order for a table to be successfully imported, the file . The chosen .csv must
contain at least two defined lines. The first line must be filled with the set of descriptions of
the constants and variables to be analyzed and the later lines will represent the values or
labels assumed by the constants / variables in each test. In cases where these requirements
are not met, the software returns the message below to the user.
It is Important to note that all lines in the file. Imported CSV must contain the same
number of columns (this number is defined by the number of columns observed when
reading the first line of the file .CSV), that is, every constant / variable must be associated
with a value / label in each test. Thus, if a line with a different number of columns is read, it
will be neglected.
After you add a file, table will be generated with all their values. Each column of
data can be classified as: Variable, Constant or None. When a column is defined as
"Constant", the software will perform a processing in order to check all the possibilities
of values assumed by the constant in the tests of the inserted table, aiming to make available
such values for use in Plot XY analysis, if the constant is selected. When a column is
defined as "Variable", the data in this column will be part of the sample space used by the
software. Thus, it can be used s in all kinds of software analysis. Finally, if the option
chosen is “None”, the column data will be neglected during the analysis. The figure below
illustrates the import of a file .CSV , where the “LocFalta” and “SIR” columns were selected
as constants, the “TRIPZ1”, “TRIPZ2” and “TRIPZ1B” columns as variables and the “X /
R” column was neglected.
It is Important to note that the software performs a special treatment when importing
non-numeric values (defined by labels) and blank gaps. This treatment is performed in a
window where the user will define values for the labels found in the file .CSV. For example,
suppose the following content in a file .CSV.
Line Column 1 2 3
1 Nº Protection Zone Time TRIP
2 1 ZONA 1 0,053
3 2 ZONA 1 0,055
4 3 ZONA 1 0,000
5 4 ZONA 2 0,212
6 5 ZONA 2 0,214
7 6 ZONA 2 0,218
8 7 ZONA 3 0,311
9 8 ZONA 3 0,316
10 9 ZONA 3 0,314
When importing this file .CSV the software will identify both the blank gap (row 3
(counted from the description lines), column 3) and the values in column 2 that were defined
by the labels: ZONE 1, ZONE 2 and ZONE 3. The user must then define numerical values
for both blank gaps and for values defined by labels, thus enabling subsequent analyzes to be
carried out. The open window will be, in this case, the one shown below.
Definition of
the initial row
and column for
reading the .csv
file.
List of labels
found in the
.csv file and
their respective
values to be
defined.
List of blank
gaps in the .csv
file and their
respective
values to be
defined.
Figure 1177 – Statistical Analysis – Tables Tab - Definition of values for variables not defined or defined
by labels.
Assuming that the filling was done according to the figure above, the software will
import the file and build the table as shown in the figure below.
Note that the blank space was filled by the value “0” chosen by the user. The
variables defined by labels follow their respective values in parentheses for user recognition.
Move up.
Delete Analysis.
Figure 1180 – Statistical Analysis – Analysis Tab - Filters / Weights List - Filters Tab
In the "Weights" tab, each value of a "Constant" can have an assigned weight. In this
way, it is possible to assign a relevance to certain data directly influencing the resulting
graph. If for example, we have only one test with the constant LocFalta assuming the value
10 and its assigned weight is 3, the software will consider that this test was repeated 3 times.
Figure 1181 – Statistical Analysis – Analysis Tab - Filters / Weights List - Weights Tab
On this screen we will have all the options for handling XY graphics. Such options
are listed below.
Assuming the situation above, where the selected constant was “LocFalta”, the
selected variable was “TRIPZ1”, all statistical options were marked and the option to plot
curves was chosen as “All ”, we will have the following situation:
1) - The X axis will consist of all possible variations of the “LocFalta” constant, in
ascending order.
2) - Each checked statistical option will give rise to a different curve. The
“Maximum” curve, for example, at point X = 0 (LocFalta = 0) will have on the Y axis the
maximum value found for the variable “TRIPZ1” in all tests where the constant “LocFalta”
assumes a value equal to 0 The point X = 10 (LocFalta = 10) will have on the Y axis the
maximum value found for the variable “TRIPZ1” in all tests where the constant “LocFalta”
assumes a value equal to 10. And so on until all are found the maximum values for all
variations of the “LocFalta” constant.
3) - The procedure will be the same for the generation of the Minimum, Arithmetic
Average, Geometric Average, Median, Standard Deviation and Variance curves.
Finally, assuming the same situation as above, where the selected constant was
“LocFalta”, the selected variable was “TRIPZ1”, all statistical options were checked, but
now the option to plot curves was chosen as “SIR” (or any other constant available in the
table ), we will have the following situation:
1) - The X axis will consist of all possible variations of the “LocFalta” constant, in
ascending order.
CONPROVE INDÚSTRIA E COMÉRCIO LTDA 835
CTC SOFTWARE MANUAL
2) - Each statistic option checked will give rise to a curve , for each possibility of
constant change "SIR", i.e. in this case will be 8 options for statistical calculations are
multiplied s by two variation possibilities for the constant "SIR" , which will result in a total
of 16 curves plotted on the graph .
3) - The “Maximum” curve, for example, at point X = 0 (Missing Loc = 0) with the
value of the constant “SIR” equal to 5, will have on the Y axis the maximum value found for
the variable “TRIPZ1” in all tests where the “LocFalta” constant assumes a value equal to
0 and the “SIR” constant assumes a value equal to 5 . At point X = 10 (LocFalta = 10) the
maximum value found for the variable “TRIPZ1” will be on the Y axis in all tests where the
constant “LocFalta” assumes a value equal to 10 and the constant “SIR” assumes a value
equal to 5 . And so on until you found all the maximum values for all variations
of s constant s "LocFalta" and "SIR" .
4) - The procedure will be the same for the generation of the Minimum, Arithmetic
Average, Geometric Average, Median, Standard Deviation and Variance curves.
Clicking on the “Table ” sub-tab, you can see in which table and line each point was
removed to compose each curve of the graph.
Figure 1184 – Statistical Analysis – Analysis - Results ( PlotXY Graph ) - Sub Tab Table
It is also Important to point out that all the variations of the Constants that are used
for the construction of the curves respect what was defined in the Window “List Filters /
Weights”, that is, if in the Filters window the value “0” of the constant “LocFalta” is
disabled, it will be disregarded for all curves in the current analysis.
If an analysis of the selected type " Graph of bars ", the displayed screen will look
as shown below.
On this screen we will have all the options for handling bar graphs. Such options are
listed below.
Visualization of the Shows the number of
graph in 2D or 3D. occurrences on the graph.
If it is changed to Class
Types of analysis: Interval in this setting,
1 Variable 1 Interval; the user starts to define
2 Variables 1 Interval; the number of classes no
2 Variables 2 Intervals. longer, but the period of
each interval.
The user will have two
possibilities to define the Here the user defines the
number of intervals for the Variable to be analyzed
selected variables. With the as well as its unit.
Number of Classes option,
the difference between the
highest and lowest value of Here the user can define
the analyzed variable will minimum and maximum
be defined and this value values of the selected
will be divided by the value variable that will be
considered in the analysis.
defined in the text box by
the user. The result will be
the number of bars shown in
Results obtained by the
the graph (defined
software with the
intervals).
processing of the
selected variable.
Figure 1186 – Statistical Analysis – Bar Graph– Definitions for graph construction
CONPROVE INDÚSTRIA E COMÉRCIO LTDA 837
CTC SOFTWARE MANUAL
Assuming the above situation, where the selected variable was “TRIPZ1”, the
interval defined by the number of classes equal to 5, 3D graphic, showing the number of
occurrences and the limitation of the variable between the values -1000000 and 1000000, we
will have the following situation.
1) - All values assumed by the variable “TRIPZ1” will be acquired in all tests of all
inserted and enabled tables.
2) - In possession of these values, values that are not in the defined range of -
1000000 to 1000000 will be disregarded. The disregarded values are also counted and can
be viewed in the bar graph later as “Out of Range” values.
3) - With the remaining values, the maximum and minimum values will be checked,
with the result of subtracting them divided by 5 (number of classes) and with this final
result, the period of each interval to be represented in the bar graph.
4) - If the final result is equal to 2, for example, the intervals of the bar graph would
be: Interval 1 {0 to 2}, Interval 2 {2 to 4}, Interval 3 {4 to 6}, Interval 4 {6 to 8}, Range 5
{8 to 10}.
5) - With the definition of the intervals, all values of the variable “TRIPZ1” are
traversed and their distribution is carried out between such intervals, so that the user can
view a frequency distribution through the bar graph .
6) - Finally, statistical calculations are performed with the sample space of the
selected variable and the results are made available to the user.
If selected an analysis of the " Graphic Pizza ", the displayed screen will be similar
to the one below.
On this screen we will have all the options for handling the pie charts. Such options
are the same as those observed for the bar graph, however, this type of analysis does not
allow the user to represent two variables on the same graph as is made possible by bar graph
analysis.
Assuming the same selection observed for the example of the bar graph, where the
selected variable was “TRIPZ1”, the interval defined by the number of classes equal to 5,
3D graph, showing the number of occurrences and the limitation of the variable between the
values -1000000 and 1000000, we will have the same processing situation for building the
graph.
The user will be allowed to choose which analyzes will be presented, if graphs and /
or tables will be included, if the sample space of each analysis will be detailed and if the
general information will also appear in the report. Having made the choice of what will be
included in the report, the user just needs to select the language of the report and click OK.
AUTODIAG-
NÓSTICO
SOFTWARE FOR ANALYZING THE
OPERATION OF THE TEST SET
Create, open or save a The generation starts, generates Clears a selected run, clears all
new file. from or only the selected test and runs and creates a report.
for generation.
Shows all Displays the Shows the general Shows the list of errors
rehearsals. history of settings and parameters in the test set.
problems. of the test set.
REMOT
ACCESS
SOFTWARE TECHNICAL SUPPORT FROM
CONPROVE GET REMOTE ACCESS
GPS
INSTRUCTIONS ON OPERATING THE
CE-GPS
45.1 Description
The CE-GPS is a lightweight and portable device, to receive the signals from the
GPS Satellites (24 satellites in total) specially thinking about End -to- End tests , to be used
with the Conprove Test Bag model CE600X . It allows programming the firing with up to
100 na in seconds of resolution allowing synchronization of two or more units at a distance
with an accuracy equal to or better than 1 micro second. The unit has a cable 15 meters long
for connection with its antenna, as well as a communication cable between the test case and
the CE-GPS. The equipment is controlled by the test bag operation software itself,
simplifying the adjustments for the user.
APPENDIX A
POWER AMPLIFIER USAGE MANUAL CE-4000
A1 - Introduction
In this manual the user finds all the information necessary to work with the External
Amplifier CE-4000. It was designed in an objective way, aiming to put the user immediately
in contact with the equipment.
The external single-phase current amplifier CE-4000 was designed to work with the
CE-6000, the CE-6006 and the CE-6003, allowing greater current and power in the tests.
Some software of the CE-6006/3 has been changed to work with the amplifier, they
are:
CALIBRAC. The new version allows calibration of the amplifier with a standard
user-defined ammeter, similar to the calibration process of the CE-6006/3.
MANUAL. In the new version, the user will be able to work with the amplifier,
having total freedom of signal generation and control of the range.
TRANSIENT. In this version, the amplifier can be easily used in tests that require
higher currents and power.
SEQUENC. In the new version, the user will be able to work with the amplifier,
having total freedom of signal generation and control of the range.
RAMP. In the new version, the user will be able to work with the amplifier, having
total freedom of signal generation and control of the range.
MASTER. In this version, the amplifier can be easily used in tests that require higher
currents and power.
SOBRECOR. In this version, the amplifier can be easily used in tests that require
higher currents and power. The software automatically controls the amplifier's range,
switching it, in order to better adjust to the needs of the in-process tests
A2 - Hardware
The CE-4000 Current Amplifier allows the generation of up to 100 A. For better
quantization of the signal and use of power, the amplifier has several tapes. The maximum
power value is achieved when the current generated is equal to the maximum value of the
selected tape. Thus, the user must select the tape whose maximum value is closest to the
value of the current that will be generated.
The CE-4000 also has two power modes switched by key, which allows the
equipment to work with 200 W in steady state or up to 400 W transient. The figure below
shows the transient power curve.
NOTE: The above graph is valid for s frequency fundamental s 60 and 50 Hz, measures the current output
terminal. For frequency s harmonics the power drops considerably.
The CE-4000 panel layout is shown in the figure below, each panel field is
commented below..
In the CE-4000 panel, the output field is located that informs the user which tape the
amplifier is working on, that is, which tape is selected: 7, 10, 15, 20, 30, 40, 60, 80 or 100
THE.
On the left part of the panel, there is the terminal where the communication cable
with the CE-6006/3 (CE-6006/3 EXCITATION) is connected , reset push- button and the
OVERLOAD indicator led, the power switch. output, 200 W permanent or 400 W transient
and the light indication whether the amplifier is on or not (ON).