Package: Effluent Treatment Plant Project: 1 X 660 MW Bhusawal Tps Unit 6
Package: Effluent Treatment Plant Project: 1 X 660 MW Bhusawal Tps Unit 6
Title:
Data sheet For EQMS Analyzer
21-02/M (92-EQMS) 18.04.2024
A GENERAL
1 Location Open to Sky; Installed close to Treated water Disposal
Pumps; Installation frame to be mounted on CMB Tank
wall as per site condition
2 Quantity 01; One Multi parameter extractive type Analyzer with
external sensor for pH and Conductivity complete with
sensors; sensor adopter cable; Cleaning Solution Drum
3 Spare NIL
4 Total 1 No.
5 Application Effluent Quality Monitoring System (EQMS) for Treated
Effluent Water quality monitoring
6 Ambient Temperature Operating Temp.; (5 -550C)
7 Surrounding Atmosphere Dusty, Humid & Corrosive atmosphere in a Thermal
Power Plant
8 Relative Humidity < 90% RH
9 Design Condition To meet System Requirement; Monitor the Treated Water
Parameters
10 Scope of Supply i) SS 316 L enclosure Analyzer with UV module for COD /
BOD / Turbidity in treated effluent water;
ii) External pH Sensor with 3 M Cable
iii) External Conductivity Sensor with 3 M cable
iv) External Temp Sensor with 3 M cable ;
v) Cleaning Accessories as per Mfr standard
vi) Standard Solution for Calibration of pH and
Conductivity Sensor
vii) FRP Housing Panel weather proof full cover Canopy;
Stand with Accessories
11 Make Hemera; France marketed and serviced by sole and
exclusive distributor M/s. Agua Instruments Pvt Ltd ,
Mumbai
12 Mounting Type Analyzer - panel mounted; Sensor - Flow Through
mounted
13 Model Number L- 800
14 Analyzer Tag No. GTA 24 (CQ001)
B TECHNICAL PARAMETER
1 Method of Measurement/ i)) COD / BOD / Turbidity: = UV – visible Absorption
Operational Principal Spectroscopy with double beam full spectrum scanning.
ii) pH : Electrochemical Sensor
iii ) Conductivity : Electrochemical Sensor
iv) Temp : RTD Sensor
2 Parameter Range i) COD – 0 to 1000 ppm
ii) BOD – 0 to 1000 ppm
Project: PT Plant; 1x660 MW
BHUSAWAL THERMAL POWER
00 Submitted for approval 18.04.2024 PROJECT
Rev Description Date Client: BHEL/NTPC
21-02/M (92-EQMS) 18.04.2024
NC
M
E
selective
L800
Safe area
z
L800 UV-SPECTROSCOPY MULTI-PARAMETER ANALYZER
SENSOR & METHOD RANGE ACCURACY REPEATABILITY RESOLUTION RESPONSE TIME
pH
0-14 pH ± 1 % FS ± 1 % FS ± 1 % FS ≤ 5s.
Electrochemical Sensor
Turbidity— Turb
0.1 ... 200 NTU ± 5 % FS ± 5 % FS ± 5 % FS ≤ 10s.
UV Visible Spectroscopy
Electrical Conductivity — EC
0.1 ... 1000 mS/cm ± 1 % FS ± 1 % FS ± 1 % FS ≤ 5s.
Electrochemical Sensor
Temperature — Temp.
0 ... 100 °C ± 1 % FS ± 1 % FS ± 1 % FS ≤ 5s.
PTD Sensor
E IN FRA
AD
years
WARRANTY
• Indoor / Outdoor
›U
V-Visible Spectroscopy with Full Spectrum • Stainless Steel (IP65 in option)
scanning with double beam • High quality material
› Spectrum range — 180 nm to 750 nm • No spare parts needed to be changed
with 0.10 nm. Scanning rate with turbidity • Multiplexing system in option
and colour compensation • Integrated touch screen
• Once a year calibration
APPLICATIONS
WASTE WATER DRINKING WATER PROCESS WATER SURFACE WATER SEA WATER
The measuring principle is based on UV spectroscopy. The absorption spectrum calculation is the difference
between incident light (IT0) on zero and transmitted light (ITS) on the sample.
A = log IT0/ITS
The concentration of molecules (c) are linear to the absorbance spectrum (A) and the optical path (l) of the
measuring cell.
ε = A/lc
Spectrograph
340 mm 195 mm
HEIGHT:
430 mm
WIDTH:
340 mm
DEPTH:
195 mm
430 mm
ENCLOSURE
Material SS316L
Weight ± 15 kg
SAMPLE CONDITION
CONTROLLER
Memory 16 GB SD card
COMMUNICATION OUTPUT
Digital RS485 / Modbus (Slave or Master) - Show value, calibration status, fault status...
POWER SUPPLY
DO NOT HESITATE
TO CONTACT US
THROUGH OUR SALES
REPRESENTATIVE
Hemera Analyzers
www.hemera.fr
E IN FRA
AD
sales@hemera.fr • +33 6 77 87 63 00 • +33 4 76 63 27 52