Manual Mini200
Manual Mini200
User Manual
Content
Preface ........................................................................................................................................ I
Guarantee...........................................................................................................................II
Product manual................................................................................................................ IV
7.2 QC ........................................................................................................................... 85
9.3.2 Purified water, waste water and cooling alarm checking ............................. 101
9.4.2 The Question of closed software can not be opened .................................... 102
9.6.1 The way to deal with the failure of alarm .................................................... 129
Preface
I
Auto Chemistry Analyzer User Manual
Guarantee
Warranty period
One year from the date of completion of the installation.
Ensure that content
In the warranty period, if any failure caused by the design, manufacturing defects,
will be repaired free of charge, our company will take the appropriate content of the
corresponding countermeasures.
Statement
The company has the final interpretation of this manual.
The company shall be responsible for the safety, reliability and performance of the
product only if all of the following requirements happened:
1. Assembly operations, expansion, re-adjustment, improvement, repair and
replacement of parts are performed by professionals recognized of the company.
2. All repairs involving replacement parts and supporting the use of accessories,
supplies are original of the company (original) or approved by the company.
3. The related electrical equipment is according to national standards and the use of
the manual requirements.
4. Product operation is carried out according to the instruction manual.
Disclaimer
The company shall not be liable for any equipment failure, damage, any direct or
indirect damage that may occur during the use of the equipment.
1. Malfunction or damage due to violation of the instructions, precautions, and
intended use of this manual.
2.Failure and damage caused by operations performed by inspection personnel,
doctors or laboratory technicians who are not trained by the company or the
company's designated agents
3. Malfunction or damage caused by repair or alteration of the other company.
4. Malfunction or damage caused by use instruments of other company at the same
time.
5. Malfunction or damage caused by operating environment not corresponding to the
specified operating environment (power conditions, installation environment, etc).
6. Malfunction or damage caused by natural disasters such as earthquakes and floods.
II
Auto Chemistry Analyzer User Manual
3
Auto Chemistry Analyzer User Manual
Product manual
4
Auto Chemistry Analyzer User Manual
Transport
In the packaging states, the instruments conduct transportation according to the
requirements of the transport contract, in the transport process to prevent rain and sun
exposure, to prevent severe impact, weight , heavy pressure and dumping.
Note: If the instrument has been unpacked, before moving the instrument, please re-
packaging equipment before transport.
Store
Packaged instruments should be stored at -10℃ to 40℃and relative humidity not
more than 85%, non-corrosiveness gases and in a well-ventilated environment.
5
Auto Chemistry Analyzer User Manual
6
Auto Chemistry Analyzer User Manual
Safety information
This chapter introduces the safety symbols and their meanings, summarizes the safety
hazards and precautions for using the instrument, and the specific meaning of the
label affixed to the instrument. It also lists the content of toxic and hazardous
substances contained in each part of the instrument is in accordance with the relevant
standards or not.
Safety symbols
Various safety symbols are used in this instruction manual to remind you of the
precautions to be taken during operation. As shown in the following sheet:
Symbols Slogan Introduction
Indicates a risk of biological infection, if not
Biological infection in accordance with the instructions, there may
be the risk of biological infection.
Protective
Please ensure reliable grounding protection.
grounding
VII
Auto Chemistry Analyzer User Manual
VIII
Auto Chemistry Analyzer User Manual
Safety precautions
For the safe use of this instrument, please carefully read the following safety
precautions. Any violation of the following safety precautions may result in personal
injury or damage to the instrument.
Warning:
IX
Auto Chemistry Analyzer User Manual
• Do not open the back cover or the side cover when the
main power is turned on.
• Reagent and sample are spilled in the instrument, which
may cause malfunction and electric shock. Do not place
samples or reagents on the instrument. If splash occurs,
please turn off the power immediately and contact the
company service.
Burn protection
To prevent burns from halogen lamps, observe the following precautions.
Burn protection:
10
Auto Chemistry Analyzer User Manual
11
Auto Chemistry Analyzer User Manual
Waste treatment
To prevent environmental pollution and personal injury from waste, please
observe the following precautions when handling waste.
Biological infection:
Instrument treatment
Please dispose of the discarded instruments as follows
Warning:
XII
Auto Chemistry Analyzer User Manual
Operation precautions
To use the instrument correctly and effectively, please read the following precautions
carefully.
Intended use
Warning:
Operator
Warning:
Environment
Attention:
13
Auto Chemistry Analyzer User Manual
Instrument operation
Attention:
(2) Before using the instrument for the first time, please carry
out calibration, quality control, to confirm the instrument is
working properly.
(4) Do not open the sample / reagent tray cover during working.
14
Auto Chemistry Analyzer User Manual
Computer viruses can destroy software and data. Do not use the
computer for other purposes or connect to the Internet.
(7) Do not touch the monitor, the mouse, or the keyboard with
wet or sticky chemicals.
15
Auto Chemistry Analyzer User Manual
Instrument maintenance
Attention:
Before cleaning, turn off all power to the instrument and unplug
the power cord. Take necessary measures to prevent water
droplets from entering the instrument during cleaning. Doing so
may cause damage to the instrument or personal injury.
Parameter setting
Attention:
16
Auto Chemistry Analyzer User Manual
Sample
Attention:
(1) Use the full separation serum samples and urine samples
that do not contain suspended solids. If the serum sample
contains fibrin, or urine samples containing suspended solids
and other insoluble impurities, are likely to block the needle,
affecting the analysis results.
XVII
Auto Chemistry Analyzer User Manual
(7) Before testing, make sure the sample is placed in the correct
sample position, otherwise the correct result may not be
obtained.
Reagents, calibrator, QC
Attention:
18
Auto Chemistry Analyzer User Manual
Data backup
Attention:
19
Auto Chemistry Analyzer User Manual
Auto Chemistry Analyzer is mainly used for the quantitative determination of serum,
plasma, urine, cerebrospinal fluid of clinical biochemistry project, immunization
programs, therapeutic drug monitoring and drug abuse monitoring. The system
consists of analyzer, computer, LCD monitor, keyboard, mouse and printer. The
appearance of the Auto Chemistry Analyzer as figure1-1.
1
Auto Chemistry Analyzer User Manual
Reagent bottle
20mL, 50mL
specifications
2
Auto Chemistry Analyzer User Manual
3
Auto Chemistry Analyzer User Manual
The structure is based on the "two trays + one probe + one stirring"(reaction tray and
reagent&sample tray, one dispensing probe, one mixing arm). The dispensing probe
is used for adding sample, reagent 1 and reagent 2. Mixing arm is used for mixing
reagent and sample. Optical measurement part adopts after spectrophotometry to the
cuvettes for photoelectric collection. During the test, the cuvettes will be washed
automatically by the cleaning needle.
4
Auto Chemistry Analyzer User Manual
10
11
7 8 9
5
Auto Chemistry Analyzer User Manual
12 14 13
115
116 17
8
9
10 14
1 15
16
17
12 13
No Name Remark
4 Left side panel Power switch is located on the left side panel.
6
Auto Chemistry Analyzer User Manual
Communication
12 Used to connect analyzer and PC.
interface
Connect grounding line, water alarm float switch and
13 L-type support
water tube connection.
6 3
5
2
2 4
Figure 1-6 Internal structure
7
Auto Chemistry Analyzer User Manual
No Name Remark
Reagent&Sampl
1 Put reagent bottle, sample cup and test tube.
e tray
2 Cleaning cup Clean dispensing probe.
Extraction reagent and sample from
3 Sample system reagent&sample tray, and distribute to the
cuvettes.
4 Mixing system Mix reagent and sample in the cuvettes.
Fixed cuvettes, keep suitable temperature and
5 Reaction system
provide reaction conditions.
6 Cleaning system Used to clean cuvettes.
1 2 3 4 5 6 7 8 9
10 12 14 16 17
11 13 15
1-Sample pump, 2- Cleaning arm, 3- Water L frame, 4- Reaction tray, 5- sample arm,
6- Cleaning cup, 7- Reagent&sample tray, 8-Circuit board assembly,9-Power panel,
10- Pump, 11- Solenoid valve, 12- Fan support, 13- Halogen lamp,14- Cleaning cup,
15- Optical component, 16- Stirring arm, 17- Circuit board
Figure 1-7 Instrument structure
8
Auto Chemistry Analyzer User Manual
The analyzer consists of operation part and analysis part, which are connected by
Internet. As shown in figure 1-8.
Operation part consists of computer, monitor, keyboard, mouse and printer etc.
Operation system: Windows7 or above, 32 bits or 64 bits. Already installed
Microsoft .NET FrameWork 3.5 and software of the analyzer. It is recommended to
install Microsoft Office Access software.
Computer configuration: Must be a branding machine CPU frequency ≥2.80GHz,
hard disk≥80G, memory≥2G, Internet and stable USB interface.
Monitor: 17 inches or more, not less than 1280 x 1024 resolution.
Printer: Inkjet or laser printer, print the test report.
Mouse: Able to complete the operation of the software.
Keyboard: Able to edit each function of analyzer.
9
Auto Chemistry Analyzer User Manual
As shown in figure 1-9, sample tray is divided into reagent ring and sample positions,
total is 37 sample positions. We can put the standard cup that contain sample,
calibration liquid, quality control liquid in the designated position, through the
sample disc rotating to send samples to dispensing position of the sample arm.
2
3
10
Auto Chemistry Analyzer User Manual
1
3
2 4
2 3
11
Auto Chemistry Analyzer User Manual
When check movement, find the "Instrument Check", input target value in "Target
Position", and click “Y” to check, "Zero Position" is back to original position.
As figure1-14 is shown, sample arm consist of dispensing probe, cross arm, linear
optical axis, drive motor and photoelectric switch etc.
2
7
3
8
4
9
10
11
12
5
12
Auto Chemistry Analyzer User Manual
In the process of taking reagent, in case of reagent dilution, the probe takes reagent in
accordance with the "set quantitative+ allowance" rule and add in the cuvettes.
Reagent volume is 20μl-350μl generally. The setting unit can be 1μl.
After turn on the power, the probe will first rise and move to the top of the cleaning
cup and drop down for washing, then rise up.
13
Auto Chemistry Analyzer User Manual
During testing, the probe follow the order sample position, reaction position, and
cleaning cup.
Nut
Up fixed screw
Spring
Adapter sleeve
Anti-collision base
14
Auto Chemistry Analyzer User Manual
5
1
3
4
15
Auto Chemistry Analyzer User Manual
1
2 5
3
6
16
Auto Chemistry Analyzer User Manual
1
2 6
7
8
3
4 9
5
1-Stirrer cap, 2- Stirrer, 3- Mixing arm drive seat,4- Up and down belt wheel,
5- Mixing arm frame, 6- Spline shaft, 7- Left and right belt wheel, 8- Left and right driving motor,
9- Up and down driving motor
Figure 1-22 Mixing arm
When power on, the stirring probe rises up, swing left and right, stops at the top of
cleaning position, then fall down to the cleaning cup, when finish, back to upward
side.
Initialization process is the same motion state after turning on the power supply.
When run test, the mixing arm fall vertical to washing position first, then swing to 32
position in reaction tray to stir reaction fluid.
When check movement, click "Instrument Check", click "
17
Auto Chemistry Analyzer User Manual
1 6
7
8
2 9
3
4 10
11
5
12
As figure 1-23 shown, cleaning arm consists of washing probe, drive motor,
photoelectric switch etc.
When turning on the power supply, cleaning arm rises up and then drop into the
cuvettes, cleaning cuvettes, after cleaning up and stopped above the cuvettes.
When check movement, click "Instrument check", choose "
18
Auto Chemistry Analyzer User Manual
Reagent incubator
Peltier
Cooling Piece
Cooling Fan
Semiconductor cooling piece is divided into cold and hot surface, please notice the
difference when installation.
Cooling control system is mainly composed of reagent&sample tray, through cooling
piece and cooling fan to keep reagent&sample tray temperature within 2℃~15℃.
1.4.2.7 Optical inspection system
Optical detection system is used to measure the absorbance of the reaction fluid in
the cuvettes. The system consists of optical system and signal detection, its main
function is to detect the change of light intensity through reactant, by the method of
photoelectric converting optical signal to electrical signals, by measuring electrical
signal variation amount to reflect the change of light intensity. Its working principle
diagram is shown in figure 1-25.
Optical system is composed of light source, colorimetric system and fiber optic
components, used to provide enough intensity monochromatic light, after the light
source produce stable and continuous spectrum, polychromatic light divide into
monochromatic light through the fiber optic components, then after filtering the
specific wavelengths of monochromatic light into the signal detection system.
Signal detection system includes photoelectric conversion section and AD collection
and processing. Its main function is to convert the light of the transmission to
electrical signals, then amplify, analysis and processing, acquiring the light path
signal.
19
Auto Chemistry Analyzer User Manual
Cuvettes
Optical signal
Light source detection
20
Auto Chemistry Analyzer User Manual
Chapter 2 Installation
In order to ensure the normal operation of the instrument after installation, only
person authorized by the company is allowed to doing the initial installation and
setting.
2.1 Unpacking
After the goods arrival, please carefully check the instrument packaging, to see if
there is physical damage. If any damage, please contact with the company or local
distributor. After confirming no external damage, unpacking follow steps below:
●Open the accessory box and check whether object is complete according to the
packing list or not. If there is any missing, please contact with the company or local
distributor.
●Carefully check the instrument appearance, if there is any damage, please contact
with the company or local distributor.
●Ensure all parts of instrument are in good appearance without damage before
transporting.
●All moving and transportation must keep instrument upright direction, do not tilt or
side up.
●Avoid vibration while transporting, please check and debug after transporting. If all
is OK, please follow chapter 2.2.
21
Auto Chemistry Analyzer User Manual
Power supply:AC220V/110V±10%,
50H/60Hz±1Hz
Instrument power: 300VA
Instrument is equipped with a three
core power line, red line, blue
yellow green ground zero line.
Grounding method:
first the wire ends
stripping, then red
ground bolt nut loose,
then the wire end
passes through the
grounding bolt hole,
UPS: power is
and screw the nut on
more than 1000W
the ground. The
ground should be
pressed into the
building ground.
The power socket
The waste bucket: analyzer using needs
Medical good grounding (at
through a flexible
pure least one 20A
pipe, a silicone tube
water switchboard and three
connected with
machine 5A socket can be
biochemical analyzer
L fra me under the used).
two water outlet
joint, float switch by Power
AC Voltage
waste liquid level. Outlet Regulator
Monitor: 17 inches Mainfra me computer: Must be a branding machine CPU Printer: ordinary
or more, not less frequency is more than 2.80GHz, the hard disk is more than 80, inkjet or laser
than 1280 x 1024 more than 2G of memory, Ethernet interfa ce and stable USB printer that
resolution interface, install Windows7 and above version of the operating prints the
system, the 32 bit or 64 bit version of Microsoft installed.NET measured data
FrameWork 2 and immunoanalys er exclusive software.It is
recommended to install Microsoft Office Access software.
In figure 2-2, the connecting lines indicate the corresponding piping or lines.
22
Auto Chemistry Analyzer User Manual
To convenient for the equipment operation maintenance and repair, auto chemistry
analyzer need to satisfy the following conditions when installation.
a) The distance between the left and right sides against the wall should be greater than
500mm
b) The distance between the rear panel against the wall should be greater than 500mm.
c) The distance between front side of the analyzer and other instruments should be
greater than 1000mm.
d) The space of the waste liquid discharging device and the pure water supplying
device should be ensured while installation.
Wall
Minimum50
0
Analyzer
425
Minimum 1000
Front
Unit:mm
23
Auto Chemistry Analyzer User Manual
Note:
The instrument will not guarantee the accuracy of the normal operation and test data
in the rugged environment mentioned above. If the temperature and humidity cannot
meet the above requirements, please use the air conditioning equipment.
In the working process, the instrument can produce heat and it is discharged through
the instrument back. Working environment should be kept well ventilation and
ventilation equipment can be used when necessary. Direct flow of air should be
avoided to the instrument, or it may affect the accuracy of the instrument test.
24
Auto Chemistry Analyzer User Manual
The following requirements of water supply and drainage must be met before the
instrument is delivered:
Instrument requires deionized water, the conductivity below 1μs/cm, water
consumption is 4L per hour. We recommend to use 15L water machine.
Ensure that the instrument water supply hole and pipe installation should be
unimpeded. In addition, the inlet hole of L rack should be higher than the pure water
barrel, and difference of feed water level should be less than 50cm.
Ensure that the instrument drain hole and pipe installation should be unimpeded, and
the equipment L rack outlet hole should be higher than the waste barrel (or waste
discharge), waste pipe length should not exceed 2m.
Step 1: Click the button in the lower left corner of the computer, as shown
below, find the control panel.
25
Auto Chemistry Analyzer User Manual
26
Auto Chemistry Analyzer User Manual
Windows7 system generally comes with Microsoft. NET Framework 3.5 framework,
if the computer does not come with the Microsoft. NET Framework 3.5 framework,
you need to manually install the installation steps are as follows:
27
Auto Chemistry Analyzer User Manual
Step 1: the instrument application software installation CD into the computer's drive,
find the biochemical control software files .
Step 3: Open auto chemistry analyzer installation wizard, as shown below, click
“Next” to enter the selected installation folder.
Analyzer Software
Analyzer software
Analyzer software
Step 4: Select the software installation path, the default is D:\\Program Files \\
Analyzer software \, click Next to confirm the installation.
28
Auto Chemistry Analyzer User Manual
Analyzer Software
Analyzer software
Analyzer software
Analyzer software
Step 5: confirm the installation, click “Next” to install, the installation process shown
in figure 2-14.
Analyzer software
Step 6: Installation is complete, as shown in figure 2-15, click the "Close" button to.
close the form
29
Auto Chemistry Analyzer User Manual
Analyzer Software
Analyzer software
Step 8: In the installation directory D:\\Program Files \ Analyzer software in the open
and paste, as shown in figure 2-17:
30
Auto Chemistry Analyzer User Manual
If the program need uninstall, click on the Control Panel\Programs and Features\
Uninstall the program, find “Analyzer software”. Or click the "Start" menu to select
"All Programs", choose to open “Analyzer software” analyzer folder. Uninstall
analyzer software to complete the uninstallation of the program, as shown in figure 2-
18.
After the software is uninstalled, the default database is retained. If you want to
upgrade the software version, after the software is uninstalled, install the software in
the original software path, and retain the original data.
Before using the operating software for printing, please check the following
information:
(1) Check whether the printer driver is installed or not.
31
Auto Chemistry Analyzer User Manual
(2) Check whether the data cable between biochemical analyzer and printer is correct
or not.
(3) Make sure that the printer has been installed with the specific paper, you can print
after opening the printer switch.
Auto Chemical Analyzer operating software is divided into two types of closed and
open. The enclosed instrument needs to be equipped with a code scanner. Open
instruments do not require a code scanner
Auto Chemical Analyzer scanner use USB port.
After installation, install the pure water float switch and washing solution float switch,
so that the floats to keep floating state. Connected to the following power:
Power supply, computer power supply, computer display supply, printer power, open
the master switch (on the rear board of machine), the instrument master and the
cooling switch (on the left board of machine).
The initialization operations are as follows:
1. Mixing arm rising, cleaning arm rising.
2. Mixing arm swing then back to the cleaning position.
3. Sample tray initialization.
4. Repeat one time of step 1 and 2.
5. Sample arm rising, then swing.
6. Reagent&sample tray initialization then clockwise rotation.
7. Repeat one time of step 5 and 6 then machine goes into standby mode.
Note: When the various positions are not in the zero position with the instrument in
the standby state, it needs to enter the software and click on the initialization
command to perform an initialization operation. Then they will performed!
After about 20 minutes of power on (wait for temperature and light source is stable),
the instrument can be tested.
When the instrument is installed for the first time or placed for a long time, please
perform the water pipe exhaust 20 times before use to ensure that the air bubbles in
the pipe are drained.
32
Auto Chemistry Analyzer User Manual
Run the desktop icon . When you run the software for the first time, it will
tip On line setup, as shown in figure 3-1. Click "OK", the dialog box shown in figure 3-2 will pop
up and then click the software configuration. If the following figure 3-3 dialog box appears,
the software
configuration is successful.
Run the desktop icon again, open the software, enter the login interface,
as shown in figure 3-4, input the initial user name: 1000, the initial password: 1000, click the
button and the
software opens.
33
Auto Chemistry Analyzer User Manual
Click "OK" and enter the main interface, first is the instrument initialized operation. After the initial
operation completed, run three times exhaust operations. When the exhaust operation is completed,
the instrument enters standby state. If you choose "Cancel", the software will directly into the main
interface, and the instrument into the standby state. The main interface of the software is shown in
figure 3-6.
Software interface consists of several major parts: the menu bar, toolbar, workspace, the status bar,
as shown in figure 3-7.
34
Auto Chemistry Analyzer User Manual
Workspace
Menu
bar
1) Menu bar
Menu Bar: Displays the function menu of the instrument software. Click the menu bar to display the
submenu, Click the submenu to display the software operation interface
The menu bar is located on the left side of the main interface, display the function menu for the
operation of the software, the menu bar icons and functions are shown in sheet 3-1.
Sheet 3-1 Menu bar icons and functions
Enter the sample information that needs to be tested and carry out
1
calibration and QC entry for new items, and re-test the samples.
Test curve, test result query, QC result query, report printing, data
4
maintenance, export.
Sample tray, reagent tray, reaction tray status monitoring during the
7
test.
35
Auto Chemistry Analyzer User Manual
2) Status bar
Located in the top right of the main interface, real-time display the current user, display computer
system time, as well as current working state or testing process. Display as shown in figure 3-8:
: Display the current operator's name, the user can be added and deleted in "User
setting".
: Display the current computer system time.
4) Workspace
According to user's choice, it can appear the corresponding functions in the interface window. Such
as in the menu bar, click on the "project input" button, the display as shown in figure 3-10 sample
input software interface.
36
Auto Chemistry Analyzer User Manual
5) Version information
Software version shows in the status bar, as figure 3-11 shown. The way of
viewing the hypogynous machine version: click , workspace shows the user setting
interface, click "About" button, as figure 3-11 shown, click "Get Drive" to get the corresponding
driver board version number, click "Get Version" to get the machine number and the version number
of the hypogynous machine.
37
Auto Chemistry Analyzer User Manual
Specific Item
Edit special item parameters
Parameters’ Setting
Data Processing Export data Export test data and other information
Sample Status Display the status, location, and type of the sample
38
Auto Chemistry Analyzer User Manual
Operate software by clicking mouse. Input number&word by keyboard entry (Input method
according to Windows system, can transform by Shift+Ctrl button).
Forms are divided into 2 kinds: modal form and non-modal form.
Modal form: Other form cannot be operate before close the form. As shown in figure 3-13.
39
Auto Chemistry Analyzer User Manual
Radio box: Choosing only one function among several functions, as shown in figure 3-16:
40
Auto Chemistry Analyzer User Manual
Check boxes: Choosing two or more functions at the same time. As shown in figure 3-17:
41
Auto Chemistry Analyzer User Manual
Test finished
Water blank
Reagent R1
Reagent R2
Cleaning
Initialize
Sample
Air out
Start
… …
End
Stir
Stir
First Circle Second Circle Sixteenth Circle
There are three washing probes, including top two probes are double probe, needle A
is responsible for washing inlet, needle B is responsible for washing outlet, the last
one probe only responsible for washing outlet.
The washing steps of the cuvettes are:
First reaction tray move to the position, cleaning arm vertically downward, washing
probe B start to draw the liquid. Then probe A add water, two probe work at the same
time when washing arm rise, then arm stops for a short period at the top of the
cuvettes, the probe A&B stops working, the washing arm back to its original position.
42
Auto Chemistry Analyzer User Manual
This analyzer uses the whole absorbance testing mode, when the reaction tray is
rotated, the absorbance of pure water or reaction solution in the cuvettes is determined
continuous.
The reaction tray rotates in 1 laps for about 18 seconds, and the absorbance value of
48 cuvettes are measured when the optical axis passed. 36 times (36 metering points)
were determined for each cuvettes.
Light emitted from the light source is focused by a lens, first pass the cuvettes, then to
the photoelectric receiver transformed into electrical signals. After the two-stage
amplification and Log conversion obtained after the absorbance or absorbance rate of
change. Single/dual wavelength test available.
Note 1: When the main power supply access, the analysis of the power supply and
cooling system power respectively through the independent switch control. Even
though the analysis department does not work, the cooling system can also be in the
working state.
Note 2: The usage, storage and handling of reagents need to follow the instructions
strictly
43
Auto Chemistry Analyzer User Manual
Attention: After the machine electricity and into the standby mode, each
position is not at the initial position, entry the software and click initialization,
then can be corresponding to the initial position.
The relative position of the probe and cuvette is: Dispensing probe 1 position, mixing
probe 32 position, cleaning probe 12 positions.
This analyzer uses the whole absorbance testing mode, the absorbance of pure water
or reaction solution in the cuvettes is determined continuous.
44
Auto Chemistry Analyzer User Manual
The reaction tray rotates in 1 laps for about 18 seconds, 36 times (36 metering points)
were determined for each cuvettes.
Light emitted from the light source is focused by a lens, first pass the cuvettes, then to
the photoelectric receiver transformed into electrical signals. After the two-stage
amplification and Log conversion obtained after the absorbance or absorbance rate of
change
Among them:
A - Absorbance is absorbed by solution.
T - The Ratio of transmitted light intensity and incident light intensity, transmittance.
I0 - Intensity of incident light.
It - Intensity of transmitted light.
ε - Molar extinction coefficient (ml × mmol-1 × cm-1).
c - Molar concentration of the solution (mmol/ ml).
b - Solution layer thickness (cm).
Solution layer thickness (b), that is, the optical path is fixed and known. Solution
molar extinction coefficient (ε) is the wavelength, the solution and the solution
temperature correlation coefficient, when the guaranteed solution temperature
stability, in its single wavelength, the concentration of the solution and absorbance
has linear relationship ( ε is given directly in manufacturers reagent kit).
When the test sample is a homogeneous distribution of the solution, it was limited to
the role of incident monochromatic light absorption process, and fluorescence,
scattering and photochemical phenomenon does not occur, and in the absorption
45
Auto Chemistry Analyzer User Manual
Note 2: For reading points not used, be sure to enter "0." in the software.
Absorbance
measure range
B
R1 S R2 C
Time
AX AL + AL −1 + ... + AM +1
= + AM
(L − M )
(c) oncentration calculation
C X = {K ( AX − A1 )} IFA + IFB
46
Auto Chemistry Analyzer User Manual
R2
Cup blank
B1
S
R1
B1
Time
47
Auto Chemistry Analyzer User Manual
S+
Rj j =1
k= b
S+ Ri
i =1
CX = K ( AX − B ) + C1 IFA + IFB
B is water blank, R1~R2 is add location of reagent, Ax is the absorbance difference
B1
Time
48
Auto Chemistry Analyzer User Manual
( AM 2+ AM −1 ) ( A + AL−1 )
− 2 L
AX =
t
Among them:
t: Interval of L and M metering points (minutes).
(c) alculation of Concentration
CX = K ( AX − B ) + C1 IFA + IFB
B is cup blank, R1~R2 is add position of reagent, Ax is the average metering point
4. Rate method
Measuring method according to the rate of change in absorbance per minute to
calculate the concentration or activity values between the two metering points, known
as rate method, the reaction curve show as below:
49
Auto Chemistry Analyzer User Manual
Absorbance
R1 S
R2
Cup blank
B1
AL AM
Reaction boundary level
B1
Time
Figure 5-4 Rate method reaction curve
CX = K ( AX − B ) + C1 IFA + IFB
B is water blank, R1~R2 is additional reagent position, A(M − is the average
L)
50
Auto Chemistry Analyzer User Manual
Absorbance
Ax
B STD(1)
C1 Cx Concentration
51
Auto Chemistry Analyzer User Manual
Absorbance
A2 STD(2)
Ax Sample
B STD(1)
C1 Cx C2 Concentration
52
Auto Chemistry Analyzer User Manual
A10
A9
A8
A7
A6
A5
A4
A3
Ax
A2
B
C1 C2 Cx C3 C4 C5 C6 C7 C8 C9 C10 Concentration
X = − Cr ) (Ai − A)
(Cri
i=1
53
Auto Chemistry Analyzer User Manual
Y=
i =1
(Cri − Cr ) 2
A= Ai /n
i =1
n
Cr = Cri /n
i =1
A1, A2 is the two measured values of calibration solution (1), n is the number of
calibration solution N×2, Cri is the concentration calibration solution (i) .
CX = K ( AX − B ) + C1
IFA + IFB
Cx is the sample concentration to be test, Ax is the sample absorbance or absorbance
change per minute, IFA and IFB is the instrument constant, are represented by the
slope and intercept.
(d) Applicable analysis method
One endpoint method, fixed time method, two endpoint method, rate method.
AN STD(N)
Ax Sample
A3 STD(3)
A2 STD(2)
B
STD(1)
C1 C2 C3 Cx CN Concentration
54
Auto Chemistry Analyzer User Manual
Calibration model:[Logit-log4P]
Calibration point:[4~6](Calibration fluid quantity)
minute.
K: value of the absorbance of calibration fluid 1 or change rate minus B
a, b: Approximation of coefficient, calculate automatically.
S1ABS, K, a, b displayed in the calibration result interface.
(c) alculation of concentration
CX = (C + C1 ) IFA + IFB
K
AX = B
1+
+
aC b
1 K − ( AX − B)
C= b
a AX − B
Cx is concentration of testing sample, C1 is blank concentration. Ax is absorbance of
sample of the change value per minute. K is approximation of coefficient. When Cx is
closer to ∞, Ax is closer to B. If K<0, Ax≤B+K or K>0, Ax≥B+K, so C1=0,
(Aij-Ai’) is d-value of absorbance between Ai’ from fitting equation and Aij from
testing or the d-value between Aij and A12. Every calibration fluid test twice, and the
maximum value of Aij is 12.
(e) Applicable methods
One endpoint method, fixed time method, two endpoint method, rate method.
55
Auto Chemistry Analyzer User Manual
AN
A3
Ax
A2
C1 C2 Cx C3 CN Concentration
minute.
K, a, b, c: Approximation of coefficient, calculate automatically.
S1ABS, K, a, b, c displayed in the calibration result interface.
(c) alculation of concentration
AX − =0
a+ b lnC + c C B
− ln
K − ( AX − B )
Get C from Newton Approximation
CX = (C + C1) IFA + IFB
K
AX = B
1+ exp ( −a − b l nC − c C)
+
56
Auto Chemistry Analyzer User Manual
(Aij-Ai’) is d-value of absorbance between Ai’ from fitting equation and Aij from
testing or the d-value between Aij and A12. Every calibration fluid test twice, and the
maximum value of Aij is 12.
(e) Applicable methods
One endpoint method, fixed time method, two endpoint method, rate method.
Absorbance
STD(6)
STD(2)
STD(1)
C1 C2 CA C6 Concentration
57
Auto Chemistry Analyzer User Manual
C2 − C 1
K=
A2 − B
B: Absorbance or change rate of Calibration fluid 1.
A2: Absorbance or change rate of Calibration fluid 2.
C1: Concentration of Calibration fluid 1.
C2: Concentration of Calibration fluid 2.
Calculate K2, K3, K4, K5 with same method
(c) alculation of concentration
CX = KN ( AX − AN ) + CN IFA + IFB
(d) Applicable methods
One endpoint method, fixed time method, two endpoint method, rate method.
AN
Ax
A3
A2
C1 C2 C3 Cx CN Concentration
58
Auto Chemistry Analyzer User Manual
(Aij-Ai’) is d-value of absorbance between Ai’ from fitting equation and Aij from
testing or the d-value between Aij and A12. Every calibration fluid test twice, and the
maximum value of Aij is 12.
(e) Applicable methods
One endpoint method, fixed time method, two endpoint method, rate method.
59
Auto Chemistry Analyzer User Manual
The reagent blank check is used to test the absorbance of the cuvette containing the
reagent. The reagent blank test value can be used to correct the absorbance of the
sample, resulting in more accurate test results.
3. Sample blank testing
The sample blank is the concentration of a sample used as a zero reference in a test
procedure. The sample blank can counteract the positive error caused by the color or
turbidity of the sample itself before the addition of the test reagent. Due to hemolysis,
lipidemia, jaundice, etc., the sample’s absorbance will affect the testing results. So,
measuring the absorbance of sample can remove the impacts. Measurement method is
based on the normal test reagents and sample size, changing the reagent into distilled
water or saline, and then testing.
60
Auto Chemistry Analyzer User Manual
The brief sequence of operation is shown in sheet 6-1, and the details in "6.2 detailed
operations."
Sheet 6-1 Schematic instruments operation sequence sheet.
Procedure
Operation steps Operation
Form/key
The power should be checked before
1. The pre-test check ——
the test.
Connect the water inlet, outlet and
waste water pipes, and turned on the
2. Connect the purified water Log power of pure water device and
and power, analyzer analyzer.
Login Software system Input operator's user name and
password in the login form of the
system software.
61
Auto Chemistry Analyzer User Manual
Procedure
Operation steps Operation
Form/key
5. Prepare reagent (reagent
information) Confirm the residual reagent volume and
(1) Confirm of the residual Reagent testing times of each items of reagent.
reagent volume information Put the reagent that testing required o the
(2) Colorimetric analysis proper position.
items
Confirm the project name required
Calibration
6. Set calibration and QC calibration analysis.
information
items Confirm the project name required QC
QC
tests.
To carry out a single or batch of routine
sample entry and a single emergency
7.Sample entry and testing Sample entry
sample entry, patient information editing,
information modification and deletion.
8. Start testing
(1) Preparation of sample, Put sample, calibration liquid and QC
calibration solution and QC liquid on the sample tray.
Start condition
solution. Select test conditions, then click "start
(2) Set starting conditions test".
and send test instructions.
System
9. Testing process monitor
Test status of real time monitoring
(1) System monitoring Add pause /
instruments.
(2) Sample adding continue
Edit the sample in the test process and
(3) Emergency stop Emergency
click start testing.
(4) Additional sample stop
Sample entry
Query, modify, delete, review and print the
10. Confirmation of
Testing result test results.
the testing result (result
Sample response curve.
data)
62
Auto Chemistry Analyzer User Manual
Procedure
Operation steps Operation
Form/key
Confirm the condition for re-examining
11. Determination of re-test
Testing result items in the form of "adding the re-
samples
examine items".
12. The analysis finished Confirm, check and print the result of re-
(1) The result of the re testing items.
confirmation Regular database backup, suggest that the
Testing result
(2) Database backup normal case of a week.
System
(3) System shutdown Cut off the power supply for instrument,
management
(4) Cut off the power supply computer and pure water device.
(5) Preparation before the Clean up the analyzer to prepare for next
next work use.
63
Auto Chemistry Analyzer User Manual
(a) Power on the Auto Chemical Analyzer. The switch panel is located in the right
side of the analysis apparatus. When a reagent on the reagent tray the main power
switch should be in the open state, to ensure the normal operation of the cooling
system. Power switch is power supply for analysis apparatus.
(b) Login the Auto Chemical Analyzer application software, if you want to test, you
should first execute online instruction.
In the light case, adjust the instrument and amplifier voltage, A/D readings is about
48000, it is required that the range of delta OD is about ± 0.0015.
64
Auto Chemistry Analyzer User Manual
Click the software Toolbar " " menu, open the temperature display, observe the
(37±0.1)℃ for approximately 20 minutes. And the stability of the light source
also takes a few minutes, so after the instrument is turned on, you can view the
form of information, the entry of the parameter of the item, the alarm
information, etc., and after
temperature stability, the sample can be tested.
Before the test, first of all to carry out the analysis of the item to add, and then carry
out the item parameter, calibration parameter, QC parameter setting.
Analysis the parameter setting, using methods, precautions and save of used reagents,
calibration liquid, QC solution please refer to the instructions or to the relevant
manufacturers, vendors consulting.
6.2.5.1 Item parameter confirmation
In the menu bar " ", click" ", enter the item parameter
interface, determine the item parameter, as shown in figure 6-3.
65
Auto Chemistry Analyzer User Manual
The setting of reagent item parameter is according to the reagent insert. Some
parameters are needed for testing, such as: test item name, analysis method, the
main/auxiliary reading points, the main/sub-wavelength, sample volume, reagent
volume. Although some of the parameters do not affect the results, but important for
measuring the reliability of the test results, such as the normal range, absorbance limit,
it is recommended that the operator one by one input parameter correctly and then test.
6.2.5.2 Setting of item portfolio
Item portfolio is the combination of the relevant items together, such as a full set of
liver function, full set of renal function, just click on the name of the portfolio item
can complete a number of item registration function to facilitate the sample
registration of the fast entry.
Click the "special items" button on the toolbar, enter the "special item parameter
setting" form and click the "increase" option at the right side of the portfolio item, fill
in the relevant items as needed.
As shown in figure 6-4: Click the first increase, fill in the "item number" according to
the serial number, fill in "item short", "full Chinese name", "full English name"
according to the need to select the item in "Optional Items", and click" ".
Added to "Selected Items", click on" " cancel related project operation,
after choose you click "Save" to complete the operation.
66
Auto Chemistry Analyzer User Manual
67
Auto Chemistry Analyzer User Manual
68
Auto Chemistry Analyzer User Manual
If the reagent input sequence is wrong, use the "Delete" key to delete the item and re-
scan.
2. Open reagent information entry
In the "reagent information" form, click the " ", the instrument
automatic detection reagent remaining volume, the test times of rest reagents, the
results of the test information is displayed in the "reagent information" list box.
During the testing process, as long as the reagent is used to absorb the reagent, the
residual amount of the reagent is detected, and the remaining amount of the reagent
and the remaining test number are updated.
Factor — CK-MB
70
Auto Chemistry Analyzer User Manual
6.2.8 Login QC
Click the menu bar " ", the "Calibration Par" operation interface is
displayed, click on to select the lower right side of the window " ", the
"QC Parameter" interface is displayed, as shown in figure 6-10.
Add a new QC item steps: First click on the "QC parameter" window "
", fill in the lot number according to the lot number of the QC manual,
as shown in figure 6-11,select the QC item, click on " " it is selected,
71
Auto Chemistry Analyzer User Manual
For example: Select the newly QC 123, the need items for QC on the right side,
according to the QC 123 to add the "Target Value" and "Standard Deviation", modify
the window to complete the operation.
Click the " " button in the toolbar, enter the "entry" form, as
shown in the figure below:
72
Auto Chemistry Analyzer User Manual
Single sample entry: Click , and choose test items in the left column, click "
" to move test item to "selected test items" column, the sample number is
generally from the beginning of 1, followed by accumulation, a number can not be
repeated, and then fill in the information of patients "name", "gender", "age", edit
"inspection department" and "sample type", "submission of doctors' information,
the fill in the quantity and starting position number that to be tested,
sample position number imputing is from 1 to 37, from 37 on, sample position
number plus 1, increasing gradually start from 1, save the selected items to be tested,
complete the operation.
Project portfolio login: If in the process of sample inputting, the project selected is the
combination that has been set up, you can directly select in the combination bar that
in the left side of interface, click directly add to the "selected project
determination" column.
You can edit the information you have input. In the landing list column, you can
choose one sample number to choose, which display the patient information and
testing items information, press " " or " " button, to edit, delete the
73
Auto Chemistry Analyzer User Manual
a) b)
Figure 6-14 Start sample testing
Notes: When the button in the start interface displays “ ”, the mouse is
placed on the button to display the corresponding prompt, as shown in figure 6-14 b),
Check the
message prompt. After completion, click "Start" again in the main interface. Conduct
sample testing.
During the test, the status of the sample testing is displayed in the status bar, as shown
in figure 6-15a).
During the sample testing, click on the menu bar to perform real-time
monitoring of the instrument sample tray, reagent tray, reaction tray status, as shown
in figure 6-15 b).
74
Auto Chemistry Analyzer User Manual
b) Status monitor
Figure 6-15 Sample test
Sample tray monitoring: Click "sample trays status" under the "monitor" form, as
shown in figure 6-16, wherein the sample tray with different colors to distinguish the
status of samples.
Reagent tray monitoring: Click the "reagent tray status" in the "monitor" form,
respectively, the reagent tray test status will display in the left of the form, as shown
in figure 6-17:
75
Auto Chemistry Analyzer User Manual
Reagent tray by serial number indicates that the reagent placed number, when the
analyzer starts testing, reagent&sample probe after taking the reagent, the software
will automatically calculate the residual of the reagent, and the reagent information
displayed on the inside. In the reagent tray: white indicates empty, yellow indicates
less reagent, green indicates that the reagent sufficient, red means no reagent.
The reaction tray detection: The reaction tray in real time display the status of the
reaction at the time of sample testing. As shown in figure 6-18, wherein the different
colors represent the cuvette in different working conditions.
76
Auto Chemistry Analyzer User Manual
Sample test in progress, click push button in the editor toolbar " ", "
The test result values are displayed in the "Print" interface of "Data processing", and
in the "Test Result Query" interface of "User Setting", can query the test results. As
shown in figure 6-19 and figure 6-20.
77
Auto Chemistry Analyzer User Manual
As shown in the figure, in the "Print" or "Test Result Query", select "Date", "Name",
"Patient Information", can query the history test results.
6.2.11 ta statistics
In "User Setting", you can use "Date" function to make a statistical query for the
historical test data. Query process, choose the corresponding date in the "date", then,
click "Statistics", get the test result.
Data statistics include test statistics, workload statistics and cost statistics, as shown
below.
a) Test statistics
78
Auto Chemistry Analyzer User Manual
b) Workload statistics
c ) Expense statistics
Figure 6-21 Test statistics
In the "Sample Input" interface, select "retest" as shown in figure 6-22 a), and select
re-test the sample number, click "Edit", and select re-test items. Then, click "Save",
Complete the retesting sample or item input.
Or in the "Print" interface of "Data Processing", click the "Save retest", and select the
retested sample or items, as shown in figure 6-22 b).
79
Auto Chemistry Analyzer User Manual
a) Sample retest
If the sample is testing, the retest sample will automatically enter the test item. If you
have finished the sample detection, you need to click the start button" ",and
perform sample retest and detection tasks.When the sample retest is not performed, if
the sample retest needs to be cancelled, click "reset retest".
Note : Only the test data of the day, which can be re-tested.
Test finished, if there are no objection for the data, select "Washing&&background"
in the menu bar "Maintenance", and in the "Washing&&background" interface ,
select "Cleaning" function ,turn off the power of biochemical analyzer after cleaning.
80
Auto Chemistry Analyzer User Manual
Biochemical reagents needed be save separately, take out the extra reagent and use
reagent bottle caps cover it, then put it into the medical refrigerator . Remove the
already tested sample, calibration and control solution where in the sample tray. If
you use waste container, dispose of the liquid in a safety way. Detecting the
equipment table is dirty or not, and clean the table.
81
Auto Chemistry Analyzer User Manual
In software interface, click the , then, click the "Calibration Par", you can
check and change the calibration parameter.
7.1 Calibration
Click "Item Setting", then click the "Calibration Par" , input the calibrator parameter.
After we finished the input reagents parameter in this interface, select test items
which need be calibration. Then click on the right " ". After input the correct
parameter, in "Calibration type", "Standard counts", "Repeat counts", "Sample type",
"Concentration", "Cup No.", finally click "Save". "Concentration" and "Cup No",
need according to the calibrator and the calibrators location number in reagents tray,
then click "Edit".
", and select the need calibrator test items, click the . Then, click "Save" ,
saved successfully. As shown in figure 7-2.
82
Auto Chemistry Analyzer User Manual
After the calibration items input is completed,click the Start button " ", and
then do calibration test, and the status bar displays the testing process, as shown in
figure
7-3.
83
Auto Chemistry Analyzer User Manual
In Curves interface, at "Test Date", select the calibration date and the need queried
items. Then, we can get the calibration curve.
As shown in figure 7-4, in response curve interface, select the "calibration" column
query calibration curve, select the right side of the interface "test date" calibration
time, Select the items you want to query on the right side of the interface ,the
calibration curve can be directly inquired according to the specific time of calibration.
2. Calibration result query
Under "Item setting" window, left-click to enter "Calibration Par" screen, select the
desired view of the direct results of the item, as shown in figure 7-5.
As shown in figure 7-5, in "Calibration Par" screen, first in the "item list" column
select the name of the item you want to view, and then select the specific calibration
time on the right side of the interface to query the calibration results.
84
Auto Chemistry Analyzer User Manual
7.2 QC
In the "Data Processing" menu, open the "QC Result Query" window and select the
corresponding QC rules in the QC Rules drop-down list, as follows:
85
Auto Chemistry Analyzer User Manual
1. Westgard multi-rule QC
According to the test needs, select Westgard multi-rule QC, pitch on
The operator according to the test needs, by clicking the way to select the appropriate
QC rules, click the "Save" button to return to the QC results query window, while the
QC rules to save the setting. After setting, in the "day QC" and "month QC" window
will be set in accordance with the rules of QC data out of control analysis.
According to Westgard multi-rule judgment benchmark for the measured QC results,
perform out of control analysis. figure 7-8 shows the Westgard multi-rule logic
diagram.
QC data
No
1 2S In control
Yes No
No No No No 10 -
1 3S 2 2S R 4S 4 1S
x
Yes Yes Yes Yes Yes
Out of control
86
Auto Chemistry Analyzer User Manual
87
Auto Chemistry Analyzer User Manual
(5) Calculate, accumulate and clear when accumulating and just changing the symbol
(positive or negative) and accumulate the sum for the subsequent data points.
(6) When the accumulated and over control limit H (greater than the upper limit or
less than the lower limit), then judged to be out of control.
3. Twin-Plot (two-dimensional) QC
Twin-Plot (two-dimensional) chart to the QC X test results for the horizontal axis, QC
Y test results for the vertical axis, will participate in the joint test results for the
vertical axis, will participate in joint judgment of the two QC, and the same batch of
test results are plotted to determine system errors and random errors.
7.2.2 QC test
After the QC information is registered, click "QC parameter" in the menu item "Item
parameter", click "Add" in the QC area on the right side of the interface, select the lot
number in the "Control Lot", "QC number" in the input control of the sample placed
in the sample tray number, in the "QC test items" need to select the QC of the item,
"QC List" shows the number of editing QC test, click "Save" to complete the
operation. As shown in figure 7-9.
After the QC item input is completed, click the start button to test the QC item.
QC test is completed, the instrument automatically calculate the measured QC target
(average), standard deviation, coefficient of variation and other data.
N
Xi
i =1
Target( X ): N
88
Auto Chemistry Analyzer User Manual
( Xi − average )2
i =1
Standard deviation(SD): N −1
SD
100%
average
Coefficient of variation(CV%):
Deviation: Xi-(average)
Deviation
100%
%Error: average
Among them:
N is the number of measurements, Xi is the test result.
1. QC curve query
As shown in the figure, in the "QC results query" interface, first, select the item to be
inquired and the QC batch number, and then according to the QC test time to find the
QC results. Choose different QC rules and view the corresponding QC trend, as
shown in figure 7-11.
89
Auto Chemistry Analyzer User Manual
90
Auto Chemistry Analyzer User Manual
At the same time the system supports manual entry of "new QC value", can do the out
of control analysis for the QC results , click "Add", enter the new QC value, click
"Save" to display the QC results status. Click "Out of control analysis" to display out
of control results.
3. QC chart printing
In the QC results query, click " ", and then print daily QC or
monthly QC chart.
4. QC report printing
In the QC results query, click " ", and then print daily QC or
91
Auto Chemistry Analyzer User Manual
The calibration and QC test are completed, the query results of QC in the QC target
range, then the scaling factor(Ki)can be directly used for sample testing. If the QC
results of deviation from the target range, it is necessary to control the scaling factor
of QC in the target range.
The method of correcting factor is calculated according to the QC value and target
value, and the formula is as follows(Ki:Old factor,Ki+1:New factor):
Formula:
𝐾𝑖× Target value
QC value
𝐾𝑖+1
92
Auto Chemistry Analyzer User Manual
After adjusting the factor, we must re-QC, and determine the value of QC in the range
93
Auto Chemistry Analyzer User Manual
of target value, to determine the correct adjustment of the factor, in order to carry out
the sample test.
Calibration
No Factor
QC
correction
Yes
Or
QC value in the
range of target value QC control
value off target Or
Sample test
94
Auto Chemistry Analyzer User Manual
Click the "User Setting", can add or delete the account, change user password. As
shown in figure 8-1.
As shown in figure 8-1, in the user code, user name, respectively, enter the
corresponding name, and select the user level, click "Add" to increase a system user.
If the advanced user login software, you can select a user left button, click "Delete" to
delete the current user.
Click the "Change User Password", can change the being used account’s password.
95
Auto Chemistry Analyzer User Manual
8.5 About
"About" is a software version of the introduction.
96
Auto Chemistry Analyzer User Manual
Chapter 9 Maintenance
9.1 Prepare before system maintenance
To ensure the precision and accuracy of the auto chemistry analyzer, the operator
must obey "Auto Chemistry Analyzer User Manual" for the operation and required
maintenance. Thus the instrument will provide reliable results and ensure longer
service life of the instrument.
In the daily operation and maintenance of the system, use deionized water, electrical
conductivity is less than 1 μs/cm. When using pure water equipment, please do not
forget the regular maintenance and inspection of water purification equipment. Please
refer to the manual of the pure water device, or contact the manufacturer or distributor
of the water purifier.
9.2.1 Initialization
Click the software Toolbar "initialization", the instrument automatically back to zero
position. In the reset process, do not allow emergency stop. Other operations can be
carried out after the initialization is completed.
96
Auto Chemistry Analyzer User Manual
Attention: When the instrument is powered into the standby state, each
position is not in the zero position, and it is necessary to enter the software click
initialization command to perform an initialization operation to return to zero!
Attention:
1. Vertical cleaning position of the dispensing probe: Thinner part of the needle
should be completely in the cleaning cup. Dispensing probe vertical depth of cuvette:
Point of probe is 30 steps distance to the bottom of the cuvette. Reagent position
depth is 20 steps distance to the bottom of the reagent bottle. Sample position depth is
10 steps distance to the bottom of sample cup.
2. Cleaning position of the stirrer should be enough deep but not touch the bottom of
the cleaning cup. Reaction position of stirrer should be 20 steps distance to the bottom
of the cuvette.
3. Cleaning arm position adjustment should make sure all the probe except the one
with dry unit, just reach the bottom of the cuvettes. The last needle should 2mm lower
than other probes.
97
Auto Chemistry Analyzer User Manual
forth to see the brightest spot on the track cup campaign. Spot shape around a long
5mm, rectangular spot width of about 1.5mm, If there is a problem spot brightness,
check halogen installation location is correct (This position is generally not adjusted).
9.2.3.2 ne key gain calibration
One-key gain calibration enables automatic calibration of A/D readings, as follows:
In the "Maintenance" form, open the "Instrument Detection" interface, click the
"Parameter Setting" button, enter the password in the password box: 00000000, you
can jump to a key gain calibration interface, enter the target AD value of 48000, click
"One Key Auto Gain" button, you can achieve auto adjustment, as shown below:
Attention: In order to the calibration results, please take out all cuvettes
before one key auto gain calibration.
9.2.3.3 AD stability adjustment
After adjusting the AD board voltage. Click software toolbar "Instrument Check"
button, enter "Instrument Check" interface. If the optical system ΔOD has larger
changes, Please adjust as follows:
(a) Check the lamp is uniform, the spot size is suitable or not.
(b) 12V input voltage is stable, if separate power supply or not.
(c) Grounding line is good. Requirements to external and internal ground only on a
single-point ground power supply. External power supply on the ground must be
connected.
98
Auto Chemistry Analyzer User Manual
In the "Absorbance Test" interface, click "Test" button, the reaction tray rotates two
circles, read the absorbance of each cuvette. Finished reading, corresponding
absorbance value with cuvette number display in the interface. Pull the scroll bar to
read the absorbance value with all cuvettes.
99
Auto Chemistry Analyzer User Manual
As shown in figure 9-3, cleaning needle includes two injection pipe and three suction
pipe, 2 injection pipe corresponding solenoid valve V5, V4, 3 suction pipe
respectively miniature diaphragm pump P3, P4, P5.
Miniature diaphragm pump P2 connect dispensing piston arm. Miniature diaphragm
pumps P1 connect cleaning needle injection pipe, dispensing probe cleaning cup and
stirrer cleaning cup.
L frame has three waterways adapter, one small linker are water inlet. The installation
of water pipes should be as short as possible, and install water fall head. Ensure the
machine waterway is normal.
Attention: When installation of water pipes make sure the smooth flow of
waste water outlet.
Each cleaning needle connected with a corresponding waterway through a
polyethylene tube.
When added head or pumping head is abnormal, sequentially check the pipe joints are
connected solenoid valves and micro diaphragm pump.
When finished the installation of all parts and water system, please power on the
machine and check the conditions:
(1) The rotation of reaction tray, reagent&sample tray.
(2) Moving conditions of mechanical arms left&right, and up&down.
(3) Working conditions of syringe pump, solenoid vale and diaphragm pump.
(4) Reagent&sample tray cooling condition .
As below figure 9-4 (a) (b) shows: “Instrument power supply port” and “Internet port”
100
Auto Chemistry Analyzer User Manual
Attention: Before connect Internet wire, please install Internet device drive.
If instrument is alarming when power on, please follow the alarm indicator and check
if waste water is full, less purified water. Change waste container, adding purified
water.
a)
101
Auto Chemistry Analyzer User Manual
b)
Figure 9-5 Water Pipe
Open the software installation folder, find " " and open it.
Find [print page], set the print paper typeA4, A5 or B6. For example the print paper
If the auto biochemistry analyzer appears below interface figure 9-6, please use below
"Repair Tool" in the "bin" folder. The "Repair Tool" icon is shown as figure 9-7.
Operation procedure as below: Double click the repair tool, appear figure 9-8
operation interface, chose "OK" button, appears figure 9-9 interface. Click confirm to
102
Auto Chemistry Analyzer User Manual
complete the repair. When finish repairing, will indicate that the repair is completed,
please re login the software.
Click login button, appears below interface, software and instrument not match.
Please click cancel to close the interface.
Click "login" button and at the same time press the "Ctrl" button. Enter software
operation interface. Click "Ctrl + Alt + R", then click "OK" as figure 9-11 shows and
click "OK" button to exist the software.
Click login again. Appears figure 9-12 interface. Click “software configuration”,
appears figure 9-13 interface, that means the software configuration is successful.
103
Auto Chemistry Analyzer User Manual
When the software show database operation failed, please use the database repair tool
"access repair wen test" as the figure 9-14 shows, in the "bin" sub folder of the
software installation folder. Open this tool with double click, show the interface as
figure 9-15, click the "Repair" button, waiting interface as figure 9-16, finish interface
as figure 9-17.
104
Auto Chemistry Analyzer User Manual
Regular clean, check and replace part, please see sheet 9-2 (Supposed using the
instrument 5 hours per day):
(○: Regular clean, check ●: Regular replace, adding)
Sheet 9-2 Regular replacement of accessories
Period
Needs Yearly Dai- Tim- Wee- Mon- Every Every Yea
No Items for one consu- ly ely kly thly 3 6 -
time mption mont- mon- rly
hs ths
1 Sample cup ●
2 Dispensing ○
probe
3 Cleaning cups ○
of dispensing
105
Auto Chemistry Analyzer User Manual
Period
Needs Yearly Dai- Tim- Wee- Mon- Every Every Yea-
No Items for one consu- ly ely kly thly 3 6 rly
time mption mont- mon-
hs ths
probe and stirrer
4 Cuvettes 6 24 ○ ●
(Note (8units/group) groups groups
a)
5 Reaction tray ○
and reaction
tray heating
6 Halogen lamp 1 2 ●
7 Cleaning needle ○
8 Stirrer ○
9 Dispensing ●
pump
10 Water supply ○
filter net
11 Reagent tray ○
cooling system
12 Cooling fan ○
13 Purified water ○ ●
system
14 Waste water ○
exhaust
Note:
(a) Number in the sheet is maximum.
(b) Halogen lamp life is 2000 hours. To ensure the accuracy and precision, suggest
replace the lamp after using for 1500 hours.
106
Auto Chemistry Analyzer User Manual
(c) Instrument is suitable for wire printers, inkjet printers, and laser printers, the user
according to the printer select printer supplies.
(d) Execute save the blank weekly. Otherwise will indicate cell blank abnormal alarm.
(e) If conductivity of purified water is more than 1μs/cm, please replace water purifier
consumables.
Please prepare below accessories, in case any faults need replacement, please check
sheet 9-3:
Sheet 9-3 Regular accessories sheet
Suggest
Stock
No. Accessories Name Remarks
quantity
/Year
1 Halogen Lamp 6V 10W 1
(8 units/group×6
2 Cuvettes 24 group
group)
3 TPU tube 3.2mm×6.4mm 10m
Water supply filter
4 net, water supply Used for water tube 2
filter holder
Dispensing reagent
5 Dispensing probe 1
and sample
6 Stirrer Used for mixing 1
7 Cleaning needle Clean the cuvettes 1
Attention:
●Do not drop water, reagent to any mechanical or electrical parts, avoid damage to
the instrument.
107
Auto Chemistry Analyzer User Manual
●When operating, do not touch dispensing probe, stirrer and cleaning needle. Avoid
the risk of infection or injury.
●During operation, should take preventive measures, with protective gloves, wear
work clothes. Otherwise, it is possible to contact contaminated areas, contaminated
fluid and infected, or contact with corrosive liquids and damage the skin. If there is
contamination or corrosive liquid accidentally touching the body, please immediately
washing by water and disinfection.
9.5.3.1 Sample adding system maintenance
If internal and external of the probe is polluted, it is easy to attach serum, reagent and
water drops, also easy to cause the blockage of the syringe tube inside, thus affecting
the instrument test results. Therefore, it is necessary to regularly check and timely
cleaning.
1. Clean dispensing probe outside
(a) lose the total power switch of the instrument.
(b) Scrub the outer wall with cotton swab dipped in alcohol (During the process of
cleaning, do not bend the probe). As shown in figure 9-18.
(c) After start up again, dispensing probe automatically return to the position of the
reset.
2. Dispensing probe jam clean up
When the instrument probe blocking, or probes out of flow discontinuity and not
vertical, dripping water, should cleaning the blocked probe.
(a) lose the total power switch of the instrument.
108
Auto Chemistry Analyzer User Manual
(b) Remove the reagent&sample tray cover, revolve the dispensing probe by hand
under the arm, make it moved to the above reagent sample tray. As shown in figure 9-
19.
Warning: Do not hold dispensing probe cap top with hands to revolve the
probe. Faulty operation as shown in figure 9-20.
(c) Gently holding probe cap with one hand, move the cover to the side, then loose on
the other side, can take down the cap after lift, as shown in figure 9-21.
109
Auto Chemistry Analyzer User Manual
(d) Remove the liquid level signal lines, need to hold cross arm from below with
one hand, hold 2P terminal with other hand, remove the liquid level signal lines.
Note, cannot pull liquid level signal lines, to prevent the terminal loss, as shown in
figure 9- 22.
(e) Stack up absorbent cotton pad on the liquid level board, prevent water droplets on
the liquid level board when remove hoses, as shown in figure 9-23.
110
Auto Chemistry Analyzer User Manual
(f) Unplug the pipeline of one side of the dispensing probe. As shown in figure 9-24.
(g) Penetrate from the lower end of the probe, place a cotton cloth on the side of the
level plate, and put the needle (0.3mm/0.5mm) into the probe to clean up from the
bottom of the probe. As shown in figure 9-25.
111
Auto Chemistry Analyzer User Manual
(h) After clean up the probe with needle (0.3mm/0.5mm), screw the probe from the
light blocking plate to one side of the reagent arm/sample arm, and the probe is
moved to the cleaning cup. Extract 10mL of pure water by injection syringe. Put the
pure water in the syringe injection probe from the upper, water flows from
needlepoint place, all 10mL of water discharge, and the water flow unhindered, as
shown in figure 9-26.
(i) Install the probe on the cross arm in the reverse order, and installed water pipe,
line, etc.
3. The adjustment and confirmation of the position of sample needle
(a) Open the instrument of the total power switch, initialize the instrument.
(b) Click the menu bar of the "maintenance", open the instrument detection window,
in the test project list to find sample arm. As shown in figure 9-27.
112
Auto Chemistry Analyzer User Manual
(c) A check of dispensing probe horizontal position: Click on the sample position "Y",
dispensing probe turn left, when the dispensing probe stop at the upward side of
sample position of reagent sample tray, check whether the pinpoint of dispensing
probe is in the center of the sample cup, as shown in figure 9-28, then click the zero
position "Y", dispensing probe is placed back to the initial position.
Attention: If the sample with the tip of the needle is not in the center of the
sample cup, please contact maintenance personnel.
(d) Reagent outer ring level examination: Click the reagent outer ring "Y", dispensing
probe turn left, when the dispensing probe stop at the upward side of reagent outer
ring of reagent sample tray, check whether the pinpoint of dispensing probe is in the
center of the reagent position of reagent outer ring, then click the zero position "Y",
dispensing probe is placed back to the initial position.
Attention: If the sample with the tip of the needle is not in the center of the
reagent bottle of reagent outer ring, please contact maintenance personnel.
(e) Cuvette level examination: Click the cuvette position "Y", dispensing probe turn
right, when the dispensing probe stop at the upward side of cuvette of the reaction
tray, check whether the sample with the tip of the needle is in the center of the cuvette,
113
Auto Chemistry Analyzer User Manual
as shown in figure 9-29, then click the zero position "Y", dispensing probe is placed
back to the initial position.
反应杯
Cuvettes
The针的前端
front end of
dispensing probe
Attention: If the sample with the tip of the needle is not in the center of the
cuvette, please contact maintenance personnel.
(f) Sample vertical position examination: Click the sample position "Y", dispensing
probe turn left and stop at the top of the sample position, click the vertical position
"Y", the dispensing probe will continue to drop until the needle tip touches the bottom
of the sample cup, then click the vertical position "Y", the needle is lifted above the
sample position, then click the zero position "Y", dispensing probe turn right and
placed back to the initial position.
(g) Reagent outer ring vertical position examination: Click the reagent outer ring "Y",
dispensing probe turn left and stop at the top of the reagent outer ring, click the
vertical position "Y", the dispensing probe will continue to drop until the needle tip
touches the bottom of the reagent cup, then click the vertical position "Y", the needle
is lifted above the reagent outer ring, then click the zero position "Y", dispensing
probe turn right and placed back to the initial position.
(h) Cuvette vertical position examination: Click the cuvette position "Y", dispensing
probe turn right and stop at the top of the cuvette, click the vertical position "Y", the
dispensing probe will go down, put the part of the probe which coated with Teflon
into the cuvette, then click the vertical position "Y", the needle is lifted above the
cuvette, then click the zero position "Y", dispensing probe turn left and placed back to
the initial position.
114
Auto Chemistry Analyzer User Manual
(b) Then, about 100mL of pure water was poured into each of the washing tank to be
washed.
9.5.3.2 Reaction tray maintenance
If the cuvette or thermostat is contaminated, may result in inaccurate test results. In
addition, the cuvette will occur for a long time aging, so the cuvette should be cleaned
regularly, and check the cuvette absorbance, when the absorbance value is abnormal,
the cuvette should be replaced in time.
During the sample test, click on the menu bar , can realize real-time
monitoring of the reaction state of the instrument, if the dirty cup is detected during
the test, the
115
Auto Chemistry Analyzer User Manual
instrument will automatically skip the dirty cup and use the next cup. In the
monitoring, the cup will show red, as shown in figure 9-31.
shows if more than 1/3 cuvette in reaction tray is dirty, need to replace new one.
Below is the procedure:
(a) Turn off the instrument main power switch, remove the reaction tray.
(b) Wear protective gloves and unscrew the fastening screws, as shown in figure 9-32.
116
Auto Chemistry Analyzer User Manual
(d) Install the six new cuvettes in the reverse order onto the reaction tray.
Attention1: If the used cuvette is exposed to air for long period of time,
contaminants may hangs on the cuvette wall, So we should cover the reaction
tray cover promptly. In addition, if the emergency stop happened during the test,
should clean cuvettes which not be cleaned or rinse them with pure water to
avoid the reaction fluid remain in the cuvette for a long time.
117
Auto Chemistry Analyzer User Manual
(e) After cleaning the reaction tank, install the cuvette and cover the reaction tray
cover.
9.5.3.3 Maintenance of the light source lamp
Halogen-lamp inspection: If light source lamp aging, light energy will deviate from
test range of light. During the sample test, it won’t test correctly because of interfere.
In the "instrument detection" interface for non-blocking AD readings test, if less than
20,000, should replace the halogen lamp. To replace the halogen lamp, proceed as
follows:
When the halogen lamp reaches the service life, after entering the main interface of
the software, the “Halogen Life Limit to Limit” alarm will be displayed, that is, click
the red alarm button, and the alarm information window will be displayed as shown in
Figure 9-35. The halogen lamp should be replaced.
Figure 9-35
118
Auto Chemistry Analyzer User Manual
Operation is as follows:
(a) Prepare a new halogen lamp, as shown in figure 9-36:
Attention: Do not touch the surface of the halogen lamp, otherwise the
amount of light will be affected. If you find fingerprints and other stains on the
surface, you can wipe with alcohol gauze.
(b) Turn off the instrument's main power switch, and after about 30 minutes (wait the
lamp compartment to cool completely), perform the next operation to avoid burns.
(c) Unscrew the one screw on the plate 1 with a phillips screwdriver. Operation as
shown in figure 9-37.
Plate 1
119
Auto Chemistry Analyzer User Manual
Terminal
block
Fastening screw
(e) Remove the protective cover from the terminal block. Use a Phillips screwdriver
to unscrew the two fixing posts of the halogen lamp lead. Remove the white silicone
leads and cut the one tie strips with a slanting mouth to facilitate replacement of the
halogen lamp. As shown in figure 9-39.
Wire 1
Tie
Wire 2
(f) Unscrew the fastening screw , and remove the halogen lamp as shown in figure 9-
40.
120
Auto Chemistry Analyzer User Manual
(g) According to the above converse steps to replace the new halogen lamps, pay
attention to tighten the screws. Lead wire should not be loose or cocked.
(h) Turn on the power. After the instrument is in standby, check the spot to meet the
5mm size. On the “Instrument Check” interface, AD values should be higher than
40000 without the block light, and then it can be tested.
121
Auto Chemistry Analyzer User Manual
(c) With one 0.5mm needle penetrate from the bottom of nozzle to clean up the block,
as shown in figure 9-43.
122
Auto Chemistry Analyzer User Manual
123
Auto Chemistry Analyzer User Manual
124
Auto Chemistry Analyzer User Manual
(d) Loosen the two setscrews with a phillips screwdriver for a round, as shown in
figure 9-46.
(e) Wipe the tip of the new mixer probe with a cotton swab dampened with an
alkaline cleaning fluid, and wipe off the cleaning fluid on the surface of probe with a
cotton cloth dampened with pure water (Do not bend the stirrer during wiping).
(f) When installing a new mixer probe, insert the stirring needle into the root of the
motor shaft and fix it with M2 screws, as shown in figure 9-47.
125
Auto Chemistry Analyzer User Manual
Mixing arm
M2 screw Stirrer
See the location of the pin in relation to the cuvette in a top-down view, as shown in
figure 9-49:
Stirrer
Cuvettes
126
Auto Chemistry Analyzer User Manual
Note: If not in the center of the cuvette, please contact the service staff.
(j) Probe’s vertical position check: In the "Instrument Check" window, click the
cuvette position "Y", the probe will move to the top of the cuvette, then click the
vertical position "Y", the probe down. Click the vertical position "Y" again, the probe
lift. Finally click the zero position "Y", the probe will return to the top of the cleaning
cup.
9.5.3.6 Reagent&sample tray maintenance
The instrument reagent sample tray has reagents refrigerated storage reagent function.
Reagent sample tray is divided into sample position, reagent outer ring. Including 37
sample position, 28 reagent position, as shown in figure 9-50.
HiHgihgehsetst
reraegaegnetnt
ppooosssiiittiiiooonnn 50ml
reagent
20ml Reagent bottle
reagent bottle
bottle fixing
block
Figure 9-50 Reagent position
Attention: Reagent can not be too full to ensure that below the top of the
reagent location as above figure, otherwise it will affect the level detection.
1. Replace the reagent bottle
When reagent less than the remaining time, please replace the reagent bottle. 50mL
reagent bottle can only be placed directly in the reagent outer circle, 20mL reagent
bottle can be set with reagent bottle fixing block in the reagent outer circle. Place the
reagent bottle as follows:
(a) 0mL reagent bottle placed: bottle near the sample circle.
(b) 20mL reagent bottle placed in the reagent outer circle: The narrower side of bottle
near the sample circle, and need to be fixed (Fix the block in the box) as shown in
127
Auto Chemistry Analyzer User Manual
figure 9-51. After placing the reagent bottle, fix the reagent bottle with the fixing
piece and then insert the fixing block vertically into the bottom.
reagent bottle
fixing block
The narrower
side of bottle
Attention: Must put the reagent bottle first, and then put a fixed block, to
prevent friction damage to the reagent bottle bar code.
2. Reagent&sample tray clean
The sample tray will be contaminated with reagents, samples and dust for a long time
use, and condensation water will be generated during the cooling process, so it should
be cleaned once a day.
(a) emove the reagent sample tray cover.
(b) Unscrew the reagent tray handle counterclockwise to remove the reagent sample
tray.
Reagent&sample
tray handle
128
Auto Chemistry Analyzer User Manual
(c) ipe the inside of the reagent sample pan with a wet gauze and wipe the
condensate from the reagent sample pan to the condensate drain.
Condensate
drain
Starting up operation:
1. Check the pure and waste water bucket before the test.
2. Check the dispensing probe, whether or not blocked.
3. Check the cleaning cup and waste water pipe whether or not blocked.
Turn off operation:
1. Clean the cuvette, empty the sample tray to test the sample, the reagent covered
with the corresponding cap, then cold storage.
2. Treatment of waste water bucket.
3. Add pure water to the cuvette, then save the background.
4. Turn off the machine power supply, cover the machine cover.
The fault of the instrument can be divided into failure and alarm failure according to
different problems.
129
Auto Chemistry Analyzer User Manual
130
Auto Chemistry Analyzer User Manual
Fault
Main reason Solution
information
1. Bad contact of liquid level
1. Determine if the interface line of the
board
Abnormal liquid level board is in good contact
2. Instrument grounding
liquid level 2. Check if the ground is connected
problems
detection 3. Check if there is a large elec-
3. There is a large electro-
tromagnetic interference around
magnetic interference
1. Make sure the reagent is prepared and
the position is correct
2. Check if has impurity in sample
The absorbance of the 3. Check the reaction tank for water or
Absorbance
reaction liquid is beyond impurities
beyond
3.3Abs 4. Check the cuvette whether or not
cracks and scratches
5. Check whether the optical window is
clean or has water
1. Confirm the optocoupler under the
1. Reaction tray Unable to covered on the disc tray without
Reaction tray locate stop position abnormal reaction
anomaly 2. The reaction tray is not in 2. Check the optocoupler and the
the specified position electrical wiring whether or not off
and abnormal contact
1. Check if the optocoupler block is
abnormal
1. Left and right motion
Sample 2. Check if the corresponding optical
abnormalities
probe coupler and the electrical wiring of the
2. Up and down motion
moves Electric machinery is abnormal
abnormalities
abnormal 3. Check if the corresponding drive tray
installation is abnormal
131
Auto Chemistry Analyzer User Manual
Fault
Main reason Solution
information
1. Confirm the optocoupler plate under
the Reagent sample tray is normal
2. Check if the optocoupler and the
electrical wiring is off or abnormal
1. Reagent sample tray
Reagent contact
moves anomaly
sample tray 3. Reagent sample tray Pallet is loosing
2. Reagent sample tray is not
anomaly 4. The reagent sample tray adjustment
initialized at zero position
position is unreasonable, and the zero
position should be on the dispensing
probe horizontal within the range of
90 degrees
1. Do not maintain the
equipment regularly
1. Maintain the equipment regularly
2. Chemical reagents go bad,
according to user's Manual
There are chemical
2. Replace the new reagent, and correct
substances Precipitation or
storage and use of reagents
has impurity substance
3. The conductivity of pure water should
3. Purifier water quality is
be below 1μs/cm
not good
Reproducibilit 4. Replace reagent
4. Cleaning is not thorough
y not good 5. Spacing placement the reagent which
5. Reagent crystallization
has Cross contamination, or use the
6. Analysis of projects have
Cross contamination program to avoid
Cross contamination
it
7. Sample unqualified(Fibrin
6. The samples unqualified were
in the sample)
centrifuged again
8. There is a large
7. Remove interference source
electromagnetic
interference
132
Auto Chemistry Analyzer User Manual
Fault
Main reason Solution
information
The analysis and solution of the instrument failure of alarm as sheet 9-5:
Sheet 9-5 The analysis and solution of the instrument failure of alarm sheet
Arm Alarm
Describe Treatment recommendations
number source
512 Sample tray The sample tray is not moved Press the shortcut key [Ctrl]+[F3]
to the specified position to reset the device and try again.
If the problem persists, contact
your service person
1290 Sample arm Not down Press the shortcut key [Ctrl]+[F3]
to reset the device
1802 R2 arm Not down Press the shortcut key [Ctrl]+[F3]
to reset the device
1803 Dispensing Dispensing probe’s pipetting Make complete the sample
probe not enough, position[{0}]. quantity and re test
1804 R1 arm Not detect the reagent, Recommend the implementation
position[{0}]. of a reagent residue detection to
ensure that there are enough
reagents in the reagent bottle
1805 R2 arm Not detect the reagent, Recommend the implementation
position[{0}]. of a reagent residue detection to
ensure that there are enough
reagents in the reagent bottle
1806 Dispensing Dispensing probe block Recommend the implementation
probe of a reagent residue detection to
ensure that there are enough
reagents in the reagent bottle
133
Auto Chemistry Analyzer User Manual
Arm Alarm
Describe Treatment recommendations
number source
2059 AD module Unable to receive absorbance Please contact service person
data of module
2060 AD module Data acquisition module Please contact service person
interrupt abnormal
2304 Purified Net bucket deficiency Supply Adequate purifier water
Water outside the instrument
Bucket
2305 Purified Lack of water outside the Supply Adequate purifier water
Water instrument leads to outside the instrument
Bucket insufficient water in Purified
Water Bucket
2306 Purified Purifier water bucket is too Suspended water purifier
Water full
Bucket
3335 System Send data to LIS serial port Please check whether normal link
timeout to LIS
3584 Reaction The water blank is not Cleaning cuvette
tray standard, the cuvette may is
dirty
4097 Bulb Bulb energy abnormal Check the bulb, retest, if still
abnormal alarm, please contact the
maintenance person
4098 Data Did not receive the maximum Proposed check line
acquisition number of cups
communicat
ion
abnormal
4099 Data OD module does not adjust Proposed check line
acquisition the gain
communicat
ion
abnormal
4353 The device Start test signal timeout Check the sample and reagent
break down margin, reset and run again
4354 The device Reaction tray stop signal 1 Proposed reset and run again
break down timeout
4355 The device R1 arm action timeout Check the sample and reagent
break down margin, reset and run again
4356 The device Cleaning arm action timeout Proposed reset and run again
break down
4357 The device Stirring arm movement 1 Proposed reset and run again
break down timeout
4358 Machine Timer 1 times out Advice after resetting , run again
malfunction
134
Auto Chemistry Analyzer User Manual
Arm Alarm
Describe Treatment recommendations
number source
4359 Machine Reaction tray stop signal 2 Advice after resetting ,run again
malfunction timed out
4360 Machine Sample arm action timeout. Check the sample and reagent
malfunction margin, reset and run again
4361 Machine Get Reagent R2 action Check the sample and reagent
malfunction timeout margin, reset and run again
4362 Machine Stirring arm movement 2 Advice after resetting , run again
malfunction timeout
200015 Reagent Reagent position={0},bar Please re-scan after the new item
tray code={1}:Unknown new parameter added
reagent
200016 Reagent Reagent Position={0},bar Please confirm the reagent bar
tray code={1}:Reagent bar code is code and then scan it again
invalid
300000 Data Error while querying Check the corresponding reagent
processing reference value parameter setting to ensure that
the project reference values are set
up.
300001 Data An error occurred while Check the corresponding reagent
processing calculating the test parameter setting
application result
300002 Data An error occurred while Please check the database service
processing generating the test application is correct, or contact technical
support personnel
300003 Data An error occurred while Check the corresponding reagent
processing calculating the sample result parameter setting
of the sample
300004 Data Error while updating sample Please check the database service
processing status is correct, or contact technical
support personnel
300007 Data The test results exceeded the Please check the status of the
processing valid range, time={0}, sample on the sample tray
sample number={1}, Item
code={2}, Repeat number=
{3}.
300008 Data An error occurred during the Please check the previous project
processing automatic calibration of fitting parameter and status, and
reagent {0}: {1}. to ensure that the standard
quantity is sufficient for fitting the
current
300009 Data Reagent bottle empty, Please check the location of the
processing position: {0} reagent bottles
300018 Data Electrolyte calibration was Please apply for electrolyte
processing not passed calibration again
135
Auto Chemistry Analyzer User Manual
Arm Alarm
Describe Treatment recommendations
number source
300020 Data The requested barcode {0} Make sure the LIS return
processing does not contain any test information is correct
items
300023 Data Failed to send data to LIS via Please check whether the LIS link
processing database. Sample number = is normal.
{0}
400022 Communica The host computer and the Contact after sales staff
tion board slave computer version does
not match, the PC version
{0}, the next version of the
{1}
400070 Data {0} position sample value is Add samples
processing insufficient
400071 Data {0} position no sample Add samples
processing
400072 Data Lack of reagent Add reagent
processing
400073 Data No reagent Add reagent
processing
400074 Sample Sample driver module Please check Sample driver
drive communication abnormal module
module
400075 Reagent R1 Reagent R1 driver module Please check Reagent R1 driver
drive communication abnormal module
module
400076 Reaction Reaction tray module Please check Reaction tray
tray module communication abnormal module
400077 AD module AD module communication Please check AD module
abnormal
400078 Electric Exhaust not completed Please check electric circuit
circuit
60001 Adapter Network connection error, Please turn off the host computer
network card is disabled or software, and then start the device
not connected until the network connection is
normal, and then rerun the host
software
60002 Communica Communication abnormal! Please turn off the host software,
tion board and then restart the device, and
then re run the host software
60003 Communica Data return length error Please check the communication
tion board line or communication board
60004 Adapter Offline Please check the communication
line, and then run the host
software
136
Auto Chemistry Analyzer User Manual
Arm Alarm
Describe Treatment recommendations
number source
60005 Adapter Port number off Please turn off the host software,
and then restart the device, and
then re run the host software
60006 Adapter Network not connect Please turn off the host software,
and then start the device until the
network connection is normal, and
then re run the host software.
60010 Adapter Software and instrument Please contact the engineer to
number does not match reconfigure the software
60011 Strike Strike needle. The instrument Please contact the service person
needl movement is abnormal and
e stop adding sample.
100000 Reagent Reagent information not Please check if the parameter of
tray found. Reagent code ={0}. reagent is existence and determine
that the reagent is also set on that
reagent tray
100001 Test An error occurred while Please check the communication
application downloading the test line is normal, and determine
application whether the instrument is in
standby or normal working state
100002 Test An error occurred when the Please check the communication
application instrument began testing the line is normal, and determine the
sample instrument is in standby mode
100010 Data An error occurred while Please Check database parameter
processing loading and refreshing the and status, contact technical
sample tray state support person
100011 Data Automatic re-check failure Please Check database parameter
processing and status, contact technical
support person
100012 Data Invalid fitting type code. Item Please check the previous project
processing = {0}. fitting parameter and state
100013 Reaction The reaction tray temperature
tray is too low to ensure that the
test results are correct, test
has stopped
100014 Reaction The reaction tray temperature
tray is too high to ensure that the
test results are correct, test
has stopped
100015 Reaction Reaction tray’s temperature is
tray low, test results may be
affected
100016 Reaction Reaction tray’s temperature is
tray high, test results may be
affected
137
Auto Chemistry Analyzer User Manual
Arm Alarm
Describe Treatment recommendations
number source
100017 Reaction The reaction tray temperature
tray sensor not linked or damaged.
100022 Data Invalid instrument type Try reinstalling the program or
processing contact your service provider
100023 Sample arm Sample arm initialization Please restart, or contact technical
error. support person
100024 Reaction The initial error of the Please restart, or contact technical
tray moving part of the reaction support person
tray
100025 Mixing arm Mixing arm initialization Please restart, or contact technical
error support person
100026 Others Unable to get the basic Please check the database service
information from the database is correct, or contact technical
<= Alarm ID={0} support person
100031 Date Check failed after Try reinstalling the program or
processing initialization contact your service provider
100032 Others Instrument initialization Proposed to restart the
failed slave computer
200000 Reaction The cup blank value is lower Please clean the cuvette and
tray than the lower limit, the replace the cuvette if necessary
cuvette number is ={0}
200001 Reaction The cup blank value is lower Please clean the cuvette and
tray than the lower limit, the replace the cuvette if necessary
cuvette number is ={0}
400079 Reagent R2 Reagent R2 driver module Please check the reagent R2 driver
drive communication abnormal module
module
400080 Stirring Communication module Please check the stirring module
drive Abnormal
module
138
Auto Chemistry Analyzer User Manual
139
Auto Chemistry Analyzer User Manual
140
Auto Chemistry Analyzer User Manual
141
Auto Chemistry Analyzer User Manual
142
Auto Chemistry Analyzer User Manual
143
Auto Chemistry Analyzer User Manual
The reagent formula’s changing in cross-contamination, so the above test is only for
reference, if not, please refer to the actual test situation.
144