Download as pdf or txt
Download as pdf or txt
You are on page 1of 6

TAG project data

Project : Oleoplan IRQ


Customer: C.Project No.: 0
Contact person: --- Phone: --- Fax: ---
eMail: ---

TAG: SPEC-AID BIO9402 / BIO9041


Timestamp: 15.07.2019 08:13 AM
Review number: 1
Product: Promass A 200
Manufacturer: Endress+Hauser
Extended Order Code: 8A2B02- SAD6SAA1
Data
TAG remarks:
Customer remarks:
Critical: No
Process medium:
Process application:
Measurement principle: Coriolis (Promass)
Bus address:
Sub location:
User name:
Price:
Delivery time (days):
Offer text:
Online Shop remarks:
Equipment remarks:
Comment 1:
Comment 2:
Person:
Attachments:

Print date: 15.07.2019 08:13 AM -1/6-


Applicator®: 11.05.02 / 12.02.31

Under no circumstances is Endress+Hauser liable for errors, neither in the Software and in its documentation, nor for any errors
and consequential damage which may arise out of their use. The results in Applicator apply to parameters entered by the user. A
change in these parameters could lead to different results. Mandatory data are in the according technical information (TI).
Applicator Sizing - Flow
Project : Oleoplan IRQ
Customer: C.Project No.: 0
Contact person: --- Phone: --- Fax: ---
eMail: ---

TAG : SPEC-AID BIO9402 / BIO9041


Timestamp: 15.07.2019 08:13 AM Review number: 1
Sales order number:

Sizing Sheet
General Parameters
Fluid 0 Dummy-Liquid (Water) based on SP -
MODIFIED-
State Liquid
Character Clean Atmospheric Pressure 1.0133 bar_a
Abrasivity Not abrasive Standard EN/DIN/ISO
Fluid Group (PED) Dangerous Fluid (Fluid group 1)
Fluid type Newtonian

Operating Conditions
minimum nominal maximum
Requested Flow 300 3 000 6 000 g/h
Pressure 3 3 3 bar_g
Temperature 25 25 25 °C
Density 920 920 920 kg/m3
Viscosity 2 2 2 cP
Vapor Pressure 0.0274 0.0274 0.0274 bar_a
Design pressure (min/max) 3 3 bar_g
Design Temp. (min/max) 25 25 °C

Flowmeter : Promass A 200 Generation: 3 Device model: 1


Flow Principle Coriolis (Promass)
Meter Size DN 2
Operating range min. 0 g/h
Operating range max. 100 000 g/h
Material (sensor) * Stainl. Steel 1.4435 (316/316L)
Process connection* DN15 PN40 groove, 1.4404/316/316L, flange EN1092-1-D
PED category ** : Flowmeter is generally excluded from PED

Extended Order Code


Qty Item Description Extended Order Code
1 Flowmeter Promass A 200 8A2B02-******SAD6SAA1

*The user is responsible for the selection of process-wetted materials in view of their corrosion resistance. Endress+Hauser makes no guarantees and
assumes no liability for the corrosion resistance of the materials selected here for the application described above.
** The PED category is an Endress+Hauser recommendation and depends on the fluid category, process data as well from the max. permissible
pressure of the selected pressure rating. The fluids of the Applicator data base are classified to 67/548/EWG.

Print date: 15.07.2019 08:13 AM -2/6-


Applicator®: 11.05.02 / 12.02.31

Under no circumstances is Endress+Hauser liable for errors, neither in the Software and in its documentation, nor for any errors
and consequential damage which may arise out of their use. The results in Applicator apply to parameters entered by the user. A
change in these parameters could lead to different results. Mandatory data are in the according technical information (TI).
Applicator Sizing - Flow
Project : Oleoplan IRQ
Customer: C.Project No.: 0
Contact person: --- Phone: --- Fax: ---
eMail: ---

TAG : SPEC-AID BIO9402 / BIO9041


Timestamp: 15.07.2019 08:13 AM Review number: 1
Sales order number:

Sizing Sheet
Sizing and Calculated Results
minimum nominal maximum
Requested Flow 300 3 000 6 000 g/h
Velocity 0.029 0.288 0.577 m/s
Velocity Max. 0.018 0.185 0.369 m/s
Pressure loss 0 0.01 0.02 bar
Measurement error Mass*** 1.67 0.17 0.1 %
Reynolds No. 21 212 424
***For error calculation, the specified reference conditions for the calibration of the flowmeter according to ISO/IEC 17025 apply. Further information in
technical documentation.

Warning(s)

Notice(s)

Message(s)

Print date: 15.07.2019 08:13 AM -3/6-


Applicator®: 11.05.02 / 12.02.31

Under no circumstances is Endress+Hauser liable for errors, neither in the Software and in its documentation, nor for any errors
and consequential damage which may arise out of their use. The results in Applicator apply to parameters entered by the user. A
change in these parameters could lead to different results. Mandatory data are in the according technical information (TI).
Applicator Sizing - Flow
Project : Oleoplan IRQ
Customer: C.Project No.: 0
Contact person: --- Phone: --- Fax: ---
eMail: ---

TAG : SPEC-AID BIO9402 / BIO9041


Timestamp: 15.07.2019 08:13 AM Review number: 1
Sales order number:

Fluid properties sheet


Fluid
Fluid name 0 Dummy-Liquid (Water) based on SP - State Liquid
MODIFIED-
Chemical formula Calculation standard Supporting Points

Fluid description
Solid content 0 %
Medium character Clean
Abrasivity Not abrasive
Conductivity more than 50 µS/cm
Fluid group (PED) Dangerous Fluid (Fluid group 1)
Fluid type Newtonian

Basic fluid parameters


Tc (Critical temperature) 374.1 °C Tm (Melting point) 0.04 °C
Pc (Critical pressure) 221.2 bar Tb (Boiling point) 100 °C
Heating value 0 kWh/Nm3
Temperature / Viscosity Temperature / Density
Temperature Viscosity Temperature Density
1 20 °C 1 cP 1 20 °C 998 kg/m3
2 80 °C 0.35 cP 2 80 °C 970 kg/m3
Temperature / Heat capacity
Temperature Thermal capacity
1 0 °C 4.215 kJ/(kg*K)
2 50 °C 4.177 kJ/(kg*K)
3 100 °C 4.215 kJ/(kg*K)

Calculated results
Density nom. 920 kg/m3 Pressure nom. 3 bar_g
Viscosity nom. 2 cP Temperature nom. 25 °C
Sound velocity nom. 1 502 m/s
Vapor pressure nom. 0.0274 bar_a
Thermal expansion coeff. Beta P 0.0005
Viscosity coefficient AL -6.179
Viscosity coefficient BL 1 811.51

Reference values: Normal conditions (SI): Standard conditions (US):


Atmospheric pressure 1.0133 bar_a Atmospheric pressure 1.0133 bar_a

Print date: 15.07.2019 08:13 AM -4/6-


Applicator®: 11.05.02 / 12.02.31

Under no circumstances is Endress+Hauser liable for errors, neither in the Software and in its documentation, nor for any errors
and consequential damage which may arise out of their use. The results in Applicator apply to parameters entered by the user. A
change in these parameters could lead to different results. Mandatory data are in the according technical information (TI).
Applicator Sizing - Flow
Project : Oleoplan IRQ
Customer: C.Project No.: 0
Contact person: --- Phone: --- Fax: ---
eMail: ---

TAG : SPEC-AID BIO9402 / BIO9041


Timestamp: 15.07.2019 08:13 AM Review number: 1
Sales order number:

Compare sensors sheet


General Parameters
Fluid 0 Dummy-Liquid (Water) based on SP -
MODIFIED-
State Liquid
Character Clean Atmospheric Pressure 1.0133 bar_a
Abrasivity Not abrasive Standard EN/DIN/ISO
Fluid Group (PED) Dangerous Fluid (Fluid group 1)
Fluid type Newtonian

Sizing and Calculated Results


Alternative 1 Current Alternative 2
Flowmeter Promass A 200
Generation/Device model 3/1
Flow Principle Coriolis (Promass)
Meter Size DN 2
Operating range min. 0 g/h
Operating range max. 100 000 g/h
Velocity at req. Flow min. 0.029 m/s
Velocity at req. Flow nom. 0.288 m/s
Velocity at req. Flow max. 0.577 m/s
Sensor velocity at req. Flow min. 0.018 m/s
Sensor velocity at req. Flow nom. 0.185 m/s
Sensor velocity at req. Flow max. 0.369 m/s
Pressure loss at req. Flow min. 0 bar
Pressure loss at req. Flow nom. 0.01 bar
Pressure loss at req. Flow max. 0.02 bar
Meas. error Vol. at req. Flow min.*** 1.67 %
Meas. error Vol. at req. Flow nom.*** 0.17 %
Meas. error Vol. at req. Flow max.*** 0.1 %
Meas. error Mass at req. Flow min.*** 1.67 %
Meas. error Mass at req. Flow nom.*** 0.17 %
Meas. error Mass at req. Flow max.*** 0.1 %
Warning(s)
*The user is responsible for the selection of process-wetted materials in view of their corrosion resistance. Endress+Hauser makes no guarantees and
assumes no liability for the corrosion resistance of the materials selected here for the application described above.
***For error calculation, the specified reference conditions for the calibration of the flowmeter according to ISO/IEC 17025 apply. Further information in
technical documentation.

Print date: 15.07.2019 08:13 AM -5/6-


Applicator®: 11.05.02 / 12.02.31

Under no circumstances is Endress+Hauser liable for errors, neither in the Software and in its documentation, nor for any errors
and consequential damage which may arise out of their use. The results in Applicator apply to parameters entered by the user. A
change in these parameters could lead to different results. Mandatory data are in the according technical information (TI).
Applicator Sizing - Flow
Project : Oleoplan IRQ
Customer: C.Project No.: 0
Contact person: --- Phone: --- Fax: ---
eMail: ---

TAG : SPEC-AID BIO9402 / BIO9041


Timestamp: 15.07.2019 08:13 AM Review number: 1
Sales order number:

Compare sizes Sheet


General Parameters
Fluid 0 Dummy-Liquid (Water) based on SP -
MODIFIED-
State Liquid
Character Clean Atmospheric Pressure 1.0133 bar_a
Abrasivity Not abrasive Standard EN/DIN/ISO
Fluid Group (PED) Dangerous Fluid (Fluid group 1)
Fluid type Newtonian

Sizing and Calculated Results


Next Smaller Size Current Size Next Bigger Size
Flow meter Promass A 200 Promass A 200 Promass A 200
Generation/Device model 3/1 3/1 3/1
Flow Principle Coriolis (Promass) Coriolis (Promass) Coriolis (Promass)
Meter Size DN 1 DN 2 DN 4
Process connection* DN15 PN40 DN15 PN40 DN15 PN40
Operating range min. 0 0 0 g/h
Operating range max. 20 000 100 000 450 000 g/h
Velocity at req. Flow min. 0.095 0.029 0.007 m/s
Velocity at req. Flow nom. 0.953 0.288 0.072 m/s
Velocity at req. Flow max. 1.906 0.577 0.144 m/s
Sensor velocity at req. Flow min. 0.095 0.018 0.008 m/s
Sensor velocity at req. Flow nom. 0.953 0.185 0.076 m/s
Sensor velocity at req. Flow max. 1.906 0.369 0.152 m/s
Pressure loss at req. Flow min. 0.01 0 0 bar
Pressure loss at req. Flow nom. 0.21 0.01 0 bar
Pressure loss at req. Flow max. 0.56 0.02 0 bar
Meas. error Vol. at req. Flow min.*** 0.33 1.67 6.67 %
Meas. error Vol. at req. Flow nom.*** 0.1 0.17 0.67 %
Meas. error Vol. at req. Flow max.*** 0.1 0.1 0.33 %
Meas. error Mass at req. Flow min.*** 0.33 1.67 6.67 %
Meas. error Mass at req. Flow nom.*** 0.1 0.17 0.67 %
Meas. error Mass at req. Flow max.*** 0.1 0.1 0.33 %
Reynolds No. 482 212 136
Warning(s)
*The user is responsible for the selection of process-wetted materials in view of their corrosion resistance. Endress+Hauser makes no guarantees and
assumes no liability for the corrosion resistance of the materials selected here for the application described above.
***For error calculation, the specified reference conditions for the calibration of the flowmeter according to ISO/IEC 17025 apply. Further information in
technical documentation.

Print date: 15.07.2019 08:13 AM -6/6-


Applicator®: 11.05.02 / 12.02.31

Under no circumstances is Endress+Hauser liable for errors, neither in the Software and in its documentation, nor for any errors
and consequential damage which may arise out of their use. The results in Applicator apply to parameters entered by the user. A
change in these parameters could lead to different results. Mandatory data are in the according technical information (TI).

You might also like