Professional Documents
Culture Documents
5991 8993en Icp Ms Journal 71
5991 8993en Icp Ms Journal 71
Page 7
Japan Hosts Another
Successful Asia Pacific Winter
Plasma Conference
Page 8
On-Demand Webinars; Agilent We welcome your input, whether you’d like to contribute an article, send feedback or
ICP-MS Publications ask a question. Please contact the Journal editor at icpms@agilent.com
Agilent ICP-MS Journal February 2018, Issue 71
2
February 2018, Issue 71 Agilent ICP-MS Journal
Results and Discussion Note that an accurate and precise split ratio is assumed
for such mass balance calculations. Moreover, the
An arsine in nitrogen standard was used for the
arsenic peak coelutes with the sample matrix in the total
optimization of the system operating conditions.
analysis (column bypass) measurement, so sensitivity
Chromatograms of a 68 ppb (v/v) arsine standard in
may not be consistent for both direct and GC conditions.
nitrogen and a N2 blank are shown in Figure 1. The N2
These two facts could explain the significant uncertainty
blank chromatogram is shown on a y-axis scale that is
(12% RSD) observed for the calculated recovery.
approximately two orders of magnitude lower than the
scale used for the arsine sample. As can be seen, the Calibration: Calibration curves were generated for
first peak in the arsine standard chromatogram appears “total As” and speciation (AsH3) analysis based on AsH3
at 0.2 min, corresponding to the total arsenic eluted standard levels from 0.2 ppb to 68 ppb, as shown in
from the inert transfer line (column bypass flow). A Figure 2. The method showed very good linearity within
second peak is observed at 3.8 min corresponding to the the tested range of concentration for both total (r2 =
speciation analysis (arsine eluted from the column). The 0.9996) and speciation analysis (r2 = 0.998). Detection
chromatograms demonstrate that total and speciation limits of 2 ppt for total As and 12 ppt for AsH3 by
analysis can be performed simultaneously within the speciation analyses were obtained. To the best of our
same injection. knowledge, these are the lowest DLs for direct total As
and AsH3 speciation analysis in gaseous samples ever
reported in the literature.
is expected in the inert transfer line. Therefore the Synthetic sample analysis: Liquefied butane doped
comparison between the experimentally obtained and with a non-certified level of arsine was analyzed as a
the expected theoretical GC inlet split ratio can be used synthetic sample. He cell mode was used for analysis of
to assess if the total As observed is exclusively in the the samples to minimize the occurrence of polyatomic
form of arsine or if other As species are present in the interferences. The vaporization chamber was set at
sample. Integration of peak areas for the “total As” and 120 ºC and connected on-line with the gas dilutor for
arsine in each point of a calibration curve from 0.2 to 68 vaporization of the sample. A chromatogram of this
ppb (n = 13) showed a mean recovery for the arsine peak sample diluted 1:7 in Ar is shown in Figure 3. Results
of 89 ± 11 % of the total As in the standard. This indicates obtained for total and speciation analysis are given in
that, as expected, As is exclusively present as arsine in Table 2.
the standard.
3
Agilent ICP-MS Journal February 2018, Issue 71
Table 2. Quantitative results (ppb As, v/v) obtained for liquefied butane
and propylene samples. Expanded uncertainties (95% confidence level)
are given (not including the uncertainty of the arsine standard).
Total analysis 92 ± 4 54 ± 4 34 ± 3
Speciation
95 ± 4 53 ± 13 31 ± 5
analysis
4
February 2018, Issue 71 Agilent ICP-MS Journal
Introduction and WO3+, shifting to higher masses. Hg+ does not react
Many mercury (Hg) compounds are toxic, causing and so remains at m/z 200. Q2 is set to 200 u, allowing
symptoms ranging from skin irritation, headaches, and
200
Hg+ ions to pass to the detector free of interference. In
tremors, through to nervous system damage and renal MS/MS, reaction chemistry is controlled and consistent
failure. Because Hg compounds are easily absorbed because only the target analyte mass enters the CRC.
through the skin, their use in cosmetics is controlled. For This capability offers a much more predictable and
example, the US Food and Drug Administration (FDA) reliable approach to resolving interferences on a wide
does not allow Hg in cosmetics, except under specific range of elements.
conditions where there are no other safe and effective Experimental
preservatives available [1].
Standards and samples: Mercury standards were
Increasingly, however, Hg has been found in cosmetic prepared in 0.5 % high purity hydrochloric acid (TAMA-
products such as skin creams, soap, and lotions sold as Pure-AA-100, Kanagawa, Japan).
“anti-aging” or “skin lightening”.
A tungsten-rich cosmetic sample was bought from a
Analysis of Hg by ICP-MS: Hg is a challenging element local store in Shanghai. The liquid sample was weighed
to determine at low levels by ICP-MS. It has a high first to the nearest 0.100 g, and then diluted 100-fold with
ionization potential (10.44 eV), so is relatively poorly Milli-Q de-ionized water acidified with 0.5 % HCl. The
ionized in the plasma, leading to low sensitivity. Also, Hg sample was shaken for a couple of minutes to ensure it
has seven naturally occurring isotopes, each with relatively was fully homogenized.
low % abundance, further reducing sensitivity. Despite Instrumentation: An Agilent 8900 Standard configuration
these difficulties, ICP-MS can still be used successfully ICP-QQQ was used. The standard sample introduction
to perform trace-level analysis of Hg. However, some system was used, comprising a glass concentric nebulizer,
cosmetics contain large amounts of tungsten (W). This quartz double-pass spray chamber, 2.5 mm injector quartz
leads to polyatomic interferences from WO+ and WOH+ torch, and Ni interface cones. The ICP-QQQ was operated
that affect all the Hg isotopes, making Hg measurement in no gas mode, with He cell gas, and with O2 cell gas in
even more challenging. For example, the most abundant both single quad (SQ) and MS/MS modes.
Hg isotopes, 200Hg and 202Hg, suffer overlaps from
Results and Discussion
184
W16O+ and 186W16O+, respectively. Conventional single
quadrupole ICP-MS (ICP-QMS) fitted with a collision/ Multiple isotope analysis study: To demonstrate the
reaction cell (CRC) is unable to resolve the WO+ and ability of MS/MS to resolve interferences on multiple
WOH+ interferences sufficiently to allow the accurate isotopes, a scan spectrum comparison was made.
determination of Hg at trace levels. The mass range of the Hg isotopes was acquired for a
simple Hg standard and a solution containing the same
In this study, the superior interference removal capability
concentration of Hg spiked into a high W matrix. The
of triple quadrupole ICP-MS (ICP-QQQ) in MS/MS mode
overlaid spectra are shown in Figure 1, confirming that
with O2 cell gas was used for on-mass detection of Hg
the measured isotopic abundances match the natural Hg
in a tungsten matrix. 200Hg+ and WO+ ions at m/z 200
isotope pattern despite the presence of the high W matrix
enter the CRC. WO+ reacts with O2 cell gas to form WO2+
in the second sample.
5
Agilent ICP-MS Journal February 2018, Issue 71
6
February 2018, Issue 71 Agilent ICP-MS Journal
7
On-Demand Webinars