ITSD
CONTENTS
01
PART ONE
System Overview
05
PART FIVE
Calibration
02
PART TWO
Switch On/Off 06
PART SIX
Control
03
PART THREE
Preparation 07
PART SEVEN
Sample Test
04
PART FOUR
System Test 08
PART EIGHT
Daily Maintenance
Alarm Light
RFID Reader
Consumable Indicator
Liquid Joints
Cuvette Storage
Cuvette Dustbin
Starter Area
FOCUS MAKES PROFESSION
01 System Overview
PART ONE
Comsumables
Code NAME STORAGE Usage Function
No sunlight or humidity.
1 Reaction module Disposable CLIA Reaction place
Shelf life:2 years.
15-30°C, Shelf life :1 year. 1bottle (714mL)+distill water= 10L Wash the magnetic micro-
2 Wash Concentrate
On board stability: 4 weeks. Wash liquid beads and needles.
Measure
Wash
Incubation
Sampling
Add reagent
Username: snibe
Password: snibe
3.When status of the machine marked by red rectangle show “Standby”,tests can be run.
A. Software:Click “STOP”,
then click “EXIT”
Scan and load the reagents will be used to the reagent area. The analyzer will shake
the magnetic micro-beads Component for 30 minutes automatically.
Green : normal
Red : abnormal
5
If stop using the analyzer ≥5 days, 30 times for Pipettor, Washer and Chamber Prime.
If only replace the wash liquid, 30 times for Pipettor and Washer Prime, 10times for Chamber Prime.
If only replace the starters, 30 times for Chamber Prime, 10times for Pipettor and Washer Prime.
If replace both wash liquid and starters, 30 times for Pipettor, Washer and Chamber Prime
FOCUS MAKES PROFESSION
05
PART FIVE
Calibration
3
4
If you want to run batch calibration, select all reagents need to be calibrated at 2nd step.
1
CV:<15%
4 Dev:
-70%<Dev<70%
2 click once
5
6
2
②Set the QC information
① ③ ⑥
4
5 Select the QC information
8 6 Select the QC assay
3
4
① Collect and deal with samples gently and avoid shaking violently; fibrin, hemolysis, lipemia, jaundice
and bubbles might affect the results.
② Avoid repeated freezing and thawing.(< 2 times)
③ Centrifugation for serum samples :10-15 min with over 4000r/min.
④ Use the required sample tube.
1
①Load the STAT sample in
the sample rack.
5 Select the assay
2
Select the rack
3
Select the position
3 6 Note:
Select the diluted assay 4 Select the preferred dilution ratio The sample volume is
at least 10μL.
2
After finishing editing sample
4
information, click “Dilution” to
1
choose the diluted ratio you
5
have selected or set.
6
3
7
Standard tube is
recommended.
MAGLUMI X3-1.2.3.0
Fixtures
Sample rack installation
Precautions
The tube will be stuck as it rotates more than
Same sample tube for one rack, or it will 90°. Rotate in reverse until label faces rack
be an error. opening and unplug it out.
1) Scan identification
In user, the sample tube is default (code-named A), When scanner identifies the code, it will upgrade the
and you can add new type by two ways : second letter for the type of the tube in [ work list].
And it will restore back to the Manual setting code or
1) Scan identification 2) Manual setting Normal code as the sample rack is unplugged.
Priority for the type of Sample tube to identify: Notice:
Scan identification > Manual setting > Normal. Load the rack with fixture and make sure
the type can be identified accurately each
As the scanner identify the different type from the time.
Manual setting, the software will give priority to the
Adjust the fixture for a little inclined as if
type identified by scanner. it could not be identified every time.
2) Manual setting
Click <General Setting> in <setting> , and select the type for each column then
Save. It will change immediately in [ working list ].
2) Manual setting
The type of manual setting is fixed and effective that
it will not be changed as the rack is unplugged, and
the subsequent tube loaded in this column will be
defaulted to the type you set.
( It can only be temporarily covered by the type
identified by scanner, and it will restore back to the
default type when the rack is unplugged.)