MISPA PLUS
Dear Customers,
Yours truly,
ENGINEERING EXCELLENCE
CASTING TRUST
PACKING LIST
Item /x Item /x Item /x Item /x
Operation Manual (1 No.) Pipette Holder (1 No.) Print Roll (2 Nos.) Power Cord (1 No.)
Installation Report
Warranty Card (1 No.) Pipette (100-1000µl) (1 Pipette (5 - 50µl) (1 No.)
(1No.)
No.)
1
MISPA PLUS
[ADL/RDE/P016/044 -R00]
Forwards
1. Intellectual Property
AGAPPE Diagnostics Ltd owns the intellectual property right to this AGAPPE product and
this manual. This manual may refer to information protected by copyright or patents and
does not convey any license under the patent rights or copyright of AGAPPE or of others.
AGAPPE intends to maintain the contents of this manual as confidential information.
2. Safety Precautions
Page 2 of 35
[ADL/RDE/P016/044 -R00]
4. Warranty
Agappe guarantees that all its instruments are free from manufacturing defects or faults.
Agappe undertake to repair or substitute free of charge any part, which may be found to
have manufacturing defects.
Warranty Conditions
• Such parts of the instruments that may have defects due to materials or manufacturing
defects will be repaired or substituted free of charge.
• The warranty does not cover those parts that are subject to normal wear and tear and
which are normally changed from time to time during maintenance procedure of the
instrument.
• Repairs and interventions carried out during the period of the warranty do not extend
or renew said period of warranty. The repairs of the instrument will be carried out by
authorized technician on request by the customer to the company.
Termination of Warranty
This warranty shall be terminated at the end of the warranty period and in the following
cases:
• Where attempts to make repairs or alternations have been made by unauthorized
person and/or with spare parts, which are not originals.
• Alternations have been made to the serial number of the product or of the certificate.
• The instrument is transferred to a new location without prior written approval from
Agappe.
Elements not covered by warranty
The following damages and faults are not covered by warranty.
• Damage derived from insufficient electric circuit
• Breakdown caused by imprudence
• Damage caused by force majeure such as fire and earthquake
• Consumable items such as fuses, lamps, all tubing etc. are not covered
Page 3 of 35
[ADL/RDE/P016/044 -R00]
TABLE OF CONTENTS
Forwards............................................................................................................ 2
1. Intellectual Property ........................................................................................ 2
2. Safety Precautions ........................................................................................... 2
3. Caution for Biohazard Management ....................................................................... 2
4. Warranty ...................................................................................................... 3
5. Customer service department.............................................................................. 3
1 Overview ..................................................................................................... 7
1.1 Introduction............................................................................................. 7
1.2 Features ................................................................................................. 8
1.3 Specifications ........................................................................................... 9
2 Installation ................................................................................................. 10
2.1 Unpacking and Installation ......................................................................... 10
2.2 Installation Environment ............................................................................ 10
2.3 Printer and paper installation ...................................................................... 10
2.4 Power supply connection ........................................................................... 11
2.5 Connecting waste bottle ............................................................................ 12
2.6 Test trial .............................................................................................. 13
3 Basic Operating Instructions ............................................................................ 13
3.1 How to aspirate liquid ............................................................................... 13
3.2 Instructions for the correct aspiration ............................................................ 13
3.3 Reset and Turn off ................................................................................... 13
3.4 Automatic display sleep function .................................................................. 13
3.5 Common icons in all screens ....................................................................... 14
4 Analyser Settings and Functions ....................................................................... 15
4.1 Start Up ............................................................................................... 15
4.2 Home .................................................................................................. 16
4.3 Test .................................................................................................... 17
4.3.1 Test List ......................................................................................... 17
4.3.2 Begin test ....................................................................................... 18
4.3.3 Quality Check ................................................................................... 20
4.3.4 Program test .................................................................................... 22
4.4 Maintenance .......................................................................................... 26
4.4.1 General .......................................................................................... 26
4.4.2 Temperature .................................................................................... 27
4.4.3 Pump Calibration ............................................................................... 27
4.5 Absorbance ........................................................................................... 28
4.6 History ................................................................................................. 29
Page 4 of 35
[ADL/RDE/P016/044 -R00]
Page 5 of 35
[ADL/RDE/P016/044 -R00]
LIST OF FIGURES
Page 6 of 35
[ADL/RDE/P016/044 -R00]
1. Overview
1.1 Introduction
The Mispa Plus, true semi-automated clinical chemistry analyser is a symbol of superior
technology and convenience, designed to deliver high accuracy, quality results with
least turnaround time and low maintenance.
Mispa Plus, a technically superior semi-automated analyser which ensures consistent accuracy
and convenience to the user. Because of the embedded system, the equipment can switch off
while at any menu or at any stage. Mispa Plus comes with 7-inch TFT colour display with
capacitive touch screen for user friendly operation. Also, Mispa Plus provides USB and
Ethernet connectivity options to the user.
Mispa Plus gives convenient option to work through your computer facilitating convenience in
programming, sample testing, QC analysis, reporting etc.
Page 7 of 35
[ADL/RDE/P016/044 -R00]
1.2 Features
• It can perform End Point, Kinetic, Fixed time and Multi standard assays.
• 7-inch TFT colour display with capacitive touch screen for user friendly operation of
the analyser.
Page 8 of 35
[ADL/RDE/P016/044 -R00]
1.3 Specifications
Heads Specifications
Dimension(mm) 450 x 350 x 385 (LXWXH)
Weight 9.7 Kg
Graphic display 7 inches with capacitive touch
Patient name length 24 Characters
Maximum t e s t r e s u l t storage 5000 test results
capacity
Light source Halogen lamp 12V /20W
Photo detector Silicon photodiode (range 300-1000nm)
Minimum Aspiration Volume 250 µL
Interface USB and ETHERNET
Spectral Range 340-630nm
Wave-length selection Automatic, via 8-position filter wheel. 6 standard
interference filter-340nm, 405nm, 505nm, 546nm,
578nm, 630nm.
Page 9 of 35
[ADL/RDE/P016/044 -R00]
2. Installation
2.1 Unpacking and Installation
Place the shipment containing the analyser in upright position and remove the BOPP tape.
Carefully lift the analyser with the supporting cushion packing and place on clean table-top.
Do not use any sharp objects or gadgets to unpack the shipper as they can cause damage to
the analyser inside.
After unpacking the shipping package, please verify that the product does not have any
damage from shipment and make sure all listed components are included.
Power supply cable, Ethernet cable, Waste Bottle with Holding Clamp, Paper roll (2 Nos),
User manual, Warranty Card, Dust Cover, Pipette Holder, Pipette 100-1000 µL & 5-50 µL.
(a)
(b)
Page 10 of 35
[ADL/RDE/P016/044 -R00]
(c) (d)
1. Before connecting the power cord, check the AC power supply corresponding to the value
that is stated on the analyser label.
2. Make sure that your AC main line has an efficient ground line. Bad ground line connection
may affect analysis result and damage the analyser.
3. Connect power cord to the port, at the rear panel of the analyser.
1. Connect the power plug to a good, grounded AC wall outlet, preferably one that is not
shared with other electric appliance and with low function of line voltage compared to
the standard voltage specified (10-15%).
2. Keep the analyser away from other appliances that generate high frequency electric noise
(eg: radiological instrument).
3. Before installing the analyser or connecting the power, make sure that the analyser is
turned off.
4. Make sure that your AC main line has an efficient ground line. Bad ground line connection
may affect analysis result and damage the instrument.
Page 11 of 35
[ADL/RDE/P016/044 -R00]
1. To avoid the spilling of waste water around the analyser, make sure that waste outlet
tube is connected to waste bottle properly.
Page 12 of 35
[ADL/RDE/P016/044 -R00]
Do not operate the touch screen with sharp objects such as pin, pen tip, nails
etc as they can cause scratches and damage on display screen
• Turn “ON” the power switch, present at the back of the analyser.
• Analyser gets switched on. The analyser will perform the start-up function and
display the main screen after successfully completing the start-up function.
• Perform wash function by selecting the wash function from main menu.
• Once the analyser is stabilized, kindly perform1-2 tests and evaluate the test results.
• Confirm that the analyser is working as per specifications.
Page 13 of 35
[ADL/RDE/P016/044 -R00]
Icon Description
Warning
Temperature unstable
Temperature stable
To stop wash
Page 14 of 35
[ADL/RDE/P016/044 -R00]
Page 15 of 35
[ADL/RDE/P016/044 -R00]
4.2 Home
Page 16 of 35
[ADL/RDE/P016/044 -R00]
4.3 Test
4.3.1 Test List
• All 300- test names will be displayed, 20 tests per screen and we can move to the next
set of tests by moving the horizontal scroll.
• Analyser will load and display the selected test run screen. (as shown above)
It displays the program details of the test.
Page 17 of 35
[ADL/RDE/P016/044 -R00]
Page 18 of 35
[ADL/RDE/P016/044 -R00]
• The right side Calibrate column will change based on fixed, single
or multi standard test. For fixed test, the Calibrate button will be disabled.
• to feed paper
Page 19 of 35
[ADL/RDE/P016/044 -R00]
• The Quality Check on the test detail screen brings to Quality Check home screen.
• In this screen, there is a dropdown to select the QC level. By default, level 1 is selected,
and the corresponding values for that particular level will be displayed along with the
graph.
• You can change the QC level from the dropdown and the corresponding values will be
shown along with graph.
• The right-side column shows the number of QCs performed for the selected QC level.
4.3.3.1 Run QC
Page 20 of 35
[ADL/RDE/P016/044 -R00]
4.3.3.2 QC History
• Displays the history of quality checks performed for the selected test.
• At a time, 10 entries will be shown, click on left/right arrow keys to move to next list
of entries.
• <- Back to return to QC test run page.
4.3.3.3 QC Set Up
Page 21 of 35
[ADL/RDE/P016/044 -R00]
4.3.4.1 General
4.3.4.2 Result
• Enter the Aspiration volume, check the radio button for wavelength. Select the
dropdowns for Wavelength 1 and Wavelength 2. If wavelength ‘Double’ is checked,
select the Decimal and Unit dropdowns to set the result format.
• Print – to print the details
• Save – to save the entered data.
• <-Back to return to test detail page.
Page 22 of 35
[ADL/RDE/P016/044 -R00]
4.3.4.3 Time
4.3.4.4 Calibration
With fixed factor
Page 23 of 35
[ADL/RDE/P016/044 -R00]
• For single and multi-standard tests, choose the standard and select the type. Use
the dropdown for Number of Standards to choose single or multi standard. Enter
the concentration and factor range (for single standard only) and Absorbance will
be calculated.
• If number of standards is greater than 5, click on the down arrow “6-10” to display
the next series from 6 to 10.
4.3.4.5 Reference
Page 24 of 35
[ADL/RDE/P016/044 -R00]
Page 25 of 35
[ADL/RDE/P016/044 -R00]
4.4 Maintenance
4.4.1 General
Page 26 of 35
[ADL/RDE/P016/044 -R00]
4.4.2 Temperature
• Click ‘Calib’ and an instruction will be displayed asking the user to aspirate 3 mL of
water. Press the aspiration switch and aspirate 3mL of water. Once the water is
empty, release the switch. Be sure, no air gets aspirated. The observed value will
be shown. For fine tuning, click the upper or lower values.
• Once done, click the ‘Check calibration’ button and aspirate 500 mL of water to
confirm the calibration.
Page 27 of 35
[ADL/RDE/P016/044 -R00]
4.5 Absorbance
• Select filter and press the button. The button will toggle to , so
that the user can stop when required.
• Follow the instructions shown in the display panel at the right-hand side top.
• After aspiration, analyser will plot the graph along with the ABS value.
Page 28 of 35
[ADL/RDE/P016/044 -R00]
• If the user wants to know the absorbance of any new sample with respect to the current
aspirated sample, click on so ABS will be set to 0.00 and the running sample
will be considered as blank for the next sample.
ABS Settings
• Enter the delay time, print interval, aspiration volume and temperature and click ‘Save’
option to save the ABS settings.
4.6 History
Page 29 of 35
[ADL/RDE/P016/044 -R00]
• Select a particular test to display the results calculated for that test.
• We can also perform search directly by either entering the patient id or by date range.
• Selecting a row highlights that particular result and we can print it.
Page 30 of 35
[ADL/RDE/P016/044 -R00]
• Analyser will load and display the Serial No, Date, Pt ID, Test Name, Result, unit and flag.
• At a time, 10 entries will be displayed. To see more, make use of left/right arrow keys to
scroll the entries.
4.7 Help
• You can view the system details, guidelines, operations and contact details by clicking on
the corresponding tabs.
Page 31 of 35
[ADL/RDE/P016/044 -R00]
5. Maintenance
5.1 Wash
After the completion of each test/QC run, it is recommended that the flow cell should be
washed with distilled water. This is possible every time even when analyser is in RUN mode.
• Select the “Wash” button from the main menu. The pump will drain out the water at high
speed from the flow cell.
• It is necessary to repeat the operation without water so that the water remaining inside
the analyser can be expelled completely.
5.2 Daily Maintenance
• Flow cell cleaning using distilled water or any recommended cleaning solution through
wash function.
• Select “Wash” key from the main menu for the flow cell washing
• The peristaltic pump will aspirate the distilled water/cleaning solution.
• Use wet cloth to remove the spillage if any on the instrument.
5.3 Periodical Maintenance
• Every week perform at least two complete washing cycle using DI Water.
• Wipe the external surface of the analyser every month with non-abrasive detergent using
soft sponge.
• Wash the waste bottle interior with clean water or disinfectant if necessary.
• In the case of keeping the Analyser standstill for more than two weeks keep the rubber tube
of the Peristatic pump outside of pump and while using the Analyser again connect rubber
tube in position with some silica gel as lubricant.
Page 32 of 35
[ADL/RDE/P016/044 -R00]
6. Messages
6.1 Error Messages
Connection Lost Communication Error/Network Turn off the main power and
Problem wait for 3 sec then turn on
the power switch, if the
problem remains, contact
service engineer.
Page 33 of 35
[ADL/RDE/P016/044 -R00]
7. Abbreviations
• CAL / Calib - Calibration
• ABS - Absorbance
• CONC - Concentration
• SD - Standard deviation
• CV - Coefficient of variation
• AVG - Average
• Vol - Volume
• Asp - Aspiration
• Deg - Degree
• PT ID - Patient ID
• Intrvl - Interval
• H - High
• L - Low
• N - Normal
• RES - Result
• STD -Standard
• Factor calculation for End point method
• Mean =
• SD =
Note: Any disposal of the equipment should be done based on the local regulation requirement.
Page 34 of 35
HAND BOOK
WRITING NOTES
35
MISPA PLUS
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
OFFICE COPY HAND BOOK
Name of Institution/Hospital/Laboratory:
Address:
2. Free from dust, mechanical vibrations, loud noises and electrical interference like
electrical motors and high-power rating equipment (hot air ovens)
YES NO
3. Direct sunlight or any other form of heat does not fall on the instrument YES NO
Pre-installation Check
Date & Signature
Done By
36
Applicable only for installation in India MISPA PLUS
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
OFFICE COPY HAND BOOK
INSTALLATION REPORT
CUSTOMER DETAILS
Installation Address:..........................................................................................................................
..................................................................................................................PIN:................................
EQUIPMENT DETAILS
Invoiced from AGAPPE / Distributor (If through distributor , please mention the name)
SITE DETAILS
Comments if any:..............................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
Declaration: The items as per the order have been delivered, installed, programmed and handed over in
38
Applicable only for installation in India MISPA PLUS
CUSTOMER COPY HAND BOOK
INSTALLATION REPORT
CUSTOMER DETAILS
Installation Address:..........................................................................................................................
..................................................................................................................PIN:................................
EQUIPMENT DETAILS
Invoiced from AGAPPE / Distributor (If through distributor , please mention the name)
SITE DETAILS
Comments if any:..............................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
Declaration: The items as per the order have been delivered, installed, programmed and handed over in
40
Applicable only for installation in India MISPA PLUS
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
OFFICE COPY HAND BOOK
“Agappe Hills”, Pattimattom (PO), Dist. Ernakulam, Kerala - 683 562, India.
TEL: + 91 484 2867000 | customersupport@agappe.in | www.agappe.com
Model:.....................................................................Serial Number:.......................................................
Purchased from:................................................................................................................................
Date of installation:................................................................................................................................
Customer name:.............................................................................................................
Address:............................................................................................................Pin:....................
Phone:..........................................Fax:.............................E-mail:.........................................................
1. WARRANTY CONDITIONS:
1.1. To activate the warranty, the customer must fill all the columns and
return the slip to AGAPPE.
1.2. AGAPPE, guarantees that all its instruments are free from
manufacturing defects or faults, as are the materials used. AGAPPE
undertakes to repair and substitute free of charge any part which may be 4. TERMINATION OF WARRANTY
found to have manufacturing defects. 4.1. This warranty shall be terminated at the end of the guarantee period
2.VALIDITY AND DURATION and also in the following cases:
2.1 The warranty will be valid and provided only if the original invoice or a. Where attempts to make repairs or alteration have been made by
sales receipt and the warranty card which must be properly completed and unauthorised person and/or with spare parts which are not originals.
indicating the date of purchase, model name, dealer's name, and serial b. Alteration have been made to the serial number of the product or of the
number of the product that are presented with the defective product within certificate.
the warranty period. c. The instrument is transferred to a new location without prior written
2.2. Agappe will refuse the Warranty if these documents are not presented or approval from ADL.
if they are incomplete. Moreover this warranty will not apply if the model 5. ELEMENTS NOT COVERED BY THE WARRANTY
name and the serial number on the product have been deleted or altered and 5.1. The following damage and faults are not covered by warranty.
if the date of the installation, model name, serial number and the dealers a. Damage deriving and/or originating from an insufficient or inadequate
name on the warranty card have been altered or deleted. electric circuit or from the power input in the area where the instrument
2.3. The warranty is valid for a period of 12 months from the date of is set up and used.
installation or 13 months from the date of purchase(billing), whichever is b. Breakdowns caused by careless handling, imprudence, lack of expertise
earlier. and in any case caused by lack of skill or any degree of negligence on the
3.WARRANTY AND CONDITIONS part of the operator.
3.1. Such parts of the instruments as may have defects due to materials or c. Damage defects and faults deriving from unexpected events, accidents
manufacture will be repaired and/or substituted free of charge. during transport by purchaser, due to FORCE MAJORE and in any case,
3.2. The warranty does not cover defects of parts which are subject to normal due to situation which can in no way be attributed to manufacturing
wear and tear and which are normally changed from time to time during and/or material defects.
maintenance procedures for the instrument. 5.2. Substitution of the defective instrument is never foreseen.
In any case, parts which are normally changed during maintenance 5.3. AGAPPE shall accept no responsibility whatsoever for damage either
procedures will also be substituted or repaired if they have manufacturing directly or indirectly to persons or materials from the use of the
and/or material defects such as to make it impossible to use the instrument instrument.
or such as to alter the proper functions of the instrument. 5.4. Consumable item such as fuses, lamps all tubings etc are not covered
3.3. Repairs and interventions carried out during the period of warranty do under this warranty condition.
not extend or renew said period of warranty. The repairs of the instrument 6. GOVERNING LAW
will be carried out by authorised technicians, on request by the customer to 6.1. The court of law of Ernakulam shall have exclusive jurisdictions for
the company. any controversies that may arise regarding this warranty certificate.
42
Applicable only for installation in India MISPA PLUS
CUSTOMER COPY HAND BOOK
“Agappe Hills”, Pattimattom (PO), Dist. Ernakulam, Kerala - 683 562, India.
TEL: + 91 484 2867000 | customersupport@agappe.in | www.agappe.com
Model:.....................................................................Serial Number:.......................................................
Purchased from:................................................................................................................................
Date of installation:................................................................................................................................
Customer name:.............................................................................................................
Address:............................................................................................................Pin:....................
Phone:..........................................Fax:.............................E-mail:.........................................................
1. WARRANTY CONDITIONS:
1.1. To activate the warranty, the customer must fill all the columns and
return the slip to AGAPPE.
1.2. AGAPPE, guarantees that all its instruments are free from
manufacturing defects or faults, as are the materials used. AGAPPE
undertakes to repair and substitute free of charge any part which may be 4. TERMINATION OF WARRANTY
found to have manufacturing defects. 4.1. This warranty shall be terminated at the end of the guarantee period
2.VALIDITY AND DURATION and also in the following cases:
2.1 The warranty will be valid and provided only if the original invoice or a. Where attempts to make repairs or alteration have been made by
sales receipt and the warranty card which must be properly completed and unauthorised person and/or with spare parts which are not originals.
indicating the date of purchase, model name, dealer's name, and serial b. Alteration have been made to the serial number of the product or of the
number of the product that are presented with the defective product within certificate.
the warranty period. c. The instrument is transferred to a new location without prior written
2.2. Agappe will refuse the Warranty if these documents are not presented or approval from ADL.
if they are incomplete. Moreover this warranty will not apply if the model 5. ELEMENTS NOT COVERED BY THE WARRANTY
name and the serial number on the product have been deleted or altered and 5.1. The following damage and faults are not covered by warranty.
if the date of the installation, model name, serial number and the dealers a. Damage deriving and/or originating from an insufficient or inadequate
name on the warranty card have been altered or deleted. electric circuit or from the power input in the area where the instrument
2.3. The warranty is valid for a period of 12 months from the date of is set up and used.
installation or 13 months from the date of purchase(billing), whichever is b. Breakdowns caused by careless handling, imprudence, lack of expertise
earlier. and in any case caused by lack of skill or any degree of negligence on the
3.WARRANTY AND CONDITIONS part of the operator.
3.1. Such parts of the instruments as may have defects due to materials or c. Damage defects and faults deriving from unexpected events, accidents
manufacture will be repaired and/or substituted free of charge. during transport by purchaser, due to FORCE MAJORE and in any case,
3.2. The warranty does not cover defects of parts which are subject to normal due to situation which can in no way be attributed to manufacturing
wear and tear and which are normally changed from time to time during and/or material defects.
maintenance procedures for the instrument. 5.2. Substitution of the defective instrument is never foreseen.
In any case, parts which are normally changed during maintenance 5.3. AGAPPE shall accept no responsibility whatsoever for damage either
procedures will also be substituted or repaired if they have manufacturing directly or indirectly to persons or materials from the use of the
and/or material defects such as to make it impossible to use the instrument instrument.
or such as to alter the proper functions of the instrument. 5.4. Consumable item such as fuses, lamps all tubings etc are not covered
3.3. Repairs and interventions carried out during the period of warranty do under this warranty condition.
not extend or renew said period of warranty. The repairs of the instrument 6. GOVERNING LAW
will be carried out by authorised technicians, on request by the customer to 6.1. The court of law of Ernakulam shall have exclusive jurisdictions for
the company. any controversies that may arise regarding this warranty certificate.
44
Applicable only for installation in India MISPA PLUS
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
46
MISPA PLUS
OFFICE COPY
MISPA PLUS
CUSTOMER COPY HAND BOOK
MISPA PLUS
The above maintenance is designed in an assumption that the water quality used is of type II as per international standards.
48
Applicable only for installation in India MISPA PLUS
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
50
OFFICE COPY
MISPA PLUS
MISPA PLUS
CUSTOMER COPY HAND BOOK
MISPA PLUS
The above maintenance is designed in an assumption that the water quality used is of type II as per international standards.
52
Applicable only for installation in India MISPA PLUS
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
54
MISPA PLUS
OFFICE COPY
MISPA PLUS
CUSTOMER COPY HAND BOOK
MISPA PLUS
The above maintenance is designed in an assumption that the water quality used is of type II as per international standards.
56
Applicable only for installation in India MISPA PLUS
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
58
OFFICE COPY
MISPA PLUS
MISPA PLUS
CUSTOMER COPY HAND BOOK
MISPA PLUS
The above maintenance is designed in an assumption that the water quality used is of type II as per international standards.
60
Applicable only for installation in India MISPA PLUS
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
OFFICE COPY HAND BOOK
E-Mail ID
Signature: HQ :
Date:
62
Applicable only for installation in India MISPA PLUS
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
OFFICE COPY HAND BOOK
Hardware Version
M echanical version
Peripheral Software
Version
LCD Software Version
S/ No
After Gain Calibration, From Filter Status,
Filter ADC Value Gain
340 nm
405 nm
505 nm
546 nm
578 nm
630 nm
From Tem perature Service Screen,
Actual Offset
37 0 C
30 0 C
25 0 C
From Pum p Calibration Screen
No of Steps
Control Value for any two available paramet ers
QC M AKE
LOT EXP
ITEM RANGE RESULT
64
Applicable only for installation in India MISPA PLUS
CUSTOMER COPY HAND BOOK
Hardware Version
M echanical version
Peripheral Software
Version
LCD Software Version
S/ No
After Gain Calibration, From Filter Status,
Filter ADC Value Gain
340 nm
405 nm
505 nm
546 nm
578 nm
630 nm
From Tem perature Service Screen,
Actual Offset
37 0 C
30 0 C
25 0 C
From Pum p Calibration Screen
No of Steps
Control Value
for any tw o available param eters
QC M AKE
LOT EXP
ITEM RANGE RESULT
66
Applicable only for installation in India MISPA PLUS
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
OFFICE COPY HAND BOOK
TRAINING REPORT
Person in Charge:
Operator:
Trainer Name:
Equipment details
Training Hours Trainer Sign/-
From: To:
Customer Name:
Signature:
Date:
68
Applicable only for installation in India MISPA PLUS
CUSTOMER COPY HAND BOOK
TRAINING REPORT
Person in Charge:
Operator:
Trainer Name:
Equipment details
Training Hours Trainer Sign/-
From: To:
Customer Name:
Signature:
Date:
70
Applicable only for installation in India MISPA PLUS
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
OFFICE COPY HAND BOOK
FEEDBACK FORM
S/ No: ................................................................................................................................................
6. Overall, how satisfied are you with the amount of communication between you and Agappe?
7. Would you recommend Agappe's services to colleagues or contacts within your industry?
72
Applicable only for installation in India MISPA PLUS
HAND BOOK
WRITING NOTES
74
MISPA PLUS
76
MISPA PLUS
ADL/MCD/REC/009/2020/R00 Company reserves the right to change any design and technical features of the product at any time, If needed.
Corp Address : “Agappe Hills”, Pattimattom (PO), Dist. Ernakulam, Kerala - 683 562, India.
www.agappe.com