Professional Documents
Culture Documents
915 BL
915 BL
The 915-BL Dual-Frequency Doppler is two Dopplers in one case. The low-frequency nominal 2 MHz
precordial probe may be used to detect the passage of air emboli in the heart. The high-frequency nominal
8-9 MHz pencil probe is used for systolic pressures at sites where a stethoscope is not used, as well as
to detect blood pressures that are too low for a stethoscope to auscultate. This probe can also be used to
listen for blood flow and pulses distal to arterial repair.
A built-in cautery suppressor with a controllable threshold shuts off the sound when the interference gets
too high.
Safety
This instrument is intended for use by health care professionals only.
The following symbols are used throughout this manual:
CAUTION: Indicates a potentially hazardous situation that, if not avoided, could result in personal
injury or damage to the instrument.
DANGEROUS VOLTAGE: Indicates a potential electrical hazard that, if not avoided, could result in
personal injury or damage to the instrument.
DUAL-FREQUENCY DOPPLER
MODEL 915-BL
1 0197
PARKS MEDICAL ELECTRONICS, INC.
ALOHA, OREGON U.S.A.
FLASHING LIGHT
INDICATES LOW
BATTERY TYPE B
APPLIED PART
MEDICAL EQUIPMENT
2 LOW
OW H
HIGH
VOLUME
ON
N OFF O
ON
11
3 SUP. ON
S 10
9
PROBES
SUP. OFF
SUP
CAUTERY SUPPRESSOR
24 V CHARGER HEADPHONE
HEADPHONES
4 2.1 MHZ 8.3 MHZ
0 63 A
0.63
KEEP ON CHARGE
HARGE WHEN NOT
OT IN USE
5 6 7
Front Panel
1. Manufacturer’s Label 7. Battery Charger Jack
Lists serial number (SN), part number, 24 V
0.63 A
The battery charger plugs into this jack.
model, and date of manufacture. 8. Headphone Jack
2. Battery Indicator Lamp Optional headphones plug into this jack.
This lamp lights when the unit is turned on. 9. Cautery Suppressor Control Knob
Blinking lamp warns battery needs charging. This knob controls the amount of cautery
3. Power Control Knob noise that is cut out.
The center position turns instrument off. The full counterclockwise position turns the
Turn switch to the left for 2 MHz probes. cautery suppressor off (SUP OFF).
Turn switch to the right for 8 MHz probes. Turning the knob clockwise cuts out more
4. 2.1-2.25 MHz (Specified) Probe Jacks noise.
The low-frequency precordial probe 10. Cautery Suppressor Indicator Lamp
plugs into these jacks. This indicator lamp is lit only when the
5. 8.0-9.9 (Specified) MHz Probe Jacks cautery suppressor is on (SUP ON).
The high-frequency pencil probes and flat 11. Volume Control Knob
probes plug into these jacks. This knob controls the Doppler sound level.
6. Battery Charger Indicator Lamp Turning the knob clockwise increases the
This lamp lights when the unit is charging. volume.
Flat Probes
Infant Adult
(an option with kit)
Frequency: 8.0-9.9 MHz (Specified)
Size: Adult 5/8 in X 3/4 in; Infant 1/2 in X 5/8 in
Cable length*: 5 ft standard
Crystals are set into the plastic so that the ultrasound beam goes into the vessel at about 15 degrees
from perpendicular. Flat probes are easily taped into place for taking repeated measurements.
*7 and 10 ft cable lengths are available by special order. Doppler must be tuned to probes with longer cables.
Battery Life
Battery life between charges for a new battery will be more than 16 hours on a full charge. The battery
life will decrease as the battery ages or if it is not kept fully charged.
With normal service and care, a battery can be expected to last two to three years. Leaving the battery
discharged for days will shorten the life of the battery.
Cautery Suppressor
Under normal operating conditions, the cautery suppressor is off.
Use of the Doppler in the presence of a cautery can cause interference. A cautery generates sound over
a wide band of frequencies that cannot be filtered by the Doppler. The frequencies close to the probe
frequency will be picked up by the probe wires and transmitted to the Doppler speaker. Experimenting
with placement of the probe cable and the Doppler can minimize cautery interference, but not eliminate
it. The cautery suppressor is designed to shut off sound to the Doppler speaker or headphones when the
volume of the cautery interference exceeds that of the blood flow or air emboli detected by the probe.
The control knob for the cautery suppressor is on the front panel of the Doppler. The cautery suppressor
is off when the cautery suppressor control knob is turned all the way counterclockwise. Turning the knob
clockwise increases the level of suppression.
Headphones
The speaker is disconnected when the optional low impedance stereo headphones are plugged in. You will
always hear more through the headphones, especially when checking weak flow or veins.
*See Diagnostic Procedures section for placement of precordial probe and flat probe.
The low-frequency precordial probe may be used to detect the passage of air emboli in the heart. The
high-frequency pencil probe is used for systolic pressures at sites where a stethoscope is not used, as
well as to detect blood pressures that are too low for a stethoscope to auscultate. This probe can also
be used to listen for blood flow and pulses distal to arterial repair. A built-in cautery suppressor with a
controllable threshold shuts off the sound when the interference gets too high.
Popliteal Artery
1. Arm, using highest brachial (or radial) artery reading, if ankle/brachial indices are indicated.
2. Ankle, using the doralis pedis or posterior tibial artery, whichever gives the higher reading.
3. Below knee (BK) or calf, using the doralis pedis or posterior tibial artery, whichever gives the higher
reading.
4. Above knee (AK), using popliteal artery if readings are difficult to obtain with dorsalis pedis or
posterior tibial artery.
5. High thigh (HT), using popliteal artery if readings are difficult to obtain with dorsalis pedis or
posterior tibial artery.
Venous Evaluation
Venous Doppler testing is the most subjective test done in the vascular laboratory. To provide consistent,
reproducible results, the technologist must be thoroughly familiar with venous anatomy (including
deviations from normal) and the subtleties of venous Doppler signals.
Veins are located by first finding the artery, and then moving the probe slightly to either side of the arterial
signal until the characteristic windstorm-like sound of venous flow is heard.
Doppler studies can be used to assess patients for the presence of occlusions, deep vein thrombosis
(DVT), and valvular incompetence associated with varicose veins. Doppler studies can detect
spontaneous venous flow, which is phasic with respiration, and augmentation with distal compression.
See standard textbooks for more information.
Tuning
The Doppler unit is tuned by the manufacturer to match the frequency of the probe, and the tuning of
the circuit is very stable. If sensitivity problems are suspected, do not attempt to adjust the instrument;
contact your sales representative or Parks Medical Electronics, Inc.
Routine Maintenance
The Doppler unit does not require regularly scheduled maintenance. The battery may need to be
replaced every two or three years. The probe jacks can wear out with extended use.
To Check the Battery:
1. Determine how long, at most, the Doppler is kept on continuously.
2. Recharge the battery for at least 8 hours.
3. Connect a dry probe to the unit, turn the unit on, and turn the volume 2/3 of the way up.
4. After one hour or twice the longest period of continuous use, whichever is greater, check the
battery indicator lamp. If the lamp is blinking, replace the battery.
To Check the Probe Jacks:
1. Bend back the four sections of the ground flange of the plug of a defunct probe.
2. Insert the center pin of the plug into each probe jack.
3. If there is no friction or resistance, replace the probe jack.
4. If you do not have a defunct probe, contact technical support.
Troubleshooting Guide
This table lists common problems and suggestions for troubleshooting them. If a problem persists after
these actions have been taken, contact Parks Medical Electronics, Inc.
Problem Situation Suggested Actions
(continued) (continued)
3.0 Hard to hear 3.1 Too much noise 3.1.4 Check for noisy probe:
blood sounds (poor 1. Clean and dry the crystal end of the probe.
signal: noise ratio) 2. Connect the dry probe to the Doppler unit, turn
the Doppler unit on, and turn the volume up.
3. Wiggle the probe plugs.
If you hear a crackling sound, the connection
is noisy.
4. Clean and tighten the probe plugs:
• Wipe the pin in the center of the plug and
the flange around it with contact cleaner or
isopropyl alcohol.
• Squeeze the flange with your fingers (or
gently with pliers). Make the flange form
a smaller circle so that it grips the jacks on
the Doppler panel more tightly.
5. Check for noise again.
6. If there is still noise, bend the cable here:
Carrying Case with Lid Aluminum case with handle, removable lid, and space for accessories.
(Optional)
Parks 915-BL UN Manual Operating manual with maintenance and service section.
Probes
Precordial Probe: Low-frequency specified 2.1-2.25 MHz; double-shielded cable; 3/4 in
diameter circular crystal; 5 ft standard cable length.
Standard Pencil Probe: High-frequency specified 8.0-9.9 MHz; 3/8 in diameter; 5 ft standard
OR cable length.
Adult Flat Probe: High-frequency specified 8.0-9.9 MHz; crystals transmit 3/8 in
beam width at 15 degrees from perpendicular; 5 ft standard cable length.
Environmental Conditions for Transport and Storage
Ambient Temperature: -40° C to 70° C
Relative Humidity: 10% to 100%, including condensation
Atmospheric Pressure: 500 hPa to 1060 hPa
Operating Conditions
Ambient Temperature: 10° C to 40° C
Relative Humidity: 30% to 75%
Atmospheric Pressure: 700 hPa to 1060 hPa
IPXO Rating: Degree of protection against ingress of water... none provided.
Implanted Devices
Implanted devices such as cardiac pacemakers should be avoided due to the possibility of affecting
their operation. Some plastics used in replacement surgery may be affected by absorption of ultrasonic
energy.
Environmental Hazards
Battery and Instrument
Batteries must be recycled or disposed of according to your local ordinances.
Do not dispose of old instruments or lead acid batteries in land fills.
Radio Interference
This instrument is intended for use by health care professionals only. This instrument may cause
radio interference or may disrupt the operation of nearby equipment. It may be necessary to take
precautionary measures, such as reorienting or relocating equipment, or shielding the location.
This Doppler may experience a high pitched tone or buzzing noise from radio interference caused by a
cell phone, mobile service, neon sign, an electrocautery, fluorescent lighting, or another Doppler.
Warranty
Parks Medical Electronics, Inc. warrants this Doppler against defects in materials and workmanship
for a period of one year, probes for six months. Parks will, at its discretion, replace or repair free of
charge, including labor, all parts which prove to be defective and subject to such warranty.
This warranty does not apply to any instrument or probe not used according to instructions or
damaged by abuse, accident, alteration, misuse, and/or tampering.
Headphones Optional low impedance stereo headphones with 1/4 in. plug.
Available from Parks Medical Electronics, Inc.
When ordering replacement parts, refer to the Instrument Information Section of this manual, where user
recorded the specifics about the Doppler.
Contact Information
Parks Medical Electronics, Inc.
Address
Mailing Address: PO Box 5669 Aloha OR 97006-0669 USA
Shipping Address: 19460 SW Shaw St Aloha OR 97007-1242 USA
Phone/Fax
Phone: 503-649-7007 1-800-547-6427
Fax: 503-591-9753
Web
e-mail: Info@parksmed.com
Web Site: www.parksmed.com
Parks’ products do not have adjustable output power. Our products are offered over the
frequency ranges of 3.9 - 5.3 MHz (low frequency), and 8.0 MHz - 9.9 MHz (high
frequency). The power output increases with frequency due to the design of the
ultrasound transmitter. Parks uses the same transmitter circuit design for all Doppler
products; therefore, this report applies to all of them.
To insure that the acoustical output power evaluation would show the absolute worse
case acoustical output parameters, Parks had the test done using a model 811-B tuned to
9.8 MHz. The output power was 75 mW, which translated to a maximum P of 20.8 mW
for the probe tuned to the unit. This exhibits an optimal coupling efficiency of 27% for
the crystal transducer.
ONDA report T-4PME02 shows that the maximum global output power was 675.3 mW.
The maximum allowed by the FDA is 720 mW. Parks Medical Electronics, Inc. is
exempt from acoustical output labeling for the US market. Parks has decided to provide
the output power for each probe type even though it is not required.
The uncertainties listed above and in the following tables are from a test report done by
Onda Labs for Parks’ Medical Electronics, Inc. The test report number is T-5PME01.
Leg distress⎯ cramps, fatigue, or just vague foot, in the heel area, or in the region of the and the superficial veins in the distal portion
pain⎯ often signal the presence of a peripheral metatarsophalangeal joint. The patient of the foot will fill in 10-15 seconds. If arterial
vascular disease. It could be arterial ⎯ the describes the pain as a severe ache or throbbing insufficiency is marked, the normal color may
first sign of arteriosclerosis obliterans. Simple which often wakes him after several hours of take 40-60 seconds to return; in severe cases
office evaluation, including the patient’s sleep. Relief is sometimes obtained by rubbing more than 2 minutes. Reactionary rubor of the
description of leg pain, and an examination of the affected foot or placing it in a dependent foot (a burgundy red color) may occur after the
the affected limbs, will often give a clear picture position over the edge of the bed. Elevating limb becomes dependent; this indicates the
of the underlying vascular problem⎯ where the limb often increases the pain. In some presence of advanced limb ischemia and often
it is, how extensive it is, which vessels are patients, relief is afforded by sleeping in a portends ulceration and “ gangrene of the foot.”
involved, and how adequate is the collateral chair with the leg in a dependent position. An absent or greatly diminished pulse is a
circulation. Leg pain and even “pseudoclaudication,” a diagnostic finding of major importance. The
A typical patient with arteriosclerosis form of leg pain that can mimic true presence of a pulse, however, does not always
obliterans, the most common of the arterial intermittent claudication, can occur with a indicate a normal arterial flow; its absence is
occlusive diseases, develops pain in one or variety of processes other than arteriosclerosis far more significant.
both legs that requires him to stop and rest obliterans such as degenerative disorders, In a patient who complains of intermittent
after walking a short distance. He may call it for example, osteoarthritis of the hip or claudication and has palpable resting pulses,
a cramp, a charley horse, or it may be just a spondylolysis. Compression of the cauda don’t discard the diagnosis of arteriosclerosis
feeling of tiredness in the limb after walking a equina also can produce claudication-like obliterans without investigating the
certain distance. An elderly patient often symptoms. In these patients, the typical cycle “disappearing pulse” phenomenon. Have the
admits to having calf pain for months or of exercise-pain-rest-relief is not present. patient exercise until he experiences
sometimes years with “no reason to mention The presence of normal pulses in the leg claudication; examination of the ankle pulses
it since aches and pains are bound to occur as and a normal ankle systolic blood pressure may then reveal an absent pulse associated
you grow older.” However, where he aids in distinguishing the pain of with leg pallor. The physiologic explanation
pseudoclaudication from arteriosclerosis is simple: exercise causes a marked dilation of
previously could walk 4-5 blocks before the
obliterans. the arteriolar beds within the exercising
calf pain occurred and made him stop, he’s
Examination of the affected foot yields a muscles. During exercise blood is shunted to
concerned because the pain now occurs after
great deal of information. Skin color, texture, these muscle groups causing a drop in the ankle
walking only a block or so.
and consistency all depend on arterial blood systolic blood pressure which is distal to the
Arterial occlusive disease due to
flow and can indicate the presence of an site of the arterial block. As the ankle systolic
arteriosclerosis obliterans, is insidious in onset
impaired circulation. blood pressure falls, the distal pulses
and often present in a patient for many years disappear.
If the patient has only mild, generally
before any ischernic symptoms occur. Rarely Except when life expectancy or surgical risk
asymptornatic occlusive arterial disease, the
will anyone complain of intermittent contravenes due to other systemic diseases,
color and nutrition of the leg and foot appears
claudication, the most commonest symptom angiography and surgical revascularization is
normal. As the ischemic process becomes
of arterial occlusive disease, while indoors. indicated for all patients with symptomatic
more severe, the skin appears shiny and
However, walking outdoors causes the pain arteriosclerosis obliterans exhibiting cutaneous
smooth, and hair is often absent from the toes
to occur. There is a typical pattern to this skin changes, such as ischernic ulcers or
and the dorsal region of the foot. Muscle
symptom of limb ischemia: exercise⎯ pain gangrene; the presence of rest pain or
atrophy, loss of subcuataneous fat, and pallor
⎯ rest⎯ relief. intermittent claudication that handicaps the
all indicate a severe degree of ischernia.
Patients characterize the pain of intermittent patient economically or socially.
A fairly simple and accurate clinical test that
claudication in various ways. One will Angiography is necessary to evaluate the
can measure the degree of arterial insufficiency
describe it as a sensation of cramping or extent of the arterial lesion. Typical arterial
in the affected leg is the elevation-dependency
tightness, “as if the leg is in a vise”. Another maneuver performed while the patient is on lesions due to arteriosclerosis obliterans are
will describe increasing fatigue, eventually the examining table. With the patient lying on segmental, occurring in areas of branching,
forcing the patient to stop walking and rest. his back, place his heels in the palms of your narrowing, or bifurcation sites of an artery.
However, in all of these patients, resting for a hands and elevate both legs 24-36 inches off The commonest occurring at the bifurcation
few minutes is sufficient to relieve the pain. the table. Hold the legs in this position for at of the aorta, or the iliac and femoral artery
If the need to sit down or elevate the extremity least 45-60 seconds. Observe the color of the divisions. Another frequent site is the distal
is a feature of a patient’s complaint, or if it feet and legs, particularly the soles of the feet. superficial femoral artery as it emerges from
takes more than a few minutes for the pain to If the arterial circulation is normal the skin the adductor canal of Hunter and the distal
abate, suspect a disease process other than coloration will decrease only slightly. Skin pallor popliteal artery as it branches into the anterior
arterial insufficiency as the cause of the pain. will develop in the affected limb according to and posterior tibial and peroneal arteries.
As the disease progresses, a different type the degree of arterial insufficiency. If pallor Certain patients are not candidates for
of pain occurs in the toes or heel. Termed rest occurs in both feet, suspect either an angiography; these include the elderly patient
pain or night pain, it is an ominous symptom arteriosclerotic block in the abdominal portion with moderate disability and associated cardiac
of advanced arterial occlusive disease due to of the aorta or similar blocks in the major limb disease, and generally any patient with the
multisegmental blocks in the major limb arteries. After the period of leg elevation, presence of another life-threatening disease. A
arteries and an inadequate collateral generally no more than several minutes, have satisfactory angiographic study provides a
circulation around these blocks. This pain the patient immediately stand up. In a patient visual study of the anatomic lesion, indicating
characteristically occurs in the distal portion with normal arterial circulation in the legs, color the extent of the lesion and to a degree its
of the foot, the toes, over the dorsum of the will return to the foot in 10 seconds or less, severity.
The angiogram, however, does not yield claudication. However, minor ischemic
information regarding the hemodynamic ulcerations may heal and rest pain may
significance of the lesion. For this information occasionally be relieved following
the Doppler flow detector is a useful and sympathectomy.
accurate instrument. The importance of the A limited number of patients with advanced
ankle/arm systolic blood pressure ratio in the arteriosclerosis obliterans may show
diagnosis and objective assessment of arterial improvement over periods of weeks from
disease underscores the value of the Doppler nonsurgical measures including rest, the use of
ultrasound blood flow detector. Although the vasodilators, and avoidance of tobacco smoking,
auscultatory method with a stethoscope is permitting the development of an additional
one of the most common blood pressure collateral circulation. However, the long period
measurements in clinical medicine, it is seldom before results can be evaluated is risky:
used in the lower exremity because it is gangrene may supervene, calling for emergency
difficult to obtain the Korotkoff sounds in surgical reconstruction of the occluded
the distal portion of a limb, especially when vascular bed if amputation is to be avoided.
arterial occlusive disease is present. The
Doppler flow detector, when used with a
sphygmomanometer and a pneumatic occlusion
cuff, can measure the lower extremity systolic
blood pressure easily and accurately.
Normally, the ankle systolic blood pressure is
slightly higher than the arm systolic blood
pressure. Any pressure gradient or difference CONCLUSION
that exists between the arm and ankle systolic There are both humanitarian, economic and
blood pressure provides a valuable, objective social reasons for mounting a strong effort to
hemodynamic assessment of the arterial lesion. salvage the lower limbs in patients, who are
Certainly, the lower the ankle systolic blood often elderly, and debilitated, but obvious
pressure when compared to that in the arm, candidates for a revascularization procedure.
the greater the gradient and therefore, the more The physical, psychological and economic
advanced the arterial occlusive lesion in the burdens following amputation preclude a useful
distal abdominal aorta or lower extremity or comfortable life in the years remaining to
arteries. them.
Two of the most important aspects of Limb salvage, the desideraturn of any age, is
managing patients with Arteriosclerosis particularly important in the geriatric patient
obliterans are the need to avoid any form of whose care then places a heavy burden on a
injury to the affected foot; this includes the spouse of similar age or another family member.
avoidance of any form of home surgery on the A patient confined to a wheelchair or bed is
feet, and the avoidance of tobacco smoking. A unable to attend to even the simplest personal
recent study of ours indicates that 40 percent needs; he often becomes withdrawn from
of major lower extremity amputations might human contacts, psychologically as well as
have been avoided by taking simple physically damaged.
precautions against infections stemming Partial limb revascularization, by saving the
from mechanical, thermal, or chemical extremely valuable knee joint, is often a
injuries to the feet. Meticulous, periodic reasonable alternative, allowing salvage of as
podiatric care is necessary for any foot with much as possible of a limb affected by end-stage
an impaired circulation with its known ischernia.
increased susceptibility to infection. Even The goal should be treatment of the disease
fungal infections such as athlete’s foot can put without loss of the lower limb or, if amputation
the ischernic foot at serious risk. Certain becomes necessary, a below-knee procedure or
over-the-counter pharmaceutical preparations even a transmetatarsal amputation, permitting
for removing corns and calluses that contain the patient to lead a more normal life.
caustic chemicals can be harmful to 222 East 19th Street
surrounding tissues, especially when used in New York, New York 10003
the presence of an impaired circulation and
should be avoided!
Thromboendarterectomy of the occluded
artery or the use of bypass grafts are the
operative procedures most frequently used for
revasculatization of the ischernic limb.
*Attending Vascular Surgeon, Cabrini Medical Center, New
Sympathectomy may be helpful in certain York, N.Y.; Associate Professor of Surgery, New York,
patients, but is is rarely considered for patients University School of Medicine, New York, N.Y.
who are candidates for arterial surgery and it is
not recommended for treatment of intermittent
Practical Office Technics for
Physiologic Vascular Testing*
COL CLYDE 0. HAGOOD, JR., MC, USAF, LTC DAVID J.
MOZERSKY, MC, USAF, and SSGT RANDAL N. TUMBLIN, BS, USAF, †
Lackland AFB, Tex
After the routine history and physical examination in the nature of the arterial signal can be
have been completed, the peripheral vessels are detected easily,5 after gaining some experience
surveyed with the Doppler ultrasonic velocity with the instrument.
detector. The quality of the arterial signals is An objective and reproducible determination of
evaluated at the common femoral, superficial the extent of the occlusive disease process can
femoral, popliteal posterior tibial, and dorsalis be made by measuring the systolic arterial
pedis arteries bilaterally. If the vessels are patent, pressure at the ankle. 7 Standard blood pressure
these signals will be composed of at least two, cuffs are placed around the ankle and the arterial
and often three, distinct sounds. The first is flow signal is monitored in the posterior tibial or
high-pitched and is distinctly separated from the dorsalis pedis artery (Fig 2). The cuff is then
second (Fig 1, A). inflated until the arterial signal is no longer
When mild proximal arterial stenosis is present, detected. With the probe still in place, the
there is a slight decrease in the frequency of the pressure is gradually lowered in the cuff until
first sound and the second sound is no longer flow signals are audible once again. The point at
detected (Fig 1, B). As the degree of occlusion which flow is re-established is the systolic
becomes progressively more severe and pressure at the level of the cuff. Under normal
collateralization develops, the signal becomes circumstances, the systolic pressure in the ankle
more monotonous with only small fluctuations should be equal to or above the systolic pressure
with each heart beat (Fig 1, C). These changes in the brachial artery. If the artery is occluded,
FIG I
Graphic representation of audible Doppler ultrasonic flow velocity signals obtained from the posterior tibial artery of three
separate patients. A - normal; B - obstructed; C - severely obstructed.
FIG 2
Technic of obtaining ankle systolic blood pressure. The angle of the probe in relation to the long axis of the posterior tibial
artery should be optimized to obtain the best signal.
the pressure will be lowered in the vessels distal immediately after the activity, using the method
to the occlusion. An arm-to-ankle pressure gradient previously described. The form of exercise which
of more than 20 to 30 mm Hg is suggestive of is selected will depend upon the interest,
occlusion at more than one level. It has been requirements, and resources of the physician. In
found that the ankle/arm pressure ratio correlates our clinic, all patients have been exercised on a
well with the severity of the obstructive process. 4 level, motorized treadmill at 3 mph to the point of
When the ratio is one or greater, it is unlikely that claudication, or a distance of 1,000 yards. This is
significant obstruction is present. Similarly, rest an indication of how far the subject can walk at a
pain is rare when the ankle/arm pressure ratio is marching pace on flat ground and is an objective
greater than 0.5. assessment of his complaints of claudication.
There are some patients in whom the ankle pressure Strandness and Bell 3 have found that by angling
is normal or nearly normal and who h a v e the treadmill at 12% grade and at a speed of 2
significant complaints of claudication. In these mph the amount of work involved is increased and
patients, exercise testing is an extremely valuable pressure gradients are accentuated. Normal
method of evaluation.8 subjects can walk five minutes with no difficulty
When the blood flow to a limb is increased by and no drop in ankle pressure. A patient propelled
exercise, the pressure gradient between the arm and treadmill may also be used with a metronome. The
ankle is accentuated. 2 In severe cases of obstructive patient is instructed to step each time the
vascular disease, the ankle systolic pressure may be metronome beats and, in this inexpensive way, the
unobtainable following exercise. Depending upon functional and provocative test of the patient’s
the extent and location of the disease and the type exercise capacity can be carried out.
of exercise, there may be a period of up to 20 to 30
minutes before the ankle pressure returns to normal Tw o l e s s e l a b o r a t e m e t h o d s o f e x e r c i s i n g
levels. The routine vascular examination should patients are available and may be used in office
include stressing the patient with some form of practice. Carter 8 has recently used active plantar
exercise and measuring the arm and ankle pressure and dorsiflexion of the foot against a
resistance provided by the attendant’s hand. Alternatively, Doppler survey and ankle pressure measurements not
the patient can be instructed to rapidly perform toe lifts only suggested the correct diagnosis preoperatively,
until fatigue or calf pain develops. These methods are but also confirmed the salutory effect of the surgical
less standardized than treadmill exercise, but they have procedure post-operatively. Patients with symptoms
the advantage of requiring no special equipment. suggestive of claudication and intact pulses may be
If the initial ankle systolic pressure after exercise is equal mistakenly treated for arthritis, neuritis, or emotional
to or higher than pre-exercise values, then the test is problems. As was shown in this case, the correct
normal and no further measurements are required. If the diagnosis can be quickly and accurately made, using
initial ankle pressure is low, the measurements are repeated simple testing procedures.
every minute for the next ten minutes. It should be Case 2. A 62-year-old man with many complicated
emphasized that the Doppler survey, a nkl e pre ssure medical problems came to our clinic with complaints
measurements, and exercise testing can be done in less of pain in the right calf and night pain. Pain in the
than 20 minutes by paramedical office personnel. calf was precipitated by walking less than 100 ft and
The following cases are presented to illustrate the relieved by rest. Physical examination of the right
usefulness of vascular testing methods in the clinical leg showed a normal pulse in the groin. No other
situation. pulses were palpable. The right brachial systolic blood
pressure was 206 mm. Hg. The right ankle systolic
Case Reports pressure was 78 turn Hg. Survey with the Doppler
C a s e 1 . A 5 2 - y e a r- o l d m a n c a m e t o t h e m e d i cal ultrasonic velocity detector showed a slightly abnormal
clinic complaining of pain involving the right leg, thigh, flow signal high in the right groin. This signal became
and buttock. The pain was precipitated by walking one high-pitched and continuous at a point about two inches
block and relieved by rest. Physical examination of below the inguinal ligament, and low-pitched,
the right leg showed 2+ femoral and dorsalis pedis pulses monotonous signals were noticed distal to that level.
and a weakly palpable posterior tibial pulse. No trophic He walked for 108 yards on the flat treadmill at 3
changes or temperature differences between the two legs mph. There was a 66.5% decrease in ankle systolic
were observed. It was the physician’s initial impression pressure after exercise. An arteriogram done by the
that the slightly diminished pulses probably were not Seldinger technic showed minimal decrease in the aorta
responsible for the symptoms, but the patient was and the right iliac artery. The superficial femoral artery
referred to the vascular surgery c l i n i c f o r f u r t h e r was occluded in Hunter’s canal. Close inspection of the
s t u d i e s a t t h a t t i m e . T h e s u r v e y with the Doppler bifurcation of the common femoral artery suggested
ultrasonic velocity detector revealed abnormal signals significant occlusive disease, involving the origin of the
at the right femoral, popliteal, dorsalis pedis, and deep femoral artery.
posterior tibial areas. The brachial systolic pressure was Due to the patient’s poor state of general health, the
112 turn Hg. The right ankle systolic pressure was 86 right groin was explored under local anesthesia. A large
mm. Hg. The patient was able to walk 115 yards at 3 occlusive plaque was located in the common femoral
mph on the treadmill. artery and a tight stenosis of the deep femoral orifice
These studies suggested the presence of severe occlusive was observed. A common femoral and deep femoral
atherosclerosis of the right iliac artery. Angiography endarterectomy were done. The patient was examined
confirmed a high-grade stenosis of the e n t i r e r i g h t in the laboratory three months after operation.
common iliac artery and its bifurcation. An aortoiliac Brachial systolic pressure was 132 mm. Hg. The right
endarterectomy was done subsequently. Postoperatively ankle systolic pressure was 70 mm. Hg. He walked 1,000
the patient was completely asymptomatic. Survey with y a r d s o n t h e f l a t t r e a d m i l l a t 3 m p h w i t h a 61%
the ultrasonic velocity detector showed normal flow in decrease in ankle pressure after exercise.
both lower extremities. The brachial systolic pressure Comment. The history and physical findings were not
was 140 mm Hg. The ankle systolic pressures were 137 compatible with an isolated, superficial femoral artery
turn Hg on the right and 140 mm Hg on the left. obstruction. Disabling claudication and night pain are
Comment. At the time the patient was initially seen in usually indicative of multiple arterial occlusions. The
the general medicine clinic, his complaints were 128 mm Hg pressure gradient between the arm and
suggestive of severe occlusive arterial disease. The fact ankle suggested that this was the case. Since a full
that pulses were palpable in all areas, however, was femoral pulse was palpated, and only a slightly abnormal
confusing to the physicians who first saw him. When he Doppler flow signal was heard in the upper common
was examined in the vascular laboratory using the femoral artery, it was thought that the o b s t r u c t e d
Doppler ultrasonic velocity detector the lesion was arteries were located in the thigh or calf or both.
quickly localized to the right iliac artery. The 34 The presence of arterial signals in the popliteal artery
turn Hg arm/ankle gradient demonstrated the severity indicated its patency. Doppler examination of the
of the problem, and the functional disability was dorsalis pedis and posterior tibial arteries showed arterial
confirmed by his performance on the treadmill. Since the flow signals similar to those obtained in the popliteal
patient’s job required a great deal of walking he was artery. This suggested no significant obstructive lesion
essentially disabled by his condition. On the basis of between these levels. On the basis of these findings,
the objective tests, proper, diagnosis and treatment t h e o b s t r u c t i v e l e s i o n s could be localized to the
were begun. superficial and deep femoral arteries.
FIG 3 Conclusions
Palpation of pulses and inspection of the limb are
subjective methods of evaluating complaints that
may be related to arterial insufficiency. Since the
physician whose practice includes patients with
many varying problems may feel insecure about
his ability to evaluate pulses, objective testing
methods have a real place in the diagnostic
armamentarium. The Doppler ultrasonic velocity
detector may be thought of in the same manner as
one thinks of a stethoscope, as an instrument
which extends clinical perception. It is a rapid
and safe tool which can provide relevant, and at
times indispensable, information.
Measurement of ankle systolic pressure by the
Doppler ultrasonic technic has proven to be an
objective and repeatable test which correlates well
Angiogram showing a large plaque in the common femoral with the anatomic situation. Exercise testing
artery involving the orifice of the profunda. provides both a functional assessment of the
patient’s complaints and a diagnostic challenge by
Results of angiography confirmed the superficial femoral which small, resting pressure gradients can be
artery obstruction and suggested severe stenosis of the deep accentuated. Resting pressure measurement,
femoral orifice (Fig 3). Since the hemodynamically
exercise testing, and Doppler ultrasonic survey are
significant lesions were confined to the thigh, it seemed
logical to increase the collateral circulation distal to extremely simple methods of evaluation by which
the occluded superficial femoral artery. Therefore, even minor degrees of occlusive arterial disease
an endarterectomy of the distal common femoral may be detected. The equipment necessary to
artery and deep femoral artery orifice was done. evaluate vascular problems can be purchased for
It is significant to note that there were no changes less than $1,000.
in the patient’s physical examination after operation
and, without vascular testing, only his testimony could
be used to indicate a beneficial result. The preopera-
tive ankle/arm ratio was 0.38. After operation, it
r o s e t o 0 . 5 3 . I n a d d i t i o n , t h e p a t i e n t ’s e x e r c i s e
tolerance increased nearly tenfold. These concrete
data should be used as the criteria for success or
failure.
C a s e 3 . A 5 4 - y e a r- o l d d e n t i s t w a s s e e n o n e y e a r
after bypass graft for severe aortoiliac occlusive
disease. Three months after operation, he was References
evaluated in the laboratory. Brachial systolic pressure 1 . Satomura S: Ultrasonic blood rheograph. J Acoust Soc
was 108 mm Hg. Ankle systolic pressures were 114 Japan 15:151-154
m m H g o n t h e r i g h t a n d 111 m m H g o n t h e l e f t . 2 . Winsor T, Hyman C, Payne JH: Exercise and limb circulation
Both ankle pressures increased after exercise. There in health and disease. Arch Surg 78:184-192
3 . Strandness DE Jr, Bell JW: An evaluation of the hemo-
were no signs or symptoms of ischemia. He was dynamic response of the claudicating extremity to exercise.
studied again eight m o n t h s l a t e r. H i s p u l s e s w e r e Surg Gynecol Obstet 119:1237-1242
intact. The arm pressure was 114 min Hg, and the 4 . Yao ST, Hobbs IT, Irvine WT: Ankle systolic pressure
measurements for arterial disease affecting the lower
ankle pressures were 126 mm Hg on the right and 96 extremities. Br J Surg 56:676-679
mm Hg on the left. He walked 1000 yards on the flat 5 . Strandness DE Jr, Schultz RD, Sumner DS, et al: UItrasonic
treadmill at 3 mph with no symptoms. There was a flow detection: a useful technique in the evaluation of
peripheral vascular disease. Am J Surg 113:311-320.
24% decrease in the left ankle pressure after exercise.
6 . Wi l d J J , R e i d J M : E f f e c t o f u l t r a s o u n d o n biological
Comment. This patient, though asymptomatic, is tissues. J Acoust Soc Am 25:270-280
exhibiting disease progression, which will require closer 7. Strandness DE Jr: Management of Arterial Occlusive
Disease. Chicago, Year Book Publishers, Inc.
follow-tip. This disease progression could not have 8 . Carter SA: Response of ankle systolic pressure to leg
been detected on a clinical basis, but was readily exercise in mild or questionable arterial disease.
detectable using simple vascular testing technics. N Engl J Med 287:578-582