Download as pdf or txt
Download as pdf or txt
You are on page 1of 4

T 264 METHODS OF SAMPLING AND TESTING

Standard Method of Test for

Collection and Preservation of Water Samples

AASHTO DESIGNATION: T 264-78

1. SCOPE vary from as little as 5 minutes to as long hardness. Use time-composite samples
as 1 hour or more. only for determining components that can
When the composition of a source varies be demonstrated to remain unchanged
1.1 The objective of sampling is to under the existing conditions of sample
collect a portion of material small enough in space rather than time, a set of samples
collected from appropriate locations with collection and preservation.
in volume to be conveniently transported
to and handled in the laboratory while less emphasis on timing may provide the Take individual portions in a widemouth
still accurately representing the material most useful information. bottle having a diameter of at least 35
being sampled. This implies, first, that Use great care in sampling wastewater mm at the mouth and capacity of at least
the relative portions or concentrations of sludges, sludge banks, and muds. No 120 mL. Collect these portions each
all pertinent components must be the definite procedure can be given, but hour-in some cases half hour or even
same in the sample as in the material every possible precaution should be taken every 5 minutes-and mix at the end of the
being sampled, and second, that the to obtain a representative sample. sampling period or combine in a single
sample must be handled in such a way bottle as collected. If preservatives are
2.2 Composite Samples-In most cases, used, add them to the sample bottle
that no significant changes in the term composite sample refers to a
composition occur before the tests are initially so that all portions of the
mixture of grab samples collected at the composite are preserved as soon as
performed. Complete and unequivocal same sampling point at different times.
preservation of samples, either domestic collected. Analysis of individual samples
Sometimes the term time-composite is may sometimes be necessary.
sewage, industrial waste, or natural used when it is necessary to distinguish
waters, is a practical impossibility. this type of sample from others. Time- It is desirable, and often absolutely
Regardless of the nature of the sample, composite samples are most useful for essential, to combine the individual
complete stability for every constituent observing average concentrations that are samples in volumes proportional to the
can never be achieved. At best, used, for example, in calculating the volume of flow. A final volume of 2 to 3
preservation techniques can only retard loading or the efficiency of a wastewater liters is sufficient for sewage, effluents,
the chemical and biological changes that treatment plant. As an alternative to the and wastes.
inevitably continue after the sample is separate analysis of a large number of Automatic sampling devices are available
removed from the parent source. samples, followed by computation of but should not be used unless the sample
1.2 The values stated in SI units are to be average and total results, composite is preserved as described below. Clean
regarded as the standard. samples of this type represent a sampling devices, including bottles, daily
substantial saving in laboratory effort and to eliminate biological growths and other
expense. For these purposes, a composite deposits.
2. TYPES OF SAMPLES sample representing a 24-hour period is 2.3 Integrated Samples-For certain
considered standard for most determina- purposes, the information needed is
2.1 Grab or Catch Samples-Strictly tions. Under certain circumstances, provided best by analysis of mixtures of
speaking, a sample collected at a however, a composite sample represen- grab samples collected from different
particular time and place can represent ting one shift, or a shorter time period, or points simultaneously, or as nearly so as
only the composition of the source at that a complete cycle of a periodic operation, possible. Such mixtures sometimes are
time and place. However, when a source may be preferable. Evaluation of the called integrated samples. An example of
is known to be fairly constant in effects of special, variable, or irregular the need for such sampling occurs in a
composition over a considerable period discharges and operations may require river or stream that varies in composition
of time or over substantial distances irr composite samples representing the across its width and depth. For evaluation
all directions, then the sample may be period during which such discharges of average composition or total loading, a
said to represent a longer time period or a occurs. mixture of samples representing various
larger volume, or both, than the specific For determination of components or points in the cross-section, in proportion
point at which it was collected. In such characteristics subject to significant and to their relative flows, may be useful. The
circumstances, some sources may be unavoidable changes on storage, need for integrated samples also may
quite well represented by single grab composite samples cannot be used. exist if combined treatment is proposed
samples. Examples are some water Perform such determinations on for several separate wastewater streams,
supplies, some surface waters, and rarely individual samples as soon as possible the interaction of which may have a
some wastewater streams. after collection and preferably at the significant effect on treatability or even
When a source is known to vary with sampling point. Analyses for all the composition of the mixture.
time, grab samples collected at suitable dissolved gases, residual chlorine, soluble Mathematical prediction of the
intervals and analyzed separately can be sulfide, temperature, and pH are interactions may be inaccurate or
of great value in documenting the extent, examples of determinations of this type. impossible and testing of a suitable
frequency, and duration of these Changes in such components as dissolved integrated sample may provide more
variations. Choose sampling intervals on oxygen or carbon dioxide, pH, or useful information.
the basis of the frequency with which temperature may produce secondary Both natural and artificial lakes often
changes may be expected, which may changes in certain inorganic components shown variations of composition with
such as iron, manganese, alkalinity, or
592
T 264 METHODS OF SAMPLING AND TESTING

both depth and horizontal location. for streams not likely to be uniformly them to stand full of distilled water for several
However, under most conditions, neither mixed, more samples are desirable and days. Conditioning may be hastened by a
total nor average figures are especially are usually taken at a number of points at preliminary treatment with dilute hydrochloric
acid solution.
significant in these situations. The local the surface across the entire width and at
variations are of more importance, and a number of depths at each point. Take
the samples are examined separately care, when boats are used, to avoid 5.2 The closures for the sample
rather than integrated. collecting samples where the turbulence containers shall be glass stoppers that
The preparation of integrated samples caused by a propeller or by oars has have been thoroughly washed, or plastic
usually requires special equipment to disturbed the characteristics of the water. caps with suitable liners.
collect a sample from a known depth, Ordinarily, samples are taken at these 5.3 Refer to Table 1 for the
without contamination by the overlying points and then combined to obtain an recommended container for individual
water. Prior knowledge about the integrated sample of such a stream of parameters.
volume, movement, and composition of water. Alternatively, test the single grab
the various parts of the water being samples, for example, to determine the
sampled usually is required. Therefore, point of highest bacterial density. 6. GENERAL PRECAUTIONS
the collection of the integrated samples 4.3 Choose the location of the sampling
becomes a complicated and specialized point with respect to the information 6.1 For sampling of unconfined water at
process that cannot be described in desired and in conformity to local condi- any specified depth in ponds, lagoons,
complete detail here. tions. Allow sufficient distance downs- reservoirs, etc., during which contact
tream, with respect to stream flow at the with air or agitation of the water would
time of sampling, from a tributary or cause a change in concentration of
3. FREQUENCY AND DURATION source of pollution to permit thorough
OF SAMPLING characteristics of a constituent to be
mixing. If this is not possible, it is better determined, use a sampling apparatus so
to sample the stream above the tributary constructed that the solution at the depth
3.1 A reasonably accurate estimate of the or source of pollution and, in addition, to to be sampled flows through a tube to the
composition of a raw water piped from a sample the tributary or source of pollu- bottom of the container, and that a
large body of water, such as the Great tion. In general, a distance of 1.5 to 5 km volume of sample equal to 4 to 10 times
Lakes, far enough from the shoreline to (1 to 3 miles) below the tributary is the volume of the receiving container
avoid variation from inflowing tributaries sufficient. passes through it. When no determi-
and waste discharges, may be made by 4.4 Collect samples at least 0.8 km (0.5 nations of dissolved gases are to be made,
taking individual samples at infrequent miles) below dams or waterfalls to allow any less complicated apparatus may be
intervals, such as biweekly or monthly, time for the escape of entrained air. used that will permit the collection of a
sufficient to cover seasonal changes. If When lakes, reservoirs, or other bodies of sample at a desired depth, or of an
samples are taken from near the shoreline water are sampled, it is necessary to integrated sample containing water from
of such a body of water or from a river, avoid non-representative areas such as all points in a vertical section.
take them at shorter intervals, for instan- those created by inlet streams, more 6.2 Some test methods require adjust-
ce daily, to provide more exact know- stagnant areas, or abrupt changes in ment to other than ambient temperature.
ledge of the variations in composi-tion shorelines, unless determining the effect Such adjustment should be carried out
where these are of importance in the use of such local conditions is a part of the when indicated.
to which the water is to be put. If greater sampling program.
variations or cycles of pollution occur or 6.3 Normally, samples are taken without
4.5 It is desirable to take a series of separation of particulate matter. If cons-
closer surveillance of plant intake water samples from any source of water to
is required, collect more frequent tituents are present in colloidal or
determine whether differences in flocculent suspension take the sample so
samples, for example, at hourly intervals. composition are likely to exist before that they are present in representative
3.2 Water undergoing continuous or final selection of the sampling point. portion.
intermittent treatment must be sampled
with such frequency that adequate control 6.4 Volume of sample:
is assured. The interval between samples 5. SAMPLE CONTAINERS 6.4.1 Collect a minimum volume of 2
is directly related to the rate at which liters; 4 liters is preferable.
critical characteristics can reach 5.1 Sample containers shall be made of 6.4.2 The estimates in Table 1 cover
intolerable limits. materials that will not contaminate the the volume of sample required for the
sample and, before use, shall be cleaned usual determinations that are made on a
4. POINT OF SAMPLING thoroughly to remove all extraneous water sample as well as for several tests
surface dirt. Chemically resistant glass that are made for special purposes.
and polyethylene are suitable materials 6.4.3 The sample volume required for
4.1 Choose the point of sampling with for the containers. Only polyethylene bacteriological analysis varies greatly
extreme care so that a representative containers shall be used for samples in with the bacterial density of the water.
sample of the water to be tested is which small amounts of hardness, silica, Serial dilutions are necessary for high
obtained. Avoid surface scum. sodium, or potassium are to be density samples.
4.2 Because of a wide variety of con- determined. The collection, storage, and
subsequent analytical determination in 6.5 When samples are to be shipped, do
ditions found in streams, lakes, reservo- not fill the bottle entirely in order to
irs, and other bodies of water, it is not plastic containers will result in erroneous
pH values due to the permeability of allow some room for expansion when
possible to prescribe the exact point of subjected to a change in temperature. An
sampling. Where the water in a stream is plastic to gases such as CO2.
air space of 10 to 25 mL usually suffices
mixed so as to approach uniformity, a for this purpose, although this does not
sample taken at any point in the cross NOTE-New chemically resistant glass protect against bursting of the container
section is satisfactory. For large rivers or containers shall be conditioned by allowing due to freezing.
593
T 264 METHODS OF SAMPLING AND TESTING

6.6 If contact with air would cause a TABLE 1 Recommendation for Sampling and Preservation of Samples According to
change in the concentration or Measurement6
Volume
characteristics of a constituent to be Measurement
Req., mL
Container Preservative Holding Time°
determined, secure the sample without Acidity 100 P,G2 Cool, 4°C 24 hours
contact with air and completely fill the Alkalinity 100 P, G Cool, 4°C 24 hours
container. Arsenic 100 P, G HNO3 to pH < 2 6 months
BOD 1000 P. G Cool, 4°C 6 hours'
6.7 Microbiological samples: Bromide 100 P, G Cool, 4°C 24 hours
6.7.1 When taking a sample from a COD 50 P. G H2SO4 to pH < 2 7 days
Chloride 50 P, G None required 7 days
sample line or tap, allow the water to run
Chlorine req. 50 P, G Determine on site No holding
for at least 5 minutes or long enough to Color 50 P, G Cool, 4°C 24 hours
flush, with 6 to 10 times its volume, the Cyanides 500 P, G Cool, 4°C 24 hours
entire part of the system that has been NaOh to pH<2
stagnant for 2 hours or more. Dissolved oxygen:
Probe 300 G only Determine on site No holding
6.7.2 Turn off the sample outlet and Winkler 300 G only Fix on site 4 to 8 hours
empty it of water without touching the Fluoride 300 P. G Cool, 4°C 7 days
inside. Flame the outlet with a suitable Hardness 100 P. G Cool, 4°C 7 days
HNO3 to pH < 2
torch or other devise that will avoid the Iodide 100 P, G Cool, 4°C 24 hours
deposition of soot, which is undesirable. MBAS 250 PG Cool, 4°C 24 hours
This flaming procedure may be omitted Metals:
if the outlet is carefully cleaned by use of Dissolved 200 PG Filter on site 6 months
HNO3 to pH < 2
a swab of cotton or other suitable Suspended Filter on site 6 months
material thoroughly saturated with Total 100 HNO3 to pH < 2 6 months
denatured ethyl alcohol (70 percent). Mercury:
Dissolved 100 P. G Filter 38 days (glass)
6.7.3 Choose a sterile sample bottle HNO to pH < 2 13 days (hard plastic)
containing Na2S2O3 if the water being Total 100 P. G HNO3 to pH 2 38 days (glass)
sampled contains residual chlorine, has 13 days (hard plastic)
been chlorinated, or contains any free or Nitrogen:
Ammonia 400 P. G Cool. 4°C 24 hours4
combined available oxidizing agent H2SO4 to pH < 2
intended to sterilize it. If such sterilizing Kjeldahl 500 P. G Cool, 4°C 7 days
agents are not present, the thiosulfate Total H2SO4 to pH < 2
may be omitted. In cases where Nitrate 100 P. G Cool, 4°C 24 hours4
thiosulfate interferes with subsequent H2SO4 to pH < 2
Nitrite 50 P. G Cool, 4°C 24 hours4
examination, such as examination for NTA 50 P. G Cool, 4°C 24 hours
sulfatereducing bacteria, etc., omit the Oil and grease 1000 G only Cool, 4°C 24 hours
use of thiosulfate in the sample bottle H2SO4 to pH < 2 or HCl
even if such sterilizing agents are Organic carbon 25 P. G Cool, 4°C 24 hours
H2SO4 to pH < 2
present. In this case, if such sterilizing pH 25 P. G Cool, 4°C Determine on site 6 hours4
agents are present, the examination must Phenolics 500 G only Cool, 4°C 24 hours
be performed as soon as possible. H3PO4 to pH < 4
1. O g CUSO4/I
6.7.4 Remove the stopper from the Phosphorous:
sample bottle. Grasp the stopper by the Orthophosphate, dissolved 50 P. G Filter on site 24 hours4
dust cover so as not to contaminate it by Cool, 4°C
touching it; do not lay it down. Hold the Hydrolyzable 50 P, G Cool. 4°C 24 hours4
Total 50 P, G Cool, 4°C 7 Days
bottle by the bottom to avoid touching H2SO4 to pH < 2
the neck. Do not rinse the bottle with the Total, dissolved 50 P, G Filter on site 24 hours'
sample. Quickly hold the bottle under the Cool, 4°C
flowing water to be sampled until it is Residue:
Filterable 100 PG Cool, 4°C 7 days
about three-fourths full to permit mixing Nonfilterable 100 PG Cool, 4°C 7 days
by shaking prior to testing. Replace the Total 100 PG Cool, 4°C 7 days
stopper and promptly crimp the metal Volatile 100 PG Cool, 4°C 7 days
dust cover in place over the neck of the Settleable matter 1000 PG None required 24 hours
bottle. Take care that the stopper and Selenium 50 PG HNO3 to pH < 2 6 months
Silica 50 P only Cool, 4°C 7 days
bottle neck are not touched during this Specific conductance 100 PG Cool, 4°C 24 hours5
operation and that no dust blows into the Sulfate 50 PG Cool, 4°C 7 days
bottle, insofar as possible. Sulfide 500 P.G 2 mL zinc acetate 24 hours
Sulfite 50 PG Determine on site No holding
Temperature 1000 PG Determine on site No holding
7. PRESERVATION OF SAMPLES Treshold odor 200 G only Cool, 4°C 24 hours
Turbidity 100 P.G Cool, 4°C 7 days
1 More specific instructions for preservation and sampling are found with each procedure as detailed in the
7.1 Add chemical preservatives to appropriate reference.
2 Plastic or glass.
samples for chemical, physical, or 3 If samples cannot be returned to the laboratory in less than 6 hours and holding time exceeds this limit, the
radiological examination only as final reported data should indicate the actual holding time.
specified in Table 1. Quick freezing has 4 Mercuric chloride may be used as an alternate preservative at a concentration of 40 mIJL, especially if a
longer holding time is required. However, the use of mercuric chloride is discouraged whenever possible.
been found to be beneficial in preserving 5 If the sample is stabilized by cooling. it should be warmed to 25°C for reading, or temperature correction
some organic constituents. Note any made and results reported at 25°C.
preservatives added on the label. 6 It has been shown that samples properly preser, ed may be held for extended periods beyond the
recommended holding time.

594
T 264 METHODS OF SAMPLING AND TESTING

7.2 Samples for bacteriological 8.2 On the statement of an analysis, Type and quantity of preservative
examination require refrigeration if they specify the length of time elapsed added,
cannot be processed within 1 hour after between collection and analysis of the
sample. Results of field tests made on the
collection. sample,
7.3 If the sample is to be examined 8.3 Make the determination of dissolved
gases, such as oxygen, hydrogen sulfide, Signature of sampler.
within 1 hour of collection, store it in a
cool place without icing. and carbon dioxide at the source, except 9.2 Fix the stoppers closing the sample
that in some cases such constituents may containers in place by wire, tape, or cord
7.4 If the sample is held more than 1 be fixed and determined later as specified to prevent leakage in transit. The sample
hour before examination, store it in a in the specific test methods. containers shall be of such size that when
refrigerator or ice chest at a temperature filled with the desired amount of sample,
of not more than 4°C. The interval 8.4 When sampling for radioactivity
determinations, note the exact time of space roughly equivalent to 1 percent of
between collection and examination of the volumetric capacity of the containers
the sample must in no case be more than sample collection. If short-lived activity
is of interest, analysis should be made as will be available for expansion of the
12 hours, or 6 hours if the sample is liquid. Exceptions to this are for those
suspected to contain large numbers of rapidly as practical to minimize loss of
activity by radioactive decay. If only constituents noted in Section 8.3.
organisms. Field examination procedures
should be considered for longer periods. long-lived activity is of interest, 9.3 The sample shipping container shall
measurement of the radioactivity be a case (wood preferred) having a
7.5 If the sample is transported, ship it in sometimes can be simplified by allowing separate compartment for each sample
an insulated, iced container to maintain sufficient time before analysis for the container. Line the compartment around
the temperature between 0 and 4°C to decay of the short-lived radionuclides. each sample container with corrugated
allow it to be examined within 12 hours paper, felt, or similar material; or hold
of collection. the sample containers in place with
7.6 If the conditions in Section 7.4 or 7.5 9. LABELING AND TRANSPORTA- spring clips, sawdust, excelsior, foamed
cannot be met, note the actual conditions TION OF SAMPLES plastic, or similar material. Use insulated
on the examination report. containers for quick-frozen samples,
9.1 Provide space for the following usually shipped with solid carbon dioxide
information on an etched area of the to maintain the sample in frozen
8. TIME INTERVAL BETWEEN condition.
COLLECTION AND ANALYSIS bottle, a gummed label, or a cardboard or
OF SAMPLES linen tag securely affixed to the 9.4 Print the addresses of consignee and
container: consignor plainly upon two sides of the
Sample number, outer container, or attach firmly thereon
8.1 In general, allow as short a time as by cards or labels. Attach warning and
possible to elapse between the collection Date and time of sampling, Source of descriptive labels to the outer container
of a sample and its analysis. Under some sample, such as "Fragile," "Liquid," "Glass,"
conditions, analysis in the field is Point of sampling (designated in "Handle With Care," "This Side Up,"
necessary to secure reliable results. The sufficient detail to enable anyone to etc., when applicable. In cold weather,
actual time that may be allowed to collect a second sample from the attach the label "Keep from Freezing" to
intervene between the collection and identical spot rom which the first sample the outer container, except for those
analysis of a sample varies with the type was taken), samples which are intentionally frozen.
of examination to be conducted, the Temperature and rate of flow of the
character of the sample, and the time fluid in the equipment from which the
interval allowable for applying corrective sample as taken,
treatment.
Temperature of sample,

595

You might also like