Download as pdf or txt
Download as pdf or txt
You are on page 1of 42

h

c
rd
F
s
n
m
ie
S

SIREMOBIL Compact / L
SP

Maintenance Instructions
System

Compact L and Compact, serial number 02001 and


higher

The protocol SPR2-135.832.01.02.02 is required for these


instructions
10409105
03780629
03776494

© Siemens 2008
The reproduction, transmission or use
of this document or its contents is not
permitted without express written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.

Print No.: SPR2-135.831.01.02.02 English


Replaces: SPR2-135.831.01.01.02 Doc. Gen. Date: 05.09
2 Revision / Disclaimer
1Revision / Disclaimer

Document revision level


The document corresponds to the version/revision level effective at the time of system
delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.

Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise autho-
rized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with
or authorized by Siemens or one of its affiliates are directed to contact one of the local
offices of Siemens or one of its affiliates before attempting installation or service proce-
dures.

SIREMOBIL Compact / L SPR2-135.831.01.02.02 Page 2 of 42 Siemens


05.09 CS PS SP
Table of Contents 3

0 Table of Contents
1 _______ General Information______________________________________________ 5

Prerequisites. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Required documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Systems equipped with SIMOMED N X 2080 monitor . . . . . . . . . . . . . . . . . . . . . . . . 6
Systems equipped with the standard 100/120Hz monitor. . . . . . . . . . . . . . . . . . . . . . 6
Systems equipped with TFT displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Systems equipped with an external I.I. laser light localizer. . . . . . . . . . . . . . . . . . . . . 6
For systems equipped with an integrated I.I. laser light localizer . . . . . . . . . . . . . . . . 6
When an integrated laser light localizer is present (close to tube) . . . . . . . . . . . . . . . 7
When DICOM bridge is installed. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
When Codonics printer is present. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Required tools, measurement and auxiliary devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Spare parts which may be needed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Text emphasis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
General safety information (in existing documents) . . . . . . . . . . . . . . . . . . . . . . . . . 12
General electrical safety information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Radiation safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Mechanical safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Safety information - risk of infection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Integrated I.I. laser light localizer option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
I.I. laser light localizer option or single tank laser light localizer option . . . . . . . . . . . 15
Information on the protective conductor resistance test . . . . . . . . . . . . . . . . . . . . . . . . . 16
System leakage current measurement information . . . . . . . . . . . . . . . . . . . . . . . . . 17
Explanation of abbreviations in the maintenance protocol . . . . . . . . . . . . . . . . . . . . . . . 21
Maintenance interval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

2 _______ Inspection of exterior and surroundings____________________________ 23

Visual inspection of exterior . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23


Inspection of surroundings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Power outlets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

3 _______ Safety inspection _______________________________________________ 25

Mechanical safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
External I.I. laser light localizer (if present) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Integrated I.I. laser light localizer (if present) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Integrated laser light localizer, close to tube - (if present). . . . . . . . . . . . . . . . . . . . . 26
DICOM bridge (if present). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Codonics printer (if present) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Monitor(s) / displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Electrical safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Monitor(s) / displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30

Siemens SPR2-135.831.01.02.02 Page 3 of 42 SIREMOBIL Compact / L


05.09 CS PS SP
4 Table of Contents

Conductive rubber. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32

4 _______ Maintenance, operating values and funct. inspect. ___________________ 33

Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
DICOM bridge (if present) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Codonics printer (if present) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Operating values, inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Functional inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36

5 _______ Final result / quality inspect. & general maint. _______________________ 39

Final work steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40

6 _______ Changes to Previous Version _____________________________________ 41

SIREMOBIL Compact / L SPR2-135.831.01.02.02 Page 4 of 42 Siemens


05.09 CS PS SP
General Information 5
1-
1 General Information

1.1 Prerequisites
The requirements described in chapter 1 of the Service Instructions also apply to mainte-
nance.

Siemens SPR2-135.831.01.02.02 Page 5 of 42 SIREMOBIL Compact / L


05.09 CS PS SP
6 General Information

1.2 Required documents

• Safety information according to ARTD, part 2


• Maintenance Protocol SPR2-135.832.01...
• Service Instructions SPR2-130.061.02...
or
SPR2-135.840.01...
• Image quality quick test SPR2-135.820.01..
• Technical safety checks - protocol* SP00-000.834.01...

* Within the purview of DIN VDE 0751-1, we recommend documenting the results of the
maintenance both in the maintenance protocol and in the TSC protocol. The protocols
should be filled out completely and handed over to the client after maintenance is complete.

1.2.1 Systems equipped with SIMOMED N X 2080 monitor

• Maintenance check list Z35


• Adjustment instructions RX52-060.061.02...

1.2.2 Systems equipped with the standard 100/120Hz monitor

• Maintenance Instructions SPR2-230.101.01...


• Maintenance Protocol SPR2-230.105.01...

1.2.3 Systems equipped with TFT displays

• Replacement instructions for the TFT display on hand

1.2.4 Systems equipped with an external I.I. laser light localizer

• Installation and setting instructions RXR2-120.814.01

1.2.5 For systems equipped with an integrated I.I. laser light localizer

• Installation and setting instructions SP00-000.814.07...

SIREMOBIL Compact / L SPR2-135.831.01.02.02 Page 6 of 42 Siemens


05.09 CS PS SP
General Information 7

1.2.6 When an integrated laser light localizer is present (close to tube)

• Setting instructions (when laser light localizer is present, RXR2-130.815.01...


delivered until March 2006)
• Installation and setting instructions (when laser light local- SP00-000.814.08...
izer is present, delivered starting in April 2006)

1.2.7 When DICOM bridge is installed

• DICOM Bridge Installation and Setting Instructions SPR2-130.033.04...

1.2.8 When Codonics printer is present

• Installation and Setting Instructions Codonics Printer EP SPR2-130.814.01...


1660

Siemens SPR2-135.831.01.02.02 Page 7 of 42 SIREMOBIL Compact / L


05.09 CS PS SP
8 General Information

1.3 Required tools, measurement and auxiliary devices


NOTE The indicated articles are listed in the STC (Service Tools
Catalog) unless otherwise stated (the STC is a component of
the Spare Parts Catalog), except for those items identified
with "*".

Tools For example Material No.:


• Standard tool kit*
• Set of Allen wrenches*
• Spring balance rated for up to 200 N 44 15 113 RH090
• Leakage current measuring device Unimet 1100 safety tester 51 38 727
• Ground wire test meter Unimet 1100 safety tester 51 38 727
• Internal line resistance meter 84 28 104 Y4337
• DVM Fluke 187 99 94 831
• Luminance meter SPOTMETER for SMfit ACT 77 52 848
• Densitometer (if MULTISPOT 2000 97 02 416 Y1996
option is installed)
• Dose meter PTW Diados 97 17 612 Y0388
• Dynamic test kit 37 90 156 X1963
• Precision X-ray filter 99 00 598 XE999
• Set of resolution tests 28 71 820 RE999
• Set of radiation filters (10 x 0.3 mm) 44 06 120 RV090
• Set of radiation filters 97 98 596 G5321
• 25 mm AL measuring stand, type 26765 in accordance with DIN
6868 Part 50*
or
1.2 mm Cu (e.g., from set of radiation filters 44 06 120) plus

11 67 662
• 17 Micrometer Cu strip
• Centering cross (only with Diamen- 96 60 051 RE999
tor)
• Service PC
• Cable for connecting service PC to 99 00 440 RE999
host
• Service software* (see logbook)

SIREMOBIL Compact / L SPR2-135.831.01.02.02 Page 8 of 42 Siemens


05.09 CS PS SP
General Information 9

Tools For example Material No.:


• Sealing compound 20 49 716
• Special oil (Viscogen kl 300) 72 79 107

Siemens SPR2-135.831.01.02.02 Page 9 of 42 SIREMOBIL Compact / L


05.09 CS PS SP
10 General Information

1.4 Spare parts which may be needed

• Cable deflector for large steering castor, for SIREMO-


BIL stand
• Cable deflector for small wheels, for SIREMOBIL stand
• Conductive rubber for SIREMOBIL stand and monitor
trolley
• Touch-up paint 34 44 403
• Spray paint 84 27 734 RH999

SIREMOBIL Compact / L SPR2-135.831.01.02.02 Page 10 of 42 Siemens


05.09 CS PS SP
General Information 11

1.5 Text emphasis


Tab. 1 GEFAHR / DANGER

Bei einer unmittelbar drohenden Gefahr, die bei Nichtvermeidung


zum Tod oder zu einer schweren Körperverletzung führt.

Indicates when there is an immediate danger that leads to death


or serious physical injury.

Tab. 2 WARNUNG / WARNING

Bei einer Gefahr, die bei Nichtvermeidung zum Tod oder zu einer
schweren Körperverletzung führen kann.

Indicates a risk of danger that may lead to death or serious phys-


ical injury.

Tab. 3 VORSICHT / CAUTION

Bei einer Gefahr, die bei Nichtvermeidung zu einer leichten oder


mittleren Körperverletzung und/ oder zu einer Sachbeschädigung
führt oder führen kann.
Indicates a risk of danger that leads to slight or moderate physical
injury and/or damage to property.

Tab. 4 ACHTUNG / NOTICE

Bei einer Gefahr, die bei Nichtvermeidung zu einem unerwünscht-


en Ergebnis oder Zustand führt oder führen kann (nicht Tod,
Körperverletzung oder Sachbeschädigung).
Indicates a risk of danger that if disregarded leads or may lead to
a potential situation which may result in an undesirable result or
state (not death, physical injury or property damage).

Tab. 5 HINWEIS / NOTE

Ist als Tipp zu verstehen. Der Anwender muss diese Anweisung


nicht unbedingt beachten. Er erfährt jedoch Vorteile, wenn er dies
tut.
Should be understood as a tip. The user does not absolutely have
to observe these instructions. However, there will be advantages
if he does.

Siemens SPR2-135.831.01.02.02 Page 11 of 42 SIREMOBIL Compact / L


05.09 CS PS SP
12 General Information

1.6 Safety instructions

1.6.1 General safety information (in existing documents)

WARNING Danger of injury, death, or material damage.


Note
¹ the product-specific safety notes in these instructions,
¹ the general safety information in TD00-000.860.01...
and
¹ the safety information in accordance with ARTD part 2.
¹ Non-compliance can lead to death, injury, or material
damage.

1.6.2 General electrical safety information

WARNING Electrical safety!


Non-compliance can lead to severe injury or even death, as
well as material damage.
¹ After the cover panels are opened, parts under voltage
are accessible. To avoid danger, disconnect the sys-
tem from the power supply before opening the covers.
Pull out the power supply plug.
¹ If an uninterruptible power supply (UPS) is installed in
the system, the voltage output of the UPS must also be
deenergized or the voltage output plug must be discon-
nected.
¹ If work has to be performed under electrical voltage,
the general safety information according to
TD00-000.860.01... must be complied with.

CAUTION Electrical voltage!


Non-compliance can result in material damage.
¹ When working on the system, ESD regulations must be
observed.

SIREMOBIL Compact / L SPR2-135.831.01.02.02 Page 12 of 42 Siemens


05.09 CS PS SP
General Information 13

1.6.3 Radiation safety information

WARNING X-ray radiation!


Non-compliance can lead to illness, irreversible damage to
body cells and the genotype, severe injury and even death.
When performing work on the system during which radiation
must be released, the radiation protection directives and the
rules for radiation protection according to ARTD 002.731.02
must be complied with.
¹ Please note:
¹ Use available radiation protection devices.
¹ Wear radiation protection clothing (lead apron).
¹ Stay as far away as possible from the radiation source.
¹ Release radiation only if necessary.
¹ Set the radiation activity as low as possible. (low kV
and mA values, short radiation time)
¹ Release radiation for as short a time as possible.
¹ Checks in which radiation must be released are identi-
fied by the radiation warning symbol.

1.6.4 Mechanical safety information

CAUTION Risk of burns from hot parts or components! Failure to


observe may result in minor to more severe burns, especially
on the hands.
Parts and components (e.g., power components, cooling
fins, electromagnetic brakes) that can exceed 50 degrees
Celsius during operation are accessible after the covers are
opened.
¹ To avoid burns, switch the system off prior to touching
parts or components and allow at least 5 minutes for
cooling.

Siemens SPR2-135.831.01.02.02 Page 13 of 42 SIREMOBIL Compact / L


05.09 CS PS SP
14 General Information

CAUTION Risk of injury from mechanical parts! Failure to observe may


result in minor to more severe injury, especially to the hands.
Once the cover panels are opened, parts such as flat plugs,
threaded bolts, cut-off cable ties and component edges are
exposed. If you are not careful, these can cause crushing,
scrapes, and cuts to the skin, particularly to the hands.
¹ Perform the required work with special care and atten-
tion to detail.
¹ If necessary, wear work gloves.

1.6.5 Safety information - risk of infection

WARNING Risk of infection due to pathogens! Non-compliance can lead


to severe injury and even death.
This product can be contaminated by infected blood or other
bodily fluids.
¹ Avoid all contact with blood or other bodily fluids!
¹ Strictly observe the safety information in
ARTD-002.731.37... regarding prevention of infectious
diseases during customer service calls.

1.6.6 Integrated I.I. laser light localizer option

CAUTION Laser emissions!


This product contains lasers of the class 1M.
Disregarding safety precautions can lead to bodily injury,
especially to the retina of the eye, resulting in irreversible
damage to vision.
¹ Observe the safety information in ARTD-002.731.03...
When working with the laser light localizer, do not view
the laser beams directly with optical instruments.

SIREMOBIL Compact / L SPR2-135.831.01.02.02 Page 14 of 42 Siemens


05.09 CS PS SP
General Information 15

1.6.7 I.I. laser light localizer option or single tank laser light localizer option

CAUTION Laser emissions!


This product contains class 2 lasers. (USA: class 2 laser).
Disregarding safety precautions can lead to bodily injury,
especially to the retina of the eye, resulting in irreversible
damage to vision.
¹ Observe the safety information in ARTD-002.731.03...
When working with the laser light localizer, do not look
directly into the laser beam.

NOTE Laser emissions!


There is no direct hazard to the eye (blinking reflex). Never-
theless, do not look directly into the laser beam.

Siemens SPR2-135.831.01.02.02 Page 15 of 42 SIREMOBIL Compact / L


05.09 CS PS SP
16 General Information

1.7 Information on the protective conductor resistance test


Observe the instructions in the safety rules for installation and repair (ARTD-002.731.17
...).
The protective conductor resistance must be measured, documented, and evaluated dur-
ing maintenance.

NOTE For evaluation purposes, the first measured values and the
values documented during previous maintenance or safety
checks must be compared to the measured values.
A sudden or unexpected increase in the measured values
may indicate a defect in the protective conductor connec-
tions - even if the limit value of 0.2 ohms is not exceeded.
(Protective conductor or contacts).

The measurement must be performed in accordance with DIN VDE 0751, Part 1 (refer to
ARTD Part 2). The protective conductor resistance for all touchable conductive parts must
be measured during the normal operating state of the system.
Make sure that control cables or data cables between the components of the system are
not mistaken for protective conductor connections.
During the measurement the power cable and additional connection cables which also
make the ground wire connection between parts of the system (e.g., monitor cable between
basic unit and monitor trolley) must be moved section by section to detect broken conduc-
tors.
The protective conductor resistance must not exceed 0.2 ohms.

Fig. 1: Measuring circuit for measuring the protective conductor resistance in units/systems that
are permanently connected to the power supply net (according to DIN VDE

SIREMOBIL Compact / L SPR2-135.831.01.02.02 Page 16 of 42 Siemens


05.09 CS PS SP
General Information 17
0751-1:2001-10, Fig. C3).
Pos. 1 System
Pos. 2 Application part (not available)
Pos. 3 Measurement setup (integrated into measuring device)

The determined values, including the measuring points, must be recorded and assessed
in the protective conductor resistance report.
The measuring procedure and the measuring device used (designation and serial number)
must also be documented.

1.7.1 System leakage current measurement information

NOTE During maintenance, the system leakage current measure-


ment must be performed and recorded as a repeat measure-
ment.
However, the first measured value must be newly determined
and a new protocol be must created under the following con-
ditions:
Lack of documentation for system leakage current measure-
ment
Local line voltage or line frequency deviating from the line
voltage and line frequency documented in the report (e.g. in
the event of a site change or operator change)
When a different procedure for measuring the system leak-
age current than the one documented in the protocol is used.
For the purpose of traceability, reference to the new report is
to be written on the old report. The reason for newly deter-
mining the first measured value must be documented and
confirmed with a name and signature.

Observe the instructions in the safety rules for installation and repair (ARTD-002.731.17
...).

Siemens SPR2-135.831.01.02.02 Page 17 of 42 SIREMOBIL Compact / L


05.09 CS PS SP
18 General Information

WARNING Electrical voltage!


Non-compliance can lead to severe injury and even death.
¹ The system leakage current measurement may be per-
formed on systems of protection class I only after the
protective conductor test has been passed.

First measured value


The first measured value was already determined and documented in the system leakage
current report. The measuring procedure was also recorded.
The measurement was performed with the recorded line voltage, line frequency and with
the recorded measuring equipment.
Measurement
Perform the measurement in accordance with DIN VDE 0751, Part 1 (see
ARTD-002.731.17...), and record the value determined.
The measuring procedure indicated in the protocol must be used.
If the first measured value has to be newly determined (see note), a measuring procedure
can be selected (direct measurement or differential measurement).
Measurement of the system leakage current according to the differential current method
(measurement setup according to (Fig. 2 / p. 18)) must be given preference, since this
method is not dangerous to the person performing the measurement and other persons.
However, please note the minimum resolution of the system leakage current measuring
instrument and any additional manufacturer's data restricting the use of the measuring
device.

Fig. 2: Measuring circuit for measuring the system leakage current according to the differential
current method in compliance with DIN VDE 0751-1:2001-10, Fig. C6 for protection class

SIREMOBIL Compact / L SPR2-135.831.01.02.02 Page 18 of 42 Siemens


05.09 CS PS SP
General Information 19
I.
Pos. 1 System
Pos. 2 Application part (not available)
Pos. 3 Measurement setup (integrated into measuring device)

If the direct measurement of the system leakage current is used (Fig. 3 / p. 19)), the system
must be insulated during the measurement and must not be touched.

Fig. 3: Measuring circuit for direct measurement of the system leakage current in compliance with
DIN VDE 0751-1:2001-10, Fig. C5 for protection class I.

Pos. 1 System
Pos. 2 Application part (not available)
Pos. 3 Measurement setup (integrated into measuring device)

Siemens SPR2-135.831.01.02.02 Page 19 of 42 SIREMOBIL Compact / L


05.09 CS PS SP
20 General Information

WARNING Electrical voltage!


Non-compliance can lead to severe injury and even death.
No housing parts of the system may be touched during direct
measurement of the system leakage current (Fig. 3 / p. 19).
¹ Third-person access to the system must be prevented.

The system must be switched on during measurement. Measuring devices with automated
measuring sequences must therefore be set to manual measurement.
The highest value must be entered in the system leakage current report.
This value must not exceed the permissible leakage current values according to DIN VDE
0751-1:2001-10, Table F.1, line "System leakage current for devices according to remarks
1 and 3", of 2.5 mA.
Measure and record the current line voltage. If the measured line voltage deviates from the
nominal voltage, correct the measured value to the value corresponding to a measurement
at the nominal value of the line voltage. This must also be documented.
Document the measuring procedure (differential measurement or direct measurement)
and the measuring device used (designation and serial number).
In the case of repeat measurements, the measured value must also be evaluated.

NOTE For evaluation purposes, the first measured values and the
values documented during previous maintenance or safety
checks must be compared to the measured values.
A sudden or unexpected increase of the measured values
may indicate that a fault has occurred in the primary circuit
of the power supply (damaged insulation, damage caused by
water ingress or humidity, defective interference suppressor,
etc.) - even if the limit value of 2.5 mA is not exceeded.

The evaluation is not necessary in the case of a new determination.


File the protocol sheet in the system binder or log book.

SIREMOBIL Compact / L SPR2-135.831.01.02.02 Page 20 of 42 Siemens


05.09 CS PS SP
General Information 21

1.8 Explanation of abbreviations in the maintenance protocol

Abbrevia- Description
tion
SI Safety Inspection
SIE Electrical safety
SIM Mechanical safety
PM Preventive Maintenance
PMP Periodic preventive maintenance
PMA Preventive maintenance adjustments
PMF Preventive check of operating values and functions
Q System quality, image quality
QIQ Image quality
QSQ System quality
SW Software Maintenance
CSE Customer service engineer
KSK No. Customer-specific code
IVK Installed volume component
WE Maintenance unit

The work steps identified by these abbreviations are listed as check-off points in the main-
tenance protocol and should be checked off upon completion.

Siemens SPR2-135.831.01.02.02 Page 21 of 42 SIREMOBIL Compact / L


05.09 CS PS SP
22 General Information

1.9 Maintenance interval


• 12 months

SIREMOBIL Compact / L SPR2-135.831.01.02.02 Page 22 of 42 Siemens


05.09 CS PS SP
Inspection of exterior and surroundings 23
2-
2 Inspection of exterior and surroundings

2.1 Visual inspection of exterior


PMP Damage to the system
• Inspect the entire system for damage, such as damage to the housing or paint.
PMP Damage to the network connection
• Check all installed network cables, network plugs, socket panel, and sockets for dam-
age.

Siemens SPR2-135.831.01.02.02 Page 23 of 42 SIREMOBIL Compact / L


05.09 CS PS SP
24 Inspection of exterior and surroundings

2.2 Inspection of surroundings

2.2.1 Power outlets


SIE Damage
• Inspect for damage all power outlets used for system operation.
SIE Line voltage
• Measure the line voltage and compare it to the line voltage label on the monitor trolley.
SIE Internal line resistance
• Measure the internal line resistance.

SIREMOBIL Compact / L SPR2-135.831.01.02.02 Page 24 of 42 Siemens


05.09 CS PS SP
Safety inspection 25
3-
3 Safety inspection

3.1 Mechanical safety


SIM Cover panels
• Remove the covers from the SIREMOBIL stand and monitor trolley.
• Check cover panels for mechanical damage.
SIM Cable deflectors
• Inspect the cable deflectors on the stand and on the monitor trolley, and replace them if
necessary.
SIM Cassette holder (if present)
• Check cassette holder for mechanical damage.
• Install the cassette holder on the I.I. Ensure that it locks into place and seats properly.
SIM Foot brake
• Test the function of the foot brakes on the SIREMOBIL stand and monitor trolley on a
level floor.
SIM Brakes
• Using the spring balance, check to make sure the orbital brake, angulation brake,
swivel brake and horizontal lift brake achieve the specified braking value in both braked
and unbraked state.
SIM C-arm
• Perform all C-arm movements, checking for play in the bearings and bearing noises.
SIM Wheels
• Move the SIREMOBIL in a straight line on a level floor.
• Evaluate the straight and quiet movement of the SIREMOBIL.
• Replace any defective wheels or castors.
• Check the steering function.
SIM Lifting column
• Switch the system on.
• Electrically raise and lower the lifting column over its entire range,
- while listening for noises and checking for play in the bearings.
- The lifting column movement must switch off automatically when the end positions
are reached.
- On the SIREMOBIL Compact L, an acoustical warning sounds during downward
motion and the lifting column does not move. The acoustical warning signal contin-
ues to sound as the downward motion moves to the end position.
• Switch off the system.
• Lubricate the lifting column.
- To do so, move the lifting column approx. 20 cm upward. Remove the plastic cap.
- Oil the spindle (approx. 2 cm3 special oil).

Siemens SPR2-135.831.01.02.02 Page 25 of 42 SIREMOBIL Compact / L


05.09 CS PS SP
26 Safety inspection

SIM Emergency stop switch (Compact / L)


• Press the emergency stop switch. The lifting column must not be movable when it is
being raised or lowered.
• Now unlock the emergency stop switch. The lifting column can be moved again.

3.1.1 External I.I. laser light localizer (if present)


SIM Mechanics
• Inspect the I.I. laser light localizer for mechanical damage.
• Install the I.I. laser light localizer on the I.I. and ensure that it locks into place and seats
properly. When doing this, pay special attention to the tension band and its closure.
• Check the I.I. ring for damage.
SIM Function
• Perform the necessary maintenance of the I.I. laser light localizer according to Start-Up
Instructions RXR2-120.814.01...

3.1.2 Integrated I.I. laser light localizer (if present)


SIM Mechanics
• Inspect the integrated I.I. laser light localizer for mechanical damage.
• Check the I.I. ring for damage and to make sure it is screwed in all the way.
SIM Function
• Check the function of the laser diodes.
• Check the adjustability of the laser light fan beams on the focal point of the tube hous-
ing.

3.1.3 Integrated laser light localizer, close to tube - (if present)


SIM Mechanics
• For systems equipped with I.I. cap (delivered until March 2006): Inspect the I.I. cap of
the laser light localizer for mechanical damage.
• For systems equipped with I.I. cap (delivered until March 2006): Install the I.I. cap of the
laser light localizer on the I.I. and ensure that it locks into place and seats properly.
• For systems equipped with I.I. cap (delivered until March 2006): Check the I.I. ring for
damage and to make sure it is screwed in all the way.
SIM Function
• When laser light localizer is present, delivered until March 2006: Perform maintenance
on the laser light localizer according to Setting Instructions RXR2-130.815.01...
• When laser light localizer is present, delivered in or after April 2006: Perform mainte-
nance on the laser light localizer according to Setting Instructions SP00-000.814.08...

SIREMOBIL Compact / L SPR2-135.831.01.02.02 Page 26 of 42 Siemens


05.09 CS PS SP
Safety inspection 27

3.1.4 DICOM bridge (if present)


SIM Attachment
• Check to make sure the DICOM bridge is mounted properly in the monitor trolley.

3.1.5 Codonics printer (if present)


SIM Mechanics
• Ensure that the sliding rails are firmly secured (with screws) to the mounting compo-
nents of the framegrabber (PC).
• Check to make sure the printer is firmly secured (with screws) to the sliding rails. All fas-
tening screws must be present.
• Check to make sure the framegrabber (PC) is firmly secured (with screws) to the
mounting bracket. All fastening screws must be present.
SIM Function
• Check the guide rails for correct functioning and smooth operation.
• The guide rails must not be damaged.
• When pushed in, the guide rails must lock correctly in the end position and be mechani-
cally blocked.
• If necessary, use a cleaning brush to remove dirt and dust from the rails.

3.1.6 Monitor(s) / displays


SIM Monitor(s) / displays
• Check the monitor(s) or displays for damage.
• Ensure that the monitor(s) or displays are correctly mounted on the monitor trolley.

3.1.7 Labels
SIM Warning labels
• Ensure that all required warning labels are attached and in good condition.
- Replace any illegible labels or markings.
SIM Identification labels
• Ensure that all required ID labels are attached and in good condition.
- Replace any illegible labels or markings.

Siemens SPR2-135.831.01.02.02 Page 27 of 42 SIREMOBIL Compact / L


05.09 CS PS SP
28 Safety inspection

3.2 Electrical safety


SIE Cables and plugs
• Check visible system cables and plugs for damage.
SIE Fluoroscopic timer
• Check: See section on "Compulsory radiation switch-off".
SIE Audible warning signal
• Check: See section on "Compulsory radiation switch-off".
SIE Radiation displays
• Check: See section on "Compulsory radiation switch-off".
SIE Radiation release switch
• Check: See section on "Compulsory radiation switch-off".
SIE Compulsory radiation switch-off
Test the functioning of the audible warning signal and compulsory radiation switch-off (if
required) according to the country-specific regulations. Perform the following checks as
well:
X • Release fluoroscopy for one minute. Use a wristwatch to check proper functioning of
the fluoroscopic timer.
• All radiation indicators on the system operating console as well as the monitor trolley
must illuminate during radiation release.
X • Check the radiation release switches (hand and foot switches) for proper functioning.
• Inspect the radiation release switches for damage.
• Check the cables of the hand and foot switches for mechanical damage.
• Move the cables of the hand and foot switches to check for breaks or short circuits.

NOTE • Depending on the programming, the audible warning sig-


nal has to sound 4.5, 5, or 9.5 minutes after the fluoros-
copy has started. It sounds again every 5 or 10 minutes.
Pressing the -0- button once turns it off.
• Pressing this button again resets the fluoroscopic timer
to "0".
• Depending on the country-specific programming, either
the required radiation shut-off does not occur or it occurs
after 5 or 10 minutes and then every 5 or 10 minutes after
that.

SIE Dose rate


• Switch the system on.
X • Check the dose rate (see the SIREMOBIL Compact / L service instructions for this pur-
pose).

SIREMOBIL Compact / L SPR2-135.831.01.02.02 Page 28 of 42 Siemens


05.09 CS PS SP
Safety inspection 29

SIE Iris collimator


• Check the iris collimator and correct it if necessary.
X - Select the full I.I. format and release fluoroscopy briefly.
- The collimator blades must be clearly visible at the edges of the image.
- Select a zoom format and release fluoroscopy briefly.
- The collimator blades must be clearly visible at the edges of the image.
SIE Dose area product meter (if present)
X • Check the area dose product measurement device and its accuracy (see the SIREMO-
BIL Compact/L service instructions).

SIE Air kerma meter (if present)


X • Check the air kerma measurement device and its accuracy (see the SIREMOBIL Com-
pact/L service instructions).

SIE Codonics printer - cable breaks (if present)


• Check all connection cables between the printer and framegrabber (PC) for damage or
cable breaks.
• Check the cable run of the energy chain beneath the printer for damage and correct
functioning. Ensure that no cable is crushed.

3.2.1 Monitor(s) / displays


SIE Monitor(s) / displays
• Check the SIMOMED N monitor as prescribed in the checklist.
• Check the luminance of the 100/120 Hz standard monitor (Maintenance instructions
SPR2-230.101.01...).
• Check the luminance of the TFT displays. Check the ambient light sensor (if present).

Siemens SPR2-135.831.01.02.02 Page 29 of 42 SIREMOBIL Compact / L


05.09 CS PS SP
30 Safety inspection

3.3 Operation
SIE Check the operating functions.
• Check whether the operating icons and button labels are clearly legible and recogniz-
able.
• Check whether the light indicators of the operating elements and displays are properly
functioning.
• Check all system operating functions.

SIREMOBIL Compact / L SPR2-135.831.01.02.02 Page 30 of 42 Siemens


05.09 CS PS SP
Safety inspection 31

3.4 Conductive rubber


SIE SIREMOBIL stand voltage discharge rubber
• Inspect the voltage discharge rubber on the SIREMOBIL stand for damage or dirt. If
necessary, clean or replace it.
SIE Monitor trolley voltage discharge rubber
• Inspect the voltage discharge rubber on the monitor trolley for damage or dirt. If neces-
sary, clean or replace it.

Siemens SPR2-135.831.01.02.02 Page 31 of 42 SIREMOBIL Compact / L


05.09 CS PS SP
32 Safety inspection

3.5 Measurements
SIE Protective conductor test (performed under "Final work steps")
• Measure the protective conductor with the protective conductor meter.
• The protective conductor resistance must not exceed 0.2 Ω.

NOTE Observe the guidelines given here: ((Information on the pro-


tective conductor resistance test / p. 16)).

NOTE The protective conductor test is part of the safety inspection.


However, it cannot be performed until all maintenance work
has been completed and must be performed with all covers
closed. See (Final work steps / p. 40).

SIE Leakage current (performed under "Final work steps")


• Measure the leakage current with all system cover panels closed. Document and evalu-
ate the determined values. Document the measuring procedure. The leakage current
must not exceed 2.5 mA.

NOTE Observe the guidelines given here: ((System leakage current


measurement information / p. 17)).

NOTE The leakage current test is part of the safety inspection. How-
ever, it is recommended to perform it only after all mainte-
nance work has been completed, and with all system cover
panels closed. See (Final work steps / p. 40).

SIREMOBIL Compact / L SPR2-135.831.01.02.02 Page 32 of 42 Siemens


05.09 CS PS SP
Maintenance, operating values and funct. inspect. 33
4-
4 Maintenance, operating values and funct. inspect.

4.1 Maintenance
PMP Cleaning the system
• Clean the entire system:
- Visible cables
- Exterior surfaces
- Contact surface of wheels
- Interior space
PMP System ventilation
• Clean the ventilation slots of the SIREMOBIL stand and monitor trolley.
PMP Memoskop ventilation
• Clean the ventilation slots on the Memoskop.
• Check to make sure the fan is functioning properly.

4.1.1 DICOM bridge (if present)


PMP Ventilation
• Clean the ventilation slots on the DICOM bridge.
• Check to make sure the fan is functioning properly.

4.1.2 Codonics printer (if present)


PMP Ventilation
• Clean the ventilation slots on the printer and Framegrabber.
• Check to make sure the fan is functioning properly.
PMP Cleaning

NOTE According to the operating instructions for Codonics printer


EP1660, the customer is responsible for cleaning the thermal
printhead and the film feed mechanism. The cleaning interval
is also specified in the operating instructions for the printer.
See the maintenance chapter of the Codonics operating
instructions. Annual cleaning of the printer as part of the
maintenance work described here does not release the cus-
tomer from the obligation to conduct cleaning work in accor-
dance with the specifications of the printer manufacturer. If
errors or image quality problems are identified during main-
tenance, refer the customer to the maintenance work and
maintenance intervals specified in the Codonics printer oper-
ation instructions.

Siemens SPR2-135.831.01.02.02 Page 33 of 42 SIREMOBIL Compact / L


05.09 CS PS SP
34 Maintenance, operating values and funct. inspect.

WARNING Danger of burns from the hot thermal printhead!


Non-compliance can result in burns, particularly on the
hands.
¹ Allow the thermal printhead to cool prior to cleaning.

• Clean the printer as described in the maintenance chapter of the Codonics operating
instructions. Use the customer's cleaning cloths (several cleaning cloths are included in
the delivery scope of the printer).

SIREMOBIL Compact / L SPR2-135.831.01.02.02 Page 34 of 42 Siemens


05.09 CS PS SP
Maintenance, operating values and funct. inspect. 35

4.2 Operating values, inspection


PMF Error memory
• Read out the error memory of the system and analyze the results.

Siemens SPR2-135.831.01.02.02 Page 35 of 42 SIREMOBIL Compact / L


05.09 CS PS SP
36 Maintenance, operating values and funct. inspect.

4.3 Functional inspection


PMF Video printer (if present)
• Check the function of the video recorder.
• Check the remote control (Print button).
PMF Video recorder (if present)
• Check the function of the video recorder.
• Check the remote control (REC-Pause-Start / REC-Pause-Stop).
PMF Sneaker Net laser camera system (if present)
• Check the function of the “Sneaker Net” laser-camera system.
PMF DICOM bridge (if present)
• Check the function of the DICOM bridge.
PMF DICOM Connect (if present)
• Check the function of the DICOM Connect network connection.
PMF Codonics printer, manual control panel (if present)
• Check the function of the printer.
• Check the function of the manual control panel.
• Inspect the cable of the manual control panel for damage and breaks.
PMF 4 View (if present)
• Check the plug-in connections for damage.
• Check the function of the 4 View monitor extensions.
• Check the function of the additional monitors and the additional keyboard (if present).
PMF Laser camera connection (if present)
• Check the functioning of the laser camera connection.
• Check the remote control (Print button).
PMF Image rotation
Check the image rotation (camera rotation).
X • Check the 0 setting and the entire rotational range (± 220) under fluoroscopy.
• Check the preview display of the image rotation.
PMF Slot diaphragm
• Select I.I. full format.
X • Open the slot diaphragm completely and briefly release fluoroscopy.
The slot diaphragm blades are no longer visible on the screen.
• Select I.I. zoom format.
X • Release fluoroscopy briefly.
The slot diaphragm blades are no longer visible on the edge of the image.
X • Close the slot diaphragm.
Rotate the slot diaphragm ± 360 under fluoroscopy. The slot diaphragm can be rotated
360.

SIREMOBIL Compact / L SPR2-135.831.01.02.02 Page 36 of 42 Siemens


05.09 CS PS SP
Maintenance, operating values and funct. inspect. 37

PMF Monitor display of the iris collimator aperture


• Select I.I. full format.
• Close the iris collimator (X-iris) completely.
• Mark the diameter of the collimated iris collimator aperture on the monitor.
X • Release fluoroscopy briefly.
The actual diameter of the iris collimator is visible.
The previously indicated monitor display of position and diameter must coincide with
the actual iris collimator aperture.
• Open the iris collimator completely.
• Mark the diameter of the collimated iris collimator opening on the monitor.
X • Release fluoroscopy briefly.
The actual diameter of the iris collimator is visible.
The previously indicated monitor display of position and diameter must coincide with
the actual iris collimator aperture.
• Select I.I. zoom format.
• Close the iris collimator (X-iris) completely.
• Mark the diameter of the collimated iris collimator aperture on the monitor.
• Release fluoroscopy briefly.
The actual diameter of the iris collimator is visible. The previously indicated monitor dis-
play must match the position and diameter of the actual iris collimator aperture.
• Open the iris collimator completely.
• Mark the diameter of the collimated iris collimator opening on the monitor.
X • Release fluoroscopy briefly.
The actual diameter of the iris collimator is visible.
The previously indicated monitor display must match the position and diameter of the
actual iris collimator aperture.
PMF Monitor display of the slot diaphragm positions
• Select I.I. full format.
• Completely close the slot diaphragm and rotate the slot diaphragm from its basic posi-
tion.
• Mark the distance and angle of rotation of the displayed slot diaphragm position on the
monitor.
X • Release fluoroscopy briefly.
The actual position (distance of the blades and angle of rotation) of the slot diaphragm
is visible. The previously displayed monitor display must match the position and angle
of rotation of the actual slot diaphragm position.
• Open the slot diaphragm completely and rotate it again.
• Mark the distance and angle of rotation of the displayed slot diaphragm position on the
monitor.

Siemens SPR2-135.831.01.02.02 Page 37 of 42 SIREMOBIL Compact / L


05.09 CS PS SP
38 Maintenance, operating values and funct. inspect.

X • Release fluoroscopy briefly.


The actual position (distance of the blades and angle of rotation) of the slot diaphragm
is visible. The previously displayed monitor display must match the position and angle
of rotation of the actual slot diaphragm position.
• Select I.I. zoom format.
• Completely close the slot diaphragm and rotate the slot diaphragm from its basic posi-
tion.
• Mark the distance and angle of rotation of the displayed slot diaphragm position on the
monitor.
X • Release fluoroscopy briefly.
The actual position (distance of the blades and angle of rotation) of the slot diaphragm
is visible.
• The previously displayed monitor display must match the position and angle of rotation
of the actual slot diaphragm position.
• Open the slot diaphragm completely and rotate it again.
• Mark the distance and angle of rotation of the displayed slot diaphragm position on the
monitor.
X • Release fluoroscopy briefly.
The actual position (distance of the blades and angle of rotation) of the slot diaphragm
is visible.
The previously displayed monitor display must match the position and angle of rotation
of the actual slot diaphragm position.

NOTE Perform this check with the lowest possible kV values so that
the front edges of the slot diaphragm blades are effectively
visualized.

SIREMOBIL Compact / L SPR2-135.831.01.02.02 Page 38 of 42 Siemens


05.09 CS PS SP
Final result / quality inspect. & general maint. 39
5-
5 Final result / quality inspect. & general maint.
SIE Image quality (IQ) quick test
• Perform the IQ quick test according to the SIREMOBIL Compact / L image quality quick
test procedure.
X • Include additional monitors and displays in the test.

SIE DICOM bridge IQ test (if present)


• Perform the IQ test according to chapter 6 of the Installation and Setting Instructions
SPR2-130.033.04... in the course of the system IQ quick test.
SIE Codonics printer IQ test (if present)
• Inspect the image quality according to the chapter “Testing the image quality” in the
installation instructions SPR2-130.814.01...
PMP Upkeep
• Entire system: Touch up any paint damage as needed.

Siemens SPR2-135.831.01.02.02 Page 39 of 42 SIREMOBIL Compact / L


05.09 CS PS SP
40 Final result / quality inspect. & general maint.

5.1 Final work steps


SIE Protective conductor test
SIE Protective conductor resistance
• Perform the protective conductor test with all system cover panels closed. Document
and evaluate the determined values, and indicate the measuring points used. The
protective conductor resistance must not exceed 0.2 Ohms.

NOTE Observe the guidelines given here: ((Information on the pro-


tective conductor resistance test / p. 16)).

NOTE Evaluate the results by comparing the first measured value


and the values documented during preceding maintenance
procedures or safety checks to the measured values.
A sudden or unexpected increase in the measured values
may indicate a defect in the protective conductor connec-
tions (ground wire or contacts) - even if the limit value of 0.2
ohms is not exceeded.

SIE Leakage current


• Measure the leakage current. Document and evaluate the determined values. The
limit value of 2.5 mA must not be exceeded.

NOTE Observe the guidelines given here: ((System leakage current


measurement information / p. 17)).

NOTE Evaluate the results by comparing the first measured value


and the values documented during preceding maintenance
procedures or safety checks to the measured values.
A sudden or unexpected increase of the measured values
may indicate that a fault has occurred in the primary circuit
of the power supply (damaged insulation, damage caused by
water ingress or humidity, defective interference suppressor,
etc.) - even if the limit value of 2.5 mA is not exceeded.

SIREMOBIL Compact / L SPR2-135.831.01.02.02 Page 40 of 42 Siemens


05.09 CS PS SP
Changes to Previous Version 41
6-
6 Changes to Previous Version
Chapter 1, Section 10 (Safety checks) deleted and "Forms4Mobile" template incorporated.

Siemens SPR2-135.831.01.02.02 Page 41 of 42 SIREMOBIL Compact / L


05.09 CS PS SP
42 Changes to Previous Version

SIREMOBIL Compact / L SPR2-135.831.01.02.02 Page 42 of 42 Siemens


05.09 CS PS SP

You might also like