AN 42218 EA CHNSO Coals AN42218 EN 5

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 1

Table 5. CHN data of coal samples in duplicate using CHN reactor.

Series Elem. Coal Samples

A B C D E F
% Diff. % Diff. % Diff. % Diff. % Diff. % Diff.
88.36 82.58 80.04 78.34 65.82 79.93
1 C 0.08 0.04 0.38 0.27 0.02 0.06
88.44 82.54 79.66 78.61 65.80 79.87
3.72 4.24 4.45 4.36 4.60 4.30
H 0.02 0.04 0.07 0.05 0.13 0.02
3.74 4.20 4.38 4.31 4.47 4.28
1.88 1.17 1.01 1.71 0.80 1.15
N 0.02 0.03 0.02 0.03 0.00 0.04
1.90 1.14 0.99 1.74 0.80 1.11
88.04 82.25 79.46 77.82 66.20 80.09
2 C 0.09 0.19 0.30 0.56 0.17 0.09
87.95 82.06 79.73 78.38 66.03 80.18
3.75 4.19 4.38 4.27 4.45 4.28
H 0.02 0.01 0.00 0.03 0.04 0.02
3.73 4.18 4.38 4.30 4.41 4.30
1.88 1.13 0.98 1.71 0.80 1.15
N 0.00 0.02 0.01 0.03 0.00 0.00
1.88 1.11 0.99 1.74 0.80 1.15
88.28 82.35 80.06 77.96 65.94 80.10
3 C 0.24 0.02 0.45 0.50 0.07 0.09
88.04 82.37 79.61 78.46 66.01 80.01
3.74 4.20 4.44 4.27 4.49 4.29
H 0.02 0.01 0.06 0.03 0.01 0.02
3.72 4.19 4.38 4.30 4.48 4.27
1.83 1.14 1.00 1.71 0.80 1.18
N 0.02 0.00 0.01 0.03 0.00 0.01
1.85 1.14 0.99 1.74 0.80 1.17

The accuracy of the FlashSmart EA for CHN Conclusions


determination was demonstrated by performing three For CHNS/O determination in coal samples the
series of analysis of the six coal samples, which were Thermo Scientific FlashSmart EA Analyzer provides
analyzed twice. The calibration of the system was accurate data, which meet the requirements of the
performed with 2–3 mg of acetanilide. Coal sample ASTM D 5373-02 Method.
A was weighed at 2–2.5 mg and coal samples B-F
2.5–3 mg. Table 5 shows the CHN results and the No matrix effect was observed when changing the
difference (Diff.) calculated between both data. All sample, meaning that samples were entirely combusted.
differences are acceptable and fall within the the
Repeatability Limit, which is indicated in the Official The Thermo Scientific FlashSmart EA Analyzer allows
method, meaning a good homogeneity and the total CHNS and oxygen determinations with no need for any
combustion of the samples. extra modules, reducing operational cost, while increasing
speed of analysis.

Find out more at www.thermofisher.com/OEA

©2017 Thermo Fisher Scientific Inc. All rights reserved. ASTM is a trademark of the American Society for Testing and Materials.
All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an example
of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that
might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are
available in all countries. Please consult your local sales representatives for details. AN42218-EN 0717S