420 Datasheet
420 Datasheet
This compact 3 U high 19” rack enclosure houses both the NOx converter as well as a self contained heated filter
and calibration gas selection valve. It is generally used with the Signal Group S4 PULSAR multi-gas analyser or to
check converter efficiencies.
When gas is passing through the converter the analyser The Model 4000 converter has been designed to
is said to be in NOx mode, when gas is by-passing the overcome this undesirable side reaction.
converter, the instrument is said to be in NO mode.
Following prolonged use of the Carbon converter it
It is important, when designing a converter, that will be necessary to replace the Carbon material as
operating conditions and materials are chosen to this is slowly eroded by Oxygen and Nitrogen dioxide.
prevent undesirable side reactions occurring which may Replacement of this Carbon will therefore be necessary
destroy the Nitric oxide in the sample gas. after about six months of continuous use. Replacement
Carbon monoxide is known to cause problems in is a simple job and a spare charge of Carbon is available
certain types of converters due to the in a Spares Kit.
SPECIFICATION
RANGE: CATALYST TEMPERATURE: HEATED FILTER TYPE:
0-1,000ppm Factory set for optimum performance Quick release via rear panel.
0 - 600ºC Hydrocarbon free. Fibreglass 0.1 micron.
EFFICIENCY: Nominal 400ºC Heated to 200ºC
>95%
NO2 CALIBRATION SOLENOID MAINS VOLTAGE:
CONVERTER MATERIAL: TEMPERATURE: 220V/240V/50Hz 110V/120V/60Hz
Vitrified Carbon 200ºC
POWER:
750 watts
Authorised Representative: